April 2015 Rev.1, 08/15 © 2015 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
725S Multifunction Process Calibrator
Users Manual
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that Service Center with a description of the problem. THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands
11/99
i
Table of Contents
Title Page
Introduction .................................................................................................................... 1 Contact Fluke ................................................................................................................. 1 Safety Information .......................................................................................................... 3 Symbols ......................................................................................................................... 5 Standard Equipment....................................................................................................... 6 Getting Acquainted with the Product .............................................................................. 8
Input and Output Terminals ....................................................................................... 8 Keys .......................................................................................................................... 10 Display ....................................................................................................................... 13
Configuration Menus ...................................................................................................... 14 Contrast Adjustment .................................................................................................. 14 Shut Down Mode ....................................................................................................... 15 CJC ........................................................................................................................... 15 Celcius and Fahrenheit (°C and °F) ........................................................................... 15 Frequency Pulse Output Voltage ............................................................................... 15 Pulse Output Frequency ............................................................................................ 15 HART Functionality .................................................................................................. 16 Step Time .................................................................................................................. 16 Ramp Time ................................................................................................................ 16
Get Started ..................................................................................................................... 16
725S Users Manual
ii
Voltage to Voltage Test ............................................................................................. 16 Measurement Mode ....................................................................................................... 18
Electrical Parameters (Upper Display) ...................................................................... 18 Current Measurement with Loop Power .................................................................... 18 Electrical Parameters (Lower Display) ...................................................................... 20 Temperature Measurements ..................................................................................... 21
Thermocouples .................................................................................................... 21 Resistance-Temperature Detectors (RTDs) ......................................................... 24
PRT Custom Curves ................................................................................................. 24 Pressure Measurements ........................................................................................... 27 Zeroing with Absolute Pressure Modules .................................................................. 28
Source Mode ................................................................................................................. 30 Source 4 mA to 20 mA .............................................................................................. 30 Simulate a 4 mA to 20 mA Transmitter ..................................................................... 30 Source Other Electrical Parameters .......................................................................... 32 Simulate Thermocouples .......................................................................................... 34 Simulate RTDs .......................................................................................................... 36 Source Pressure ....................................................................................................... 38
0 % and 100 % Output Parameters ............................................................................... 41 % Error Functionality ................................................................................................. 41
Step and Ramp the Output ............................................................................................ 41 Manually Step the mA Output ................................................................................... 42 Auto Ramp the Output .............................................................................................. 42
Store and Recall Setups ................................................................................................ 42 Store a Setup ............................................................................................................ 42 Recall a Setup........................................................................................................... 43
Store and Recall Data .................................................................................................... 43 Store Data ................................................................................................................. 43 Recall Data ............................................................................................................... 44
Contents (continued)
iii
Pulse Train Source/Read ............................................................................................... 44 Calibrate a Transmitter ................................................................................................... 45 Calibrate a P/I Transmitter ............................................................................................. 47 Calibrate an I/P Transmitter ........................................................................................... 49 Pressure Switch Test ..................................................................................................... 51 Test an Output Device .................................................................................................... 51 Remote Control Commands ........................................................................................... 52 Maintenance ................................................................................................................... 53
Replace the Batteries ................................................................................................ 54 Clean the Product ...................................................................................................... 54 Service Center Calibration or Repair ......................................................................... 54 Replacement Parts .................................................................................................... 54
Accessories .................................................................................................................... 56 External Fluke Pressure Module Compatibility .......................................................... 56
Specifications ................................................................................................................. 60 DC Voltage Measurement ......................................................................................... 60 DC Voltage Source .................................................................................................... 60 Millivolt Meaurement and Source .............................................................................. 60 DC mA Measurement and Source ............................................................................. 61 Ohms Measurement .................................................................................................. 61 Ohms Source ............................................................................................................ 62 Frequency Measurement ........................................................................................... 62 Frequency Source ..................................................................................................... 62 Temperature, Thermocouples ................................................................................... 63 Loop Power Supply ................................................................................................... 64 RTD Excitation (Simulation) ...................................................................................... 64 Temperature, RTD Ranges, and Accuracies (ITS-90) ............................................... 65 Pressure Measurement ............................................................................................. 65 General Specifications ............................................................................................... 66
725S Users Manual
iv
v
List of Tables
Table Title Page
1. Summary of Source and Measure Functions ........................................................................ 2 2. Symbols ................................................................................................................................. 5 3. Input/Output Terminals and Connectors ................................................................................ 9 4. Key Functions ........................................................................................................................ 11 5. Thermocouple Types Accepted ............................................................................................. 22 6. RTD Types Accepted ............................................................................................................ 25 7. mA Step Values ..................................................................................................................... 42 8. Replacement Parts ................................................................................................................ 54 9. Pressure Module Compatibility .............................................................................................. 56 10. Pressure Modules ................................................................................................................. 57
725S Users Manual
vi
vii
List of Figures
Figure Title Page
1. Standard Equipment .............................................................................................................. 7 2. Input/Output Terminals and Connectors ................................................................................ 8 3. Keys ...................................................................................................................................... 10 4. Elements of a Typical Display ............................................................................................... 13 5. Adjust the Contrast ................................................................................................................ 14 6. Voltage-to-Voltage Test ......................................................................................................... 17 7. Measure Voltage and Current Output .................................................................................... 18 8. Connections for Supplying Loop Power ................................................................................ 19 9. Measure Electrical Parameters ............................................................................................. 20 10. Measure Temperature with a Thermocouple ......................................................................... 23 11. Measure Temperature with an RTD, Measure 2-, 3-, and 4-Wire Resistance ....................... 26 12. Gage and Differential Pressure Modules ............................................................................... 27 13. Connections for Pressure Measurement ............................................................................... 29 14. Connections for Simulating a 4- to 20-mA Transmitter .......................................................... 31 15. Electrical Sourcing Connections ............................................................................................ 33
725S Users Manual
viii
16. Connections for Simulating a Thermocouple ........................................................................ 35 17. Connections for Simulating 3- and 4-Wire RTD .................................................................... 37 18. Connections for Sourcing Pressure ...................................................................................... 40 19. SAVE DATA Menu ................................................................................................................ 44 20. Calibrate a Thermocouple Transmitter.................................................................................. 46 21. Calibrate a Pressure-to-Current (P/I) Transmitter ................................................................. 48 22. Calibrate a Current-to-Pressure (I/P) Transmitter ................................................................. 50 23. Calibrate a Chart Recorder ................................................................................................... 52 24. Replace the Batteries............................................................................................................ 53 25. Replacement Parts ............................................................................................................... 55
1
Introduction The Fluke 725S Multifunction Process Calibrator (the Product) is a handheld, battery-operated instrument that measures and sources electrical and physical parameters. See Table 1. In addition to the functions in Table 1, the Product also has these features and functions: • A split-screen display. The upper display allows users
to measure volts, current, and pressure only. The lower display allows the user to measure and source volts, current, pressure, resistance temperature detectors, thermocouples, frequency, and ohms.
• A thermocouple (TC) input/output terminal and internal isothermal block with automatic reference-junction temperature compensation.
• Stores and recalls setups. • Manual and automatic stepping and ramping. • Stores and recalls calibration screens. • Remote control from a PC running a terminal
emulator program.
