2015Product Catalogue
v 1.0
(ES) Equipements Scientifiques SA - Département Tests Energie Mesures - 127 rue de Buzenval BP 26 - 92380 GarchesTél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com
Who we areCINERGIA is the result of more than ten years of experience in the conception, design, manufacture and commissioning of customized power electronics solutions. Our core competencies are the development of power electronics converters, DSP-based digital control, industrial communica-tions and process automation. Our know-how is the result of continuous R+D in the �elds of electrical conversion, electromobility, renewable resources and smart grids.
Our productsCINERGIA supplies high-technology products based on an industrial hardware platform designed to work 24/7. By using di�erent DSP �rmware, developed by CINERGIA, our standard product portfolio includes DC, AC and DC&AC power sink/sources. RegenerativeThanks to its bidirectional topology all our products can regenerate the test energy back to the grid, reducing the energy consumption and optimizing the electrical installation.
Our valueWe provide cutting edge technology at cost-competitive prices thanks to the combination of a robust industrial platform with CINERGIA’s proprietary control system.
We can adapt our solutions to your speci�c needs and provide the engineering support to achieve the desired result thanks to the knowledge and experience of our engineering team.
CINERGIA develops and supplies AC and DC Regenerative Power Converters for Industrial, R&D and Academic applications
DC platform
AC platform
GridEmulator
AC programmable
sourceCustomer
equipmentElectronic load(regenerative)
Grid AC/AC DC DeviceUnder test AC/AC
GridEmulator
DC programmable
sourceCustomer
equipmentElectronic load(regenerative)
Grid AC/DC DC DeviceUnder test DC/AC
ELECTROMOBILITYElectromomobility needs expertise in energy storage, batteries, battery charging, traction conver-ters, power supplies, communications and its control. CINERGIA supplies power electronics equipments and platforms for test in production lines, prototype validation, burn-in and R&D.
SMART GRIDSIEC de�nes the smart grid as an electricity network that intelligently integrate the actions of all its users- generators, consumers and those that do both. It is related with electrical vehicles, energy storage systems (ESS), renewable generation (DER), consumers , grid.tied/islanded systems, communication and energy manage-ment. CINERGIA can provide complete test solutions for Smartgrid Labs and Original Equip-ment Manufacturers.
SOLUTIONS
ACADEMIC AND R&DCINERGIA is your partner in Academic and R&D applications by adapting our standard products to the speci�c coming from academic and research requirements.
TEST PLATFORMSCINERGIA o�ers power electronics equipments, marked in green colour in the diagrams, as well as complete turn key solutions, for your testing needs: at the laboratory, production line, prototype and pre-series validation, burn-in, quality and R&D.
GridEmulator
BatteryCharger
AC/DCGridBms
+Batteries
BatteryEmulator
TractionConverter
VscInverter
Motor Generator
TestBench
AC/AC
ElectronicLoad
GridEmulator
BatteryEmulator Batteries
WindTurbine
Emulator
Grid
ElectricalVehicle
ElectricalVehicle
ElectronicLoad
MicrogridManager
BatteryCharger
(B2C)
BatteryCharger
PvEmulator u. DER e.
AcLoad
AC/AC DC/AC
SELECTION GUIDE
DCPS EL-DC B2C BE
Constant Voltage
Constant Current
Constant Power
Constant Resistance
Battery Charge
Battery Emulation
PAGE
√
√
√
√
√
√
13
√
√
√
√
18
√
√
√
√
√
23
√
√
28
DC only
GE-AC&DC EL-AC&DC
√
√
√
33
√
√
√
√
37
AC&DC
GE-AC EL-AC
Voltage Source
Current Source
Harmonics
Disturbance generation
10 to 400Hz
360 to 800Hz
PAGE
√
√
√
√
1
√
√
√
7
√
√
√
†
AC only
GE-AC&DC EL-AC&DC
√
√
√
√
33
√
√
√
37
AC&DC
EL-AC800
† Preliminary product, contact us for further information
1
CINERGIA’s Grid Emulators are AC programmable power supplies specially designed to create AC grids and emulate its disturbances. Test pro�les can be automatically executed by creating a csv �le. Thanks to its bidirectional power hardware, the test energy can be regenerated to the electrical grid. This feature is needed to test Distributed Energy Resources (DER) or generators, where the Equipment Under Test can generate, consume or store energy.
FUNCTIONAL DESCRIPTION
It generates di�erent AC grid systems:-Three-phase system (3F+N) up to 480Vrms line-line-Single-phase system (F+N) up to 277Vrms line-neutral-Con�gurable frequency from 10 to 400Hz-Con�gurable phase angle-Con�gurable internal resistance
It creates grid disturbances:-Voltage harmonics up to 780Hz (except 10, 12 and 14th) -Flicker -Over and Undervoltage-Interruptions-Symetric and unsymetric voltage dips (types A, B, C, D).-Frequency �uctuations.
All disturbances can be con�gured in amplitude, duration and repetition time.
