SAR Measurements with cSAR3D
APPLICATION NOTESystem Check or Veri�cation
1. INTRODUCTION Application Note
1 Introduction
These instructions describe the system veri�cation (i.e., system check) procedure for cSAR3D. This procedure
conforms to IEC draft 62209-3 and IEC PAS 63151. System veri�cation provides a fast and reliable method to verify
that the SAR system is operational with no system component failure, drifts or deviation from target performance
requirements. A system veri�cation also veri�es the measurement repeatability of the SAR measurement system
against calibrated system check antenna(s) before compliance testing.
2 Objective
System veri�cation should be performed on a regular basis, especially:
upon receiving a new or recalibrated cSAR3D,
after the cSAR3D has been moved to a new location,
prior to starting a set of DUT measurements in conformance with IEC draft 62209-3 or IEC PAS 63151 and
within seven days of any measurement of the DUT,
The system veri�cation protocol is described in the cSAR3D System Handbook and conforms to IEC draft
62209-3 and IEC PAS 63151. The measured SAR values are compared against numerical targets. They are
also compared against measured reference values available in the cSAR3D calibration certi�cate. To make the
comparison easier, an Excel �le is attached at the end of this Application Note. The system veri�cation is
successful if:
the measured values are within two standard deviations of the system standard uncertainty us of the absolute
(numerical) targets, and
the measured values are repeatable within �10% of the measured reference values from the cSAR3D cali-
bration certi�cate.
3 Equipment Required
The system veri�cation test requires the use of cSAR3D devices, cSAR3D Veri�cation Dipoles (and their acces-
sories) and an accurate power reference. Use each of the cSAR3D devices (Flat, Head, Quad) that you own.
3.1 cSAR3D Veri�cation Dipoles
cSAR3D Veri�cation Dipoles from SPEAG should have a valid calibration, including a calibration certi�cate. The
dipole antennas are standardized dipole antennas whose electrical and mechanical speci�cations are described in
IEC PAS 63151, which are identical to the dipoles described in IEC draft 62209-3, IEEE 1528-2013, and IEC
62209-1 2016 (see Figure 1.1). The separation distances of the dipole antennas to the �at phantom (s = 15 mm
for f < 1 GHz, and s = 10 mm for f > 1 GHz) are also as de�ned in these standards.
The measured reference values in the cSAR3D calibration certi�cate will be used for the comparison. The Set
of cSAR3D Veri�cation Dipoles and the Set of Veri�cation Accessories include the following components. Dipoles
are also available at other frequencies.
Set of cSAR3D Veri�cation Dipoles:
1 D835V2 - cSAR3D Veri�cation Dipole, including calibration certi�cate
1 D1950V3 - cSAR3D Veri�cation Dipole, including calibration certi�cate
1 D2450V2 - cSAR3D Veri�cation Dipole, including calibration certi�cate
1 D5GHzV2 - cSAR3D Veri�cation Dipole, including calibration certi�cate
cSAR3D Application Note: System Check or Veri�cation, December 2017 i
3. EQUIPMENT REQUIRED Application Note
Figure 1.1: Mechanical details of the standard dipole. L = length of the dipole, d1 = diameter of the dipole, d2 =
Diameter of the coaxial cable at the balun, h = length of the balun choke section.
1 cSAR3D Veri�cation Dipole Spacer, 15 mm (for frequencies below 1 GHz)
3 cSAR3D Veri�cation Dipole Spacers, 10 mm (for frequencies above 1 GHz)
Set of Veri�cation Accessories:
1 VMLV1 - Veri�cation Mask for cSAR3D Left Head V1
1 VMRV1 - Veri�cation Mask for cSAR3D Right Head V1
1 VAHV1 - Veri�cation Antenna Holder for cSAR3D Heads V1
1 VDSV1 - Veri�cation Dipole Stand for cSAR3D Flat/Quad V1
cSAR3D Application Note: System Check or Veri�cation, December 2017 ii
3. EQUIPMENT REQUIRED Application Note
Figure 1.2: cSAR3D Veri�cation Kit.
cSAR3D Application Note: System Check or Veri�cation, December 2017 iii
3. EQUIPMENT REQUIRED Application Note
Figure 1.3: VAHV1 - Veri�cation Antenna Holder for cSAR3D Heads V1.
cSAR3D Application Note: System Check or Veri�cation, December 2017 iv
3. EQUIPMENT REQUIRED Application Note
3.2 Power Source
The tests require an accurate power source. The POWERSOURCE1 from SPEAG is suitable for this test. If
POWERSOURCE1 is used, follow the procedure in Application Note �POWERSOURCE1� for the setup (see also
Figure 1.4).
POWERSOURCE1 kit contains:
1 POWERSOURCE1 V1 - Broadband USB-fed source V1
1 USB Cable
1 Power Bank Battery Module including cable (optional)
Figure 1.4: Equipment setup for system check and system validation using POWERSOURCE1. The ampli�er,
�lter and directional coupler are not required when using the POWERSOURCE1. If traceability is required for
submission and the unit is not ISO/IEC 17025 calibrated, measure the output power before and after the SAR
measurement using a calibrated power meter in order to achieve the required traceability.
