Disinfection and Sterilization in Healthcare: New CDC Guidelines William A. Rutala, Ph.D., M.P.H. University of North Carolina (UNC) Health Care System and UNC at Chapel Hill, NC

11h willianrutala-novoguideline

Embed Size (px)

DESCRIPTION

cdc.guidelines

Citation preview

  • 1. Disinfection and Sterilization in Healthcare: New CDC Guidelines William A. Rutala, Ph.D., M.P.H. University of North Carolina (UNC) HealthCare System and UNC at Chapel Hill, NC

2. Disinfection and Sterilization New CDC GuidelinesProvide overviewDiscuss processes and productsEmerging pathogens and prionsSpecial instrument reprocessing issues 3. Disinfection and Sterilization in Healthcare FacilitiesRutala WA, Weber DJ and HICPAC, In press Overview Last CDC guideline in 1985 250 pages (130 pages preamble, 21 pages recommendations, glossary of terms, 15 tables, >1100 references) Evidence-based guideline (search of the literature using Medline) 4. Disinfection and Sterilization in Healthcare FacilitiesNew Sections Inactivation of emerging pathogens/bioterrorist agents/antibiotic-resistant bacteria New sterilization processes such as hydrogen peroxide gas plasma and liquid peracetic acid Toxicologic, environmental and occupational concerns associated with disinfection and sterilization processes Disinfection of medical instruments (endoscopes, probes) Inactivation of Creutzfeldt-Jakob Disease Agent 5. Efficacy of Disinfection/SterilizationInfluencing FactorsCleaning of the objectOrganic and inorganic load presentType and level of microbial contaminationConcentration of and exposure time to disinfectant/sterilantNature of the objectTemperature and relative humidity 6. Disinfection and SterilizationEH Spaulding believed that how an object will be disinfecteddepended on the objects intended use.CRITICAL - objects which enter normally sterile tissue or the vascularsystem or through which blood flows should be sterile.SEMICRITICAL - objects that touch mucous membranes or skin thatis not intact require a disinfection process (high-level disinfection[HLD]) that kills all microorganisms but high numbers of bacterialspores.NONCRITICAL -objects that touch only intact skin require low-leveldisinfection. 7. Processing Critical Patient Care ObjectsClassification:Critical objects enter normally sterile tissue or vascular system, or through which blood flows.Object:Sterility.Level germicidal action: Kill all microorganisms, including bacterial spores.Examples:Surgical instruments and devices; cardiac catheters; implants; etc.Method:Steam, gas, hydrogen peroxide plasma or chemical sterilization. 8. Critical ObjectsSurgical instrumentsCardiac cathetersImplants 9. Chemical Sterilization of Critical ObjectsGlutaraldehyde (> 2.0%) Hydrogen peroxide-HP (7.5%)Peracetic acid-PA (0.2%) HP (1.0%) and PA (0.08%) HP (7.5%) and PA (0.23%) Glut (1.12%) and Phenol/phenate (1.93%)_______________________________________________Exposure time per manufacturers recommendations 10. Processing SemicriticalPatient Care ObjectsClassification:Semicritical objects come in contact with mucous membranes or skin that is not intact.Object:Free of all microorganisms except high numbers of bacterial spores.Level germicidal action: Kills all microorganisms except high numbers of bacterial spores.Examples:Respiratory therapy and anesthesia equipment, GI endoscopes, thermometer, etc.Method:High-level disinfection 11. Semicritical ItemsEndoscopesRespiratory therapy equipmentAnesthesia equipmentEndocavitary probesTonometersDiaphragm fitting rings 12. High Level Disinfection ofSemicritical Objects Exposure Time > 12 m-30m, 20oCGermicide Concentration_____Glutaraldehyde > 2.0%Ortho-phthalaldehyde (12 m)0.55%Hydrogen peroxide* 7.5%Hydrogen peroxide and peracetic acid* 1.0%/0.08%Hydrogen peroxide and peracetic acid* 7.5%/0.23%Hypochlorite (free chlorine)* 650-675 ppmGlut and phenol/phenate**1.21%/1.93%___*May cause cosmetic and functional