KEY FEATURES
7.5 kVA – 200 kVA
4 Quadrant Power Supply
Regenerative up to 100% rated power
AC Output: 0-480 Vrms, l-l 0-277 Vrms, l-n 0-230Arms 10-400Hz Independent con�guration per phase:- Voltage, frequency and phase angle- Harmonics up to 780Hz
Disturbance con�guration:Interruptions and voltage dips, Flicker, Over and Undervoltage, Frequency Fluctuation
AC&DC version available
GE-AC Grid Emulator
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
GEA
C_2
015_
v1
1
CONCEPTUAL SCHEMATIC
TECHNICAL DIAGRAM OF GE
AC Input is connected to the grid (neutral connection or isolation transformer is required)AC Output is connected to the Equipment Under Test (EUT) and can be used as:- Three-phase system- Three independent single-phase systems- One parallelized single-phase system (3 times rated current)
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output Inverter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. The Inverter generates three independent AC voltages controlling the amplitude, phase angle, frequency and harmonic content by using a digital resonant control algorithm.
www.cinergia.coop
AC
Grid Emulator
GRID
DC AC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
-U -U
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
3 P + N 3 P + NAC IN (GRID side) AC OUT (EUT side)
GE-AC Grid Emulator
GEA
C_2
015_
v1
2
FUNCTIONALITIES
www.cinergia.coop
Undervoltage Voltage interruptions Balanced voltage dips
Frequency �uctuation400Hz Aeronautics Grid
Voltage Harmonics
Transient time 2-5ms
Independent phase harmonics Unbalanced voltage dips
Flicker
Balanced voltage dips
GE-AC Grid Emulator
GEA
C_2
015_
v1
3
USER INTERFACE
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
OptionsGalvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
Windows 7 user interface for remote operation and data acquisition.
Con�guration of harmonics Con�guration of disturbances
SOFTWARE FEATURES
GE-AC Grid Emulator
GEA
C_2
015_
v1
4
www.cinergia.coop
RANGE AND SPECIFICATIONSMAGNITUDE VALUE
Power 7.5kVA-200kVAInput side (GRID side) AC Voltage Rated 3x400Vrms+Neutral+Earth Voltage range +15% / -20 % AC Current (at rated power) 10-290Arms Frequency 48-62Hz THDi (at rated power) <3% Power Factor Typical at rated power ≥0.99 Con�gurable by user 0-1 (capacitive/inductive) E�ciency (at rated power) >92% Overload 125% for 10 min / 150% for 60 sOutput side (EUT side) AC Voltage Phase-neutral 0-277Vrms (10-100Hz) 0-210Vrms (101-200Hz) 0-150Vrms (201-400Hz) AC Current Per phase 10-230Arms Frequency Fundamental voltage 10-400Hz Harmonic Per phase 1st - 15th at 50Hz (except 10,12 & 14) 1st - 13th at 60Hz (except 10&12) 1st at 400Hz Harmonic content Maximum % 1st - 9th: 100% 11th: 50%, 13th-15th:10% Modes of operation Constant Voltage Range Resolution Ripple 0-100% <±0.1% <1% Disturbance Generation Voltage dip and interruptions Over and Undervoltage Frequency variation Flicker Harmonic sequenceGeneral Measurements Input Voltage (Vrms) and Current (Irms) Input and Output Power Output Voltage and Current Temperatures User interface 3.2” Touchscreen Local Control port: 1 analog input, 3 analog outputs, 4 inputs, 3 relays Communication Port: Ethernet (Optionals: RS485, RS232, CAN) Communication Protocol: Modbus/TCP Humidity 10-90% (Absolute maximum, without condensation) Temperature 5-35 ºC (Absolute maximum) Cooling Forced air Protections Over Current, Over Voltage, Shortcircuit, OvertemperatureStandards CE Marking Safety EN-62040-1-2, EN-60950-1 EMC EMC: EN-62040-2All speci�cations are subject to change without notice.
GE-AC Grid Emulator
GEA
C_2
015_
v1
5
www.cinergia.coop
GE MODELS
GE7.5GE10GE15GE20GE30GE40GE50GE60GE80GE100GE120GE160GE200
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENTAC rms / phase AC rms / parallel
10A15A20A25A40A50A65A80A105A130A155A185A230A
30A45A60A75A120A150A195A240A315A390A465A555A690A
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
All speci�cations are subject to change without notice.
All speci�cations are subject to change without notice.
GE-AC Grid Emulator
GEA
C_2
015_
v1
6
AC Electronic Loads are power electronics devices that behave as an AC electrical load. Typical applications are test platforms for Smartgrids, Load Emulation, AC sources, UPS, AC generators, Inverters, and electrical devices (as fuses or circuit breakers).
While most electronic loads are dissipative, CINERGIA o�ers a com-prehensive range of DC, AC and AC&DC regenerative electronic loads producing signi�cant energy savings, optimizing the electrical and cooling installation and reducing test costs.
FUNCTIONAL DESCRIPTION
Operation modes:-Constant Current (CC)-Constant Power (CP)
Three AC channels:-Independent control of each channel, as three single-phase loads-Parallel control of the three channels, as one single-phase load with 3 times the current rating-Independent control of each channel, as a star three-phase load balanced or unbalanced
80% typical energy saving by regenerating to the grid:
KEY FEATURES
7.5 kVA – 200 kVA
4 Quadrant Power Supply
Regenerative up to 100% rated power
Full 4Q: pure capacitive and inductive current
Independent con�guration per channel:- Fundamental current up to 400Hz- Harmonics up to 780Hz- Phase angle
Working (EUT) voltage: 0-480 Vrms, l-l (with neutral)0-277 Vrms, l-n 10-400Hz
AC&DC version available
EL-AC Electronic Load
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
-Constant Impedance (CI)-Automatic test from Excel �le
GRIDEquipment Under Test
(EUT) 92 kW
8 kW100 kW11 1 .1kW
η=90%11.1 kW
19.1kW AC
Electronic Load
DC AC
DC η = 92%
92 kW
ELAC
_201
5_v1
7
CONCEPTUAL SCHEMATIC
TECHNICAL DIAGRAM OF EL-AC
AC Input is connected to the grid (neutral connection or isolation transformer is required)AC Output is connected to the Equipment Under Test (EUT) and can be used as:- Three-phase star load (balanced or unbalanced)- Three independent single-phase loads- One single-phase load (3 times rated current)
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output Inverter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. The Inverter generates three independent AC currents controlling the amplitude, phase angle and harmonic content by using a digital resonant control algorithm.
www.cinergia.coop
AC
Electronic Load
GRID
DC AC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
-
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
AC IN (GRID side)3 P + N
AC OUT (EUT side)3 P + N
-
EL-AC Electronic Load
ELAC
_201
5_v1
8
FUNCTIONALITIES
www.cinergia.coop
Resistive load Pure capacitive load Pure inductive load
Phase delay + harmonicsPhase delay
Single-phase recti�er
In previous oscilloscope captures: AC source under test (yellow), EL currents phase A, B and C (green, magenta, blue)
Three-phase recti�erHarmonics
Independent phase control
EL-AC Electronic Load
ELAC
_201
5_v1
9
USER INTERFACE
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
OptionsGalvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
Windows 7 user interface for remote operation and data acquisition.
