BA381C/07/en/POD
Valid as of:
Software version 00.00.16
Operating Instructions
Liquiline M CM42
Two-wire transmitter for pH/redox measurement
Part 1: Commissioning
Overview of documentation supplied
Operating Instructions
The Operating Instructions are split into two parts:
• Commissioning
– Description of your device
– All steps that you have to carry out once only during initial commissioning
• Operation
– Software description
– Trouble-shooting
– All steps that you have to carry out during the routine operation of your device,
depending on your measuring task
! Note!
This part of the Operating Instructions is the documentation on commissioning.
CD-ROM
Contents:
• Fundamentals of pH and conductivity measurement, fieldbuses
• Overview of analytical measuring technology
• Accessories
• Tables
• Applications
• Safety
• Ordering information
In addition
• Adhesive labels:
– Replacement nameplate
– Wiring adhesive label
Endress+Hauser
Liquiline M CM42
Table of contents
1 Safety instructions . . . . . . . . . . 4
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . 4
1.2 Incoming acceptance, transport, storage . . 4
1.3 Installation, commissioning and operation 4
1.4 Operational safety . . . . . . . . . . . . . . . . . . 5
1.5 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.6 Notes on safety conventions and icons . . . 5
2 Identification . . . . . . . . . . . . . . 6
2.1 Device designation . . . . . . . . . . . . . . . . . . 6
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . 8
2.3 Reordering documentation . . . . . . . . . . . . 8
2.4 Certificates and approvals . . . . . . . . . . . . . 8
3 Device description . . . . . . . . . . 9
3.1 Housing closed . . . . . . . . . . . . . . . . . . . . . 9
3.2 Housing open . . . . . . . . . . . . . . . . . . . . . . 9
4 Installation . . . . . . . . . . . . . . . 10
4.1 Measuring system . . . . . . . . . . . . . . . . . 10
4.2 Installation conditions . . . . . . . . . . . . . . 11
4.3 Installation instructions . . . . . . . . . . . . . 12
4.4 Post-installation check . . . . . . . . . . . . . . 15
5 Wiring . . . . . . . . . . . . . . . . . . 16
5.1 Electrical connection . . . . . . . . . . . . . . . 16
5.2 Post-connection check . . . . . . . . . . . . . . 23
6 Commissioning. . . . . . . . . . . . 25
6.1 Function check . . . . . . . . . . . . . . . . . . . 25
6.2 Quick Setup . . . . . . . . . . . . . . . . . . . . . . 25
6.3 User administration . . . . . . . . . . . . . . . . 27
6.4 Calibration . . . . . . . . . . . . . . . . . . . . . . . 28
7 Accessories. . . . . . . . . . . . . . . 30
Index . . . . . . . . . . . . . . . . . . . 31
Safety instructions Liquiline M CM42
4 Endress+Hauser
1 Safety instructions
1.1 Designated use
Liquiline CM42 is a two-wire transmitter for liquid analysis in all areas of process engineering:
• Chemical processes
• Pharmaceutical industry
• Food technology
If the device is used for any purpose other than that described, this poses a threat to the safety
of people and the entire measuring system and is thus not permitted.
The manufacturer does not accept liability for damage caused by improper or non-designated
use.
1.2 Incoming acceptance, transport, storage
• Make sure the packaging is not damaged!
Inform your supplier of any damage to the packaging.
Please keep the damaged packaging until any issues have been resolved.
• Make sure the contents are not damaged!
Inform your supplier of any damage to the contents.
Please keep the damaged goods until any issues have been resolved.
• Check the scope of delivery against the delivery papers and your order to ensure it is complete
and nothing is missing.
• Pack the device in such a way as to protect it reliably against impact and moisture for storage
and transportation. Optimum protection is provided by the original packaging materials. In
addition, the permitted ambient conditions must be observed (see Technical data).
• If you have any queries, please contact your supplier or local sales center.
