nice NeOumltech Medical Systems Pvt Ltd
Presentation Flow
COMPANY OVERVIEW
CLINICAL ASPECTS OF nice PHOTOTHERAPY SYSTEMS
PRODUCT OVERVIEW
TROUBLESHOOTING
Company INFO
Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive
for growth and delighted customers
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
WHY BABIES REQUIRE PHOTOTHERAPY
1Immature liver
2Hyperbilirubin
What is jaundice
Jaundice is a common temporary and usually harmless condition in newborn
infants It affect both premature and term babies Usually appearing during the
first week of the baby life
WHAT ARE THE CAUSES OF JAUNDICE
Physiological Jaundice
This is the most common cause of newborn jaundice and occurs more than
50 of babies Because the baby has immature liver and bilirubin processed
slower
Breast feeding jaundice
Breast feeding jaundice may occur when your baby does not drink enough
breast milk It occurs 5 to 10
Breast milk Jaundice
Breast milk jaundice occurs in 1 to 2 of breast fed babies This type of
jaundice starts 4 to 7 days of age It may last 3 to 10 week
Blood Group Incompatibility
If baby and mother have different blood group sometimes mother produces
antibodies that destroy the newborn red cell This cause a sudden build up of
bilirubin in the babies blood This series type of jaundice begin during 24 hours
of life
Some other reason for neonatal jaundice
1 Abnormal blood cell shapes
2 Blood type mismatch between mother and baby
3 Bleeding underneath the scalp caused by a difficult delivery
4 High level of RBC
5 Infections
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Presentation Flow
COMPANY OVERVIEW
CLINICAL ASPECTS OF nice PHOTOTHERAPY SYSTEMS
PRODUCT OVERVIEW
TROUBLESHOOTING
Company INFO
Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive
for growth and delighted customers
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
WHY BABIES REQUIRE PHOTOTHERAPY
1Immature liver
2Hyperbilirubin
What is jaundice
Jaundice is a common temporary and usually harmless condition in newborn
infants It affect both premature and term babies Usually appearing during the
first week of the baby life
WHAT ARE THE CAUSES OF JAUNDICE
Physiological Jaundice
This is the most common cause of newborn jaundice and occurs more than
50 of babies Because the baby has immature liver and bilirubin processed
slower
Breast feeding jaundice
Breast feeding jaundice may occur when your baby does not drink enough
breast milk It occurs 5 to 10
Breast milk Jaundice
Breast milk jaundice occurs in 1 to 2 of breast fed babies This type of
jaundice starts 4 to 7 days of age It may last 3 to 10 week
Blood Group Incompatibility
If baby and mother have different blood group sometimes mother produces
antibodies that destroy the newborn red cell This cause a sudden build up of
bilirubin in the babies blood This series type of jaundice begin during 24 hours
of life
Some other reason for neonatal jaundice
1 Abnormal blood cell shapes
2 Blood type mismatch between mother and baby
3 Bleeding underneath the scalp caused by a difficult delivery
4 High level of RBC
5 Infections
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Company INFO
Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology we always strive
for growth and delighted customers
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
WHY BABIES REQUIRE PHOTOTHERAPY
1Immature liver
2Hyperbilirubin
What is jaundice
Jaundice is a common temporary and usually harmless condition in newborn
infants It affect both premature and term babies Usually appearing during the
first week of the baby life
WHAT ARE THE CAUSES OF JAUNDICE
Physiological Jaundice
This is the most common cause of newborn jaundice and occurs more than
50 of babies Because the baby has immature liver and bilirubin processed
slower
Breast feeding jaundice
Breast feeding jaundice may occur when your baby does not drink enough
breast milk It occurs 5 to 10
Breast milk Jaundice
Breast milk jaundice occurs in 1 to 2 of breast fed babies This type of
jaundice starts 4 to 7 days of age It may last 3 to 10 week
Blood Group Incompatibility
If baby and mother have different blood group sometimes mother produces
antibodies that destroy the newborn red cell This cause a sudden build up of
bilirubin in the babies blood This series type of jaundice begin during 24 hours
of life
Some other reason for neonatal jaundice
1 Abnormal blood cell shapes
2 Blood type mismatch between mother and baby
3 Bleeding underneath the scalp caused by a difficult delivery
4 High level of RBC
5 Infections
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Timeline-nice Neotech
2005bullFirst New Factory Unit 1 Opened
bullIncrease in production Line
2001bullRegistered as Pvt ltd
bullProduction started for different models
2013bullSecond New Factory Unit 2 Opened
1999bull Neotech Logo
Designed
bull Designed first incubator amp open
care prototype
2003bull Export to
Vietnam
2009-2011
bull CE certification for products
2015bull 125+skilled
employee strong firm