Con�guration of harmonics Con�guration of disturbances
SOFTWARE FEATURES
EL-AC Electronic Load
ELAC
_201
5_v1
10
www.cinergia.coop
RANGE AND SPECIFICATIONSMAGNITUDE VALUE
Power 7.5kVA-200kVAInput side (GRID side) AC Voltage Rated 3x400Vrms+Neutral+Earth Voltage range +15% / -20 % AC Current (at rated power) 10-290Arms Frequency 48-62Hz THDi (at rated power) <3% Power Factor Typical at rated power ≥0.99 Con�gurable by user 0-1 (capacitive/inductive) E�ciency (at rated power) >92% Overload 125% for 10 min / 150% for 60 sOutput side (EUT side) AC Voltage Rated maximum, ch-neutral 0-277Vrms (10-100Hz) 0-210Vrms (101-200Hz) 0-150Vrms (201-400Hz) AC Current 3ch mode, per channel 10-230Arms 1ch mode 30-690Arms Frequency Fundamental current 10-400Hz Power factor Con�gurable by user 0-1 (capacitive/inductive) Harmonic Con�gurable by user 1st - 15th at 50Hz (except 10,12 & 14) 1st - 13th at 60Hz (except 10&12) 1st at 400Hz Harmonic content Maximum % 1st - 9th: 100% 11th: 50%, 13th-15th:10% Crest factor <3Modes of operation Range Resolution Ripple Constant Current 0-±100% <±0.1%† <1% Constant Power 0-±100% <±0.1% <1% Constant Impedance (R,L,C) min.-100% <±0.1% <1%General Measurements Input Voltage (Vrms) and Current (Irms) Input and Output Power Output Voltage (Vrms) and Current (Irms) Temperatures User interface 3.2” Touchscreen Local Control port: 1 analog input, 3 analog outputs, 4 inputs, 3 relays Communication Port: Ethernet (Optionals: RS485, RS232, CAN) Communication Protocol: Modbus/TCP Humidity 10-90% (Absolute maximum, without condensation) Temperature 5-35 ºC (Absolute maximum) Cooling Forced air Protections Over Current, Over Voltage, Shortcircuit, OvertemperatureStandards CE Marking Safety EN-62040-1-2, EN-60950-1 EMC EMC: EN-62040-2† FS of current measurements is 3 times the rated curent (crest factor)All speci�cations are subject to change without notice.
EL-AC Electronic Load
ELAC
_201
5_v1
11
www.cinergia.coop
EL-AC MODELS
EL7.5-ACEL10-ACEL15-ACEL20-ACEL30-ACEL40-ACEL50-ACEL60-ACEL80-ACEL100-ACEL120-ACEL160-ACEL200-AC
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENTAC rms / phase AC rms 3ch mode 1ch mode10A15A20A25A40A50A65A80A105A130A155A185A230A
30A45A60A75A120A150A195A240A315A390A465A555A690A
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
All speci�cations are subject to change without notice.
All speci�cations are subject to change without notice.
EL-AC Electronic Load
ELAC
_201
5_v1
12
CINERGIA’s Programmable DC Power Supplies are power electronics equipment designed to generate a DC constant voltage, current or power. Thanks to its bidirectional hardware the DCPS can source or sink the test energy while the grid-tied converter will actively regenerate the energy back to the electrical grid consuming a sinusoidal current with unity power factor and low harmonic distortion.
FUNCTIONAL DESCRIPTION
Operation modes:-Constant Voltage (CV)-Constant Current (CC)-Constant Power (CP)-Automatic test from Excel �le
It provides three DC channels:-The three channels can be controlled independently, allowing di�erent voltage, current or power setpoints-The three channels can be controlled in parallel, sharing the same setpoint and providing 3 times the current-In Unipolar 2Q applications, the loads will be connected between one channels and the common negative. -In Bipolar 4Q applications, the loads will be connected between two channels
KEY FEATURES
6.75 – 160 kW
2/4 Quadrant Power Supply
Regenerative up to 100% rated power
1 channel Output: 0 to 750V, 0 to ±690A
3 channels Output:0 to 750V, 0 to ±230A/ch
Bipolar Output-750 to +750V, 0 to ±230A
CV, CC, CP modes
DCPS Programmable DC Power Supply
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
DC
PS_2
015_
v1
13
CONCEPTUAL SCHEMATIC
DCPS Programmable DC Power Supply
TECHNICAL DIAGRAM
AC Input is connected to the grid (neutral connection is required). Galvanic isolation is recommended.AC Output is connected to the Equipment Under Test (EUT) and can be used as:- Three independent 2Q channels- One 2Q channel (3 times rated current)- One 4Q bipolar channel (connecting the load between two channels)
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output DCDC converter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. The DCDC converter generates three independent DC voltages controlling the voltage, current or power.
www.cinergia.coop
AC
DC Power Supply
GRID
DC DC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
-U -U
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
3 P + N 3 CH-COM
CH1CH2CH3Com_neg
AC IN (GRID side) DC OUT (EUT side)
DC
PS_2
015_
v1
14
USER INTERFACE
DCPS Programmable DC Power Supply
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
OptionsGalvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
Windows 7 user interface for remote operation and data acquisition.