1.3 Installation, commissioning and operation
Observe the following points:
• Installation, commissioning, operation and maintenance of the measuring system must only
be carried out by trained technical personnel.
The technical personnel must be authorized to perform the tasks by the owner-operator.
• The electrical connection may only be established by an electrical technician.
• The technical personnel must have read and understood these Operating Instructions and
must follow the instructions they contain.
• Prior to commissioning the entire measuring point, check that all connections are correct.
Make sure that electric cables and hose connections are not damaged.
• Do not commission damaged products. Protect them against unintentional start-up. Label and
identify the damaged product as defective.
• Faults at the measuring point may only be rectified by authorized and properly trained
personnel.
Liquiline M CM42 Safety instructions
Endress+Hauser 5
• If the faults cannot be eliminated, take the products out of service and protect them against
unintentional start-up.
• Repairs not described in these Operating Instructions may only be carried out directly at the
manufacturer's or by the Service Organization.
1.4 Operational safety
The transmitter is designed to meet state-of-the-art safety requirements, has been tested and left
the factory in a condition in which it is safe to operate. The applicable regulations and European
standards have been taken into account.
As the user, you are responsible for observing the following safety regulations:
• Installation guidelines
• Local standards and regulations
1.5 Return
If returning the transmitter, please send it cleaned to your sales center.
Use the original packaging when returning the device.
1.6 Notes on safety conventions and icons
1.6.1 Warnings
1.6.2 Document symbols
#Warning!
This symbol alerts you to dangers.
If the warning is not observed, this can result in serious injury to people or damage
to property.
"Caution!
This symbol alerts you to possible malfunctions resulting from incorrect operation.
If not observed, this could result in damage to property.
! Note!
This symbol indicates important items of information.
È ä 1 This symbol stands for a cross-reference to a certain page (e.g. Page 1).
È å 2 This symbol stands for a cross-reference to a certain graphic (e.g. Fig. 2).
Identification Liquiline M CM42
6 Endress+Hauser
2 Identification
2.1 Device designation
2.1.1 Nameplate
Nameplates can be found:
• On the outside of the housing
• On the packaging
• On the adhesive label sheet provided for your own use
Compare the data on the nameplate with your order.
2.1.2 Serial number and order code
The order code and serial number of your device can be found in the following locations:
• On the nameplate
• Printed under the title drawing of the Operating Instructions
• In the delivery papers
! Note!
• Please note the documentation only contains information that was valid at the time of
delivery. If you have any queries, please contact your local sales center.
• You need the serial number of your device and that of the DAT memory module (inside the
display cover, adhesive label on the module) every time you reorder.
a0003281
Fig. 1: Nameplate information
The nameplate provides you with the following
information on your device:
1. Order code
2. Serial number
3. Device version
4. Measuring range(s)
5. Input, output values
6. Approvals (depending on the version ordered)
7. Protection class
8. Ambient temperature range
9. Approval symbols
10. Safety notices, warnings
Liquiline M CM42 Identification
Endress+Hauser 7
2.1.3 Product structure
Sensor input
A No module
C Conductivity, conductive measurement
I Conductivity, inductive measurement
M Digital sensor, Memosens
P pH (glass/ISFET) or redox
R pH (glass/redox redundant) or ISFET
Approval
A Non-hazardous area, CSA GP
G Atex II (1)2 G EEx ia/ib IIC T6 / II 3 D
H Atex II 3G EEx nA II T6
N NEPSI Ex nA II T6
P FM IS NI Cl. I, II, III, Div. 1&2, Groups A?G
S CSA IS NI Cl. I, II, III, Div. 1&2, Groups A?G
T TIIS
Certificate
A Not selected
B Test certificate to EN 10204, 3.1
C Test certificate to EN 10204, 3.1, factory calibration certificate
Output
0 1 current output 4 ... 20 mA, Hart®
1 2 current outputs 4 ... 20 mA, 1x Hart®
2 PROFIBUS PA
3 FOUNDATION Fieldbus
Housing
0 Plastic housing
1 Stainless steel housing 1.4404 (AISI 316L)
Cable entry
0 M20 x 1.5
1 NPT ½"
2 G½
Software
EA Standard version
EB Advanced version
EC Advanced Plus version
Device language
D De/En (Docs, German)
E En/De (Docs, English)
F Fr/En (Docs, French)
S Es/En (Docs, Spanish)
Documentation
0 Installation + operation
Additional fittings
0 Basic version
CM42- Complete order code
Identification Liquiline M CM42
8 Endress+Hauser
2.2 Scope of delivery
• 1 transmitter in the version ordered
• 1 wall securing unit incl. 4 flat head screws
• 1 sheet of adhesive labels (nameplates, terminal connection diagrams)
• 1 Operating Instructions Part 1, BA381C "Commissioning"
• 1 Operating Instructions Part 2, BA382C "Operation"
• 1 CD-ROM with additional documentation
If you have any queries, please contact your supplier or local sales center.
