Maintenance and Calibration PeriFlux 6000 | peripheral pressure made intelligent 44-00315-01

Maintenance and calibration of laser Doppler probes and pressure cuffs

Embed Size (px)

Citation preview

Page 1: Maintenance and calibration of laser Doppler probes and pressure cuffs

Maintenance and Calibration

PeriFlux 6000 | peripheral pressure made intelligent

44-00315-01

Page 2: Maintenance and calibration of laser Doppler probes and pressure cuffs

Disclaimer

The information contained in this document is intended to provide general

information only. It is not intended to be, nor does it constitute, medical advice.

Under no circumstances is the information contained in this document to be

interpreted as a recommendation for a particular treatment for specific

individuals. In all cases it is recommended that clinicians perform their own

interpretations of data in conjunction with the clinical assessment of their patient.

Due to Perimed’s commitment to continuous improvement of our products, all

specifications are subject to change without notice.

All information and content in this document is protected by copyright. All rights

are reserved. Users are prohibited from modifying, copying, distributing,

transmitting, displaying, publishing, selling, licensing, creating derivative works,

or using any information available in or through the document for commercial or

public purposes. All responsibility for any liability, loss or risk, personal or

otherwise, which is incurred as a consequence, directly or indirectly, of the use

and application of any of the material in this document is specifically disclaimed.

Page 3: Maintenance and calibration of laser Doppler probes and pressure cuffs

• Diagnosing Peripheral Arterial Disease

(PAD), Critical Limb Ischemia (CLI) and Non-

healing Wounds

• Hands on, Tips and Tricks

• Interpreting Curves and Results

• Maintenance and Calibration

Page 4: Maintenance and calibration of laser Doppler probes and pressure cuffs

Introduction

The aim of this document is to summarize the basic

maintenance requirements for laser Doppler probes and

pressure cuffs used for pressure measurements.

Page 5: Maintenance and calibration of laser Doppler probes and pressure cuffs

Cleaning laser Doppler Probes

1. Wipe down the probe/holder with a disposable cloth/paper. For

probes and probe holders that are attached using tape, tape

remnants may be removed by repeated applications of clean

adhesive tape, for instance Transpore® or Blenderm®.

2. Immerse the probe holder and/or probe head and cable in a

mild detergent and water solution. Normally approximately 15

minutes is enough.

3. Rinse the probe holder and/or probe head and cable

thoroughly under clean running water. Let dry or wipe dry with

a lint-free cloth.

4. Clean the probe connectors with a lint-free cloth moistened

with alcohol to ensure good optical transmission.

Alternatively, the probes and probe holders may be cleaned

with a cloth moistened with alcohol (<70%).

Page 6: Maintenance and calibration of laser Doppler probes and pressure cuffs

Disinfection of laser Doppler Probes

The following disinfectants are recommended for the

probes and probe holders:

• Alcohol (70%)

• Rely+On™ PeraSafe™ (DuPont™)

• CIDEX® OPA (Johnson&Johnson)

• MadaCide-FD (MADA Inc.)

• Sekusept Plus (Ecolab)

Note! Do not immerse the probe connectors in any liquid.

Page 7: Maintenance and calibration of laser Doppler probes and pressure cuffs

Cleaning of Cuffs

• If possible, all straight segmental cuffs should be

washed daily or at minimum once a week. This can

be accomplished by hand wash or by using a

washing machine with a mild detergent and gentle

cycle.

• Using a soft brush is useful to remove any organic or

inorganic material without damaging the cuff fabric.

• If there is a lot of protein contamination, such as

blood, an enzyme detergent will help remove the

organic material.

Page 8: Maintenance and calibration of laser Doppler probes and pressure cuffs

Cleaning of Cuffs cont.

1. If cuffs are grossly contaminated, keep the cuffs wet or moist to

prevent the material from drying, making cleaning more difficult.

2. Remove the bladders from the cuffs.

3. If the bladder is contaminated, wash it with soapy water and

then rinse.

Do not get water inside the bladder or tubing.

4. Add a laundry soap or mild detergent to either warm or cold

water and put the cuff covers in.

5. After washing, open the end of the cuff, where the bladder is

placed, and hang the cuff up to dry.

6. Make sure both the outside and inside are dry, including the

Velcro, before reinserting the bladders. Do not disinfect until

cuffs are completely dry.

