DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 10:59:30 AM
OUTPATIENT / PACT CLINIC - LEASE MODULEDISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 10:59:31 AM
OUTPATIENT / PACT CLINIC - LEASE MODULELEGEND
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 10:57:48 AM
OUTPATIENT / PACT CLINIC - LEASE MODULEAXONOMETRIC
TO
ILE
TT
NP
G1
PR
OC
ED
UR
ET
RP
R1
TO
ILE
TT
NP
G1
EX
AM
EX
PA
01
CL
EA
NU
CC
L1
EX
AM
EX
PA
01
EX
AM
EX
PA
01
EX
AM
EX
PA
01
STAFF CORRIDOR
ST
AF
F C
OR
RID
OR
(3
.51m
)
11
'-6
"
EX
AM
EX
PA
1
EX
AM
EX
PA
1
EX
AM
EX
PA
1
WO
ME
N'S
EX
AM
EX
W01
WO
ME
NS
EX
AM
EX
W01
TO
ILE
TT
NP
G1 T
OIL
ET
TP
G01
TO
ILE
TT
PG
01
EQ
UIP
RC
A01
ME
D A
LC
OV
ER
CA
06
(3
.81m
)
12
'-6
"
(3.05m)
10'-0"
(3.05m)
10'-0"
(4.27m)
14'-0"
(1.83m)
6'-0"
(1.83m)
6'-0"
(3.05m)
10'-0"
(3.05m)
10'-0"
(3.05m)
10'-0"
(3.05m)
10'-0"
(3.05m)
10'-0"
(3.05m)
10'-0"
(1.83m)
6'-0"
(1
.52m
)
5'-0
"
(6
.1m
)
20
'-0
"
(1
.52m
)
5'-0
"
(1.83m)
6'-0"
STAFF CORRIDOR
STAFF CORRIDOR
STAFF CORRIDOR
FL
EX
OF
FIC
EF
A01
PA
CT
TE
AM
WO
RK
SP
AC
EW
RT
M1
PA
CT
TE
AM
WO
RK
SP
AC
EW
RT
M1
AL
CO
VE
, S
CA
LE
PE
HW
2
CO
PY
AL
CO
VE
RC
A02
PATIENT CORRIDOR
PATIENT CORRIDOR
SY
ST
EM
S
FU
RN
ITU
RE
P
AR
TIT
ION
S
(1.37m)
4'-6"
(1.22m)
4'-0"
(1.37m)
4'-6"
(0.91m)
3'-0"
(0.91m)
3'-0"
(1.37m)
4'-6"
(1.22m)
4'-0"
(2.74m)
9'-0"
(1.22m)
4'-0"
(1.22m)
4'-0"
(1
.68m
)
5'-6
"
PA
CT
TE
AM
WO
RK
SP
AC
EW
RT
M1
(3.05m)
10'-0"
(3.05m)
10'-0"
(1.83m)
6'-0"
(3
.05m
)
10
'-0
"
(1
.68m
)
5'-6
"
(3
.51m
)
11
'-6
"
(3
.81m
)
12
'-6
"
SY
ST
EM
S
FU
RN
ITU
RE
P
AR
TIT
ION
S
SY
ST
EM
S
FU
RN
ITU
RE
P
AR
TIT
ION
S
(1.52m)
5'-0"
EQ
UIP
RC
A01
(1.83m)
6'-0"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/8" = 1'-0"
6/28/2018 10:57:50 AM
OUTPATIENT / PACT CLINIC - LEASE MODULEFLOOR PLAN
EX
AM
EX
PA
01
TO
ILE
TT
NP
G1
PR
OC
ED
UR
ET
RP
R1
TO
ILE
TT
NP
G1
FL
EX
OF
FIC
EF
A01
EX
AM
EX
PA
01
CL
EA
NU
CC
L1
EX
AM
EX
PA
01
EX
AM
EX
PA
01
PA
CT
TE
AM
WO
RK
SP
AC
EW
RT
M1
EQ
UIP
RC
A01
ME
D A
LC
OV
ER
CA
06
WO
ME
NS
EX
AM
EX
W01
TO
ILE
TT
NP
G1
EX
AM
EX
PA
1
EX
AM
EX
PA
1
EX
AM
EX
PA
1
WO
ME
N'S
EX
AM
EX
W01
TO
ILE
TT
PG
01
TO
ILE
TT
PG
01
PA
CT
TE
AM
WO
RK
SP
AC
EW
RT
M1
AL
CO
VE
, S
CA
LE
PE
HW
2
CO
PY
AL
CO
VE
RC
A02
EQ
UIP
RC
A01
PA
CT
TE
AM
WO
RK
SP
AC
EW
RT
M1
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/8" = 1'-0"
6/28/2018 10:57:51 AM
OUTPATIENT / PACT CLINIC - LEASE MODULEREFLECTED CEILIING PLAN
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 1:47:06 PM
CONFERENCE ROOM (CFR01)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:47:06 PM
CONFERENCE ROOM (CFR01)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 1:47:07 PM
CONFERENCE ROOM (CFR01)AXONOMETRIC
LEASING
DD
60"
1
24
3
COMPUTER WITH FLAT
PANEL MONITOR
M1801
(5.84m)
19'-2"
TELEVISION, HDTV, LARGE
SCREEN, 60"
M0512
BRACKET, TELEVISION,
WALL-MNTD, TILT ANGLE
A5212
240 NSF
CONFERENCE ROOM
(CFR01)
WORKSTATION,
COMPUTER, ENCLOSED,
WITH LOCK
F0690
5'-6
"
(3
.81m
)
12
'-6
"
CONFERENCE CHAIRF0220
TELEPHONE,
CONFERENCING SYSTEM
A1020
BASKET, WASTEPAPERF2000
CLOCK, BATTERY, 12"F3200
TO STAFF
CORRIDOR
VIDEO CONFERENCEM0507
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:47:07 PM
CONFERENCE ROOM (CFR01)FLOOR PLAN (240 NSF / 22.30 NSM)
LEASING
C C C
C C C
B BB
(5.84m)
19'-2"
240 NSF
CONFERENCE ROOM
(CFR01)
(3
.81m
)
12
'-6
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:47:08 PM
CONFERENCE ROOM (CFR01)REFLECTED CEILING PLAN
LEASING
WORKSTATION,
COMPUTER, ENCLOSED,
WITH LOCK
F0690
BASKET, WASTEPAPERF2000
TELEVISION, HDTV, LARGE
SCREEN, 60"
M0512
WORKSTATION,
COMPUTER, ENCLOSED,
WITH LOCK
F0690
COMPUTER WITH FLAT
PANEL MONITOR
M1801
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:47:08 PM
CONFERENCE ROOM (CFR01)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
VIDEO CONFERENCEM0507
CLOCK, BATTERY, 12"F3200
BASKET, WASTEPAPERF2000
CONFERENCE CHAIRF0220
CLOCK, BATTERY, 12"F3200
TELEPHONE,
CONFERENCING SYSTEM
A1020
TELEVISION, HDTV, LARGE
SCREEN, 60"
M0512
BRACKET, TELEVISION,
WALL-MNTD, TILT ANGLE
A5212
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:47:09 PM
CONFERENCE ROOM (CFR01)ELEVATIONS
ELEVATION 4
ELEVATION 3
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 1:47:09 PM
CONFERENCE ROOM (CFR01)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
m: RWC at 4’ - 0”
m: RB at h: 4" (100mm)
LVT
--
STC 40
m:Wood t:19 (Note 2) s:S
--
Yes
Yes
--
--
--
--
--
--
--
--
--
--
Yes
--
Design in accordance with project requirements.
--
1) Provide dimmable lighting for a variety of tasks (meetings, A/vpresentations, & conferences).
Connect to receptacles and equipment
Not Required
--
--
--
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) Provide acoustical/tackable wall covering on one wall.2) 40 STC rated acoustical door
2) Consider placement and optics to mitigate glare and ceilingreflections
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 1:59:18 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:59:19 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 1:59:19 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)AXONOMETRIC
LEASING
D
D
+48"
+48"
125 NSF
EXAM ROOM-PATIENT
ALIGNED CARE TEAM
(EXPA1)
LIGHT, EXAM, MOBILEM7401
OTOSCOPE /
OPHTHALMOSCOPE
M4200
SPHYGMOMANOMETER,
ANEROID
M4100
WORKSTATION,
COMPUTER
M1803
CHAIR, SIDE WITH ARMSF0205SINK, SS, SINGLE
CS010
COUNTERTOP, SOILD
SURFACE
CT020
HOOK, GARMENTA5145
BOARD, BULLETINF3025
HOOK, GARMENTA5145
TRACK, CUBICLE, WITH
CURTAIN
A5180
CLOCK, BATTERY, 12"F3200
TELEPHONE, WALL-MNTD,
1 LINE
A1012
RAIL, ACCESSORYA1132
TABLE, EXAM/TREATMENTM9025
1
24
3
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
(3
.05
m)
10
'-0
"
(3.81m)
12'-6"
5'-6
"
CHAIR
EXTENSION
CLEARANCE
BASKET, WASTEPAPERF2000
DISPENSER, PAPER
TOWEL
A5080
CART, STORAGEE0948
DISPENSER, SOAPA5075
WASTE DISPOSAL UNIT,
SHARPS
A5108
COMPUTER WITH FLAT
PANEL MONITOR
M1801STOOL, SELF-ADJUSTING
F0340
TO
ST
AF
F
CO
RR
IDO
R
TO
PA
TIE
NT
CO
RR
IDO
R
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:59:20 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)FLOOR PLAN (125 NSF / 11.61 NSM)
LEASING
A
A
125 NSF
EXAM ROOM-PATIENT
ALIGNED CARE TEAM
(EXPA1)
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"
TRACK, CUBICLE, WITH
CURTAIN
A5180
LIFT SYSTEM, OVERHEADA1203
TRACK, CUBICLE, WITH
CURTAIN
A5180
OPTIONAL PATIENT LIFT
NON-BARIATRIC PATIENT LIFT
OCCURS IN FIRST EXAM ROOM IN
PACT MODULE. SUBSEQUENT
EXAM ROOMS IN PACT MODULE
SHALL NOT HAVE A PATIENT LIFT
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:59:20 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)REFLECTED CEILING PLAN
LEASING
RAIL, ACCESSORYA1132
OTOSCOPE /
OPHTHALMOSCOPE
M4200
SPHYGMOMANOMETER,
ANEROID
M4100
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
BOARD, BULLETINF3025
TRACK, CUBICLE, WITH
CURTAIN
A5180
HOOK, GARMENTA5145
CART, STORAGEE0948
TRACK, CUBICLE, WITH
CURTAIN
A5180
TABLE, EXAM/TREATMENTM9025
LIGHT, EXAM, MOBILEM7401
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:59:21 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
CABINETCD040
COUNTERTOP, SOILD
SURFACE
CT020
SINK, SS, SINGLECS010
TRACK, CUBICLE, WITH
CURTAIN
A5180
HOOK, GARMENTA5145
CHAIR, SIDE WITH ARMSF0205
CLOCK, BATTERY, 12"F3200
COMPUTER WITH FLAT
PANEL MONITOR
M1801
WORKSTATION,
COMPUTER
M1803
TELEPHONE, WALL-MNTD,
1 LINE
A1012
STOOL, SELF-ADJUSTINGF0340
TRACK, CUBICLE, WITH
CURTAIN
A5180
CABINETCD040
DISPENSER, SOAPA5075
COUNTERTOP, SOILD
SURFACE
CT020
BASKET, WASTEPAPERF2000
WASTE DISPOSAL UNIT,
SHARPS
A5108
SINK, SS, SINGLECS010
DISPENSER, PAPER
TOWEL
A5080
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 1:59:21 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 1:59:22 PM
EXAM ROOM, PATIENT ALIGNED CARE TEAM (EXPA1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: RB at h: 4" (100mm)
LVT
--
STC 40
m:Wood t:19 (Note 1) s:S
--
Yes
Yes
--
Yes
--
--
--
Yes
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door
Provide ballasts per fixture for desired switching configuration. To provide a uniform lighting level, switch inner lamp(s) on first switch and outer lamps on second switch.
