Upload
valentine-freeman
View
218
Download
3
Embed Size (px)
Citation preview
Wilkes University - CHM 341
2-D Gas Chromatography (or 3D with MS Detector)
A powerful separations tool for complex volatile mixtures of heat-stable samples
[information shamelessly taken fromWikipedia
andother sources – see last slide]
History Comprehensive 2-D gas chromatography (GC)
(abbreviated GC X GC) a mature technique with commercial products on the market technique has tremendous separation power uses simple robust hardware similar analysis times to temperature-programmed high-
resolution capillary chromatography
Comprehensive 2-D LC (LC X LC) still in its infancy more complex to perform driven by user needs, especially in proteomics
The Concept additional physico-chemical criterion employed for the separation of the
mixture of analytes (sample) resolution and quality of chromatographic separation can be increased higher specifity of the separation capability is obtained separation of compounds indistinguishable by 1-D chromatography
Gas-Phase Chromatography, 2-D [an illustrative example] coupling a second, short column to the first long column shock-freeze eluents in order of elution from the 1st column reheat them in order of elution for release into the 2nd column transit time through the 2nd column needs to be shorter than the time until the
next sample is reheated & released Gas-Phase Chromatography, 2-D [an illustrative example]
one column is used to separate analytes followed by Time-of-Flight Mass Spectrometer (TOFMS) detection as the
second dimension TOF-Mass Spectrometers used in gas chromatography can be very short [ due to the limited range of m/z required]
The Instrument - Background
The Instrument Requirements two pieces of hardware are added to a
conventional GC a second column an interface between the second column and the
first (the modulator) the time for analysis on the second column is
very fast (short column) the interface repetitively samples the effluent
from the first column, and injects it onto the second
The Instrument Requirements Rt on the two columns may be thought of as lying
perpendicular to one another 2nd column stationary phase different from the 1st
retention mechanism different from that of the 1st retention mechanism are “uncorrelated”, “independent”,
or “orthogonal” molecules are separated on the basis of independent
chemical properties in the 1st and 2nd columns Ex. 1st column separates based on the basis of “molecular size”
(volatility/b.p.) 2nd column separates on the basis of polarity
the molecular property “polarity” is largely independent of the molecular property “size,” or volatility
Polar funct. groups can be attached to compounds of any size
The Instrument Requirements Modulator performs 3 tasks (in repetitive cycle):
accumulates sample eluting from 1st column period of time equal to 1/3 to 1/5, of the duration of an
individual peak from the first column Ex. a first column peak is 9 seconds wide at the base: modulator
will accumulate material every 2 (or 3) seconds, thereby “chopping” the peak eluting from the first column into “cuts”
focuses the material collected from each cut into a narrow “band”, “plug”, or “chemical pulse”. by flash-freezing (with, for ex., a cold jet of CO2)
“launches” or injects the sharp chemical pulses sequentially onto the second column a series of high speed gas chromatographic separations occur one separation for each chemical pulse launched onto the second
column
Data interpretation
each vertical column of the image may be integrated and plotted as a function of 1st column elution time the conventional gas chromatogram, or “first-dimension
chromatogram,” appears scanning downward from each “peak” in the first
dimension chromatogram, one can count the number of coeluents, visible as discernable second-dimension peaks, of which
the conventional one-dimensional peak actually consists
Data interpretation
Most of the colored spots -- the chromatographic peaks -- in images such as Figure 4 (above) are believed to represent one, or a very few, of the chemical species present in the sample.
In the case of the diesel oil appearing in Figure 4, some 5,000 peaks are discernable.
Even with this very high peak count, co-elutions still occur at the higher carbon number region on the right hand side of the image. Nonetheless, valuable information is still available from rich and chemically significant peak patterns.
In the example of Figure 4, chemical classes are clearly visible. Column bleed products eluting from the first column are also clearly
distinguishable from the sample matrix. Diagonal sub-bands appear throughout the chromatogram,
corresponding to groups of isomers – the so-called “roof-tile” effect.
Data interpretation – the 3rd D
Data interpretation – the 3rd D
Data interpretation – the 3rd D
Liver – Drug induced damage
Data interpretation – the 3rd D
Data interpretation – the 3rd D
For more info . . .
Visit Zoex Corporation for info taken from their technical note http://www.zoex.com/technote_kt030505-1.html
LECO Corporation – product flyer PegasusHThttp://www.leco.com/products/sep_sci/pegasus_ht/pegasus_ht.htm
LECO Corporation – technical noteshttp://www.leco.com/resources/application_note_subs/
separation_science_application_notes.htm
Visit this site for info on comprehenisive LCxLChttp://www.chromatographyonline.com/lcgc/article/articleDetail.jsp?id=187969