[1] DS-004E TLG-837 Sulfur Pit Analyzer
An ultra-safe sulfur pit analyzer for monitoring H2SSO2 accumulation
The TLG-837 continuously measures the concentrations of H2S and SO2 in the headspace above a liquid sulfur pit Using the patented in situ DEMISTER sampling probe and a full-spectrum UV-Vis spectrophotometer this system provides extremely fast accurate response for prompt response to any dangerous buildup of H2S or SO2 which are early warnings for potential fires and explosions Keeping a vigilant watch on sulfur pit gases the TLG-837 keeps the entire plant safe
Features
raquo Continuously measures concentrations of H2S and SO2 in vapor using dispersive UV-Vis spectrophotometer
raquo Patented DEMISTER sampling probe with internal sulfur vapor removal
raquo Totally solid state with no moving parts sample lines or heat tracing mdash modern design for low maintenance
raquo Ultra-safe fiber optic design mdash no toxicexplosive sample gas in analyzer enclosure
H2S SO2
TLG-837 Sulfur Pit AnalyzerApplied Analytics Data Sheet No DS-004E
[2]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
The Sulfur Pit Analysis Need
The liquid sulfur produced in the process of refining petroleum is often stored in a concrete lsquosulfur pitrsquo in the ground The vapor headspace above the liquid sulfur often contains H2S which is extremely toxic and explosive In order to ensure a safe operating environment this area needs to be continuously purged of H2S
To validate this purging process and provide round-the-clock safety real-time H2S measurement is required along with an alarm when the H2S concentration approaches its lower explosive limit (LEL) in air The SO2 measurement should also be incorporated in order to detect smoldering fires in the liquid sulfur since these sulfur fires tend to spray SO2 fumes upward a sensitive detector can recognize early warnings of spreading fire
Sulfur Recovery Unit Schematic
In the diagram below the sulfur pit analysis application is identified by AT7
[3]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
TLG-837 Principle of Operation
xenon light source
holographicgrating
diode array
fiber optic cables
flow cell disk
o-ring
collimator
cooling extension
steam tracing
The TLG-837 measurement cycle is instantaneous but it can be helpful to visualize it in stages
(1) The white light signal originates in the pulsed Xe lamp that functions as the light source
(2) The signal travels via fiber optic cable to the flow cell A collimator narrows the light beam
(3) The signal travels directly across the flow cell interacting with the continuously drawn process sample
(4) The signal exits the flow cell through a collimator now containing the distinct absorbance imprint of the current chemical composition of the sample
(5) The signal travels via fiber optic cable to the nova II
(6) The signal is dispersed by the holographic grating Each differentiated wavelength is focused onto a designated photodiode within the diode array The nova II provides this rich data to the HMI for real-time visualization of the absorbance spectrum
[4]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
The Worldrsquos Safest Sulfur Pit Analyzer
Applied Analytics design centers on intrinsic safety The major safety flaw of other sulfur pit analyzers is that they bring the toxic sample fluid into the analyzer enclosure for analysis Not only does this practice expose the system electronics to higher corrosion effects it also poses a lethal threat if there is any leak in the instrument mdash especially inside a shelter mdash the human operator is placed at enormous risk
The key difference between the TLG-837 and other sulfur pit analyzers is the use of fiber optic cables we bring the light to the sample instead of bringing the sample to the light The toxic sample only needs to circulate through the probe and never enters the analyzer electronics enclosure
Gas-FreeAnalyzerEnclosure
= sample transport route= wetted parts
Key safety features include
raquo No danger of leaks inside the analyzer because the sample gas does not enter the analyzer enclosure
raquo No need for a shelter mdash system designed for outdoor environment
raquo Custom fiber length up to 6 meters allows for distance between analyzer and probe
