DesignofLowProfile,ModularLowerExtremityExoskeletons
By
YoonJungJeong
Adissertationsubmittedinpartialsatisfactionofthe
requirementsforthedegreeof
DoctorofPhilosophyin
Engineering–MechanicalEngineering
inthe
GRADUATEDIVISION
ofthe
UNIVERSITYOFCALIFORNIA,BERKELEY
Committeeincharge:
PROFESSORHOMAYOONKAZEROONI,CHAIRPROFESSORDENNISLIEUPROFESSORKIMIKORYOKAI
SPRING2014
DesignofLowProfile,ModularLowerExtremityExoskeletons
Copyright2014byYoonJungJeong
1
Abstract
DesignofLowProfile,ModularLowerExtremityExoskeletons
by
YoonJungJeong
DoctorofPhilosophyinEngineering‐MechanicalEngineering
UniversityofCalifornia,Berkeley
ProfessorH.Kazerooni,Chair
Studieshaveshownthatwalkingforaslittleas30minutesofadaycanimproveoverallhealth.However,morethan250,000peopleintheUnitedStatesareparalyzedduetospinalcordinjuries(SCI),andcannolongerwalkwithoutsupport.MostoftheSCIpatientsrelyonwheelchairsaftertheirinjuries,andsufferfromsecondaryinjuriescausedbyprolongedsitting.Moreover,thepatients’reducedmobilityoftenbringsnegativeeffectstotheirindependenceandsociallife.
Theprimaryresearchobjectiveistodeveloppracticalandeffectiveexoskeletontechnologythatenablesparalyzedindividualstoachievemobilityandgainindependenceintheirdailylives.Inthisdissertation,Idiscussdesignsoflowprofileandmodularexoskeletons,withanemphasisondifferentdesignattributesthatincreasedeviceusability.
Aminimally‐actuatedmedicalexoskeletonwasdevelopedwithpoweredhipjointsandpassivekneejoints.Thelowprofileactuationunitsusedforthisexoskeleton’shipactuationallowedthewearertoachievepropulsionandwalkthroughnarrowpassageways.Infact,thisexoskeletonbecamethefirstpoweredmedicalexoskeletonthatweighslessthantwentypounds.Thisexoskeletoncontinuedtoevolveintoasystemwithincreasedmodularitythatmeetsvarioususerneedsbasedondifferentphysicalconditions.Themodulardesignallowseasycustomizationviaitsversatilesupportlevelconfigurationsfordifferentindividuals.Asapartofthisresearch,comprehensiveuserstudiesofthesedevicesandtheuserinterfacewereconductedtocreateapositiveuserexperienceandacomfortablelinkbetweentheuserandtheexoskeleton.
Paralyzedpatientswhoparticipatedinthisstudycorroboratedthatthecompactandlightweightdesignofthesystemenhancestheirmobilityandmaneuverability.Themodularityindesignopensassistiveopportunitiestovariouspeople,suchastheelderlyandotherpeoplewithintactmobility.Iexpectthatassistiveexoskeletontechnologywillcontinuetoadvanceandenhancethequalityoflifeforawidespectrumofcommunitiesinouragingsociety.
i
TableofContents
TableofContents.......................................................................................................................................................i
ListofFigures............................................................................................................................................................iv
ListofTables...............................................................................................................................................................v
Acknowledgements................................................................................................................................................vi
1.Introduction...........................................................................................................................................................1
1.1Motivation......................................................................................................................................................................1
1.2ResearchObjectives...................................................................................................................................................1
1.3OutlineoftheDissertation......................................................................................................................................2
2.Background............................................................................................................................................................3
2.1 SpinalCordInjury...............................................................................................................................................3
2.2 ConventionalRehabilitationwithPassiveOrthoses..............................................................................4
2.2.1Knee‐Ankle‐Foot‐Orthoses(KAFO).............................................................................................................4
2.2.2ReciprocatingGaitOrthoses(RGO).............................................................................................................4
2.3 PoweredMedicalExoskeletons......................................................................................................................5
2.3.1PoweredMedicalExoskeletonsontheMarket......................................................................................5
2.3.2PoweredMedicalExoskeletonsinAcademicDevelopment..............................................................6
2.4OverviewofHumanWalking..................................................................................................................................7
2.4.1BodyPlanes...........................................................................................................................................................7
2.4.2LowerExtremityJointAngles........................................................................................................................8
2.4.3GaitCycle................................................................................................................................................................8
2.4.4ClinicalGaitAnalysis(CGA)............................................................................................................................9
2.4.5PowerConsumedinaJointduringWalking..........................................................................................10
3.LowProfileExoskeleton................................................................................................................................12
3.1HipActuationUnitsHardwareDesignOverview.........................................................................................13
3.1.1 BrushlessDCMotorandStrainWaveGear...................................................................................14
3.1.2ShaftDesignandSensorMounting............................................................................................................20
3.1.3 BearingstoHoldExternalLoads........................................................................................................20
3.2AssemblyProcessandDesignConsiderationforSafety...........................................................................22
3.2.1AssemblyofaHipActuationUnit..............................................................................................................22
3.2.2DesignConsiderationsforUserSafety.....................................................................................................26
3.3IntegrationintoaMinimally‐ActuatedExoskeleton...................................................................................27
ii3.4LowProfileHipActuatorDesign–SecondRevision..................................................................................29
4.ModularExoskeleton......................................................................................................................................31
4.1MobilitySpectrum.....................................................................................................................................................31
4.2HardwareDesignOverview..................................................................................................................................34
4.3AssessmentoftheModularSystem...................................................................................................................37
5.UserInterfaceModuleDesign.....................................................................................................................40
5.1SchematicsofExplicitUserInterface................................................................................................................41
5.2OverviewofUIPrototype......................................................................................................................................43
5.3GloveUIEvaluation:Methods..............................................................................................................................44
5.3.1PatricipantsOverview....................................................................................................................................44
5.3.2Interviews............................................................................................................................................................44
5.3.3UsabilityTests:MaterialsandProcedure...............................................................................................45
5.4ResultsandDesignRecommendations............................................................................................................46
5.4.1GloveType...........................................................................................................................................................46
5.4.2NumberofButtons...........................................................................................................................................46
5.4.3WalkingOperationScheme..........................................................................................................................47
5.4.4DisplayModulePosition................................................................................................................................47
5.5ImplementationoftheUserStudy.....................................................................................................................48
6.DiscussionsonExoskeletonUserTesting..............................................................................................50
6.1Minimally‐ActuatedMedicalExoskeletonUserTesting............................................................................51
6.1.1TestParticipantOverview............................................................................................................................51
6.1.2UserTestingProceduresandGuideline..................................................................................................52
6.2TrunkExoskeletonUserTesting.........................................................................................................................55
6.2.1TestParticipantOverview............................................................................................................................55
6.2.2UserTestingProcedureandGuideline....................................................................................................55
6.3DesignConsiderationsforFutureUserTesting............................................................................................56
6.4ArchetypalPersonasinEachUserCategories...............................................................................................57
6.4.1Kevin,apoweredwheelchairuser............................................................................................................57
6.4.2Jennifer,awheelchairuser...........................................................................................................................57
6.4.3.Samuel,acrutchuser.....................................................................................................................................58
6.4.4Mary,aseniorcitizen......................................................................................................................................58
6.4.5Tom,anindustryworker...............................................................................................................................58
6.4.6Andrew,amilitarysoldier............................................................................................................................58
7.LimitationsandFutureWork......................................................................................................................59
7.1HipActuators..............................................................................................................................................................60
iii7.2Modularity....................................................................................................................................................................60
7.3UserInterface..............................................................................................................................................................60
8.ConcludingRemarks.......................................................................................................................................62
Bibliography............................................................................................................................................................64
iv
ListofFigures
Figure1.Spinalcordinjurylevelsandremainingextremityfunction.............................................3 Figure2.Mostcommonpassivelegorthoses.............................................................................................5 Figure3.Poweredmobilemedicalexoskeletonsonthemarket........................................................6 Figure4.(a)AUSTINExoskeleton(b)MinaExoskeleton.......................................................................7 Figure5.(a)Referenceplanesofthebody(b)Jointangledefinitions...........................................8 Figure6.Illustrationofanormalwalkingcycle.........................................................................................9 Figure7.Variationsintheswingandstanceperiods..............................................................................9 Figure8.ClinicalGaitAnalysisdataatthe(a)knee,(b)hip...............................................................10 Figure9.Kneeandhipjointpowersduringlevelwalking..................................................................11 Figure10.Lowprofileactuationunitdesignprocessflowchart.....................................................13 Figure11.TheBLDCmotorandstrainwavegearusedfortheactuatordesign.......................14 Figure12.Actuatorperformancelinesatisfyingtherequirementforwalking.........................14 Figure13.Motorperformanceplotoftorqueversusangularvelocity...........................................16 Figure14.Strainwavegearcomponents.....................................................................................................18 Figure15.Strainwavegearoperation..........................................................................................................19 Figure16.Ratchetingphenomenon..............................................................................................................19 Figure17.(a)Cross‐sectionalview(b)Apatientwearingasystem...............................................21 Figure18.Loaddiagramforanactuationunitwithapairofangularbearings........................21 Figure19.Anassembledlowprofilehipactuationunit........................................................................23 Figure20.(a)Thesidecross‐sectionalview(b)Sideviewofassembledunit...........................24 Figure21.Explodedviewoflowprofilehipactuationunit................................................................25 Figure22.Experimentalsetupfortestingthehipactuationunit....................................................26 Figure23.Mechanicalhardstopsforhipextensionandflexion......................................................27 Figure24.Lowprofile,minimally‐actuatedmedicalexoskeleton...................................................28 Figure25.Thetestpilotwasabletowalkthroughnarrowaisles...................................................28 Figure26.Cross‐sectionalviewofanactuationunitwiththeMaxonEC90...............................30 Figure27.Mobilityspectrumoftheinjuredusergroup......................................................................32 Figure28.Extendedmobilityspectrum......................................................................................................32 Figure29.(a)HumanUniversalLoadCarrier(HULC)(b)StrideManagementAssist...........33 Figure30.Overviewofthehipmodulewithadditionaljoints..........................................................35 Figure31.Abduction,adduction,androtationjoints............................................................................36 Figure32.Exampleusesofthetrunkexoskeletonmodule................................................................37 Figure33.Thetrunkexoskeletonmodulewornbyahealthy,non‐injuredindividual..........38 Figure34.ThetrunkexoskeletonmodulewornbyaSCIpatient....................................................38 Figure35.Overviewofthehipmodule........................................................................................................39 Figure36.Anexampleofadevice‐coupledUI..........................................................................................41 Figure37.Anexoskeletonfinitestatemachine(Singleton)...............................................................42 Figure38.Alternatingfinitestatemachine................................................................................................42 Figure39.(a)ThecomponentsoftheGloveUI(b)Aparticipantwearingtheprototype....43 Figure40.UsabilitytestofthegloveUIprototype.................................................................................46 Figure41.FingergloveUIoverview.............................................................................................................49 Figure42.FingerGloveUIwithvisualandhapticfeedback..............................................................49
vFigure43.Sceneoffirsttimeexoskeletontrial........................................................................................51 Figure44.(a)Checkingkneejointalignmentbeforestanding.(b)Thefirsttimestandingup.........................................................................................................................................................................................54 Figure45.(a)Checkingkneejointalignmentafterthefirsttimestandingup.(b)Afteradjustingtorsoandankleangles,thepatientstaysbalancedmoreeasily....................................54 Figure46.(a)Aparaplegicpatienttakingherfirststeps.(b)Anotherparaplegicpatientwearinganexoskeletonwalkingonthestreetwithoutanyassistance.........................................55
ListofTables
Table1.PoweredexoskeletonsandRGOcomparisonchart.................................................................6 Table2.MotorandTransmissionCharacteristics...................................................................................15 Table3.MaxonEC90FlatMotorCharacteristics...................................................................................29 Table4.ComparisonofTwoMotorDesigns...............................................................................................30 Table5.DifferentExoskeletonFeaturesforDifferentUserGroups...............................................34 Table6.SummaryofUIAssessment..............................................................................................................48 Table7.ArchetypalPersonas...........................................................................................................................59
vi
Acknowledgements
Foremost,Iwouldliketoexpressmysinceregratitudetomyadvisor,ProfessorHomayoonKazerooni.ProfessorKazerooniwasthereasonthatIcametoBerkeley,andIamthankfulthatIhavehadsuchaninspiring,enthusiastic,supportive,andcaringadvisor.Hisguidancehelpedmethroughoutmygraduatestudiesandresearchendeavors.Iamlookingforwardtocontinuingmycollaborationwithhimafterthisfive‐yearjourney.
Iwouldalsoliketothankthemembersofmythesiscommitteeandqualifyingcommittee:ProfessorsAliceAgogino,DavidDornfeld,DennisLieu,andKimikoRyokai,fortheirencouragementandinsightfulcomments.Ireallyappreciatethetimetheyspentpatientlyguidingmethroughgraduateschool.
MysincerethanksalsogotoallmycolleaguesandfriendsintheRoboticsandHumanEngineeringLaboratory(HEL).TheyhavemademyPh.D.journeytremendouslyfun,memorable,andenriching.Inparticular,IwanttothankMinervaPillai,MichaelMcKinley,andWayneTungforourstimulatingdiscussions,countlessendeavors,andallthefunwehavehadthesepastfiveyears.IamsogratefulthatIhavehadsuchtalented,fun,andkindcolleagueswhosharedtheirdedicationandcamaraderiewithme.IamalsogratefulforthehardworkandfriendshipofJoshCherian,Lee‐HuangChen,StephenMcKinley,JasonReid,KyunamKim,DongjinHyun,NickUm,andPatrickBarnes;andthenewgenerationofHEL‐ChristinaYee,BradleyPerry,MeitingWu,NicholasEricco,LoganVanEngelhoven,GraceLiao,LilyWu,andOctavioNarváezAroche.Myappreciationfurtherextendstoallofthecurrentandpreviousundergraduatestudents,specificallyAlexanderWen,NathanPoon,JonathanMcKinley,andYaseminSarigul‐Klijn.Theyhavebeentremendouslyhelpfulwithmachining,maintaininghardware,andperformingcountlessothertasks.
Mysinceregratitudealsogoestotheamazinginstructorsofthestudentmachineshop–MickFranssen,GordonLong,DennisLee,ScottMcCormick,andJesseLopez.Theyalwayspatientlyhelpedmewithmyhardwarechallenges.Theirincrediblesupportandadvicemadethebestofmylimitedmachiningskills.
Furthermore,IwanttothankDr.LoraOehlberg,Dr.ElizabethGoodman,ProfessorDorAbrahamson,EuiyoungKim,JenniferWang,andSungtaekLimforinsightfuldiscussions.Itwasveryhelpfultoreceivetheirfeedbackonmyresearchfromtheirdifferentpointsofview.Ireallyappreciatetheirprompt,helpfulresponsestomycoldemails,aswellasthetimetheyspentmeetingandtalkingwithme.
Iwanttoexpressmysincerethankstoallofthetestpilotswhopatientlyworkedwithmeandprovidedinvaluablefeedbackalongtheway.AlthoughIcannotnameeveryoneherebecauseoftheprotocols,theirkindparticipationandhonestfeedbackmadealltheamazingprojectspossible.Thankyouallsomuch.
Andfinally,Iamendlesslygratefultomyfamily‐theirguidance,support,andlovethroughoutmylifemademyPh.D.possible.
1
1.Introduction
1.1MotivationAsof2013,thereweremorethan273,000individualssufferingfromspinalcordinjuries(SCI)intheUnitedStates.Morethan100,000ofthemareparaplegia,paralyzedintheirlowerbodywithouttheabilitytowalkorstandupright.Oncetheylosetheabilitytowalkorstand,theymainlyrelyonawheelchairformobility[1].However,duetothenatureofthehumanbody,SCIpatientswhospendextensiveamountoftimesittinginwheelchairssignificantlyincreasetheirrisksofincurringsecondaryinjuries,whichincludethefollowing[2]:
Urinarytractinfections. Bloodclots. Reducedcardiovascularfunction. Decreasedbonemineraldensity,bonedensityloss,andosteoporosis. Acutepressureulcerdevelopment. Muscularatrophy,musclespasticity,anddecreasedjointrangeofmotion. Reduceddigestiveandbowlfunctions.
Ithasbeenshownthatthebestwaytodelaytheonsetofsecondaryinjuriesistodelaywheelchairusageandstartrehabilitationassoonaspossibleafteraninjury[3].Moreover,beingabletostanduprightandwalkbringspositivepsychologicalandsocialimpacts.Beinginanuprightpositionforseveralhoursadaynotonlydecreasesthechancesofsustainingsecondaryinjuries,butalsoincreasestheSCIpatients’overallqualityoflifeandlifeexpectancy[4].Forthisreason,rehabilitationtherapy,whichcomprisesvarioustypesofequipmentandlowerextremityorthoses,isstronglyrecommendedforSCIpatients.