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 2:24:21 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 2:24:21 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 2:24:21 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)AXONOMETRIC
LEASING
+44"
+48"
+48"
1
24
3
CLOCK, BATTERY, 12"F3200
SPHYGMOMANOMETER,
ANEROID
M4100
OTOSCOPE /
OPHTHALMOSCOPE
M4200
RAIL, ACCESSORYA1132
TABLE, EXAM/TREATMENTM9025
HOOK, GARMENTA5145
BOARD, BULLETINF3025
CHAIR, SIDE WITH ARMSF0205
125 NSF
WOMENS HEALTH
EXAM ROOM
(EXW01)
LIGHT, EXAM, MOBILEM7401
CHAIR EXTENSION
CLEARANCE
TO PATIENT
TOILET
TO
ST
AF
F
CO
RR
IDO
R
TO
ST
AF
F
CO
RR
IDO
R TRACK, CUBICLE, WITH
CURTAIN
A5180
(3.81m)
12'-6"
5'-6
"
(3
.05
m)
10
'-0
"TELEPHONE, WALL-MNTD,
1 LINE
A1012
BASKET, WASTEPAPERF2000
WASTE DISPOSAL UNIT,
SHARPS
A5108
COMPUTER WITH FLAT
PANEL MONITOR
M1801
HOOK, GARMENTA5145
CART, STORAGEE0948
STOOL, SELF-ADJUSTINGF0340
COMPUTER WITH FLAT
PANEL MONITOR
M1801
WORKSTATION,
COMPUTER
M1803
DISPENSER, SOAPA5075
SINK, SS, SINGLECS010
COUNTERTOP, SOLID
SURFACE
CT020
DISPENSER, PAPER
TOWEL
A5080
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 2:24:22 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)FLOOR PLAN (125 NSF / 11.61 NSM)
LEASING
A
A
125 NSF
WOMENS HEALTH
EXAM ROOM
(EXW01)
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"LIFT SYSTEM, OVERHEADA1203
TRACK, CUBICLE, WITH
CURTAIN
A5180
TRACK, CUBICLE, WITH
CURTAIN
A5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 2:24:22 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)REFLECTED CEILING PLAN
LEASING
CLOCK, BATTERY, 12"F3200
LIGHT, EXAM, MOBILEM7401
RAIL, ACCESSORYA1132
TABLE, EXAM/TREATMENTM9025
OTOSCOPE /
OPHTHALMOSCOPE
M4200
BOARD, BULLETINF3025
HOOK, GARMENTA5145
CART, STORAGEE0948
SPHYGMOMANOMETER,
ANEROID
M4100
STOOL, SELF-ADJUSTINGF0340
TRACK, CUBICLE, WITH
CURTAIN
A5180
TRACK, CUBICLE, WITH
CURTAIN
A5180
BOARD, BULLETINF3025
RAIL, ACCESSORYA1132
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 2:24:23 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
COMPUTER WITH FLAT
PANEL MONITOR
M1801
TRACK, CUBICLE, WITH
CURTAIN
A5180
DISPENSER, SOAPA5075
TELEPHONE, WALL-MNTD,
1 LINE
A1012
CHAIR, SIDE WITH ARMSF0205
WORKSTATION,
COMPUTER
M1803
COUNTERTOP, SOLID
SURFACE
CT020
SINK, SS, SINGLECS010
HOOK, GARMENTA5145
DISPENSER, PAPER
TOWEL
A5080
WASTE DISPOSAL UNIT,
SHARPS
A5108
COUNTERTOP, SOLID
SURFACE
CT020
SINK, SS, SINGLECS010
DISPENSER, SOAPA5075
BASKET, WASTEPAPERF2000
TRACK, CUBICLE, WITH
CURTAIN
A5180
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 2:24:23 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 2:24:24 PM
EXAM ROOM, WOMEN'S HEALTH (EXW01)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: RB at h: 4" (100mm)
LVT
--
STC 40
m:Wood t:19 (Note 1) s:U
--
Yes
Yes
--
Yes
--
--
--
Yes
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
Not Required
--
Yes
Yes
Yes
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door
Provide ballasts per fixture for desired switching configuration. To provide a uniform lighting level, switch inner lamp(s) on first switch and outer lamps on second switch.
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 3:08:16 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:08:16 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 3:08:16 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)AXONOMETRIC
LEASING
+44
"+
44"
+44"
+44"+48"
+44
"
+48"
+44
"
4
1
2
3
210 NSF
LABORATORY,
GENERAL(LBGM1)
ANALYZERL0990
STAINER, SLIDE, AUTO,
HEMATOLOGY
L8580
CENTRIFUGEL1502
SHAKER, ROTATINGL8520
COPIER/FAX COMBOM1840
COMPUTER WITH FLAT
PANEL MONITOR
M1801
PRINTER, LABEL,
PHARMACY
M1830
DISPENSER, PAPER
TOWEL
A5080
SINK, SS, SINGLE
COMPARTMENT
CS150
DISPENSER, SOAPA5075
ANALYZER, HEMATOLGOYL1115
TELEPHONE, DESK,
CONFERENCE
A1011
(4
.6m
)
15
'-1
"
HOOK, GARMENTA5145
BASKET, WASTEPAPERF2000
BOARD, BULLETINF3010
INFECTIOUS WASTE BAGM3072
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
DISPENSER, SOAPA5075
SINK, SS, SINGLECS010
DISPENSER, PAPER
TOWEL
A5080
COUNTERTOP, SOLID
SURFACE
CT020
REFRIGERATOR-FREEZER,
BIOLOGICAL UPRIGHT
R6070
(4.27m)
14'-0"
5'-6
"WASTE DISPOSAL UNIT,
SHARPS
A5108
EYEWASH, HANDS-FREEP1965
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:08:17 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)FLOOR PLAN (210 NSF / 19.51 NSM)
LEASING
B B
B
BB
B
210 NSF
LABORATORY,
GENERAL(LBGM1)
(4.27m)
14'-0"
(4
.6m
)
15
'-1
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:08:17 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)REFLECTED CEILING PLAN
LEASING
STAINER, SLIDE, AUTO,
HEMATOLOGY
L8580
ANALYZER, HEMATOLGOYL1115
REFRIGERATOR-FREEZER,
BIOLOGICAL UPRIGHT
R6070
CABINET, U-C-B, 1 SHELF,
2 HALF DR, 2 DO
C04F0
STAINER, SLIDE, AUTO,
HEMATOLOGY
L8580
ANALYZERL0990
SHAKER, ROTATINGL8520
DISPENSER, SOAPA5075
SINK, SS, SINGLE
COMPARTMENT
CS150
DISPENSER, PAPER
TOWEL
A5080
CENTRIFUGEL1502
CABINET, U-C-B, 1 SHELF,
2 HALF DR, 2 DO
C04F0
CABINET, SINK, U-C-B, 2
DOOR
C05P0
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:08:17 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, SOAPA5075
COPIER/FAX COMBOM1840
TELEPHONE, DESK,
CONFERENCE
A1011
COMPUTER WITH FLAT
PANEL MONITOR
M1801
PRINTER, LABEL,
PHARMACY
M1830
CABINET, U-C-B, 1 SHELF,
2 HALF DR, 2 DO
C04F0
CABINET, ADA, SINK
SUPPORT
A1018
EYEWASH, HANDS-FREEP1965
DISPENSER, PAPER
TOWEL
A5080
CABINET, SINK, U-C-B, 2
DOOR
C05P0 DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
HOOK, GARMENTA5145
BASKET, WASTEPAPERF2000
INFECTIOUS WASTE BAGM3072
BOARD, BULLETINF3010
SINK, SS, SINGLECS010
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
WASTE DISPOSAL UNIT,
SHARPS
A5108
CABINET, ADA, SINK
SUPPORT
A1018
COUNTERTOP, SOLID
SURFACE
CT020
EYEWASH, HANDS-FREEP1965
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:08:17 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 3:08:18 PM
LABORATORY, GENERAL OUTPATIENT CLINIC (LBGM1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT-SP
--
m: GWB f:SC
--
WSF at 0’ - 4”
WSF
--
STC 40
m:Wood t:1 (Note 1) s:V
--
Yes
Yes
--
--
Yes
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment.
