Page 1: Development of Hard X-ray Detector with GEM

Development of Hard X-ray Detectorwith GEM

Shoji Uno (KEK)M. Sekimoto, T. Murakami, T. Uchida (KEK)

T. Koike (Tokyo University of Science)M. Kuroishi (Shinshu University)KEK Detector Technology Project

MPGD2009  Crete, Greece12 June 2009

Page 2: Development of Hard X-ray Detector with GEM

GEM foils made by Japanese company

• New method (plasma etching) was tried in several years ago. – Not chemical etching (CERN) – M. Inuzuka, et al., NIM A 525(2004) 529-534

• Laser + Plasma – To reduce sparks– T.Tamagawa, et al., NIM A560(2006) 418

• It is convenient for us to make new types of GEM foils.– Fine pitch/small hole : 50m/30m

• Better resolution Cosmic X-ray polarimeter• RIKEN

– Thicker/thinner : 100(150)m/ 25(12)m • Higher gas gain : a part of my talk

– S.Uno, et al., NIM A581 (2007) 271• Higher transparency : Gating grid for TPC

– Coated GEM• Boron : Neutron detection • Gold : Hard X-ray my talk• CsI : Hamamatsu Sekiya’s talk

10cm

10cm

Scienergy Co., Ltd. (Japanese company)http://www.scienergy.jp/

Page 3: Development of Hard X-ray Detector with GEM

Ar-CO2

P10

Triple GEM with 50mt

Single GEM with 100mt

100mmt GEM : Hole diameter 70m (90m)Pitch 140m

Page 4: Development of Hard X-ray Detector with GEM

Double GEM with 100mt GEM 70m

Page 5: Development of Hard X-ray Detector with GEM

Thickness(m)

VGEM1

(V)VGEM2

(V)Sum(V)

Gain Comments

100/100 557 557 1114 1.3x104

100/100 579 579 1158 4.3x104

100/100 651 506 1157 5.4x104

100/50 675 375 1050 4.8x104 Stable

100/50 650 400 1050 5.0x104 Stable

Triple 390 × 3 1170 3.0x104

Diameter:90m for 100mt GEM :70m for 50mt GEM

Combination of different thickness GEM

Page 6: Development of Hard X-ray Detector with GEM

Hard X-ray detector• Hard X-ray can even penetrate a thick iron plate and a concrete

block. Therefore, it is useful for non-destructive inspection in industrial purpose.

• Usually, Xenon gas has been used for detection of X-ray.

• But, efficiency is not so high for hard X-ray ( >~60keV).

• Solid converter (Gold) is used for detection of hard X-ray.

• Range of produced electron is very short. Gold layer should be thin. Then, many thin gold layers are necessary to get higher efficiency.

Page 7: Development of Hard X-ray Detector with GEM

Gold plated GEMto get higher efficiency

for hard X-ray with higher energy

Au   3m Cu   1m

Polyimide   50m

X-Rays

Electrons

Standard GEM+Gold Plating

Page 8: Development of Hard X-ray Detector with GEM

141keV γ-ray

0.00

0.05

0.10

0.15

0.20

0.25

0.30

0.00 0.05 0.10 0.15

Electron Energy (MeV)

Rel

ativ

e In

tens

ity

8

Simulation Study

Energy spectrum in gas volume for produced electron in gold layer

Page 9: Development of Hard X-ray Detector with GEM

0

5

10

15

20

25

30

0 10 20 30 40 50

Number of GEMs

Eff

icie

ncy(

%)

9

Optimization of Gold thickness

0.0

0.2

0.4

0.6

0 5 10 15 20 25

Thickness of Gold Layer (m)

Effic

ienc

y (%

)

Photon Energy: 141keV

1m

2m

3m

4m

5m

10m

Simulation

Page 10: Development of Hard X-ray Detector with GEM

Chamber structure

Au-GEM3m

2mm

2mm

Readout board X-Y strip0.8mm pitch 120+120

Multiplication

X-ray conversion layersEffective gas gain ~ 1

Gas:Ar-CO2(70-30)

Ground

100mt GEM1mm1mm1mm

1mm

1mm

  50mt GEM

Drift planeAu 3m

Page 11: Development of Hard X-ray Detector with GEM

0

1

2

3

4

0 1 2 3 4 5

Number of Au-GEM

Rel

ativ

e V

alu

e

11

Experimental result

57Co ( 122keV )

Thickness of Gold Layer: 3mon both surfaces

Gold coated GEMC

oun

tin

g ra

te (

rela

tive

val

ue)

Page 12: Development of Hard X-ray Detector with GEM

GEMChamber

• I/F– One HV cable– Five LV cables– One Ethernet cable

• Electronics– Four ASIC boards– One FPGA board

• FE2007 ASIC– By Y. Fujita (KEK)

• Data transfer and Control through Ethernet– SiTCP ( by T. Uchida)– Using Note-PC

Present Detector System

Low Voltage

Electronics

Note PC

Ethernet

Detector size 150mm×150mm×510mm

High Voltage

Compact and Portable System T.Uchida et. al., "Prototype of a Compact Imaging System for GEMdetectors," was published on IEEE TNS 55(2008)2698.

Page 13: Development of Hard X-ray Detector with GEM

0

0.2

0.4

0.6

0.8

0 20 40 60 80 100 120 140

Photon Energy (KeV)

Rel

ativ

e In

tens

ity

X-ray irradiation test

• Medical X-ray Generator

– Voltage   120kV

– Current   1mA 0

200

400

600

800

1000

1200

0 1 2 3 4 5

(mA)管電流値

(kHz)

計数

Current (mA)

Cou

ntin

g ra

te (

kHz)

Energy Spectrum

Data Taking Rate

Without filterWith Mo 0.3mmt

Max. rate1.1MHz

Page 14: Development of Hard X-ray Detector with GEM

2D Image for Hard X-ray

Clear 2D image could be seen.

96mm96

mm

Page 15: Development of Hard X-ray Detector with GEM

Pinhole ( 0.5mm ) Irradiation time:100sec

0.8mm

0.8m

m

Page 16: Development of Hard X-ray Detector with GEM

Iron Plate and Bars• On the Detector

1. Grid (Laminate structure of Lead and Aluminum)

To reduce scattering X-Rays 2. Iron plate : 10mm thickness 3. Four bars with different diameter

0.4mm2mm

0.9mm4.2mm

Page 17: Development of Hard X-ray Detector with GEM

Iron Platewith holes

9mm 8mm 7mm

6mm 5mm 4mm

3mm 2mm1mm

Depth:

Iron plate with 10mm thickness Hole diameter : 6mm

Page 18: Development of Hard X-ray Detector with GEM

Iron Bars inside Concrete

Thickness of Concrete :10cmIrradiation Time : 100sec

9~10mm 6mm 15~17mm

Four types of Iron Barsbefore molding

Concrete sample with Iron Bars

10cm

Page 19: Development of Hard X-ray Detector with GEM

Summary• Japanese company (Scienergy Co., Ltd.) can produce various

GEM foils.

• Flexible 100mt GEM is effective to get higher gas gain for single GEM structure.

• Gold coated GEM foils were manufactured.

– Higher efficiency was obtained with more GEM foils.

• A prototype of hard X-ray detector with gold coated GEM was

constructed.• Performance study was done with X-ray generator.

– Clear 2D image could be seen.

– Hard X-ray could be observed through the thick iron plate with 10mm thickness.

– Iron bars could be seen inside concrete block.