1.2ResearchObjectivesBothpassiveandpoweredstate‐of‐the‐artexoskeletonshavelimitationsthatpreventtheirwidespreaduse.Whilepassiveorthosesprovidehealthbenefitsatlowercosts,boththeirperformanceandinherentcharacteristicswhichdemandmoreupperbodystrengthinhibitlong‐termuse.Poweredexoskeletonshavebeendevelopedinanefforttoimprovetheperformanceofpassivedevicesthatrequirelessuserstrength.However,drasticallyincreasedcostslimittheiraccessibility,whilesignificantlyhigherdeviceweightsandsizesdecreasesystemportability.Moreimportantly,larger,heavierdevicesimpedeusers’maneuverabilityinconfinedspaces.Thisnotonlyreducesdeviceusability,butalsosignificantlyaffectsusers’independence.
Inadditiontotheabovementionedcharacteristics,existingexoskeletonshavelowusabilityduetotheirlimitedcustomizablefeatures.Theirdesignsareintendedforpatientswithspecificphysicalconditions.Therefore,abroaderpatientrangewithvaryinglevelsofinjuriesandphysicalconditionsneedstobeaddressedtoallowthistechnologytobenefitawiderpopulation.
2Thus,theprimaryresearchobjectiveistoincreasetheusabilityofassistiveexoskeletonsystems.Toachievethisgoal,theresearchexploresthreemajordesigntopics:
1)Developmentoflowprofilepoweredexoskeletonsthatprovideimprovedmobilityanduserindependenceindailylife.
2)Increasesinmodularityandadjustabilityintheexoskeletondesigntoenhancecustomizationandsatisfycomprehensiveuserneeds.
3)Explorationofuserinterface(UI)designtoprovideanintuitive,comfortablelinkbetweentheuserandthemachine.
1.3OutlineoftheDissertationThisdissertationisorganizedasfollows:
1) Chapter2providesbackgroundonthetopicofspinalcordinjuries(SCI)andexoskeletonrehabilitation.Thischapteralsoprovidesabriefoverviewofthebiomechanicshumanwalking.
2) Chapter3presentsthedesignofalowprofileexoskeletonsystem.Itfocusesontheroleoflowprofilehipactuationunits.Thedesignprocess,hardwaredesign,andanassemblyoverviewarediscussed.
3) Chapter4presentsthedesignofamodularexoskeletonsystem.Itelaboratesonthedesignoflockable,modularjointsystemsthatprovidecustomizablesupportforuserswithdifferentneeds.
4) Chapter5discussestheexoskeletonUIdesign.Itpresentstheuserstudywithaglove‐typeUIprototypeforpeoplewithvaryingneeds.Designrecommendationsfromin‐depthuserinterviewsandusabilitytestsarealsodiscussed.
5) Chapter6describesusertestingproceduresandanalysesbasedonHuman‐Computer‐Interaction(HCI)theories.Interactionsamongagentsinvolvedinusertestingaswellasthepatients’“activitiesofre‐learningtowalk”arediscussed.
6) Chapter7discusseslimitationsandfutureworkforlowprofiledesign,modularity,andUIdesign.
7) Chapter8concludesthisdissertationwithasummaryofresults,contributions,andfutureworkontheassistiveexoskeletontechnology.
3
2.Background
2.1 SpinalCordInjuryASCIconsistsofdamageortraumatothespinalcordandresultsinfunctionimpairmentorloss,whichcausesreducedmobilityorfeeling.Commoncausesofdamagecanbecaraccidents,gunshots,falls,sportsinjuries,orvariousdiseases,whichincludeTransverseMyelitis,Polio,SpinaBifida,andFriedreich’sAtaxia[5].
Thespinalcordisamajorbundleofnervesthatcarriesnerveimpulsestoandfromthebraintotherestofthebody[6].Itisabout18inchesinlengthandextendsfromthebaseofthebraintothewaist,alongthemiddleoftheback.Fig.1showsthespinalcordandtheeffectsofthelocationofinjuryonremainingextremityfunction.Thelevelofinjurycanbemorespecificallydescribedbyindicatingtheinjuredvertebra.“C”indicatescervical,“T”indicatesthoracic,“L”indicateslumbar,and“S”indicatessacral.TheclassificationofcompletenessinSCIisprovidedbytheAmericanSpinalInjuryAssociation(ASIA)andisbasedonneurologicalresponses.“A”indicatesa"complete"spinalcordinjuryinwhichnomotororsensoryfunctionispreservedatS4‐S5.“B”indicatesan"incomplete"spinalcordinjurywheresensorybutnotmotorfunctionispreservedatS4‐S5.“C”indicatesan"incomplete"spinalcordinjurywheremotorfunctionisreduced.“D”indicatesan"incomplete"spinalcordinjurywheremotorfunctionismarginallyreducedbelowthelevelofinjury.“E”indicates"normal"motorandsensoryfunction[7].
Figure1.Spinalcordinjurylevelsandremainingextremityfunction.
4
2.2 ConventionalRehabilitationwithPassiveOrthosesPassiveorthosesareoftenstronglyrecommendedbydoctorsforSCIpatients.Here,twoofthemostcommonlyusedpassiveorthosesareintroduced,withafocusontheirbenefitsandlimitations.
2.2.1Knee‐Ankle‐Foot‐Orthoses(KAFO)
Themostbasictypeofpassivelowerlimborthosesareknee‐ankle‐foot‐orthoses(KAFO),whicharealsoknownaslonglegbraces.KAFOaresomeoftheearliestandstillmostcommonassistivedevicesdesignedtorestorebipedalmobility.Duetotheirmultiplehealthbenefitsandrelativelylowercosts,KAFOarecommonlyprescribedtoSCIpatientsandhighlyrecommendedbydoctors[8].AsshowninFig.2(a),thesebracesarerigidlyfastenedtotheuser’slegstolockboththekneeandanklejointsagainstkneeflexion,ankledorsiflexion,andplantarflexion.Theusercanachieveaquasi‐bipedalgaitwiththeseconstrainedjointdegreesoffreedomandtheassistanceofcrutches,walkers,orparallelbars.However,swingingoneleginfrontoftheotherwhilewearingKAFOrequiresagreatdealofupperbodystrengthandabdominalcontrol.
2.2.2ReciprocatingGaitOrthoses(RGO)
Reciprocatinggaitorthoses(RGO)areanadvancedvariationofKAFO,withatorsounitextendedtotheuser’supperbody(Fig.2(b)).RGOhaveametallinkmountedatthebackofthetorsounitthatactsasatorque‐transferringunit,whichreciprocallycouplesthemotionofthetwohipjoints.Thispassivetransmissionsystemprovidesthecoupledhipmotion–i.e.,whenonehipextends,theenergyisusedtodrivetheotherhiptoflex,andviceversa.Withthismechanism,alongerstridelengthcanbeattainedwithreducedenergyconsumption,ascomparedtothemanuallegswingwithKAFO[9‐11].However,studieshaveshownthatambulationwiththeassistanceofRGOisstillenergeticallyinefficient,requiringapproximately14timesmoreworkincomparisontonormalambulation[12,13].Itappearsthatuserfatigueandslowambulationcontributetoalowlong‐termadoptionrate,despiteahighlevelofphysicalhealthimprovements.Sykesetal.showedthatonly29percentofRGOuserswerestillusingthedevicesafteranaverageof5.4years[14].Theshortcomingsofpassiverehabilitationdevicesoftenovershadowtheirbenefits,preventingalong‐termadoptionrate.
5
(a)(b)
Figure2.Mostcommonpassivelegorthoses.(a)Longlegbraces(KAFO),(b)RGO.
2.3 PoweredMedicalExoskeletonsEventhoughpassiveorthosesprovidehealthbenefitswithrelativelylowcosts,thestate‐of‐the‐arttechnologydoesnotprovidealevelofperformanceandusabilitythatenablestheirwidespreaduseandlong‐termadoptionamongparalyzedpatients.Asanalternative,researchgroupshaveinvestigatedpoweredmedicalexoskeletontechnology,andsomeofthesedeviceshavemadesignificantadvancementsthatareemerginginthecommercialmarket.
2.3.1PoweredMedicalExoskeletonsontheMarket
Fig.3showsthefourmostwell‐knownpoweredexoskeletonsonthemarket:RexbyRexBionicsinNewZealand,ReWalkTMbyArgoMedicalTechnologiesinIsrael,Indego®byIndegogroup,whichisaVanderbiltUniversityspin‐off,andEksoTMbyEksoBionicsinCalifornia.Morespecifically,EksoTMevolvedfrommilitaryexoskeletondevelopmentoftheBerkeleyLowerExtremityExoskeleton(BLEEX),andHumanUniversalLoadCarrier(HULC)[15‐17]intheBerkeleyRoboticsandHumanEngineeringLaboratoryattheUniversityofCaliforniaatBerkeley.Theaforementionedpoweredmedicalexoskeletonsaredesignedforrehabilitationandwalkingoutsidetheclinic.Theyhave4to10poweredelectricjointsthataligntoapatient’sbiologicaljoints.Thesepoweredjointsarepre‐programmedtomimicthehumangaitandambulatethewearerinamannersimilartothatofnaturalwalkingastheusercommands.ThebiggestdifferencesamongtheseexoskeletonslieintheiractuationsystemsandUIs.ReWalkTMcontainspoweredkneesandhips,andexecutesstepsbysensingthetiltoftheuser’storso[18].Indego®operatesinasimilarfashion,byutilizinghipandkneeactuation.Ittakesstepsbymeasuringthewearer’scenterofpressurechangeaccordingtohis/herupperbodymovement[19].EksoTMalsocontainspoweredkneeandhipjoints,andhasvariousoperatingmethods.Whilepushbuttonsareusedforinitial
6trainings,moreimplicitwaysofdetectinguserintentexisttoexecutesteps,usingmultiplesensorsonthepilot'sarmsandcrutches[20].Rex,ontheotherhand,requiresnobalancingaidandhasajoystick‐typeUIlocatedonthearmrest.Withatotaloftenactuatedjointstopowerthehip,knee,andanklealongboththefrontalandsagittalplanes,Rexprovidesself‐balancingofthedeviceandwearer.However,theseadditionalfeaturesalsocomewithtradeoffssuchasslowerambulation,increasedweightandpriceduetothemorecomplexactuationsystems[21,22].Table1summarizesthespecificationsoftheabovementionedpoweredexoskeletonsaswellasthoseoftheRGOforcomparison.
(a)(b)(c)(d)Figure3.Poweredmobilemedicalexoskeletonsthatareavailableonthemarket(a)ReWalkTM,(b)Indego®,(c)EksoTM,(d)Rex.
Table1.PoweredexoskeletonsandRGOcomparisonchart.
Name PoweredDegreesofFreedom Weight ApproximateCost
ReWalkTM 4(1Hip,1Knee) 45Pounds $100,000
Indego® 4(1Hip,1Knee) 27Pounds $140,000
EksoTM 4(1Hip,1Knee) 45Pounds $100,000
Rex 10(2Hip,1Knee,2Ankle) 85Pounds $150,000
RGO 0 15Pounds $10,000
2.3.2PoweredMedicalExoskeletonsinAcademicDevelopment
Severalresearchinstitutionshavedevelopedassistiveexoskeletonsforvariouspurposes.Fig.4showstwoexoskeletonsusedforambulatingparalyzedindividuals.AUSTIN(Fig.4(a))wasdevelopedintheRoboticsandHumanEngineeringLabattheUniversityofCaliforniaatBerkeley.AUSTINcontainsapairofpoweredhipactuatorswithacoupledhip‐and‐knee
7gaitgenerationmechanism.Thekneejointsaredrivenbypoweredhipjointsthroughthecoupledgaitgenerationmechanism,whichisoperatedbyabutton‐integratedUIonawalkerhandle.Thisisthefirstexoskeletonsystemequippedwithonlytwopoweredjointsthathasambulatedaparaplegicpatient[23,24].AUSTINexecutesstepswhenthewearercommandseachstepviabuttonsonthewalkergrip.Mina(Fig.4(b))wasdevelopedbytheFloridaInstituteforHumanandMachineCognition.Minacontainsfouridenticalactuatorsforthehipandkneejoints.Theseactuatorsarepre‐programmedwithgaittrajectories,whichareobtainedbyanable‐bodiedpersonwalkswhilewearingMina.AtMina’scurrentdevelopmentstage,stepsarecommandedbyacontroloperator(practitioner),whilethewearerprovidesverbalandgesturalcues.ThepractitionerusesaMusicalInstrumentsDigitalInterface(MIDI)tocontroleachstep,andthecontrollercantriggerasinglesteporcontinuoussteps,aswellasadjustthewalkingspeeds[25].
(a)(b)Figure4.(a)AUSTINExoskeletonfromUCBerkeley(b)MinaExoskeletonfromtheFloridaInstitute
forHumanandMachineCognition
2.4OverviewofHumanWalkingInthissection,abriefintroductionofhumanwalkingiscoveredtoaidtheunderstandingofthefollowingchapters.Morespecifically,wediscussbodyplanesandjointangledefinitions,thehumangaitcycle,clinicalgaitanalysis,andpowerconsumptionduringthegaitcycle.
2.4.1BodyPlanes
AsshowninFig.5(a),threereferenceplanesareusedtodescribevariousjointmotions.Thecoronalplane,alsoknownasthefrontalplane,verticallyslicesthroughthebodywidth‐wise,andpassesthroughboththearmsandlegs.Thetransverseplanehorizontallyslicesthroughthebodyparalleltothegroundplane.Finally,thesagittalplaneisaverticalplanethatorthogonallyslicesthecoronalplane,andrunsthroughfromthefronttothebackofthebody.
82.4.2LowerExtremityJointAngles
Flexionandextensionaretermscommonlyusedtodescribejointmotion.Theydescribewhetheraparticularrelativejointangleincreasesordecreasesinthesagittalplane.ThesestandardsareillustratedinFig.5(b)forthekneeandhip.
(a)(b)
Figure5.(a)Referenceplanesofthebody[26].(b)Jointangledefinitions.
2.4.3GaitCycle
Humanwalkingisacyclicalprocess.Therearetwofundamentalstatesthateachlegcanhave:stanceandswing.Inthestancephase,thefootisplantedontheground.Duringtheswingphase,thelegswingsforwardwiththefootoffoftheground.Thefirstpartoftheswingischaracterizedbyapositiveaccelerationofthehipflexion(forward),andflexingofthekneebackwardtoallowtheswingfoottocleartheground.Thelattersegmentoftheswinginvolvesnegativeaccelerationofthehipasthekneeextendstoprepareforheel‐strike,whichiswhenthelegagainentersthestancephase.ThiswalkingcycleisillustratedinFig.6.Throughouttheprocess,thetwolegsalternaterolesbetweenstanceandswing.Whenwalkingatslowerspeeds,thereareperiodswhenbothlegsareinstance.Theseperiodsarereferredtoasdouble‐stanceordouble‐support.Therearetwodouble‐stanceperiodsthatconstituteabout10%ofthegaitcyclewhenwalkingat1meterpersecond.Thus,thepureswingperiodconstitutesabout40%ofthewalkingcycle.Asaperson’swalkingspeedincreases,double‐stancetimedecreasesuntilthereisnotimespentindouble‐stance.Runninginvolvesperiodsofflightwherebothlegsareinswing.TherelationshipbetweentheswingandstanceperiodsatdifferentspeedsofambulationisshowninFig.7.Foreachcondition,thebargraphbeginsattheinitialcontactoftheleftleg
9andrepresentstwocompletegaitcycles.Whitebarsindicatetimespentinswing,andbluebarsindicatetimespentinstance[27].
2.4.4ClinicalGaitAnalysis(CGA)
Thejointangledatathroughoutthegaitcycleareavailablefrommanybiomechanicslaboratories.Typically,avideoiscapturedasaparticipantwalks,andlegjointmotionsinthesagittalplaneareextractedusingamotioncapturesystem.ThisisreferredtoasClinicalGaitAnalysis(CGA)data.Fig.8showsCGAdataforkneeandhipjointanglesinthesagittalplanethroughoutthegaitcycle,ascollectedbyWinter,Kirtley,andLinskell[28‐30].
Figure6.Illustrationofanormalwalkingcycle[26].
Figure7.Variationsintheswingandstanceperiodsaswalkingspeedincreases.Imageadaptedfrom
[27].
10
(a)(b)
Figure8.Winter,Kirtley,andLinskellClinicalGaitAnalysisdataatthe(a)knee,(b)hip.
2.4.5PowerConsumedinaJointduringWalking
D.A.Winterandotherresearchershaveinvestigatedpowerconsumptioninjointsduringwalking.Ajoint’sangularvelocityiscalculatedfromjointangledatarecordedovertime,whileajoint’smomentisderivedfrominversedynamics.Jointpoweristhencomputedastheinnerproductofthevelocityandmomentvectors.