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) Optional partial lite as shown. Louver as required.
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 3:35:27 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:35:27 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 3:35:28 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)AXONOMETRIC
LEASING
+48
"+
48"+
48"
80 NSF
BLOOD SPECIMEN
COLLECTION ROOM
(LBVP1)
1
2
3
COMPUTER WITH FLAT
PANEL MONITOR
M1801
CART, STORAGEE0948
4
BASKET, WASTEPAPERF2000
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
BULLETIN BOARDF3025
CHAIR, LAB, BLOOD
DRAWING W STORAGE
M1410
HOOK, GARMENTA5145
(2.44m)
8'-0"
(3
.05
m)
10
'-0
"
PRINTER, LABEL,
PHARMACY
M1830
WORKSTATION,
COMPUTER
M1803
TELEPHONE, WALL-MNTD,
1 LINE WITH SPEAKER
A1014
CABINETCD040
DISPENSER, HAND
SANITIZER
A5077WASTE DISPOSAL UNIT,
SHARPS
A5108
5'-6
"
DISPENSER, PAPER
TOWEL
A5080
LAVATORY, VITREOUS
CHINA
P3100
LAVATORY PART OF OVERALL
PHLEBOTOMY AREA,
LOCATE PER PLANNING
CONFIGURATIONS.
COUNTERTOP, SOLID
SURFACE
CT020
TRACK WITH CURTAINA5180
HOLLOW METAL
CASED OPENING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:35:28 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)FLOOR PLAN (80 NSF / 7.43 NSM)
LEASING
A
80 NSF
BLOOD SPECIMEN
COLLECTION ROOM
(LBVP1)
(2.44m)
8'-0"
(3
.05
m)
10
'-0
"
TRACK WITH CURTAINA5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:35:29 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)REFLECTED CEILING PLAN
LEASING
HOOK, GARMENTA5145
COMPUTER WITH FLAT
PANEL MONITOR
M1801
TELEPHONE, WALL-MNTD,
1 LINE WITH SPEAKER
A1014
WORKSTATION,
COMPUTER
M1803
TRACK WITH CURTAINA5180
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
BASKET, WASTEPAPERF2000
COMPUTER WITH FLAT
PANEL MONITOR
M1801
WORKSTATION,
COMPUTER
M1803
PRINTER, LABEL,
PHARMACY
M1830
CABINETCD040
WASTE DISPOSAL UNIT,
SHARPS
A5108
TRACK WITH CURTAINA5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:35:29 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
CHAIR, LAB, BLOOD
DRAWING W STORAGE
M1410
BULLETIN BOARDF3025
BASKET, WASTEPAPERF2000
PRINTER, LABEL,
PHARMACY
M1830
CABINETCD040
CHAIR, LAB, BLOOD
DRAWING W STORAGE
M1410
BULLETIN BOARDF3025
TELEPHONE, WALL-MNTD,
1 LINE WITH SPEAKER
A1014
TRACK WITH CURTAINA5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 3:35:30 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 3:35:30 PM
BLOOD SPECIMEN COLLECTION ROOM (LBVP1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
GWB-P; AT
--
m: GWB f:SC
--
m: WSF at h: 4" (100mm)
WSF
--
STC 40
m:Wood t:1 (Note 1) s:V
--
Yes
Yes
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) Optional Partial lite as shown. Louver as required.
Provide ballasts per fixture for desired switching configuration. To provide a uniform lighting level, switch inner lamp(s) on first switch and outer lamps on second switch.
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 4:03:22 PM
CONSULT ROOM (OFDC1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:03:23 PM
CONSULT ROOM (OFDC1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 4:03:23 PM
CONSULT ROOM (OFDC1)AXONOMETRIC
LEASING
+44
"
125 NSF
CONSULT ROOM
(OFDC1)
1
2
3
4
(3
.05m
)
10
'-0
"
(3.81m)
12'-6"
CHAIR, SIDE, BARIATRIC,
WITH ARMS
F0206
TELEPHONE, DESK,
MULTI-LINE
A1015
BASKET, WASTEPAPERF2000
CLOCK, BATTERY, 12"F3200
BRACKET, TELEVISION,
WALL-MNTD, TILT ANGLE
A5212
TELEVISION BACKINGA5220
CHAIR, SWIVELF0300
DISPENSER, HAND
SANITIZER, HANDS FREE
A5077
5'-6
"
WHITE BOARD, BULLETIN
BOARD COMBINATION
F3065
WORKSTATIONE0063
COMPUTER WITH FLAT
PANEL MONITOR
M1801
COUNTERTOP, SOLID
SURFACE
CT020
CABINETCD040
DISPENSER, SOAPA5075
DISPENSER, PAPER
TOWEL
A5080
PRINTER, COMPUTERM1825
CABINET, FILING,
LATERAL, HALF HEIGHT
F0420
HOOK, GARMENTA5145
CHAIR, SIDE, BARIATRIC,
WITH ARMS
F0206
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:03:23 PM
CONSULT ROOM (OFDC1)FLOOR PLAN (125 NSF / 11.61 NSM)
LEASING
C C
CC
125 NSF
CONSULT ROOM
(OFDC1)
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:03:23 PM
CONSULT ROOM (OFDC1)REFLECTED CEILING PLAN
LEASING
CHAIR, SIDE, BARIATRIC,
WITH ARMS
F0206
CLOCK, BATTERY, 12"F3200
TELEPHONE, DESK,
MULTI-LINE
A1015
BRACKET, TELEVISION,
WALL-MNTD, TILT ANGLE
A5212
CHAIR, SWIVELF0300
CABINETCD040
DISPENSER, SOAPA5075
COUNTERTOP, SOLID
SURFACE
CT020
COMPUTER WITH FLAT
PANEL MONITOR
M1801
SINK, SS, SINGLECS010
CHAIR, SIDE, BARIATRIC,
WITH ARMS
F0206
CHAIR, SIDE WITH ARMSF0205
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:03:24 PM
CONSULT ROOM (OFDC1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
WHITE BOARD, BULLETIN
BOARD COMBINATION
F3065
HOOK, GARMENTA5145
PRINTER, COMPUTERM1825
CABINET, FILING,
LATERAL, HALF HEIGHT
F0420
DISPENSER, HAND
SANITIZER, HANDS FREE
A5077
CABINETCD040
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, SOAPA5075
BASKET, WASTEPAPERF2000
COUNTERTOP, SOLID
SURFACE
CT020
PRINTER, COMPUTERM1825
CABINET, FILING,
LATERAL, HALF HEIGHT
F0420
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:03:25 PM
CONSULT ROOM (OFDC1)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 4:03:25 PM
CONSULT ROOM (OFDC1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
PRB
CPT
--
STC 40
m:Wood t:19 (Note 1) s:U
--
Yes
Yes
--
--
--
--
--
--
--
--
--
Yes
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door2) Aesthetic specialty materials may be used as accents where thedesign and budget permits.3) Profile Base height will vary depending on profile base selected.4) Carpet tile is preferred especially in high traffic areas. Wherebudget and design conditions do not allow carpet tile; broadloomcarpet may be substituted.
Provide ballasts per fixture for desired switching configuration. To provide a uniform lighting level, switch inner lamp(s) on first switch and outer lamps on second switch.
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 4:36:56 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:36:57 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 4:36:57 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)AXONOMETRIC
LEASING
+44
"
1
3
125 NSF
CONSULT ROOM
(OFDC2)
4 5'-6
"
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"
2
CHAIR, SIDE, BARIATRIC,
WITH ARMS
F0206
CHAIR, SIDE WITH ARMSF0205
BRACKET, TELEVISION,
WALL-MNTD, TILT ANGLE
A5212
CLOCK, BATTERY, 12"F3200
TELEVISION BACKINGA5220
COMPUTER WITH FLAT
PANEL MONITOR
M1801
TELEPHONE, DESK,
MULTI-LINE
A1015
WORKSTATIONE0063
DISPENSER, HAND
SANITIZER, HANDS FREE
A5077
DISPENSER, SOAPA5075
COUNTERTOP, SOLID
SURFACE
CT020
SINK, SS, SINGLECS010
DISPENSER, PAPER
TOWEL
A5080
BASKET, WASTEPAPERF2000
CHAIR, SWIVELF0300
HOOK, GARMENTA5145 T
O S
TA
FF
CO
RR
IDO
R
TO
PA
TIE
NT
CO
RR
IDO
R
WHITE BOARD, BULLETIN
BOARD COMBINATION
F3065
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:36:58 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)FLOOR PLAN (125 NSF / 11.61 NSM)
LEASING
C
C C
C
125 NSF
CONSULT ROOM
(OFDC2)
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:36:58 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)REFLECTED CEILING PLAN
LEASING
BASKET, WASTEPAPERF2000
COUNTERTOP, SOLID
SURFACE
CT020
SINK, SS, SINGLECS010
DISPENSER, SOAPA5075
CABINETCD040
DISPENSER, HAND
SANITIZER, HANDS FREE
A5077
CHAIR, SIDE, BARIATRIC,
WITH ARMS
F0206
WORKSTATIONE0063
COMPUTER WITH FLAT
PANEL MONITOR
M1801
BRACKET, TELEVISION,
WALL-MNTD, TILT ANGLE
A5212
CLOCK, BATTERY, 12"F3200
CHAIR, SWIVELF0300
CHAIR, SIDE, BARIATRIC,
WITH ARMS
F0206
CHAIR, SIDE WITH ARMSF0205
WHITE BOARD, BULLETIN
BOARD COMBINATION
F3065
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:36:59 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
DISPENSER, SOAPA5075
SINK, SS, SINGLECS010
DISPENSER, PAPER
TOWEL
A5080
CABINETCD040
COUNTERTOP, SOLID
SURFACE
CT020
BASKET, WASTEPAPERF2000
HOOK, GARMENTA5145
HOOK, GARMENTA5145
WHITE BOARD, BULLETIN
BOARD COMBINATION
F3065
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/27/2018 4:36:59 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)ELEVATIONS
ELEVATION 4
ELEVATION 3
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/27/2018 4:37:00 PM
CONSULT ROOM, MENTAL HEALTH (OFDC2)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
PRB
CPT
--
STC 40
m:Wood t:19 (Note 1) s:U
--
Yes
Yes
--
--
--
--
--
--
--
--
--
Yes
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door2) Aesthetic specialty materials may be used as accents where thedesign and budget permits.3) Profile Base height will vary depending on profile base selected.4) Carpet tile is preferred especially in high traffic areas. Wherebudget and design conditions do not allow carpet tile; broadloomcarpet may be substituted.
