Image Cytometry System for 20 µl Cell-Based Assays Cellometer ® Spectrum Cell Cycle Apoptosis Cell Mediated Cytotoxicity Cell Counting and Viability Interchangeable Fluorescent Modules

Which Instrument is Right for Me? Advantages of Cellometer ... · Simple 20 µl Cell-Based Assays • Pre-qualified reagents • Small20µlsamplesize ... • Determine cell viability,

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Which Instrument is Right for Me? Advantages of Cellometer ... · Simple 20 µl Cell-Based Assays • Pre-qualified reagents • Small20µlsamplesize ... • Determine cell viability,

Image Cytometry System for 20 µl Cell-Based Assays

Cellometer

® Spectrum

Nexcelom Bioscience360 Merrimack Street, Building 9Lawrence, MA 01843, USA

Nexcelom products are for RESEARCH USE ONLY and are not approved for diagnostic or therapeutic use. © Copyright 2018 Nexcelom Bioscience LLC. All Rights Reserved.

Innovation and Expertise in the Science of Cell CountingSchedule a FREE demonstration or technical seminar with a Nexcelom Applications Specialist today.

Call 978-327-5340 or E-mail [email protected]

1001465 Rev B 7/18

For Europe Call 44 (0) 161 232 4593 orE-mail [email protected]

For China Call +86 21 5886 0038 orE-mail [email protected]

Cell CycleApoptosisCell Mediated CytotoxicityCell Counting and ViabilityInterchangeable Fluorescent Modules

Accurate Cell Counting

R2=0.99899

1.00E+04

1.00E+05

1.00E+06

1.00E+07

1.00E+06 1.00E+07 1.00E+08

Theo

retic

al C

once

ntra

tion

[log]

Measured Concentration [log]Counted Cells Trend Line

Broad Dynamic Range1.00E+08

1.00E+051.00E+04

• AutomaticidentificationofAcridineOrange(AO)stainedcells(shownabove)

• Image and count up to 2x107 cells/mL

• Image acquisition and analysis: 30 seconds/sample for a 3-channel assay

• Versatile platform - use for routine cell countingandcomplexcell-basedassays

Bright Field StainedAO+Cells CountedAO+Cells

Cell Imaging• Visually check cell morphology

• Ensure only cells of interest are counted

• Archiveandre-analyzecellimages

• Exportimagesforpublication

Proprietary Pattern-Recognition Software• Countindividualcellsinclusters

• Count irregular-shaped cells

• Countcellsbasedonsize

• Eliminatedebrisfromcellcounts

Non-Fluidic Platform• Disposablecountingchambers-nowashing

• Compatiblewithfragilecells

• Maintenance-free

• RobustopticsmodulesandLEDlightsources

IQ/OQ Validation and GMP/GLP Accessories• InstallationQualificationreagents/protocol

• OperationalQualificationreagents/protocol

• On-siteIQorOQPerformance

• GMP/GLPSoftwareModule

Which Instrument is Right for Me?

Features Bright Field Cell Counters Fluorescent Viability Cell Counters Image Cytometers

Mini Auto T4 Auto 1000

Auto 2000 X1 X2 K2 Spectrum Spectrum

(10x)Celigo

BFCeligo 4 Channel

Celigo 5 Channel

Cell / Sample TypeCell Line X X X X X X X X X

CulturedPrimaryCells X X X X X X X X XAlgae X

Platelets X XLow Concentration Cell Lines X X X X X X

Yeast (Clean Sample) X X XYeast (Messy Sample) X X

Primarycells(MessySample*) X X X X XPBMCs,Splenocytes,StemCells X X X X X

Hepatocytes X X X XAdipocytes*** X X X X X X

Cell-BasedAssay** X X X X X X X XApoptosis(AnnexinV-FITC/PI) X X X X XApoptosis(CaspaseActivity) X X X X X

Autophagy (CytoID-green) X XCellProliferation(CFSE) X X X X

CellCycle(PI) X X X X X X XGFPTransfection X X X X X X XRFPTransfection X X X X

MitochondrialPotential(JC-1) X X X XMulti-drug Resistance (ABC

Transporter) X X X X

Surface Marker Analysis X X X XVitality(Calcein-AM/PI) X X X X X X

Vitality (CFDA-AM) X XImage Cytometry** X X X X X

* A messy sample is a heterogeneous sample containing unwanted cell types, such as red blood cells, in addition to the cells of interest.** FCS Express license must be purchased in order to perform Cell Based Assay or Image Cytometry analysis*** Cellometer CHT4-PD300 slides are required for cells greater than 80µm in diameter