raquo User can safely perform service on the analyzer while process is running mdash no exposure to sample gas
raquo Digital link (eg Modbus) provides additional process data during any upset conditions mdash personnel do not need to physically visit the analyzer during potentially dangerous situations
raquo Full port 2rdquo ball valve sealsunseals the process allowing user to isolate and remove probe while process is running
[5]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Analyzer Internal Components
human machine interface (HMI)
nova IItrade spectrophotometer
IO board
powersupplies
circuitbreakers
Auto Zero Functionality
The Auto Zero runs on a custom schedule to normalize the spectrophotometer readings while running a zero-absorbance fluid through the probe flow cell disk
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer[6]
[7]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Technical DataGeneralMeasurement Principle Dispersive UV-Vis SW-NIR absorbance spectrophotometryDetector nova IItrade Spectrophotometer
Data sheet httpwwwa-a-inccomdocumentsAA_DS201A_novaIIpdfSpectral Range 200-800 nmLight Source Standard pulsed xenon lamp with average 5 year lifespanFiber Optic Cables Standard 600 μm core 18 meter fiber optic cables (qty = 2)
Data sheet httpwwwa-a-inccomdocumentsAA_DS206A_FiberOpticspdfSample Introduction DEMISTER in situ sampling probe
Data sheet httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_ProbepdfAnalyzer Calibration One-time calibration at factory or site with certified calibration gas no re-calibration ever
required Auto Zero normalizes spectrophotometer reading on zero-absorbance fluidReading Verification Simple verification with samples or neutral density filtersHuman Machine Interface Applied Analytics standard HMI industrial controller with touch-screen LCD display
Data sheet httpwwwa-a-inccomdocumentsAA_DS202A_HMIpdfUser Interface ECLIPSEtrade Runtime Software
Data sheet httpwwwa-a-inccomdocumentsAA_DS203A_EclipsepdfData Storage 32GB Solid State Drive
Data sheet httpwwwa-a-inccomdocumentsAA_DS204A_SSDpdfEnclosure Standard wall-mounted NEMA 4X stainless steel type 304 Enclosure
Options in data sheet httpwwwa-a-inccomdocumentsAA_DS401X_EnclosurespdfCertifications Class I Division 2
Class I Division 1 mdash optionalATEX Exp II 2(2) GD mdash optionalAny other certification mdash please inquire
Measuring ParametersAccuracy amp Repeatability Analyte Typical Range Accuracy Repeatability
H2S 0-2000 ppm plusmn1 of measurement plusmn04
SO2 0-1000 ppm plusmn1 of measurement plusmn04
Zero Drift plusmn01 after 1hr warm-up measured over 24hrs at constant ambient temperatureSensitivity plusmn01 full scaleNoise plusmn0004 AU at 220 nmOff-Ratio Range 1001 gt H2SSO2 ratio gt 120Response Time (T10 - T90) 10 secondsAmbient ConditionsAnalyzer Environment IndoorOutdoor (no shelter required)
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics For any other arrangement please inquire directly with Sales
Sub
ject
to m
odifi
catio
ns S
peci
fied
prod
uct c
hara
cter
istic
s an
d te
chni
cal d
ata
do n
ot s
erve
as
guar
ante
e de
clar
atio
ns
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
[2]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
The Sulfur Pit Analysis Need
The liquid sulfur produced in the process of refining petroleum is often stored in a concrete lsquosulfur pitrsquo in the ground The vapor headspace above the liquid sulfur often contains H2S which is extremely toxic and explosive In order to ensure a safe operating environment this area needs to be continuously purged of H2S
To validate this purging process and provide round-the-clock safety real-time H2S measurement is required along with an alarm when the H2S concentration approaches its lower explosive limit (LEL) in air The SO2 measurement should also be incorporated in order to detect smoldering fires in the liquid sulfur since these sulfur fires tend to spray SO2 fumes upward a sensitive detector can recognize early warnings of spreading fire
Sulfur Recovery Unit Schematic
In the diagram below the sulfur pit analysis application is identified by AT7
[3]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
TLG-837 Principle of Operation
xenon light source
holographicgrating
diode array
fiber optic cables
flow cell disk
o-ring
collimator
