7/27/2019 Pm-51 touch screen1.pdf
1/157
Photometer 5010 V5+Version 5.x
> SN 7000
Service Manual
7/27/2019 Pm-51 touch screen1.pdf
2/157
CONTENTS
chapter headline page
1 General notes
1.1 Contact partner 1.1
1.2 Proved security 1.1
1.3 Precautions and hazards 1.2
2 Service concepts
2.1 Service level 2.1
2.2 Conditions 2.2
3.1 Block diagram 3.1.1
3.2 Short instruction guide 3.2.1
3.3 Service tools 3.3.1
3.4 Plug identification 3.4.1
4.1 Mains power circuit 4.1.1
4.2 Mains connector 4.2.1
4.3 AC / DC power supply 4.3.1
4.4 Cable power supply / printer 4.4.1
4.10 Optical unit 4.10.1
4.11 Filter wheel 4.11.1
4.12 Halogen lamp 4.12.1
4.13 Unit Photodiode 4.13.1
4.14 Unit cuvette adaptor 4.14.1
4.15 Unit Peltier element 4.15.1
4.16 Step motor pump 4.16.1
4.18 Step motor filter wheel 4.18.14.19 Light barrier filter wheel 4.19.1
4.20 Bubble detector with plug 4.20
4.21 Flow-through cuvette adaptor 4.21
4.22 Standard cuvette adaptor 4.22
4.23 Connector bubble detector 4.23
4.24 Lamp holder 4.24
4.26 Tubes 4.26.1
4.30 Top case 4.30.1
4.31 Top case pre-assembled 4.31.1
4.32 Metal tube inlet 4.32.1
RIELE BERLIN Photometer 5010 V5+ 1 23.07.2013
7/27/2019 Pm-51 touch screen1.pdf
3/157
CONTENTS
chapter headline page
4.33 Microswitch with plug 4.33.1
4.34 Plastic foil touchscreen 4.34.1
4.35 Printer mechanism 8NAD 4.35.1
4.40 PCB RR-120 Mother board
Function description 4.40.1
Component diagram 4.40.2
Voltages 4.40.4
Voltage 12 VDCof halogen lamp 4.40.5
Fuses F1, F2 and F3 4.40.5
Amplifier for photocurrent and ADC 4.40.6
Adjustment of amplifier stage 4.40.7
Temperature control 4.40.9
Adjustment of temperature control 4.40.10
Thermal adjustment 4.40.11
Installation instruction 4.40.12
Circuit diagrams 4.40.14
4.41 Touchscreen 4.41.1
4.43 PCB RR-122 Display, printer, touch 4.43.1
4.44 Printer controller 8NAD 4.44.1
4.45 PCB RR_121 Motor driver 4.45.1
4.50 Fan 4.50.1
4.51 Bottom case 4.51.1
4.52 Carton Photometer 5010 4.52.1
5 Assembling / Dismantling
6.1 Communication protocol 6.1.1
6.2 Download of operating system:
Resume after malfunction 6.2.1
6.3 Initialization of bootstrap 6.3.16.4 Configuration of operating system 6.4.1
6.5 Download of operating system using 6.5.1a programmed flash memory
7 Trouble shooting 7.1.1
8 Spare parts 8.1.1
9.1 Calibration procedure of Photometer 5010Subject: Cuvette temperature 9.1.1
RIELE BERLIN Photometer 5010 V5+ 2 23.07.2013
7/27/2019 Pm-51 touch screen1.pdf
4/157
1 GENERAL NOTES
The data and information provided in this manual correspond to the state of knowledge existing atthe time of introducing the Photometer 5010 V5+ on the market. Any important changes will betaken into account in the next edition of this manual.
In any case, the respective packaging leaflet should be regarded as authoritative.
This service manual was created for the telephone service and technical service staff.
1.1 Contact Partner
Technical product management and support:
ROBERT RIELE GmbH & Co KG
Address: Kurfuerstenstrae 75 - 79D - 13467 BerlinGermany
Phone: +49 / 30 / 404 40 87Fax: +49 / 30 / 404 05 29
E-mail: [email protected]
1.2 Proved Security
This device has been constructed under the regulations of EN 61010-1 andEN 61010-2-101.
This device was examined and left the factory in perfect technical condition. To preserve this and
protect safe and faultless operation, the user must follow the instructions and remarks of thisservice manual.
To preserve this condition and to guarantee a safe working, information and working advices,which are included in these working instructions, have to be heeded by the applier.
This device meets protection 1 (earth conductor wire).
1.3 Precaution and Hazards
All electrical equipment is potentially hazardous. Never remove covers without first ensuring that it
is isolated from the AC supply, unless specific maintenance instructions or repairs are being carriedout by authorized ROBERT RIELE personnel.
The hard- and software is subject to a program of continuous evaluation and improvement and,therefore, may be changed in the future. This also concerns service requirements.
All samples and reagents should be treated with caution accorded to those known to containpathogenic organisms. Similarly, the cleaning of component part of Photometer 5010 V5+shouldbe carried out in compliance with health and safety.
RIELE BERLIN Photometer 5010 V5+ 1.1 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
5/157
1 GENERAL NOTES
Warning: All components used on Photometer 5010 V5+ must be regarded as potentiallydangerously contaminated when doing repair work. Use rubber gloves or double gloves whenevercleaning or sterilizing components. The most frequently used components (tubing, cuvette,measuring chamber flap, sip. lever and waste container) should be cleaned and treated with asuitable disinfecting solution (75% alcohol) prior to doing any service work.
Disinfect and wash hands after work is completed.
This device has been constructed and checked in accordance with standards EN 61010-1 andEN 61010-2-101. When the device leaves our factory it is in perfect condition from the point of viewof work safety. To keep it that way and to ensure safe operation the user must follow theinstructions and warnings given in the operation instruction manual.
The electrical protection of the device corresponds to Class 1 (it has a protective earth).
The power plug must only be inserted into a socket that has a protective contact. The protectionmust not be abolished by using an extension lead that does not have a protective earth wire.
Warning:Any break in the earth wire inside or outside the device and any loose connection of the earth wirecan make the operation of the device dangerous. A deliberate break or interruption is not allowed.
When the housing is opened or the components are being removed (except when this can be doneby hand), live parts may be exposed. Connection may also be live.
Therefore, if the carrying out of an adjustment, service or repair on the open device in the live stateis unavoidable, this must be done by an expert who is familiar with the dangers involved.
Make sure you only use fuses of the specified type and rated amperage when replacing the oldones. Repaired fuses must not be used and the fuse holder must not be bypassed.
If you have any reason to believe that the device can no longer operate safely, take it out of useand make sure no one can use it accidentally.
It must be assumed that safe operation is no longer possible when the device:
shows visible signs of damage
fails to operate
has been stored under unfavorable conditions for a fairly long time
has been transported under rough conditions
The Photometer 5010 V5+should be used by appropriately qualified persons only.
RIELE BERLIN Photometer 5010 V5+ 1.2 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
6/157
2 SERVICE CONCEPTS
2.1 Service Level
From the early stage of development, the Photometer 5010 V5+ was designed for simple errordetection and easy exchangeability of modules. This gives the service workshops the possibility ofa fast and easy repair of the device on service level A (module level). No big stock or expensiveequipment is necessary and service technicians are easier to train. Also, a permanent technicalimprovement in layout and components took and takes place for better productivity and economicproduction.
On repairable modules the quality and function is always provided by the manufacturer accordingto the latest technology. This keeps Photometer 5010 V5+always on the highest quality level.
The exchange price for modules will be kept on a low level to guarantee repairs, on an economicalbasis.
2.2 Conditions
Warranty periodfor devices ...
The warranty period for devices is 16 months starting with the date ofshipment ex works Berlin/Federal Republic of Germany or 12 monthsstarting with the date of the first installation, whichever period is shorter.
... and spare parts For spare parts a warranty of 6 months from installation date of the partapplies. Prerequisite for this is that the warranty period for spare parts shallnot fall short of the warranty period of the device where a spare part hasbeen installed. A further prerequisite is, that this period shall not exceed aperiod of 24 months starting with the delivery of the spare part ex works
Berlin/Federal Republic of Germany.
Hint:In case the device has a remaining warranty period of more than 6 months,the parts remain under warranty until the warranty period of the deviceexpires.
Handling ofwarranty c laims
Replacement of partsThe defective parts have to be handled via Return Authorization (RA)procedure. Please answer all the questions on the RA form with the greatestcare.Especially a detailed fault description is needed or the warranty claim will notbe accepted by the manufacturer.Complete devices are notacceptedunless this has been agreed with theservice department of ROBERT RIELE GmbH & Co KG.
Important information:
Only parts marked with "A" in the price list are generally accepted underwarranty.
Only return those parts marked with "R" in the spare part list.
All defective parts ( non-"R" and "A" parts ) should be kept on stock fora period of 7 months. In case the manufacturer needs the part forinvestigation it will be requested from Berlin.
All parts returned to Berlin and not requested by Berlin will be sent back
at the expense of the dealer/customer.
RIELE BERLIN Photometer 5010 V5+ 2.1 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
7/157
2 SERVICE CONCEPTS
Exclusion ofwarranty
The aforementioned warranties do not apply in case of improper use,handling, transportation or storage, faulty installation, repair or maintenance,Chemical influence or contamination as well as damages resulting from that,failure to follow operating instructions, alterations or modifications of devicesor parts thereof not authorized or recommended by ROBERT RIELE GmbH& Co KG and resulting damages, normal wear and tear and in case of other
circumstances beyond the control of ROBERT RIELE GmbH & Co KG.
Handling ofrepairs
As a general rule, all device repairs should be carried out by authorized andtrained personnel only.
Repair of parts marked with " R"Parts which are economically worth repairing are marked with "R" in thespare parts price list. The defective parts have to be handled via ReturnAuthorization procedure. For correct handling of the exchange it isabsolutely necessary to return the complete filled in RA form (preferable viaelectronic mail), giving full details of the defect.
Repair of devicesComplete devices are not accepted for replacement or repair unless thishas been agreed with ROBERT RIELE GmbH & Co KG.Before replacement or repair can take place, the validity of the request mustbe examined and the question of costs must be settled in a writtenagreement with ROBERT RIELE GmbH & Co KG.