damage; **efficacy not verified 13. Processing NoncriticalPatient Care ObjectsClassification:Noncritical objects will not come in contact with mucous membranes or skin that is not intact.Object:Can be expected to be contaminated with some microorganisms.Level germicidal action: Kill vegetative bacteria, fungi and lipid viruses.Examples:Bedpans; crutches; bed rails; EKG leads; bedside tables; walls, floors and furniture.Method:Low-level disinfection or detergent cleaning 14. Low-Level Disinfection forNoncritical ObjectsExposure time > 1 minGermicide Use ConcentrationEthyl or isopropyl alcohol70-90%Chlorine100ppm (1:500 dilution)Phenolic UDIodophor UDQuaternary ammoniumUD___________________________________________________________________UD=Manufacturers recommended use dilution 15. Disinfection and Sterilization of Emerging Pathogens 16. Disinfection and Sterilization of Emerging PathogensHepatitis C virusClostridium difficileCryptosporidiumHelicobacter pyloriE.coli 0157:H7Antibiotic-resistant microbes (MDR-TB, VRE, MRSA)SARS Coronavirus, avian influenza, norovirusBioterrorism agents (anthrax, plague, smallpox) 17. Disinfection and Sterilization of Emerging PathogensStandard disinfection and sterilization proceduresfor patient care equipment are adequate to sterilizeor disinfect instruments or devices contaminatedwith blood and other body fluids from personsinfected with emerging pathogens 18. Creutzfeldt Jakob Disease (CJD):Disinfection and Sterilization 19. Decreasing Order of Resistance of Microorganisms toDisinfectants/SterilantsPrionsSpores Mycobacteria Non-Enveloped Viruses Fungi Bacteria Enveloped Viruses 20. Risk Assessment: Patient, Tissue, Device Patient Known or suspected CJD or other TSEs Rapidly progressive dementia Familial history of CJD, GSS, FFI History of dura mater transplant, cadaver-derived pituitary hormone injection Tissue High risk-brain, spinal cord, eyes Device Critical or semicritical 21. CJD: Disinfection and Sterilization Conclusions Critical/SC-cleaning with special prion reprocessing NaOH and steam sterilization (e.g., 1N NaOH 1h, 121oC 30 m) 134oC for 18m (prevacuum) 132oC for 60m (gravity) No low temperature sterilization technology effective* Noncritical-four disinfectants (e.g., chlorine, Environ LpH) effective (4 log decrease in LD50 within 1h)*VHP reduced infectivity by 4.5 logs (Lancet 2004;364:521) 22. Endoscopes/AERS 23. TRANSMISSION OF INFECTIONGastrointestinal endoscopy >300 infections transmitted 70% agents Salmonella sp. and P. aeruginosa Clinical spectrum ranged from colonization to death (~4%)Bronchoscopy 90 infections transmitted M. tuberculosis, atypical Mycobacteria, P. aeruginosa Spach DH et al Ann Intern Med 1993: 118:117-128 and Weber DJ, Rutala WA Gastroint Dis 2002;87 24. ENDOSCOPE DISINFECTIONCLEAN-mechanically cleaned with water andenzymatic/detergent cleanerHLD/STERILIZE-immerse scope and perfuseHLD/sterilant through all channels for at least 12 minRINSE-scope and channels rinsed with sterile water,filtered water, or tap water followed by alcoholDRY-use forced air to dry insertion tube and channelsSTORE-prevent recontamination 25. Minimum Effective Concentration Chemical SterilantDilution of chemical sterilant occurs during useTest strips are available for monitoring MECTest strips for glutaraldehyde monitor 1.5%Test strip not used to extend the use-life beyond theexpiration date (date test strips when opened)Testing frequency based on how frequently the solutionsare used (used daily, test at least daily)Record results 26. Endoscope SafetyEnsure policies equivalent to guidelines from professionalorganizations (APIC, SGNA, ASGE);policies = practicesAre the staff who reprocess the endoscope specificallytrained in that job?Are the staff competency tested at least annually?Conduct IC rounds to ensure compliance with policy 27. Comparison of Glutaraldehyde and OPA>2.0% Glutaraldehyde 0.55% Ortho-phthalaldehydeHLD: 45 min