cooling extension
steam tracing
The TLG-837 measurement cycle is instantaneous but it can be helpful to visualize it in stages
(1) The white light signal originates in the pulsed Xe lamp that functions as the light source
(2) The signal travels via fiber optic cable to the flow cell A collimator narrows the light beam
(3) The signal travels directly across the flow cell interacting with the continuously drawn process sample
(4) The signal exits the flow cell through a collimator now containing the distinct absorbance imprint of the current chemical composition of the sample
(5) The signal travels via fiber optic cable to the nova II
(6) The signal is dispersed by the holographic grating Each differentiated wavelength is focused onto a designated photodiode within the diode array The nova II provides this rich data to the HMI for real-time visualization of the absorbance spectrum
[4]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
The Worldrsquos Safest Sulfur Pit Analyzer
Applied Analytics design centers on intrinsic safety The major safety flaw of other sulfur pit analyzers is that they bring the toxic sample fluid into the analyzer enclosure for analysis Not only does this practice expose the system electronics to higher corrosion effects it also poses a lethal threat if there is any leak in the instrument mdash especially inside a shelter mdash the human operator is placed at enormous risk
The key difference between the TLG-837 and other sulfur pit analyzers is the use of fiber optic cables we bring the light to the sample instead of bringing the sample to the light The toxic sample only needs to circulate through the probe and never enters the analyzer electronics enclosure
Gas-FreeAnalyzerEnclosure
= sample transport route= wetted parts
Key safety features include
raquo No danger of leaks inside the analyzer because the sample gas does not enter the analyzer enclosure
raquo No need for a shelter mdash system designed for outdoor environment
raquo Custom fiber length up to 6 meters allows for distance between analyzer and probe
raquo User can safely perform service on the analyzer while process is running mdash no exposure to sample gas
raquo Digital link (eg Modbus) provides additional process data during any upset conditions mdash personnel do not need to physically visit the analyzer during potentially dangerous situations
raquo Full port 2rdquo ball valve sealsunseals the process allowing user to isolate and remove probe while process is running
[5]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Analyzer Internal Components
human machine interface (HMI)
nova IItrade spectrophotometer
IO board
powersupplies
circuitbreakers
Auto Zero Functionality
The Auto Zero runs on a custom schedule to normalize the spectrophotometer readings while running a zero-absorbance fluid through the probe flow cell disk
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer[6]
[7]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Technical DataGeneralMeasurement Principle Dispersive UV-Vis SW-NIR absorbance spectrophotometryDetector nova IItrade Spectrophotometer
Data sheet httpwwwa-a-inccomdocumentsAA_DS201A_novaIIpdfSpectral Range 200-800 nmLight Source Standard pulsed xenon lamp with average 5 year lifespanFiber Optic Cables Standard 600 μm core 18 meter fiber optic cables (qty = 2)
Data sheet httpwwwa-a-inccomdocumentsAA_DS206A_FiberOpticspdfSample Introduction DEMISTER in situ sampling probe
Data sheet httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_ProbepdfAnalyzer Calibration One-time calibration at factory or site with certified calibration gas no re-calibration ever
required Auto Zero normalizes spectrophotometer reading on zero-absorbance fluidReading Verification Simple verification with samples or neutral density filtersHuman Machine Interface Applied Analytics standard HMI industrial controller with touch-screen LCD display
Data sheet httpwwwa-a-inccomdocumentsAA_DS202A_HMIpdfUser Interface ECLIPSEtrade Runtime Software
Data sheet httpwwwa-a-inccomdocumentsAA_DS203A_EclipsepdfData Storage 32GB Solid State Drive
Data sheet httpwwwa-a-inccomdocumentsAA_DS204A_SSDpdfEnclosure Standard wall-mounted NEMA 4X stainless steel type 304 Enclosure
Options in data sheet httpwwwa-a-inccomdocumentsAA_DS401X_EnclosurespdfCertifications Class I Division 2