Terms of delivery All shipping conditions for spare parts to the dealer/customer areex works Berlin.
Emergency shipments require additional costs to be charged.
Emergencyshipment
In case of emergency, a copy of the complete RA form should be sent by fax(+49 / 30 / 404 05 29) or electronic mail (e-mail: [email protected]) toROBERT RIELE GmbH & Co KG, Berlin
marked with EMERGENCY
If fax was used for Emergency RA the original RA form has to be mailed orsent by electronic mail to Berlin.
Handling of costs Replacements for R-parts are shipped at a repair price. In case the parts are
no longer repairable, the dealer/customer will be charged later on with thedifference of the new price.Replacements for warranty requests are shipped free of charge.In case the manufacturer does not accept the warranty request, thedealer/customer will be charged later on with the R-price for R-parts and thenew price for non R-parts.
RIELE BERLIN Photometer 5010 V5+ 2.2 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
8/157
2 SERVICE CONCEPTS
RA form (seeform at next page)
Return AuthorizationPlease answer all the questions on the RA form with the greatest care andsign the form.
Date (problem date)
Type of device
Serial No. (of the device)
Installation date (of device)
mark Spare Part (devices should not be returned)
Mat.-No. of the spare part
Fault description (incl. error code and job codes)
mark request e.g. Repair and Replacement
fill in RA-number
give place and date (when filled in form) and sign RA form
Note: RA sent via electronic mail to Berlin is valid without asignature.
In case of thedevice is out ofwarranty
Installation date of spare part when part was installed within the last6 months.
All returned parts should be individually labeled with the corresponding RANo. and shipped together with the completed RA form (if not sent viaelectronic mail) to:
ROBERT RIELE GmbH & Co KG
Kurfrstenstrae 75 79D-13467 BerlinGermany
RIELE BERLIN Photometer 5010 V5+ 2.3 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
9/157
3.1 BLOCK DIAGRAM (up to SN 8001)
RIELE BERLIN Photometer 5010 V5+ 3.1.1 29.08.2007
7/27/2019 Pm-51 touch screen1.pdf
10/157
3.1 BLOCK DIAGRAM (from SN 8002)
RIELE BERLIN Photometer 5010 V5+ 3.1.2 29.08.2007
7/27/2019 Pm-51 touch screen1.pdf
11/157
3.2 SHORT INSTRUCTION GUIDE
Utility programs
Main menu:
Utility programs are necessary for the adjustmentand maintenance of the photometer.
Page 1 of utilit y programs:
Scrolling through all utility programs is possibleby [PAGE]. The current page is shown at the rightupper screen corner. By [EXIT] the programreturns to the main menu.
A utility program is selected by pressing therelating key.
Page 5 of utilit y programs:
The last key is representing the service tools.
M A I N M E N U
M E A S U R E W I T H P R O G R . M E T H O D S
M E A S U R E W I T H B A S I C M E T H O D S
M E A S U R E N E W / C H A N G E / C O P Y
U T I L I T I E S L A M P L F
3 7 . 0 3 C 2 2 . 1 1 . 0 7 1 1 : 3 9
U T I L I T I E S P A G E 1 / 5
D A R K A D J U S T M E N T P A G E
M U L T I - S T A N D A R D
P R I N T E R
P U M P / B U B B L E D E T E C T E X I T
3 7 . 0 1 C 2 2 . 1 1 . 0 7 1 3 : 0 5
U T I L I T I E S P A G E 5 / 5
L A N G U A G E P A G E
A D C C O U N T S O P T I C )
S E R V I C E T O O L S E X I T
3 7 . 0 1 C 2 2 . 1 1 . 0 7 1 3 : 0 5
RIELE BERLIN Photometer 5010 V5+ 3.2.1 12.03.2009
desplazamiento
7/27/2019 Pm-51 touch screen1.pdf
12/157
3.2 SHORT INSTRUCTION GUIDE
The service tools are reserved for trainedspecialists only and therefore protected by apassword.
The password is given as
[ 1 3 4 6 7 ]
Stop function by [ESC].
S E R V I C E T O O L S
P A S S W O R D
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
Utilities Description inchapterof OM
Dark adjustment 7.2.1
Multi-standard function 7.2.2
Printer ON / OFF 7.2.3
Pump / bubble detector 7.2.4
Menu serial COM 7.2.5Quality control 7.2.6
Settings printout 7.2.7
Data logging 7.2.8
Temperature ON / OFF 7.2.9
Temperature adjustment 7.2.10
Laboratory name 7.2.11
User name 7.2.12
Error list 7.2.13
Key signal ON / OFF 7.2.14
Touch panel adjustment 7.2.15
Date / Time 7.2.16
Language 7.2.17
ADC counts (Optic) 7.2.18
Send results 7.2.19
Service tools 7.2.20
RIELE BERLIN Photometer 5010 V5+ 3.2.2 12.03.2009
registro de datos
7/27/2019 Pm-51 touch screen1.pdf
13/157
3.3 SERVICE TOOLS
SELECTION OF SERVICE TOOLS
Within the service program level there are functions
influencing special basic parameters of the device.Because they are allowed to be changed divergentfrom the work configuration only in exceptions and onlyby trained staff, their unintentional selection isprotected by a password.
Change in service program with[SERVICE TOOLS].
Enter password:
[ 1 3 4 6 7 ]
Confirm with [E].
The service level will be opened until to the nextshutdown of the device.
U T I L I T I E S P A G E 5 / 5
L A N G U A G E P A G E
A D C C O U N T S O P T I C )
S E R V I C E T O O L S E X I T
S E R V I C E T O O L S
P A S S W O R D
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
RIELE BERLIN Photometer 5010 V5+ 3.3.1 11.03.2009
Dentro
personal calificado
7/27/2019 Pm-51 touch screen1.pdf
14/157
3.3 SERVICE TOOLS
Selection of a desired service tool is possible with[PAGE].
The current page is shown on the right upper corner ofthe display.
Abort of service tool selection and change back toprogram selection with [EXIT].
A service tool will be invoked by pushing thecorresponding button.
S E R V I C E T O O L S P A G E 1 / 6
S Y S T E M I N I T I A L I S A T I O N P A G E
M E M O R Y
D I S P L A Y T E S T
L A M P E X I T
Item no. Service tool Descriptionin chapter
1 SYSTEM INITIALISATION S1
2 MEMORY S2
3 DISPLAY TEST S3
4 LAMP S4
5 SERIAL NO. S5
6 PCB LAYOUT S6
7 SET WAVELENGTH S7
8 TEST FILTER WHEEL S8
9 TEMPERATURE ON/OFF S910 TEMPERATURE VALUE S10
11 CORRECTION VALUE S11
12 TEMPERATURE CALIBRATION S12
13 ADC CORRECTION S13
14 ADC COUNTS (OPTIC) S14
15 ADC COUNTS (DARK ADJ.) S15
16 BOOST WAVELENGTH S16
17 CALIBRATION 12V ADC S17
18 DISTRIBUTOR TEXT INIT S18
19 AUTOMATIC RESULTS S19
20 BCC OF METHOD / CURVE S20
21 CLEAR FACTORS + RB S21
22 CLEAR ERROR LIST S22
23 PRINTER TEST S23
24 STACK TEST S24
RIELE BERLIN Photometer 5010 V5+ 3.3.2 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
15/157
3.3 SERVICE TOOLS
DESCRIPTION OF SERVICE TOOLS
S1 System Initiali sationWith [START] the default parameters will be set.
Several settings (see table) are then typed over in thebattery powered RAM. They remain active, even afterswitching the device off/on, until they are specificallychanged.
[EXIT] leaves the function without any changes.
After a print-out of the program release the date andtime can be set (chapter OM 7.2.16).
The action ends with a long beep.
S Y S T E M I N I T I A L I S A T I O N
E X I T S T A R T
D A T E T I M E
m m . d d . y y h h : m m : s s
1 0 . 1 2 . 0 5 1 4 : 5 4 : 3 4
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
Beeper ON
Printer internal ON
Printer external OFF
Real time clock ON
Boost mode up to 390 nm
Language English
Analog-digital converter correction 150
Temperature control: Correction values 3000
EDP OFF
EDP protocol STX / ETX / BCC
Resulting factors of methods based on standard not affected
Error list not affected
Laboratory name not affected
User name not affected
Data logging not affected
Quality control not affected
Multi-standard functions not affected
Bubble detector ON
RIELE BERLIN Photometer 5010 V5+ 3.3.3 11.03.2009
ellos permanecen acyivos incluso
7/27/2019 Pm-51 touch screen1.pdf
16/157
3.3 SERVICE TOOLS
S2 Memory
The submenu shows different kinds of memory.S E R V I C E T O O L S S U B M E N U
D E L E T E F R E E M E T H O D S
M U L T I - S T D M E M O R Y
D E L . L O G G E R M E M O R Y
D E L . Q C M E M O R Y E X I T
S2.1 Delete Free Methods
Attention! With [START] all methods edited by the userare deleted without a call-back. In the case of doubtthese methods should be checked before: A print out ofall stored methods can be done for an optional laternew programming.
The action ends with a long beep.
[EXIT] leaves the function without any changes.
D E L E T E F R E E M E T H O D S
E X I T S T A R T
S2.2 Multi-Standard Memory
Attention! With [DEL] the memory assigned to nonlinearmethods is deleted without a call-back.
The contents of the memory can be exported and
imported by special remote commands (chapter 6.1 Communication protocol).
The action ends with a long beep.
[EXIT] leaves the function without any changes.
M U L T I - S T D M E M O R Y
E X I T D E L S E T
RIELE BERLIN Photometer 5010 V5+ 3.3.4 11.03.2009
registro
7/27/2019 Pm-51 touch screen1.pdf
17/157
3.3 SERVICE TOOLS
S2.3 Delete Logger Memory
Attention! With [START] the memory assigned to storedresults is deleted without a call-back.
The action ends with a long beep.
[EXIT] leaves the function without any changes.
