Upload
anu-antony
View
92
Download
1
Embed Size (px)
Citation preview
DEEPAK.P
UNIT 5
Instrumentation for clinical laboratory
1
DEEPAK.P2
Respiratory Transducers and Instruments
RespirationIt is responsible for bringing oxygen in to the body and discharging waste particles from body.
3 DEEPAK.P
Respiratory System Measurements1. Vital capacity
2. Functional residual capacity
3. Inspiratory capacity
4. Total Lung capacity
4 DEEPAK.P
Respiratory System MeasurementsTidal volume = volume of air inhaled during normal quiet
breathing. Residual volume (RV) = volume of air in the lungs after
maximal expiration. Functional residual capacity (FRC) = volume of air in the
lungs after exhalation during quiet breathing. Inspiratory capacity = volume of air from FRC to maximal
inhalation. Vital lung capacity from RV to maximal inhalation.
5 DEEPAK.P
Respiratory System Measurements1. Pneumograph is used to detect respiration.
2. Spiro meters are used for the following measurements.
1. Tidal Volume (TV)
2. Inspiratory Reserve Volume (IRV)
3. Expiratory reserve volume (ERV)
4. Residual Volume (RV)
5. Minute Volume
3. Pulmonary Capacities are also calculated by respiratory machine.
6 DEEPAK.P
DEEPAK.P7
Pneumograph
PneumographA Pneumograph, also known as a pneumatograph or spirograph, is a device for recording velocity and force of chest movements during respiration.
There are various kinds of pneumographic devices, which have different principles of operation.
1.In one mechanism, a flexible rubber vessel is attached to the chest and the vessel is equipped with sensors.
2.Others are impedance based.
•In first case a very thin elastic tube filled with mercury is stretched across patient chest.
8 DEEPAK.P
PneumographImpedance pneumography is a commonly-used technique
to monitor a person ’s respiration rate, or breathing rate. The idea behind the Impedance pneumograph is that, the AC
impedance across the chest of a subject changes as respiration occurs.
It is mainly used in neonatal respiration monitors.
9 DEEPAK.P
PneumographIn second case it is implemented by either using two
electrodes or Four electrodes The objective of this technique is to measure changes in the
electrical Impedance of the person ’s thorax caused by respiration or breathing.
10 DEEPAK.P
Impedance Pneumograph
11 DEEPAK.P
Impedance Pneumograph Eqlt. Ckt.
12 DEEPAK.P
Pneumograph• There are two type of pneumograph that uses piezoresistive
strain gauge transducers.
1. Mercury strainguage
2. Same wire foil or semiconductor piezoresistive devices.
• Thermistors are used as flow detectors in some pneumographs.
• In one type a bed thermistor is placed just inside the patients nostril.
• In another type thermistors are mounted on a patient who is fitted with an endotracheal tube or is on a respirator or ventillator.
13 DEEPAK.P
Thermistor airway Pneumograph
14 DEEPAK.P
Pneumograph• In some transducer , the thermistor is placed with a thin
platinum wire stretched taut across a short section of tubing.
15 DEEPAK.P
DEEPAK.P16
Spirometer
Spirometer
17 DEEPAK.P
Spirometer• A bell or jar is suspended from top in a tank of water.
• An air hose leads from a mouthpiece to the space inside of the bell above the water level.
• The weight maintains the bell at atmospheric pressure.
• When the patient exhales, the pressure inside the bell increases above atmospheric pressure causing the bell to rise.
• Similarly, When the patient inhales, the pressure inside the bell decreases causing the bell to comes down.
18 DEEPAK.P
DEEPAK.P19
Bio Electric Amplifiers
Bio Electric Amplifiers
20 DEEPAK.P
Amplitudes and spectral ranges of some important bio-signals
21 DEEPAK.P
DEEPAK.P22
Instrumentation Amplifiers
Instrumentation AmplifierAn instrumentation (or instrumentational) amplifier is a type of differential amplifier that has been outfitted with input buffer amplifiers.The input buffer eliminate the need for input impedance matching and thus make the amplifier particularly suitable for use in measurement and test equipment.