Contact Fluke To contact Fluke, call one of the following telephone numbers: • Technical Support USA: 1-800-44-FLUKE (1-800-
443-5853) • Calibration/Repair USA: 1-888-99-FLUKE (1-888-
993-5853) • Canada: 1-800-36-FLUKE (1-800-363-5853) • Europe: +31 402-675-200 • Japan: +81-03-6714-3114 • Singapore: +65-6799-5566 • Mainland China: +86-400-810-3435 • Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To see, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.
725S Users Manual
2
Table 1. Summary of Source and Measure Functions
Function Measure Source
dc V 0 V to 30 V 0 V to 20 V
dc mA 0 to 24 mA 0 to 24 mA
Frequency 2 CPM to 15 kHz 2 CPM to 15 kHz
Resistance 0 Ω to 4000 Ω 5 Ω to 4000 Ω
Thermocouple Types E, J, K, T, B, R, S, L, U, N, C, XK, BP
RTD
(Resistance-Temperature Detector)
Pt100 Ω (385) Pt100 Ω (3926) Pt100 Ω (3916) Pt200 Ω (385) Pt500 Ω (385) Pt1000 Ω (385)
Ni120 (672) CU10
Pressure 29 modules ranging from 1.0 in. H2O to 10,000 psi
Pulse 1-100,000
Freqency Max 15 kHz
1-10,000
Frequency Range 2 CPM to 15 kHz
Other functions Loop supply, HART resistor, pressure switch test, save screen, step, ramp, memory, cold junction compensation
Multifunction Process Calibrator Safety Information
3
Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
International electrical symbols used on the Product and in this manual are explained in Table 2.
Warning To prevent possible electrical shock, fire, or personal injury:
• Carefully read all instructions.
• Read all safety information before you use the Product.
• Use the Product only as specified, or the protection supplied by the Product can be compromised.
• Do not use the Product around explosive gas, vapor, or in damp or wet environments.
• Do not use the Product if it is damaged.
• Disable the Product if it is damaged.
• Examine the case before you use the Product. Look for cracks or missing plastic. Carefully look at the insulation around the terminals.
• The battery door must be closed and locked before you operate the Product.
• Remove all probes, test leads, and accessories before the battery door is opened.
• Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
• Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
• Do not apply more than the rated voltage, as marked on the Product, between the terminals, or between any terminals and earth ground (30 V max all terminals).
• Disconnect test leads from a source before changing function.
• Use the proper terminals, mode, and range for the measuring or sourcing applications.
725S Users Manual
4
• Turn off circuit power before connecting the Product’s mA and COM terminals to a circuit. Place the Product in series with the circuit.
• When using a pressure module, make sure the pressure line is shut off and depressurized before disconnecting it from the pressure module.
Caution To avoid possible damage to the Product or to equipment under test:
• Disconnect the power and discharge all high-voltage capacitors before testing resistance or continuity.
• The Product’s MEASURE/SOURCE terminals are ESD (electro-static discharge) sensitive to levels above ± 4 kV. The Product can experience temporary loss of measurement or source functionality, which may require operator intervention to restore product function, or even cause permanent damage. In general, a disruptive ESD event will only occur during connection of the test leads to the circuits being measured or if the operator is carrying a large static charge and touches the Calibrator terminals. The most common cause of ESD is the user carrying the Product across a carpet, or other similar triboelectric activity, before they connection to the circuit being measured.
Multifunction Process Calibrator Symbols
5
Symbols Table 2. Symbols
Symbol Description Symbol Description
AC (Alternating Current) DC (Direct Current)
Double insulated Pressure
Consult user documentation. Earth
WARNING.RISK OF DANGER. See Manual. WARNING. HAZARDOUS VOLTAGE. Risk of electric
shock.
Static awareness. Static discharge can damage part(s) Certified by CSA Group to North American safety
standards.
Conforms to relevant Australian EMC standards
Conforms to relevant South Korean EMC Standards.
Conforms to European Union directives.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste.
725S Users Manual
6
Standard Equipment If the Product is damaged or something is missing, contact the place of purchase immediately. To order replacement parts, see Table 8. The items listed below and shown in Figure 1 are included with the Product.
• TL75 test leads
• Stackable alligator clip test leads
• 725S Safety Sheet (not shown in Figure 1)
• 725S CD-ROM (contains Users Manual; not shown in Figure 1)
• 4 AA Batteries (installed)
Multifunction Process Calibrator Standard Equipment
7
mA+
mA-
TRIGGER/STOP
StackableTest Leads
TL75Test Lead Set
hxl01.eps
Figure 1. Standard Equipment
725S Users Manual
8
Getting Acquainted with the Product Input and Output Terminals
Figure 2 shows the Product input and output terminals. Table 3 explains their use.
8
7
6 5 43
2
1
hxl05.eps
Figure 2. Input/Output Terminals and Connectors
Multifunction Process Calibrator Getting Acquainted with the Product
9
Table 3. Input/Output Terminals and Connectors
Item Name Description
Pressure module connector/serial connector
Connects the Product to a pressure module or to a PC for a remote control serial connection.
, MEASURE V, mA terminals Input terminals for measuring voltage, current, supplying loop power, HART resistance, switch test options.
Thermocouple (TC) input/output
Terminal for measuring or simulating thermocouples. This terminal accepts a miniature polarized thermocouple plug with flat, in-line blades spaced 7.9 mm (0.312 in) center to center.
,
SOURCE/ MEASURE V, RTD, Pulse, Hz, Ω terminals
Terminals for sourcing or measuring voltage, resistance, pulse, frequency, and RTDs.
, SOURCE/ MEASURE mA terminals, 3W, 4W
Terminals for sourcing and measuring current and performing 3 W and 4 W RTD measurements. HART resistor option in mA mode.
725S Users Manual
10
Keys Figure 3 shows the Product keys and Table 4 explains their use.
2
1
4
3 20
19
18
17
1514
13
12 11
10
9
8
7
6
16
hxl41.eps
Figure 3. Keys
Multifunction Process Calibrator Getting Acquainted with the Product
11
Table 4. Key Functions
No Name Description
Turns the power on or off.
%Error
Toggles voltage, mA, or Loop Power and % Error measurement functions in the upper display.
ZERO
3 Seconds
Selects the pressure measurement function in the upper display. Repeated pushes cycle through the different pressure units. Zeros pressure when held for 3 seconds.
OPEN/CLOSE Activates the switch test.
Turns backlight on or off.
Selects frequency sourcing or measurement.
Selects pulse sourcing or measurement.
EXIT CONFIG
Recalls a source value from memory corresponding to 100 % of span and sets it as the source value. Push and hold to store the source value as the 100 % value. Exits Configuration Menu.
Increments output by 25 % of span.
Decrements output by 25 % of span.
Recalls from memory a source value corresponding to 0 % of span and sets it as the source value. Push and hold to store the source value as the 0 % value. Push and hold when powering up to identify the firmware version. The firmware version is shown in the upper display for about 1 second after initialization.
725S Users Manual
12
Table . Key Functions (cont.)
No Name Description
TRIGGER/STOP
Cycles through : Slow repeating 0 % - 100 % - 0 % ramp Configurable repeating 0 % - 100 % - 0 % ramp Configurable repeating 0 % - 100 % - 0 % ramp in 25 % steps Used for the pulse train and totalizer functions.