D E L . L O G G E R M E M O R Y
E X I T S T A R T
S2.4 Delete QC MemoryAttention! With [START] the memory assigned to QCdata is deleted without a call-back.
The action ends with a long beep.
[EXIT] leaves the function without any changes.
D E L . Q C M E M O R Y
E X I T S T A R T
RIELE BERLIN Photometer 5010 V5+ 3.3.5 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
18/157
3.3 SERVICE TOOLS
S3 Display Test
The display test is invoked with [START].
For three seconds all pixels of the touch screen areilluminated.
The test terminates by itself.
[EXIT] leaves the function without any changes.
D I S P L A Y T E S T
E X I T S T A R T
S4 Lamp
The function shows the current lamp voltage. The valueis shown in the second line (e.g. 12.13 V).
With [MIN.] the lamp voltage can be temporarilyreduced. In case of the adjustment of the optical paththe light spot is better to be seen.
With [MAX.] the lamp voltage arises to its normal value.
[EXIT] leaves the function without any changes.
L A M P
1 2 . 1 3 V
E X I T M I N .
S5 Serial No.
The current serial number of the device can be readand set. The number (e.g. 7001) is shown on the right
upper corner of the window.
A new number can be entered with maximum fivedigits.
The input has to be confirmed with [E].
S E R I A L N O . 7 0 0 1
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
RIELE BERLIN Photometer 5010 V5+ 3.3.6 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
19/157
3.3 SERVICE TOOLS
S6 PCB Layout
The attribute of the PCB layout can be read and set.The attribute (e.g. c) is shown on the right upper cornerof the window.Software applications may be depend on this attribute.The attribute is labelled on the PCB RR_120 near plugST13.
The attribute can be changed with [SELECT].
The input has to be confirmed with [OK].
[EXIT] leaves the function without any changes.
P C B L A Y O U T c
E X I T O K S E L E C T
S7 Set Wavelength
The assignment of wavelength and filter position canbe done.
The current filter position is blinking.
With [SELECT] the filter position can be set.
With [NEW] the wavelength can be edited.
[EXIT] leaves the function without any changes.
In the first line the current filter position is shown.
The new value of the wavelength has to be entered.
The input has to be confirmed with [E].
S E T W A V E L E N G T H
1 : 3 4 0 6 : 6 2 3
2 : 4 0 5 7 : 9 9 9
3 : 4 9 2 8 : 9 9 9
4 : 5 4 6 9 : 9 9 9
5 : 5 7 8
E X I T N E W S E L E C T
N E W W A V E L E N G T H 1
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
RIELE BERLIN Photometer 5010 V5+ 3.3.7 11.03.2009
disposicion
cerca
7/27/2019 Pm-51 touch screen1.pdf
20/157
3.3 SERVICE TOOLS
S8 Test Filter Wheel
The rotation of the filter wheel can be controlled.
With [F0] up to [F9] the discrete positions of the filterwheel can be selected.
A specific value (e.g. 399MS) shows the rotation of thefilter wheel. The abbreviation MS is used for motorstep.
T E S T F I L T E R W H E E L
F 1 3 4 0 n m 3 9 9 M S
F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8
E S C F 9 F 0
S9 Temperature ON/OFF
The temperature regulation can be switched on orswitched off. The current setting is shown in the firstline as ON or OFF.
The display shows additionally the followinginformation:
a D in case of an inserted flow through cuvetteadaptor
an E in case of an inserted standard cuvetteadaptor
the required value of temperature regulation
the actual value of temperature regulation
the stability value
the current value of ADC
With [SELECT] the temperature level can be chosen.
With [OK] the temperature regulation is started.
[EXIT] leaves the function without any changes.
T E M P E R A T U R E 3 7 C O N D
O F F R E Q U I R E D 1 4 1 6 5
2 5 C A C T U A L 1 4 0 5 3
3 0 C S T A B I L I T Y 4
3 7 C C O N V E R T E R 9 4 7
E X I T O K S E L E C T
RIELE BERLIN Photometer 5010 V5+ 3.3.8 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
21/157
3.3 SERVICE TOOLS
S10 Temperature Value
The temperature values of the device can be read andset. A new value can be entered with maximum fivedigits.
With [SELECT] a specific temperature value can beset. The temperature values are marked as E25, E30,E37, D25, D30 and D37 corresponding to the type ofadaptor (E or D) and the temperature level (25 C,30 C and 37 C).
The cursor of the current temperature value is blinking.With [NEW] a temperature value can be edited.
Enter temperature value. Confirm with [E].
(chapter 4.40 - Thermal adjustment)
T E M P E R A T U R E V A L U E
E 2 5 8 3 2 2 D 2 5 8 2 9 9
E 3 0 9 5 5 2 D 3 0 9 4 8 5
E 3 7 1 1 2 7 3 D 3 7 1 1 1 6 5
E X I T N E W S E L E C T
N E W V A L U E
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
S11 Correction Value
The correction values of the device can be read andset. A new value can be entered with maximum fivedigits.
With [SELECT] a specific correction value can be set.The correction values are marked as E25, E30, E37,D25, D30 and D37 corresponding to the type ofadapter (E or D) and the temperature level (25 C,30 C and 37 C).
The cursor of the current correction value is blinking.
With [NEW] a correction value can be edited.
Enter correction value. Confirm with [E].
C O R R E C T I O N V A L U E
E 2 5 3 0 0 0 D 2 5 3 0 0 0
E 3 0 3 0 0 0 D 3 0 3 0 0 0
E 3 7 3 0 0 0 D 3 7 3 0 0 0
E X I T N E W S E L E C T
N E W V A L U E
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
RIELE BERLIN Photometer 5010 V5+ 3.3.9 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
22/157
3.3 SERVICE TOOLS
S12 Temperature Calibration
The correction of the temperature regulation is possibleby input of the temperature value which is measured inthe cuvette with an external measuring system(Thermistor and DMM).
This function is available only when the temperatureregulation is stabilized. In an unstable condition anerror message is shown on the display.
Enter the actual temperature value always four-digit in C e.g. [3] [7] [.] [2] [0] measured by an externalmeasuring system.
Confirm with [E].
The system calculates and stores the new setting forthe selected temperature level and starts working with itat once.
T E M P . C A L I B R A T I O N
T E M P . A C T U A L :
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
S13 ADC Correction
The current ADC correction of the device can be readand set.
The range of this value is 0 up to 255. The defaultvalue is 150.
The input has to be confirmed with [E].
A D C C O R R E C T I O N 1 5 0
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
S14 ADC Counts (Optic )
Indicated is the current value of the optical analogue-digital converter. The value is proportional to the light-
current depending on the selected amplification leveland the boost setting.
To a key actuation the system reacts possibly only afterthree seconds.
The functions V0 to V7 set the amplification level. Thefunction BO increases (boost ON) or reduces (boostOFF) the time slot. The functions F0 to F9 place thefilter wheel into the positions 0 to 9. The position 0corresponds to the filter wheel position at the dark leveladjustment.
Stop function by [ESC].
A D C C O U N T S O P T I C )
B O O S T O F F
V 0 F 1 3 4 0 n m 1 8 8 1 0
V 0 V 1 V 2 V 3 V 4 V 5 V 6 V 7 B 0 .
F 0 F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8 F 9
E S C
RIELE BERLIN Photometer 5010 V5+ 3.3.10 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
23/157
3.3 SERVICE TOOLS
S15 ADC Counts (Dark Adjus tment)
The ADC values of the dark adjustment can be read.
The display shows the following information:
a D in case of a preselected flow through cuvetteadaptor
an E in case of a preselected standard cuvetteadaptor
the ADC values assigned to the amplification levelsV0 up to V7
With [E/D] the display shows alternately the values incombination with the different adaptors E or D.
A D C C O U N T S D A R K A D J . )
E
0 : 1 8 4 2 5 1 : 1 8 4 2 8
2 : 1 8 4 3 3 3 : 1 8 4 4 2
4 : 1 8 4 5 7 5 : 1 8 4 8 3
6 : 1 8 5 3 4 7 : 1 8 6 3 0
E X I T E / D
S16 Boost Wavelength
To improve the low light gain at small wavelengths, themeasurement time is extended when operating below adefinite wavelength. This definite wavelength will bedisplayed on the right upper corner of the window.
The default value is 390 nm.
The input has to be confirmed with [E].
B O O S T W A V E L E N G 3 9 0 n m
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
RIELE BERLIN Photometer 5010 V5+ 3.3.11 11.03.2009
para mejorar
7/27/2019 Pm-51 touch screen1.pdf
24/157
3.3 SERVICE TOOLS
S17 Calibration 12V ADC
In the first line the actual value of the voltage 12 VDCisto be seen (e.g. 12.02V).
The internal measuring unit can be calibrated bypushing [START].
It is assumed that the adjustment of the voltage 12 VDCof halogen lamp is done properly (chapter 4.40 Voltage 12 VDCof halogen lamp).
The procedure is finished by [EXIT].
C A L . 1 2 V A D C 1 2 . 0 2 V
E X I T S T A R T
S18 Distributor Text Init
The distributor text shown after switching on thePhotometer 5010 V5+will be erased by invoking[START].
[EXIT] leaves the function without any changes.
A new distributor text can be set by software SerialMonitor 5010 (chapter 4.9).
D I S T R I B U T O R T E X T I N I T
E X I T S T A R T
RIELE BERLIN Photometer 5010 V5+ 3.3.12 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
25/157
3.3 SERVICE TOOLS
S19 Automatic Results
This operation works only wi th standard cuvetteadapter!
The function invokes by [START] a test based on thecalculation procedure 1C/F (OM chapter 5.4.1).
For each setting parameter a leading identificationnumber is shown. If the identification number isselected on the keyboard, the corresponding settingparameter becomes configurable.
After setting parameters like wavelength, factor, delaytime, unit, the procedure is started by [OK].
By setting delay time the intervals between themeasurements can be defined.
The procedure is working in an open loop andproduces results as long as the [EXIT] is pushed.
By [ZERO] an automatic zero adjustment can be done.To a key actuation the system reacts possibly only afterthree seconds.