Fig.9showsthepoweroftheankle,kneeandhipjointsduringlevelwalking.Jointpowercanbeanegativeorpositivequantity.Powerispositivewhenthejoint’smomentofforceanddirectionofmovement(rotation)areinthesamedirection,meaningtheenergyisgeneratedbyaconcentricactioninmusclescrossingatthatjoint.Incontrast,powerisnegativewhenthejoint’smomentanddirectionofmovementareinoppositedirections.Inthiscase,energyisabsorbedbyaneccentricmuscleactionand/orelongationofothersofttissuescrossingthatpoint.Becausethekneejointisusedformotiondamping,kneepowerisrelativelylowandoftennegative.Meanwhile,asignificantportionofthegaitcyclecomprisesahighlypositivehippowerthatprovidesthepropulsiveforce[31,32].
11
Figure9.Kneeandhipjointpowersduringlevelwalking.Powersarenormalizedbybodyweight.Imageadaptedfrom[31].
12
3.LowProfileExoskeleton
Inadditiontoaddressingthemechatronicconstraints,thedesignmustaccountforacceptanceamonguserswhendevelopingassistiveexoskeletons.Hereareafewquotesfromtheparaplegicpatientswhohaveparticipatedinusabilitytesting:
“Ihadtocrawltomyseatintheairplanebecausemywheelchairdidn’tfitthenarrowaisle.”
“Iwouldn’tweartheexoonthestreetbecauseIdon’twanttolooklikeIamwearingarobot.”
Fromthefeedbackabove,itisclearthattheformfactorofexoskeletonsplaysahugeroleindeviceusability.Thebulkinessofthedevicenotonlyrestrictsauser’sabilitytonavigatenarrowpassages,butalsohis/herabilitytoblendinwithacrowdofablebodiedpeople.Onlylowprofileexoskeletonscanprovideindependencethroughoutthedaybyallowingthewearertopassthroughnarrowaisles,useregularcomputerchairswitharmrests,andbemobileindailylifewithoutassistance.Besidesprovidingindependence,exoskeletonswithsmallerformfactorsallowpatientstowearthedeviceundertheirclothes–respectingone’sformofpersonalexpression.Moreover,minimalhardwareoftheexoskeletonreducesitsweight,whichcanincreasetheefficiencyoftheoverallsystemasamobiledevice.
AsdiscussedinChapter1,humanhipandanklejointsprovidepositivepropulsivepower,whereaskneejointsareusedtodampenthemotionwithnegativepower.However,inbelow‐kneeamputeegaits,itwasobservedthatpropulsionfroma(prosthetic)ankleisverysmall,andcompensatedbypropulsiveenergyfromhipextension[33].Fromthisgaitdata,kneepowercurvesareveryclosetozeroinbelow‐kneeamputeegaits,despitethepresenceofkneemuscles.Also,studieshavedemonstratedthattheamplitudesofmuscleactivationaroundtheanklejointscarcelychangewhenwalkingwitheitheramechanicalorpoweredankle–footorthoses(AFO)[34].Itappearsthatahighlevelofbipedalmobilityisattainableusinghipjointsasthemainsourceofpropulsivepower.Therefore,toachievealowprofile,theexoskeletondesigninthischapterusespoweredelectrichipactuators,andlacksanypoweredactuatorsatthekneeandanklejoints.Thislowprofileexoskeletonhasasimilarambulationstrategytothatoftheamputeegait,andmaximizestheuseofpropulsivepowerfromthehipjoints.
Astheonlypoweredunitforpropulsion,hipactuatorsarethesignificantweightandsizefactorsforthedevice.Therefore,todeveloplowprofileexoskeletons,itiscriticaltoachievealightweight,slimformfactorforthehipactuationunits.Thischapterdiscussesthedesignofminimallyprofiledhipactuationunitsthatpermitambulationofparalyzedindividuals.Theactuationunitsareusedwithpassivekneejointsina“minimally‐actuatedmedicalexoskeleton”.Withlowprofilehipactuators,passivekneejoints,andstandardankle‐foot‐orthoses,thisexoskeletonbecamethefirstpoweredexoskeletonthatweighslessthan20pounds.Furthermore,itscompactformfactormorecloselyresemblesanRGOthanotherpoweredexoskeletons.
13
3.1HipActuationUnitsHardwareDesignOverviewThelowprofilehipactuationunitsdiscussedinthischaptercompriseabrushlessDC(BLDC)motor,transmissionsystem,sensors,load‐bearinginterfacingcomponents,andenclosurecomponents.BarecomponentsofaBLDCmotorandstrainwavegearsystemsareintegratedintoapancakeactuationunitwithcustom‐madeinterfacingparts,andpairsofoff‐the‐shelfradialandangularballbearings.Asasingleintegratedunitwithoutanyredundantbearingstructures,thesystem’sloadratingandformfactorareoptimizedandthisminimizesthesystem’ssizeandweight.TheoveralldesignprocedureisillustratedinFig.10.DesigndecisionsduringeachprocessandabriefoverviewofDCmotordynamicsandstrainwavegearingmechanicsarecoveredinthissection.
Figure10.Lowprofileactuationunitdesignprocessflowchart.
143.1.1BrushlessDCMotorandStrainWaveGear
Therequiredperformancecurvefora200‐poundparaplegicperson’swalkingwasderivedfromCGAdata,amputeegaitdata,andempiricalanalysis.Assumingthatanominal50voltsareappliedusinglithium‐ionbatteries,aBLDCmotorandstrainwavegearwereselectedtofulfilltheperformancecriteria.
Outofalltheoff‐the‐shelfBLDCmotorsthatmeettherequiredspecifications,theEmoteqHT03000‐J01‐Zframelesselectricmotorwasselectedduetoitsrelativelysmallsize.ThisBLDCmotorconsistsofa3‐phase,27‐slotstator,anda12‐polerotor(Fig.11.(a),(b)).Meanwhile,theHarmonicDriveTM(CSD‐25‐160‐2A‐GR‐SP)160‐to‐1unhousedstrainwavegearcomponents(Fig.11.(c))werechosenforthetransmissionsystem.AsillustratedinFig.12,theactuatordynamicslineencompassestherequiredperformancecurveintheplot,whichimpliesthattheselectedcomponentssatisfythespeedandtorquerequirementsofa200‐poundperson’swalking.TechnicalspecificationsofthesecomponentsfrommanufacturersaresummarizedinTable2.Thefundamentalsofmotordynamicsaredetailedbelowtoaidunderstandingofthisplot.
(a)(b)(c)
Figure11.TheBLDCmotorandstrainwavegearusedfortheactuatordesign.(a)12‐polerotormagnet.(b)3‐phase,27‐slotstator.(c)Strainwavegearcomponents[35].
Figure12.Actuatorperformancelinesatisfyingtherequirementforwalking.
15
Table2.MotorandTransmissionCharacteristics.
R(Ω) KT(Nm/A)
KB(V/rad/s)
KM(Nm/√ )
TPR(oC/W)
GearRatio
6 0.2 0.2 0.082 2.89 160
3.1.1.1DCMotordynamics ThedynamicsofaDCmotoraretypicallycharacterizedbythefirstorderdifferentialequation:
V , (1)
whereVisthevoltageacrossamotorwinding,Listhemotorinductance,Ristheresistance, isthebackelectromotiveforceconstant("backEMF"),and istherotationalvelocityofthemotorshaft.Excludingtheeffectofinductanceatsteady‐state,thevoltageequation(1)becomes V . (2)TheBLDCmotorispoweredbyacurrentcontrolamplifierwithaninternalfeedbacklooptotrackacommandedcurrentwithinanacceptablerange.Thereforeitisassumedthatthemotorcurrentcanbedirectlycontrolled.Themotortorquegeneratedisafunctionofthesuppliedcurrent T , (3) whereTisthetorqueandKTisthetorqueconstant,whichisamotor‐specificparameteralongwithK .Substitutingfor inequation(2)fromequation(3)resultsinequation(4)
,(4)
where: .
isreferredtoasthemotorconstant. istheslopeofthemotorwindinglineinFig.13,whichisamotor‐specificparameter.Theinterceptionofthelinewiththehorizontalaxisiscalledtheno‐load‐speed,whichisthespeedwhenthereiszerotorqueonthemotor.Theinterceptofthelinewiththeverticalaxisisthepeaktorque,whichisthetorquethatcanstopthemotorshaftfromspinning.
Themotorwindinglinethatconnectstheno‐load‐speedandthepeaktorquedefinestheboundariesofthemotorperformance.Notethatforagivenmotor,theslopeoftheline
16
doesnotchangewhendifferentvoltagesareapplied.Thisisbecause isafunctionofmotorvolume,andstatesthemotor’sabilitytodissipateheatforagivenamountoftorque.Ahigher representsabetterabilitytodissipateheat.Foragivenmotor,themotorperformancelinetranslatesaccordingtodifferentvoltages,whilemaintainingthesameslope.Theslopeofthislinechangeswhenagearreductionisappliedtothemotor.Motorperformancecanbeimprovedbyapplyingahighervoltagewithinitsheatdissipationability.Toensureaproperlydesignedactuationunit,thedesiredsystemperformance‐torqueversusangularvelocityprofile‐shouldbeencompassedunderthemotorperformanceline(Fig.13).
Figure13.Motorperformanceplotoftorqueversusangularvelocity.
Ontheperformancelimit,excessiveheatmaycausedamagetothemotor’sinternalcomponents.Toaccountforthis,powerdissipation(W)inthemotormustbecalculatedas
W
1. (5)
Acompleteformofequation(5)canbewrittenas
W
1, (6)
where isthedampingthroughthemotor.Theaveragepowerdissipationoverthetimeinterval∆tis
17
W1 1∆
∆ 1∆
∆
, (7)
W1
, (8)
where istherootmeansquared(RMS)torqueand istheRMSspeed.Continuouspowerlimitscanbeobtainedwithasimplerelationship:
∆, (9)
where∆TisanadmissibletemperatureincreaseandTPRisthetemperatureriseperwatt.
Inthislowprofilehipactuationunit,theselectedBLDCmotorallowsapeaktorqueof266.7Nmandano‐load‐speedof1.4rad/secwitha160:1gearreduction.However,whenoperatingthemotor,maximumcontinuousstalltorqueofthemotorshouldalsobeconsideredaccordingtotheheatdissipationlimit.Maximumcontinuousstalltorqueisdefinedas:
Maximumcontinuous stall torque
∆∙ .
(10)
TPRisgivenbyamotor’sspecificationsand∆Tisoftendecidedbytheinsulationratingofthewire.IntheEmoteqHT03000‐J01‐Z,atemperatureriseto125 isadmissibleforanRMSof1.8amperes.Thisresultsinamaximumcontinuousstalltorqueof82.07Nm.Inadditiontothemaximumcontinuousstalltorque,themomentarymaximumtorquepermittedtothetransmissionsystemshouldalsobeconsidered.ThemaximummomentarypeaktorqueoftheHarmonicDriveCSD‐25‐160‐2A‐GR‐SPis152Nm,whichislimitedbytheflexspline[35].Furtherinformationonthestrainwavegearsystemisdiscussedbelow.
3.1.1.2StrainWaveGearingSystemAstrainwavegear,theHarmonicDriveCSD‐25‐160‐2A‐GR‐SP,wasusedinthelowprofilehipactuationunit.Themainreasonsforchoosingastrainwavegearingsystemareitscapabilityofahighgearreductionratioinasmallvolumeanditslightweight.Agearreductionratioupto320:1ispossiblewithstrainwavegears,whileplanterygearscanonlyproducea10:1gearratiointhesamespace.Othertypesofgearssuchaswormgearswereexcludedfromconsiderationduetotheirlowefficiencyandlimitedgearratioinaconfinedspace.
Astrainwavegearcomprisesthreecomponents:awavegenerator,aflexspline,andacircularspline.Thewavegeneratorisathin‐racedballbearingfittedontoanellipticalplug,whichservesasatorqueconvertor.Theflexsplinehasanon‐rigid,shallowcupstructurewithexternalteeth.Thesidesoftheflexsplineareverythin,butthebottomisthickerand
18rigid.Thisresultsinwallflexibility.Theflexsplinefitstightlyoverthewavegenerator,andisheldinanellipticalshape.Thecircularsplineisarigidringwithteethontheinnerside.Theseteethengagewiththeteethontheoutersideoftheflexsplineacrossthemajoraxisofthewavegenerator.
Figure14.Strainwavegearcomponents[35].
Whiletherearemultipleconfigurationsofstrainwavegearstructures,thefollowingconfigurationwasusedinthelowprofileactuationunitdiscussedinthischapter:thewavegeneratorisconnectedtotherotor,whilethecircularsplineisrigidlyattachedtotheenclosingstructure.Therotationoftheflexsplineisusedastheoutputrotationoftheactuationunit.ThemechanismofthisconfigurationisillustratedinFig.15.
Asthewavegeneratorrotates,theflexsplineteethengagementwiththecircularsplineteethchange;i.e.,themajoraxisoftheflexsplinerotates.Becausetherearefewerteethontheflexsplinethanthecircularspline,everyfullrotationofthewavegeneratorresultsinaslightbackwardrotationoftheflexsplinerelativetothecircularspline.Thus,therotationspeedoftheflexsplineismuchslowerthanthatofthewavegenerator,whichcreatesahighgearreduction.Thegearreductionratioiscalculatedby:
GearReductionRatio
.
Forexample,ifthereare202teethonthecircularsplineand200teethontheflexspline,thegearreductionratiois‐0.01.Thismeanstheflexsplinespinsat1/100ofthespeedofthewavegeneratorintheoppositedirection.
Inadditiontothestrainwavegears’lightweight,compactdesign,andhighgearreductionratio,otheradvantagesincludenobacklash,highefficiency,back‐drivability,goodresolution,andhighrepeatability.Meanwhile,disadvantagesincludehighcostandflexibilityintheirmechanicalstructures,whichcanresultinpotentialdegradationsuchastheratchetingphenomenonfromenvironmentalandmechanicalshocks[36,37].Theratchetingphenomenoncanoccurwhenanexcessivetorque(atorabovetheratcheting
19torque)isappliedwhilethegearisinmotion.Theteethbetweenthecircularsplineandflexsplinemaynotengageproperlyandlosetheirconcentricity(Fig.16).Thismaydamagetheflexsplineandloweritslifespan,aswellastheratchetingtorque.Ratchetingtorqueisacriticalfactorthatmustbeconsideredwhendesigninganactuationunitwithastrainwavegear.
Figure15.Strainwavegearoperation.
Figure16.Ratchetingphenomenon[35].
203.1.2ShaftDesignandSensorMounting
Therearetwoshaftsinthisactuationsystem,the“innershaft”andthe“outershaft”.Theoutershaft,illustratedinblueinFig.17(a),transmitstorqueandrotationfromtherotor(coloredinyellow)tothestrainwavegearinput,whichisthewavegenerator(coloredinlightgreen).Fourscrewsclamptheoutershaft;thescrewscrossthroughthewavegenerator,outershaft,androtorwiththreadedholesontherotor’ssteelbody.Rotationissupportedbytworadialbearingslocatedontheouterandinnersideoftheoutershaft(coloredinblack).Theinnershaft,illustratedinorange,isusedtoencodetherotationalangledataoftheactuationunit’sfinaloutput(i.e.,thehipjointangle).Amagnet(coloredinred)fortheabsoluterotarymagneticencoderispressedonthetipofinnershaft.ThismagnetprovidesangularrotationdatatoAustriaMicrosystem’sAS5145B12‐bitencoder,whilerotatingwiththeenclosingcomponent(coloredinpink)andtheflexspline(coloredindarkgreen)‐theoutputoftheactuationunit.Theinnershaftlocatesitselfinsidetheoutershaft,andbecomesasingleentitywiththeoutputoftheentireactuationunitwhenfullyassembled.Locatingtheinnershaftinsidetheoutershafthastwoadvantages:
1) Itreducestheoverallthicknessoftheactuationsystembyutilizingthespaceinsidethehollowoutershaftfortheencoderversusaddinganexternalencodermountingstructure.
2) ItallowsthesystemtoefficientlyroutewiresbyplacingthewiresfromtheabsoluteencoderandBLDCmotorclosetothewearer’slowerback,wherethecontrollercircuitboardandbatteriesarelocated.Fig.17(b)showsacompleteassemblyofalowprofilehipactuationunitinstalledontheexoskeleton.Withthisconfiguration,theencoderandmotorconnecttothecontrollercircuitboardandbatterieswithoutexposinganywires.
3.1.3BearingstoHoldExternalLoads
Inadditiontoapairofradialballbearingsthattaketheinternalloads,apairofangularcontactballbearingswasusedtowithstandexternalloads.Here,weassumethattheworst‐caseexternalloadingscenariooccurswhenthewearerleanstothesidewithhis/herbodyweightinthefrontalplane(Fig18).Inthiscase,calculationofthemomentappliedtotheactuationunitisrathersimple.Westartbyassumingthatthewearer’sweightis200poundsandthedistancebetweenthemid‐planesofthehipactuatorsoneachside(paralleltothesagittalplane)is20inches.Inthisloadingscenario,weneglecttheaxialloadappliedtothesystem.Therefore,themomentappliedtothesystempurelyresultsfromtheradialloadfromthewearer’supperbody.Knowingthatthewearer’supperbodyweightis55%ofthewholebodyweight[38]andthemomentarmlengthis10inches,wecalculatethemomentappliedtothebearingstobe1,100lb‐in,orabout125Nm.Apairofangularcontactbearings,KaydonBearingKA035AR0,waschosenandinstalledwithaback‐to‐backorO‐configurationsuchthattheactuationunitssafelysustainthesystemunderthisloadingandpreventstructuralfailure,whichcanbeinjurioustothewearer.Theseselectedbearingsareabletotake552lbofdynamicradialloadand1,350lbofdynamicaxialload.