Page 9: Maintenance and calibration of laser Doppler probes and pressure cuffs

Disinfection of Cuffs

The following disinfectants may be used on straight

segmental and digit cuffs:

• Hydrogen peroxide

• Hydrogen peroxide with silver (Sanosil®)

• Hydrogen peroxide

• Peroxyacetic acid

• Silver (Steriplex™)

• Silver with citric acid (PureGreen24™ & SpectraSan™ 24)

• Octyl decyl dimethyl ammonium chloride with dioctyl dimethyl

ammonium chloride with didecyl dimethyl ammonium chloride with

dimethyl benzyl ammonium chloride (Protex™)

• 99 % isopropyl alcohol

• T-Spray™

• Steriplex SD

Page 10: Maintenance and calibration of laser Doppler probes and pressure cuffs

Cleaning of Tubing

The tubing (Tygon S-50-HL Hospital and Surgical

tubing) can be cleaned with soap and water, or

sterilized with autoclave (steam 30 minutes at 15 psi,

121 °C) or radiation.

Page 11: Maintenance and calibration of laser Doppler probes and pressure cuffs

Checking laser Doppler Calibration

To obtain reliable measurements, a monthly check of the calibration of

the probe and the PF 6010 unit is recommended.

1. Start the instrument at least 5 minutes before check of calibration.

2. Select MENU|Maintenance|PF6010 Check Calibration.

3. In the dialog that opens, select the function unit(s) to be checked.

In case of several units, they will be checked in sequence.

Page 12: Maintenance and calibration of laser Doppler probes and pressure cuffs

Checking laser Doppler Calibration cont.

4. Check that the serial number on the connected probe is the

same as the serial number indicated in the dialog. If not, close

the dialog and go to the Change Probe procedure. Otherwise,

tap Next.

5. Place the probe in the motility standard:

– Shake the bottle containing the motility standard gently. Carefully fill

the glass container with motility standard. Avoid bubbles. Place on

a vibration free surface. •

– Place the probe tip 5 mm into the liquid. Tap cautiously to remove

potential air bubbles at the tip.

Note! When checking calibration of a PROBE 407, the probe

should be placed in the motility with a probe holder attached in

order to achieve an accurate result.

Page 13: Maintenance and calibration of laser Doppler probes and pressure cuffs

Checking laser Doppler Calibration cont.

6. Tap Start to start the check. Wait for the first step to complete.

7. If the first step is completed successfully, place the probe on

the zeroing disc using a double adhesive tape strip. Tap Start.

8. Wait for the second step to complete.

9. In case of successful verification, continue with the next unit

(step 4–9). When finished with all units, close the wizard by

tapping Close.

10. If the check fails, follow the instructions on the screen.

11. When finished, drain the motility standard back into the bottle

and secure the container lid tightly.

Page 14: Maintenance and calibration of laser Doppler probes and pressure cuffs

Changing laser Doppler Probe

When swapping laser Doppler probes, the correct probe, identified by

probe must be identified by probe type and serial number, and set,

using the Change Probe procedure.

To change probe:

1. Start the instrument at least 5 minutes before calibration.

2. Insert the probe to be used.

3. Select MENU|Maintenance|PF6010 Change Probe.

4. In the dialog that opens, select the function unit(s) to be changed.

Tap Next.

5. In the dialog that opens, select the serial number of the new probe

using the drop down menu. Tap OK.

6. Perform a check of calibration of the probe by following the

instructions on the screen.

Page 15: Maintenance and calibration of laser Doppler probes and pressure cuffs

Verification of Pressure Measurements

Enter the PF 6050 Verification Mode.

1. Select MENU|Settings|Advanced.

2. Check the Use verification mode check box. Close the

Advanced dialog by tapping OK.

3. Connect an external pump with manometer to one of the

outlets.

4. Check that the instrument displays 0 ± 0.5 mmHg.

5. Pump up to a value of 150 ± 30 mmHg. Check that the

external manometer and the instrument display the

same value ± 3 mmHg.

Page 16: Maintenance and calibration of laser Doppler probes and pressure cuffs

Verification of Pressure Measurements cont

6. Pump up to a value of 280 ± 10 mmHg. Check that the

external manometer and the instrument display the

same value ± 6 mmHg.

7. Repeat the procedure for all pressure outlets.

8. Select MENU|Settings|Advanced and un-check the

Verification mode check box.

9. Close the Advanced dialog by tapping OK.

If the external manometer and the instrument display

different values outside the given intervals, contact your

Perimed representative.

Page 17: Maintenance and calibration of laser Doppler probes and pressure cuffs

Thank You!

PeriFlux 6000 | peripheral pressure made intelligent