The gain of the circuit is given by
23 DEEPAK.P
Instrumentation Amplifier
24 DEEPAK.P
Instrumentation AmplifierThe two amplifiers on the left are the buffers.
The rightmost amplifier, along with the resistors is standard differential amplifier circuit.
25 DEEPAK.P
DEEPAK.P26
Isolation Amplifiers
Isolation AmplifierIsolation amplifiers are a form of differential amplifier that allow measurement of small signals in the presence of a high common mode voltage by providing electrical isolation and an electrical safety barrier.
They protect data acquisition components from common mode voltages, which are potential differences between instrument ground and signal ground.
Isolation amplifiers are used in medical instruments to ensure isolation of a patient from power supply leakage current.Amplifiers with internal transformers eliminate external isolated power supply.
27 DEEPAK.P
Isolation Amplifier
28 DEEPAK.P
Symbol of Isolation Amplifier
29 DEEPAK.P
Isolation Amplifier
30 DEEPAK.P
Isolation Amplifier
31 DEEPAK.P
Isolation Amplifier
32 DEEPAK.P
Isolation Amplifier
33 DEEPAK.P
Isolation Amplifier
34 DEEPAK.P
Isolation AmplifierThe 3650 and 3652 are optically coupled integrated circuit
isolation amplifiers.Compared to these earlier isolation amplifiers, the 3650 and
3652 have the advantage of smaller size, lower cost, wider bandwidth and integrated circuit reliability.
35 DEEPAK.P
Isolation Amplifier
36 DEEPAK.P
DEEPAK.P37
Chopper Stabilized Amplifiers
Chopper Stabilized amplifiersChopper stabilization constantly corrects input offset voltage
errors, including both errors in the initial input offset voltage and errors in input offset voltage due to time, temperature, and common-mode input voltage.
A chopper-stabilized amplifier is actually two amplification paths in parallel.
A high-accuracy, low-frequency path (A2) incorporates high gain and chopping, while high-frequency signals are amplified by the parallel wideband amplifier A1.
The outputs of both stages are subtracted in a summer amplifier, whose output is fed back to the inputs of both amplifiers through a feedback resistor.
38 DEEPAK.P
Chopper Stabilized Amplifiers
39 DEEPAK.P
Chopper Stabilized amplifiers
40 DEEPAK.P
Chopper Stabilized amplifiers
41 DEEPAK.P
Chopper Stabilized amplifiers Two problems arise when we tries to record low level bio-potentials.
Noise
DC Drift
These are worse , if we are using high gain amplifiers to amplify the weak bio-potentials.
Noises produced in amplifier circuit and human body makes the problem worse.
Drift is the change in gain or dc offset caused by thermal effects on amplifier components.
42 DEEPAK.P
Chopper Stabilized amplifiers Drift can be minimized with the use of negative feedback.
This can be avoid by convert a DC (or near dc, low frequency analog) signal to an AC signal that will pass through the amplifier.
The solution is to chop or sample the analog signal at a frequency that will pass through the AC coupled amplifier.
43 DEEPAK.P
DEEPAK.P44
Input Guarding
Input Guarding Physiological signals are low amplitude signals.
In most cases physiological signals accompanied by large CM signals.
If Op Amps are used, both differential (E)and common mode (Ecm) signals are present.
The op amp cannot distinguish artifact from real signal and C.M signal.
To avoid these problem, input guarding is used.
Here , we are placing a shield at the CM signals.
Input cable is shielded to avoid CM signals.
45 DEEPAK.P
Input Guarding Technique for increase both the input impedance of the amplifier of bio-potentials and the CMRR.
Instrumentation amplifier providing input guarding
46 DEEPAK.P
Driven-right-leg circuit reducing common-mode interference.
Input Guarding
47 DEEPAK.P
DEEPAK.P48
Flow volume Transducers
Flow volume Transducer• It measures the flow volume in litres per minute.
• The transducer assembly consists of differential pressure transducer and an airway containing a wire mesh obstruction.
• The wire mesh produces a pressure drop when it placed in an airway.
• This pressure drop is measured as a differential pressure across the mesh.