SOFTWARE FEATURES
0.0
100.0
200.0
300.0
400.0
500.0
600.0
700.0
800.0
0 1ch mode3ch mode-3ch mode-1ch mode
Minimumvoltage atrated power
Out
put v
olta
ge (V
)
Output current (A)
OPERATION AREA: 1/3 CHANNELS OPERATION AREA: BIPOLAR
-800.0
-600.0
-400.0
-200.0
0.0
200.0
400.0
600.0
800.0
-1ch mode 0 +1ch mode
Out
put v
olta
ge (V
)
Output current (A)
DC
PS_2
015_
v1
15
DCPS Programmable DC Power Supply
www.cinergia.coop
RANGE AND SPECIFICATIONSMAGNITUDE VALUE
Power 7.5kVA-200kVAInput side (GRID side) AC Voltage Rated 3x400Vrms+Neutral+Earth Voltage range +15% / -20 % AC Current (at rated power) 10-290Arms Frequency 48-62Hz THDi (at rated power) <3% Power Factor Typical at rated power ≥0.99 Con�gurable by user 0-1 (capacitive/inductive) E�ciency (at rated power) >92% Overload 125% for 10 min / 150% for 60 s
Output side (EUT side) DC Voltage Channel-Com_neg 0-750V Channel-Channel -750 to 750V Minimum voltage at rated power† 220V DC Current 1 channel output 0 to ±690A 3 channels output 0 to ±230A/ch Bipolar output 0 to ±230A Modes of operation Range Resolution Ripple Constant Voltage 0-100%‡ <±0.1% <1% Constant Current 0-±100% <±0.1% <1% Constant Power 0-±100% <±0.1% <1%Response time Rated resistance load 1-5ms (10-90%) General Measurements Input Voltage (Vrms) and Current (Irms) Input and Output Power Output Voltage and Current Temperatures User interface 3.2” Touchscreen Local Control port: 1 analog input, 3 analog outputs, 4 inputs, 3 relays Communication Port: Ethernet (Optionals: RS485, RS232, CAN) Communication Protocol: Modbus/TCP Humidity 10-90% (Absolute maximum, without condensation) Temperature 5-35 ºC (Absolute maximum) Cooling Forced air Protections Over Current, Over Voltage, Shortcircuit, OvertemperatureStandards CE Marking Safety EN-62040-1-2, EN-60950-1 EMC EMC: EN-62040-2
† Below minimum voltage the power is derated due to the current limitation. See operation area for further detail‡ 0-±100% in bipolar mode (galvanic isolation requested)All speci�cations are subject to change without notice.
DC
PS_2
015_
v1
16
DCPS Programmable DC Power Supply
www.cinergia.coop
MODELS
DCPS7.5DCPS10DCPS15DCPS20DCPS30DCPS40DCPS50DCPS60DCPS80DCPS100DCPS120DCPS160DCPS200
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENT3channels 1channel Bipolar0-750v 0-750v -750 to 750V ±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
±30A±45A±60A±75A±120A±150A±195A±240A±315A±390A±390A±465A±555A
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
All speci�cations are subject to change without notice.
All speci�cations are subject to change without notice.
DC
PS_2
015_
v1
17
DC Electronic Loads are devices that behave as a load and are designed to test electric and electronic DC sources. While most electronic loads are dissipative, CINERGIA o�ers a com-prehensive range of DC, AC and AC&DC regenerative electronic loads producing signi�cant energy savings, optimizing the electrical and cooling installation and reducing test costs.
FUNCTIONAL DESCRIPTION
Operation modes:-Constant Current (CC)-Constant Power (CP)-Constant Resistance (CR)
Three DC channels:-The three channels can be controlled independently, allowing di�erent setpoints on each channel-The three channels can be controlled in parallel, sharing the same setpoint and providing 3 times the current-In Unipolar 2Q applications, the sources will be connected between one channel and the common negative. -In Bipolar 4Q applications, the source will be connected between two channels
80% typical energy saving by regenerating to the grid:
KEY FEATURES
6.75 – 160 kW
2/4 Quadrant Power Supply
Regenerative up to 100% rated power
1 channel Output: 0 to 750V, 0 to ±690A
3 channels Output:0 to 750V, 0 to ±230A/ch
Bipolar Output-750 to +750V, 0 to ±230A
CC, CP, CR, CV modes
AC&DC version available
EL-DC DC Electronic Load
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
-Constant Voltage (CV)-Automatic test from Excel �le
GRIDEquipment Under Test
(EUT) 92 kW
8 kW100 kW11 1 .1kW
η=90%11.1 kW
19.1kW DC
Electronic Load
DC AC
DC η = 92%
92 kW
ELD
C_2
015_
v1
18
CONCEPTUAL SCHEMATIC
TECHNICAL DIAGRAM
AC Input is connected to the grid (neutral connection is required). Galvanic isolation is recommended.AC Output is connected to the Equipment Under Test (EUT) and can be used as:- Three independent 2Q channels- One 2Q channel (3 times rated current)- One 4Q bipolar channel (connecting the load between two channels)
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output DCDC converter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. The DCDC converter generates three independent DC voltages controlling the voltage, current or power.
www.cinergia.coop
AC
DC Electronic Load
GRID
DC DC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
-U -U
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
3 P + N 3 CH-COM
CH1CH2CH3Com_neg
AC IN (GRID side) DC OUT (EUT side)
EL-DC DC Electronic Load
ELD
C_2
015_
v1
19
USER INTERFACE
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
OptionsGalvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
Windows 7 user interface for remote operation and data acquisition.