2.3 Reordering documentation
• Reordering documentation (both parts of BA, pdf files):
– Via Internet (www.endress.com/liquiline-documentation)
You need the serial number of your device to reorder items.
You receive the same documentation as received with the delivery.
• Initial ordering of appropriate documentation (both parts of BA, pdf files) for devices which
you have assembled yourself by inserting or replacing modules:
– Via Internet (www.endress.com/liquiline-documentation)
You need the serial number of your device and of the installed modules to reorder items. You
also have to indicate the order code the device corresponds to after installing the module.
You receive the documentation to suit your newly modified device.
2.4 Certificates and approvals
2.4.1 CE mark: Declaration of conformity
With this declaration, the manufacturer guarantees that the product conforms to the regulations
of European Directive 89/336/EEC. This is proven by observing the standards listed in the
declaration of conformity.
! Note!
The EC type-examination certificate can be found under the "Certificates" category on the
CD-ROM supplied.
Liquiline M CM42 Device description
Endress+Hauser 9
3 Device description
3.1 Housing closed
3.2 Housing open
a0001068
Fig. 2: CM42 with closed display cover
1
2
3
4?7
8
9
Wall securing plate
Display
Navigator
Soft keys,
assignment depends on menu
Alarm LED
PML terminal
a0002755
Fig. 3: With open display cover (without wiring)
1
2
3
DAT memory modules
CPU module
Transmitter module
4
5
6
DIN rail
Threaded joints
CDI socket (service interface)
2
3
4567
1
9
8
1 2 3
456
Installation Liquiline M CM42
10 Endress+Hauser
4 Installation
4.1 Measuring system
A complete measuring system consists of:
• Liquiline M CM42
• An assembly:
– immersion assembly, e.g. CPA140 or
– flow assembly, e.g. CPA240 or
– retractable assembly, e.g. CPA471
each with or without potential matching pin (PML)
• A pH/redox sensor:
– pH/redox combined electrode, e.g. CPS71 or
– pH?ISFET sensor, e.g. CPS471
• A relevant measuring cable:
– CPK9 (pH)
– CPK12 (ISFET)
a0002122
Fig. 4: Example of a measuring system
1 Liquiline M CM42
2 pH cable CPK9
3 pH/redox sensor
4 Retractable assembly CPA471 (manual version)
1
23
4
Liquiline M CM42 Installation
Endress+Hauser 11
4.2 Installation conditions
4.2.1 Dimensions
a0001073–en
Fig. 5: Dimensions
145
/5.7
1
134
/5.2
8
110
/4.3
3
132.5 / 5.22
136.5 / 5.37
160
/6.3
0
134.5 / 5.30
136.5 / 5.37
26 /1.02
174 / 6.85
174
/6.8
5
10/
0.3
9
3/0.1
2
220 / 8.66
95
/3.7
4
mm / inch
Installation Liquiline M CM42
12 Endress+Hauser
4.2.2 Mounting plate
4.2.3 Weather protection cover
! Note!