bull Multiple range of NICU products
1997bullStarted Neotech Medical Systems with 2 employees
bullFirst Infant Radiant Warmer
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
WHY BABIES REQUIRE PHOTOTHERAPY
1Immature liver
2Hyperbilirubin
What is jaundice
Jaundice is a common temporary and usually harmless condition in newborn
infants It affect both premature and term babies Usually appearing during the
first week of the baby life
WHAT ARE THE CAUSES OF JAUNDICE
Physiological Jaundice
This is the most common cause of newborn jaundice and occurs more than
50 of babies Because the baby has immature liver and bilirubin processed
slower
Breast feeding jaundice
Breast feeding jaundice may occur when your baby does not drink enough
breast milk It occurs 5 to 10
Breast milk Jaundice
Breast milk jaundice occurs in 1 to 2 of breast fed babies This type of
jaundice starts 4 to 7 days of age It may last 3 to 10 week
Blood Group Incompatibility
If baby and mother have different blood group sometimes mother produces
antibodies that destroy the newborn red cell This cause a sudden build up of
bilirubin in the babies blood This series type of jaundice begin during 24 hours
of life
Some other reason for neonatal jaundice
1 Abnormal blood cell shapes
2 Blood type mismatch between mother and baby
3 Bleeding underneath the scalp caused by a difficult delivery
4 High level of RBC
5 Infections
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Why nice Neotech
User friendly range of products Customized solutions Ethical Business practices Stringent quality control management Complete customer satisfaction
WHY BABIES REQUIRE PHOTOTHERAPY
1Immature liver
2Hyperbilirubin
What is jaundice
Jaundice is a common temporary and usually harmless condition in newborn
infants It affect both premature and term babies Usually appearing during the
first week of the baby life
WHAT ARE THE CAUSES OF JAUNDICE
Physiological Jaundice
This is the most common cause of newborn jaundice and occurs more than
50 of babies Because the baby has immature liver and bilirubin processed
slower
Breast feeding jaundice
Breast feeding jaundice may occur when your baby does not drink enough
breast milk It occurs 5 to 10
Breast milk Jaundice
Breast milk jaundice occurs in 1 to 2 of breast fed babies This type of
jaundice starts 4 to 7 days of age It may last 3 to 10 week
Blood Group Incompatibility
If baby and mother have different blood group sometimes mother produces
antibodies that destroy the newborn red cell This cause a sudden build up of
bilirubin in the babies blood This series type of jaundice begin during 24 hours
of life
Some other reason for neonatal jaundice
1 Abnormal blood cell shapes
2 Blood type mismatch between mother and baby
3 Bleeding underneath the scalp caused by a difficult delivery
4 High level of RBC
5 Infections
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
WHY BABIES REQUIRE PHOTOTHERAPY
1Immature liver
2Hyperbilirubin
What is jaundice
Jaundice is a common temporary and usually harmless condition in newborn
infants It affect both premature and term babies Usually appearing during the
first week of the baby life
WHAT ARE THE CAUSES OF JAUNDICE
Physiological Jaundice
This is the most common cause of newborn jaundice and occurs more than
50 of babies Because the baby has immature liver and bilirubin processed
slower
Breast feeding jaundice
Breast feeding jaundice may occur when your baby does not drink enough
breast milk It occurs 5 to 10
Breast milk Jaundice
Breast milk jaundice occurs in 1 to 2 of breast fed babies This type of
jaundice starts 4 to 7 days of age It may last 3 to 10 week
Blood Group Incompatibility
If baby and mother have different blood group sometimes mother produces
antibodies that destroy the newborn red cell This cause a sudden build up of
bilirubin in the babies blood This series type of jaundice begin during 24 hours
of life
Some other reason for neonatal jaundice
1 Abnormal blood cell shapes
2 Blood type mismatch between mother and baby
3 Bleeding underneath the scalp caused by a difficult delivery
4 High level of RBC
5 Infections
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Breast milk Jaundice
Breast milk jaundice occurs in 1 to 2 of breast fed babies This type of
jaundice starts 4 to 7 days of age It may last 3 to 10 week
Blood Group Incompatibility
If baby and mother have different blood group sometimes mother produces
antibodies that destroy the newborn red cell This cause a sudden build up of
bilirubin in the babies blood This series type of jaundice begin during 24 hours
of life
Some other reason for neonatal jaundice
1 Abnormal blood cell shapes
2 Blood type mismatch between mother and baby
3 Bleeding underneath the scalp caused by a difficult delivery
4 High level of RBC
5 Infections
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Side Effects of Jaundice