SOFTWARE FEATURES
0.0
100.0
200.0
300.0
400.0
500.0
600.0
700.0
800.0
0 1ch mode3ch mode-3ch mode-1ch mode
Minimumvoltage atrated power
Out
put v
olta
ge (V
)
Output current (A)
OPERATION AREA: 1/3 CHANNELS OPERATION AREA: BIPOLAR
-800.0
-600.0
-400.0
-200.0
0.0
200.0
400.0
600.0
800.0
-1ch mode 0 +1ch mode
Out
put v
olta
ge (V
)
Output current (A)
EL-DC DC Electronic Load
ELD
C_2
015_
v1
20
www.cinergia.coop
RANGE AND SPECIFICATIONSMAGNITUDE VALUE
Power 7.5kVA-200kVAInput side (GRID side) AC Voltage Rated 3x400Vrms+Neutral+Earth Voltage range +15% / -20 % AC Current (at rated power) 10-290Arms Frequency 48-62Hz THDi (at rated power) <3% Power Factor Typical at rated power ≥0.99 Con�gurable by user 0-1 (capacitive/inductive) E�ciency (at rated power) >92% Overload 125% for 10 min / 150% for 60 s
Output side (EUT side) DC Voltage Channel-Com_neg 0-750V Channel-Channel -750 to 750V Minimum voltage at rated power† 220V DC Current 1 channel output 0 to ±690A 3 channels output 0 to ±230A/ch Bipolar output 0 to ±230A Modes of operation Range Resolution Ripple Constant Voltage 0-100%‡ <±0.1% <1% Constant Current 0-±100% <±0.1% <1% Constant Power 0-±100% <±0.1% <1% Constant Resistance Min.-100% <±0.1% <1%Response time Rated resistance load 1-5ms (10-90%) General Measurements Input Voltage (Vrms) and Current (Irms) Input and Output Power Output Voltage and Current Temperatures User interface 3.2” Touchscreen Local Control port: 1 analog input, 3 analog outputs, 4 inputs, 3 relays Communication Port: Ethernet (Optionals: RS485, RS232, CAN) Communication Protocol: Modbus/TCP Humidity 10-90% (Absolute maximum, without condensation) Temperature 5-35 ºC (Absolute maximum) Cooling Forced air Protections Over Current, Over Voltage, Shortcircuit, OvertemperatureStandards CE Marking Safety EN-62040-1-2, EN-60950-1 EMC EMC: EN-62040-2
† Below minimum voltage the power is derated due to the current limitation. See operation area for further detail‡ 0-±100% in bipolar mode (galvanic isolation requested)All speci�cations are subject to change without notice.
EL-DC DC Electronic Load
ELD
C_2
015_
v1
21
www.cinergia.coop
MODELS
EL7.5-DCEL10-DCEL15-DCEL20-DCEL30-DCEL40-DCEL50-DCEL60-DCEL80-DCEL100-DCEL120-DCEL160-DCEL200-DC
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENT3channels 1channel Bipolar0-750v 0-750v -750 to 750V
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
EL-DC DC Electronic Load
All speci�cations are subject to change without notice.
All speci�cations are subject to change without notice.
±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
±30A±45A±60A±75A±120A±150A±195A±240A±315A±390A±390A±465A±555A
±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
ELD
C_2
015_
v1
22
CINERGIA’s Bidirectional Battery Chargers are Programmable DC Power Supplies designed speci�cally to charge and discharge batteries or other storage systems. Based on the DCPS hardware and �rmware, the B2C add a IU Battery Charge mode to the Volta-ge, Current, Power and Resistance modes. The equipment is regenerative so the energy discharged from the battery will be injected back to the electrical grid.
FUNCTIONAL DESCRIPTION
Operation modes:-Battery Charge (BC) -Constant Voltage (CV)-Constant Current (CC) -Constant Power (CP)-Constant Resistance (CR) -Automatic test from Excel �le
It provides three DC channels:-The three channels can be controlled independently, allowing di�erent charge parameters, voltage, current or power setpoints-The three channels can be controlled in parallel, sharing the same setpoint and providing 3 times the current
Battery charge is controlled with an IU algorithm. Equaliza-tion, Boost and Float charge are possible. Battery discharge can be done at constant current, power, resistance or by programming a sequence in an Excel �le.
KEY FEATURES
6.75 – 160 kW
2 Quadrant Power Supply
Regenerative up to 100% rated power
1 channel Output: 0 to 750V, 0 to ±690A
3 channels Output:0 to 750V, 0 to ±230A/ch
BC, CV, CC, CP, CR modes
Equalization, Boost and Float charge
Constant current or power discharge
B2C Bidirectional Battery Charger
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
B2C
_201
5_v2
23
CONCEPTUAL SCHEMATIC
TECHNICAL DIAGRAM
AC Input is connected to the grid (neutral connection is required). Galvanic isolation is recommended.AC Output is connected to the Equipment Under Test (EUT) and can be used as:- Three independent 2Q channels- One 2Q channel (3 times rated current)
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output DCDC converter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. The DCDC converter generates three independent DC voltages controlling the voltage, current or power.
www.cinergia.coop
AC
B2C
GRID
DC DC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
-U -U
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
3 P + N 3 CH-COM
CH1CH2CH3Com_neg
AC IN (GRID side) DC OUT (EUT side)
B2C Bidirectional Battery Charger
B2C
_201
5_v2
24
USER INTERFACE
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
OptionsGalvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
Windows 7 user interface for remote operation and data acquisition.