When installing outside, always use the weather protection cover (see 'Accessories', CD-ROM)
to protect the transmitter against rain and snow.
a0001676?en
Fig. 7: Weather protection cover
4.3 Installation instructions
4.3.1 Wall or field mounting
There are many ways of mounting the device:
• Wall mounting
• Mounting on a vertical pipe or post (round or square)
• Mounting on railing or on a horizontal pipe (round or square)
a0001075–en
Fig. 6: Mounting plate
320
/12.6
0
270 / 10.63
mm / inch
300 / 11.81
Liquiline M CM42 Installation
Endress+Hauser 13
! Note!
You require the field mounting kit to mount on a pipe or post. This is an accessory and is not
included in the scope of delivery (see "Accessories").
a0002161
Fig. 8: Mounting on a vertical pipe/post
1 CM42 (as delivered, incl. a mounting plate)
2 Securing screws (part of the field mounting kit, see Accessories)
a0003090
Fig. 9: Side view
3 Mounting plate (field mounting
kit)
4 Pipe or post
a0002163
Fig. 10: Mounting on a horizontal pipe/railing
1 CM42 (as delivered)
2 Securing screws (field mounting kit)
a0003091
Fig. 11: Side view
3 Mounting plate (field mounting
kit)
4 Pipe or railing
1 2
34
1
4 3
2
2
3
1
4
1
4 3
2
Installation Liquiline M CM42
14 Endress+Hauser
4.3.2 Panel mounting
! Note!
• For panel mounting, you need the installation kit consisting of tensioning screws and a front
seal. This is an accessory (see "Accessories") and is not included in the scope of delivery.
• If installing the units above one another, you must observe a minimum distance for the
cable glands of the upper device.
• If installing the units beside one another, you must observe a minimum distance for opening
the front of the housing.
• If arranging in a square, you must take into account the lengths of the mounting plates at the
rear of the device or the cable glands for minimum spacing distance.
a0002165
Fig. 12: Wall mounting (optional weather protection cover)
a0002396?en
Fig. 13: Panel mounting several devices above one another: left side view, right front view
Liquiline M CM42 Installation
Endress+Hauser 15
4.4 Post-installation check
• After installation, check the transmitter for damage.
• Check whether the transmitter is protected against moisture and direct sunlight.
a0002630?en
Fig. 14: Panel mounting several devices next to one another: left side view, right front view
Wiring Liquiline M CM42
16 Endress+Hauser
5 Wiring
5.1 Electrical connection
# Warning!
• The electrical connection may only be established by an electrical technician.
• The electrical technician must have read and understood these Operating Instructions and
must follow the instructions they contain.
• Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
5.1.1 Opening the housing
1. Loosen the four screws on the front with a Phillips screwdriver:
2. Open the housing with the aid of the suction button supplied:
" Caution!
Do not use any sharp or pointed objects, such as a screwdriver, knife etc., to open the housing.
Otherwise, you might damage the housing seal, scratch the housing etc.
Liquiline M CM42 Wiring
Endress+Hauser 17
5.1.2 Housing grounding
" Caution!
You must connect the outer ground connection of the housing to the foundation ground with a
separate 4 mm2 line (GN/YE).
5.1.3 Cable grounding in housing
! Note!
• If possible, only use terminated genuine cables.