Provide ballasts per fixture for desired switching configuration. To provide a uniform lighting level, switch inner lamp(s) on first switch and outer lamps on second switch.
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:34:52 AM
ALCOVE, SCALE (PEHW2)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:34:52 AM
ALCOVE, SCALE (PEHW2)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:34:52 AM
ALCOVE, SCALE (PEHW2)AXONOMETRIC
LEASING
1
80 NSF
ALCOVE, SCALE(PEHW2)
(0
.46
m)
1'-6
"
(1
.22
m)
4'-0
"
(4.42m)
14'-6"
SCALE, WITH
WHEELCHAIR RAMP
M4020
5'-6
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:34:53 AM
ALCOVE, SCALE (PEHW2)FLOOR PLAN (60 NSF / 5.57 NSM)
LEASING
C C C
80 NSF
ALCOVE, SCALE(PEHW2)
(1
.22
m)
4'-0
"
(4.42m)
14'-6"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:34:53 AM
ALCOVE, SCALE (PEHW2)REFLECTED CEILING PLAN
LEASING
SCALE, WITH
WHEELCHAIR RAMP
M4020
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:34:53 AM
ALCOVE, SCALE (PEHW2)ELEVATIONS
ELEVATION 1
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:34:54 AM
ALCOVE, SCALE (PEHW2)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
RWC at 4’ - 0”
m: RB at h: 4" (100mm)
LVT
--
--
OPEN
--
--
--
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
Yes
--
--
--
--
--
--
--
Recessed pendant, center of tile
1) Ref. Note 6 and 64, PG-18-14
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 12:40:32 AM
ALCOVE, MEDICATION (RCA06)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 12:40:33 AM
ALCOVE, MEDICATION (RCA06)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 12:40:33 AM
ALCOVE, MEDICATION (RCA06)AXONOMETRIC
LEASING
+44
"
30 NSF
ALCOVE, MEDICATION(RCA06)
1
COUNTERTOP, SOILD
SURFACE
CT020
BASKET, WASTEPAPERF2000
SINK, SS, SINGLECS010
DISPENSER, SOAPA5075
DISPENSER, PAPER
TOWEL, SENSOR, HANDS
FREE
A5082
COMPUTER WITH FLAT
PANEL MONITOR
M1801
REFRIGERATOR, U-CR6200
DISTRIBUTION SYSTEM,
MEDICATION, AUTOMATIC
M3150
WORKSTATION,
COMPUTER
M1803
68" SYSTEMS FURNITURE
(0
.76
m)
2'-6
"
(3.66m)
12'-0"WASTE DISPOSAL UNIT,
SHARPS
A5108
AUXILARY MEDICATION CABINET
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 12:40:34 AM
ALCOVE, MEDICATION (RCA06)FLOOR PLAN (30 NSF / 2.79 NSM)
LEASING
DISPENSER, SOAPA5075
DISPENSER, PAPER
TOWEL, SENSOR, HANDS
FREE
A5082
SINK, SS, SINGLECS010
REFRIGERATOR, U-CR6200
COMPUTER WITH FLAT
PANEL MONITOR
M1801
WORKSTATION,
COMPUTER
M1803
BASKET, WASTEPAPERF2000
DISTRIBUTION SYSTEM,
MEDICATION, AUTOMATIC
M3150WASTE DISPOSAL UNIT,
SHARPS
A5108
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
68" SYSTEMS FURNITURE
SYSTEMS FURNTER
COUNTER @ 36"
OPEN TO BEYOND
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 12:40:34 AM
ALCOVE, MEDICATION (RCA06)ELEVATIONS
ELEVATION 1
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 12:40:35 AM
ALCOVE, MEDICATION (RCA06)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
--
LVT
--
--
OPEN
--
Yes
--
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
--
--
--
--
--
--
--
--
--
--
Recessed pendant, center of tile
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:05:47 AM
RECEPTION (RCP03)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:05:47 AM
RECEPTION (RCP03)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:05:47 AM
RECEPTION (RCP03)AXONOMETRIC
LEASING
1
2
3
(2
.13m
)
7'-0
"
(1
.52
m)
5'-0
"
(0
.91
m)
3'-0
"
(1
.52
m)
5'-0
"
(1
.52
m)
5'-0
"
(5.49m)
18'-0"
5'-6
"234 NSF
RECEPTION
(RCP03)BASKET, WASTEPAPER
F2000
CHAIR, SWIVEL, HIGH
BACK
F0275
TELEPHONE, DESK,
MULTI-LINE
A1015
RECEPTION DESK,
BUILT-IN
E0130
COMPUTER WITH FLAT
PANEL MONITOR
M1801
SHREDDER, PAPER
HEAVY DUTY
F2550
COPY STORAGE
CLOCK, BATTERY, 12"F3200
WORKSPACE
CABINET, U-C-B, 1 SHELF,
2 HALF DR, 2 DO
C04F0
CABINETCD040
COUNTERTOP, SOLID
SURFACE
CT020
RECEPTION DESK,
BUILT-IN
A6105
CHAIR, SWIVEL, HIGH
BACK
F0275
TELEPHONE, DESK,
MULTI-LINE
A1015
COMPUTER WITH FLAT
PANEL MONITOR
M1801
COPIER, FLOOR
STANDING
M1810
SYSTEMS FURNITURE;
60" HEIGHT
ADA ACCESSIBLE
TRANSACTION COUNTER
PEDESTAL, FILE, MOBILEF0404
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:05:48 AM
RECEPTION (RCP03)FLOOR PLAN (234 NSF / 21.74 NSM)
LEASING
B
B B
B B
B B B
BB B
(5.49m)
18'-0"
(2
.13m
)
7'-0
"
(1
.52
m)
5'-0
"
(3
.96
m)
13
'-0
"
234 NSF
RECEPTION
(RCP03)
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:05:48 AM
RECEPTION (RCP03)REFLECTED CEILING PLAN
LEASING
CLOCK, BATTERY, 12"F3200
RECEPTION DESK,
BUILT-IN
E0130
TELEPHONE, DESK,
MULTI-LINE
A1015
CHAIR, SWIVEL, HIGH
BACK
F0275
BASKET, WASTEPAPERF2000
COMPUTER WITH FLAT
PANEL MONITOR
M1801
CHAIR, SIDE WITH ARMSF0205
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:05:48 AM
RECEPTION (RCP03)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
BASKET, WASTEPAPERF2000
RECEPTION DESK,
BUILT-IN
E0130
TELEPHONE, DESK,
MULTI-LINE
A1015
CHAIR, SWIVEL, HIGH
BACK
F0275
SYSTEMS FURNITURE
60" HEIGHT
PEDESTAL, FILE, MOBILEF0404
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:05:49 AM
RECEPTION (RCP03)ELEVATIONS
ELEVATION 3
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:05:49 AM
RECEPTION (RCP03)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: PRB
CPT ; LVT
--
--
m:Wood t:1 (Note 1) s:S
--
Yes
Yes
--
Yes
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
--
--
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) Ref. PG 18-14 Notes 6 and 8
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:02:45 AM
ALCOVE, COPY (RPR03)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:02:45 AM
ALCOVE, COPY (RPR03)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:02:45 AM
ALCOVE, COPY (RPR03)AXONOMETRIC
LEASING
1
30 NSF
ALCOVE, COPY(RPR03)
(4.57m)
15'-0"
BASKET, WASTEPAPERF2000
COPIER, FLOOR
STANDING
M1810
SHREDDER, PAPER
HEAVY DUTY
F2550
COUNTERTOP, SOLID
SURFACE
CT020
(0
.61
m)
2'-0
"
42" SYSTEMS FURNITURE
TELEPHONE, DESK,
CONFERENCE
A1011
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:02:46 AM
ALCOVE, COPY (RPR03)FLOOR PLAN (30 NSF / 2.79 NSM)
LEASING
COUNTERTOP, SOLID
SURFACE
CT020
CABINET, U-C-B, 1 SHELF,
2 HALF DR, 2 DO
C04F0
SHREDDER, PAPER
HEAVY DUTY
F2550
BASKET, WASTEPAPERF2000
COPIER, FLOOR
STANDING
M1810
TELEPHONE, DESK,
CONFERENCE
A1011
42" SYSTEMS FURNITURE
OPEN TO BEYOND
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:02:46 AM
ALCOVE, COPY (RPR03)ELEVATIONS
ELEVATION 1
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:02:46 AM
ALCOVE, COPY (RPR03)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: RB at h: 4" (100mm)
LVT
--
--
OPEN
--
Yes
Yes
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
--
--
--
--
--
--
--
--
--
--
Recessed pendant, center of tile
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:26:37 PM
TOILET, GENERAL (TNPG1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:26:37 PM
TOILET, GENERAL (TNPG1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:26:38 PM
TOILET, GENERAL (TNPG1)AXONOMETRIC
LEASING
60 NSF
TOILET, GENERAL(TNPG1)
2
3
4
TOILET, WALL HUNGP9050
DISPOSAL, SANITARY
NAPKIN
A5090
DISPENSER, TOILET ROLLA5200
(2.44m)
8'-0"
(2
.29
m)
7'-6
"
1
12" MINIMUM
DISPENSER, SOAPA5075
BASKET, WASTEPAPERF2000
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
LAVATORY, VITREOUS
CHINA
P3100
DISPENSER, PAPER
TOWEL
A5080
HOOK, GARMENTA5145
GRAB BARA5109
5'-6
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:26:38 PM
TOILET, GENERAL (TNPG1)FLOOR PLAN (60 NSF / 5.57 NSM)
LEASING
C
C
60 NSF
TOILET, GENERAL(TNPG1)
(2.44m)
8'-0"
(2
.29
m)
7'-6
"
12" MINIMUM
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:26:38 PM
TOILET, GENERAL (TNPG1)REFLECTED CEILING PLAN