SOFTWARE FEATURES
0.0
100.0
200.0
300.0
400.0
500.0
600.0
700.0
800.0
0 1ch mode3ch mode-3ch mode-1ch mode
Minimumvoltage atrated power
Out
put v
olta
ge (V
)
Output current (A)
OPERATION AREA: 1/3 CHANNELS BATTERY CHARGE MODE
B2C Bidirectional Battery Charger
B2C
_201
5_v2
25
www.cinergia.coop
RANGE AND SPECIFICATIONSMAGNITUDE VALUE
Power 7.5kVA-200kVAInput side (GRID side) AC Voltage Rated 3x400Vrms+Neutral+Earth Voltage range +15% / -20 % AC Current (at rated power) 10-290Arms Frequency 48-62Hz THDi (at rated power) <3% Power Factor Typical at rated power ≥0.99 Con�gurable by user 0-1 (capacitive/inductive) E�ciency (at rated power) >92% Overload 125% for 10 min / 150% for 60 s
Output side (EUT side) DC Voltage Channel-Com_neg 0-750V Channel-Channel -750 to 750V Minimum voltage at rated power† 220V DC Current 1 channel output 0 to ±690A 3 channels output 0 to ±230A/ch Bipolar output 0 to ±230A Modes of operation Range Resolution Ripple Constant Voltage 0-100% <±0.1% <1% Constant Current 0-±100% <±0.1% <1% Constant Power 0-±100% <±0.1% <1% Constant Resistance Min.-100% <±0.1% <1% Battery Charge 0-±100% <±0.1% <1%Response time Rated resistance load 1-5ms (10-90%) General Measurements Input Voltage (Vrms) and Current (Irms) Input and Output Power Output Voltage and Current Temperatures User interface 3.2” Touchscreen Local Control port: 1 analog input, 3 analog outputs, 4 inputs, 3 relays Communication Port: Ethernet (Optionals: RS485, RS232, CAN) Communication Protocol: Modbus/TCP Humidity 10-90% (Absolute maximum, without condensation) Temperature 5-35 ºC (Absolute maximum) Cooling Forced air Protections Over Current, Over Voltage, Shortcircuit, OvertemperatureStandards CE Marking Safety EN-62040-1-2, EN-60950-1 EMC EMC: EN-62040-2
† Below minimum voltage the power is derated due to the current limitation. See operation area for further detailAll speci�cations are subject to change without notice.
B2C Bidirectional Battery Charger
B2C
_201
5_v2
26
www.cinergia.coop
MODELS
B2C7.5B2C10B2C15B2C20B2C30B2C40B2C50B2C60B2C80B2C100B2C120B2C160B2C200
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
B2C Bidirectional Battery Charger
All speci�cations are subject to change without notice.
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENT3channels / 0-750V 1channel / 0-750V
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
All speci�cations are subject to change without notice.
±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
±30A±45A±60A±75A±120A±150A±195A±240A±315A±390A±390A±465A±555A
B2C
_201
5_v2
27
A Battery Emulator is a power electronics equipment that behaves as a real battery pack. CINERGIA’s BE is based on a regenerative power DC supply. When emulating a battery charge, the energy will be injected back to the electrical grid consuming a sinusoidal cu-rrent with unity power factor and low harmonic distortion. Using a BE will save space in the laboratory, avoiding safety issues of real batteries and allow a high �exibility in the tests.
FUNCTIONAL DESCRIPTION
Operation modes:-Constant Voltage (CV)-Static-Dynamic
Three DC channels:-The three channels can be controlled independently, allowing di�erent battery emulation-The three channels can be controlled in parallel, emulating the same battery and providing 3 times the current
Static mode: the emulated battery voltage will depend on the open-circuit voltage and internal charge and discharge resistances.
Dynamic mode: the emulated battery voltage will depend, additionally, on the SOC and the rate of discharge.
KEY FEATURES
6.75 – 160 kW
2 Quadrant Power Supply
Regenerative up to 100% rated power
1 channel Output: 0 to 750V, 0 to ±690A
3 channels Output:0 to 750V, 0 to ±230A/ch
CV, Static, and Dynamic modes
BE Battery Emulator
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
BE_2
015_
v1
28
CONCEPTUAL SCHEMATIC
TECHNICAL DIAGRAM
AC Input is connected to the grid (neutral connection is required). Galvanic isolation is recommended.AC Output is connected to the Equipment Under Test (EUT) and can be used as:- Three independent 2Q channels- One 2Q channel (3 times rated current)
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output DCDC converter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. The DCDC converter generates three independent DC voltages controlling the voltage, current or power.
www.cinergia.coop
AC
Battery Emulator
GRID
DC DC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
-U -U
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
3 P + N 3 CH-COM
CH1CH2CH3Com_neg
AC IN (GRID side) DC OUT (EUT side)
BE Battery Emulator
BE_2
015_
v1
29
USER INTERFACE
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
Windows 7 user interface for remote operation and data acquisition.
SOFTWARE INTERFACE
0.0
100.0
200.0
300.0
400.0
500.0
600.0
700.0
800.0
0 1ch mode3ch mode-3ch mode-1ch mode
Minimumvoltage atrated power
Out
put v
olta
ge (V
)
Output current (A)
OPERATION AREA: 1/3 CHANNELS
BE Battery Emulator
STATIC MODE
Voc
Rdisch
Rch
Vout
+
-
DYNAMIC MODE
CABINET
The battery is emulated through a simply�ed model based on �xed Voc and charge/discharge resistors.
A model based on battery manufacturer data allows realistic voltage and SOC behaviour.