• Ground all the cables in the housing of the transmitter as shown in the following diagrams
(sample cable, does not necessarily correspond to the original cable).
a0003616
Fig. 15: Grounding the housing
1
2
Outer ground connection
4 mm2 line (GN/YE)
1
2
Wiring Liquiline M CM42
18 Endress+Hauser
5.1.4 Cable terminals
a0002679
Fig. 16: Terminated cable (example)
1 Outer shield (exposed)
2 Cable cores with ferrules
3 Cable sheath (insulation)
a0002680
Fig. 17: Inserting the cable
4 Grounding clip
a0002681
Fig. 18: Grounding the cable
(functional
ground)
a0003602
Fig. 19: Opening the terminal
1. Insert a suitable screwdriver into the opening
of the terminal spring (square opening) until
the stop.
a0003603
Fig. 20: Inserting the cable
2. Insert the terminated cable end into the
terminal opening (round opening).
a0003604
Fig. 21: Releasing the terminal spring
3. Remove the screwdriver. Make sure the cable
is securely positioned in the terminal.
1 2
3
4
Liquiline M CM42 Wiring
Endress+Hauser 19
5.1.5 Signal output / power supply
! Note!
Connect the transmitter with two-wire cabling shielded on both sides.
5.1.6 Sensor connection
Explanation of abbreviations in the following diagrams:
! Note!
• Cable colors indicated as per IEC 757 (see CD–ROM).
• You must connect shielded connections and terminals with functional ground ()) (there is
no protective ground (*) for plastic housings).
a0002379
Fig. 22: View in device (CPU module)
a0001092
Fig. 23: Wiring diagram
Abbreviation Meaning
pH Signal from pH membrane glass
Ref Signal from reference electrode
Src Source
Drn Drain
PM Potential matching
ϑ Signal of the temperature sensor
d.n.c. Do not connect!
+
–
23
42
33
23
22
31
13
41
33
13
21
31
133
134 +
–
133
134
4..
.2
0m
A
Wiring Liquiline M CM42
20 Endress+Hauser
With PML (symmetrical)
Glass electrodes
a0002385
Fig. 24: View in device (sensor module)a0001079
Fig. 25: Wiring diagram
PM
BN
GN
WH
YE
31
73
18
32
03
16
31
5111
11
311
23
19
112
113
111
320
318
317
BK
BN
BK
BN
GN
WH
YE
CPK9
318
320
112
113
111
317
CM42
319
316
315
J
PM
Ref
pH
Sensor
CPK1
Liquiline M CM42 Wiring
Endress+Hauser 21
ISFET sensors
a0002380
Fig. 26: View in device (sensor module)a0001076
Fig. 27: Wiring diagram
PM
BN
317
318
320
316
315
111
113
112
319
GN
WH
YE
112
113
111
320
318
316
BK
BN
RD315
BK
BN
GN
WH
YE
CPK12
317
318
112
113
111
319
CM42
315
316
320
RD
Src
J
Ref
PM
Drn
Sensor
Wiring Liquiline M CM42
22 Endress+Hauser
Without PML (asymmetrical)
Glass electrodes
a0002386
Fig. 28: View in device (sensor module)a0001080
Fig. 29: Wiring diagram
d.n.c.
BN
31
73
18
32
03
16
31
5111
11
311
23
19
BK
GN
WH
YE
112
113
111
317
BN320
BK
BN
GN
WH
YE
318
320
112
113
111
317
CM42
319
315
316
CPK9 / CPK1
J
d.n.c.
Ref
pH
Sensor
pH
CPK1
Liquiline M CM42 Wiring
Endress+Hauser 23
ISFET sensors
5.2 Post-connection check
Device condition and specifications
1. Are the transmitter and cables free from damage on the outside?
Electrical connection
2. Are the mounted cables strain relieved?
3. Are the cables run without loops and cross-overs?
4. Are the signal lines correctly connected in accordance with the wiring diagram?
5. Have you positioned unused connection wires on the "EMC barrier"?
a0002381
Fig. 30: View in device (sensor module)a0001077
Fig. 31: Wiring diagram
BN
317
318
320
316
315
111
113
112
319
GN
WH
YE
112
113
111
320
316BK
RD315
d.n.c.