LEASING
DISPENSER, PAPER
TOWEL
A5080
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
GRAB BARA5109
DISPENSER, TOILET ROLLA5200
TOILET, WALL HUNGP9050
LAVATORY, VITREOUS
CHINA
P3100
DISPOSAL, SANITARY
NAPKIN
A5090
DISPENSER, SOAPA5075
BASKET, WASTEPAPERF2000
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
DISPENSER, PAPER
TOWEL
A5080
LAVATORY, VITREOUS
CHINA
P3100
DISPENSER, SOAPA5075
BASKET, WASTEPAPERF2000
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:26:39 PM
TOILET, GENERAL (TNPG1)ELEVATIONS
ELEVATION 2
ELEVATION 1
LEASING
HOOK, GARMENTA5145
DISPENSER, TOILET ROLLA5200
DISPOSAL, SANITARY
NAPKIN
A5090
GRAB BARA5109
TOILET, WALL HUNGP9050
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:26:39 PM
TOILET, GENERAL (TNPG1)ELEVATIONS
ELEVATION 4
ELEVATION 3
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:26:40 PM
TOILET, GENERAL (TNPG1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
CT; PT at 4’ - 0”
m: PT ; RES at h: 4" (100mm) (Note 2)
PT ; RES (Note 2)
3" (75mm)
STC 40
m:Wood t:19 (Note 1) s:U
--
--
--
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
Yes
--
--
--
--
--
--
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door2) Ref. PG-18-14 Notes 7 and 74
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:04:50 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:04:50 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:04:50 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)AXONOMETRIC
LEASING
60 NSF
TOILET, WOMEN'S
HEALTH, EXAM ROOM(TPG01)
(2
.29
m)
7'-6
"
(2.44m)
8'-0"
TOILET, WALL HUNGP9050
GRAB BARA5109
3
1
24
5'-6
"
12" MINIMUM
DISPOSAL, SANITARY
NAPKIN
A5090
GRAB BARA5109
DISPENSER, TOILET ROLLA5200
BASKET, WASTEPAPERF2000
DISPENSER, SOAPA5075
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
LAVATORY, VITREOUS
CHINA
P3100
DISPENSER, PAPER
TOWEL
A5080
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:04:50 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)FLOOR PLAN (60 NSF / 5.57 NSM)
LEASING
C
C
60 NSF
TOILET, WOMEN'S
HEALTH, EXAM ROOM(TPG01)
(2
.29
m)
7'-6
"
12" MINIMUM
(2.44m)
8'-0"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:04:51 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)REFLECTED CEILING PLAN
LEASING
BASKET, WASTEPAPERF2000
LAVATORY, VITREOUS
CHINA
P3100
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, SOAPA5075
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:04:51 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
DISPENSER, TOILET ROLLA5200
DISPOSAL, SANITARY
NAPKIN
A5090
TOILET, WALL HUNGP9050
GRAB BARA5109
DISPENSER, PAPER
TOWEL
A5080
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
LAVATORY, VITREOUS
CHINA
P3100
DISPOSAL, SANITARY
NAPKIN
A5090DISPENSER, TOILET ROLL
A5200GRAB BAR
A5109
TOILET, WALL HUNGP9050
BASKET, WASTEPAPERF2000
DISPENSER, SOAPA5075
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 1:04:51 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 1:04:52 PM
TOILET, WOMEN'S HEALTH, EXAM ROOM (TPG01)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
CT; PT at 4’ - 0”
m: PT ; RES at h: 4" (100mm) (Note 2)
PT ; RES (Note 2)
3" (75mm)
STC 40
m:Wood t:19 (Note 1) s:U
--
--
--
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
Yes
--
--
--
--
--
--
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door2) Ref. PG-18-14 Notes 7 and 74
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 11:55:11 AM
TOILET, SPECIMEN (TPSP1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 11:55:11 AM
TOILET, SPECIMEN (TPSP1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 11:55:11 AM
TOILET, SPECIMEN (TPSP1)AXONOMETRIC
LEASING
1
2
(2.44m)
8'-0"
(2
.29m
)
7'-6
"
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
DISPENSER, TOILET ROLLA5200
DISPOSAL, SANITARY
NAPKIN
A5090
TOILET, WALL HUNGP9050
WATER SHUTOFF FOR
SPECIMEN COLLECTION
BY DETERMINATION OF
FACILITY FUNCTION
LAVATORY, VITREOUS
CHINA
P3100
GRAB BARA5109CABINET, SPECIMEN,
PASS THRU, CRS
L1200
60 NSF
TOILET, SPECIMEN(TPSP1)
5'-6
"
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, SOAPA5075
BASKET, WASTEPAPERF2000
2
HOOK, GARMENTA5145
12" MINIMUM
1
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 11:55:12 AM
TOILET, SPECIMEN (TPSP1)FLOOR PLAN (60 NSF / 5.57 NSM)
LEASING
C
C
60 NSF
TOILET, SPECIMEN(TPSP1)
(2.44m)
8'-0"
(2
.29m
)
7'-6
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 11:55:12 AM
TOILET, SPECIMEN (TPSP1)REFLECTED CEILING PLAN
LEASING
TOILET, WALL HUNGP9050
DISPOSAL, SANITARY
NAPKIN
A5090
DISPENSER, TOILET ROLLA5200
GRAB BARA5109
CABINET, SPECIMEN,
PASS THRU, CRS
L1200
HOOK, GARMENTA5145
LAVATORY, VITREOUS
CHINA
P3100
DISPENSER, SOAPA5075
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
TOILET, WALL HUNGP9050
DISPOSAL, SANITARY
NAPKIN
A5090
DISPENSER, TOILET ROLLA5200
DISPENSER, PAPER
TOWEL
A5080GRAB BAR
A5109
BASKET, WASTEPAPERF2000
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 11:55:12 AM
TOILET, SPECIMEN (TPSP1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, SOAPA5075
LAVATORY, VITREOUS
CHINA
P3100
BASKET, WASTEPAPERF2000
CABINET, SPECIMEN,
PASS THRU, CRS
L1200
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 11:55:12 AM
TOILET, SPECIMEN (TPSP1)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 11:55:12 AM
TOILET, SPECIMEN (TPSP1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
CT; PT at 4’ - 0”
m: PT ; RES at h: 4" (100mm) (Note 2)
PT ; RES (Note 2)
3" (75mm)
STC 40
m:Wood t:19 (Note 1) s:U
--
--
--
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
Yes
--
--
--
--
--
--
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door2) Ref. PG-18-14 Notes 7 and 74
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:28:12 AM
PROCEDURE ROOM, GENERAL (TRPR1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:28:13 AM
PROCEDURE ROOM, GENERAL (TRPR1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:28:13 AM
PROCEDURE ROOM, GENERAL (TRPR1)AXONOMETRIC
LEASING
+44
"
+44"
+44
"
+48"
1
2
3
4
SPHYGMOMANOMETER,
ANEROID
M4100
OTOSCOPE /
OPHTHALMOSCOPE
M4200
STAND, IVM4255
HOOK, GARMENTA5145
CLOCK, BATTERY, 12"F3200
STAND, MAYOM8810
STOOL, SELF-ADJUSTINGF0340
MONITOR, VITAL SIGNSM4116
ELECTROCARDIOGRAPH,
PORTABLE
M7710
DISPENSER, SOAPA5075
DISPENSER, PAPER
TOWEL
A5080
WORKSTATION,
COMPUTER,
RETRACTABLE
M1802
TELEPHONE, WALL-MNTD,
1 LINE
A1012
RAIL, ACCESSORYA1132
LIGHT, EXAM,
WALL-MOUNTED
M7415
COMPUTER WITH FLAT
PANEL MONITOR
M1801
CABINETCD040
TRACK, CUBICLE,
SURFACE MOUNTED
A5180
180 NSF
PROCEDURE ROOM,
GENERAL(TRPR1)
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
TRACK, CUBICLE,
SURFACE MOUNTED
A5180
STRETCHER, MOBILEM4655
(4
.27m
)
14
'-0
"
HAMPER, LINENM3070
LOCKER, MEDICATIONE0915
SINK, SS, SINGLECS010
CART, STORAGEE0948
BOARD, BULLETINF3010
INFECTIOUS WASTE BAGM3072
WASTE DISPOSAL UNIT,
SHARPS
A5108
BASKET, WASTEPAPERF2000
COUNTERTOP, SOLID
SURFACE
CT020
60 NSF
TOILET, GENERAL
PROCEDURE(TNPG1)
5'-6
"
TO
PA
TIE
NT
CO
RR
IDO
R
TO
ST
AF
F C
OR
RID
OR
(3.91m)
12'-10"
12" MINIMUM
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:28:13 AM
PROCEDURE ROOM, GENERAL (TRPR1)FLOOR PLAN (180 NSF / 16.72 NSM)
LEASING
C C
C
C C
B
AA
C
B
180 NSF
PROCEDURE ROOM,
GENERAL(TRPR1)
(3.91m)
12'-10"
(4
.27m
)
14
'-0
"
(2
.29m
)
7'-6
"
LIFT SYSTEM, OVERHEADA1203
TRACK, CUBICLE,
SURFACE MOUNTED
A5180
12" MINIMUM
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:28:14 AM
PROCEDURE ROOM, GENERAL (TRPR1)REFLECTED CEILING PLAN
LEASING
HOOK, GARMENTA5145
GAP FOR OPTIONAL PATIENT LIFT SYSTEM
BOARD, BULLETINF3010
RAIL, ACCESSORYA1132
OTOSCOPE /