21
(a)(b)
Figure17.(a)Cross‐sectionalviewofanactuationunit.(b)Apatientwearingalowprofileexoskeletonsystem.
Figure18.Loaddiagramforanactuationunitwithapairofangularbearingsinstalledback‐to‐back.
22
3.2AssemblyProcessandDesignConsiderationforSafetyWithconsiderationsforachievingalowprofilesystemanddesignforassembly[39,40],atotalofsevencustom‐madecomponentsweredesigned.Withsystemsthatrequireassemblyprocesses,itisimportanttoconsidereaseofassemblyearlyinthedesignstage.Insection3.2.1,therecommendedassemblyprocessofthelowprofilehipactuationunitsisdiscussed.Section3.2.2coversthedesignconsiderationstakenintoaccountforsafety.
3.2.1AssemblyofaHipActuationUnit
Therearethreeenclosingcomponents(EC)andtwointerfacingcomponents(IC)inthehipactuationunit.Thesefivecustom‐madepartswerefabricatedoutofeitherAluminum7075‐T6orStainlessSteel17‐4PH.Cross‐sectionandsideviewsareshowninFig.20,whileanexplodedviewisshowninFig.21.Designforassemblywasconsideredfromanearlystageinthedesign,andtherecommendedassemblyprocessisasfollows:
1) Installtheradialbearing(large)ontheoutershaft(pressfit),andslideIC#2alongtheouterraceoftheradialbearing.
2) InstallthecircularsplineontheIC#1.3) SlideIC#2fromstep1)overIC#1.4) InstalloneangularbearingonEC#2,andtheotherangularbearingonIC#1;make
suretheorientationofangularbearingsformsaback‐to‐backconfiguration.5) Combinetheproductsofstep3)and4).6) Installtheradialbearing(small)ontheinnerraceoftheoutershaft.7) PressinthestatoronIC#2;makesurethewirealignmentiscorrect.8) Assemblethewavegenerator,outershaft(assembledwithIC#2andtheradial
bearing),androtor.9) Presstheencodermagnetontotheinnershaft.10) Slidetheproductofstep9)throughtheinnerraceoftheradialbearing,whichwas
installedinstep6).11) Assemblethebearingretainertotherotor.12) InstalltheabsoluteencodermoduleontheEC#3.13) AssembleEC#3andIC#2.14) Preloadtheangularbearing.15) AssembleEC#1andtheflexspline.16) Assembletheproductof15)withtheinnershaftandEC#2.
Whenassembled,EC#3isconnectedtothetorsoadaptorwithsafetystops,whichisexplainedinsection3.2.2.
23
Figure19.Anassembledlowprofilehipactuationunit.
24
Figure2
0.(a)T
hesid
ecross‐sectionalview
ofalowprofileh
ipactu
ationunit.(b
)Sideview
ofalowprofileh
ipactu
ationunit
assembly.
25
Figure2
1.Exp
lodedview
oflowprofileh
ipactu
ationunit.
26
3.2.2DesignConsiderationsforUserSafety
Multiplemotortest‐runswereconductedwithoutahumantestpilotbeforeexoskeletonsystemintegration.Tosimulatenaturalhumanwalkingandverifythehipactuationunit’sperformance,anexperimentalsetupwasarrangedasshowninFig.22.Compensatoryweightswereattachedtothethighandshanklinkstomatcha200‐poundperson’slegweight.Althoughtheexoskeletonwasintendedtohavepassivekneejoints,ahipactuatorwasusedtosimulatekneejointmovementinthissetuptofacilitatetheexperiment,whilethepassivekneejointtechnologywasunderdevelopment.Thistestprovidedempiricalvalidationthatthehipactuationunitcouldprovidesufficientspeedandtorqueduringtheswingphaseofwalking.
Asidefromtest‐runningthehipactuatorsundercomparableloads,mechanicalhardstopsweredesignedintoatorsoadaptorunit.Thisistoensurethehipactuator’srangeofmotionislimitedtoanatural,biologicalrangeofmotioninthesagittalplane(Fig.23).However,unexpectedcollisionswiththehardstopscancreatehighmomentarytorquestothehipactuator,whichcanpotentiallycauseratchetingofthestrainwavegear.Therefore,inadditiontomechanicalhardstops,itisimportanttohaveelectricsafetystopsinthecontrollertolimitthecurrentprovidedtotheactuator.
Figure22.Experimentalsetupfortestingthehipactuationunit.
27
Figure23.Mechanicalhardstopsforhipextensionandflexion.
3.3IntegrationintoaMinimally‐ActuatedExoskeletonApairoflowprofilehipactuationunitswasintegratedintoaminimally‐actuatedmedicalexoskeleton(Fig.24).Thisexoskeletonalsoconsistsofpassivekneejointsandacarbonfibertorsoframe,whichcontainsacontrollercircuitboardandbatteries.Also,asetofstandardlegbracescustom‐moldedforeachpatientwasused.Detailsofthepassivekneejointdesigncanbefoundin[41].
Thisdevicewastestedbyatestpilotwhowasa28‐year‐oldmale,5’10”talland170pounds,whohadsustainedacompleteT12injuryfor12years.Whenwornbythetestpilot,theactuatorswereabletoprovideaspeedofupto1secondperstep,withvaryingstepsizesfrom14to16inches.Theoverallwalkingspeedoftheexoskeletonwearervariedsignificantlyaccordingtothebalancingtimebetweeneachstep,whichappearedtoberelatedtothewearer’sphysicalconditions.Inthisstudy,thetestpilotwasabletowalkataspeedof0.22m/s.Thetorqueprovidedfromtheactuatorsprovedtobesufficientforwalking,aswellasforprovidingmoderateassistanceduringsittingandstanding(withthehelpofbalancingaids).
Thegeometryoftheactuatorswasofasufficientlylowprofiletoallowthewearertositinaregularcomputerchairwitharmrests,partiallyconcealthedeviceunderhisclothing,andwalkthroughnarrowaisles(Fig.25).Thetotalweightoftheactuatoris3.5pounds,whichenablestheweightofthewholeexoskeletonsystemtobe19.5pounds.Thisexoskeletonisthelightestpoweredexoskeleton,weighinglessthan20pounds,withacompactformfactorthatmorecloselyresemblesanRGOthanotherpoweredexoskeletons.
28
Figure24.Lowprofile,minimally‐actuatedmedicalexoskeleton.
Figure25.Withlowprofileactuators,thetestpilotwasabletowalkthroughthenarrowaisleofa
localstore.Thismedicalexoskeletoniswornpartiallyunderneathhisclothing.
29
3.4LowProfileHipActuatorDesign–SecondRevisionLowprofilehipactuationunitswiththeEmoteqHT03000‐J01‐ZandHarmonicDriveTMCSD‐25‐160‐2A‐GR‐SPallowacompactprofile,lightweight,andhighusermaneuverability.Infact,giventheirspecifications,theactuationunitsdiscussedinthischapterarethemostcompactunitscomparedtootheractuationunitsonthemarket.Meanwhile,thedesign’sdisadvantagesincludeitscomplicatedassemblyprocessandhighcostduetomanycustom‐madecomponents.Toimproveupontheseshortcomings,anothersetofhipactuatorswasdeveloped(Fig.26).Theseactuatorsuseoff‐the‐shelfhousedcomponents:aMaxonEC90FlatmotorandHarmonicDriveTMSHD‐20‐100‐2SH.Asaresult,theactuatorcontainsfewercustom‐madeparts,andhasaneasierassemblyanddisassemblyprocess.Asatradeoff,adoptingtheseactuationunitsresultedinanincreaseintheexoskeletonhipwidthofabout1.9inchesandadecreaseduserweightlimitduetothestrainwavegear’sbearings.TechnicalspecificationsaresummarizedinTable3.ThecharacteristicsoftheMaxonEC90FlatmotorarecomparabletotheEmoteqHT03000‐J01‐Z,andthegearratiowasselectedtobe100:1insteadof160:1,toachieveamoreoptimaldynamicswingofthepassivekneejoints.Although,neitheractuationunitrequiresanymodification,exceptforthedisassemblyandre‐assemblyprocessforinterchangingthestrainwavegearsbetween100:1and160:1.
Whilethedesignusesanoff‐the‐shelfBLDCmotorandstrainwavegearthathavetheshaftandbearingsintegrated,somemodificationsontheBLDCmotorstructureweremadetooptimizeinterfacingwithminimalthickness.Therecommendedassemblyandmodificationprocessofthisactuationunitisdescribedbelow.
1) MillouttheholepatternandshoulderonthemetalinterfaceoftheMaxonEC90flatmotorforthestrainwavegear.
2) CuttheshaftlengthoftheMaxonEC90to0.592in.3) PressIC#2ontotheshaftoftheMaxonEC90.4) AssembleIC#1andthewavegeneratorofthestrainwavegear.5) Aligntheflexsplineandcircularsplineassemblywiththehole‐patternfrom1).6) InstallIC#3ontheproductofstep5).7) Slidetheproductofstep4)ontotheproductofstep3),andfastenthescrews.8) PressthemagnetontotheOutputLink.9) Installtheproductofstep8)ontotheproductofstep5)10) InstalltheabsolutemagneticencoderonEC#1.11) Installtheproductofstep10)ontoIC#3.
AcomparisonofthetwohipactuationunitsissummarizedinTable4.Futureworkremainstofurtherdevelopcost‐effectiveandlowprofileactuationunitsthatcanbeusedbyuserswhoweighmorethan170pounds.
Table3.MaxonEC90FlatMotorCharacteristics
R(Ω) KT(Nm/A)
KB(V/rad/s)
KM(Nm/√ )
TPR(oC/W)
2.26 0.217 0.217 0.144 2.99
30
Figure26.Cross‐sectionalviewofanactuationunitwiththeMaxonEC90.
Table4.ComparisonofTwoMotorDesigns
DCMotor Emoteq,Frameless MaxonEC90,Housed
StrainWaveGear UnhousedComponents Housed,AssembledGearset
Assembly Complicated Simple
Diameter 4.5” 3.8”
Thickness 1.8” 2.5”+cover(notshown,0.25”)
Weight 3.5lb 3lb
UserWeightLimit 200lb 170lb
31
4.ModularExoskeleton
Targetusers’specificconditionsandneedsmustbethouroughlyconsideredtodevelopaneffective,widelyacceptedmobileexoskeletonsystem.Amongparalyzedindividuals,onecrucialuserneedthatvariessignificantlyistheamountofsupportrequiredfromtheexoskeletonforbalancing.Evenwithinaparaplegicpatientgroup,thereisabroadrangeofpatientabilitiesforsustainingupperbodycontrolafterinjury.Whileagivenamountofsupportfromtheexoskeletonmightnotsufficeforcertainpatients(inwhichcasethepatientmaynotbeabletousethedevice),thesameamountofsupportcouldbeexcessiveandimpedingforotherpatientswhosustainmoremusclecontrol.
Humanhipjointsareoftendescribedasballjoints,whichhavedegreesoffreedominallthreebodyplanes:flexionandextensioninthesagittalplane,adductionandabductioninthefrontalplane,andexternalandinternalrotationinthetransverseplane.Whilehipextensionandflexionarethemainsourcesofbipedallocomotion,ab/adductionandinternal/externalrotationalsocontributetonumerouskindsofmaneuversandbalancing,aslongastheindividualretainsthecontrol.Beingabletoprovideanadequatelevelofsupportandfreedombasedonthewearer’sphysicalconditioniscriticalfordeviceusability.
Existingpoweredexoskeletonshavea“oneexoskeletonforall”approach,withafixedamountofsupport.Thislimitstheusergrouptothosewhoneedtheexactlevelofsupportthedeviceprovides,whichoftentimesisnotoptimalforpatients.Moreover,beingabletochangethesupportlevelsisbeneficialforbothusersandpractitioners.Thisallowspractitionerstorecommendfirsttimeuserstostartwithfullsupport,thengraduallylowerthesupportlevelformorenaturalmovementasthepatientgetsaccustomedtothedeviceanddevelopsbalancingskills.
Inlightofthis,atrunkexoskeletonmodulethatcomprisespoweredhipactuatorsandlockablejointsthatprovidedifferentsupportlevelswasdeveloped.Thisexoskeletonmodulecanbeusedeitherbyitselforwithotherkneemodules.Itsconfigurablearchitecturecreatesversatilesupportlevelsforpatientsandnon‐injuredusers.Thetrunkexoskeletonmodule,apairofkneemodules,andapairofAFOcanbecombinedtoformafullbodyexoskeleton.Thischapterdiscussesanoverviewofthehardwaredesignandpossibleconfigurationsofthedesign,whichopensassistiveopportunitiestoseveraltypesofuserswhichincludesnon‐injuredpeople.
4.1MobilitySpectrumAsbrieflyexplainedabove,itappearsthateachparalyzedpatient’ssustainingmobilitylevelvariessignificantly.Therefore,therecommendeddevice’srangeofmotionhastobefittedaccordingtoanindividual’sinjurylevelandphysicalconditions.Fig.27illustratesthe“mobilityspectrum”toaidintheunderstandingofthedifferentuserconditionsandneeds.Thecolordensityofthespectrumstandsforthemobilityoftheuser.Forexample,aparalyzedusergroupwithahighinjurylevelandlowresidualmusclecontrolfallsonthe
32leftsideofthespectrum.Patientsinthisusersegmentmayrequirealargeramountofsupportfromthedevicewhichcanconstraintheiruncontrolledmovementswithhigherdevicerigidity.Thisisachievedusinglockedjoints(exceptforthehipandkneeduringextensionandflexion),shoulderstraps,acorset,shinguards,etc.Patientswithlowerinjurylevelswhosustaingreatermusclecontrolfallontherightsideofthespectrum.Alimitedrangeofmotion,lockedjoints,andadditionalstrappingsmayimpedetheirnaturalmovements.
Figure27.Mobilityspectrumoftheinjuredusergroup.ItwasnotedthatpatientsattaininghighmobilitywhoareontheveryrightsideofthemobilityspectruminFig.27havesimilarneedsastheelderly,orthosewhoarenotseverelyinjuredyetrequireanassistiveexoskeletonsystem.Thisoverlapinneedsandpreferencesdespitebeingunderdifferentphysicalconditionsnaturallyraisesaquestiontous‐whatifwefurtherextendthismobilityspectrumtotherightside?Totherightsideofthismobilityspectrum,therewillbeindividualswithhighermobility,suchasindustryworkersandsoldiers(Fig.28).
Figure28.Extendedmobilityspectrum.Currently,technologyexistsonlyforafewdiscretesegmentsofthisspectrum.However,designmodularitycanfulfillthefullrangeofthespectrumandfillgapsindeviceavailabilitybetweensegmentsonthespectrum.
ExoskeletonsforparalyzedpatientswerediscussedinSection2.3andChapter3.Asmentionedinprevioussections,theyhavelimiteduseforpatientswithcertainconditions.
33Fig.29(a)showstheHumanUniversalLoadCarrier(HULC)[42],anexoskeletonforsoldiers.HULCevolvedfromtheUCBerkeley’sBerkeleyLowerExtremityExoskeleton(BLEEX)[15‐17],andislicensedtotheLockheedMartinCorporation.TheHULCuseshydraulicactuatorswithhighpowerdensitytohelpsoldiers’heavyloadscarrying.Fig.29(b)showsHonda’sStrideManagementAssist.Ithastwoelectrichipactuatorswithoutkneejoints,andassistsflexionduringwalking[43].Itisdesignedfor“individualswithweakenedmuscleswhoarestillabletowalk”[44],andiscurrentlyundergoingaclinicaltrialforstrokerehabattheRehabilitationInstituteofChicago.EventhoughHonda’soriginaltargetusergroupincludespeoplewithvariousphysicalconditions,thedevicedoesnotincludecustomizablefeaturesintermsofrangeofmotionandlevelofsupportfordifferentindividuals[45].