BN
BK
BN
GN
WH
YE
317
318
112
113
111
319
CM42
315
316
320
RD
Src
Ref
Drn
J
d.n.c.
CPK12 Sensor
Wiring Liquiline M CM42
24 Endress+Hauser
6. Are all the connection wires securely positioned in the cable terminals?
7. Are all the cable entries installed, tightened and sealed?
# Warning!
Only put the transmitter into operation if you can answer yes to all these questions. Otherwise,
safe operation cannot be guaranteed. In such instances, the manufacturer does not accept any
liability for the device.
Liquiline M CM42 Commissioning
Endress+Hauser 25
6 Commissioning
6.1 Function check
# Warning!
• Check that all connections have been established correctly.
• Make sure that the supply voltage matches the voltage indicated on the nameplate!
6.2 Quick Setup
1. Connect the supply voltage.
2. Wait for the initialization to be complete.
3. Completely work through the "Quick Setup" submenu in "PARAM".
6.2.1 Menu structure
6.2.2 Configuration options
PARAM
Quick Setup
Language
TAG number
Measurand
Potential matching
Sensor type (only pH)
Internal buffer entry (only if sensor type = "Glass (internal buffer entry)")
Damping
Temperature sensor
Temperature unit
Buffer manufacturer
Temperature compensation
Current output 1
Output value
Low value (4 mA)
Up value (20 mA)
Function Options Info
Language Options
• English
• Language ordered
Factory setting
Language ordered
! Note!
""Language ordered" refers to the language that you
selected for your device using the order code ("device
language").
If you select the other language in question, all the
other settings remain intact.
TAG number Enter any text
Commissioning Liquiline M CM42
26 Endress+Hauser
Measurand Options
• pH
• ORP mV
• ORP %
• rH
Factory setting
pH
Potential matching Options
• With PM
• Without PM
Factory setting
With PM
Indicate whether you want to measure symmetrically
(= with PM) or asymmetrically (= without PM).
! Note!
More information on symmetrical and asymmetrical
measurement on the CD-ROM.
Sensor type pH Options
• Glass 7.0
• Glass (internal buffer entry)
• ISFET
• Antimony
• Ceramic
Factory setting
Glass 7.0
"Ceramic" stands for electrodes with a
high-impedance reference (= Pfaudler electrodes).
Internal buffer entry –2.0 ... 16.0 pH
Factory setting
7.0
Only available if you have selected "Sensor type" =
"Glass (internal buffer entry)".
Damping 0.0 ... 60.0 s
Factory setting
0 s
Time constant
A value of 5.0 s means that the measured values are
interpolated over the period of 5 seconds.
Temperature sensor Options
• None
• Pt100/Pt1000
• NTC 30K
Factory setting
Pt100/Pt1000
If "Pt100/Pt1000" is selected, the system automati-
cally detects whether a Pt100 or Pt1000 is connected.
Temperature unit Options
• °C
• °F
Factory setting
?C
Buffer manufacturer Options
• E+H (NIST)
• Ingold/Mettler
• DIN 19267
• DIN 19266
• Merck/Riedel
Factory setting
E+H (NIST)
The menu is not available unless the buffer
recognition is set to "Automatic" or "Fixed" in the
PARAM/CAL-settings menu. The factory setting is set
here during initial commissioning.
You can define special buffers in the
PARAM/Cal-settings menu.
Function Options Info
Liquiline M CM42 Commissioning
Endress+Hauser 27
6.3 User administration
To protect the device against unwanted or unauthorized access, you can assign an access code
to the menu.
! Note!
The factory setting for the access code is an empty field (= no entry). The prompt to enter a
password is deactivated. Configuration is possible for everyone.
Assigning/changing access codes:
1. Go to PARAM menu, "General settings"/"User administration".
2. Select the type of protection:
– No password protection (=factory setting).
– Password protection per code entry.
3. If you select "Enter code", select the user role and assign a password to each of the user
roles.