Class I Division 1 mdash optionalATEX Exp II 2(2) GD mdash optionalAny other certification mdash please inquire
Measuring ParametersAccuracy amp Repeatability Analyte Typical Range Accuracy Repeatability
H2S 0-2000 ppm plusmn1 of measurement plusmn04
SO2 0-1000 ppm plusmn1 of measurement plusmn04
Zero Drift plusmn01 after 1hr warm-up measured over 24hrs at constant ambient temperatureSensitivity plusmn01 full scaleNoise plusmn0004 AU at 220 nmOff-Ratio Range 1001 gt H2SSO2 ratio gt 120Response Time (T10 - T90) 10 secondsAmbient ConditionsAnalyzer Environment IndoorOutdoor (no shelter required)
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics For any other arrangement please inquire directly with Sales
Sub
ject
to m
odifi
catio
ns S
peci
fied
prod
uct c
hara
cter
istic
s an
d te
chni
cal d
ata
do n
ot s
erve
as
guar
ante
e de
clar
atio
ns
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
[3]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
TLG-837 Principle of Operation
xenon light source
holographicgrating
diode array
fiber optic cables
flow cell disk
o-ring
collimator
cooling extension
steam tracing
The TLG-837 measurement cycle is instantaneous but it can be helpful to visualize it in stages
(1) The white light signal originates in the pulsed Xe lamp that functions as the light source
(2) The signal travels via fiber optic cable to the flow cell A collimator narrows the light beam
(3) The signal travels directly across the flow cell interacting with the continuously drawn process sample
(4) The signal exits the flow cell through a collimator now containing the distinct absorbance imprint of the current chemical composition of the sample
(5) The signal travels via fiber optic cable to the nova II
(6) The signal is dispersed by the holographic grating Each differentiated wavelength is focused onto a designated photodiode within the diode array The nova II provides this rich data to the HMI for real-time visualization of the absorbance spectrum
[4]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
The Worldrsquos Safest Sulfur Pit Analyzer
Applied Analytics design centers on intrinsic safety The major safety flaw of other sulfur pit analyzers is that they bring the toxic sample fluid into the analyzer enclosure for analysis Not only does this practice expose the system electronics to higher corrosion effects it also poses a lethal threat if there is any leak in the instrument mdash especially inside a shelter mdash the human operator is placed at enormous risk
The key difference between the TLG-837 and other sulfur pit analyzers is the use of fiber optic cables we bring the light to the sample instead of bringing the sample to the light The toxic sample only needs to circulate through the probe and never enters the analyzer electronics enclosure
Gas-FreeAnalyzerEnclosure
= sample transport route= wetted parts
Key safety features include
raquo No danger of leaks inside the analyzer because the sample gas does not enter the analyzer enclosure
raquo No need for a shelter mdash system designed for outdoor environment
raquo Custom fiber length up to 6 meters allows for distance between analyzer and probe
raquo User can safely perform service on the analyzer while process is running mdash no exposure to sample gas
raquo Digital link (eg Modbus) provides additional process data during any upset conditions mdash personnel do not need to physically visit the analyzer during potentially dangerous situations
raquo Full port 2rdquo ball valve sealsunseals the process allowing user to isolate and remove probe while process is running
[5]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Analyzer Internal Components
human machine interface (HMI)
nova IItrade spectrophotometer
IO board
powersupplies
circuitbreakers
Auto Zero Functionality
The Auto Zero runs on a custom schedule to normalize the spectrophotometer readings while running a zero-absorbance fluid through the probe flow cell disk
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer[6]
[7]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Technical DataGeneralMeasurement Principle Dispersive UV-Vis SW-NIR absorbance spectrophotometryDetector nova IItrade Spectrophotometer