A U T O M A T I C R E S U L T S
E X I T S T A R T
M E T H O D N O . 1 C / F
1 - W A V E L E N G T H 3 4 0 n m
2 - F A C T O R 1 . 0 0 0 E X I T
3 - T E M P E R A T U R E 3 7 C
4 - D E L A Y 0 s
5 - U N I T O K
6 -
7 -
8 -
1 2 3 4 5 6 7 8 P 1 / 3
D E L A Y 5 C / F
4 2
E X I T M O D E W A S H Z E R O R E S U L T
1.234
S20 BCC of Method/Curve
To check the correct storing of a user-defined methodor curve, the special test byte BCC concerning the datain the memory is used. With this service program theBCC of a method can be determined.
With [M] the BCC of a method can be tested.
With [C] the BCC of a curve can be tested.
B C C O F M E T H O D / C U R V E
E X I T M C
RIELE BERLIN Photometer 5010 V5+ 3.3.13 11.03.2009
bucle abierto
almacenamiento
7/27/2019 Pm-51 touch screen1.pdf
26/157
3.3 SERVICE TOOLS
S21 Clear factors + RB
The determined resulting factor of a standardmeasurement and the reagent blank (RB) is storedtogether with the corresponding method number. Afterrenewed selection of this method the last resultingfactor is offered as "OLD STD", the reagent blank (RB)is offered as OLD RB.
All resulting factors and reagent blanks (RB) will beerased by invoking [START].
[EXIT] leaves the function without any changes.
C L E A R F A C T O R S + R B
E X I T S T A R T
S22 Clear error list
The last 10 serious errors are shown or printed.
The complete error list will be erased by invoking[START].
[EXIT] leaves the function without any changes.
C L E A R E R R O R L I S T
E X I T S T A R T
S23 Printer test
A special printout for testing the printer ribbon or thehardware is produced.
The printout is started by invoking [START].
[EXIT] leaves the function without any changes.
P R I N T E R T E S T
E X I T S T A R T
RIELE BERLIN Photometer 5010 V5+ 3.3.14 11.03.2009
7/27/2019 Pm-51 touch screen1.pdf
27/157
3.3 SERVICE TOOLS
S24 Stack test
The address pointer of the stack can be controlled.
The attribute can be changed with [SELECT].
ON = only for development purposes
OFF = normal position
The setting has to be confirmed with [OK].
[EXIT] leaves the function without any changes.
S T A C K T E S T O F F
E X I T O K S E L E C T
RIELE BERLIN Photometer 5010 V5+ 3.3.15 11.03.2009
solo para fines de desarrollo
7/27/2019 Pm-51 touch screen1.pdf
28/157
3.4 PLUG IDENTIFICATION
RIELE BERLIN Photometer 5010 V5+ 3.4.1 22.11.2007
In the device all plug-in connections are marked with P1 ... P23 as described below:
PLUG FROM TO PLUG
Terminal AC / DC power supply motherboard pcb RR-120 P1
P2 motherboard pcb RR-120 sip control peristaltic pump
P3 motherboard pcb RR-120 fan
P4 motherboard pcb RR-120 halogen lamp
P5 motherboard pcb RR-120 light barrier filter wheel
P7 motherboard pcb RR-120 photodiode
P8 motherboard pcb RR-120 Peltier element
P9 motherboard pcb RR-120 Pcb RR-122 for display, printer, touch
P10 motherboard pcb RR-120 Pcb RR-122 for display, printer, touch
P11 motherboard pcb RR-120 Pcb RR-121 motor filter wheel
P13 motherboard pcb RR-120 Interface RS-232
P14 motherboard pcb RR-120 Pcb RR-121 motor peristaltic pump
P20 motherboard pcb RR-120 Pcb RR-126 cuvette adaptor
P21 motherboard pcb RR-120 Pcb RR-126 cuvette adaptor
Terminal AC / DC power supply Pcb printer controller CN2
7/27/2019 Pm-51 touch screen1.pdf
29/157
4.1 MAINS POWER CIRCUIT
Function DescriptionThe mains power circuit shows the design of the complete primary circuit. Components of the circuit are:
Mains connector
Mains fuses Fa and Fb REF 5010-018
Mains switch
AC/DC power supply REF 5010-016 (terminal block with 7-pins)
or REF 5010-086 (terminal block with 9-pins)
Line filter
Ground connection
Wires
LocationBackside of Photometer 5010 V5+, base plate
Replacement see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
see chapter 5.6 Power supply
InstallationFor installation proceed in reverse order.
FusesBetween the mains switch and the mains cable connector there is a cartridge with two fuses Fa and Fb withfollowing specifications:
dimensions [mm] : 5 * 20
standard: IEC 60127-2/V
time-current characteristic: time lag (T)
voltage rating: 250 V
rated current: 1.6 A
marking: T 1.6 A H
RIELE BERLIN Photometer 5010 V5+ 4.1.1 18.08.2010
7/27/2019 Pm-51 touch screen1.pdf
30/157
4.1 MAINS POWER CIRCUIT
power input90 V~ - 264 V~
gnye gnye
fuses Fa and Fb
groundconnection
wh gr
mains switch
br bk
line filter
br bk
PE N L
AC / DCpower supply
REF 5010-016(terminal blockwith 7-pins)
RIELE BERLIN Photometer 5010 V5+ 4.1.2 18.08.2010
7/27/2019 Pm-51 touch screen1.pdf
31/157
4.1 MAINS POWER CIRCUIT
power input90 V~ - 264 V~
gnye gnye
fuses Fa and Fb
groundconnection
wh gr
Mains switch
br bk
line filter
br bk
PE N L
AC / DCpower supply
REF 5010-086(terminal blockwith 9-pins)
RIELE BERLIN Photometer 5010 V5+ 4.1.3 18.08.2010
7/27/2019 Pm-51 touch screen1.pdf
32/157
4.2 MAINS CONNECTOR
Function DescriptionThe unit connects the power input with the AC/DC power supply.
LocationBackside of Photometer 5010 V5+, base plate
fan
fuses Fa and Fb
mains switch
power input
90 V~ - 264 V~
RIELE BERLIN Photometer 5010 V5+ 4.2.1 13.11.2006
7/27/2019 Pm-51 touch screen1.pdf
33/157
4.3 AC / DC POWER SUPPLY REF 5010-016/-086
Function Descript ion of REF 5010-016The AC / DC power supply is designed for producing following voltages:
V1: + 5 VDC/ 7 A
V2: + 15 VDC/ 3 A
V3: - 15 VDC/ 1 A
LocationBase plate of Photometer 5010 V5+
Replacement
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.6 Power supply
see chapter 4.4 Cable power supply / printer
InstallationFor installation proceed in reverse order.See also exploded view, chapter Dismantling
Terminal block
RIELE BERLIN Photometer 5010 V5+ 4.3.1 27.08.2010
7/27/2019 Pm-51 touch screen1.pdf
34/157
4.3 AC / DC POWER SUPPLY REF 5010-016/-086
P1
Cable AC / DC power supply
Capacitor10000 F/25 V
RIELE BERLIN Photometer 5010 V5+ 4.3.2 27.08.2010
7/27/2019 Pm-51 touch screen1.pdf
35/157
4.3 AC / DC POWER SUPPLY REF 5010-016/-086
Function Descript ion of REF 5010-086The AC / DC power supply is designed for producing following voltages:
V1: + 5 VDC/ 7 A
V2: + 15 VDC/ 3 A
V3: - 15 VDC/ 1 A
LocationBase plate of Photometer 5010 V5+
Replacement
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.6 Power supply
see chapter 4.4 Cable power supply / printer
InstallationFor installation proceed in reverse order.See also exploded view, chapter Dismantling
Terminal block
RIELE BERLIN Photometer 5010 V5+ 4.3.3 27.08.2010
7/27/2019 Pm-51 touch screen1.pdf
36/157
4.3 AC / DC POWER SUPPLY REF 5010-016/-086
CableCN2 printercontrol ler /power supply
Cable AC / DCpower supply
Capacitor10000 F/25 V
RIELE BERLIN Photometer 5010 V5+ 4.3.4 27.08.2010
7/27/2019 Pm-51 touch screen1.pdf
37/157
4.3 AC / DC POWER SUPPLY REF 5010-016/-086
Installation notes
Cable with line filter (brown, black) and ground connection cable (yellow/green) are shown above only
for the sake of completeness and are not included in delivery.
In case of servicing a device previous to SN 8002 be sure that the cable PCB motherboard P23 / PCB
printer controller CN2 has been replaced by power supply cable (REF 5010-077) as described in
chapter 4.4.
A 15R / 5W wirewound resistor must be soldered at ST23 between pins 1 and 3 as shown below:
Devices f romSN 8002 withou tpin header.
Devices up toSN 8002 withpin header.
After installing AC / DC power supply it is strongly recommended to update to the newest operating
system version (see Operating Instructions Photometer 5010, chapter 7.2.5.4).
RIELE BERLIN Photometer 5010 V5+ 4.3.5 27.08.2010
7/27/2019 Pm-51 touch screen1.pdf
38/157
4.4 CABLE POWER SUPPLY / PRINTER REF 5010-077
RIELE BERLIN Photometer 5010 V5+ 4.4.1 29.08.2007
Function DescriptionOn devices previous to SN 8002, after a certain timeof operation, may occur error 52 (see chapter 7) dueto the printer controller power supply plug connection(P23).
To prevent this, it is possible to replace P23 with thenew power supply cable described in this chapter.The printer controller power supply CN2 will beconnected then directly to the AC / DC power supply,eliminating the connection through the motherboard.
Beginning with SN 8002 all devices are equippedwith this cable at factory, thus no replacement isrequired.
LocationThe printer controller cable supply should be placedas shown here.
7/27/2019 Pm-51 touch screen1.pdf
39/157
4.4 CABLE POWER SUPPLY / PRINTER REF 5010-077
RIELE BERLIN Photometer 5010 V5+ 4.4.2 29.08.2007
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
In case of servicing a device previous to SN 8002 theconnection has to be removed:
Cable PCB motherboard P23 / PCB printer controllerCN2.
Remove the plastic cover of the terminal.
Unscrew the two lens head screws labeled with V1and GND.