Advantages of Cellometer Image Cytometry

Page 2: Which Instrument is Right for Me? Advantages of Cellometer ... · Simple 20 µl Cell-Based Assays • Pre-qualified reagents • Small20µlsamplesize ... • Determine cell viability,

Cell Types for Many Research Areas

• Clinical Immunology: PBMCs

• Diabetes / Obesity: Adipocytes

• Immunotherapy: Leukocytes

• Microbiology: Yeast (Spectrum 10x)

• Oncology: Cell Lines

• Regenerative Medicine: Stem Cells

• Toxicology: Hepatocytes

• Transplantation: Nucleated Cells

• Vaccine Development: Splenocytes

All-in-One System Basiccellcounting,primarycellviability,andcell-basedassays.

Dual-Fluorescence for Accurate Primary Cell Viability No interference from red blood cells.Analyzebonemarrow,peripheralblood,andcordbloodwithoutlysing.

Unique Algorithms for Advanced Cell AnalysisDetermine concentration and viabilityof hepatocytes, adipocytes, and othersophisticated cell types.

Fast ResultsObtain cell images, counts, sizemeasurementsandviabilitycalculationsin < 30 seconds per sample.

Features of the SpectrumTM Image Cytometry System

Optimizedforprimarycellanalysis

PBMCsStem CellsAdipocytesNeural CellsHepatocytesDendritic Cells

Epithelial CellsKeratinocytesLymphocytesSplenocytesMonocytes

Simple 20 µl Cell-Based Assays• Pre-qualifiedreagents

• Small20µlsamplesize

• Simple,image-basedanalysis

• Pre-definedinstrumentsettings

• Assay-specificdatatemplates

• Accurate,consistentresults

1.Pipette20µl 2. Insert slide & count

3. Get images & data

Cell Population

% of Gated Cells

Concentration (106 cells/mL)

Total 100 3.18

SubG1 3.8 0.12

G0/G1 61.9 1.97

S 15.3 0.49

G2/M 19 0.60

Population gate: 94.7% of total cells countedNumber of gated cells: 4475 cellsConcentration of gated cells: 3.18 x10^6 cells/mL

Assay Infor ation

Assay Name: FCS_Cell Cycle-PIDescription:Date of experiment: 06/15/2018 10:55:09Instrument: Spectrum-101-0002Optics Module 1: S1-655-527 Exposure Time: 300.0 msec

CC4_8 images.NXDat

0 7500 15000 22500 300000

6.25

12.50

18.75

25.00

Population

Export to FCS ExpressTM* for Flow-Like Data

• Exportimagedataintoflowcytometrysoftware: FCS ExpressTM

• Simpleworkflow:Nofluid-stream,NoPMTvoltages,Noforward/sidescatter

• Easily perform data analysis using pre-designed templates

• Quickly plot cell population data as a: histogram,scatter,dotorcontourplot

Size(m

icron)

FL 1 (intensity)

* FCS ExpressTM Flow Cytometry software is a product of De Novo SoftwareTM and is included with the Cellometer SpectrumTM

CountedPI+Cells ContourPopulationPlot

User-Changeable Fluorescence Optics Modules*

*This table is a partial list of compatible fluorophores, nucleic acid stains, and fluorescent proteins. Please contact Nexcelom technical support regarding compatibility of other reagents.Sytox, AlexaFluor, and Cy are trademarks of Life Technologies.

Cellometerassaysusespecificopticmodulesformaximumperformanceanddiscrimination between fluorescence channels. Example filtermodules arelistedbelow.Additionalfiltermodulesavailableuponrequest.Pleasecheckourwebsiteforthemostrecentmodulesavailable.