Return to Recall
Increases or decreases the source level. Cycles through the 2-, 3-, and 4-wire selections. Moves through the memory locations of Product setups. Moves through the configuration menus.
ENTER
Saves and recalls setups & data. ENTER is used in the configuration menus.
Used to enter and navigate the configuration menus.
Cycles the Product through MEASURE and SOURCE modes in the lower display.
Selects TC (thermocouple) measurement and sourcing function in the lower display. Repeated pushes cycle through the thermocouple types.
HART
Toggles between voltage, mA sourcing, or mA simulate functions in the lower display. Inserts a 250 Ω resistor when in mA.
Selects RTD (resistance temperature detector) measurement and sourcing function in lower display. Repeated pushes cycle through the RTD types. Selects resistance mode.
Selects the pressure measurement and sourcing function. Repeated pushes cycle through the different pressure units.
Multifunction Process Calibrator Getting Acquainted with the Product
13
Display
Figure 4 shows the elements of a typical display.
LoopIndicator
Memory Locationsfor Calibrator Setups
UnitsDisplay
Auto Ramp
ModeIndicator
ModeIndicator
Low BatterySymbol
HART ModeIndicator
% ErrorIndicators
hxl07.eps
Figure 4. Elements of a Typical Display
725S Users Manual
14
Configuration Menus Use the configuration menus to set or change these parameters of the Product:
• Contrast Adjustment
• Shut Down Mode
• CJC on/off
• °C/°F
• Frequency/Pulse output voltage
• Pulse output frequency
• HART resistor on/off
• Step time
• Ramp time
To enter the configuration menus, push . Push , to save new configuration. Push /EXIT CONFIG to exit configuration.
Configuration menus are explained below.
Contrast Adjustment
To adjust the contrast (see Figure 5):
1. Push until Contst Adjust appears on the display.
2. Use and to adjust the contrast up and down.
3. Push to save the setting.
1
3
2
hxl06.eps
Figure 5. Adjust the Contrast
Multifunction Process Calibrator Configuration Menus
15
Shut Down Mode
The Product has a shut down mode set for 30 minutes (displayed for about 1 second when the Product is initially turned on). When shut down mode is enabled, the Product automatically shuts down after the elapsed time from when the last key was pushed.
1. Push until SHUT DOWN appears on the screen.
2. Use and to increase or decrease the time.
3. Use and to turn on and off.
4. Push to save the setting.
CJC
Cold Junction Compensation (CJC) is a value for the cold end of a thermocouple at the Meter’s end.
1. Push until SELECT CJC appears on the display.
2. Use and to select ON or OFF.
3. Push to save the setting.
Celcius and Fahrenheit (°C and °F)
1. Push until SELECT UNIT °C (or °F) appears on the display.
2. Use and to select °C or °F.
3. Push to save the setting.
Frequency Pulse Output Voltage
1. Push until FREQ OUTPUT V Adjust appears on the display.
2. Use , , and to adjust the frequency pulse output voltage from 1 to 20 V.
3. Push to save the setting.
Pulse Output Frequency
1. Push until PULSE OUTPUT Hz FREQ Adjust appears on the display.
2. Use , , and to adjust the pulse output frequency from 2 CPM to 15 kHz.
3. Push to save the setting.
725S Users Manual
16
HART Functionality The Product has a user-selectable 250 Ω HART to facilitate use with HART communication devices. The resistor can be switched in or out using config selection menus. Use a HART communicator when measuring mA with loop power or sourcing mA.
To put HART Resistor ON/OFF:
1. Push until SELECT HART ON or OFF appears on the display.
2. Use or and to toggle ON and OFF.
3. Push to save the setting.
Note When HART mode is selected, the 250 Ω resistor is turned on both mA channels.
Step Time Step time sets the ramp step time from 1 second to 99 seconds. 1. Push until STEP TIME shows on the screen. 2. Use and to set the step time. 3. Push to save the setting. Ramp Time Ramp time set the ramp time from 5 seconds to 99 seconds. 1. Push until RAMP TIME shows on the screen. 2. Use and to set the ramp time. 3. Push to save the setting.
Get Started This section details some basic operations of the Product.
Voltage to Voltage Test
To perform a voltage-to-voltage test:
1. Connect the Product’s voltage output to its voltage input as shown in Figure 6.
2. Push to turn on the Product. Push to select dc voltage (upper display).
3. If necessary, push for SOURCE mode (lower display). The Product is still measuring dc voltage, the active measurements are visible in the upper display.
4. Push to select dc voltage sourcing.
5. Push and to select a digit to change. Push to select 1 V for the output value. Push and hold to enter 1 V as the 0 % value.
6. Push to increase the output to 5 V. Push and hold to enter 5 V as the 100 % value.
7. Push and to step between 0 and 100 % in 25 % step increments.
Multifunction Process Calibrator Get Started
17
100%
25%
25%
MEAS
SOURCE
0%
SAVERECALL
FREQ
ENTER
EXIT CONFIG
HART
TC RTD PULSE
V mA
V mALOOP SWITCH
TEST
%ErrorZERO
3 Seconds OPEN/CLOSE
TRIGGER/STOP
CONFIGSELECTION
hxl39.eps
Figure 6. Voltage-to-Voltage Test
725S Users Manual
18
Measurement Mode Electrical Parameters (Upper Display)
To measure the current or voltage output of a transmitter, or to measure the output of a 750 Series pressure module, use the upper display and proceed as follows:
1. Push to select volts or current. LOOP should not be on.
2. Connect the leads as shown in Figure 7.
Current Measurement with Loop Power
The loop power function activates a 24 V supply in series with the current measuring circuit, allowing you to test a transmitter when it is disconnected from plant wiring. To measure current with loop power:
1. Connect the Product to the transmitter current loop terminals as shown in Figure 8.
2. Push while the Product is in current measurement mode. LOOP appears and an internal 24 V loop supply turns on.
Note
When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels.
hxl42.eps
Figure 7. Measure Voltage and Current Output
Multifunction Process Calibrator Measurement Mode
19
Red
TEST DC PWR+ –+–
–+
BlackHARTCommunicator
hxl18.eps
Figure 8. Connections for Supplying Loop Power
725S Users Manual
20
Electrical Parameters (Lower Display)
To measure electrical parameters using the lower display, proceed as follows:
1. Connect the Product as shown in Figure 9.
2. If necessary, push for MEASURE mode (lower display).
3. Push for dc voltage or current, for frequency, and for resistance.
Black
Red
hxl43.eps
Figure 9. Measure Electrical Parameters
Multifunction Process Calibrator Measurement Mode
21
Temperature Measurements
Thermocouples
The Product supports 13 standard thermocouples. Table 5 summarizes the ranges and characteristics of each.
To measure temperature using a thermocouple:
1. Select Celsius or Fahrenheit, depending on the desired measurement. Refer to “Configuration Menus” for more information.
2. Attach the thermocouple leads to the appropriate TC miniplug, then to the TC input/output as shown in Figure 10.
Caution One pin is wider than the other. Do not try to force a miniplug into the wrong polarization.
Note
If the Product and the thermocouple plug are at different temperatures, wait one minute or more for the connector temperature to stabilize after plugging the miniplug into the TC input/output.