In babies whose bilirubin blood levels reach harmful levels bilirubin may get into the brain and cause reversible damage (called acute bilirubin encephalopathy) or permanent damage (called kernicterus) Frequent monitoring and early treatment of infants at high risk for jaundice can help to prevent severe hyperbilirubinemia
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
JAUNDICE TREATMENT
The goal of jaundice treatment is to quickly and safely reduce the level of bilirubin
Phototherapy mdash Phototherapy (light therapy) is the most common medical treatment for jaundice in newborns In most cases phototherapy is the only treatment required It consists of exposing an infants skin to a special blue light which breaks bilirubin down into parts that are easier to eliminate in the stool and urine Treatment with phototherapy is successful for almost all infants
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
What is Phototherapy
Phototherapy (light treatment) is the process of using light to eliminate bilirubin in the blood babys skin and blood absorb these light waves These light waves are absorbed by your babys skin and blood and change bilirubin into products which can pass through their urine system
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Phototherapy System
Bilirubin is a neurotoxic breakdown product of red blood cell haemoglobin which if allowed to concentrate in the blood causes the yellow skin coloration associated with jaundice and can result in irreversible brain damage
Exposing the affected infant to light at the blue end of the visible spectrum(420-480nm) has been shown to reduce the concentration of bilirubin by causing it to become more soluble and therefore more easily excreted from the bodyThe process of breaking down of bilirubin is by PHOTOISOMERIZATION the light-initiated process of change from one isomeric form of a compound radical or ion to another
outcome
The Phototherapy provides light of suitable wave-length and intensity for the treatment of hyperbilirubinaemia
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Advantages of Phototherapy Systems
Breakdown Bilirubin
Non-Invasive Treatment
Prevent brain damage from very high bilirubin level
Ensure baby is stabilized after birth
Phototherapy Systems
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Treat Jaundice Safer and Faster Introducing nice NeOumltech LEDCFL series Phototherapy Systems
Because every baby deserve nice care
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
nice
400
0 se
ries
LED
CF
L P
hoto
ther
apy
Sys
tem
s
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
bullLow energy consumptionbullLamp usage for more than 20000hrsbullNo infra red rays in the LEDbullNo ultra violet rays in the LEDbullNo water lossbullFeather touch keysbullMicro processor controlledbullVertical pole with height adjustmentbullIrradiance level adjustment from 25 to 100bullProgrammable treatment timer bullLamp usage hour meterbullTri color indication for lamp usagebullLarge coverage area with uniform light distributionbullEffective treatment with minimum loss of light
Features
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
How to operate
Control Panel Operation Light Source Alarms Effective Surface Area
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Control Panel Operation
LED Phototherapy unit ndash Sapphire 4000
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Control Panel Operation
LED Phototherapy unit ndash Seefar 4000
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Light Source
Treatment light- LED and CFLLight source height Adjustment Light Source tilting
Treatment Light- LED phototherapy unitHigh bright 924 Blue LED Lamps - The light provides an average intensity of 66μWcm2nm at a distance of 30cm (12 inches) from the baby
Treatment Light- CFL Phototherapy unit
bull4 Blue Lamps and 2 White Lamps - The bulllight provides an average intensity of 18-25 μWcm2nm at a distance of 30 cm (12 inches) distance from the babybull 6 Blue Lamps - The light provides an averagebull intensity of 25 ndash 30 μWcm2nm at a distance bullof 30 cm (12 inches) distance from the baby
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Light source height Adjustment
This knob allows adjusting the height of the light Source First loosen the knob then adjust the height of the light Source by holding the ellipse pipe and finally tighten the knob to lock the height
Light Source tiltingLED Light Source can be tilted by grasping the device on either side and torquing to desired angle Even when the baby is in the incubator or warmer effective phototherapy treatment can be given by tilting the light source to the required angle (plusmn90ordm)
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Effective Area
The following graph shows the normalized spectrum of blue lamp and the spectral sensitivity of the spectrophotometer
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Alarms
Timer elapsed After elapsing the set time
Lamp Replacement gt 20000 hours
Over temperature If light source get overheat