(a)(b)
Figure29.(a)HumanUniversalLoadCarrier(HULC)[42].(b)StrideManagementAssist[44].Table5showsthemajordifferencesamongexoskeletonsystemsfordifferentusergroupswhichrequirevaryingrangesofmotion,levelsofbalancingsupport,speedsofwalking,andtypesofuserinterfaces.Forexample,alimitedrangeofmotionandasubstantialamountofbalancingsupportisrequiredforaparaplegicpatientwithahighinjurylevel.Forthispatient,therecommendedwalkingspeedwouldbeslowerthanthatofasoldier.Thesuggesteduserinterfacesforthesetwousersarealsohighlydifferent.Fortheparaplegicpatient,itisrecommendedthatmorecautiousstepsbetakenwhentheuserorpractitionerexplicitlycommandsthestep,forinstance,bypressingbuttons.Incontrast,asoldierwithcompletemusclecontrolandhighagilitywouldprefertowalkwithoutexplicitstepcommandstothedevice.Amongthesefourvaryingdevicecharacteristics,threefeatures–recommendedrangeofmotion,levelofsupport,andspeed/power–arecloselyrelatedtothehipjointdesign.Therefore,thetrunkexoskeletonmodulewithlockablejoints,whichessentiallycreatesvaryinglevelsofsupportandrangesofmotion,wasdesignedtosatisfydifferentuserneedswithoutrequiringcompletelyre‐customizedsystemsforeachuser.
34
Table5.DifferentExoskeletonFeaturesforDifferentUserGroups.
*RefertoChapter5formoreinformationonexplicitandimplicitUIs.
4.2HardwareDesignOverviewTodifferentiatedevicesupportlevelsforuserswithvaryingmobility,thedeviceshouldincludeallbiologicaldegreesoffreedom,butallowuserstodisableanydegreesoffreedomatanytime.Onewayofachievingthisfunctionisthroughpoweredactuatorswhicharecontrolledbyauserorpractitioner.Mostpoweredexoskeletons,includingtheminimally‐actuatedlowprofileexoskeletondiscussedinChapter3,alreadyhavethispoweredactuationforhipextensionandflexion.However,addingadditionalpowereddegreesoffreedomincreasessize,weight,andcostofthedeviceduetoadditionalrequiredinfrastructuresuchasbatteries,sensors,andcontrollers.Sincehipadduction,abduction,androtationhavelittleinfluenceonbipedalambulation,itwasdecidedtoachievetheseadditionaldegreesoffreedomviapassive,unactuatedmechanicaljointswithlockingfeatures.Thisminimizesthesize,weight,andcostofthedevicewhileallowingdifferentdevicesupportlevels.
Ideally,alljointsthatprovidedegreesoffreedomshouldbelocatedwithinthebiologicalhipjointtoachievenaturalmovement.However,orthoticdevices(unlikeprosthetics)havegeometriclimitationswhichrequireoffsettingthedevicejointfromthebiologicaljointinthefrontalplane.Also,itisverychallengingtocreateaballjointsystemwithcomparablesizeandweightthatmimicsthehumanhipjointwithpartiallypoweredandpartiallypassivecomponents.Therefore,itwasdecidedtolocatethehipadductionandabductionjointsattwodifferentplacesontheexoskeleton–oneonthelowerbackandtheotherontheupperthighascloseaspossibletothehipjoint.Whilethelowerbackabductionjoint
35provideslimitedabductionandadduction,theadditionaljointontheupperthighisequippedwithahigherrangeofmotiontocompensateforinaccuraciesindevicemovements.Arotationaljointisalsolocatedontheupperthighinconjunctionwiththeabductionjoint(Fig.30,Fig.31).
Thesejointsareallfree,passive,mechanicaljointsthatcomplywiththewearer’smotionwithoutimpedingoramplifyingmovement.Specificdegreesoffreedomcanbeconstrainedbyinsertingpinsintopre‐designedlocationsonthedevice.
Figure30.Overviewofthehipmodulewithadditionalabduction,adduction,androtationjoints.
36
Figure31.Abduction,adduction,androtationjoints.Thistrunkexoskeletonmodule,whichisequippedwithadditionallockablejoints,canbeworneitherbyitselforincombinationwithmultipledifferentkneetechnologies,whichweredevelopedintheBerkeleyRoboticsandHumanEngineeringLabatUCBerkeley.Certaintypesofpatients(suchaspatientswhoretaincontrolintheirlowerlegsortheelderly)canbenefitfromthetrunkexoskeletonmodulebyitself,withoutaddedkneeandanklesupports.Similarly,anyable‐bodieduserwithmobilitycanreceivegaitassistanceandlowerenergyconsumption,asthesimilareffectwasachievedinpreviousmilitaryexoskeletonsystemsusinghydraulicactuators[15‐17].Thedevicecanalsoprovideassistanceoraugmentvarioususermaneuversassociatedwithindustrialwork.Theconfigurablearchitectureandprogrammablecontrollerofthesystemallowindustryworkerstowearthisexoskeletontogainsupportforrepetitive,strenuousworkmovements–forexample,liftingworkinawarehouseorassemblyworkinamanufacturingfacilitywheretheworkerneedstobendhis/herback.However,furtherdevelopmentofdifferentmodesofoperationaccordingtotheuser’spostureisrequiredfortheseapplications.
37Fig.32summarizesexampleoftheusesofthetrunkexoskeletonmodulewithadditionallockablejoints.Thedesignaccommodatesvariouscombinationsofactuationunitswithdifferentpowerlevels,andseveralkindsofkneemodules.
1. GaitAssistKneeModuleisapassivekneemoduledevelopedfor
alowprofileexoskeleton,discussedinChapter3.2. SquatAssistKneeModuleisapassivekneemoduledevelopedin
theBerkeleyRoboticsandHumanEngineeringLab.Figure32.Exampleusesofthetrunkexoskeletonmodule,byitselfandincombinationwithtwo
differentkneemodules.
4.3AssessmentoftheModularSystemThetrunkexoskeletonmodulewastestedbya32‐year‐oldmalewhoretainedpartialmusclecontrolinhislegsafterhisT9incompleteSCI.Forthisindividual,theabductionandrotationjointontheupperthigh(6inFig.30andFig.31)weredisabled,whilethead/abductionjointonthelowerback(1)wasactivated.Fig.34showsphotosofthehipmodulewornbythetestpilotwhilewalkingwiththetrunkexoskeletonmodule.
Thedevice’srangeofmotionanddegreesoffreedomwereexperimentallyverifiedbytwohealthyindividualswithoutmobilitydisorders.Whenjoints1,5,and6wereactivated,able‐bodiedwearerswereabletoperformvariousmovementsandachievedifferentpostureswithoutdeviceimpedance.SomeoftheposturesareshowninFig33.
38
Figure33.Thetrunkexoskeletonmodulewornbyahealthy,non‐injuredindividual.
Figure34.ThetrunkexoskeletonmodulewornbyanincompleteT9SCIpatient.
However,whentheable‐bodiedwearerfullysquattedwithhis/herhipsabducted,thedeviceconformationwasnotideal.Withjoints1,5,and6activated,thedevicewasnotabletofullycomplywithasquattingpostureeventhoughthewearerwasabletoachieveandmaintaintheposture.Throughoutthisempiricalanalysis,itwasfoundthatadditional
39abductionjointsalongaxis9inFig.35withalimitedrangeofmotionsignificantlyincreaseconformitytothewearerduringsquatting.Thebenefitofafullrangeoffreedomonjoints9and1remainsdebatable.
Figure35.Overviewofthehipmodulewithacompensatedabductionjointaxisindicated.Systemmodularitysimplifiesthecustomizationforeachuser’sspecificneedsbymerelyaddingorswappingdifferentmodulesbasedonuserabilities.Accuratelyaligningdevicejointstothewearer’sbiologicalhipjointswhileprovidingpropulsionassistancewithpoweredactuatorsappearstobeachallengingproblem.Especiallyinregardtohipabductionandadductionforable‐bodiedusers,itiscriticaltocomplywithjointmovementstoaugmentendurance.Therefore,exploringmorepossibilitiesforjointlocationstopermitvariousmovements,suchassquatting,lunging,andhipabductionwalking,anddiversifyingactuationunitsremainasfuturework.Enhancingmodularityindesignwillnotonlyaugmentavailablemobilitysolutionsforvariouskindsofindividuals,butalsofillthegapsofavailabledevicesacrossthemobilityspectrum.
40
5.UserInterfaceModuleDesign
State‐of‐the‐artmedicalexoskeletonsusevarioususerinputreceivingmethodswhichcanbecategorizedintotwodifferentgroups:
1) Sensingauser’sintenttowalkfromhis/herposture.Forexample,thesystemcanmeasurethegroundreactionforce,centerofpressure,etc.
2) Receivingexplicitcommandsfromthewearerviadirectuserinput,suchasbuttons.
Theadvantageofthefirstmethodisthatitrequiresminimaluserefforttoissuecommands.However,becauseofthelimitedmobilityandmusclecontroloftheintendedusers,thisimplicitUIhasthepotentialriskofmisinterpretingthewearer’sposture,whichcanresultinfalling.StudieshaveshownthattheimplicitUIofReWalkTMresultsinauniquepatternofcontrolforwalkinginindividualswithSCI;patientswithlowerinjurylevelsshowbetterwalkingperformancesandalsoprogressmorerapidly[46].Meanwhile,explicitUIssuchasthesystemsusedinRexandAUSTIN,requiremoreuserefforttocommandstepsviajoysticksorbuttons.Still,theyarethepreferredexoskeletonoperationmethodsforpatientswithhigherinjurylevelsbecausetheyminimizethemistriggeringrisksresultingfromtheuser’sweakenedmotorcontrol.Yet,thisdevice‐coupledUIdesignalsocouldresultinlimitedusabilitybydifferentuserswhosehand/fingersizesvary(Fig.36).
AsmentionedinChapter4,designmodularityincreasesdeviceusabilitybyprovidingopportunitiestoaugmentsolutionsfordifferentusergroups.Therefore,thischapterdiscussesanexplicit,user‐coupledUIdesignasanexclusivemoduleforparalyzedindividuals.Herethetermcoupledderivesfromthetermcouplingofmechanicalcomponentsinmechanicalengineering.Inparticular,wefocusonauserstudythatconsistsofinterviewsandusabilitytestsofaprototype.Basedonobservationsandinterviewsconductedasapartofthisstudy,theprimaryuserneedsforaUIaredefinedasfollows:
1) Intuitiveness:Theinterfaceshouldbeeasytolearnanduse.
2) Comfort:Thephysicalinterfaceshouldbecomfortableandcomplywiththeuser’sbodyshapeandskeletalstructure.
3) Informationfeedback:Informationfeedbackallowsthewearertobetterengagewiththeexoskeletonsystem.Unliketheclosed‐loopnervoussystemofthehumanbody,exoskeletonsareexternalandartificial.Ideally,thefeedbacksystemwillcompensateforauser’simpairedperceptionofhis/herownposition,pressure,andorientationwithrespecttogravityforbalancing.Moreover,forsafety,theusershouldbeabletomonitorandmaintainfullcontroloverthemotionoftheexoskeleton.
4) Independence:TheUIshouldinterfereminimallywiththeuser’sothernaturalmovementsandallowtheusertodon,doff,andoperatetheexoskeletonbyhimself/herself.
41
Figure36.Anexampleofadevice‐coupledUI.Atwo‐waybuttonisattachedonawalkerhandle.Notethattheswitchisintegratedintoawalkerhandle(device‐coupled).Thetwo‐wayswitchgivestwo
differentsignalsaccordingtothedirectionausertriggerstheswitch(upordown).
5.1SchematicsofExplicitUserInterfaceTheexplicitUIoftheexoskeletonswitchesthestateofthemachinefromonestatetotheotherwhentheuserissuesinputcommandsviameansofinput,suchasbuttons.Inasimpleformofwalkingoperationwithexoskeletons,therearefourdifferentstatesofthemachine:1)leftlegforward,2)rightlegforward,3)feettogether(standing),and4)seateddown.WiththewalkingschemeshowninFig.37,atleasttwobuttons‐thesimplestwayofgivingtwodifferentsignals‐arerequiredfortheusertoswitchbetweenstates.Forexample,whenthewearerisstanding(feettogether),he/shecanpresseitherbuttonAtostartwalkingbymovinghis/herleftlegforward,orbuttonBtositdown.Withthiswalkingscheme,theusercansitdownonlyifhis/herfeetaretogether,andtheweareralwaysmuststartwalkingwithhis/herleftleg.Clearly,therearemultipleapproachestomappingthesecommands,possiblyusingmorebuttons.
Twodifferentexoskeletonfinitestatemachines(i.e.,walkingoperationschemes)weresuggestedandevaluatedviausabilitytestsandinterviews.Thefirstfinitestatemachineiscalled“Singleton”,asshowninFig.36.IntheSingletonmethod,theuserpressesthesamebuttonforeachstep.Forexample,theuserpressesbuttonAforthefirststep→buttonAagainforthesecondstep→(continue)→buttonBtostandup(feettogether)→buttonBagaintositdown.
Thesecondfinitestatemachineiscalled“Alternating”,becausetheuseralternatesbuttonsas his/her stepping legs alternate. Essentially, different buttons are used to step withdifferentlegs.Forexample,inanalternatingscheme,theuserpressesbuttonAfortheleftstep→buttonBfortherightstep→buttonAagainfortheleftstep→(continue)→bothbuttons(oranallocatedthirdbutton,buttonC)tobringthefeettogetherandsitdown.ThisprocedureisillustratedinFig.38.
42
Figure37.Anexoskeletonfinitestatemachine(Singleton).
Figure38.Alternatingfinitestatemachine.
Thetwoaforementionedmethodsweresuggestedandevaluatedbyeightindividualswithmobilitydisorders,viainterviewsandusabilitytestswiththelow‐fidelityUIprototype,whichisdescribedinthefollowingsections.Becausethepotentialusers’preferenceinwalkingschemesmayhighlyaffectthehardwaredesign,itwasimportanttoidentifythepreferredmappingpriortohigh‐fidelityprototypedevelopment.
43
5.2OverviewofUIPrototypeSeveralversionsoflow‐fidelityprototypeshavebeendevelopedtosimulateanewUIandelicitfeedbackfrompotentialusersduringthetestingphase.Internalconceptscoringwasconductedtoselectthenew,user‐coupledUIconcept.Thescoringusedcriteriabasedontheprimaryuserneeds–intuitiveness,comfort,feedback,andindependence.Throughoutthisprocess,a“gloveUI”featuringaglovewithbuttonsattachedtothefingertipsandanLEDdisplaymountedontheinsideofthewrist(Fig.39(a))wasselectedasaproposednextgenerationUIformedicalexoskeletons.ThisgloveUIretainstheconceptofusingthewearer’sfingerstooperateanexoskeleton,whileitisnolongerdevice‐coupledasitwasinthepreviousUIdesign.Fig.39(b)showsawheelchairuserwearingtheprototype.Thisprototypeuseswhitestripesforbettervisualizationoftheblackglove,whilebuttonsareattachedtoallfivefingerstodeterminepreferredbuttonlocations.
ThisnewUIconceptisdesignedtobemoreuser‐centeredthanpreviousdevice‐coupledUIs,andaccommodatesdifferentusers’handsizesandgrippingpreferences.Also,itisintendedtoprovidevisualfeedbacktotheuserviaadisplaymonitor.InformationshownonthemonitorincludesOn/Offstatus,batterylife,andquickinstructionsforexoskeletonoperation.
(a)(b)Figure39.(a)ThethreecomponentsoftheGloveUIprototype.(b)Aparticipantwearingthe
prototype.
44
5.3GloveUIEvaluation:MethodsThegoalofthisevaluationwastofindthepotentialusers’preferencesregardingglovetypes,buttonnumbers,displaylocations,andwalkingschemes.Ourhypothesiswasthatthepotentialuserswouldprefertohavethesimplesthardwareandwalkingschemewithlimitedmaneuvers,ratherthanmorecomplexwalkingschemeswithmoreelaboratemaneuvers.
5.3.1PatricipantsOverview
Atotalofeightindividualswithmobilitydisordersparticipatedinthisevaluation.Abriefprofileoftheparticipantsisasfollows:
1) Participant1:30‐year‐oldmaleandrelativelynewcrutchuserduetohisrecentankleinjury.
2) Participant2:28‐year‐oldmaleand12‐yearcrutchuserafterSCIresultingfromasnowboardingaccident.
3) Participant3:34‐year‐oldfemaleandlifetimecaneorsinglecrutchuserduetocongenitallegabnormalities.
4) Participant4:24‐year‐oldmaleand5‐yearmanualwheelchairuserafterSCIresultingfromacaraccident.
5) Participant5:27‐year‐oldmaleand12‐yearwheelchairuserafterSCIresultingfromamotorcycleaccident.
6) Participant6:24‐year‐oldfemaleand2‐yearwheelchairuserafterSCIresultingfromacaraccident.
7) Participant7:42‐year‐oldmaleand20‐yearpoweredwheelchairuserafterSCIresultingfromacaraccident.
8) Participant8:33‐year‐oldfemaleand2‐yearsinglecrutchandKAFOuserafterkneeinjury.
Feedbackwascollectedfromalltheparticipantsviainterviewsandusabilitytestswithparticipants1),2),3),and4).