Optics Module Fluorophores Nucleic Acid Stains

S1-534-470Ex: 470nm (452-498nm)Em: 534nm (510-560nm)

GFPCalceinFITCAlexaFluor® 488

AO(acridineorange)SYTO®9,SYTO®13SYTOX®GreenSYTO®BC

S1-605-527Ex: 527nm (498-550nm)Em: 605nm (570-638nm)

AlexaFluor® 546AlexaFluor® 555Cy3®

PE(R-phycoerythrin)RFP

PI(propidiumiodide)EB(ethidiumbromide)SYTOX®Orange

S1-655-527Ex: 527nm (525-558nm)Em: 655nm (642-665nm)

RFP7-AADNile Red

PI(propidiumiodide)EB(ethidiumbromide)

R2=0.99498

0.00%10.00%20.00%30.00%40.00%50.00%60.00%70.00%80.00%90.00%

100.00%

0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100%

Expe

cted

% V

iabi

lity

Measured % Viability

Measured Jurkat Cell Viability

Proven Results

Bright Field StainedAO+Cells

StainedPICells CombinedCountedCells

PI (intensity)

Cou

nts

0 7000 14000 21000 28000 350000

17

34

52

69

86

Legend Sub G1 G0/G1 S G2/M

Propidium Iodide (PI) Cell Cycle

CD3 Bright Field CD3PE

FL1 ( int ensit y )

Coun

t

0 101 1020

12

23

35

46

CD3+90.14%

Bright Field GFP/RFPPositiveCells

GFP/RFP Population Analysis

• Determinecellviability,forcelllinesorprimarysamples,usingAO/PIinseconds

• Accuratelymeasurecellsampleswithvaryingviability (0 - 100%)

Measured Jurkat Cell Viability

Histogram of PE CD3+ Jurkat cells

Contour map of Jurkat cells, showing GFP+ and RFP+ intensity profiles.

Accurately count Live (AO) versus Dead (PI) cells.

RFPsity)

FL1 (intensity)

FL2

(inte

nsity

)

101

102

103

104

101

102

103

104

17.57%45.00%

2.72%34.60% Top Left: RFP+Top Right: RFP+ and GFP+Bottom Right: GFP+Bottom Left: FL-

Page 3: Which Instrument is Right for Me? Advantages of Cellometer ... · Simple 20 µl Cell-Based Assays • Pre-qualified reagents • Small20µlsamplesize ... • Determine cell viability,

Image Cytometry System for 20 µl Cell-Based Assays

Cellometer

® Spectrum

Nexcelom Bioscience360 Merrimack Street, Building 9Lawrence, MA 01843, USA

Nexcelom products are for RESEARCH USE ONLY and are not approved for diagnostic or therapeutic use. © Copyright 2018 Nexcelom Bioscience LLC. All Rights Reserved.

Innovation and Expertise in the Science of Cell CountingSchedule a FREE demonstration or technical seminar with a Nexcelom Applications Specialist today.

Call 978-327-5340 or E-mail [email protected]

1001465 Rev B 7/18

For Europe Call 44 (0) 161 232 4593 orE-mail [email protected]

For China Call +86 21 5886 0038 orE-mail [email protected]

Cell CycleApoptosisCell Mediated CytotoxicityCell Counting and ViabilityInterchangeable Fluorescent Modules

Accurate Cell Counting

R2=0.99899

1.00E+04

1.00E+05

1.00E+06

1.00E+07

1.00E+06 1.00E+07 1.00E+08

Theo

retic

al C

once

ntra

tion

[log]

Measured Concentration [log]Counted Cells Trend Line

Broad Dynamic Range1.00E+08

1.00E+051.00E+04

• AutomaticidentificationofAcridineOrange(AO)stainedcells(shownabove)

• Image and count up to 2x107 cells/mL

• Image acquisition and analysis: 30 seconds/sample for a 3-channel assay

• Versatile platform - use for routine cell countingandcomplexcell-basedassays

Bright Field StainedAO+Cells CountedAO+Cells

Cell Imaging• Visually check cell morphology

• Ensure only cells of interest are counted

• Archiveandre-analyzecellimages

• Exportimagesforpublication

Proprietary Pattern-Recognition Software• Countindividualcellsinclusters

• Count irregular-shaped cells

• Countcellsbasedonsize

• Eliminatedebrisfromcellcounts

Non-Fluidic Platform• Disposablecountingchambers-nowashing

• Compatiblewithfragilecells

• Maintenance-free

• RobustopticsmodulesandLEDlightsources

IQ/OQ Validation and GMP/GLP Accessories• InstallationQualificationreagents/protocol

• OperationalQualificationreagents/protocol

• On-siteIQorOQPerformance

• GMP/GLPSoftwareModule

Which Instrument is Right for Me?