Temp. compensation Options
• Auto. comp. (ATC)
• Man. comp. (MTC)
• Off
Factory setting
Auto. comp.
Temperature entry –20 ... +150 °C
Factory setting
25 °C
Only available if you have selected "Temperature
compensation" = "Manual comp.".
Current output 1
Output value Options
• Main measured value (1)
• rH
• Sensor mV (1)
• Temperature
Factory setting
Main value
Low value (4 mA) –2 ... 11.5 pH or
–700 ... +400 mV
Factory setting
2 pH
Specify the measured value that should correspond to
the 4 mA or 20 mA value.
The smallest possible spread between4 mA and
20 mA value is:
1 for pH, 1 mV for ORP mV
Up value (20 mA) –1.5 ... 16 pH or
–600 ... +700 mV
Factory setting
12 pH
Function Options Info
Commissioning Liquiline M CM42
28 Endress+Hauser
6.3.1 User roles and access authorization
If the password protection has been defined and activated, you have to log in with the
appropriate user role before you can make changes to the settings in the menu.
Here, you will be prompted by the device to enter an access code as soon as you want to
configure a value:
User roles and access authorization (factory setting):
! Note!
If you are logged in as the "Specialist" (=Administrator), you can change the access
authorization of all other user groups:
1. Select a submenu where you want to assign authorization to, or take authorization from,
another user group.
2. Press and hold down the navigator button for at least 3 seconds.
The display shows all the user groups and the current access authorization to the menu
selected.
3. Use the navigator to change the authorization in question and confirm with "OK".
4. Make a note of the changes in your operating logbook or similar.
6.4 Calibration
Calibration is necessary:
• At initial commissioning
• After replacing a sensor
Maintenance team Specialist Operator
• Read access in the PARAM and
DIAG menus
• Read and write access in the
calibration menu
• Read and write access in all menus
• Change the access authorization
for other user groups
• Read access in the PARAM and
DIAG menus
• Read authorization in the
calibration menu
Liquiline M CM42 Commissioning
Endress+Hauser 29
• After periods of standstill
• At sensible, process-dependent intervals1).
6.4.1 Types of calibration
• Automatic buffer recognition (only pH):
– "PARAM"/"Sensor"/"Cal-settings"/"Buffer manufacturer" or
"PARAM"/"Quick Setup"/"Buffer manufacturer"
– You select the buffer type (e.g. DIN 19266).
– The device automatically recognizes the buffer during calibration.
• Fixed buffer (only pH):
– "PARAM"/"Sensor"/"Cal-settings"/"Buffer recognition"
– You select two buffer solutions from the menu.
• Manual:2)
– You enter the buffer values during calibration.
• Numeric input:
– "PARAM"/"Sensor"/"Cal-settings"/"Type of calibration"
– You enter the slope, zero point and temperature (for pH).
– You enter the mV value (for ORP).
6.4.2 Calibrate
Proceed as follows to calibrate the sensor:
1. Go through the menu "Param"/"Quick Setup" completely 3).
2. Remove the sensor from the process.
3. Clean the sensor.
4. Press the soft key for "CAL".
5. Follow the instructions in the menu.
6. Finish calibrating by switching back to the measuring mode.
7. Install the sensor back into the process.
Your measuring point is now ready for operation.
! Note!
• If calibration is aborted using ESC, or if the calibration is faulty, the system continues to use
the original calibration data. A calibration error is shown as plain text on the display.
• During each calibration, the device automatically switches to Hold (factory setting).
• Any offset set is automatically deleted after accepting the calibration.
1) Depending on the process conditions, the intervals can range from several times daily to once quarterly.
2) Not available during initial commissioning if you are working with the factory settings.