7/27/2019 Pm-51 touch screen1.pdf
40/157
4.4 CABLE POWER SUPPLY / PRINTER REF 5010-077
RIELE BERLIN Photometer 5010 V5+ 4.4.3 29.08.2007
Insert the two screws through the solder lugs of thenew power supply cable.
Both cable terminals have to be screwed at the AC /DC power supply, with V1 (red wires) and GND (bluewires).
Set the plastic cover of the terminal.
The new power supply cable has to be placed underthe corner of the motherboards base plate. This willprevent the cable to interfere with the fan operationafter mounting / dismounting the device.
Plug the cable with CN2 at the printer controller.
7/27/2019 Pm-51 touch screen1.pdf
41/157
4.10 OPTICAL UNIT
Function DescriptionOptical unit
The optical unit consists of the following components:
Filter wheel with filters
Step motor filter wheel
Light barrier filter wheel
Step motor pump
Pump wheel
Connector bubble detector
Halogen lamp
Unit photodiode
Lenses
Unit Peltier element
Unit cuvette adaptor
LocationBase plate
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.10.1 11.05.2006
7/27/2019 Pm-51 touch screen1.pdf
42/157
4.11 FILTER WHEEL
RIELE BERLIN Photometer 5010 V5+ 4.11.1 09.04.2009
Function DescriptionThe filter wheel has nine filter positions called F1 to F9.Six filter positions are equipped with optical filters of standard wavelengths.The wavelengths are: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm and 623 nm
Location
Optical unit
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
see chapter 5.5 Optical unit: Filter wheel, motor filter wheel, motor peristaltic pump, light barrier
InstallationFor installation proceed in reverse order.
Filter wheel with nine filter positionscalled F1 to F9
Filter wheel equipped with interferenceline filters of standard wavelengths
Motor mounting plate
7/27/2019 Pm-51 touch screen1.pdf
43/157
4.11 FILTER WHEEL
RIELE BERLIN Photometer 5010 V5+ 4.11.2 09.04.2009
Installation of additional optical filters / Replacement of optical filtersThe filter wheel can be equipped with additional optical filters in the positions 7 to 9.In the standard equipment the not used positions are optically locked by metal slices.The respective metal slice is replaced with a filter at installation of an additional filter. In the memory of thePhotometer 5010 V5+ the value of the wavelength (e.g. 670 nm) for the corresponding position (e.g. 7)becomes changed by a service software routine.
Mechanical operations:LocationOptical unit
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate page 5.3.1
The filter wheel is turned by hand so that the corresponding hole is at the cut-out of the motor mounting plate
(see page 4.11.1).
With the tool kit filter exchange ( RIELEREF 5010-001 ) a filter exchange can beeasily carried out.
The tool kit filter exchange consists of the
tools A and B.
Tool B
Tool A
7/27/2019 Pm-51 touch screen1.pdf
44/157
4.11 FILTER WHEEL
RIELE BERLIN Photometer 5010 V5+ 4.11.3 09.04.2009
With the tool A and a pipe wrench themetal slice can be pushed from the holein the filter wheel.
The cheeks of the pipe wrench grasp atthe tool and at the filter wheel.
With a screwdriver the Seeger ring andthe metal slice can be taken out.
Left picture:Optical filter with marking340 nm
Right picture:Optical filter with markingarrow. The arrow showsthe direction of the light.
Two kinds of optical filterscan be used in the filterwheel:
diameter 12 mm
diameter 10 mm
In case of using filters with adiameter 10 mm an
additional spacer ring isneeded.
Spacer ring for Optical filter Complete unit:optical filter with 10 mm diameter with 10 mm diameter Optical filter with spacer ring
7/27/2019 Pm-51 touch screen1.pdf
45/157
4.11 FILTER WHEEL
RIELE BERLIN Photometer 5010 V5+ 4.11.4 09.04.2009
Put the filter into the corresponding holeof the filter wheel.
If the filter has only one reflectingsurface, this reflecting surface must bejudged to the light source.
If the filter is marked with an arrow itshows the direction of the light.
The Seeger ring can loosely be insertedinto the hole in the filter wheel.
With the tool B (left illustration) and apipe wrench the filter and the Seeger ringcan be fixed.
7/27/2019 Pm-51 touch screen1.pdf
46/157
4.11 FILTER WHEEL
RIELE BERLIN Photometer 5010 V5+ 4.11.5 09.04.2009
1. Run UTILITIES SERVICE TOOLS - SET WAVELENGTH(see chapter 3.3).
The assignment of wavelength and filterposition can be done.
The current filter position is blinking.
With [SELECT] the filter position can be set.
With [NEW] the wavelength can be edited.
In the first line the current filter position isshown.
The new value of the wavelength has to beentered.
The input has to be confirmed with [E].
2. Run UTILITIES SERVICE TOOLS - ADC COUNTS (OPTIC).
Indicated is the current value of the opticalanalogue-digital converter. The value isproportional to the light-current depending onthe selected amplification level and the boostsetting.
To a key actuation the system reactspossibly only after three seconds.The functions V0 to V7 set the amplificationlevel. The function BO increases (boost ON)or reduces (boost OFF) the time slot. Thefunctions F0 to F9 place the filter wheel into
the positions 0 to 9. The position 0 corre-sponds to the filter wheel position at the darklevel adjustment.
Select amplification level V0 and boost OFF.Stop function by [ESC].
3. The ADC counts of the new wavelength should be within a range of 15.000 and 600.000.
4. Finish the operation with [ESC].
S E T W A V E L E N G T H
1 : 3 4 0 6 : 6 2 3
2 : 4 0 5 7 : 9 9 9
3 : 4 9 2 8 : 9 9 9
4 : 5 4 6 9 : 9 9 9
5 : 5 7 8
E X I T N E W S E L E C T
N E W W A V E L E N G T H 1
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
S E R V I C E T O O L S P A G E 4 / 5
A D C C O U N T S O P T I C )
B O O S T O F F
V 0 F 1 3 4 0 n m 2 1 3 3 6
V 0 V 1 V 2 V 3 V 4 V 5 V 6 V 7 B O .
F 0 F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8 F 9
E S C
7/27/2019 Pm-51 touch screen1.pdf
47/157
4.12 HALOGEN LAMP REF 0554871001
RIELE BERLIN Photometer 5010 V5+ 4.12.1 12.05.2007
Function DescriptionThe halogen lamp is working with 12 VDCand anelectrical power of 20 W.
LocationOptical unit
The Photometer 5010 V5+ has to be dismantled(see chapter 5.1 Opening item 3 and item 5).
Unscrew the cylinder head screw.
Remove the lamp shield.
ReplacementUnscrew the knurled head screw.
Remove the holding plate.
7/27/2019 Pm-51 touch screen1.pdf
48/157
4.12 HALOGEN LAMP REF 0554871001
RIELE BERLIN Photometer 5010 V5+ 4.12.2 12.05.2007
The halogen lamp can be easily removed.
Attention:
The halogen lamp may be hot.
Use tweezers for removing.
The picture shows the lamp socket.
Attention: The lamp socket is adjusted in the factory.The two screws (A) are not to be solved without aspecial reason.
A
7/27/2019 Pm-51 touch screen1.pdf
49/157
4.12 HALOGEN LAMP REF 0554871001
RIELE BERLIN Photometer 5010 V5+ 4.12.3 12.05.2007
The plastic bag (B) of the halogen lamp can be usedfor inserting of the lamp. Do not touch the glass ofthe halogen lamp with fingers.
The picture shows the plastic bag, the halogen lamp,the holding plate and the lamp shield.
B
7/27/2019 Pm-51 touch screen1.pdf
50/157
4.12 HALOGEN LAMP REF 0554871001
RIELE BERLIN Photometer 5010 V5+ 4.12.4 12.05.2007
Installation
1. For installation proceed in reverse order.
2. In case of working with the flow-through cuvette adaptor the cell must be filled with distilled water. Incase of working with the standard cuvette adaptor the cuvette compartment must be empty.
3. Switch on the Photometer 5010 V5+.
4. Check the voltage 12 VDCof halogen lamp (see chapter 4.40.4).
5. Run UTILITIES ADC COUNTS (OPTIC).
Indicated is the current value of the opticalanalogue-digital converter. The value isproportional to the light-current depending onthe selected amplification level and the boostsetting.
To a key actuation the system reactspossibly only after three seconds.The functions V0 to V7 set the amplificationlevel. The function BO increases (boost ON)or reduces (boost OFF) the time slot. Thefunctions F0 to F9 place the filter wheel intothe positions 0 to 9. The position 0 corre-sponds to the filter wheel position at the darklevel adjustment.
Select amplification level V0 and boost OFF.
Stop function by [ESC].
6. Check one after another the used positions of the filter wheel.
7. The ADC counts should be within a range of 20.000 and 600.000(see chapter 4.40.6 Adjustment of amplifier stage).
8. Finish the operation with [ESC].
U T I L I T I E S P A G E 5 / 5
A D C C O U N T S O P T I C )
B O O S T O F F
V 0 F 1 3 4 0 n m 2 1 3 3 6
V 0 V 1 V 2 V 3 V 4 V 5 V 6 V 7 B O .
F 0 F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8 F 9
E S C
7/27/2019 Pm-51 touch screen1.pdf
51/157
4.13 UNIT PHOTODIODE REF 5010-056
Function DescriptionThe unit photodiode contains a silicon photodiode, alense, a lense holder and a cable with plug.
Lense
Plug P7
The photodiode is covered by a black label.
LocationThe photodiode is situated at the optical unit. Theplug P7 is connected with the socket ST7 on themother board RR-120.
Unit photodiode
Plug P7
RIELE BERLIN Photometer 5010 v5+ 4.13.1 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
52/157
4.13 UNIT PHOTODIODE REF 5010-056
ReplacementThe Photometer 5010 v5+ has to be dismantled:
see chapter 5.1 Opening item 1 and item 3
Unplug the unit photodiode from the mother board RR-120.Unscrew the two screws attaching the unit photodiode at the optical unit.
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 v5+ 4.13.2 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
53/157
4.14 UNIT CUVETTE ADAPTOR REF 5010-068
Function DescriptionA power Darlington transistor TIP 132 is used as aheating unit.