Features Bright Field Cell Counters Fluorescent Viability Cell Counters Image Cytometers

Mini Auto T4 Auto 1000

Auto 2000 X1 X2 K2 Spectrum Spectrum

(10x)Celigo

BFCeligo 4 Channel

Celigo 5 Channel

Cell / Sample TypeCell Line X X X X X X X X X

CulturedPrimaryCells X X X X X X X X XAlgae X

Platelets X XLow Concentration Cell Lines X X X X X X

Yeast (Clean Sample) X X XYeast (Messy Sample) X X

Primarycells(MessySample*) X X X X XPBMCs,Splenocytes,StemCells X X X X X

Hepatocytes X X X XAdipocytes*** X X X X X X

Cell-BasedAssay** X X X X X X X XApoptosis(AnnexinV-FITC/PI) X X X X XApoptosis(CaspaseActivity) X X X X X

Autophagy (CytoID-green) X XCellProliferation(CFSE) X X X X

CellCycle(PI) X X X X X X XGFPTransfection X X X X X X XRFPTransfection X X X X

MitochondrialPotential(JC-1) X X X XMulti-drug Resistance (ABC

Transporter) X X X X

Surface Marker Analysis X X X XVitality(Calcein-AM/PI) X X X X X X

Vitality (CFDA-AM) X XImage Cytometry** X X X X X

* A messy sample is a heterogeneous sample containing unwanted cell types, such as red blood cells, in addition to the cells of interest.** FCS Express license must be purchased in order to perform Cell Based Assay or Image Cytometry analysis*** Cellometer CHT4-PD300 slides are required for cells greater than 80µm in diameter

Advantages of Cellometer Image Cytometry

Page 4: Which Instrument is Right for Me? Advantages of Cellometer ... · Simple 20 µl Cell-Based Assays • Pre-qualified reagents • Small20µlsamplesize ... • Determine cell viability,

Cell Types for Many Research Areas

• Clinical Immunology: PBMCs

• Diabetes / Obesity: Adipocytes

• Immunotherapy: Leukocytes

• Microbiology: Yeast (Spectrum 10x)

• Oncology: Cell Lines

• Regenerative Medicine: Stem Cells

• Toxicology: Hepatocytes

• Transplantation: Nucleated Cells

• Vaccine Development: Splenocytes

All-in-One System Basiccellcounting,primarycellviability,andcell-basedassays.

Dual-Fluorescence for Accurate Primary Cell Viability No interference from red blood cells.Analyzebonemarrow,peripheralblood,andcordbloodwithoutlysing.

Unique Algorithms for Advanced Cell AnalysisDetermine concentration and viabilityof hepatocytes, adipocytes, and othersophisticated cell types.

Fast ResultsObtain cell images, counts, sizemeasurementsandviabilitycalculationsin < 30 seconds per sample.

Features of the SpectrumTM Image Cytometry System

Optimizedforprimarycellanalysis

PBMCsStem CellsAdipocytesNeural CellsHepatocytesDendritic Cells

Epithelial CellsKeratinocytesLymphocytesSplenocytesMonocytes

Simple 20 µl Cell-Based Assays• Pre-qualifiedreagents

• Small20µlsamplesize

• Simple,image-basedanalysis

• Pre-definedinstrumentsettings

• Assay-specificdatatemplates

• Accurate,consistentresults

1.Pipette20µl 2. Insert slide & count

3. Get images & data

Cell Population

% of Gated Cells

Concentration (106 cells/mL)

Total 100 3.18

SubG1 3.8 0.12

G0/G1 61.9 1.97

S 15.3 0.49

G2/M 19 0.60

Population gate: 94.7% of total cells countedNumber of gated cells: 4475 cellsConcentration of gated cells: 3.18 x10^6 cells/mL

Assay Infor ation

Assay Name: FCS_Cell Cycle-PIDescription:Date of experiment: 06/15/2018 10:55:09Instrument: Spectrum-101-0002Optics Module 1: S1-655-527 Exposure Time: 300.0 msec

CC4_8 images.NXDat

0 7500 15000 22500 300000

6.25

12.50

18.75

25.00

Population

Export to FCS ExpressTM* for Flow-Like Data

• Exportimagedataintoflowcytometrysoftware: FCS ExpressTM

• Simpleworkflow:Nofluid-stream,NoPMTvoltages,Noforward/sidescatter

• Easily perform data analysis using pre-designed templates

• Quickly plot cell population data as a: histogram,scatter,dotorcontourplot

Size(m

icron)

FL 1 (intensity)

* FCS ExpressTM Flow Cytometry software is a product of De Novo SoftwareTM and is included with the Cellometer SpectrumTM

CountedPI+Cells ContourPopulationPlot

User-Changeable Fluorescence Optics Modules*

*This table is a partial list of compatible fluorophores, nucleic acid stains, and fluorescent proteins. Please contact Nexcelom technical support regarding compatibility of other reagents.Sytox, AlexaFluor, and Cy are trademarks of Life Technologies.