5.3.2Interviews
Interviewswereconductedatseverallocationswherepeoplewithmobilitydisordersfrequentlyvisit.OneoftheseplacesistheEdRobertsCampuslocatedinBerkeley,California.Bothguerillaandscheduledinterviewswereconductedwithavarietyofpotentialuserswithmobilitydisorders.Ashortlistofquestionnaireswaspreparedtoobtainbasicdemographicdatafromtheparticipantsbeforetheinterviews.Giventherelativelysmallsamplesize,itwascriticaltocoordinatein‐depthinterviewssuchthattheywereimmediatelyfollowedbyusabilityteststothoroughlyexploreopinionsonthegloveUIprototype.
455.3.3UsabilityTests:MaterialsandProcedure
Usabilitytestswereconductedwithfourindividualswithmobilitydisorders–participants1),2),3),and4).Forusabilitytesting,theparticipantsweregiventhegloveUIprototypeshowninFig.39,andaskedtodemonstratehowtheywouldwalkwiththeirmobilityaidwhilewearingthegloveUI.Detailedfociontheusabilitytestaredescribedasfollows:
1)GloveTypeThegloveprototypeusedforthisusabilitytestisshowninFig.39.Theparticipantswereaskedfortheiropinionsontheglovematerial.Inparticular,whetherthefabricwastoothickorthin,andwhetherthecoverageofthegloveoverthehandwastoomuchortoolittle.
2)GripandButtonsThegripusedbyindividualswithmobilitydisorderstoholdtheircrutchesorwalkersvarieswidelywithhandsize,mobilityaidtype,andgriptype.Dependingonthesefactors,auser’spreferenceforbuttonlocationsmaydifferforeaseofuseandaccuratetriggering.Todeterminetheirpreferences,participantswereaskedtoholdthegripoftheirmobilityaidwhilewearingthegloveUIprototype,andtoselecttheirpreferredbuttonsamongthoseattachedtothefingertipsofthegloveUIprototype.
Theparticipantswerealsoaskedfortheiropinionsonthenumbersandtypesofbuttons.Theywereaskedtochooseatleasttwobuttonsforsimpleoperations,andmorebuttonsforotherpossiblefunctionsormoreelaborateexoskeletonmaneuvers.
3)UIScheme,orWalkingMethodPreferenceTwodifferentUIschemesdescribedinSection5.1wereexplainedtotheparticipants.Then,participantswereaskedtosimulatewalkingwiththetwodifferentschemesusingtheirbuttonpreferencesfromthepreviouspartoftheusabilitytest.TheparticipantswereaskedtofolloweachwalkingschemeinFig.37andFig.38threetimes.Afterthissimulation,theparticipantswereaskedtodescribetheirpreferencesandopinionsregardingthetwodifferentschemes.
4)FeedbackDisplayModulePositionThedisplaymodulecontainsanLEDscreenthatiscapableofdisplayingvarioustypesofinformation,suchasthesystemstatus,batterylife,wirelessconnectivity,anderrordiagnosticswhenapplicable.Theparticipantswereaskedtofindthepositionofthedisplaymodulewheretheycouldreceiveinformationataglancewhileoperatingtheexoskeleton.Thistestwasintendedtolearntheusers’preferencesforlocationsofthedisplaymodule,whichmaydifferfromthetypeofmobilityaidanditsgrip.
46
Figure40.UsabilitytestofthegloveUIprototype.
5.4ResultsandDesignRecommendationsBasedonfeedbackfromtheinterviewandusabilitytestuserresearch,designsuggestionsregardingfourdifferentcomponentsofthenextgenerationgloveUIaresummarizedinthissection(Table6).
5.4.1GloveType
Basedonfindingsfromtheusabilitytestsandinterviews,adoptingonetypeofall‐inclusivegloveforthisapplicationtomeetvarioususerneedsappearstobechallengingduetoconflictingpreferencesamongpotentialusers.Manualwheelchairuserspreferredthick,durableglovestoprotecttheirhandswhileusingawheelchairandwearingamedicalexoskeleton.However,afullglove,evenwiththinmaterialwasnotfavoredbyotherusergroupsbecauseofitsunnecessarycoveragewhichcancauseunpleasantheatingoftheuser’shand.Therefore,fingersleevesarerecommendedinsteadofafullglove.Fingersleevesbetterfacilitatevariationsinusers’handsizes,grippreferences,andpreferencesforglovematerialforbothwheelchairusersandnon‐wheelchairusersbyofferingtheoptionofwearingthegloveUIontopofanyothergloves.
5.4.2NumberofButtons
Allparticipantsrespondedfavorablytowardtactilefeedbackfrommomentaryswitchesontheprototype.Giventhepossiblefinitestatemachinesofanexoskeleton,atleasttwobuttonsarerequired(asstatedinSection5.1).Formoreelaboratemaneuveringofthemachine,threebuttonscanbeconsidered.However,itappearsthatpotentialusersprefer
47twobuttonsduetoitsinherentsimplicityandminimalinterface.Sixoutofeightparticipantsrespondedthattheywouldprefertohaveasimple,minimalUIwithlimitedfunctionratherthanmorebuttonswithadditionalmaneuvers.Regardingthepotentialusers’limitedcontroloftheirlowerbodyandlackofexperiencewithexoskeletontechnology,simplicityshouldbeconsideredasoneofthemostimportantattributesinUIdesign.
5.4.3WalkingOperationScheme
TheSingletonmethodwasthepreferredwalkingoperationscheme;amajorityoftestparticipantsperceivedtheSingletonmethodasmoreintuitiveandeasiertooperate.Despiteitsflexibility,theAlternatingmethodwasdeemedtoocomplicatedforbeginners.AnargumentremainstokeeptheAlternatingmethodasanoptionformoreexperiencedusers.Keyquotationsfromtheparticipantsonthesepotentialwalkingschemesareasfollows:
“I’mthinkingaboutacomputergame.Somepeoplepreferusingfewerbutmoreintuitivecommands,butIliketousemoregranularones.Butforrightnow,IliketheSingletonoptionsinceIamnewtothissystem.”
“AfterawhileifitwassomethingthatIhadtodoregularly,like,say,driveacar,itwouldprobablyjustbecomesecondnaturebecauseIwouldtrainmyselftoclickonlyone.”
Similartotheresultoftheprevioususertesting,itappearedthatsimplicityandintuitivenesswerethekeyfactorsforthepotentialusers’preferences,especiallyforthosewhohavenoexperiencewearingexoskeletons.
5.4.4DisplayModulePosition
Participantsshoweddifferentpreferencesregardingthedisplaymodule’sangleandposition.Walker,crutch,andwheelchairusersviewthedisplayfromdifferentanglesduetotheirdifferentgrips.Therefore,itisrecommendedtohaveanadjustabledisplaymodulepositionwitheasyattachmentmethods,suchasVelcro.
48
Table6.SummaryofUIAssessment
5.5ImplementationoftheUserStudyFingergloves,twobuttons,thesingletonwalkingmethod,andanadjustabledisplaypositionwerethefinaldesignrecommendationsbasedontheUIassessment.Ahigh‐fidelityprototypewasdevelopedtoincorporatemostoftheseresultsofthisuserstudy.Detailsoftheelectriccomponentsusedinthisdesigncanbefoundin[47].AsshowninFig.41and42,thisUIcontainstwobuttons(1)attachedtofingersleeves(2),andisaSingletonfinitestatemachine.Itfurthercomprisesadisplaymodule(5)andcontrollerbox(6)whichenclosesacontrollerthatwirelesslycommunicateswiththemaincontrolleroftheexoskeleton(3).Thecontrollerboxalsocontainsavibrationmotortoprovidehapticfeedbacktothewearer(notshown).Twofingersleevescanbewornonanyfingersoftheuser’spreferredhand.Thedisplaymodule(5)providesthestatusoftheexoskeletontothewearerasvisualfeedback.Informationmayincludeswing/stancelegstates,kneejointstatus,andbatterylife.
Withfurtherimplementationofsensors,thisUIcanprovidefurtherinformationonthewearer’scenterofpressure,whichcouldaidwithbalanceforwearerswithhigherinjurylevels.OtherfutureworkonthisUIdesignincludespreventingmistriggerviaatime‐outlockand/orobject‐dependenttrigger,andmodifyingthedesigntoeasilyrelocatethedisplaymoduletoeachuser’spreferredlocation.
49
Figure41.FingergloveUIoverview.
Figure42.FingerGloveUIwithvisualandhapticfeedback.
50
6.DiscussionsonExoskeletonUserTesting
Thischaptercoversusertestingoftheminimally‐actuatedmedicalexoskeletonandthetrunkexoskeletonmodulewithparaplegicindividuals.Abriefoverviewoftestparticipantsisgiven,anddetailsofusertestingproceduresarediscussedtoprovideausertestingguideline.Sixarchetypalpersonasbasedonusertestingareintroducedattheendofthischapter.
Tothebestofourknowledge,thereisnofundamentalanalysisofusertestingprocedureformedicalexoskeletonswithparalyzedpatients.Thegoaloftheusertestingistonotonlytesttheperformanceoftheexoskeletonsdevices,butalsotoprovideuserexperienceof“re‐learningtowalk”viafieldofpromotedaction.Thus,theimportanceoftheeffectiveandstructuredearlystageusertesting,orpatients’learningprocesses,shouldbeacknowledged.Also,theusertestingprocessaswellastheoutcomesshouldbeconsideredduringthedesignstagetofacilitatelearningandtesting.
Exoskeletonwearerstendtolearnattheirownpace,whichcanrangefromrelativelyfasttorelativelyslow.Learningcurvescanbeinfluencedbyvariousfactors,suchasusers’injurylevels,physicalconditions,timesofinjury,pastrehabilitationexperiences,andmotivations.Duringmedicalexoskeletontraining,patientsdevelopnewutilizationschemasviascaffoldedinteractionsthataremediatedbythetechnology,andguidedbythetrainersconsistingofengineerswhodevelopedthemedicalexoskeleton.Herethetermscaffoldedisusedfromscaffoldingtheory[48]introducedbyJ.Bruner,acognitivepsychologist.Inhisscaffoldingtheory,instructionalscaffoldingisalearningprocessdesignedtopromoteadeeperleveloflearning.Scaffoldingtheorywasdevelopedtobetterunderstand,explainandfacilitatechildren’slearningprocedures.Scaffoldingisthesupportgivenduringthelearningprocess,whichistailoredtotheneedsofthestudentwiththeintentionofhelpingthestudentachievehis/herlearninggoals[49].Whenappliedtoexoskeletontraining,scaffoldingtheoryallowsengineers’/trainers’trainingsessionstobecomebetterstructuredandfacilitateapatient’slearning.Inmedicalexoskeletonusertesting,apatient(student)uses(orlearnshowtouse)anexoskeletontostandup,walkandachievehealthbenefits(thelearninggoals).
Inthisprocessoflearningactivitywithexoskeletons,thefollowingagentsparticipateasscaffolding,andinteractwiththepatient(student).ThisisalsoshowninFig.43.
1) Tether;theweareristetheredwiththeexoskeletonstructuretoensurefallsafety.2) Parallelbars;thewearerholdsontotheparallelbarstobalancehimself/herself.3) Twotrainers;tospotthewearerinfrontandbehind.4) Apractitioner;tooperatetheexoskeleton,andtunethegaittrajectory.5) Designersofthetechnology.6) Anexoskeleton.
Notethat3),4),and5)canbeinterchangeable,forexample,oneofthetrainersin3)canbeapractitioneroradesignerofthetechnology.Throughouttheusertesting,orthepatients’relearningtowalkprocedure,humanagentsfulfillavarietyofroles;theytakesomeofthe
51distributedactions,suchasmanipulatingthedevice,spotting,takingactions,andcueingforactions.
Figure43.Sceneoffirsttimeexoskeletontrial.
6.1Minimally‐ActuatedMedicalExoskeletonUserTestingInmanysenses,theminimally‐actuatedmedicalexoskeletonusertestingisakintotheprocessoflearningtorideabicycle.Bothinvolvelearningviamotorproblemsolving,ratherthanrepetitionthroughoutthescaffoldedlearningprocess,untilthewearerachievestheabilitytobalancehimself/herselfandwalk(ride).
6.1.1TestParticipantOverview
Fortheminimally‐actuatedmedicalexoskeleton,itiseasiertoworkwithindividualswithoutjointcontracture,spasticity,andmuscletone.Theyactasexternalfactorsthatincreaseanatomicjointdamping,andoftennegativelyaffectthebehaviorofpassivekneejointsofminimally‐actuatedmedicalexoskeletons.Thekneejointsmaycreateaggressive,moderate,orhamperedswingextensiondependingonthewearer’sanatomicjointdamping.Aggressiveswingextensioncanbetunedbyaddingdampingelementstothekneejoints[41].Whilehamperedswingextensioncanbetunedbymodifyingthehipswingtrajectorytomatchthepatient’sspecificconditionsandcancompensateforanatomicdampingtoacertaindegree,anatomicjointdampingshouldnotbesignificantenoughtolockupthekneejointstoachieveeffectiveambulation.
Threeindividualswithparalysisintheirlowerbodyhavetriedtheminimally‐actuatedmedicalexoskeleton.Theyhavevaryingdegreesofinjurylevels,upperbodystrength,physique,andexperiencewithpreviousrehabilitation.
1) Testpilot#1‐isa28‐year‐oldmale,5’10”tall,and170lbs.HesustainedacompleteT12injury12yearspriortothetesting.Insteadofawheelchair,hismainmobility
52
comprisesaswing‐throughgaitusingcrutchesandpassivelegbraces.Becauseofthispriorexperience,hisbalancingskillsareexceptionallyhigh.Hehasamoderateamountofmuscletone,buthasnosignsofspasticityorjointcontracture.
2) Testpilot#2‐isa27‐year‐oldmale,6’tall,and110lbs.AlthoughhesustainedacompleteT10injuryeightyearspriortothefirsttesting,hemaintainedabdominalcontrol.Hehasasubstantialmuscleatrophy,whichmeanshiskneejointshaveminimalmuscletoneresistance.Sincediscontinuinghisrehabilitationwithlegbracesaboutsixyearsagoduetoexcessivephysicalexertion,hehasprimarilyreliesonawheelchair.
3) Testpilot#3‐isa24‐year‐oldfemale,5’2”tall,and140lbs.ShesustainedacompleteT5injurylevelthreeyearspriortothefirsttesting.Shehasverylittletoalmostnoabdominalcontrol,andacomparativelyhighamountofmuscletone.Sheiscurrentlyparticipatinginastemcellclinicaltrial.Sheistestingtheexoskeletonasapotentialmethodforrehabilitation.
6.1.2UserTestingProceduresandGuideline
Whenapatientdonstheminimally‐actuatedmedicalexoskeletonforthefirsttime,therearethreecriticalpointsthatneedtobesatisfiedbeforethepatientcanstarttakingsteps.
First,thedistancesbetweenthepatient’shipandkneejoints,aswellaskneeandanklejointsneedtobemeasuredinadvance.Inaddition,thelengthoftheexoskeleton’sthighandshanklinksmustbeadjustedaccordingly.Thetrainercanverifytheaccuracybycheckingthekneejointalignmentbetweentheexoskeletonandthepersoninthesagittalplanewhilethepatientissitting,asshowninFig.44(a).Thisalignmentcanbeachievedonlyifboththighandshanklengtharecorrect,andiscriticalforenablingthewearertobothstandupandusetheexoskeletonproperly.
Thesecondcriticalpointistore‐checkthisalignmentwhenthepatientstandsup.Standingwithanexoskeletonatfirstmayrequiresomeeffortbyutilizingthemechanicaladvantageofexoskeletonlegsandauxiliarybalancingwithhis/herarms(Fig44(b)).Whenthewearerisstandingwiththeexoskeleton,thedevice’skneejointshouldbefullyextendedandalignedtothebiologicalkneejoint.Ifthedevice’skneejointisnotfullyextended,thethighandshanklinklengthsoftheexoskeletonneedtobeadjustedaccordingly.Thedevice’skneejointalignmentshouldbeverifiedinboththefrontalandsagittalplanes.Ifthejointsarenotalignedwhenextended,strappingsorpaddingonthethighandcalfbracemayhavetobeadjusted.Ideally,thereshouldbenogapbetweenthewearer’sthighandtheexoskeleton’sthighbrace.Thesameholdsforthecalf.UsingapairofKAFObracingsthatiscustommadeforthepatientisrecommendedforthebestresult.Utilizingaccessibletechnologies,suchas3‐dimensionalscanningandprinting,isalsoanoptionforprovidinghighlyfunctionalbracings.Theamountofpaddingshouldbeaddressedsothatwearingtheexoskeletondoesnotcauseanyexcessivepressure.Especiallyforparalyzedpatientswho
53lacktactilesenseintheirlowerbodies,unnoticedexcessivepressurecanbeprolonged,causingsecondaryinjuriessuchaspressuresores.