3. If necessary, push for MEASURE mode.
4. Push for the thermocouple display. Continue pushing this key to select the desired thermocouple type.
725S Users Manual
22
Table 5. Thermocouple Types Accepted
Type Positive Lead Material Positive Lead (H)
Color Negative Lead Material Specified Range (°C) ANSI* IEC**
E Chromel Purple Violet Constantan -200 to 950
N Ni-Cr-Si Orange Pink Ni-Si-Mg -200 to 1300
J Iron White Black Constantan -200 to 1200
K Chromel Yellow Green Alumel -200 to 1370
T Copper Blue Brown Constantan -200 to 400
B Platinum (30 % Rhodium) Gray Platinum (6 % Rhodium) 600 to 1800
R Platinum (13 % Rhodium) Black Orange Platinum -20 to 1750
S Platinum (10 % Rhodium) Black Orange Platinum -20 to 1750
L Iron Constantan -200 to 900
U Copper Constantan -200 to 400
C Tungsten 5% Rhenium White None Tungsten 26% Rhenium 0 to 2316
BP 90.5 % Ni + 9.5 % Cr GOST
56 % Cu + 44 % Ni -200 to 800 Violet or Black
XK 95 % W + 5 % Re Red or Pink 80 % W + 20 % Re 0 to 2500
*American National Standards Institute (ANSI) device negative lead (L) is always red.
**International Electrotechnical Commission (IEC) device negative lead (L) is always white.
Multifunction Process Calibrator Measurement Mode
23
Warning
30 V maximum to
TC
hxl12.eps
Figure 10. Measure Temperature with a Thermocouple
725S Users Manual
24
Resistance-Temperature Detectors (RTDs)
The Product accepts RTD types shown in Table 6. RTDs are characterized by their resistance at 0 °C (32 °F), which is called the “ice point” or R0. The most common R0 is 100 Ω. The Product accepts RTD measurement inputs in two-, three-, or four-wire connections, with the three-wire connection the most common. A four-wire configuration provides the highest measurement precision, and two-wire provides the lowest measurement precision.
To measure temperature using an RTD input:
1. If necessary, push for MEASURE mode.
2. Push for the RTD display. Continue pushing this key to select the desired RTD type.
3. Push or to select a 2-, 3-, or 4- wire connection.
4. Attach the RTD to input terminals as shown in Figure 11.
PRT Custom Curves
Up to three custom curves can be named and CVD coefficients can be entered through the serial port. Names can be up to six characters. For more information, see the Application Note on the 725S CD.
Multifunction Process Calibrator Measurement Mode
25
Table 6. RTD Types Accepted
RTD Type Ice Point (R0) Material α Range (°C)
Pt100 (3926) 100 Ω Platinum 0.003926 Ω/°C -200 to 630
Pt100 (385) 100 Ω Platinum 0.00385 Ω/°C -200 to 800
Ni120 (672) 120 Ω Nickel 0.00672 Ω/°C -80 to 260
Pt200 (385) 200 Ω Platinum 0.00385 Ω/°C -200 to 630
Pt500 (385) 500 Ω Platinum 0.00385 Ω/°C -200 to 630
Pt1000 (385) 1000 Ω Platinum 0.00385 Ω/°C -200 to 630
Pt100 (3916) 100 Ω Platinum 0.003916 Ω/°C -200 to 630
Cu10 10 Ω Copper 0.0042 Ω/°C -100 to 250
The IEC standard RTD commonly used in U.S. industrial applications is the Pt100 (385), α = 0.00385 Ω/°C.
Pt100 (3916), α = 0.003916 Ω/°C is also designated as JIS curve.
Custom RTDs may also be added, see PRT Custom Curves
725S Users Manual
26
2W
3W
4W
hxl15.eps
Figure 11. Measure Temperature with an RTD, Measure 2-, 3-, and 4-Wire Resistance
Multifunction Process Calibrator Measurement Mode
27
Pressure Measurements
Many ranges and types of pressure modules are available from Fluke, see “Accessories”. Before using a pressure module, read its instruction sheet. The modules vary in use, media, and accuracy.
Figure 12 shows the gage and differential modules. Differential modules also work in gage mode by leaving the low fitting open to atmosphere.
To measure pressure, attach the appropriate pressure module for the process pressure to be tested, and proceed as follows:
Warning To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before attaching the pressure module to the pressure line.
15 PSID/G
BURST PRESSURE 45 PSIG
PRESSURE MODULE700P04
1 bar100 kPaRANGERANGE 100 PSIG
BURST PRESSURE 300 PSIG
PRESSURE MODULE700P06
7 bar700 kPa
gj11f.eps
Figure 12. Gage and Differential Pressure Modules
725S Users Manual
28
Caution To avoid mechanically damaging the pressure module:
• Never apply more than 10 ft.-lb. (13.5 Nm) of torque between the pressure module fittings, or between the fittings and the body of the module. Always apply appropriate torque between the pressure module fitting and connecting fittings or adapters.
• Never apply pressure above the rated maximum printed on the pressure module.
• Only use the pressure module with specified materials. Refer to the printing on the pressure module or the pressure module instruction sheet for the acceptable material compatibility.
1. Connect a pressure module to the Product as shown in Figure 13. The threads on the pressure modules accept standard ¼ NPT pipe fittings. Use the supplied ¼ NPT to ¼ ISO adapter if necessary.
2. Push either or . The Product automatically senses which pressure module is attached and sets its range accordingly.
3. Zero the pressure module as described in the module’s Instruction Sheet. Modules vary in zeroing procedures depending on module type, but all require pushing for 3 seconds.
Continue pushing to change pressure display units to psi, mmHg, inHg, cmH2O@4 °C, cmH2O@20 °C, inH2O@4 °C, inH2O@20 °C, inH2O@60 °F, mbar, bar, kg/cm2, or kPa.
Zeroing with Absolute Pressure Modules
To zero, adjust the Product to read a known pressure. This can be barometric pressure, if it is accurately known, for all but the 750PA3 module. The maximum range of 750PA3 is 5 psi; therefore the reference pressure must be applied with a vacuum pump. An accurate pressure standard can also apply a pressure within range for any absolute pressure module. To adjust the Product reading, proceed as follows:
1. Push , REF Adjust appears to the right of the pressure reading.
2. Use to increase or to decrease the Product reading to equal the reference pressure.
3. Push again to exit the zeroing procedure.
The Product stores and automatically reuses the zero offset correction for one absolute-pressure module so that the module is not rezeroed every time you use it.
Multifunction Process Calibrator Measurement Mode
29
DifferentialModule
Tank
HL
IsolationValve
GageModule
hxl37.eps
Figure 13. Connections for Pressure Measurement
725S Users Manual
30
Source Mode In SOURCE mode, the Product:
• generates calibrated signals for testing and calibrating process instruments
• supplies voltages, currents, frequencies, and resistances
• simulates the electrical output of RTD and thermocouple temperature sensors
• measures gas pressure from an external source, creating a calibrated pressure source.