The unit contains a transistor TIP 132, the PCBRR-126 cuvette adaptor, cables and two plugs P20
and P21.
The plug P21 is connected with the socket ST21 onthe mother board RR-120.
The plug P20 is connected with the socket ST20 onthe mother board RR-120.
Lead with crimped connector belonging to plugP2 pin 4 (type of adaptor)
LocationOptical unit
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
Unscrew the two cylinder head screws of thePCB RR-126.
Unscrew the cylinder head screw of the transistor.
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.14.1 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
54/157
4.15 UNIT PELTIER ELEMENT REF 5010-062
Function DescriptionAn Peltier element is an electrical element based on the Peltier effect designated after Jean Peltier, whichproduces a temperature difference between two plates with current flow. The current pumps warmth from aside on the other one and produces a temperature difference between the plates. The element is used to cooldown the cuvette, cuvette adaptor and cuvette compartment.
LocationOptical unit
Unit Peltier element
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
see chapter 5.4 Optical compartment, peristaltic pump, heating unit
Two clamp screws are glued.
Heat conducting paste
Heat sink
Mica washer
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.15.1 27.04.2006
dos tornillos de fijacion
intensidad de las bombas de calor
7/27/2019 Pm-51 touch screen1.pdf
55/157
4.16 STEP MOTOR PUMP REF 5010-060
Function DescriptionThe peristaltic pump is used for sample aspirationinto the flow-through cell. A step motor drives therotor of the pump.
LocationOptical unit
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
see chapter 5.5 Optical unit: Filter wheel, motor filter wheel, motor peristaltic pump, light barrier
InstallationFor installation proceed in reverse order.
To fit the step motor the cover plate of the workingarea needs to be modified.
To remove the cover plate see chapter 5.4.2 ASSEMBLING / DISMANTLING.
Modified cover plate is part of the step motor sparepart.
RIELE BERLIN Photometer 5010 V5+ 4.16.1 23.07.2013
7/27/2019 Pm-51 touch screen1.pdf
56/157
4.18 STEP MOTOR FILTER WHEEL REF 5010-061
Function DescriptionThe step motor is used for driving the filter wheel.
LocationOptical unit
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
see chapter 5.5 Optical unit: Filter wheel, motor filter wheel, motor peristaltic pump, light barrier
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.18.1 10.05.2006
7/27/2019 Pm-51 touch screen1.pdf
57/157
4.19 LIGHT BARRIER FILTER WHEEL REF 5010-063
Function DescriptionThe light barrier is the position control of the filterwheel. The unit is delivered with a plug.
The plug P5 is connected with the socket ST5 on themother board RR-120.
Holder of the light barrier
Light barrier
LocationOptical unit.
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening item 1, item 2, item 3, item 5
see chapter 5.5 Optical unit: Filter wheel, motor filter wheel, motor peristaltic pump, light barrier
Adjusting procedure of the light barrier:
Loosen the cylinder head screw.
Hole on the mounting plate
RIELE BERLIN Photometer 5010 V5+ 4.19.1 11.05.2006
7/27/2019 Pm-51 touch screen1.pdf
58/157
4.19 LIGHT BARRIER FILTER WHEEL REF 5010-063
InstallationFor installation proceed in reverse order.
1. Switch on the Photometer 5010 V5+.
2. Run UTILITIES SERVICE TOOLS - TEST FILTER WHEEL.
The rotation of the filter wheel can be controlled.
With [F0] up to [F9] the discret positions of the filterwheel can be selected.
A specific value (e.g. 399MS) shows the rotation ofthe filter wheel. The abbreviation MS is used formotor step.
Move the holder up or down as long as each holesituated on the border line of the filter wheel is
centered to the hole situated on the mounting pF
T E S T F I L T E R W H E E L
F 1 3 4 n m 3 9 9 M S
F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8
E S C F 9 F
late.
ix the cylinder head screw.
3. Check one after another the used positions of the filter wheel.
4. Finish the operation with [ESC].
RIELE BERLIN Photometer 5010 V5+ 4.19.2 11.05.2006
7/27/2019 Pm-51 touch screen1.pdf
59/157
4.20 BUBBLE DETECTOR WITH PLUG REF 1706870001
Function DescriptionThe bubble detector controls the aspirated volume byan integrated light barrier. The detector gives ananalogue signal showing the status AIR or WATER.The signal is controlled by the microcontroller andproduces in case of a wrong handling or technical
defective an error message.
For further information see OPERATINGINSTRUCTIONS, chapter 9 Error message /Correction.
Location
Bubble detector
ReplacementOpen the lid. Disconnect the bubble detector. Withdraw the aspiration tubing from the metal tube. Dont bendthe apiration tubing. Replace the bubble detector.
InstallationFor installation proceed in reverse order.For further information see OPERATING INSTRUCTIONS, chapter 7.2.4 Pump / bubble detector.
RIELE BERLIN Photometer 5010 V5+ 4.20.1 11.05.2006
mal manejo
tapa
Para mas informacion
retirar
7/27/2019 Pm-51 touch screen1.pdf
60/157
4.21 FLOW-THROUGH CUVETTE ADAPTOR REF 1704796001
Function DescriptionThe flow-through cuvette adaptor consists of:
Plastic adaptor
Flow-through cuvette with a volume of 32 l
Heating unit
Plug Potentiometer for factory calibration
LocationOptical unit
ReplacementLoosen clamp screw, open tube connection, remove adaptor.
InstallationFor installation proceed in reverse order.
After installation of a new cuvette adaptor a dark level adjustment has to be done by
[UTILITIES] [DARK LEVEL ADJUSTMENT] [START].
RIELE BERLIN Photometer 5010 V5+ 4.21.1 28.04.2006
unidad de calefaccion
7/27/2019 Pm-51 touch screen1.pdf
61/157
4.22 STANDARD CUVETTE ADAPTOR REF 1704800001
Function DescriptionThe standard cuvette adaptor is used for disposable cuvettes or for working with a discrete flow-through cell.
The standard cuvette adaptor consists of:
Plastic adapter
Heating unit Plug
Potentiometer for factory calibration
LocationOptical unit
ReplacementLoosen clamp screw, remove adaptor.
InstallationFor installation proceed in reverse order.
After installation of a new cuvette adaptor a dark level adjustment has to be done by[UTILITIES] [DARK LEVEL ADJUSTMENT] [START].
Holding spring for disp. cuvetteREF 1707159001
RIELE BERLIN Photometer 5010 V5+ 4.22.1 28.04.2006
7/27/2019 Pm-51 touch screen1.pdf
62/157
4.23 CONNECTOR BUBBLE DETECTOR REF 5010-069
Function DescriptionThe connector bubble detector connects the bubbledetector with PCB RR-120 motherboard.
The plug P2 is connected with the socket ST2 on themother board RR-120.
LocationOptical unit
Connector bubble detector
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening item 1, item 2, item 3, item 5
Loosen milled nut and remove the connector bubble detector.Remove the green lead with crimped connector belonging to plug P2 pin 4 (type of adaptor).
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.23.1 11.05.2006
7/27/2019 Pm-51 touch screen1.pdf
63/157
4.24 LAMP HOLDER REF 5010-059
Function DescriptionThe lamp holder is used for fixing the halogen lamp.
Ondular washer
The plug P4 is connected with the socket ST4 on themother board RR-120.
LocationOptical unit
Lamp holder
Cylinder head screws with ondular washer
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening item 1, item 2, item 3, item 5
Remove halogen lamp. For further information see chapter 4.12 Halogen lamp. Unplug P4. Unscrew the twocylinder head screws.
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.24.1 12.05.2006
7/27/2019 Pm-51 touch screen1.pdf
64/157
4.26 TUBES
waste tubingREF 1704834001
outlet tube cuvetteREF 5010-066
aspiration tubingREF 5010-065
pump tube with jointsREF 5010-050
Joint inlet tube cuvetteREF 1707175001
Aspiration tubing REF 5010-065
Joint inlet tube cuvette REF 1707175001
Outlet tube cuvette REF 5010-066
Pump tube with joints REF 5010-050
Waste tubing REF 1704834001
RIELE BERLIN Photometer 5010 V5+ 4.26.1 18.05.2006
7/27/2019 Pm-51 touch screen1.pdf
65/157
4.30 TOP CASE REF 5010-064
Function DescriptionThe top case is one of the two main parts of thehousing. It is assembled with:
Microswitch (see chapter 4.33)
LocationTop case
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening item 1, item 2 and item 5
see chapter 5.2 Separating the top case from the bottom case
see chapter 4.41 Touchscreen
Note: When the plastic foil touchscreen (REF 5010-070) is removed from device it might not be reusable.
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.30.1 15.05.2006
7/27/2019 Pm-51 touch screen1.pdf
66/157
4.31 TOP CASE PRE-ASSEMBLED REF 5010-084WITH TOUCHSCREEN
RIELE BERLIN Photometer 5010 V5+ 4.31.1 15.04.2009
Function DescriptionThe top case is one of the two main parts of thehousing. It is assembled with:
Sipping lever Microswitch (see chapter 4.33)
Metal tube inlet
Lid
Touchscreen with PCB RR-122(see chapter 4.41)
Plastic foil touchscreen(see chapter 4.34)
LocationTop case
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening item 1, item 2 and item 5
see chapter 5.2 Separating the top case from the bottom case
see chapter 4.41 Touchscreen
see chapter 4.43 PCB RR-122 Display, printer, touch
see chapter 4.34 Plastic foil touchscreen
Note: When the plastic foil touchscreen (REF 5010-070) is removed from device it might not be reusable.
InstallationFor installation proceed in reverse order.
7/27/2019 Pm-51 touch screen1.pdf
67/157
4.32 METAL TUBE INLET REF 5010-055
Function DescriptionThe metal tube inlet is used to guide the aspiratetubing.
Metal tube inlet
LocationTop case
ReplacementRemove aspirate tubing from metal tube inlet.Remove metal tube inlet by turning and pulling at the same time.
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.32.1 15.05.2006
7/27/2019 Pm-51 touch screen1.pdf
68/157
4.33 MICROSWITCH WITH PLUG REF 5010-072
Function DescriptionThe microswitch is pressed by the sipping lever, itssignal starts the peristaltic pump.