Cellometerassaysusespecificopticmodulesformaximumperformanceanddiscrimination between fluorescence channels. Example filtermodules arelistedbelow.Additionalfiltermodulesavailableuponrequest.Pleasecheckourwebsiteforthemostrecentmodulesavailable.

Optics Module Fluorophores Nucleic Acid Stains

S1-534-470Ex: 470nm (452-498nm)Em: 534nm (510-560nm)

GFPCalceinFITCAlexaFluor® 488

AO(acridineorange)SYTO®9,SYTO®13SYTOX®GreenSYTO®BC

S1-605-527Ex: 527nm (498-550nm)Em: 605nm (570-638nm)

AlexaFluor® 546AlexaFluor® 555Cy3®

PE(R-phycoerythrin)RFP

PI(propidiumiodide)EB(ethidiumbromide)SYTOX®Orange

S1-655-527Ex: 527nm (525-558nm)Em: 655nm (642-665nm)

RFP7-AADNile Red

PI(propidiumiodide)EB(ethidiumbromide)

R2=0.99498

0.00%10.00%20.00%30.00%40.00%50.00%60.00%70.00%80.00%90.00%

100.00%

0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100%

Expe

cted

% V

iabi

lity

Measured % Viability

Measured Jurkat Cell Viability

Proven Results

Bright Field StainedAO+Cells

StainedPICells CombinedCountedCells

PI (intensity)

Cou

nts

0 7000 14000 21000 28000 350000

17

34

52

69

86

Legend Sub G1 G0/G1 S G2/M

Propidium Iodide (PI) Cell Cycle

CD3 Bright Field CD3PE

FL1 ( int ensit y )

Coun

t

0 101 1020

12

23

35

46

CD3+90.14%

Bright Field GFP/RFPPositiveCells

GFP/RFP Population Analysis

• Determinecellviability,forcelllinesorprimarysamples,usingAO/PIinseconds

• Accuratelymeasurecellsampleswithvaryingviability (0 - 100%)

Measured Jurkat Cell Viability

Histogram of PE CD3+ Jurkat cells

Contour map of Jurkat cells, showing GFP+ and RFP+ intensity profiles.

Accurately count Live (AO) versus Dead (PI) cells.

RFPsity)

FL1 (intensity)

FL2

(inte

nsity

)

101

102

103

104

101

102

103

104

17.57%45.00%

2. 72%34.60% Top Left: RFP+Top Right: RFP+ and GFP+Bottom Right: GFP+Bottom Left: FL-

Page 5: Which Instrument is Right for Me? Advantages of Cellometer ... · Simple 20 µl Cell-Based Assays • Pre-qualified reagents • Small20µlsamplesize ... • Determine cell viability,

Cell Types for Many Research Areas

• Clinical Immunology: PBMCs

• Diabetes / Obesity: Adipocytes

• Immunotherapy: Leukocytes

• Microbiology: Yeast (Spectrum 10x)

• Oncology: Cell Lines

• Regenerative Medicine: Stem Cells

• Toxicology: Hepatocytes

• Transplantation: Nucleated Cells

• Vaccine Development: Splenocytes

All-in-One System Basiccellcounting,primarycellviability,andcell-basedassays.

Dual-Fluorescence for Accurate Primary Cell Viability No interference from red blood cells.Analyzebonemarrow,peripheralblood,andcordbloodwithoutlysing.

Unique Algorithms for Advanced Cell AnalysisDetermine concentration and viabilityof hepatocytes, adipocytes, and othersophisticated cell types.

Fast ResultsObtain cell images, counts, sizemeasurementsandviabilitycalculationsin < 30 seconds per sample.