Source 4 mA to 20 mA
To select the current sourcing mode, proceed as follows:
1. Connect the test leads in the mA terminals (left column).
2. If necessary, push for SOURCE mode.
3. Push for current and enter the desired current by pushing ,,, and .
Simulate a 4 mA to 20 mA Transmitter
Simulate is a special mode of operation in which the Product is connected into a loop in place of a transmitter and supplies a known, settable test current. Proceed as follows:
1. Connect the 24 V loop power source as shown in Figure 14.
2. If necessary, push for SOURCE mode.
3. Push until both mA and SIM display.
4. Enter the desired current by pushing ,,, and .
Multifunction Process Calibrator Source Mode
31
LoopPowerSupply
Readout orController
+24 V
Black
Red
+
–
hxl17.eps
Figure 14. Connections for Simulating a 4- to 20-mA Transmitter
725S Users Manual
32
Source Other Electrical Parameters
Volts, ohms, and frequency are also sourced and shown in the lower display.
To select an electrical sourcing function, proceed as follows:
1. Connect the test leads as shown in Figure 15, depending on the source function.
2. If necessary, push for SOURCE mode.
3. Push for dc voltage, or for frequency, and for resistance.
4. Enter the desired output value by pushing and keys. Push and to select a different digit to change.
Multifunction Process Calibrator Source Mode
33
Red
Black
+
–
+
–
Red
Black
V
mA
Hz
HARTCommunicator
Common
Common
hxl16.eps
Figure 15. Electrical Sourcing Connections
725S Users Manual
34
Simulate Thermocouples
Connect the Product TC input/output to the instrument under test with the thermocouple wire and the appropriate thermocouple mini-connector (polarized thermocouple plug with flat, in-line blades spaced 7.9 mm [0.312 in] center to center). One pin is wider than the other.
Caution Do not try to force a miniplug into the wrong polarization.
Figure 16 shows this connection. Proceed as follows to simulate a thermocouple:
1. Attach the thermocouple leads to the appropriate TC miniplug, then to the TC input/output as shown in Figure 16.
2. If necessary, push for SOURCE mode.
3. Push for the TC display. If desired, continue pushing this key to select the desired thermocouple type.
4. Enter the temperature you want by pushing and keys. Push and to select a different digit to edit.
Multifunction Process Calibrator Source Mode
35
TEST DC PWR++–
–+
Color dependson type of TC
TC Miniplug
hxl20.eps
Figure 16. Connections for Simulating a Thermocouple
725S Users Manual
36
Simulate RTDs
Connect the Product to the instrument under test as shown in Figure 17. Proceed as follows to simulate an RTD:
1. If necessary, push for SOURCE mode.
2. Push for the RTD display.
Note
Use the 3W and 4W terminals for measurement only, not for simulation. The Product simulates a 2-wire RTD at its front panel. To connect to a 3-wire or 4-wire transmitter, use the stacking cables to provide the extra wires. See Figure 17.
3. Enter the desired temperature by pushing and . Push and to select a different digit to edit.
4. If the 725S display indicates ExI HI, the excitation current from your device under test exceeds the limits of the 725S.
Multifunction Process Calibrator Source Mode
37
100%
25%
25%
MEAS
SOURCE
0%
SAVERECALL
FREQ
ENTER
EXIT CONFIG
HART
TC RTD PULSE
V mA
V mALOOP SWITCH
TEST
%ErrorZERO
3 Seconds OPEN/CLOSE
TRIGGER/STOP
CONFIGSELECTION
Return to Recall
1
234
hxl40.eps
Figure 17. Connections for Simulating 3- and 4-Wire RTD
725S Users Manual
38
Source Pressure
The Product sources pressure by measuring pressure supplied by a pump or other sources, and displaying the pressure in the SOURCE field. Figure 18 shows how to connect a pump to a Fluke pressure module which makes it a calibrated source.
Many ranges and types of pressure modules are available from Fluke, see “Accessories”. Before using a pressure module, read its instruction sheet. The modules vary in use, media, and accuracy.
Attach the appropriate pressure module for the process pressure to be tested.
Proceed as follows to source pressure:
Warning To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before attaching the pressure module to the pressure line.
Caution To avoid mechanically damaging the pressure module:
• Never apply more than 10 ft.-lb. (13.5 Nm) of torque between the pressure module fittings, or between the fittings and the body of the module. Always apply appropriate torque between the pressure module fitting and connecting fittings or adapters.
• Never apply pressure above the rated maximum printed on the pressure module.
• Use the pressure module only with specified materials. Refer to the printing on the pressure module or the pressure module instruction sheet for the acceptable material compatibility.
Multifunction Process Calibrator Source Mode
39
1. Connect a pressure module to the Product as shown in Figure 18. The threads on the pressure modules accept standard ¼ NPT pipe fittings. Use the supplied ¼ NPT to ¼ ISO adapter if necessary.
2. Push (lower display). The Product automatically senses which pressure module is attached and sets its range accordingly.
3. Zero the pressure module as described in the module’s Instruction Sheet. Modules vary in zeroing procedures depending on module type.
4. Pressurize the pressure line with the pressure source to the desired level as shown on the display.
If desired, continue pushing to change pressure display units to psi, mmHg, inHg, cmH2O@4 °C, cmH2O@20 °C, inH2O@4 °C, inH2O@20 °C, inH2O@60 °C, mbar, bar, kg/cm2, or kPa.
Note
The units switching function is available only when the pressure module is connected to the Product.
725S Users Manual
40
Hand Pump
PressureModule
Measure mASource Pressure 24 V Loop PowerEnabled
0%
To Unit Under Test
hxl47.eps
Figure 18. Connections for Sourcing Pressure
Multifunction Process Calibrator
41
0 % and 100 % Output Parameters For current output, the Product assumes that 0 % corresponds to 4 mA and 100 % corresponds to 20 mA. For other output parameters, 0 % and 100 % points must be set before using the step and ramp functions. Proceed as follows:
1. If necessary, push for SOURCE mode.
2. Select the desired source function and use the arrow keys to enter the value. This example is temperature source using 100 °C and 300 °C values for source.
3. Enter 100 °C then push and hold to store the value.
4. Enter 300 °C then push and hold to store the value.
This setting can now be used for the following:
• Manually stepping an output in 25 % increments.
• Switch between the 0 % and 100 % span points by momentarily pushing or .
% Error Functionality Percentage Error is available for every range on the lower display. The calculation is based on a mA percentage deviation from the value sourced on the lower display 0 % mA and 100 % mA are fixed to 4 and 20 mA. 0 % and 100 % for the lower display are set in source using and , refer to “Setting 0% and 100% Output Parameters”.
Step and Ramp the Output Two additional features are available for adjusting the value of source functions:
• Stepping the output manually with the and keys, or in automatic mode
• Ramping the output
Stepping and ramping apply to all functions except pressure, which requires use of an external pressure source.
725S Users Manual
42
Manually Step the mA Output
To manually step current output:
• Use or to step the current up or down in 25 % steps.
• Touch either to go to 0 %, or to go to 100 %.
Auto Ramp the Output Auto ramping can continuously apply a varying stimulus from the Product to a transmitter, while your hands remain free to test the response of the transmitter.
When is pushed, the Product produces a repeating 0 % - 100 % - 0 % ramp in a choice of three ramp waveforms:
• 0 % - 100 % - 0 % 40-second smooth ramp
• 0 % - 100 % - 0 % configurable time smooth ramp. Use configuration menu to set ramp time.
• 0 % - 100 % - 0 % Stair-step ramp in 25 % steps, pausing 5 seconds at each step. Use configuration menu to set step time Steps are listed in Table 7.