BE_2
015_
v1
30
www.cinergia.coop
RANGE AND SPECIFICATIONSMAGNITUDE VALUE
Power 7.5kVA-200kVAInput side (GRID side) AC Voltage Rated 3x400Vrms+Neutral+Earth Voltage range +15% / -20 % AC Current 10A-290Arms Frequency 48-62Hz THDi (at rated power) <3% Power Factor Typical ≥0.99 Con�gurable by user 0-1 (capacitive/inductive) E�ciency (at rated power) >92% Overload 125% for 10 min / 150% for 60 s
Output side (EUT side) DC Voltage Channel-Com_neg 0-750V Channel-Channel -750 to 750V Minimum voltage at rated power† 220V DC Current 1 channel output 0 to ±690A 3 channels output 0 to ±230A/ch Bipolar output 0 to ±230A Modes of operation Range Resolution Ripple Constant Voltage 0-100% <±0.1% <1% Static 0-±100% <±0.1% <1% Dynamic 0-±100% <±0.1% <1%Response time Rated resistance load 1-5ms (10-90%) General Measurements Input Voltage (Vrms) and Current (Irms) Input and Output Power Output Voltage and Current Temperatures User interface 3.2” Touchscreen Local Control port: 1 analog input, 3 analog outputs, 4 inputs, 3 relays Communication Port: Ethernet (Optionals: RS485, RS232, CAN) Communication Protocol: Modbus/TCP Humidity 10-90% (Absolute maximum, without condensation) Temperature 5-35 ºC (Absolute maximum) Cooling Forced air Protections Over Current, Over Voltage, Shortcircuit, OvertemperatureStandards CE Marking Safety EN-62040-1-2, EN-60950-1 EMC EMC: EN-62040-2
† Below minimum voltage the power is derated due to the current limitation. See operation area for further detailAll speci�cations are subject to change without notice.
BE Battery Emulator
BE_2
015_
v1
31
www.cinergia.coop
MODELS
BE7.5BE10BE15BE20BE30BE40BE50BE60BE80BE100BE120BE160BE200
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENT3channels / 0-750V 1channel / 0-750V
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
BE Battery Emulator
OPTIONSGalvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
All speci�cations are subject to change without notice.
All speci�cations are subject to change without notice.
±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
±30A±45A±60A±75A±120A±150A±195A±240A±315A±390A±390A±465A±555A
BE_2
015_
v1
32
GE-AC&DC is the integration of two equipment in one: a GE-AC and a DCPS in a single cabinet. The equipment can be used in AC or in DC test and lab platforms providing a high �exibility at a very competitive price, taking advantage of the shared hardware.
FUNCTIONAL DESCRIPTION
Two main modes:- AC mode: the three output channels work as AC Voltage Sources- DC mode: the three output channels work as DC Voltage Sources (CV) or DC Current Sources (CC, CP)
Three output channels:- Independent control- Parallel control
AC mode: -Generation of AC 3-phase or 1-phase voltages-Con�gurable frequency from 10 to 400Hz-Con�gurable phase angle and internal resistance-Generation of disturbances
DC mode:-Operation modes: CV, CC and CP-Automatic test from Excel �le-Unipolar 2Q or Bipolar 4Q applications
KEY FEATURES
7.5 –200kVA / 6.75-160kW
4 Quadrant Power Supply
Regenerative up to 100% rated power
AC Output: 0-480 Vrms, l-l 0-277 Vrms, l-n 0-230Arms 10-400Hz Disturbances DC Output:
1 channel Output: 0 to 750V, 0 to ±690A
3 channels Output:0 to 750V, 0 to ±230A/ch
Bipolar Output-750 to +750V, 0 to ±230A
GE-AC&DC AC&DC Voltage Source
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
GE_
ACD
C_2
015_
v1
33
CONCEPTUAL SCHEMATIC
TECHNICAL DIAGRAM
AC Input is connected to the grid (neutral connection or isolation transformer is required)AC Output is connected to the Equipment Under Test (EUT) and can be used as AC or DC output. The loads or EUT are connected between a phase and:- AC mode: the output neutral- DC mode: the common (COM) negative pointThe AC or DC mode is selected by software.
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output Inverter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. Depending on the main mode, the output converter can work either as an AC inverter or a DCDC converter.
www.cinergia.coop
AC
Grid Emulator
GRID
DC AC or DC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
3 P + N 3 P + N + COM AC IN (GRID side) AC OUT (EUT side)
GE-AC&DC AC&DC Voltage Source
GE_
ACD
C_2
015_
v1
34
USER INTERFACE
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
OptionsBattery Charge mode (BC)Galvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
Windows 7 user interface for remote operation and data acquisition.
SOFTWARE FEATURES
Constant VoltageHarmonicsDisturbancesConstant Current Constant PowerConstant Resistance
√√√
√
√√√
AC mode DC mode
RANGE AND SPECIFICATIONSDetailed data can be found in datasheets of GE-AC and DCPS
GE-AC&DC AC&DC Voltage Source
GE_
ACD
C_2
015_
v1
35
www.cinergia.coop
GE-AC&DC MODELS
GE7.5-AC&DCGE10-AC&DCGE15-AC&DCGE20-AC&DCGE30-AC&DCGE40-AC&DCGE50-AC&DCGE60-AC&DCGE80-AC&DCGE100-AC&DCGE120-AC&DCGE160-AC&DCGE200-AC&DC
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENT3ch (per channel) 1ch (parallel mode)ACrms DC ACrms DC10A15A20A25A40A50A65A80A105A130A155A185A230A
30A45A60A75A120A150A195A240A315A390A465A555A690A
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
All speci�cations are subject to change without notice.
All speci�cations are subject to change without notice.
±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
±30A±45A±60A±75A±120A±150A±195A±240A±315A±390A±390A±465A±555A
GE-AC&DC AC&DC Voltage Source
GE_
ACD
C_2
015_
v1
36
EL-AC&DC is the integration of two equipment in one: an EL-AC and an EL-DC in a single cabinet. The equipment can be used in AC or in DC test and lab platforms providing a high �exibility at a very competitive price, taking advantage of the shared hardware.