Thefinalpointthatneedstobeaddressedatthebeginningofusertestingistheadjustmentofthedefaultankleangleandtorsoangle.Oncetheseanglesareadjusted,theintegratedsystemofthewearer,exoskeleton,andbalancingaidcanachieveastateofstaticequilibriuminthesagittalplane,sothatthewearercanholdastandingpositionmorecomfortablywithlittleeffort.Theseanglesneedtobetunedfordifferentwearers,sincethephysiqueofeachwearerandexoskeletonpaddingscreatedifferentcentersofmassfortheintegratedsystem.Fig.45(b)showstestpilot#3describedinSection6.1.1standingatthestaticbalancingpoint.
Assoonasaweareriscomfortablewiththestandingpositionandallalignmentsaretuned,takingstepscanbeinitiated.Totakeeffectivesteps,thepatientmustlearntoconstantlyshifthis/herbodyweightbetweentheswinglegandstancelegefficiently.Thisisatfirstobtainedviaexaggeratedweightshiftingmovementsfromsidetoside,requiringratherslow,exaggeratedmovementswithpausesinbetween.Fig.46(a)showsthisactivityoftestpilot#3.Atthisstageoftesting,itisrecommendedtohavethewearersupportedbyatether,withatleastonepersonspottingandanotherpersonoperatingtheexoskeletonwhilecueingthesignaltothewearerfortakingsteps.
Asthepatientgetsaccustomedtothisactivity,exaggerationsinweightshiftingandbalancingreduceandbecomemorenatural.Asthewearergainsexperiencewithside‐to‐sideweightshifting,he/shewillalsolearnhowtocontrolhis/herupperbodymovementsbackandforthinthesagittalplane.AsmentionedinChapter3,theminimally‐actuatedexoskeletonhaspassivekneejoints,whicharedynamicallydrivenbyhipactuations.Thus,placementofthewearer’storsocanchangethesteplengthbyaffectingthisdynamicswing.Forexample,stepsizewillbesmallerifthewearerleansfartherbackwardsthanthenormalstandingangle.Untilthewearerisabletoshifthis/herweightnaturally,itisrecommendedtohavethewearertetheredandinbetweenparallelbarswhileatraineroperatestheexoskeletonandcuesthewearer.
Assoonasthepatientovercomesexaggeratedweightshifting,he/shecantakestepsoutsidetheparallelbars,witheithercrutchesorawalker.Thewearercannowoperatehimself/herself.Asthewearerlearnshowtooperatethedeviceandwalkwithit,thetethercanberemovedandhe/shemayonlyneedoneindividualforspotting.Inotherwords,thenumberofsupportingagents,orscaffolds,isnowreducedcomparedtothebeginningoftesting.Theamountofthetrainingnecessarytoreachthisstagevariessubstantiallywithpatients’variousconditions.Fromthisstageoftrainingon,thedecisionmakerintheactivityofwalkingstartstobecomethewearerhimself/herself,ratherthanthetrainer.Fig.46(b)showstestpilot#1(describedinSection6.1.1)walkingonthestreetusingamedicalexoskeletonafterthreemonthsoftraining.Here,thedecisionmakerinthisactivityofcontinuouswalkingisthewearerhimself,withengineers’knowledgeencodedinthedevice.
54
(a)(b)
Figure44.(a)Checkingkneejointalignmentbeforestanding.(b)Thefirsttimestandingup.
(a)(b)Figure45.(a)Checkingkneejointalignmentafterthefirsttimestandingup.(b)Afteradjustingtorso
andankleangles,thepatientstaysbalancedmoreeasily.
55
(a)(b)Figure46.(a)Aparaplegicpatienttakingherfirststeps.(b)Anotherparaplegicpatientwearingan
exoskeletonwalkingonthestreetwithoutanyassistance.
6.2TrunkExoskeletonUserTesting
6.2.1TestParticipantOverview
ThetrunkexoskeletonmodulediscussedinChapter4wastestedbya32‐year‐oldmale,5’6”tall,and135lbs.HeisaT9incompleteparaplegicpatientwithreducedsensoryandmotorfunctionsduetoTransverseMyelitis.Thetrunkexoskeletonmodulewithoutakneemodulewasapplicabletothispatientovertheminimally‐actuatedexoskeletonfortworeasons:
1)Thefullbodyexoskeletoniscumbersomeforthepatientwhoretainscontrolsofhislowerleg.
2)Thepatientisalessidealcandidateforanexoskeletonwithpassivekneejointsbecauseheexhibitsspasticityandasubstantialamountofuncontrolledjointdampinginhisknees.Thisanatomicdampinghamperstheperformanceofpassivekneejointsusedintheminimally‐actuatedexoskeleton.
6.2.2UserTestingProcedureandGuideline
Sincethepatientexhibitssubstantiallyhighermusclecontrolcomparedtotestpilots#1,#2,and#3inSection6.1.1,fewersupportingagentswererequiredforthistesting.Also,becausethisdevicesetupdoesnotincludethekneemodule,thelengthadjustmenttoachievekneejointalignmentisnolongernecessary.However,itshouldbenotedthatwithoutanAFO,thewearersupportsthedeviceweight,asopposedtothefull‐bodymedicalexoskeleton.Thus,thelevelofcomfortintheupperbodyiscritical.Specifically,the
56lowerbackandshouldersaretheprimarycontactsurfaceswherethewearersupportstheloadofthedevice.Inthetestingwiththeaforementionedpatient,memoryfoampaddingforthelowerbackandlight‐duty,elasticshouldersuspenderswereused.Regardinghislevelofupperbodycontrol,theshoulderstrapsusedfortheminimally‐actuatedmedicalexoskeletonwerenotnecessaryandfoundtobecumbersometothewearer.
TheorderofoperationswouldbesimilartotheonefortestingthemedicalexoskeletondiscussedinSection6.1.Thelengthofthethighlinkandthefittingofbracesneedtobeverifiedbeforethewearercantakeanysteps.Also,thetorsoanglemustbeadaptedtothewearer’scomfortablestandingposture.Althoughthewearerretainssensoryandmotorfunctionsinhislowerbody,itisrecommendedtostarttestingwhiletheweareristetheredwiththeexoskeletonandinbetweenparallelbars,withindividualsspottingthewearer.Oncetheweareriscomfortablewithstandingandbalancingwhilewearingtheexoskeleton,atrainercaninitiatesteptaking.Afterthewearerisaccustomedtotakingstepsandshiftingweightside‐to‐sideandbackandforth,he/shecantrytakingstepsoutsideoftheparallelbars,operatingtheexoskeletonbyhimself/herself.Asthewearerimprovesbalancingandoperatingtheexoskeletonwhilewalking,thetethercanberemoved.Nowthewearerbecomesthedecisionmakerofthewalkingactivitybyoperatingthedeviceandinteractingwiththedeviceencodedbyengineers.
6.3DesignConsiderationsforFutureUserTestingThroughouttheusertesting,itwasfoundthatcertaindesignfeaturesandtrainingstrategiessignificantlyfacilitatetheusertesting.Theseconsiderationsshouldbetakenintoaccountforfuturegenerationsofexoskeletondesignsandusertestingsetupstoacceleratethelearningexperience.
1) Designfeaturesfortetheringandspotting:Intheearlystageofusertesting,thepatientshouldalwaysbetetheredandspotted.Afeaturefortetheringandspottingshouldbeincorporatedintoexoskeletondesign.Fortethering,itisimportanttoincludeanadapterthatcantaketheweightofboththeexoskeletonandthepatient.Forspotting,thetorsoframeshouldincludeasufficientamountofsurfaceareathatatrainercanholdduringtesting.
2) Easyadjustment:AsdescribedinSection6.1.2,therearemanyadjustmentsthathavetobemadefordifferentpatients–lengthadjustmentstoalignbiologicaljointswithdevicejoints,paddingamountadjustments,andtheangleadjustmentsforcomfortablestanding.Itiscriticalforthedesigntoincludeeasyaccessandsimpleproceduresfornecessaryadjustments.
3) Patient’sturn‐aroundmaneuver:Itshouldbenotedthattherearespaceconstraintsinthelaboratorywhenusingatetherandparallelbars.Oftentimes,thepatientneedstoturn180degreesattheendoftheparallelbarsorwhenthetetherropereachesitsendoftravel.Inthebeginningstagesoflearning,theturn‐aroundmaneuvercanbeverychallengingforpatients.Therefore,itishighlyrecommendedfortrainerstoplanthewalkingpathinadvancetominimizethismaneuver.
57
6.4ArchetypalPersonasinEachUserCategoriesThroughouttheusertesting,theexoskeletondeviceswereevaluatedbydifferentindividualswithvaryinginjurylevels.Structuredusertestingprovided“re‐learningtowalk”experiencestopatients,whileallowingpatientstoactivelyengagewiththedevices,trainersanddesigners.However,theusertestingwaslimitedtofourindividualswithvaryinglevelsofparalysis,eventhoughthereareothersegmentsofusergroupswhocanpotentiallybenefitfromthistechnology.
Here,sixarchetypalpersonasaredevelopedtorepresenteachpotentialusergroupsegment.Asoneofthehuman‐centered‐designmethodologies,creationanduseofarchetypalpersonasthatrepresentimportantclassesofusersprovideaclearwaytodefinetargetuserswhosharecommondesires,goals,andbehaviors,andwhowillbenefitfromthetechnology[50‐53].Thepersonasarebasedontheresultsofexoskeletonusertesting,medicalexoskeletonUIusabilitytesting,andotherinterviewswithindustryworkersandsoldiers.Table7summarizesthesixarchetypalpersonaswiththeirkeyattributes.
6.4.1Kevin,apoweredwheelchairuser
Kevinisa35‐year‐oldmale,anincompleteT3quadriplegicpatientwithacentralcordsyndromefromacaraccidenteightyearsago.Althoughheretainsalimitedamountofsensoryandmotorfunctioninhislowerbodyincludingabdominalcontrol,theinjuryresultedinthelossofarmfunction.Becauseofhisimpairedarmfunction,hehasbeenusingapoweredwheelchairsincetheinjury.Unfortunately,hisphysicalconditionofimpairedupperbodycontrolprecludestheuseoftheminimally‐actuatedmedicalexoskeletonorthetrunkexoskeletonmodule.Moreover,thelongerhestaysinapoweredwheelchair,thehigherhislikelihoodofmusclecontractureandreducedrangeofjointmotions,whichwillmakeKevinanevenlessidealexoskeletonwearercandidate.
6.4.2Jennifer,awheelchairuser
Jenniferisa31‐year‐oldfemale,acompleteT9paraplegicpatientfromamotorbikeaccidentoneyearago.SheiscurrentlyparticipatinginrehabilitationwithanRGO.Asofnow,herupperbodystrengthisnothighenoughtosupportandambulateherselfwithanRGO.Eventhoughsheisawareofthenecessitytocontinuerehabilitationactivitiestodelaytheonsetofsecondaryinjuries,sheisconsideringdiscontinuingtherehabduetoanexcessiveamountofphysicalexertion.SheusesawheelchairashermainmobilityaidexceptforthetimesinrehabwhensheusesanRGO,whichisfivehoursperweek.Aminimally‐actuatedmedicalexoskeletoncanpotentiallybeausefulrehabilitationtoolandmotivationforJennifertocontinueherrehabactivity,becausepoweredexoskeletonusewillnotrequireasmuchupperbodystrength.HerexperienceusinganRGOwillaccelerateherlearningexperiencewiththemedicalexoskeletonintermsofbalancingandweightshifting.
586.4.3.Samuel,acrutchuser
Samuelisa27‐year‐oldmale,acompleteT12paraplegicpatientasaresultofanaccidentwhilesnowboardingtenyearsago.HehasbeenparticipatinginvariousrehabactivitieswithaKAFOsincehisaccident.HehasbeenusingaKAFOandcrutchesashismainmobilityaidinsteadofawheelchair,whichkeepshimmoreactiveandprovideshealthbenefits.Hisathleticlifestylebeforetheinjurycoupledwithhisbalancingabilityacquiredfromhisextensive,professionalsnowboardingexperiencegreatlycontributedtohisrehabilitation.Samuelisagoodcandidatefortheminimally‐actuatedmedicalexoskeleton.AlthoughheisalreadyabletoachieveahighlevelofambulationwithhisKAFOandcrutches,theexoskeletonwillalleviatehisexhaustiveupperbodyuse,andenablehimtotravellongerdistanceswithoutgettingtired.
6.4.4Mary,aseniorcitizen
Maryisa60‐year‐oldfemaleseniorcitizenandaretiredaccountant.Shehasnotsustainedanyseriousdiseasesorinjuries,althoughshehasalightamountofarthritisduetoaging.Herpreviousoccupationdemandedsittingforlonghoursduringwork,andshehasnotbeenregularlyworkingoutforthelast35years.Asaresult,herbodystrengthisrelativelylowandshewishessheweremorefittoplaywithhergrandkidsattheplayground.ThetrunkexoskeletonmodulewithextralockabledegreesoffreedomisidealforMary,eitherwithorwithoutakneemodule,dependingonherpreference.
6.4.5Tom,anindustryworker
Tomisa40‐year‐oldmalewhoworksforalogisticscompanydeliveringpackages.Hehasmaintainedagoodphysique,but,attheendoftheday,hehaspaininhislowerbackfromrepetitivebendingandliftingactivitiesduringhisworkday.Hisjobdoesnotdemandliftingheavyboxesthataremorethan50pounds,butrepetitivebending,squatting,andwalkingthroughoutthedaytireshimout.ForTom,thetrunkexoskeletonmoduleisrecommendedforassistinghisbendingandwalking,becauseitfeaturesmoredegreesoffreedomwhichallowvariousmovementsusingmodifiedsoftwarefordifferentfinitestatemachines.
6.4.6Andrew,amilitarysoldier
Andrewisa25‐year‐oldmaleriflemanwhoservesintheU.S.ArmyInfantry.Onpatrol,theaverageweighthecarriesisabout100pounds.Hemustcarrytheseheavyweightlongdistances,whichoftenexhaustshimeventhoughhehasveryhighphysicalstrengthandendurance.Hefeelsthattheweightoftheloadhecarriesimpedeshismobilityandagility,andplaceshimatriskofinjurywhentraversinguneventerrain.TheidealexoskeletonforAndrewissimilartoTom’s(6.3.5),butwithhigherperformanceintermsofpowerandrangeofmotion.Also,forAndrew,afull‐bodyexoskeletonishighlyrecommendedtoreducetheloadhecarriesbytransferringittothegroundthroughtheexoskeleton.
59
Table7
.Arch
etypalP
ersonas
60
7.LimitationsandFutureWork
Thischapterdiscusseslimitationsandfutureworkforenhancingthepracticalityandusabilityofassistiveexoskeletons.
7.1HipActuatorsLowprofilehipactuatorsandpassivekneejointsincreasemedicalexoskeletons’practicalityandusability.AsdiscussedinChapter3,twodifferentactuationunitsweredevelopedforthispurpose:1)AdesignwithbarecomponentsofBLDCmotorsandstrainwavegears:theEmoteqHT03000‐J01‐ZandHarmonicDriveTMCSD‐25‐160‐2A‐GR‐SPand2)Adesignwithoff‐the‐shelfhousedcomponents:theMaxonEC90FlatmotorandHarmonicDriveTMSHD‐20‐100‐2SH.Thefirstdesignhasthemostcompactprofilecomparedtoavailableproductsonthemarketwithcomparablespecifications.Meanwhile,thedesign’sdisadvantagesincludeitscomplicatedassemblyprocessandhighcostduetomanycustom‐madecomponents.Theseconddesignimprovesupontheseshortcomings,withfewercustom‐madepartsandaneasierassemblyanddisassemblyprocess.Asatradeoff,adoptingtheseactuationunitsresultsinanincreaseintheexoskeletonhipwidthofabout1.9inchesandadecreaseduserweightlimitduetothestrainwavegear’sbearings.Futureworkremainstoimproveonbothdesigns’limitations.Giventhecurrentlyavailableproductsonthemarket,itisanticipatedthatincreasingtheuserweightlimitwithahousedstrainwavegearwillmostlikelyfurtherincreasetheformfactoroftheactuationunit.Therefore,reducingthenumberofcustommadepartsofthefirstdesignandsimplifyingtheassemblyprocessisrecommended.Exploringothercombinationsofmotorsandtransmissions,forexample,utilizingtheMaxonEC90motorwiththeCSDseriesstrainwavegears,isneededtofurtheroptimizethedeviceprofileandcost.
7.2ModularityAsdiscussedinChapter4,itisimportanttoprovideeasy‐to‐usecustomizationfeaturestomeetusers’differentconditionsandneeds.Thetrunkexoskeletonmodulewithadditionallockablejointsallowssimplebutversatileconfigurationsfordifferentuserswithvaryingmobilities.However,accuratelyaligningab/adductiondevicejointstothewearer’sbiologicalhipjointswhileprovidingpropulsionassistanceisachallengingproblem.Thisiscritical,especiallytoaugmenttheenduranceofable‐bodiedusers.Exploringmorepossibilitiesforjointlocationstopermitvariousmovementssuchassquatting,lunging,hipabduction,walking,anddiversifyingactuationunitsremainasfuturework.Enhancingmodularityaswellasdevelopingvariouskindsofmodulesfordifferentuserswillaugmentmobilitysolutionsandfillgapsindeviceavailabilityacrossthemobilityspectrum.