LocationTop case
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating the top case from the bottom case
Disconnect plug J/ST3 of PCB RR-122 display, printer, touch.Remove microswitch by a screw driver and clean frame from remaining glue.
Installation
For installation proceed in reverse order.Use for example glue Delomet Typ 2301.
RIELE BERLIN Photometer 5010 V5+ 4.33.1 15.05.2006
7/27/2019 Pm-51 touch screen1.pdf
69/157
4.34 PLASTIC FOIL TOUCHSCREEN REF 5010-070
Function DescriptionThe plastic foil is made of polyester and selfadhesive.
LocationTop case
ReplacementRemove the plastic foil carefully. Avoid scratches on the top case.Note: When the plastic foil is removed from device it might not be reusable.
InstallationFor installation proceed in reverse order.
RIELE BERLIN Photometer 5010 V5+ 4.34.1 12.05.2006
7/27/2019 Pm-51 touch screen1.pdf
70/157
4.35 PRINTER MECHANISM 8NAD REF 5010-052
Function DescriptionThe printer is shown with an inserted ribbon.
Printer ribbon REF 1010735001 or REF 0745235001
LocationTop case
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
Remove top cover for printer, paper and ribbon.Disconnect 18-pin plug CN6 from printer controller 8NAD (REF 5010-053).Loosen 2 front screws in the hollows and remove the other 2 screws.Remove printer mechanism.
RIELE BERLIN Photometer 5010 V5+ 4.35.1 15.05.2006
7/27/2019 Pm-51 touch screen1.pdf
71/157
4.35 PRINTER MECHANISM 8NAD REF 5010-052
InstallationFor installation proceed in reverse order.
Marked lead
Solder printer connection cable to printer mechanism.Also take care for marked lead of pin 1.
RIELE BERLIN Photometer 5010 V5+ 4.35.2 15.05.2006
7/27/2019 Pm-51 touch screen1.pdf
72/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Function DescriptionThe mother board is the central electronic unit within Photometer 5010 V5+.Following components are equipped on the PCB:
multiprocessor system based on microcontrollers SAFC517 and 87C51/52
Flash memory
battery powered ram
battery powered real time clock
ADC for photocurrent
ADC for temperature control
interface stepper motor for peristaltic pump
interface infrared bubble detector
interface stepper motor for filter wheel
interface light barrier filter wheel
interface RS 232C
interface internal printer (Centronics)
interface touchscreen
interface Peltier element
interface heating unit
halogen lamp regulator
fuses F1, F2 and F3
LocationBase plate
ReplacementThe Photometer 5010 V5+ has to be dismantled:
see chapter 5.1 Opening
see chapter 5.2 Separating top case and bottom case
see chapter 5.3 Base plate
see chapter 5.7 Motherboard
InstallationFor installation proceed in reverse order.Perform the installation instruction of this chapter.
RIELE BERLIN Photometer 5010 V5+ 4.40.1 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
73/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Component diagram (layout c)
RIELE BERLIN Photometer 5010 V5+ 4.40.2 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
74/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Component diagram (layout d)
RIELE BERLIN Photometer 5010 V5+ 4.40.3 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
75/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Voltages1. Before adjustment switch the Photometer 5010 V5+off. Open the device. Open the photometer
regarding safety rules of chapter 1 of this service manual (chapter 5.1 Opening).
2. The device plug of the mains cable must be put into the mains cable socket at the back of the deviceand the mains plug of the mains cable must be put into a grounded mains socket.
3. Switch the device on.
4. Use DVM for measuring 12 VDC.
5. Connect the negative lead of DVM to the ground connector (soldering connector).
6. Connect the positive lead of DVM to the pin 3 of IC10.
7. Check the voltage + 5 VDC(tolerance 5%).
8. Connect the positive lead of DVM to the pin 4 of IC10.
9. Check the voltage - 12 VDC(tolerance 5%).
10. Connect the positive lead of DVM to the pin 14 of IC10.
11. Check the voltage + 12 VDC(tolerance 5%).
12. Connect the positive lead of DVM to the battery BT1.
13. Check the voltage + 3 VDC(tolerance 10%).
+ 3 VDC 10%
- 12 VDC 5%+ 5 VDC 5% + 12 VDC 5% Ground connector
RIELE BERLIN Photometer 5010 V5+ 4.40.4 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
76/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Voltage 12 VDCof halogen lamp1. Before adjustment switch the Photometer 5010 V5+off. Open the device. Open the photometer
regarding safety rules of chapter 1 of this service manual (chapter 5.1 Opening).
2. Switch the device on.
3. Use DVM for measuring 12 VDC.
4. Connect the negative lead of DVM to pin 2 of plug ST4.
5. Connect the positive lead of DVM to pin 1 of plug ST4.
6. Check the voltage +12.0 VDC(tolerance 50 mVDC).
7. If necessary adjust the voltage +12.0 VDC(tolerance 5 mVDC) with potentiometer P2.Turning clockwise increases voltage.Turning counter-clockwise decreases voltage.
Potentiometer P2
Fuses F1, F2, and F3
Name Protection Specification
F1 VCC (+ 5 VDC) T 3.15 A / 250 V
F2 + 15 VDC/ + 12 VDC T 6.3 A / 250 V
F3 Power supply printer + 5 VDC T 6.3 A / 250 V
RIELE BERLIN Photometer 5010 V5+ 4.40.5 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
77/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Amplifier for photocurrent and ADCThe photocurrent of the photodiode is converted with short circuit application into a voltage signal with the aidof the amplifier IC5 (AD 711). Potentiometer P3 sets the amplification and potentiometer P4 sets the offsetvoltage of the amplifier stage.
A voltage / frequency converter IC9 (AD654) is connected to the output of the amplifier.
The microcontroller can set the amplification of the amplifier stage by the multiplexer IC10 (ADG428). Theeight different settings of amplification are called V0 up to V7. The setting V0 is equivalent to the lowestamplification.
The microcontroller converts the frequency signal of the voltage / frequency converter to a digital result in arange between 0 and ~ 600000. This result is called value of the ADC.
The value of the ADC in the case of photocurrent = 0 (darkness) is called offset. The offset is
dependent on the setting of the amplifier stage,
positive and 0, set with potentiometer P4.
The following diagram shows the ADC-counter values versus the photocurrent.
ADC
photo current
~ 600.000
0
offset
RIELE BERLIN Photometer 5010 V5+ 4.40.6 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
78/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Adjustment of amplifier stage1. Before starting adjustments switch the Photometer 5010 V5+ off and unplug from the main socket.
Open the photometer regarding safety rules of chapter 1 of this service manual. The Photometer 5010V5+has to be dismantled (chapter 5.1 Opening item 4 and item 5).
2. Insert standard cuvette adaptor for disposable cuvettes into the device, but do not insert cuvette.
3. Switch the device on.
4. Be sure that the optical unit is adjusted faultless (chapter 4.10, 5.4 and 5.5) and check the supplyvoltages and the lamp voltage (chapter 4.40.4 and 4.40.5).
5. Run UTILITIES ADC COUNTS (OPTIC).
Indicated is the current value of the opticalanalogue-digital converter. The value isproportional to the light-current depending onthe selected amplification level and the boostsetting.
To a key actuation the system reactspossibly only after three seconds.The functions V0 to V7 set the amplificationlevel. The function BO increases (boost ON)or reduces (boost OFF) the time slot. Thefunctions F0 to F9 place the filter wheel intothe positions 0 to 9. The position 0 corre-sponds to the filter wheel position at the darklevel adjustment.
Stop function by [ESC].
U T I L I T I E S P A G E 5 / 5
A D C C O U N T S O P T I C )
B O O S T O F F
V 0 F 1 3 4 0 n m 2 1 3 3 6
V 0 V 1 V 2 V 3 V 4 V 5 V 6 V 7 B O .
F 0 F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8 F 9
E S C
6. Amplification:
Select amplification level V7 and boost OFF. Turn the filter wheel to the wavelength 340 nm by selectingF1. Turn the potentiometer P3 (on the mother board) clockwise to the maximum. The ADC should showa value 150000 counts.
7. Select amplification level V0 and boost OFF. Turn the filter wheel to that position by selecting F1 up toF9, so that the shown ADC value becomes a maximum. If necessary adjust the ADC to a value
600000 counts with the potentiometer P3. The amplification can be reduced by turning P3 counter-clockwise.
8. Offset:Switch the device off.
9. Remove the the photodiode plug and connect the plug 1 MOhm with the socket ST7.
plug 1 MOhm
10. Switch the device on.
11. Run UTILITIES ADC COUNTS (OPTIC). Turn the filter wheel to the dark position by selecting F0.
12. Note the ADC value e.g. 18015.
RIELE BERLIN Photometer 5010 V5+ 4.40.7 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
79/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
13. Select amplification level V7.
14. Adjust the ADC value with potentiometer P4 to the noted value e.g. 18015 with a tolerance of 50 counts.
15. Check all sensitivity levels V0 up to V7. The ADC values shall increase or decrease step by step. Forexample valid values for the increasing case could be:
V0: 18865, V1: 18869, V2: 18872, V3: 18887, V4: 18899, V5: 18901, V6: 18923, V7: 18933
The deviation between V0 and V7 shall be smaller than absolute 200 counts.
16. Switch the device off.
17. Remove the plug 1 MOhm and connect the photodiode plug with the socket ST7.
18. Do a dark level adjustment by applying utility U1with both cuvette adaptors.(OM chapter 7.2.1 - Dark level adjustment).
P3 P4MP2 Ground connector
RIELE BERLIN Photometer 5010 V5+ 4.40.8 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
80/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Temperature controlGeneral DescriptionThe temperature control takes care of the correct temperature inside the flow-through cuvette or inside thestandard cuvette adaptor. The temperature is selectable by the program at fixed levels, i.e. 25 C, 30 C or37 C.
The temperature control circuit consists of:
microcontroller system
temperature sensor
analogue-digital-converter (ADC)
heating unit
cooling unit
Microcontroller System
The temperature control circuit is a PID-controller, a software module in the microcontroller system. The PID-controller works in the background all the time. Dependent on the cuvette adaptor the PID-controller usesdifferent parameters.