FUNCTIONAL DESCRIPTION
Two main modes:- AC mode: the three output channels work as AC Current Sources (CC, CP, CI)- DC mode: the three output channels work as DC Voltage Sources (CV) or DC Current Sources (CC, CP, CR)
Three output channels:- Independent control- Parallel control
AC mode: -Consumption of AC 3-phase or 1-phase current-100% con�gurable power factor, inductive and capacitive -Con�gurable harmonics-Operation modes: CC, CP, CI-Automatic test from Excel �le
DC mode:-Operation modes: CV, CC, CP and CR-Automatic test from Excel �le-Unipolar 2Q or Bipolar 4Q applications
KEY FEATURES
7.5 –200kVA / 6.75-160kW
4 Quadrant Power Supply
Regenerative up to 100% rated power
AC Output: 1 channel Output: 30-690Arms3 channels Output:10-230Arms per channel DC Output:1 channel Output: 0 to 750V, 0 to ±690A3 channels Output:0 to 750V, 0 to ±230A/chBipolar Output-750 to +750V, 0 to ±230A
CINERGIAPere IV 29-35, 5è 1ª 08018 BARCELONAwww.cinergia.coop [email protected]. +34 934 864 358
EL-AC&DC AC&DC Electronic Load
EL_A
CD
C_2
015_
v1
37
CONCEPTUAL SCHEMATIC
TECHNICAL DIAGRAM
AC Input is connected to the grid (neutral connection or isolation transformer is required)AC Output is connected to the Equipment Under Test (EUT) and can be used as AC or DC output. The loads or EUT are connected between a phase and:- AC mode: the output neutral- DC mode: the common (COM) negative pointThe AC or DC mode is selected by software.
BACK-TO-BACK TOPOLOGY
The converter is formed by a grid-side Active Recti�er and an output Inverter sharing a DC-link. The Active Recti�er allows sinusoidal current consumption with low harmonic distortion and unity power factor. Depending on the main mode, the output converter can work either as an AC inverter or a DCDC converter.
www.cinergia.coop
AC
Electronic Load
GRID
DC AC or DC
DC EQUIPMENT UNDER TEST
(EUT)
Optional
EMC �lter
EMC �lter
DSP control
DSP control
Drivers x 6 Drivers x 6Current sensor x 3
Voltage sensor x 2
Voltage sensor x 3
Current sensor x 3
Voltage sensor x 3
HMI
CAN
MODBUS/TCPETHERNET
3 P + N 3 P + N + COM AC IN (GRID side) AC OUT (EUT side)
EL-AC&DC AC&DC Electronic Load
EL_A
CD
C_2
015_
v1
38
USER INTERFACE
www.cinergia.coop
Local 3.2” Touchscreen panel
Local control port: -1 analog input 0-10V-3 analog outputs 0-10V-4 digital inputs-3 relay outputs-1 Emergency stop
Note: all inputs/outputs are isolated
Communications port: LAN Ethernet with Modbus/TCP protocol.
Optional communications: RS485, RS232, CAN, LabView
CoolingThe power supply is air-cooled internally.
Mechanical housingThe power supplies are housed in compact cabinets with wheels up to 120kVA for easier transportation.
OptionsGalvanic IsolationIsolation monitorIsolated analog inputsRS485, RS232, CANLabview drivers
Windows 7 user interface for remote operation and data acquisition.
SOFTWARE FEATURES
Constant CurrentHarmonicsConstant VoltageConstant PowerConstant ImpedanceConstant Resistance
√√
√√√
√
√√
√
AC mode DC mode
RANGE AND SPECIFICATIONSDetailed data can be found in datasheets of EL-AC and EL-DC
EL-AC&DC AC&DC Electronic Load
EL_A
CD
C_2
015_
v1
39
www.cinergia.coop
EL-AC&DC MODELS
EL7.5-AC&DCEL10-AC&DCEL15-AC&DCEL20-AC&DCEL30-AC&DCEL40-AC&DCEL50-AC&DCEL60-AC&DCEL80-AC&DCEL100-AC&DCEL120-AC&DCEL160-AC&DCEL200-AC&DC
REFERENCE
7.51015203040506080100120160200
RATEDkVA kW
6.75913.518273645547290108128160
RATED CURRENT3ch (per channel) 1ch (parallel mode)ACrms DC ACrms DC10A15A20A25A40A50A65A80A105A130A155A185A230A
30A45A60A75A120A150A195A240A315A390A465A555A690A
100100102105150175185185265290290540550
WEIGHTkg
770x450x1100
880x590x1320
850x900x2000
DIMENSIONSDxWxH (mm)
GALVANIC ISOLATION (optional)
IT7.5IT10IT15IT20IT30IT40IT50IT60IT80IT100IT120IT160IT200
REFERENCE
Type D - 25AType D - 32AType D - 50AType D - 63AType D - 80AType D - 100AType D - 125AType D - 160AType D - 200AType D - 250AType D - 315AType D - 400AType D - 500A
RECOMMENDED CIRCUIT BREAKER
6794125145174217280381435458514612753
WEIGHTkg
Inside the cabinet
595x415x708 (*)
789x490x865 (*)
964x684x1252 (*)
1192x744x1430 (*)
DIMENSIONSDxWxH (mm)
(*) The transformer is delivered in a stand-alone cabinet IP23
All speci�cations are subject to change without notice.
All speci�cations are subject to change without notice.
±10A±15A±20A±25A±40A±50A±65A±80A±105A±130A±130A±155A±185A
±30A±45A±60A±75A±120A±150A±195A±240A±315A±390A±390A±465A±555A
EL-AC&DC AC&DC Electronic Load
EL_A
CD
C_2
015_
v1
40
(ES) Equipements Scientifiques SA - Département Tests Energie Mesures - 127 rue de Buzenval BP 26 - 92380 GarchesTél. 01 47 95 99 45 - Fax. 01 47 01 16 22 - e-mail: [email protected] - Site Web: www.es-france.com