at 25oC HLD: 12 min at 20oCNeeds activator No activator needed14 day use life 14 day use life2 year shelf life 2 year shelf lifeACGIH ceiling limit, 0.05ppmNo ACGIH or OSHA limitStrong odor Weak odorMEC, 1.5% MEC, 0.3%Cost - $10/gallon Cost - $30/gallon 28. Ortho-phthalaldehyde (OPA)Contraindication for OPARepeated exposure to OPA, following manual reprocessing ofurological instruments, may have resulted in hypersensitivity insome patients with a history of bladder cancer undergoingrepeated cystoscopy.Out of approximately 1 million urological procedures, there havebeen reports of 24 patients who have experience anaphylaxis-likereactions after repeated cystoscopy (typically after 4-9 treatments).Risk control measures: residues of OPA minimized; andcontraindicated for reprocessing of urological instruments used onpatients with history of bladder cancer. 29. Clostridium difficile 30. C. difficileC. difficile is responsible for 15-25% of cases of antibiotic-associated diarrhea and for virtually all cases of antibiotic-associated pseudomembrananous colitis.Costs approx $3,669 per case or $1.1 billion per yearOverall mortality is 10-15%Over past 2 years, a new strain appears to be more virulentPatients can be contaminated from environmental surfaces,shared instrumentation, hospital personnel hands and infectedroommates Clin Microbiol Infect 2001;7:405; Clin Micro Rev 2004;17:863 31. Disinfectants and AntisepticsC. difficile spores at 10 and 20 min, Rutala et al, 2006~4 log10 reduction (3 C. difficile strains including BI-9)Clorox, 1:10, ~6,000 ppm chlorine (but not 1:50, ~1,200 ppm)Clorox Clean-up, ~1,910 ppm chlorineTilex, ~25,000 ppm chlorineSteris 20 sterilant, 0.35% peracetic acidCidex, 2.4% glutaraldehydeCidex-OPA, 0.55% OPAWavicide, 2.65% glutaraldehydeAldahol, 3.4% glutaraldehyde and 26% alcohol 32. Special Instrument Reprocessing Issues 33. Endocavitary ProbesProbes-Transesophageal echocardiography probes,vaginal/rectal probes used in sonographic scanningProbes with contact with mucous membranes aresemicriticalGuideline recommends that a new condom/probe covershould be used to cover the probe for each patient andsince covers may fail (1-80%), HLD (semicritical probes)should be performed 34. Endocavitary Probe CoversSterile transvaginal probe covers had a very high rate ofperforations before use (0%, 25%, 65% perforations fromthree suppliers)A very high rate of perforations in used endovaginal probecovers was found after oocyte retrieval use (75% and 81%from two suppliers) but other investigators found a lowerrate of perforations after use of condoms (0.9-2.0%)Condoms superior to probe covers for ultrasound probe(1.7% condom, 8.3% leakage for probe covers) 35. SterilizationThe complete elimination or destruction of allforms of microbial life and is accomplished inhealthcare facilities by either physical orchemical processes 36. Steam SterilizationAdvantages Non-toxic Cycle easy to control and monitor Inexpensive Rapidly microbicidal Least affected by organic/inorganic soils Rapid cycle time Penetrates medical packing, device lumensDisadvantages Deleterious for heat labile instruments Potential for burns 37. Minimum Steam Sterilization TimesTime at 132oC in Prevacuum SterilizerItem Minimum exposure Drying timeWrapped instruments 4 min 20-30 minTextile packs4 min5-20 min 38. Flash SterilizationFlash originally defined as sterilization of an unwrappedobject at 132oC for 3 min at 27-28 lbs pressure in gravityFlash used for items that must be used immediatelyAcceptable for processing items that cannot be packaged,sterilized and stored before useBecause of the potential for serious infections, implantedsurgical devices should not be flash sterilized unlessunavoidable (e.g., orthopedic screws) 39. Flash SterilizationWhen flash sterilization is used, certain parameters shouldbe met: item decontaminated; exogenous contaminationprevented; sterilizer function monitored by