OPHTHALMOSCOPE
M4200
SPHYGMOMANOMETER,
ANEROID
M4100
HOOK, GARMENTA5145
INFECTIOUS WASTE BAGM3072
ELECTROCARDIOGRAPH,
PORTABLE
M7710
WASTE DISPOSAL UNIT,
SHARPS
A5108
CART, STORAGEE0948
LIGHT, EXAM,
WALL-MOUNTED
M7415
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:28:14 AM
PROCEDURE ROOM, GENERAL (TRPR1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
CABINETCD040
COUNTERTOP, SOLID
SURFACE
CT020
SINK, SS, SINGLECS010
TELEPHONE, WALL-MNTD,
1 LINE
A1012
WORKSTATION,
COMPUTER,
RETRACTABLE
M1802
LIGHT, EXAM,
WALL-MOUNTED
M7415
LIFT SYSTEM, OVERHEADA1203
CLOCK, BATTERY, 12"F3200
COMPUTER WITH FLAT
PANEL MONITOR
M1801
BASKET, WASTEPAPERF2000
STRETCHER, MOBILEM4655
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
LOCKER, MEDICATIONE0915
HAMPER, LINENM3070
BASKET, WASTEPAPERF2000
CABINETC02Q0
CABINET, ADA, SINK
SUPPORT
Z1018
TELEPHONE, WALL-MNTD,
1 LINE
A1012
CABINETCD040
SINK, SS, SINGLECS010
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, SOAPA5075
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:28:14 AM
PROCEDURE ROOM, GENERAL (TRPR1)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:28:15 AM
PROCEDURE ROOM, GENERAL (TRPR1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT-SP
--
m: GWB f:SC
--
m: WSF at h: 4" (100mm)
WSF
--
STC 40
m:Wood t:19 (Note 1) s:U
--
Yes
Yes
--
--
--
--
--
--
--
--
--
Yes
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
--
--
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:43:04 AM
UTILITY ROOM, CLEAN (UCCL1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:43:04 AM
UTILITY ROOM, CLEAN (UCCL1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:43:04 AM
UTILITY ROOM, CLEAN (UCCL1)AXONOMETRIC
LEASING
60 NSF
UTILITY ROOM, CLEAN(UCCL1)
1
2
(3.05m)
10'-0"
(1
.83
m)
6'-0
"BASKET, WASTEPAPER
F2000
SUPPLY LOCKER ZONE -
SPECIFIC TYPE BASED ON
LOCATION REQUIREMENTS
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:43:04 AM
UTILITY ROOM, CLEAN (UCCL1)FLOOR PLAN (60 NSF / 5.57 NSM)
LEASING
A
60 NSF
UTILITY ROOM, CLEAN(UCCL1)
(3.05m)
10'-0"
(1
.83
m)
6'-0
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:43:05 AM
UTILITY ROOM, CLEAN (UCCL1)REFLECTED CEILING PLAN
LEASING
NO STORAGE WITHIN
18" OF FINISH CEILING
SUPPLY LOCKER
ZONE - SPECIFIC TYPE
BASED ON LOCATION
REQUIREMENTS
BASKET, WASTEPAPERF2000
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:43:05 AM
UTILITY ROOM, CLEAN (UCCL1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:43:05 AM
UTILITY ROOM, CLEAN (UCCL1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT-SP
--
m: GWB f:SC
RWC at 4’ – 0”
m: WSF at h: 4" (100mm)
WSF
--
STC 40
m:Wood t:11 dg:F dr: 45 min ds:Yes s:V
--
Yes
--
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) Ref. Note 9 and 54, PG-18-14
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:55:57 AM
SOILED HOLDING (USCL1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:55:57 AM
SOILED HOLDING (USCL1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:55:57 AM
SOILED HOLDING (USCL1)AXONOMETRIC
LEASING
80 NSF
SOILED HOLDING
(USCL1)
4
3
(3.81m)
12'-6"
(1
.96m
)
6'-5
"
1
DISPENSER, HAND
SANITIZER, HANDS FREE
A5077
INFECTIOUS WASTE BAGM3072
HAMPER, LINENM3070
EYEWASH, WALL-MNTD,
HANDS-FREE
P2000
DISPENSER, SOAPA5075
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
2
CAN, TRASH, 44 GALF2020
DISPENSER, PAPER
TOWEL
A5080
LAVATORY, VITREOUS
CHINA
P3100
COUNTER, STAINLESS
STEEL
CT050
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:55:58 AM
SOILED HOLDING (USCL1)FLOOR PLAN (80 NSF / 7.43 NSM)
LEASING
B
80 NSF
SOILED HOLDING
(USCL1)
(3.81m)
12'-6"
(1
.96m
)
6'-5
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:55:58 AM
SOILED HOLDING (USCL1)REFLECTED CEILING PLAN
LEASING
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
COUNTER, STAINLESS
STEEL
CT050
INFECTIOUS WASTE BAGM3072
HAMPER, LINENM3070
COUNTER, STAINLESS
STEEL
CT050
INFECTIOUS WASTE BAGM3072
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:55:58 AM
SOILED HOLDING (USCL1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
DISPENSER, HAND
SANITIZER, HANDS FREE
A5077
DISPENSER, HAND
SANITIZER, HANDS FREE
A5077
EYEWASH, WALL-MNTD,
HANDS-FREE
P2000
LAVATORY, VITREOUS
CHINA
P3100
DISPENSER, PAPER
TOWEL
A5080
DISPENSER, SOAPA5075
HAMPER, LINENM3070
COUNTER, STAINLESS
STEEL
CT050
CAN, TRASH, 44 GALF2020
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 3:55:58 AM
SOILED HOLDING (USCL1)ELEVATIONS
ELEVATION 4
ELEVATION 3
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 3:55:58 AM
SOILED HOLDING (USCL1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
GWB
--
m: GWB f:P
RWC at 4’ – 0”
m: WSF at h: 4" (100mm)
WSF
--
STC 40
m:Wood t:11 dg:F dr: 45 min ds:Yes s:V
--
Yes
--
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) Ref. Note 9 and 54, PG-18-14
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 4:32:27 AM
TELE-HEALTH ROOM (WRTM2)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 4:32:28 AM
TELE-HEALTH ROOM (WRTM2)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 4:32:28 AM
TELE-HEALTH ROOM (WRTM2)AXONOMETRIC
LEASING
125 NSF
TELE-HEALTH ROOM
(WRTM2)
1
2
3
4
DISPENSER, SOAPA5075
DISPENSER, PAPER
TOWEL
A5080
RAIL, ACCESSORYA1132
BASKET, WASTEPAPERF2000
BOARD, BULLETINF3025
CHAIR, SIDE WITH ARMSF0205
SPHYGMOMANOMETER,
ANEROID
M4100
OTOSCOPE /
OPHTHALMOSCOPE
M4200
TABLE, EXAM/TREATMENTM9025
LIGHT, EXAM, MOBILEM7401
CLOCK, BATTERY, 12"F3200
STOOL, SELF-ADJUSTINGF0340
WORKSTATION,
COMPUTER
M1803
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
TRACK WITH CURTAINA5180
HOOK, GARMENTA5145
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"
DISPENSER, HAND
SANITIZER
A5077
WASTE DISPOSAL UNIT,
SHARPS
A5108
CART, STORAGEE0948
SINK, SS, SINGLECS010
COUNTERTOP, SOLID
SURFACE
CT020
TELEPHONE, WALL-MNTD,
1 LINE
A1012
COMPUTER WITH FLAT
PANEL MONITOR
M1801
5'-6
"
TELEVISION, FLAT
SCREEN
M0506
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 4:32:28 AM
TELE-HEALTH ROOM (WRTM2)FLOOR PLAN (125 NSF / 11.61 NSM)
LEASING
A
A
TRACK WITH CURTAINA5180
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 4:32:29 AM
TELE-HEALTH ROOM (WRTM2)REFLECTED CEILING PLAN
LEASING
CLOCK, BATTERY, 12"F3200
RAIL, ACCESSORYA1132
OTOSCOPE /
OPHTHALMOSCOPE
M4200
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
HOOK, GARMENTA5145
SPHYGMOMANOMETER,
ANEROID
M4100
DISPENSER, HAND
SANITIZER
A5077
TRACK WITH CURTAINA5180
WASTE DISPOSAL UNIT,
SHARPS
A5108
CART, STORAGEE0948
RAIL, ACCESSORYA1132
BOARD, BULLETINF3025
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 4:32:29 AM
TELE-HEALTH ROOM (WRTM2)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
COMPUTER WITH FLAT
PANEL MONITOR
M1801
WORKSTATION,
COMPUTER
M1803
TELEPHONE, WALL-MNTD,
1 LINE
A1012
CHAIR, SIDE WITH ARMSF0205
CART, STORAGEE0948 SINK, SS, SINGLE
CS010
COUNTERTOP, SOLID
SURFACE
CT020
WASTE DISPOSAL UNIT,
SHARPS
A5108
TELEVISION, FLAT
SCREEN
M0506
CABINET, 1 DOORZ1020
TRACK WITH CURTAINA5180
TELEPHONE, WALL-MNTD,
1 LINE
A1012
DISPENSER, PAPER
TOWEL
A5080
COUNTERTOP, SOLID
SURFACE
CT020
DISPENSER, HAND
SANITIZER
A5077
BASKET, WASTEPAPERF2000
SINK, SS, SINGLECS010
DISPENSER, SOAPA5075
CABINET, 1 DOORZ1020
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 4:32:30 AM
TELE-HEALTH ROOM (WRTM2)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 4:32:30 AM
TELE-HEALTH ROOM (WRTM2)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: RB at h: 4" (100mm)
CPT; LVT
--
STC 40
m:Wood t:19 (Note 1) s:S
--
Yes
Yes
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
--
--