To exit ramping, push any button.
Table 7. mA Step Values
Step 4 to 20 mA
0 % 4.000
25 % 8.000
50 % 12.000
75 % 16.000
100 % 20.000
Store and Recall Setups Up to eight settings can be stored in a nonvolatile memory to recall for later use. A low battery condition or a battery change does not jeopardize the stored settings.
Store a Setup To store a setup:
1. Create the desired setup.
2. Push . The right side of the display changes to show SAVE SETUP, SAVE DATA, RECALL SETUP, and RECALL DATA..
3. Push or to highlight SAVE SETUP, and push .
4. Push or to select the desired memory location (at the top of the LCD).
5. Push to save the setup.
Multifunction Process Calibrator Store and Recall Data
43
Recall a Setup
To recall a setup:
1. Push . The right side of the screen changes to show SAVE SETUP, SAVE DATA, RECALL SETUP, and RECALL DATA.
2. Push or to highlight RECALL SETUP, and push to select RECALL SETUP.
3. Push or to select the necessary memory location (at the top of the screen).
4. Push to recall the setup from the proper memory location.
Store and Recall Data Up to 40 data samples can be stored in a nonvolatile memory to be recalled for later use. A low battery condition or a battery change does not jeopardize the stored settings.
Store Data
To store measurement data, use the following procedure, refer to Figure 19.
1. Take the desired measurement.
2. Push . The right side of the display changes to show SAVE SETUP, SAVE DATA, RECALL SETUP, and RECALL DATA.
3. Push or to highlight SAVE DATA, and push to select SAVE DATA.
4. Use or to change memory location (1-8). Use or to change the data point location (1-5).
5. Push to store the measurement and return the unit to measurement mode. Figure 19 shows memory location 1, data point 3 (3-5).
725S Users Manual
44
Memory Locations
Data Point
Save Data Mode Indicator
hxl46.eps
Figure 19. SAVE DATA Menu
Recall Data
To recall data:
1. Push . The right side of the display changes to show SAVE SETUP, SAVE DATA, RECALL SETUP, and RECALL DATA.
2. Push or to highlight RECALL DATA, and push to select RECALL DATA.
3. Use or to select the desired memory location (1-8) and use or to select data point (1-5).
4. Push to recall the data stored in that location.
Pulse Train Source/Read Pulse Train Source/Read counts input pulses or sources output pulses. Use the configuration menus to set the
frequency and output voltage. Refer to “Configuration Menus” earlier in this manual. The number of counts is set through the main display and cannot be changed while sourcing pulses. works as a trigger/stop key in this mode since ramping or stepping during a pulse train is not relevant.
Multifunction Process Calibrator Calibrate a Transmitter
45
Calibrate a Transmitter Use the measurement (upper display) and source (lower display) modes to calibrate a transmitter. This section applies to all but pressure transmitters. The following example shows how to calibrate a temperature transmitter. Use the following steps to calibrate a transmitter:
1. Connect the Product to the instrument under test as shown in Figure 20.
2. Push for current (upper display). If required, push again to activate loop power.
3. Push (lower display). If desired, continue pushing this key to select the desired thermocouple type.
4. If necessary, push for SOURCE mode.
5. Set the zero and span parameters by pushing and . Enter these parameters by pushing and holding and . For more information on setting parameters, see “Setting 0 % and 100 % Output Parameters”.
6. Perform test checks at 0-25-50-75-100 % points by pushing or . Adjust the transmitter as necessary.
Note
When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels.
725S Users Manual
46
Red
TEST DC PWR+ –+–
–+
Black
HARTCommunicator
hxl44.eps
Figure 20. Calibrate a Thermocouple Transmitter
Multifunction Process Calibrator Calibrate a P/I Transmitter
47
Calibrate a P/I Transmitter The following steps explain calibrating a pressure transmitter.
1. Connect the Product to the instrument under test as shown in Figure 21.
2. Push for current (upper display). If required, push again to activate loop power.
3. Push (lower display).
4. If necessary, push for SOURCE mode.
5. Zero the pressure module.
6. Perform checks at 0 % and 100 % of span and adjust the transmitter as necessary.
Note
When HART resistor mode is selected, the 250 Ω resistor is turned on both mA channels.
725S Users Manual
48
Hand Pump
S I G N A L
T E S T
+ –
PressureModule
Red
Black
Measure mASource Pressure 24 V Loop PowerEnabled
HARTCommunicator
hxl34.eps
Figure 21. Calibrate a Pressure-to-Current (P/I) Transmitter
Multifunction Process Calibrator Calibrate an I/P Transmitter
49
Calibrate an I/P Transmitter The following steps explain how to calibrate a device that controls pressure. Proceed as follows:
1. Connect the test leads to the instrument under test as shown in Figure 22. The connections simulate a current-to-pressure transmitter and measures the corresponding output pressure.
2. Push (upper display).
3. Push for sourcing current (lower display).
4. If necessary, push for SOURCE mode.
5. Enter the desired current by pushing and . Push and to select different digits.
725S Users Manual
50
S I G N A L
T E S T
+ –
Red
Black
MeasurePressure Source mA
HARTCommunicator
PressureModule
hxl28.eps
Figure 22. Calibrate a Current-to-Pressure (I/P) Transmitter
Multifunction Process Calibrator Pressure Switch Test
51
Pressure Switch Test
Note
This example uses a normally-closed switch. The procedure is the same for an open switch but the display reads OPEN instead of CLOSE.
To perform a switch test:
1. Connect the Product mA and COM terminals to the switch using the pressure switch terminals and connect the pump to the pressure switch. The polarity of the terminals does not matter.
2. Make sure the vent on the pump is open and, if necessary, zero the Product. Close the vent after zeroing the Product.
3. Push to enter switch-test mode. The upper display indicates the applied pressure. CLOSE is displayed to the right of the pressure reading to indicate closed contacts.
4. Slowly apply pressure with the pump until the switch opens.
Note
Apply pressure to the device slowly to ensure accurate readings. Run the test several times to confirm repeatability.
OPEN displays once the switch is open. Slowly bleed the pump until the pressure switch closes. RECALL appears on the display.
5. Push to read the pressure values for when the switch opened, for when it closed, and for the deadband.
6. Hold for three seconds to restart the test. Push OR to exit the switch test.
Test an Output Device Use the source functions to test and calibrate actuators, recording, and indicating devices. Proceed as follows:
1. Connect the test leads to the instrument under test as shown in Figure 23.
2. Push for current or dc voltage, or for frequency or resistance (lower display).
3. If necessary, push for SOURCE mode.
725S Users Manual
52
hxl25.eps
Figure 23. Calibrate a Chart Recorder
Remote Control Commands You can remotely control the Product from a PC running a terminal emulator program. The remote control commands give access to all capabilities of the Product with the exception of pressure measurement.
Go to www.fluke.com/processtools to find the Application Note about 725S Remote Programming..
.
Multifunction Process Calibrator Maintenance
53
Maintenance
Warning To prevent possible electrical shock, fire, or personal injury: • Remove the input signals before you
clean the Product. • Repair the Product before use if the
battery leaks. • Use only specified replacement parts. • Have an approved technician repair the
Product.
bec38f.eps
Figure 24. Replace the Batteries
725S Users Manual
54
Replace the Batteries
Warning To avoid false readings, which could lead to possible electric shock or personal injury, replace the batteries as soon as the battery indicator () appears.