7.3UserInterfaceAglovetypeuser‐coupledUIdesignformedicalexoskeletonsprovidesflexibilityinthegripandfeedbackfromtheexoskeletontotheuser.However,furtherimplementingsensors
61andimprovingcommunicationsbetweentheexoskeletonandtheUImoduleremainasfuturework.Utilizinginertialmeasurementunits(IMU)orpressuresensorstodetectawearer’sbalancingstatusandtransmitthesignaltotheuserinanenergeticallyefficientmannerwillbebeneficialforuserswithhighinjurylevels.Also,preventingfalsetriggerswithatime‐outlockordevice‐dependenttrigger,aswellaslocatingthedisplaymoduleattheuser’sdesiredpositionareimportantfutureworktomaturethistechnology.Currently,thedisplaymoduleisdesignedtobelocatedonauser’swrist.However,beingabletoadjustthislocationbasedontheuser’spreference,orprojectonanopticalheadmountdisplay(OHMD)willfurtherincreasetheusabilityofthisdevice.Finally,modifyingthisUItobemodular–i.e.,separatingthegloveandfeedbackdisplaymodule–sothatapractitionercanweartheglovetooperateanexoskeletonwhilefeedbackisgiventothewearerwouldbehighlybeneficialforfirsttimewearers’re‐learningtowalkactivities.
62
8.ConcludingRemarks
Thisdissertationcoveredthreedifferentdesigntopicsthatcontributetoincreasingtheusabilityofassistivelowerextremityexoskeletonsystems.Adesignofaminimally‐actuatedmedicalexoskeletonwasintroduced,withspecialemphasison“lowprofilehipactuationunits”.Thelowprofilehipactuationunitsenabledthesystemtobethemostcompactpoweredmedicalexoskeleton,andfirstofitskindtoweighlessthantwentypounds.Thisexoskeletonevolvedtoanexoskeletonsystemwithincreasedmodularityandadjustabledegreesoffreedom.Themodulardesignopensanopportunitytoexpandusersegmentstoindividualswithvariousconditionsandneedsbyallowingeasycustomizationbymeansofcombiningdifferentmodules.AcomprehensiveuserstudyoftheexoskeletonUIwasconductedinanefforttocreateacomfortable,intuitivelinkbetweentheuserandthemachine.Theintentoftheresearchwasnotonlytodevelopassistivedevices,butalsotocreatepositiveuserexperiencesbyprovidingstructuredlearningactivitieswherethepatients,devices,andengineerswhodevelopedthedevicesactivelyinteract.
Theminimally‐actuatedmedicalexoskeletonusesthelowprofilehipactuationunitsandpassivekneejoints.Thepassivekneejointsaredynamicallydrivenbythemomentumcreatedbythehipactuationunits.ThreeindividualswithcompleteSCIwithinjurylevelsrangingfromT5toT12wereabletoachievebipedallocomotionandahigherlevelofmaneuverabilitywithextendeduse.Itshouldbenotedthatthesystemisveryminimalwithonlyonepairofpoweredactuators,whichprovidessimpleambulationwithajointcouplingmechanismwithcompactandlightweighthardware.Eventhoughitdoesnotprovideanyelaboratejointmovements,suchasself‐balancing,stairascending,andstairdescending,thisambulation‐centrictechnologycanprovidehigherlevelsoffreedomandmaneuverabilitythankstoitslightweightandcompactsize.
Thetrunkexoskeletonmoduleincludessimplecustomizingfeaturesthatpermitandincreasemodularityindesign,withsimplecustomizingfeatures.Lockablefeaturesallowvarioustypesofusers–fromcompleteparaplegicpatientstoindividualswithintactmobility–tobenefitfromthistechnology.Itsmodularityandcapabilitiesforexclusiveuseenableeasycustomizationbycombiningdifferentkneemodulesfordifferentpurposes.Developingvariousmoduleswithdistinctivefinitestatemachinesandexploringmoredesignoptionsforcomplyingwithvariousmovementsassociatedwiththehumanhipjointremainasfuturework.
AconceptualUIprototypethatcontainsthekeycomponentsforevaluationwascreatedandassessedbypotentialusers.Threeresearchmethodswereusedinthisresearch:interviews,guerillatests,andusabilitytests.EightintervieweesandusertestparticipantsweredrawnfromthoseworkingwiththeBerkeleyRoboticsandHumanEngineeringLaboratoryandtheEdRobertsCampusforthedisabledinBerkeley,California.Evenwiththissmallnumber,itwaspossibletofinddistinctuserneedsandpreferences.Throughouttheusertests,designrecommendationsweredrawn,andahigh‐fidelityprototypewasdeveloped.TheUIprototypeuserstudyaddressesthenextgenerationUIdesignformedicalexoskeletons,whichcanimprovetheirusabilityandlearnability.
63Futureworkinthefieldofassistiveexoskeletonsincludesperformanceenhancementsandvariousmoduledevelopmentsthatfurtherexpandtheusergroupswhocanbenefitfromthetechnology.Currently,thetechnologyisstilllimitedtoindividualswhoretainacertainlevelofupperbodycontrol,excludingKevinfromthepersonasinSection6.3.Safetyfeatures,suchasreliableself‐balancingwhilemaintainingthelowprofileofthesystem,wouldbehighlybeneficialtothosewhosufferfromhigherlevelsofSCI.Also,increasedambulationspeedwouldincreasepracticalityforthedailyuseofmedicalexoskeletons.Applyingmachinelearningadaptationforexperiencedwearerstograduallyacceleratetheambulationspeedisapromisingsolution.Withperformanceenhancementandmodularity,theexoskeletontechnologycanfurtherextenditsbenefitstoamorediversepopulation,includingparalyzedindividualswithhigherinjurylevels,theelderly,industryworkers,andsoldiers.Ihopethattheassistiveexoskeletontechnologywillcontinueadvancingtoenhancethequalityoflifeforawidespectrumofcommunitieswithadiversearrayofneeds.
64
Bibliography
[1] "NationalSpinalCordInjuryStatisticalCenter,"2013.[Online].Available:https://www.nscisc.uab.edu/.[AccessedMarch2014].
[2] D.Sabo,S.Blaich,W.Wenz,M.Hohmann,M.Loew,andH.J.Gerner,"Osteoporosisinpatientswithparalysisafterspinalcordinjury,"ArchivesofOrthopaedicandTraumaSurgery,vol.121,p.75–78,January2011.
[3] B.Kemp,"Qualityoflifewhileagingwithadisability,"AssistiveTechnology,vol.11,pp.158‐63,1999.
[4] J.J.Eng,S.M.Levins,A.F.Townson,D.Mah‐Jones,J.Bremner,andG.Huston,"Useofprolongedstandingforindividualswithspinalcordinjuries,"Physicaltherapy,vol.81,p.1392–9,August2001.
[5] "SpinalCordTrauma,U.S.NationalLibraryofMedicine,"14May2012.[Online].Available:http://www.nlm.nih.gov/medlineplus/ency/article/001066.htm.[AccessedApril2014].
[6] "Apparalyzed,"[Online].Available:http://www.apparelyzed.com/spinal_cord_injury.html.[AccessedMarch2014].
[7] R.J.Marino,T.Barros,F.Biering‐Sorensen,S.P.Burns,W.H.Donovan,D.E.Graves,M.Haak,L.M.Hudson,M.MPriebe,"Internationalstandardsforneurologicalclassificationofspinalcordinjury,"Thejournalofspinalcordmedicine,vol.26,no.1,pp.S50‐6,2003.
[8] "CenterforOrthoticsDesign,"2013.[Online].Available:http://www.centerfororthoticsdesign.com/isocentric_rgo/.[AccessedMarch2014].
[9] L.A.Harvey,T.Newton‐John,G.M.Davis,M.B.Smith,andS.Engel,"Acomparisonoftheattitudeofparaplegicindividualstothewalkaboutorthosisandtheisocentricreciprocalgaitorthosis,"SpinalCord,vol.35,no.9,p.580–584,1997.
[10]G.Baardman,M.J.Ijzerman,H.J.Hermens,P.H.Veltink,H.B.Boom,G.Zilvold,"TheinfluenceofthereciprocalhipjointlinkintheAdvancedReciprocatingGaitOrthosisonstandingperformanceinparaplegia,"Prostheticsandorthoticsinternational,vol.21,
65
no.3,pp.210‐221,1997.
[11]W.B.Johnson,S.Fatone,B.P.O.Hons,S.A.Gard,"Walkingmechanicsofpersonswhousereciprocatinggaitorthoses,"JournalofRehabilitation,vol.46,no.3,p.435–446,2009.
[12]M.Bernardi,I.Canale,V.Castellano,L.DiFilippo,F.Felici,M.Marchetti,"Theefficiencyofwalkingofparaplegicpatientsusingareciprocatinggaitorthosis,"Paraplegia,vol.33,no.7,p.409–415,1995.
[13]L.Harvey,G.M.Davis,M.B.Smith,S.Engel,"Energyexpenditureduringgaitusingthewalkaboutandisocentricreciprocalgaitorthosesinpersonswithparaplegia,"Archivesofphysicalmedicineandrehabilitation,vol.79,no.8,p.945–949,1998.
[14]L.Sykes,J.Edwards,E.S.Powell,E.R.Ross,"Thereciprocatinggaitorthosis:long‐termusagepatterns,"Archivesofphysicalmedicineandrehabilitation,vol.76,no.8,pp.779‐783,1995.
[15]A.B.Zoss,H.Kazerooni,A.Chu,"BiomechanicalDesignoftheBerkeleyLowerExtremityExoskeleton(BLEEX),"Mechatronics,IEEEASME,vol.11,no.2,pp.128‐138,2006.
[16]A.Chu,H.Kazerooni,A.Zoss,"OntheBiomimeticDesignoftheBerkeleyLowerExtremityExoskeleton(BLEEX),"inIEEEInternationalConferenceinRoboticsandAutomation,2005.
[17]H.Kazerooni,J.Racine,L.Huang,R.Steger,"OntheControloftheBerkeleyLowerExtremityExoskeleton(BLEEX),"inIEEEInternationalConferenceonRoboticsandAutomation,2005.
[18]A.Goffer,"Gait‐locomotorapparatus".USPatent7153242B2,2006.
[19]H.A.Quintero,R.J.Farris,M.Goldfarb,"Controlandimplementationofapoweredlowerlimborthosistoaidwalkinginparaplegicindividuals,"inIEEEInternationalConferenceonRehabilitationRobotics,2011.
[20]K.A.Strausser,T.A.Swift,A.B.Zoss,H.Kazerooni,andB.C.Bennett,"MobileExoskeletonforSpinalCordInjury:DevelopmentandTesting,"inASMEDynamicSystemsandControlsConference,2011.
66[21]R.LittleandR.Irving,"Selfcontainedpoweredexoskeletonwalkerforadisabled
user".USPatentUS20110066088A1,2011.
[22]"RexBionics,"[Online].Available:http://www.rexbionics.com.[AccessedMarch2014].
[23]W.Y.Tung,M.McKinley,M.V.Pillai,J.Reid,andH.Kazerooni,"DesignofaMinimallyActuatedMedicalExoskeletonWithMechanicalSwing‐PhaseGaitGenerationandSit‐StandAssistance,"inASMEDynamicSystemsandControlConference,PaloAlto,2013.
[24]J.I.Reid,M.McKinley,W.Tung,M.PillaiandH.Kazerooni,"AMethodofSwingLegControlforaMinimallyActuatedMedicalExoskeletonforIndividualsWithParalysis,"inASMEDynamicSystemsandControlConference,PaloAlto,2013.
[25]P.D.Neuhaus,J.H.Noorden,T.J.Craig,T.Torres,J.Kirschbaum,J.E.Pratt,"DesignandEvaluationofMina,aRoboticOrthosisforParaplegics,"inIEEEInternationalConferenceonRehabilitationRobotics,Zurich,2011.
[26]V.T.Inman,H.J.Ralston,andF.Todd,Humanwalking,Williams&Wilkins,1981.
[27]T.F.Novacheck,"Thebiomechanicsofrunning,"Gait&Posture,vol.7,no.1,pp.77‐95,1998.
[28]D.A.Winter,BiomechanicsandMotorControlofHumanMovement,JohnWiley&Sons,2009.
[29]C.Kirtley,"CGANormativeGaitDatabase,"[Online].Available:http://www.clinicalgaitanalysis.com/data/.[AccessedApril2014].
[30]J.Linskell,"CGANormativeGaitDatabase,"[Online].Available:http://www.clinicalgaitanalysis.com/data/.[AccessedApril2014].
[31]D.A.Winter,TheBiomechanicsandMotorControlofHumanGait,4ed.,JohnWiley&Sons,Inc.,2009.
[32]D.A.Winter,"EnergyGenerationandAbsorptionattheAnkleandKneeduringFast,Natural,andSlowCadences,"ClinicalOrthopaedics&RelatedResearch,vol.175,1983.
[33]D.A.Winter,S.E.Sienko,"Biomechanicsofbelow‐kneeamputeegait,"JournalofBiomechanics,vol.21,no.5,pp.361‐367,1988.
67[34]G.S.Sawicki,A.Domingo,D.P.Ferris,"Theeffectsofpoweredankle‐footorthoseson
jointkinematicsandmuscleactivationduringwalkinginindividualswithincompletespinalcordinjury,"JournalofNeuroEngineeringandRehabilitation,vol.28,no.3,2006.
[35]"HarminicDrive,"[Online].Available:http://harmonicdrive.net.[AccessedMarch2014].
[36]N.Sclater,N.P.Chironis,MechanismsandMechanicalDevicesSourcebook,McGraw‐Hill,2001.
[37]T.D.Tuttle,"UnderstandingandModelingtheBehaviorofaHarmonicDriveGearTransmission,"MassachusettsInstituteofTechnology,Cambridge,1992.
[38]S.Plagenhoefup,F.GaynorEvansb,T.Abdelnourb,"AnatomicalDataforAnalyzingHumanMotion,"ResearchQuarterlyforExerciseandSport,vol.54,no.2,pp.169‐178,1983.
[39]G.Boothroyd,AssemblyAutomationandProductDesign,Florida:TaylorandFrancis,2005.
[40]G.Boothroyd,P.Dewhurst,andW.Knight,ProductDesignforManufactureandAssembly,2nded.,NewYork:MarcelDekker,2002.
[41]W.Y.Tung,"DesignandOperationofMinimallyActuatedMedicalExoskeletonsforIndividualswithParalysis,"Ph.D.Dissertation:UniversityofCalifornia,Berkeley,Berkeley,2013.
[42]"LockheedMartin,"LockheedMartin,2014.[Online].Available:http://www.lockheedmartin.com/us/products/hulc.html.[AccessedApril2014].
[43]"Freedomofmotion:Walkingassistdevicewithstridemanagementassist,"2009.[Online].Available:http://corporate.honda.com/innovation/walk‐assist/StrideManagementAssist.pdf.[AccessedApril2014].
[44]"Honda,thepowerofdreams,"AmericanHondaMotorCo.,Inc.,2009.[Online].Available:http://corporate.honda.com/innovation/walk‐assist/.[AccessedApril2014].
[45]H.S.T.K.HideakiTakahashi,"Walkingassistancedevice".USAPatentUS8663134B2,10January2011.
68[46]G.Zeilig,H.Weingarden,M.Zwecker,I.Dudkiewicz,A.Bloch,andA.Esquenazi4,
"SafetyandtoleranceoftheReWalk™exoskeletonsuitforambulationbypeoplewithcompletespinalcordinjury:Apilotstudy,"TheJournalofSpinalCordMedicine,vol.35,no.2,pp.101‐96,2012.
[47]K.Kim,"OntheDesignofaUserInterfaceDeviceforaPoweredMedicalExoskeleton,"Master'sThesis:UniversityofCalifornia,Berkeley,Berkeley,2012.
[48]A.Ninio,J.Bruner,"Theachievementandantecedentsoflabelling,"JournalofChildLanguage,vol.5,no.01,pp.1‐15,1978.
[49]R.K.Sawyer,TheCambridgeHandbookoftheLearningSciences.,NewYork:CambridgeUniversityPress,2006.
[50]A.Cooper,TheInmatesAreRunningtheAsylum,Indianapolis:MacmillanPublishingCo.,Inc,1999.
[51]A.Cooper,R.Reimann,andD.Cronin,AboutFace3:TheEssentialsofInteractionDesign,Indianapoils:WileyPublishingInc.,2007.
[52]J.Grudin,J.Pruitt,"Personas,ParticipatoryDesignandProductDevelopment:AnInfrastructureforEngagement,"inParticipationandDesignConference,Sweden,Seite,2002.
[53]J.Pruitt,T.Adlin,Thepersonalifecycle:Keepingpeopleinmindthroughoutproductdesign,SanFrancisco:MorganKaufmann,2006.
[54]F.Long,"RealorImainary:Theeffectivenessofusingpersonasinproductdesign",inIrishErgonomicsReview,Dublin,2009.