Data sheet httpwwwa-a-inccomdocumentsAA_DS201A_novaIIpdfSpectral Range 200-800 nmLight Source Standard pulsed xenon lamp with average 5 year lifespanFiber Optic Cables Standard 600 μm core 18 meter fiber optic cables (qty = 2)
Data sheet httpwwwa-a-inccomdocumentsAA_DS206A_FiberOpticspdfSample Introduction DEMISTER in situ sampling probe
Data sheet httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_ProbepdfAnalyzer Calibration One-time calibration at factory or site with certified calibration gas no re-calibration ever
required Auto Zero normalizes spectrophotometer reading on zero-absorbance fluidReading Verification Simple verification with samples or neutral density filtersHuman Machine Interface Applied Analytics standard HMI industrial controller with touch-screen LCD display
Data sheet httpwwwa-a-inccomdocumentsAA_DS202A_HMIpdfUser Interface ECLIPSEtrade Runtime Software
Data sheet httpwwwa-a-inccomdocumentsAA_DS203A_EclipsepdfData Storage 32GB Solid State Drive
Data sheet httpwwwa-a-inccomdocumentsAA_DS204A_SSDpdfEnclosure Standard wall-mounted NEMA 4X stainless steel type 304 Enclosure
Options in data sheet httpwwwa-a-inccomdocumentsAA_DS401X_EnclosurespdfCertifications Class I Division 2
Class I Division 1 mdash optionalATEX Exp II 2(2) GD mdash optionalAny other certification mdash please inquire
Measuring ParametersAccuracy amp Repeatability Analyte Typical Range Accuracy Repeatability
H2S 0-2000 ppm plusmn1 of measurement plusmn04
SO2 0-1000 ppm plusmn1 of measurement plusmn04
Zero Drift plusmn01 after 1hr warm-up measured over 24hrs at constant ambient temperatureSensitivity plusmn01 full scaleNoise plusmn0004 AU at 220 nmOff-Ratio Range 1001 gt H2SSO2 ratio gt 120Response Time (T10 - T90) 10 secondsAmbient ConditionsAnalyzer Environment IndoorOutdoor (no shelter required)
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics For any other arrangement please inquire directly with Sales
Sub
ject
to m
odifi
catio
ns S
peci
fied
prod
uct c
hara
cter
istic
s an
d te
chni
cal d
ata
do n
ot s
erve
as
guar
ante
e de
clar
atio
ns
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
[4]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
The Worldrsquos Safest Sulfur Pit Analyzer
Applied Analytics design centers on intrinsic safety The major safety flaw of other sulfur pit analyzers is that they bring the toxic sample fluid into the analyzer enclosure for analysis Not only does this practice expose the system electronics to higher corrosion effects it also poses a lethal threat if there is any leak in the instrument mdash especially inside a shelter mdash the human operator is placed at enormous risk
The key difference between the TLG-837 and other sulfur pit analyzers is the use of fiber optic cables we bring the light to the sample instead of bringing the sample to the light The toxic sample only needs to circulate through the probe and never enters the analyzer electronics enclosure
Gas-FreeAnalyzerEnclosure
= sample transport route= wetted parts
Key safety features include
raquo No danger of leaks inside the analyzer because the sample gas does not enter the analyzer enclosure
raquo No need for a shelter mdash system designed for outdoor environment
raquo Custom fiber length up to 6 meters allows for distance between analyzer and probe
raquo User can safely perform service on the analyzer while process is running mdash no exposure to sample gas
raquo Digital link (eg Modbus) provides additional process data during any upset conditions mdash personnel do not need to physically visit the analyzer during potentially dangerous situations
raquo Full port 2rdquo ball valve sealsunseals the process allowing user to isolate and remove probe while process is running
[5]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Analyzer Internal Components
human machine interface (HMI)
nova IItrade spectrophotometer
IO board
powersupplies
circuitbreakers
Auto Zero Functionality
The Auto Zero runs on a custom schedule to normalize the spectrophotometer readings while running a zero-absorbance fluid through the probe flow cell disk
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer[6]
[7]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Technical DataGeneralMeasurement Principle Dispersive UV-Vis SW-NIR absorbance spectrophotometryDetector nova IItrade Spectrophotometer