Features of the SpectrumTM Image Cytometry System

Optimizedforprimarycellanalysis

PBMCsStem CellsAdipocytesNeural CellsHepatocytesDendritic Cells

Epithelial CellsKeratinocytesLymphocytesSplenocytesMonocytes

Simple 20 µl Cell-Based Assays• Pre-qualifiedreagents

• Small20µlsamplesize

• Simple,image-basedanalysis

• Pre-definedinstrumentsettings

• Assay-specificdatatemplates

• Accurate,consistentresults

1.Pipette20µl 2. Insert slide & count

3. Get images & data

Cell Population

% of Gated Cells

Concentration (106 cells/mL)

Total 100 3.18

SubG1 3.8 0.12

G0/G1 61.9 1.97

S 15.3 0.49

G2/M 19 0.60

Population gate: 94.7% of total cells countedNumber of gated cells: 4475 cellsConcentration of gated cells: 3.18 x10^6 cells/mL

Assay Infor ation

Assay Name: FCS_Cell Cycle-PIDescription:Date of experiment: 06/15/2018 10:55:09Instrument: Spectrum-101-0002Optics Module 1: S1-655-527 Exposure Time: 300.0 msec

CC4_8 images.NXDat

0 7500 15000 22500 300000

6.25

12.50

18.75

25.00

Population

Export to FCS ExpressTM* for Flow-Like Data

• Exportimagedataintoflowcytometrysoftware: FCS ExpressTM

• Simpleworkflow:Nofluid-stream,NoPMTvoltages,Noforward/sidescatter

• Easily perform data analysis using pre-designed templates

• Quickly plot cell population data as a: histogram,scatter,dotorcontourplot

Size(m

icron)

FL 1 (intensity)

* FCS ExpressTM Flow Cytometry software is a product of De Novo SoftwareTM and is included with the Cellometer SpectrumTM

CountedPI+Cells ContourPopulationPlot

User-Changeable Fluorescence Optics Modules*

*This table is a partial list of compatible fluorophores, nucleic acid stains, and fluorescent proteins. Please contact Nexcelom technical support regarding compatibility of other reagents.Sytox, AlexaFluor, and Cy are trademarks of Life Technologies.

Cellometerassaysusespecificopticmodulesformaximumperformanceanddiscrimination between fluorescence channels. Example filtermodules arelistedbelow.Additionalfiltermodulesavailableuponrequest.Pleasecheckourwebsiteforthemostrecentmodulesavailable.

Optics Module Fluorophores Nucleic Acid Stains

S1-534-470Ex: 470nm (452-498nm)Em: 534nm (510-560nm)

GFPCalceinFITCAlexaFluor® 488

AO(acridineorange)SYTO®9,SYTO®13SYTOX®GreenSYTO®BC

S1-605-527Ex: 527nm (498-550nm)Em: 605nm (570-638nm)

AlexaFluor® 546AlexaFluor® 555Cy3®

PE(R-phycoerythrin)RFP

PI(propidiumiodide)EB(ethidiumbromide)SYTOX®Orange

S1-655-527Ex: 527nm (525-558nm)Em: 655nm (642-665nm)

RFP7-AADNile Red

PI(propidiumiodide)EB(ethidiumbromide)

R2=0.99498

0.00%10.00%20.00%30.00%40.00%50.00%60.00%70.00%80.00%90.00%

100.00%

0% 10% 20% 30% 40% 50% 60% 70% 80% 90% 100%

Expe

cted

% V

iabi

lity

Measured % Viability

Measured Jurkat Cell Viability

Proven Results

Bright Field StainedAO+Cells

StainedPICells CombinedCountedCells

PI (intensity)

Cou

nts

0 7000 14000 21000 28000 350000

17

34

52

69

86

Legend Sub G1 G0/G1 S G2/M

Propidium Iodide (PI) Cell Cycle

CD3 Bright Field CD3PE

FL1 ( int ensit y )

Coun

t

0 101 1020

12

23

35

46

CD3+90.14%

Bright Field GFP/RFPPositiveCells

GFP/RFP Population Analysis

• Determinecellviability,forcelllinesorprimarysamples,usingAO/PIinseconds

• Accuratelymeasurecellsampleswithvaryingviability (0 - 100%)

Measured Jurkat Cell Viability

Histogram of PE CD3+ Jurkat cells

Contour map of Jurkat cells, showing GFP+ and RFP+ intensity profiles.

Accurately count Live (AO) versus Dead (PI) cells.