49 DEEPAK.P
Flow volume Transducer
50 DEEPAK.P
DEEPAK.P51
Blood Cell counting
Blood Cell Counting
52 DEEPAK.P
Blood Cell Counting
53 DEEPAK.P
Blood Cell Counting
54 DEEPAK.P
Blood Cell Counting
55 DEEPAK.P
Blood Cell Counting
56 DEEPAK.P
Coulter Method: Blood Cell Counting
57 DEEPAK.P
Coulter Method: Blood Cell Counting
58 DEEPAK.P
Blood Cell CountingUsing this technology, cells are sized and counted by detecting and measuring changes in electrical resistance when a particle passes through a small aperture.
This is called the electrical impedance principle of counting cells.
A blood sample is diluted in saline, a good conductor of electrical current, and the cells are pulled through an aperture by creating a vacuum.
Electrical resistance or impedance occurs as the cells pass through the aperture causing a change in voltage.
59 DEEPAK.P
Blood Cell CountingThis change in voltage generates a pulse .
The number of pulses is proportional to the number of cells counted.
The size of the voltage pulse is also directly proportional to the volume or size of the cell.
This was the principal parameter used in earlier analyzers for characterizing all cell types, but it is now used primarily for counting and sizing red blood cells and platelets.
60 DEEPAK.P
Coulter Method: Blood Cell Counting
61 DEEPAK.P
Pico Cell Blood Cell Counting
62 DEEPAK.P
Pico Cell Blood Cell Counting
63 DEEPAK.P
Optical Blood Cell Counting
64 DEEPAK.P
Optical Blood Cell Counting
65 DEEPAK.P
Optical Blood Cell Counting
66 DEEPAK.P
DEEPAK.P67
Blood Flow Meter
Blood Flow Meter• Blood flow is the continuous circulation of blood in the
cardiovascular system.
• The science dedicated to describe the physics of blood flow is called hemodynamics.
• Usually the blood flow measurements are more invasive than blood pressure measurements / ECG
The abnormal changes in the blood flow or blood velocity gives rise to malformation of vessels.
68 DEEPAK.P
Blood Flow Meter• Blood flow is nothing but the volume of blood per time
[ml/min].Typical values for blood flow [cm/s]:1. Aorta 100 – 2502. Abdominal 1003. Vena Cava 5 – 40
69 DEEPAK.P
DEEPAK.P70
EM Blood Flow Meter
EM Blood Flow Meter• The commonly used instrument for blood flow is
Electromagnetic type.
71 DEEPAK.P
EM Blood Flow Meter• The operation principle behind the electromagnetic blood
flow meters is Faraday’s law of electromagnetic induction which states that if electrical current carrying conductor moves at right angle through a magnetic field, an electromotive force is induced in the conductor.
Electromagnetic blood flow meters can be classified in to
1.Sine wave electromagnetic blood flow meter
2.Square wave electromagnetic blood flow meters
72 DEEPAK.P
EM Blood Flow Meter• The sine wave electromagnetic blood flow meter uses
wave alternating current to generate the required magnetic field.
• The flow voltage (induced voltage) is also sinusoidal.
• Even if for lower frequency application, the circuit becomes complex and high frequency application results a problem of stray capacitance effect, it can’t be used for a wide range of frequency values.
• In the square wave electromagnetic blood flow meter, the excitation is square wave alternating current, and the induced voltage is square wave too.
73 DEEPAK.P
EM Blood Flow Meter
74 DEEPAK.P
EM Blood Flow Meter
75 DEEPAK.P
EM Blood Flow Meter
76 DEEPAK.P
Types of EM Blood Flow Meter
77 DEEPAK.P
EM Blood Flow Meter
78 DEEPAK.P
EM Blood Flow Meter
79 DEEPAK.P
DEEPAK.P80
Ultrasonic Blood Flow Meter
Ultrasonic Blood Flow Meter
81 DEEPAK.P
Ultrasonic Blood Flow Meter• An ultrasonic flow meter is a type of flow meter that
measures the velocity of a fluid with ultrasound to calculate volume flow.