The required value of temperature is ascertained by the program of the microprocessor system. Thetemperature is controlled according to the set temperature in the selected method.
The PID-controller gives information to the controlling program about the given temperature tolerances andtherefore about the readiness for measuring.
Temperature Sensor and ADCThe voltage signal of the temperature sensor LM335 is converted by the IC2 (LTC1286) to an equivalentfrequency. The microcontroller unit IC3 (87C52PLCC) converts the frequency to a digital value.
The voltage signal of the temperature sensor LM335 is typical 3.1 VDC at 37 C,3.03 VDC at 30 C and 2.98 VDCat 25 C.The ADC converts the input voltage to a digital value between 0 and 32767 as shown on following diagram:
ADC
input voltageof temperature sensor
32767
0
ca. 17000
ca. 13900
3,03 VDC2,98 VDC 3,1 VDC
ca. 15200
RIELE BERLIN Photometer 5010 V5+ 4.40.9 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
81/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
HeatingThe control variable of the controller has influence on the heating. A switched Darlington transistor in source ofpower-switching is used as heating element. The PID-controller produces a voltage at the emitter of theDarlington transistor by a pulse-width-modulation. This voltage controls the current through the transistor.Involved components are the microcontroller IC3 (87C52PLCC) and IC7A/IC7B (LM393).
CoolingFor cooling or counter cooling during heating the microcontroller system switches on a Peltier element. Thereare four power levels with following terms and meaning:
0 = Cooling out
1 = Cooling (low level)
2 = Cooling (middle level)
3 = Cooling (high level)
In case of temperature changing by method selection cooling level 0 is selected if desired temperature ishigher as selected one.In case of temperature changing by method selection cooling level 3 is selected if desired temperature is lower
as selected one.
If selected temperature is reached it is switched over to a cooling level depending on the temperature:
for 37 degrees Centigrade cooling level 1
for 30 degrees Centigrade cooling level 2
for 25 degrees Centigrade cooling level 3
The cooling element is a Peltier element. The Peltier element functions as a heat pump and transports heatout of the cuvette adaptor to the heat sink. The heat sink is provided with air by the fan of the device andtransports the heat outside to the area.The Peltier element operates with a voltage, switch able on four levels:
cooling level 0 0 VDC
cooling level 1 0.45 VDC
cooling level 2 0.9 VDC
cooling level 3 1.3 VDC
Involved components are the microcontroller unit IC3 (87C52PLCC), the OP-booster U1A (LM358), thetransistor T4 (BC547), the transistor T5 (TIP132 mounted at the base plate) and the Peltier element.
RIELE BERLIN Photometer 5010 V5+ 4.40.10 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
82/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Thermal adjustmentThe PID-controller algorithm works always as follows:
ACTUAL VALUE = REQUIRED VALUE
REQUIRED VALUE = TEMPERATURE VALUE + CORRECTION VALUE
Explanation:
ACTUAL VALUE Actual value of the ADC
REQUIRED VALUE Sum of TEMPERATURE VALUE + CORRECTION VALUE
TEMPERATURE VALUE Value assigned to a temperature level of 37 C, 30 C or 25 C. Thisvalue is automatically selected by the device according to the state ofoperating. The temperature value is board-characteristic and cuvetteadaptor specific. The values for the temperature levels of 37 C, 30 Cor 25 C are ascertained in factory and noted during the check of theboard. They are stored within the battery powered RAM.
Dependent on the used cuvette adaptor different service programs canshow and edit the values stored in battery powered RAM. The systemcan be adjusted to lower or higher temperature values. A change ofabout 25 counts means a temperature change of 0.1 C.
Service program S10 - Temperature valueis used for editing the value(chapter 3.3)
Service program S12 - Temperature calibrationcalibrates the system automatically.(chapter 3.3 / chapter 9.1)
The value is noted on a sticker. The sticker is shown on the left and puton top of the IC15 (SABC517).
The sticker shows the serial number of the board (e.g. 7000) and thetemperature values of 25 C, 30 C, 37 C of standard cuvette adaptor(S) and flow-through adaptor (FTH).
In addition the values are documented in the device card of Photometer5010 V5+and stored in a file at the manufacturer.
RR_120 7000
S FTH
25 C 8255 8230
30 C 9495 9430
37 C 11198 11113
CORRECTION VALUE Correction value in the range of 0 and 9999. The default value is 3000.After using the service program S1 - System initialization (chapter 3.3)the correction value is set to the default value 3000.The system can be adjusted to a lower or a higher temperature value. A
change of about 25 counts means a temperature change of 0.1 C. Service program S11 - Correction value
is used for editing the value(chapter 3.3)
Utility programs U10 - Temperature adjustmentadjusts the system automatically.(chapter OM 7.2.10)
RIELE BERLIN Photometer 5010 V5+ 4.40.11 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
83/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
Installation instructionIn case of changing the mother board following items have to be done after switching on the device:
1. Check the contrast of LCD(Service Manual chapter 4.42 Contrast of touchscreen).
2. Check the voltage 12 VDCof halogen lamp. The voltage was pre-adjusted at the factory(Service Manual chapter 4.40 Voltage 12 VDCof halogen lamp).
3. Check voltages + 12 VDC, - 12 VDC, + 5 VDCand + 3 VDC (battery power)(Service Manual chapter 4.40 Voltages).
4. Perform the adjustment of amplifier stage(Service Manual chapter 4.40 Adjustment of amplifier stage).
5. The display of Photometer 5010 V5+shows:
M A I N M E N U
M E A S U R E W I T H P R O G R . M E T H O D S
M E A S U R E W I T H B A S I C M E T H O D S
M E T H O D S N E W / C H A N G E / C O P Y
U T I L I T I E S L A M P L F
3 7 . 0 5 C 0 4 . 0 1 . 0 6 1 1 : 3 9
6. The serial number can be set by applying the service tool S5(Service Manual chapter 3.3 S5 - Serial no.).
7. The assignment of filter positions has to be done if additional filters has been built in
(e.g. 670 nm). Apply the service tool S7(Service Manual chapter 3.3 S7 Set wavelength).
8. Do a setting of the temperature values by performing the calibration procedure ofPhotometer 5010 V5+.(Service Manual chapter 9.1 - Subject: Cuvette temperature)(Service Manual chapter 4.40 - Thermal adjustment)
9. Do a setting of attribute of PCB layout by applying the service tool S6:Software applications may be dependent on this attribute(Service Manual chapter 3.3 S6 - PCB layout).
10. Delete allocated memory by applying the service tool S2:(Service Manual chapter 3.3 S2.1 Delete free methods)(Service Manual chapter 3.3 S2.2 Multi-standard memory)(Service Manual chapter 3.3 S2.3 Delete logger memory)(Service Manual chapter 3.3 S2.4 Delete QC memory)
11. Define the laboratory name by applying the utility U11:The name of the laboratory can be stored permanently(OM chapter 7.2.11 Laboratory name).
12. Define the user names by applying the utility U12:The names of maximum five users can be stored permanently(OM chapter 7.2.12 User name).
13. Do a dark level adjustment by applying utility U1in combination with flow-through system.(OM chapter 7.2.1 - Dark level adjustment).The utility is working specifically to the current adaptor. An adjustment of the standard cuvette adaptor
should be done later on.
14. Perform a media test by applying utility U4.4(OM chapter 7.2.4.2 Media test).
RIELE BERLIN Photometer 5010 V5+ 4.40.12 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
84/157
4.40 PCB RR-120 MOTHERBOARD REF 5010-120
15. Perform a pump calibration by applying utility U4.2(OM chapter 7.2.4.4 Pump calibration).
16. Switch off Photometer 5010 V5+. Change the cuvette adaptor. Switch on Photometer 5010 V5+.
17. Do a dark level adjustment by applying utility U1in combination with standard cuvette adapter.See OM chapter 7.2.1 - Dark level adjustment.
18. Control the settings by applying utility U7(OM chapter 7.2.7 - Printout settings).
******ACTUAL*SETS*******
DATE: 11.05.08
TIME: 11:49:06
PHOTOMETER 5010 # 7000
V5.12a 31.10.08 D /2.0
PCB LAYOUT d
ADC COUNTS (DARK ADJ.) E
0: 17970 1: 17970
2: 17971 3: 17971
4: 17971 5: 17971
6: 17972 7: 17972
ADC COUNTS (DARK ADJ.) D
0: 17970 1: 17971
2: 17971 3: 17972
4: 17972 5: 17976
6: 17980 7: 17980
TEMPERATURE E
25C 8250 3000
30C 9500 3000
37C 11250 3000
TEMPERATURE D
25C 8250 300030C 9500 3000
37C 11250 3000
FILTER
0: DDD 1: 340
2: 405 3: 492
4: 546 5: 578
6: 623 7: 999
8: 999 9: 999
PUMP
MS AIR 205
MS/50ul 61
B-DET. 4
BATTERY: OKVOLUME OPTIMIZED OFF
VDC LAMP 12.09V
ADC CORRECTION 150
STACK TEST 18
BOOST nm 390
EDP ON (STX-ETX-BCC)
LANGUAGE
1: ENGLISH
2: GERMAN
KEY SIGNAL ON
PROGR. METHODS 0
Yellow marked variables:
Those variables are specific to thedevice and created in the factory.In case of replacing the variablesget information by:
Serial number:(see item 6)
Layout of pcb motherboard:(see item 9)
Temperature values:(see item 8)
ADC correction:The default value is 150.
Boost nm:The default value is 390.
Red marked variables:ADC counts:(see item 13 and 17)
MS Air, MS/50ul and B-Det:(see item 14 and 15)
VDC Lamp:(see item 2 and chapter 3.3 S17Calibration 12 VADC)
RIELE BERLIN Photometer 5010 V5+ 4.40.13 17.03.2009
7/27/2019 Pm-51 touch screen1.pdf
85/157
4.40 PCB RR-120 MOTHERBOARD Layout c REF 5010-120
RIELE BERLIN Photometer 5010 V5+ 4.40