Data sheet httpwwwa-a-inccomdocumentsAA_DS201A_novaIIpdfSpectral Range 200-800 nmLight Source Standard pulsed xenon lamp with average 5 year lifespanFiber Optic Cables Standard 600 μm core 18 meter fiber optic cables (qty = 2)
Data sheet httpwwwa-a-inccomdocumentsAA_DS206A_FiberOpticspdfSample Introduction DEMISTER in situ sampling probe
Data sheet httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_ProbepdfAnalyzer Calibration One-time calibration at factory or site with certified calibration gas no re-calibration ever
required Auto Zero normalizes spectrophotometer reading on zero-absorbance fluidReading Verification Simple verification with samples or neutral density filtersHuman Machine Interface Applied Analytics standard HMI industrial controller with touch-screen LCD display
Data sheet httpwwwa-a-inccomdocumentsAA_DS202A_HMIpdfUser Interface ECLIPSEtrade Runtime Software
Data sheet httpwwwa-a-inccomdocumentsAA_DS203A_EclipsepdfData Storage 32GB Solid State Drive
Data sheet httpwwwa-a-inccomdocumentsAA_DS204A_SSDpdfEnclosure Standard wall-mounted NEMA 4X stainless steel type 304 Enclosure
Options in data sheet httpwwwa-a-inccomdocumentsAA_DS401X_EnclosurespdfCertifications Class I Division 2
Class I Division 1 mdash optionalATEX Exp II 2(2) GD mdash optionalAny other certification mdash please inquire
Measuring ParametersAccuracy amp Repeatability Analyte Typical Range Accuracy Repeatability
H2S 0-2000 ppm plusmn1 of measurement plusmn04
SO2 0-1000 ppm plusmn1 of measurement plusmn04
Zero Drift plusmn01 after 1hr warm-up measured over 24hrs at constant ambient temperatureSensitivity plusmn01 full scaleNoise plusmn0004 AU at 220 nmOff-Ratio Range 1001 gt H2SSO2 ratio gt 120Response Time (T10 - T90) 10 secondsAmbient ConditionsAnalyzer Environment IndoorOutdoor (no shelter required)
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics For any other arrangement please inquire directly with Sales
Sub
ject
to m
odifi
catio
ns S
peci
fied
prod
uct c
hara
cter
istic
s an
d te
chni
cal d
ata
do n
ot s
erve
as
guar
ante
e de
clar
atio
ns
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
[5]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Analyzer Internal Components
human machine interface (HMI)
nova IItrade spectrophotometer
IO board
powersupplies
circuitbreakers
Auto Zero Functionality
The Auto Zero runs on a custom schedule to normalize the spectrophotometer readings while running a zero-absorbance fluid through the probe flow cell disk
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer[6]
[7]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Technical DataGeneralMeasurement Principle Dispersive UV-Vis SW-NIR absorbance spectrophotometryDetector nova IItrade Spectrophotometer
Data sheet httpwwwa-a-inccomdocumentsAA_DS201A_novaIIpdfSpectral Range 200-800 nmLight Source Standard pulsed xenon lamp with average 5 year lifespanFiber Optic Cables Standard 600 μm core 18 meter fiber optic cables (qty = 2)
Data sheet httpwwwa-a-inccomdocumentsAA_DS206A_FiberOpticspdfSample Introduction DEMISTER in situ sampling probe
Data sheet httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_ProbepdfAnalyzer Calibration One-time calibration at factory or site with certified calibration gas no re-calibration ever
required Auto Zero normalizes spectrophotometer reading on zero-absorbance fluidReading Verification Simple verification with samples or neutral density filtersHuman Machine Interface Applied Analytics standard HMI industrial controller with touch-screen LCD display
Data sheet httpwwwa-a-inccomdocumentsAA_DS202A_HMIpdfUser Interface ECLIPSEtrade Runtime Software
Data sheet httpwwwa-a-inccomdocumentsAA_DS203A_EclipsepdfData Storage 32GB Solid State Drive
Data sheet httpwwwa-a-inccomdocumentsAA_DS204A_SSDpdfEnclosure Standard wall-mounted NEMA 4X stainless steel type 304 Enclosure
Options in data sheet httpwwwa-a-inccomdocumentsAA_DS401X_EnclosurespdfCertifications Class I Division 2
Class I Division 1 mdash optionalATEX Exp II 2(2) GD mdash optionalAny other certification mdash please inquire
Measuring ParametersAccuracy amp Repeatability Analyte Typical Range Accuracy Repeatability
H2S 0-2000 ppm plusmn1 of measurement plusmn04
SO2 0-1000 ppm plusmn1 of measurement plusmn04