mechanical,chemical, and biological monitorsDo not used flash sterilization for reasons of convenience,as an alternative to purchasing additional instrument sets,or to save time 40. New Trends in Sterilization of PatientEquipmentAlternatives to ETO-CFCETO-CO2, ETO-HCFC, 100% ETONew Low Temperature Sterilization TechnologyHydrogen Peroxide Gas PlasmaPeracetic Acid 41. Ethylene Oxide (ETO)Advantages Very effective at killing microorganisms Penetrates medical packaging and many plastics Compatible with most medical materials Cycle easy to control and monitorDisadvantages Some states (CA, NY, TX) require ETO emission reduction of 90-99.9% CFC (inert gas that eliminates explosion hazard) banned after 1995 Potential hazard to patients and staff Lengthy cycle/aeration time 42. Hydrogen Peroxide Gas PlasmaSterilizationAdvantagesSafe for the environment and health care worker; it leavesno toxic residualsFast - cycle time is 28-73 min and no aeration necessaryUsed for heat and moisture sensitive items since processtemperature 50oCSimple to operate, install, and monitorCompatible with most medical devices 43. Hydrogen Peroxide Gas PlasmaSterilizationDisadvantagesCellulose (paper), linens and liquids cannot be processedSterilization chamber is small, about 3.5ft3 to 7.3ft3Endoscopes or medical devices restrictions based onlumen internal diameter and length (see manufacturersrecommendations)Requires synthetic packaging (polypropylene) and specialcontainer tray 44. Steris System ProcessorAdvantagesRapid cycle time (30-45 min)Low temperature (50-55oC) liquid immersion sterilizationEnvironmental friendly by-products (acetic acid, O2, H2O)Fully automatedNo adverse health effects to operatorsCompatible with wide variety of materials and instrumentsSuitable for medical devices such as flexible/rigid scopesSimulated-use and clinical trials have demonstrated microbialkilling 45. Steris System ProcessorDisadvantagesPotential material incompatibility (e.g., aluminum anodizedcoating becomes dull)Used for immersible instruments onlyBiological indicator may not be suitable for routine monitoringOne scope or a small number of instruments can be processedin a cycleMore expensive (endoscope repairs, operating costs) than HLDPoint-of-use system, no long-term storage 46. ConclusionsAll sterilization processes effective in killing sporesCleaning removes salts and proteins and must precedesterilizationFailure to clean or ensure exposure of microorganismsto sterilant (e.g. connectors) could affect effectivenessof sterilization process 47. Recommendations Methods of SterilizationSteam is preferred for critical items not damaged by heatFollow the operating parameters recommended by themanufacturerUse low temperature sterilization technologies forreprocessing critical items damaged by heatUse immediately critical items that have been sterilized byperacetic acid immersion process (no long term storage) 48. ConclusionsWhen properly used, disinfection and sterilization can ensure thesafe use of invasive and non-invasive medical devices.Method of disinfection and sterilization depends on the intendeduse of the medical deviceCleaning should always precede high-level disinfection andsterilizationCurrent disinfection and sterilization guidelines must be strictlyfollowed. 49. Disinfection and Sterilization New CDC GuidelinesProvide overviewDiscuss processes and productsEmerging pathogens and prionsSpecial instrument reprocessing issues 50. disinfectionandsterilization.org 51. Thank you 52. ReferencesRutala WA, Weber DJ. CJD: Recommendations fordisinfection and sterilization. Clin Inf Dis 2001;32:1348Rutala WA, Weber DJ. Disinfection and sterilization: Whatclinicians need to know. Clin Infect Dis 2004;39:702Rutala WA, Weber DJ, HICPAC. CDC guideline fordisinfection and sterilization in healthcare facilities. In press.Rutala WA. APIC guideline for selection and use ofdisinfectants. Am J Infect Control 1996;24:313