02/28/2017 HOW TO CLEAN AND MAINTAIN YOUR ILLUMINATED DENTAL
ISOLATION SYSTEM
2. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 2 Step One: Disassembly
3. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 3 After Each Patient, Remove the Mouthpiece from the
Control Head and Safely Dispose of It. Re-use of the mouthpiece can
result in the transmission of infectious pathogens SINGLE-USE MEANS
SAFE. Mouthpieces are single-use only to eliminate the risk of
patient cross contamination.
4. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 4 Remove the Vacuum Light Pipe from the Control Head
Press and hold the rectangular release button on the underside of
the Control Head to release and disconnect the Vacuum Light
Pipe.
5. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 5 Remove the LED Smart Stick Remove the LED Smart Stick
from the Titanium Control Base by gently pulling it out in an
arcing motion.
6. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 6 Remove the Control Head from the Power Hose Carefully
remove the Control Head by grasping the end of the Power Hose and
pulling it straight off. DO NOT TWIST
7. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 7 Cap the End of the Power Hose with the Suction Line
Plug Suction Line Plug
8. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 8 Remove the Vacuum Levers from the Titanium Control
Base Remove the Vacuum Levers by rotating the levers to the off
position. Keep pushing gently until each lever pops out. Titanium
Control Base Vacuum Levers
9. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 9 Your System Is Now Properly Disassembled Vacuum Light
Pipe LED Smart Stick Titanium Control Base Vacuum Levers
10. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 10 Step Two: Surface Disinfection Between Each
Patient
11. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 11 Between Each Patient, Perform a Surface Disinfection
of the LED Smart Stick DO NOT SUBMERSE, AUTOCLAVE, CHEMCLAVE, OR
PLACE IN ULSTRASONIC CLEANER Follow the cleaning recommendations
provided in the Instructions for Use Wipe down the surface of the
LED Smart Stick with an intermediate level disinfectant wipe with a
claim to inactivate Mycobacterium tuberculosis.
12. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 12 Between Each Patient, Perform a Surface Disinfection
of the Titanium Control Base Wipe down the surface of the Titanium
Control Base with an intermediate level disinfectant wipe with a
claim to inactivate Mycobacterium tuberculosis. Titanium Control
Base Vacuum Levers
13. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 13 Between each patient, perform a surface disinfection
of the Power Hose Wipe down the surface of the Power Hose with an
intermediate level disinfectant wipe with a claim to inactivate
Mycobacterium tuberculosis.
14. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 14 Step Three: Cleaning
15. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 15 Six Vacuum Light Pipes Are Provided So One Is Always
Ready When You Need It
16. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 16 Clean the Vacuum Light Pipe Using the provided Double
End Tube Brush and a neutral pH (6.5-7.5) enzymatic cleaning
solution, remove biomatter and debris from the suction channels.
After cleaning, rinse with tap water and dry. DO NOT HOLD PARTS IN
CLEANING SOLUTION.
17. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 17 Clean the Titanium Control Base Using the provided
Double End Tube Brush and a neutral pH (6.5-7.5) enzymatic cleaning
solution, remove biomatter and debris from the suction channels.
After cleaning, rinse with tap water and dry. DO NOT HOLD PARTS IN
CLEANING SOLUTION OR USE ULTRASONIC CLEANERS
18. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 18 According to Frequency of Use, Regularly Clean the
inside of the Power Hose to Prevent Accumulation of Biofilm PRIOR
TO CLEANING, UNPLUG THE POWER ADAPTER FROM THE POWER HOSE Start by
vacuuming water through the Power Hose. Then, vacuum a dental hose
cleaner with a pH between 6 and 8, through the Power Hose per
manufacturers instructions.
19. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 19 Step Three: Sterilization
20. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 20 Follow the sterilization recommendations provided in
the Instructions for Use Before Use on a Patient, Sterilize the
Vacuum Light Pipes
21. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 21 Before Use on a Patient, Sterilize the Vacuum Light
Pipes Place the Vacuum Light Pipes into an autoclave pouch and
sterilize in a steam autoclave using these validated sterilization
cycles. Cycle Temperature Min. Exposure Time Min. Drying Time
Standard 121 C / 250 F 30 min 30 min Standard 132 C / 270 F 15 min
15 min Prevacuum 132 C / 270 F 4 min 20 min Prevacuum 134 C / 273 F
3 min 20 min
22. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 22 Follow the sterilization recommendations provided in
the Instructions for Use Weekly, or as Needed per Standard Office
Protocol, Sterilize the Titanium Control Base Titanium Control Base
Vacuum Levers
23. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 23 Weekly, or as Needed per Standard Office Protocol,
Sterilize the Titanium Control Base Place the Titanium Control Base
and Vacuum Levers into an autoclave pouch and sterilize in a steam
autoclave using these validated sterilization cycles. Cycle
Temperature Min. Exposure Time Min. Drying Time Standard 121 C /
250 F 30 min 30 min Standard 132 C / 270 F 15 min 15 min Prevacuum
132 C / 270 F 4 min 20 min Prevacuum 134 C / 273 F 3 min 20
min
24. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 24 Step Four: Lubrication and Reassembly
25. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 25 Inspect the Vacuum Lever O-rings for Damage If the
O-rings shows signs of cracking, breakage, or crusty build-up that
can not be completely cleaned with a wet cotton swab, the O-ring
will need to be removed and replaced. If properly maintained and
lubricated, the O-ring can last for many months or years.
26. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 26 If Necessary, Remove the O-ring from the Vacuum
Levers Using a probe with a point, pinch and slide the O-ring to
one side to create a small loop. Slide the O-ring off. Insert the
probe point under the O-ring.
27. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 27 Replace the O-ring on the Vacuum Levers Roll the
O-ring onto the end of the Vacuum Lever valve barrel and into the
groove. O-ring in place.
28. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 28 Lubricate the Vacuum Lever O-rings Coat the O-ring by
applying a small amount of the Dow 111 O-ring Lubricant provided in
the O-ring Maintenance Kit. THE VACUUM LEVER O-RINGS REQUIRE
REGULAR LUBRICATION TO OPERATE AT PEAK EFFICIENCY.
29. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 29 Inspect the Power Hose O-ring for Damage If the
O-ring shows signs of cracking, breakage, or crusty build-up that
can not be completely cleaned with a wet cotton swab, the O-ring
will need to be removed and replaced. The O-ring is located less
than inch inside the blue plastic fitting at the head of the Power
Hose.
30. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 30 If Necessary, Remove the O-ring from the Power Hose
Hold the head of the Power Hose and insert a probe with a point
under the O-ring. Gently guide the probe around the groove. The
O-ring should come out. Pull out the O-ring.
31. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 31 Replace the O-ring on the Power Hose Pinch the O-ring
and insert one end into the groove. Using the probe, massage the
O-ring into the groove.
32. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 32 Lubricate the Power Hose O-ring Coat the O-ring by
applying a small amount of the Dow 111 O-ring Lubricant provided in
the O-ring Maintenance Kit. THE POWER HOSE O-RING REQUIRES REGULAR
LUBRICATION TO OPERATE AT PEAK EFFICIENCY
33. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 33 Reconnect the Control Head to the Power Hose and
Reinsert the LED Smart Stick
34. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 34 Reattach the Vacuum Light Pipe to the Control
Head
35. 2017 Isolite Systems. All rights reserved. Last Revised:
02/28/2017 35 THANK YOU. PLEASE CONTACT US IF YOU HAVE ANY
QUESTIONS OR NEED ANY ADDITIONAL SUPPORT.