RFPsity)

FL1 (intensity)

FL2

(inte

nsity

)

101

102

103

104

101

102

103

104

17.57%45.00%

2. 72%34.60% Top Left: RFP+Top Right: RFP+ and GFP+Bottom Right: GFP+Bottom Left: FL-

Page 6: Which Instrument is Right for Me? Advantages of Cellometer ... · Simple 20 µl Cell-Based Assays • Pre-qualified reagents • Small20µlsamplesize ... • Determine cell viability,

Image Cytometry System for 20 µl Cell-Based Assays

Cellometer

® Spectrum

Nexcelom Bioscience360 Merrimack Street, Building 9Lawrence, MA 01843, USA

Nexcelom products are for RESEARCH USE ONLY and are not approved for diagnostic or therapeutic use. © Copyright 2018 Nexcelom Bioscience LLC. All Rights Reserved.

Innovation and Expertise in the Science of Cell CountingSchedule a FREE demonstration or technical seminar with a Nexcelom Applications Specialist today.

Call 978-327-5340 or E-mail [email protected]

1001465 Rev B 7/18

For Europe Call 44 (0) 161 232 4593 orE-mail [email protected]

For China Call +86 21 5886 0038 orE-mail [email protected]

Cell CycleApoptosisCell Mediated CytotoxicityCell Counting and ViabilityInterchangeable Fluorescent Modules

Accurate Cell Counting

R2=0.99899

1.00E+04

1.00E+05

1.00E+06

1.00E+07

1.00E+06 1.00E+07 1.00E+08

Theo

retic

al C

once

ntra

tion

[log]

Measured Concentration [log]Counted Cells Trend Line

Broad Dynamic Range1.00E+08

1.00E+051.00E+04

• AutomaticidentificationofAcridineOrange(AO)stainedcells(shownabove)

• Image and count up to 2x107 cells/mL

• Image acquisition and analysis: 30 seconds/sample for a 3-channel assay

• Versatile platform - use for routine cell countingandcomplexcell-basedassays

Bright Field StainedAO+Cells CountedAO+Cells

Cell Imaging• Visually check cell morphology

• Ensure only cells of interest are counted

• Archive andre-analyze cellimages

• Exportimagesforpublication

Proprietary Pattern-Recognition Software• Countindividualcellsinclusters

• Count irregular-shaped cells

• Countcellsbasedonsize

• Eliminatedebrisfrom cellcounts

Non-Fluidic Platform• Disposablecountingchambers-nowashing

• Compatiblewithfragilecells

• Maintenance-free

• RobustopticsmodulesandLEDlightsources

IQ/OQ Validation and GMP/GLP Accessories• InstallationQualificationreagents/protocol

• OperationalQualificationreagents/protocol

• On-siteIQorOQPerformance

• GMP/GLPSoftware Module

Which Instrument is Right for Me?

Features Bright Field Cell Counters Fluorescent Viability Cell Counters Image Cytometers

Mini Auto T4 Auto 1000

Auto 2000 X1 X2 K2 Spectrum Spectrum

(10x)Celigo

BFCeligo 4 Channel

Celigo 5 Channel

Cell / Sample TypeCell Line X X X X X X X X X

CulturedPrimaryCells X X X X X X X X XAlgae X

Platelets X XLow Concentration Cell Lines X X X X X X

Yeast (Clean Sample) X X XYeast (Messy Sample) X X

Primarycells(MessySample*) X X X X XPBMCs,Splenocytes,StemCells X X X X X

Hepatocytes X X X XAdipocytes*** X X X X X X

Cell-BasedAssay** X X X X X X X XApoptosis(AnnexinV-FITC/PI) X X X X XApoptosis(CaspaseActivity) X X X X X

Autophagy (CytoID-green) X XCellProliferation(CFSE) X X X X

CellCycle(PI) X X X X X X XGFPTransfection X X X X X X XRFPTransfection X X X X

MitochondrialPotential(JC-1) X X X XMulti-drug Resistance (ABC

Transporter) X X X X

Surface Marker Analysis X X X XVitality(Calcein-AM/PI) X X X X X X

Vitality (CFDA-AM) X XImage Cytometry** X X X X

* A messy sample is a heterogeneous sample containing unwanted cell types, such as red blood cells, in addition to the cells of interest.** FCS Express license must be purchased in order to perform Cell Based Assay or Image Cytometry analysis*** Cellometer CHT4-PD300 slides are required for cells greater than 80µm in diameter

Advantages of Cellometer Image Cytometry