Fan Fail If the fan fails
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Comparison of nice Phototherapy Systemsnice 4000 LED SAPPHIRE
nice 4000 SPOT LED SEEFAR
nice 4005 SPOT
nice 4000 CFL
nice 4005 CFL
LED Lamp Y Y Y N N
CFL Lamp N N N Y Y
Lamp wattage 3W 1W 1W 18W 18W
No of lamps 9 24 24 6 6
Height adjustment Y Y N Y N
Intensity adjustment Y Y Y N N
Under surface N N Y N Y
Observational lamp N Y Y Y Y
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
How nice Phototherapy Systemsfare over competitors
nice 4000 series LEDCFL Phototherapy Systems
User-friendly easy to access all parameters
Long life gt20000hrs
Lamp usage indicators
Plastic and aluminium base
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Comparison Graph-LED vs CFL vs Conventional Light sources
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
TROUBLESHOOTING
AND MAINTENANCE
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Cleaning Procedure
Use an Adopter or Quaternary Disinfectant Registered by the Environment Protection Agency but only after the Phototherapy is empty and disassembled such as Kleenasepticregb Germicidal Surface Cleanser may be used When using any cleaning agent follow the manufacturerrsquos directions for use After removing all solid wastes and contaminants from the disassembled parts clean them as follows
Lamp Protective shield Wipe the outer of transparent Portions and clean with a cloth dampened with a disinfectant-detergent
Light source and Pedestal Stand Use a disinfectant-detergent to clean all surfaces thoroughly and then dry with a clean cloth or paper towel
Remove dust from the exterior of the light with a soft brush or soft cloth dampened with water Sponge away remaining debris with a mild solution of detergent and water a noncaustic commercial cleaner or hospital disinfectant
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Preventive Maintenance General
bull Always disinfect and clean the unit and accessories before any maintenance ndash even when returning the unit to the supplier for repairbull Always disconnect power supply before any maintenancebull Use only nice Neotechrsquos original parts for maintenance
Warning Periodically check the insulation and the connection of the cable it may cause fire because of poor insulation and short circuits due to agingWarning Donrsquot misaligned the EMI Shielding and the beads as it may cause the EMI interference to the equipment
Safety ChecksThe following must be performed on this unit at least every 6Visual checks of the unit and accessories for mechanical damage affecting functions1048766 Clarity of Acrylic protective lens In the event of wear or poor clarity replace the same1048766 Check the Height Adjustment1048766 Check the Light source tilting1048766 Check Safety Warning notices for lsquoLegibilityrsquo1048766 Check equipment fuse links for rated current and fusing characteristic1048766 Perform functional check according to operating instructions (section 3)1048766 Check the power supply cord
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
TroubleshootingGeneral System FailurePhototherapy Failure
General System Failure
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Phototherapy Failure
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Phototherapy Failure (Contrsquodhellip)
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Cautions Use cleaning solution sparingly on a cloth when cleaning the Infant
Radiant Warmer Do not saturate the unit - excessive solution causes damage to internal
components Do not autoclave or gas sterilize the skin temperature probe Do not
immerse the probe in liquid cleaner Avoid placing excessive strain on the probe lead Do not pull on the probe lead These precautions will help avoid damage to the probe
Do not autoclave or gas sterilize the mattress Only competent individuals trained in the repair of this equipment
should attempt to service it as detailed in the service manual The Service Manual provides detailed information solely for use by individuals having proper knowledge tools and test equipment and for service representatives trained by nice Neotech
Isolation from the supply mains is Detachable Power cord provided Use of nonstandard components Consult the manufacturer for repair
and replacement of components Use of incorrect component can adversely affect Safety performance andor damage the
equipment performance
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Certifications and Approvals
Electrical Safety IEC 60601-1Particular product Electrical safety IEC 60601-2-50
EMC Safety IEC 60601-1-2
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594
Contact Us
nice Neotech Medical Systems Private Limited
No 85 Krishna Industrial Estate
Mettukuppam Vannagaram
Chennai ndash 600095 Tamil Nadu
INDIA
Tel +91-44-2476259424764608
Fax +91-44-24766920
Web wwwniceneotechcom
E-mail infoniceneotechcom marketingniceneotechcom
Overseas Enquiry overseasniceneotechcom
Toll Free +91-1800-425-2594