Zero Drift plusmn01 after 1hr warm-up measured over 24hrs at constant ambient temperatureSensitivity plusmn01 full scaleNoise plusmn0004 AU at 220 nmOff-Ratio Range 1001 gt H2SSO2 ratio gt 120Response Time (T10 - T90) 10 secondsAmbient ConditionsAnalyzer Environment IndoorOutdoor (no shelter required)
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics For any other arrangement please inquire directly with Sales
Sub
ject
to m
odifi
catio
ns S
peci
fied
prod
uct c
hara
cter
istic
s an
d te
chni
cal d
ata
do n
ot s
erve
as
guar
ante
e de
clar
atio
ns
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer[6]
[7]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Technical DataGeneralMeasurement Principle Dispersive UV-Vis SW-NIR absorbance spectrophotometryDetector nova IItrade Spectrophotometer
Data sheet httpwwwa-a-inccomdocumentsAA_DS201A_novaIIpdfSpectral Range 200-800 nmLight Source Standard pulsed xenon lamp with average 5 year lifespanFiber Optic Cables Standard 600 μm core 18 meter fiber optic cables (qty = 2)
Data sheet httpwwwa-a-inccomdocumentsAA_DS206A_FiberOpticspdfSample Introduction DEMISTER in situ sampling probe
Data sheet httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_ProbepdfAnalyzer Calibration One-time calibration at factory or site with certified calibration gas no re-calibration ever
required Auto Zero normalizes spectrophotometer reading on zero-absorbance fluidReading Verification Simple verification with samples or neutral density filtersHuman Machine Interface Applied Analytics standard HMI industrial controller with touch-screen LCD display
Data sheet httpwwwa-a-inccomdocumentsAA_DS202A_HMIpdfUser Interface ECLIPSEtrade Runtime Software
Data sheet httpwwwa-a-inccomdocumentsAA_DS203A_EclipsepdfData Storage 32GB Solid State Drive
Data sheet httpwwwa-a-inccomdocumentsAA_DS204A_SSDpdfEnclosure Standard wall-mounted NEMA 4X stainless steel type 304 Enclosure
Options in data sheet httpwwwa-a-inccomdocumentsAA_DS401X_EnclosurespdfCertifications Class I Division 2
Class I Division 1 mdash optionalATEX Exp II 2(2) GD mdash optionalAny other certification mdash please inquire
Measuring ParametersAccuracy amp Repeatability Analyte Typical Range Accuracy Repeatability
H2S 0-2000 ppm plusmn1 of measurement plusmn04
SO2 0-1000 ppm plusmn1 of measurement plusmn04
Zero Drift plusmn01 after 1hr warm-up measured over 24hrs at constant ambient temperatureSensitivity plusmn01 full scaleNoise plusmn0004 AU at 220 nmOff-Ratio Range 1001 gt H2SSO2 ratio gt 120Response Time (T10 - T90) 10 secondsAmbient ConditionsAnalyzer Environment IndoorOutdoor (no shelter required)
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics For any other arrangement please inquire directly with Sales
Sub
ject
to m
odifi
catio
ns S
peci
fied
prod
uct c
hara
cter
istic
s an
d te
chni
cal d
ata
do n
ot s
erve
as
guar
ante
e de
clar
atio
ns
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
[7]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Technical DataGeneralMeasurement Principle Dispersive UV-Vis SW-NIR absorbance spectrophotometryDetector nova IItrade Spectrophotometer
Data sheet httpwwwa-a-inccomdocumentsAA_DS201A_novaIIpdfSpectral Range 200-800 nmLight Source Standard pulsed xenon lamp with average 5 year lifespanFiber Optic Cables Standard 600 μm core 18 meter fiber optic cables (qty = 2)
Data sheet httpwwwa-a-inccomdocumentsAA_DS206A_FiberOpticspdfSample Introduction DEMISTER in situ sampling probe
Data sheet httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_ProbepdfAnalyzer Calibration One-time calibration at factory or site with certified calibration gas no re-calibration ever
required Auto Zero normalizes spectrophotometer reading on zero-absorbance fluidReading Verification Simple verification with samples or neutral density filtersHuman Machine Interface Applied Analytics standard HMI industrial controller with touch-screen LCD display
Data sheet httpwwwa-a-inccomdocumentsAA_DS202A_HMIpdfUser Interface ECLIPSEtrade Runtime Software
Data sheet httpwwwa-a-inccomdocumentsAA_DS203A_EclipsepdfData Storage 32GB Solid State Drive
Data sheet httpwwwa-a-inccomdocumentsAA_DS204A_SSDpdfEnclosure Standard wall-mounted NEMA 4X stainless steel type 304 Enclosure