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door2) Ref. Note 6 and 64, PG-18-14
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 7:29:34 AM
TELE-RETINAL ROOM (WRTM3)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 7:29:34 AM
TELE-RETINAL ROOM (WRTM3)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 7:29:35 AM
TELE-RETINAL ROOM (WRTM3)AXONOMETRIC
LEASING
125 NSF
TELE-RETINAL ROOM
(WRTM3)
1
2
3
4
CLOCK, BATTERY, 12"F3200
SPHYGMOMANOMETER,
ANEROID
M4100
OTOSCOPE /
OPHTHALMOSCOPE
M4200
RAIL, ACCESSORYA1132
STOOL, SELF-ADJUSTINGF0340
LIGHT, EXAM, MOBILEM7401
TABLE, EXAM/TREATMENTM9025
WASTE DISPOSAL UNIT,
SHARPS
A5108
CART, STORAGEE0948
CHAIR, SIDE WITH ARMSF0205
WORKSTATION,
COMPUTER
M1803
TELEVISION, FLAT
SCREEN
M0506
TELEPHONE, WALL-MNTD,
1 LINE
A1012
COMPUTER WITH FLAT
PANEL MONITOR
M1801
DISPENSER, PAPER
TOWEL
A5080
SINK, SS, SINGLECS010
COUNTERTOP, SOLID
SURFACE
CT020
DISPENSER, SOAPA5075
BASKET, WASTEPAPERF2000
TRACK, CUBICLE, WITH
CURTAIN
A5180HOOK, GARMENT
A5145
DISPENSER, HAND
SANITIZER
A5077
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
5'-6
"
RETINAL IMAGING
SYSTEM
M6050
BOARD, BULLETINF3025
CABINET, 1 DOORZ1020
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 7:29:35 AM
TELE-RETINAL ROOM (WRTM3)FLOOR PLAN (125 NSF / 11.61 NSM)
LEASING
A
A
125 NSF
TELE-RETINAL ROOM
(WRTM3)
(3.81m)
12'-6"
(3
.05
m)
10
'-0
"
TRACK, CUBICLE, WITH
CURTAIN
A5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 7:29:35 AM
TELE-RETINAL ROOM (WRTM3)REFLECTED CEILING PLAN
LEASING
CLOCK, BATTERY, 12"F3200
SPHYGMOMANOMETER,
ANEROID
M4100
OTOSCOPE /
OPHTHALMOSCOPE
M4200
RAIL, ACCESSORYA1132
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
DISPENSER, HAND
SANITIZER
A5077
HOOK, GARMENTA5145
TRACK, CUBICLE, WITH
CURTAIN
A5180
LIGHT, EXAM, MOBILEM7401
TABLE, EXAM/TREATMENTM9025
OTOSCOPE /
OPHTHALMOSCOPE
M4200
WASTE DISPOSAL UNIT,
SHARPS
A5108
BOARD, BULLETINF3025
CART, STORAGEE0948
CHAIR, SIDE WITH ARMSF0205
RETINAL IMAGING
SYSTEM
M6050
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 7:29:35 AM
TELE-RETINAL ROOM (WRTM3)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
COMPUTER WITH FLAT
PANEL MONITOR
M1801
TELEVISION, FLAT
SCREEN
M0506
WORKSTATION,
COMPUTER
M1803
TELEPHONE, WALL-MNTD,
1 LINE
A1012
SINK, SS, SINGLECS010
BASKET, WASTEPAPERF2000
CHAIR, SIDE WITH ARMSF0205
RETINAL IMAGING
SYSTEM
M6050
WASTE DISPOSAL UNIT,
SHARPS
A5108
CART, STORAGEE0948
BOARD, BULLETINF3025
CABINET, 1 DOORZ1020
COUNTERTOP, SOLID
SURFACE
CT020
DISPENSER, SOAPA5075
SINK, SS, SINGLECS010
DISPENSER, PAPER
TOWEL
A5080
TELEPHONE, WALL-MNTD,
1 LINE
A1012
BASKET, WASTEPAPERF2000
TRACK, CUBICLE, WITH
CURTAIN
A5180
COUNTERTOP, SOLID
SURFACE
CT020
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 7:29:35 AM
TELE-RETINAL ROOM (WRTM3)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 7:29:36 AM
TELE-RETINAL ROOM (WRTM3)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: RB at h: 4" (100mm)
CPT; LVT
--
STC 40
m:Wood t:19 (Note 1) s:S
--
Yes
Yes
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
--
--
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door2) Ref. Note 6 and 64, PG-18-14
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:00:51 PM
RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME
(ROOM CODE)AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:00:51 PM
RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:00:51 PM
RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)AXONOMETRIC
LEASING
1
3
4 2
300 NSF
GENERAL PURPOSERADIOLOGY ROOM
(XDR01)
MIRROR, FLOATGLASS, ADA
ACCESSIBLE
A1067
STAND, BUCKY,VERTICAL, TILT,
AUTOMATIC
M1405
DISPENSER, SOAPA5075
DISPENSER, PAPERTOWEL
A5080
FOOTSTOOL,STRAIGHT
F0355
BASKET,WASTEPAPER
F2000
STAND, IVM4255
HOOK, GARMENTA5145
TELEPHONE,WALL-MNTD, 1 LINE
A1012
CHAIR, SIDE WITHARMS
F0205
CLOCK, BATTERY, 12"F3200
PATI
ENT
ENTR
Y (4
8" D
OOR)
(1.52m)5'-0"
(5.77m)18'-11"
DISPENSER, GLOVE,SURGICAL-EXAMINATION
A5107
LAVATORY,VITREOUS CHINA
P3100
5'-6"
FUTURE EXPANSI
ON (MIRROR
ED RADIOLO
(2.59m)8'-6"
HAMPER, LINENM3070
WASTE DISPOSALUNIT, SHARPS
A5108
RADIOGRAPHIC UNIT,80 kW
X5900
(1.2
7m)
4'-2
"
(4.8
8m)
16'-0
"
CART, STORAGEE0948
RACK, APRONX3150
COUNTER TOP, SOLIDSURFACE
CT020
CONTROL SYSTEMS
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
3/13/2019 2:35:21 PMLEASING RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)FLOOR PLAN (300 NSF / 27.87 NSM)
C C C
C C C
C C C
300 NSF
GENERAL PURPOSE
RADIOLOGY ROOM
(XDR01)
(5.77m)
18'-11"
(4
.88m
)
16
'-0
"
LIFT SYSTEM, OVERHEADA1203
DASHED LINES INDICATE
METAL FRAMING SYSTEM
REQUIRED FOR CEILING
MOUNTED EQUIPMENT
*PROVIDE A MINIMUM OF
(1) OVERHEAD PATIENT
LIFT, WHERE FEASIBLE
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:00:52 PM
RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)REFLECTED CEILING PLAN
LEASING
TELEPHONE, WALL-MNTD,
1 LINE
A1012
DISPENSER, SOAPA5075DISPENSER, PAPER
TOWEL
A5080
BASKET, WASTEPAPERF2000
LAVATORY, VITREOUS
CHINA
P3100
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
CHAIR, SIDE WITH ARMSF0205
HOOK, GARMENTA5145
WASTE DISPOSAL UNIT,
SHARPS
A5108
HAMPER, LINENM3070
TELEPHONE, WALL-MNTD,
1 LINE
A1012
CART, STORAGEE0948
RACK, APRONX3150
RADIOGRAPHIC UNIT, 80
kW
X5900
RADIOGRAPHIC UNIT, 80
kW
X5900
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:00:53 PM
RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
STAND, BUCKY, VERTICAL,
TILT, AUTOMATIC
M1405
CLOCK, BATTERY, 12"F3200
CART, STORAGEE0948
RADIOGRAPHIC UNIT, 80
kW
X5900
RADIOGRAPHIC UNIT, 80
kW
X5900
RADIOGRAPHIC UNIT, 80
kW
X5900
MIRROR, FLOAT GLASS,
ADA ACCESSIBLE
A1067
LIFT SYSTEM, OVERHEADA1203
RADIOGRAPHIC UNIT, 80
kW
X5900
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 2:00:53 PM
RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 2:00:54 PM
RADIOLOGY ROOM, GENERAL PURPOSE (XDR01)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: WSF at h: 4" (100mm)
WSF
--
STC 40
m:Wood t:15 dl: LEAD s:S
--
Yes
Yes
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
Dimmable to 5 FC
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) Provide a 4’-0” wide shielded door into the General PurposeRadiographic Room.2) Provide a shielded viewing window from the Control Area to theGeneral Purpose Radiographic Room.
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 8:51:35 AM
ULTRASOUND ROOM (XDUS1)DISCLAIMER
DISCLAIMER:
ROOM TEMPLATES ARE A CRITICAL COMPONENT TO VA TECHNICAL INFORMATION LIBRARY (TIL) WHICH PROVIDES STANDARDS FOR ALL VA PLANNING, DESIGN, AND CONSTRUCTION PROJECTS. ROOM TEMPLATES ARE DEPARTMENT AND ROOM ALIGNED CRITERIA DESCRIBING SPACE, ROOM CONTENTS, AND OTHER TECHNICAL REQUIREMENTS FOR THE DEVELOPMENT OF VA PROJECTS. ROOM TEMPLATES COMMUNICATE THE BASIS OF DESIGN (BOD) AND ARE REQUIRED TO BE UTILIZED BY PROJECT TEAMS WORKING ON NEW CONSTRUCTION AND RENOVATIONS OF EXISTING FACILITIES. THE MATERIAL CONTAINED IN THE ROOM TEMPLATE CONSTITUTES A STANDARD FOR VA PLANNING, DESIGN, AND CONSTRUCTION. ANY SUBSTANTIAL VARIANCE FROM STANDARDS SHALL BE CONSIDERED ONLY AS REQUIRED TO ACCOMMODATE SPECIFIC SITE, FUNCTIONAL, AND OPERATIONAL CONDITIONS. EACH SUBSTANTIAL VARIANCE SHALL HAVE A BASIS OF RATIONALE AND BE DOCUMENTED IN THE PROJECT RECORD.
CLINICIANS, PROVIDERS, PRIMARY USERS, AND OTHER STAKEHOLDERS SHALL BE INVOLVED IN PROJECT SPECIFIC DEVELOPMENT OF ROOM TEMPLATES AND BIM TEST-FITS TO BEST ADAPT STANDARDS FOR SPECIFIC FUNCTIONAL, OPERATIONAL , AND SITE CONDITIONS AND TO PROVIDE OPTIMUM SERVICES ENVIRONMENTS FOR VETERANS. STAKEHOLDER INVOLVEMENT AND REQUIREMENTS SHALL BE DOCUMENTED IN THE PROJECT RECORD.