• The blood cells in the fluid reflects the ultrasound signal with a shift in the ultrasonic frequency due to its movement.
• Using ultrasonic transducers, the flow meter can measure the average velocity along the path of an emitted beam of ultrasound, by averaging the difference in measured transit time between the pulses of ultrasound propagating into and against the direction of the flow or by measuring the frequency shift from the Doppler effect.
82 DEEPAK.P
Ultrasonic Blood Flow Meter
83 DEEPAK.P
Ultrasonic Blood Flow Meter
84 DEEPAK.P
Ultrasonic Blood Flow Meter
85 DEEPAK.P
Ultrasonic Blood Flow Meter
86 DEEPAK.P
Ultrasonic Blood Flow Meter
87 DEEPAK.P
Ultrasonic Blood Flow Meter
88 DEEPAK.P
Ultrasonic Blood Flow Meter
89 DEEPAK.P
Ultrasonic Blood Flow Meter
90 DEEPAK.P
Ultrasonic Blood Flow Meter
91 DEEPAK.P
Doppler Blood Flow Meter
92 DEEPAK.P
DEEPAK.P93
Blood pH Measurement
Blood pH Measurement• In chemistry, pH is the negative log of the activity of the
hydrogen ion in an aqueous solution.
• Solutions with a pH less than 7 are said to be acidic and solutions with a pH greater than 7 are basic or alkaline.
• Pure water has a pH of 7.
• Mathematically, pH is the negative logarithm of the activity of the (solvated) hydronium ion, more often expressed as the measure of the hydronium ion concentration
94 DEEPAK.P
Blood pH Measurement• Blood pH is an important factor to determine the acid base
balance in the human body.
• Blood pH level plays an important role for your overall health, because if your blood pH level is acidic, your cells cannot function properly.
• One of the major contributors to acidosis is carbon dioxide, a byproduct of metabolism.
• Carbon dioxide combines with water to form carbonic acid.
• The normal pH of blood is between 7.35-7.45.
95 DEEPAK.P
Blood pH Measurement•The increase in hydrogen ion concentration causes the pH of
the body fluids to decrease.
•If the pH of the body fluids falls below 7.35, symptoms of respiratory acidosis become apparent.
•Blood gas analyzers are used to measure pH, pCO2 and pO2 etc
•pH of biological fluids is measured using a glass electrode.
96 DEEPAK.P
Blood pH Measurement
97 DEEPAK.P
Blood pH Measurement
98 DEEPAK.P
Blood pH Measurement
99 DEEPAK.P
Blood pH Measurement•A blood sample for measuring pH parameters is collected in
a glass syringe.
•The blood pH level can be determined by
1.Saliva pH test
2.Urine pH test
3.Blood pH test
100 DEEPAK.P
Saliva Blood pH Test•Saliva pH test: It is one of the fastest and simplest tests for
measuring blood pH level.
•However, it is not the most accurate.
•Before taking the test, you should wait at least two hours after eating.
•Depending on the food you eat, it can increase or decrease your blood pH level.
•To do the saliva pH test, go buy the pH paper test strip at your local drug stores and health food stores and follow the directions on the package.
101 DEEPAK.P
Urine Blood pH Test•This test works similar to the saliva pH test, but it tests urine
instead of saliva.
•The pH test strip for testing urine is the same as the saliva pH test strip.
•Blood pH test•This test is not as convenient as the saliva and urine pH test,
but it is usually more accurate.
•To do the blood pH test, you will need to visit your doctor and get your blood drawn.
102 DEEPAK.P
Blood pH Measurement•One of the best ways to bring your blood pH back to its
normal state is to eat mostly organic alkaline food. Another way is to drink distilled water.
•Distilled water is very pure water, so if you want to reduce your risk of acidosis, you will need to add some minerals to it.
103 DEEPAK.P
Blood pH Measurement
104 DEEPAK.P
Glass Electrode
105 DEEPAK.P
PH SENSITIVE FET
106 DEEPAK.P
Conductimetric PH Sensor
107 DEEPAK.P
Fiber Optic PH Sensor
108 DEEPAK.P
Fiber Optic PH Sensor
109 DEEPAK.P