3) Only for initial commissioning
Accessories Liquiline M CM42
30 Endress+Hauser
7 Accessories
• Field mounting kit
– 1 Mounting plate
– 2 Securing screws
Order No. 51518286
• Panel installation kit
– 1 Panel installation seal
– 2 Tensioning screws M6x150 mm
– 4 Hexagonal nuts M6, DIN934 A2
– 4 Spring washers, A2 DIN127, form B6
– 4 Washers A6.4, DIN125 A2
Order No. 51518284
! Note!
The CD-ROM supplied provides you with all the relevant information on all the accessories,
including descriptions, graphics and ordering information.
Additional accessories include:
• Sensors
• Assemblies
• Calibration solutions
• Cleaning solutions
• Pressure-reducing valves
• Water filters
• Additional documentation
Endress+Hauser
Liquiline M CM42
Index
AAccess authorization
Change . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
User roles. . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Access code
Factory setting . . . . . . . . . . . . . . . . . . . . . . . 27
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
CCable grounding . . . . . . . . . . . . . . . . . . . . . . . . 17
Cable terminals . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
CE mark. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Check
Electrical connection . . . . . . . . . . . . . . . . . . 23
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Post-installation . . . . . . . . . . . . . . . . . . . . . . 15
Commissioning. . . . . . . . . . . . . . . . . . . . . . . 4, 25
Connecting ground. . . . . . . . . . . . . . . . . . . . . . 17
Connection
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
With PML . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Without PML . . . . . . . . . . . . . . . . . . . . . . . . 22
DDeclaration of conformity . . . . . . . . . . . . . . . . . . 8
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Documentation
Methods of ordering . . . . . . . . . . . . . . . . . . . . 8
EElectrical connection . . . . . . . . . . . . . . . . . . . . 23
Power supply . . . . . . . . . . . . . . . . . . . . . . . . 19
Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Signal output . . . . . . . . . . . . . . . . . . . . . . . . 19
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . 16
Electrical technician . . . . . . . . . . . . . . . . . . . . . 16
FFactory setting
Access code . . . . . . . . . . . . . . . . . . . . . . 27–28
HHousing
Closed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Housing grounding . . . . . . . . . . . . . . . . . . . . . . 17
Housing opening . . . . . . . . . . . . . . . . . . . . . . . 16
IIcons
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Identification
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Serial number. . . . . . . . . . . . . . . . . . . . . . . . . 6
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . 4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 4, 10
Installation instructions
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . 11
Panel mounting . . . . . . . . . . . . . . . . . . . . . . 14
Wall or field mounting . . . . . . . . . . . . . . . . . 12
MMeasuring system. . . . . . . . . . . . . . . . . . . . . . . 10
Mounting plate. . . . . . . . . . . . . . . . . . . . . . . . . 12
NNameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
OOpening the housing . . . . . . . . . . . . . . . . . . . . 16
Operation
Safety instructions . . . . . . . . . . . . . . . . . . . . . 4
Operational safety. . . . . . . . . . . . . . . . . . . . . . . . 5
PPost-installation
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Product structure . . . . . . . . . . . . . . . . . . . . . . . . 7
QQuick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
RReturn . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Liquiline M CM42
32
SSafety conventions and icons. . . . . . . . . . . . . . . . 5
Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . 8
Sensor
Connection with PML . . . . . . . . . . . . . . . . . 20
Connection without PML . . . . . . . . . . . . . . . 22
Electrical connection. . . . . . . . . . . . . . . . . . . 19
Serial number . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Signal output . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Symbols
Document . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
TTransport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Types of calibration. . . . . . . . . . . . . . . . . . . . . . 29
UUse. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
User administration. . . . . . . . . . . . . . . . . . . . . . 27
User groups . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
User roles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
WWarnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Liquiline M CM42
Endress+Hauser
Index Liquiline M CM42
34 Endress+Hauser
Liquiline M CM42 Index
Endress+Hauser 35
www.endress.com/worldwide
BA381C/07/en/POD
Printed in Germany / FM+SGML 6.0