Figure 24 shows you how to replace the batteries.
Clean the Product
Caution To avoid damaging the plastic lens and case, do not use solvents or abrasive cleansers.
Clean the Product and pressure modules with a soft cloth dampened with water or water and mild soap.
Service Center Calibration or Repair
Calibration, repairs, or servicing not covered in this manual should be performed only by qualified service personnel. If the Product fails, check the batteries first, and replace them if needed.
To locate an authorized service center, refer to “Contact Fluke” at the beginning of the manual.
Replacement Parts Table 8 lists the part number of each replaceable part. Refer to Figure 25.
Table 8. Replacement Parts
Item Description PN Qty.
1 AA alkaline batteries 376756 4
2 Case screws 832246 4
3 Battery door 664250 1
4 Accessory mount 658424 1
5 Tilt stand 659026 1
6 Battery door ¼-turn fasteners
948609 2
7 TL75 series test leads 855742 1
8 Fluke-7XX Test Lead Set 3397308 1
9 725S Safety Sheet 4629433 1
10 725S CD ROM, contains Users Manual
4629478 1
Multifunction Process Calibrator Maintenance
55
9
3
8
7
1
4
5
10
2
6
hxl45.eps
Figure 25. Replacement Parts
725S Users Manual
56
Accessories For more information about these accessories, contact your Fluke representative. Fluke Pressure Module Compatibility is listed in Table 9. Pressure Modules and Fluke model numbers are listed in Table 10. Contact your Fluke representative about new pressure modules not listed here. • 750HTP 0 to 10,000 PSI Pump • 750PTP -11.6 to 360 PSI Pump • 750TC1 and 750TC2 Thermocouple Mini-plug Kits
External Fluke Pressure Module Compatibility
The output of Fluke 750P pressure modules can cause the 5-digit display to overflow, or produce values that are too low to be read if inappropriate units are selected. To avoid this, the Product shows OL on the screen. See Tables 9 and 10.
Table 9. Pressure Module Compatibility
Pressure Unit Module Compatibility
Psi Available on all pressure ranges
In. H20 All ranges through 3000 psi
cm. H20 All ranges through 1000 psi
Bar 15 psi and above
Mbar All ranges through 1000 psi
Kpa Available on all pressure ranges
In.Hg. Available on all pressure ranges
mm. Hg All ranges through 1000 psi
Kg/cm2 15 psi and above
Multifunction Process Calibrator Accessories
57
Table 10. Pressure Modules
Model Parameter/Range Burst Rating Hi Side Media Lo Side Media
Differential
750P00 0 to 1 in H2O (0 to 2.5 mBar) 30X Dry Air Dry Air
750P01 0 to 10 in H2O (0 to 25 mBar) 3X Dry Air Dry Air
750P02 0 to 1 psi (0 to 70 mBar) 3X Dry Air Dry Air
750P22 0 to 1 psi 0 to 1 psi (0 to 70 mBar) 3X 316 SS Dry Air
750P03 0 to 5 psi (0 to 350 mBar) 3X Dry Air Dry Air
750P23 0 to 5 psi (0 to 350 mBar) 4X 316 SS Dry Air
750P04 0 to 15 psi 0 to 15 psi (0 to 1 bar) 3X Dry Air Dry Air
750P24 0 to 15 psi (0 to 1 bar) 4X 316 SS Dry Air
Gage
750P05 0 to 30 psi (0 to 2 bar) 4X 316 SS N/A
750P06 0 to 100 psi (0 to 7 bar) 4X 316 SS N/A
750P27 0 to 300 psi (0 to 20 bar) 4X 316 SS N/A
750P07 0 to 500 psi (0 to 35 bar) 4X 316 SS N/A
750P08 0 to 1000 psi (0 to 70 bar) 3X 316 SS N/A
750P09 0 to 1500 psi (0 to 100 bar) 3X 316 SS N/A
750P2000 0 to 2000 psi (0 to 140 bar) 3X 316 SS N/A
725S Users Manual
58
Model Parameter/Range Burst Rating Hi Side Media Lo Side Media
High
750P29 0 to 3000 psi (0 to 200 bar) 3X 316 SS N/A
750P30 0 to 5000 psi (0 to 340 bar) 3X 316 SS N/A
750P31 0 to 10000 psi (0 to 700 bar) 2X 316 SS N/A
Absolute
750PA3 0 to 5 psia (0 to 350 mBar) 4X 316 SS N/A
750PA4 0 to 15 psia (0 to 1 bar) 4X 316 SS N/A
750PA5 0 to 30 psia(0 to 2 bar) 4X 316 SS N/A
750PA6 0 to 100 psia (0 to 7 bar) 4X 316 SS N/A
750PA27 0 to 300 psia (0 to 20 bar) 4X 316 SS N/A
750PA7 0 to 500 psia (0 to 35 bar) 4X 316 SS N/A
750PA8 0 to 1000 psia (0 to 70 bar) 3X 316 SS N/A
750PA9 0 to 1500 psia (0 to 100 bar) 3X 316 SS N/A
Vacuum
750PV3 -5 psi (-350 mBar) 4X 316 SS Dry Air
750PV4 -15 psi (-1 bar) 4X 316 SS Dry Air
Multifunction Process Calibrator Accessories
59
Model Parameter/Range Burst Rating Hi Side Media Lo Side Media
Dual
750PD2 -1 to 1 psi (-70 to 70 mBar) 4X 316 SS Dry Air
750PD3 -5 to 5 psi (-350 to 350 mBar) 4X 316 SS Dry Air
750PD10 -10 to 10 psi (-700 to 700 mBar) 4X 316 SS Dry Air
750PD4 -15 to 15 psi (-1 to 1 bar) 4X 316 SS Dry Air
750PD5 -15 to 30 psi (-1 to 2 bar) 4X 316 SS —
750PD50 -15 to 50 psi (-1 to 3.5 bar) 4X 316 SS —
750PD6 -15 to 100 psi (-1 to 7 bar) 4X 316 SS —
750PD7 -15 to 200 psi (-1 to 14 bar) 4X 316 SS —
750PD27 -15 to 300 psi (-1 to 20 bar) 4X 316 SS —
725S Users Manual
60
Specifications Specifications are based on a one year calibration cycle and apply from +18 °C to +28 °C unless stated otherwise. All specifications assume a five-minute warmup period.
DC Voltage Measurement Range Resolution Accuracy (% of Reading + Counts)
30 V (upper display) 0.001 V 0.02 % + 2
20 V (lower display) 0.001 V 0.02 % + 2
90 mV (Read) 0.01 mV 0.02 % + 2
Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C
DC Voltage Source Range Resolution Accuracy (% of Reading + Counts)
100 mV 0.001 V 0.02 % + 2
10 V 0.001 V 0.02 % + 2
Temperature coefficient -10 °C to 18 °C, +28 °C to 55 °C: ±0.005 % of range per °C
Maximum load: 1mA
Millivolt Meaurement and Source Range Resolution Accuracy (% of Reading + Counts)
-10 mV to 75 mV 0.01 mV ±(0.025 % + 1 count)
• Maximum input voltage: 30 V
• Temperature coefficient -10 °C to 18 °C, 28 °C to 55 °C: ±0.005 % of range per °C
• Push to select this function. The signal is available at the thermocouple miniplug connector.