ROOM TEMPLATES ARE NOT PROJECT-SPECIFIC. SITE SPECIFIC ISSUES MUST BE ADDRESSED WITH THE CONTEXT OF VA STANDARDS AND APPLIED TO EACH INDIVIDUAL PROJECT. USE OF THIS ROOM TEMPLATE DOES NOT PRECLUDE THE NEED FOR , NOR ABSOLVE PLANNERS, DESIGNERS, AND CONSTRUCTORS OF THEIR RESPONSIBILITY TO PROVIDE COMPLETE, FUNCTIONAL, SAFE, AND SECURE DESIGNS SUITED TO THE UNIQUE REQUIREMENTS OF EACH PROJECT.
EQUIPMENT AND SYSTEMS ARE SHOWN IN AN ILLUSTRATIVE, PERFORMANCE-BASED FORMAT AND ARE NOT INTENDED TO DEPICT, SUGGEST, OR OTHERWISE CONSTITUTE ENDORSEMENT OF ANY SPECIFIC PRODUCT OR MANUFACTURER. MANUFACTURERS SHOULD BE CONSULTED FOR ACTUAL DIMENSIONS, CONFIGURATIONS, AND UTILITY REQUIREMENTS.
LEASING
BA C
D2
1
1
1
1
CENTERLINE
EQUIPMENT JSN
NUMBER & NAMEELEVATION
REFERENCE
2X2 TROFFER LIGHT
FIXTURE
2X4 TROFFER LIGHT
FIXTURE
EXHAUST REGISTER RETURN REGISTER SUPPLY DIFFUSER
2X2 ACOUSTIC CEILING
SYSTEM
RECESSED CAN
LIGHT FIXTURE
CLEARANCE LINESOVERHEAD LINES
COMPOUND CEILING,
GYPSUM BOARD ON
METAL STUD
ROOM TAG
SINGLE SWITCH DIMMER SWITCH
WALL-MOUNTED
QUADRAPLEX OUTLET
FLOOR-MOUNTED
DUPLEX OUTLET
WALL-MOUNTED
DUPLEX OUTLET
FLOOR-MOUNTED
DATA OUTLET
WALL-MOUNTED
DATA OUTLET
Type CommentsJSN#
SPRINKLER HEAD
MULTIPLE SWITCH
(SUBSCRIPT INDICATES
QUANTITY)
FLOOR-MOUNTED
QUADRAPLEX OUTLET
ROOM NAME(ROOM CODE)
AREA NSF
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 8:51:35 AM
ULTRASOUND ROOM (XDUS1)LEGEND
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 8:51:36 AM
ULTRASOUND ROOM (XDUS1)AXONOMETRIC
LEASING
180 NSF
ULTRASOUND ROOM(XDUS1)
1
2
3
4
(4
.57
m)
15
'-0
"5'-6
"
(3.66m)
12'-0"
STAND, MAYOM8810
CART, COMPUTER,
MOBILE
E0945
COMPUTER WITH FLAT
PANEL MONITOR
M1801
SCANNER, ULTRASOUND,
GENERAL PURPOSE
X2100
STRETCHER, MOBILEM4655
CHAIR, SIDE WITH ARMSF0205
HOOK, GARMENTA5145
CLOCK, BATTERY, 12"F3200
STOOL, ANESTHESIA,
WITH BACK
M8940
LOCKER, SUPPLY, 6
DRAWERS, MOBILE
E0930
60 NSF
TOILET, GENERAL(TNPG1)
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
FOOTSTOOL, STRAIGHTF0355
TELEPHONE, WALL-MNTD,
1 LINE WITH SPEAKER
A1014
INFECTIOUS WASTE BAGM3072
DISPENSER, PAPER
TOWEL, SENSOR, HANDS
FREE
A5082
BASKET, WASTEPAPERF2000
OPTIONAL DOOR
DIRECT TO CORRIDOR
DISPENSER, SOAPA5075
LAVATORY, VITREOUS
CHINA
P3100
WASTE DISPOSAL UNIT,
SHARPS
A5108
LIGHT, OVERBED, DIRECT
AND INDIRECT
M7435
(2.44m)
8'-0"
(2
.29
m)
7'-6
"1
2"
MIN
IMU
M
TRACK, CUBICLE, WITH
CURTAIN
A5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 8:51:36 AM
ULTRASOUND ROOM (XDUS1)FLOOR PLAN (180 NSF / 16.72 NSM)
LEASING
C
C C
C
C
C
180 NSF
ULTRASOUND ROOM(XDUS1)
(4
.57
m)
15
'-0
"
(3.66m)
12'-0"
60 NSF
TOILET, GENERAL(TNPG1)
(2.44m)
8'-0"
(2.2
9m
)
7'-6
"
TRACK, CUBICLE, WITH
CURTAIN
A5180
12"
MIN
IMU
M
TRACK, CUBICLE,
SURFACE MOUNTED
A5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 8:51:36 AM
ULTRASOUND ROOM (XDUS1)REFLECTED CEILING PLAN
LEASING
LOCKER, SUPPLY, 6
DRAWERS, MOBILE
E0930
DISPENSER, PAPER
TOWEL, SENSOR, HANDS
FREE
A5082
LAVATORY, VITREOUS
CHINA
P3100
CLOCK, BATTERY, 12"F3200
TRACK, CUBICLE, WITH
CURTAIN
A5180
DISPENSER, SOAPA5075
COMPUTER WITH FLAT
PANEL MONITOR
M1801
CART, COMPUTER,
MOBILE
E0945
SCANNER, ULTRASOUND,
GENERAL PURPOSE
X2100
STRETCHER, MOBILEM4655
CHAIR, SIDE WITH ARMSF0205
FOOTSTOOL, STRAIGHTF0355
LIGHT, OVERBED, DIRECT
AND INDIRECT
M7435
TELEPHONE, WALL-MNTD,
1 LINE WITH SPEAKER
A1014
TRACK, CUBICLE,
SURFACE MOUNTED
A5180
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 8:51:37 AM
ULTRASOUND ROOM (XDUS1)ELEVATIONS
ELEVATION 1
ELEVATION 2
LEASING
STRETCHER, MOBILEM4655
CART, COMPUTER,
MOBILE
E0945HOOK, GARMENT
A5145
STAND, MAYOM8810
LIGHT, OVERBED, DIRECT
AND INDIRECT
M7435
TRACK, CUBICLE,
SURFACE MOUNTED
A5180
STAND, MAYOM8810
DISPENSER, PAPER
TOWEL, SENSOR, HANDS
FREE
A5082
DISPENSER, GLOVE,
SURGICAL-EXAMINATION
A5107
BASKET, WASTEPAPERF2000
WASTE DISPOSAL UNIT,
SHARPS
A5108
LAVATORY, VITREOUS
CHINA
P3100
INFECTIOUS WASTE BAGM3072
TRACK, CUBICLE, WITH
CURTAIN
A5180
DISPENSER, SOAPA5075
HOOK, GARMENTA5145
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
1/4" = 1'-0"
6/28/2018 8:51:37 AM
ULTRASOUND ROOM (XDUS1)ELEVATIONS
ELEVATION 3
ELEVATION 4
LEASING
0' 4' 8' 16'
SCALE:
SKETCH TITLE
Plot Date:
DISCLAIMER: ROOM TEMPLATES ARE GRAPHICAL REPRESENTATIONS OF SELECTED ROOM TYPES THAT ILLUSTRATE VA PLANNING REQUIREMENTS FOR SPACE, ROOM
CONTENTS, AND ROOM SPECIFIC ENGINEERING SYSTEMS. THEY PROVIDE TYPICAL CONFIGURATIONS, PLANNING CRITERIA, AND GENERAL TECHNICAL GUIDANCE, AND
ARE NOT INTENDED TO BE PROJECT SPECIFIC REQUIREMENTS.
6/28/2018 8:51:37 AM
ULTRASOUND ROOM (XDUS1)ROOM DATA
ARCHITECTURAL & INTERIOR DESIGN
Ceiling Type:
Ceiling Height:
Wall Finish:
Base:
Floor Finish:
Special Requirement:
Sound Protection:
Ceiling Finish:
Wainscot:
Slab Depression:
Doors:
Notes:
TELECOMMUNICATION / SPECIAL SYSTEMS
Data:
Telephone:
Duress Alarm:
Intercom:
Motion Intrusion Detection (MID):
Security Surveillance Television (SSTV):
Code Blue:
Cable Television:
Electronic Access and Door Control:
Nurse Call:
Public Address:
VA Satellite TV:
Video Teleconferencing (VTEL):
Special Requirement:
Notes:
LIGHTING
Special Requirement:
Luminaire Type:
Notes:
POWER
Electrical system design to comply with the latest NFPA 70 and 99
Design in accordance with project requirements.
Emergency Power:
Normal Power:
Special Requirement:
Notes:
HEATING, VENTILATING, AND AIR CONDITIONING
Refer to HVAC Design Manual.
Design in accordance with project requirements.Refer to the current version of PG-18-10 Chapter 6 Table 5 –"Mechanical Room Data Sheets" for room temperatures, humidity range, room air change requirements, and pressurization
Notes:
PLUMBING AND MEDICAL GASES
Design in accordance with project requirements.
Cold Water:
Drain:
Medical Air:
Medical Vacuum:
Special Requirement:
Hot Water:
Reagent Grade Water:
Oxygen:
Notes:
FIRE PROTECTION AND LIFE SAFETY
Design in accordance with project requirements.
Alarm Detection:
Sprinkler:
Alarm Annunciator:
Notes:
Ref. Door Type Detail #SD080000-01 for door typesRef. Door Sizes Detail #SD080000-02 for door sizes
General Requirement: Lighting system design shall be based on the guidance of applicable Illuminating Engineering Society North America (IESNA) Standards.
9'-0" (2700mm)
AT
--
m: GWB f:P
--
m: WSF at h: 4" (100mm)
WSF
--
STC 40
m:Wood t:19 (Note 1) s:X
--
Yes
Yes
--
--
--
--
--
--
--
--
--
--
--
--
Design in accordance with project requirements.
--
Connect to receptacles and equipment
--
--
Yes
Yes
--
--
--
--
--
--
Visual (15cd), center of tile
--
Recessed pendant, center of tile
1) 40 STC rated acoustical door
LEASING