Options in data sheet httpwwwa-a-inccomdocumentsAA_DS401X_EnclosurespdfCertifications Class I Division 2
Class I Division 1 mdash optionalATEX Exp II 2(2) GD mdash optionalAny other certification mdash please inquire
Measuring ParametersAccuracy amp Repeatability Analyte Typical Range Accuracy Repeatability
H2S 0-2000 ppm plusmn1 of measurement plusmn04
SO2 0-1000 ppm plusmn1 of measurement plusmn04
Zero Drift plusmn01 after 1hr warm-up measured over 24hrs at constant ambient temperatureSensitivity plusmn01 full scaleNoise plusmn0004 AU at 220 nmOff-Ratio Range 1001 gt H2SSO2 ratio gt 120Response Time (T10 - T90) 10 secondsAmbient ConditionsAnalyzer Environment IndoorOutdoor (no shelter required)
All performance specifications are subject to the assumption that the sample conditioning system and unit installation are approved by Applied Analytics For any other arrangement please inquire directly with Sales
Sub
ject
to m
odifi
catio
ns S
peci
fied
prod
uct c
hara
cter
istic
s an
d te
chni
cal d
ata
do n
ot s
erve
as
guar
ante
e de
clar
atio
ns
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
[8]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit Analyzer
DS-004E TLG-837 Sulfur Pit Analyzer
Ambient Temperature Standard 0 to 40 degC (32 to 104 degF) With optional temperature control -20 to 55 degC (-4 to 131 degF)To avoid radiational heating use of a sunshade is recommended for systems installed in direct sunlight
Physical SpecificationsDimensions Analyzer 24rdquordquo H x 20rdquordquo W x 8rdquordquo D (610mm H x 508mm W x 203mm D)
Prove Average Dimensions 36rdquo length x 12ldquo widest diameter (914mm x 305mm)Optional Utility Control Panel 24rdquo H x 24rdquo W x 8rdquo D (610mm H x 610mm W x 203mm D)
Weight Analyzer 32 lbs (15 kg)Prove Average Weight 29 lbs (13 kg)Optional Utility Control Panel 25 lbs (11 kg)
Wetted Materials Standard Stainless Steel 316316L KalrezVarious custom materials available mdash please inquire
Probe Material Standard Stainless Steel 316316LOther materials available
UtilitiesUtility Management Optional Utility Control Panel
Data sheet httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdfElectrical Requirements 85 to 264 VAC 47 to 63 HzPower Consumption 45 wattsInstrument Air 70 psig (-40 degC dew point)Steam Pressure 70 psig for DEMISTER chamber
30-50 psig for probe blowback function75-100 psig for optional ball valve steam jacket
OutputsCommunicationOutputs 1x galvanically isolated 4-20mA analog output per measured analyte
2x digital outputs for fault and utility controlOptional Modbus TCPIP RS-232 RS-485 Fieldbus Profibus HART more
IO Electronics VoltageCurrent Interface Module (ie IO Board)Data sheet httpwwwa-a-inccomdocumentsAA_DS205A_VCIMpdf
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading
is a registered trademark of Applied Analytics Group BV | wwwa-a-inccom
copy 2013 Applied Analytics Group BV Products or references stated may be trademarks or registered trademarks of their respective owners All rights reserved We reserve the right to make technical changes or modify this document without prior notice Regarding purchase orders agreed-upon details shall prevail
Europe SalesApplied Analytics Europe SpAMilan Italy | salesappliedanalyticseu
Headquarters + ManufacturingApplied Analytics IncBurlington MA | salesa-a-inccom
North America SalesApplied Analytics North America LtdHouston TX | salesappliedanalyticsus
India SalesApplied Analytics (India) Pte LtdMumbai India | salesappliedanalyticsin
Middle East SalesApplied Analytics Middle East (FZE)Sharjah UAE | salesappliedanalyticsae
Asia Pacific SalesApplied Analytics Asia Pte LtdSingapore | salesappliedanalyticscomsg
Brazil SalesApplied Analytics do BrasilRio de Janeiro Brazil | salesaadblcombr
APPLIEDA N A L Y T I C S
TM
DS-004E TLG-837 Sulfur Pit Analyzer[9]
Applied Analytics Data Sheet No DS-004E
TLG-837 Sulfur Pit AnalyzerRevised 31 July 2013
Subject LocationTLG-837 DEMISTER ProbeData sheet
httpwwwa-a-inccomdocumentsAA_DS004B_TLG837_Probepdf
TLG-837 Utility Control PanelData sheet
httpwwwa-a-inccomdocumentsAA_DS004C_TLG837_UCPpdf
Advantage of Collateral DataTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-202_CollateralDatapdf
Multi-Component AnalysisTechnical Note
httpwwwa-a-inccomdocumentsAA_TN-203_MultiComponentAnalysispdf
Further Reading