Multifunction Process Calibrator Specifications
61
DC mA Measurement and Source Range Resolution Accuracy (% of Reading + Counts)
24 mA 0.001 mA 0.02 % + 2
Temperature coefficient (-10 °C to 18 °C, 28 °C to 55 °C): ±0.005 % of range per °C
Drive capability: 1000 Ω at 20 mA.
Ohms Measurement Ohms Range Accuracy ±Ω
24 mA 4-Wire 2- and 3-Wire
0 Ω to 400 Ω 400 Ω to 1.5 kΩ 1.5 kΩ to 3.2 kΩ
0.1 0.5 1
0.15 1.0 1.5
Temperature coefficient (-10 °C to 18 °C, +28 °C to 55 °C): ± 0.005 % of range per °C
Excitation Current: 0.2 mA
Maximum input voltage: 30 V
*2-wire: Does not include lead resistance
3-wire: Assumes matched leads with a total resistance not exceeding 100 Ω
725S Users Manual
62
Ohms Source
Ohms Range Excitation Current from
Measurement Device Accuracy ±Ω
15 Ω to 400 Ω 15 Ω to 400 Ω 400 Ω to 1.5 kΩ 1.5 kΩ to 3.2 kΩ
0.15 mA to 0.5 mA 0.5 mA to 2 mA 0.05 mA to 0.8 mA 0.05 mA to 0.4 mA
0.15 0.1 0.5 1
Temperature coefficient (-10 °C to 18 °C, 28 °C to 55 °C): ± 0.005 % of range per °C
Resolution
15 Ω to 400 Ω 0.1 Ω
400 Ω to 3.2 kΩ 1 Ω
Frequency Measurement Range Resolution Accuracy
2.0 CPM to 1000.0 CPM 0.1 CPM ±(0.05 % + 1 count)
1 Hz to 1000 Hz 0.1 Hz ±(0.05 % + 1 count)
1.0 kHz to 10.0 kHz 0.01 kHz ±(0.05 % + 1 count) Sensitivity: 1 V peak-to-peak minimum Waveform: square wave
Frequency Source Range Minimum Accuracy (% of Output Frequency)
2.0 CPM to 1000.0 CPM 0.1 CPM ±0.05 %
1 Hz to 1000 Hz 0.1 Hz ±0.05 %
1.0 kHz to 10.0 kHz 0.01 kHz ±0.025 %
Waveform: 5 V p-p square wave, -0.1 V offset
Multifunction Process Calibrator Specifications
63
Temperature, Thermocouples
Type Range Measure and Source Accuracies (ITS-90)
J -200 °C to 0 °C 0 °C to 1200 °C
1.0 °C 0.7 °C
K -200 °C to 0 °C 0 °C to 1370 °C
1.2 °C 0.8 °C
T -200 °C to 0 °C 0 °C to 400 °C
1.2 °C 0.8 °C
E -200 °C to 0 °C 0 °C to 950 °C
0.9 °C 0.7 °C
R -20 °C to 0 °C 0 °C to 500 °C 500 °C to 1750 °C
2.5 °C 1.8 °C 1.4 °C
S -20 °C to 0 °C 0 °C to 500 °C 500 °C to 1750 °C
2.5 °C 1.8 °C 1.5 °C
B 600 °C to 800 °C 800 °C to 1000 °C 1000 °C to 1800 °C
2.2 °C 1.8 °C 1.4 °C
L -200 °C to 0 °C 0 °C to 900 °C
0.85 °C 0.7 °C
U -200 °C to 0 °C 0 °C to 400 °C
1.1 °C 0.75 °C
N -200 °C to 0 °C 0 °C to 1300 °C
1.5 °C 0.9 °C
725S Users Manual
64
XK -200 °C to -100 °C -100 °C to 800 °C
0.5 °C 0.6 °C
BP 0 °C to 800 °C 800 °C to 2500 °C
1.2 °C 2.5 °C
Resolution: J, K, T, E, L, N, U, XK, BP: 0.1 °C B,R,S: 1 °C
Loop Power Supply Voltage: 24 V Maximum current: 22 mA Short circuit protected
RTD Excitation (Simulation)
Allowable Excitation by RTD Type
Ni 120 0.15 mA to 3.0 mA
Pt 100-385 0.15 mA to 3.0 mA
Pt 100-392 0.15 mA to 3.0 mA
Pt 100-JIS 0.15 mA to 3.0 mA
Pt 200-385 0.05 mA to 0.8 mA
Pt 500-385 0.05 mA to 0.80 mA
Pt 1000-385 0.05 mA to 0.40 mA
Multifunction Process Calibrator Specifications
65
Temperature, RTD Ranges, and Accuracies (ITS-90)
Type Range °C Accuracy
Measure 4-Wire °C Measure 2- and 3-Wire* °C Source °C
Ni120 -80 to 260 0.2 0.3 0.2
Pt100-385 -200 to 800 0.33 0.5 0.33
Pt100-392 -200 to 630 0.3 0.5 0.3
Pt100-JIS -200 to 630 0.3 0.5 0.3
Pt200-385 -200 to 250 0.2 0.3 0.2
250 to 630 0.8 1.6 0.8
Pt500-385 -200 to 500 0.3 0.6 0.3 500 to 630 0.4 0.9 0.4
Pt1000-385 -200 to 100 0.2 0.4 0.2 100 to 630 0.2 0.5 0.2
Resolution: 0.1 °C, 0.1 °F Allowable excitation current (source): Ni120, Pt100-385, Pt100-392, Pt100-JIS, Pt200-385: 0.15 mA to 3.0 mA Pt500-385: 0.05 to 0.80 mA; Pt1000-385: 0.05 to 0.40 mA RTD Source: Addresses pulsed transmitters and PLCs with pulses as short as 5 ms. * 2-wire: Does not include lead resistance. 3-wire: Assumes matched leads with a total resistance not exceeding 100 Ω.
Pressure Measurement Range Resolution Accuracy Units
Determined by pressure module 5 digits Determined by pressure module
psi, inH2O@4 °C, inH2O@20 °C, kPa, cm H2O@4 °C, cmH2O@20 °C, bar, mbar, kg/cm2, mmHg, inHg
725S Users Manual
66
General Specifications Operating temperature -10 °C to 55 °C
Storage temperature - 20 °C to 71 °C (as limited by battery specification)
Operating altitude 3000 meters above mean sea level
Relative Humidity (% RH operating without condensation)
90 % (10 °C to 30 °C) 75 % (30 °C to 40 °C) 45 % (40 °C to 50 °C) 35 % (50 °C to 55 °C) Non-condensing <10 °C
Vibration Random, 2 g, 5 to 500 Hz
Safety IEC 61010-1 Pollution Degree 2, IEC 61010-2-030, Max 30 V to Earth
Electromagnetic Environment IEC 61326-1, Portable
Electromagnetic Compatibility Applies to use in Korea only. Class A Equipment (Industrial Broadcasting & Communication Equipment) This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.
Power requirements 4 AA alkaline batteries
Protection Class Pollution Degree II
Size 96 mm. x 200 mm. x 47 mm. (3.75 in x 7.9 in x 1.86 in)
Weight 650 gm (1 lb, 7 oz)