If POWERSOURCE1 is not available, Figure 1.5 shows the recommended setup. This setup consists of a generator,
a bi-directional coupler, attenuators and power meters. The setup is also described in international SAR measure-
ment standards such as IEEE 1528, IEC 62209-1, IEC 62209-2, IEC draft 62209-3, IEC PAS 63151 and various
national SAR measurement regulations. The requirements for this setup are described in the section �Veri�cation
Setup� in the cSAR3D System Handbook.
Figure 1.5: Equipment setup for system check and system validation according to IEC 62209-1 and other standards.
cSAR3D Application Note: System Check or Veri�cation, December 2017 v
4. VERIFICATION SOURCE LOCATIONS Application Note
4 Veri�cation Source Locations
For the �at phantom, the nominal location of the source is the center of the phantom (point A0). However, results
within �5 % are expected at any other location and rotation within the measurement �eld. For the head phantom,
all four positions will be used. Figure 1.6 shows the locations on the Head phantoms at which the dipoles will be
positioned.
Figure 1.6: Using the veri�cation mask on the cSAR3D Head. The veri�cation mask (red) indicates four locations:
Top, Mouth, Ear and Neck.
cSAR3D Application Note: System Check or Veri�cation, December 2017 vi
5. PROCEDURE Application Note
5 Procedure
The following procedure should be used for each cSAR3D device.
5.1 Preparation
Open the attached Excel �le and �ll in the data (yellow cells) as described in the following steps.
1. Go to the tab for the cSAR3D device type: Flat, Head or Quad
2. Ensure that the ambient temperature in the laboratory is within 18� to 25� and has stabilized before starting
the tests.
3. Turn on the RF source (either POWERSOURCE1 or other signal generator) and allow it to warm up.
4. Fill in the top section with the date, laboratory, operator and ambient temperature.
5. Fill in the serial number of the cSAR3D device.
6. Fill in the serial number of each cSAR3D Validation Dipole.
7. Fill in the reference SAR from the cSAR3D calibration certi�cate for each dipole frequency.
5.2 Self Test of cSAR3D sensors
The self test is an automated pass / fail test to verify that all sensors are performing as expected. In the cSAR3D
software, apply the following steps:
1. On the status bar, check that the cSAR3D device is connected.
2. Open the cSAR3D Hardware window by clicking on File / Preferences / cSAR3D Hardware (see Figure 1.7).
3. Ensure that there are no radiofrequency sources near the device.
4. Click on the `self test' button for the cSAR3D device and follow the instructions (see Figure 1.7).
5. If the self test passes (see Figure 1.8), close the window and continue with this procedure. If the self test
fails, follow the instructions given in Figure 1.9.
6. Record whether the self test passed in the Excel worksheet.
cSAR3D Application Note: System Check or Veri�cation, December 2017 vii
Flat
cSAR3D Verification Test - Flat Phantom
This spreadsheet is to be used for the System Verification Test described in the cSAR3D System Handbook
Please follow the Application Note "System Check or Verification" and fill the data in the yellow cells.
Date of Tests
Laboratory
Operator
Ambient Temp.
Notes
Location (Orientation)Antenna location and orientation as described in the cSAR3D System Handbook
Total Uncertainty (k = 2)The cSAR3D measurement uncertainty from the cSAR3D System Handbook.
SAR from cSAR3D CertificateEnter the Measured SAR in the System Validation Table (columns 6 and 7) of the cSAR3D calibration certificate.
ToleranceThe allowable repeatability tolerance, as stated in the IEC 62209-3 draft.
Test PassedCheck whether the deviation from the target is within the total uncertainty. 95% of the measured values should be within the expanded (k = 2) uncertainty and 100% should be within the k = 3 uncertainty.0.4139268516
cSAR3D TypeFlat MSL
Serial Number
Passed Self Test?Verification Against Numerical TargetsVerification Against Measured Reference Values
DipoleTypeDipoleSerial NumberfrequencyLocation (Orientation)PowerMeasured SARMeasured SARNormalized to 1 WNumerical TargetDeviation from TargetTotal Uncertainty(k = 2)Verified Against Numerical Targets?SAR from cSAR3D CertificateDeviation from Cal. Cert.ToleranceVerified Against Measured Values?
(MHz)(dBm)(W/kg)(W/kg)(W/kg)(dB)(%)(dB)(W/kg)(dB)(%)(dB)
1g10g1g10g1g10g1g10g1g10g1g10g1g10g1g10g
D835V2835A0 (0°)0.000.009.716.3825.4%1.010%0.4
D1950V31950A0 (0°)0.000.0040.321.025.4%1.010%0.4
D2450V22450A0 (0°)0.000.0051.223.725.4%1.010%0.4
D5GHzV25200A0 (0°)0.000.0071.820.128.9%1.110%0.4
D5GHzV25800A0 (0°)0.000.0074.120.528.9%1.110%0.4
cSAR3D TypeFlat HSL
Serial Number
Passed Self Test?Verification Against Absolute TargetsVerification Against Measured Reference Values
DipoleTypeDipoleSerial NumberfrequencyLocation (Orientation)PowerMeasured SARMeasured SARNormalized to 1 WNumerical TargetDeviation from TargetTotal Uncertainty(k = 2)Verified Against Numerical Targets?SAR from cSAR3D CertificateDeviation from Cal. Cert.ToleranceVerified Against Measured Values?
(MHz)(dBm)(W/kg)(W/kg)(W/kg)(dB)(%)(dB)(W/kg)(dB)(%)(dB)
1g10g1g10g1g10g1g10g1g10g1g10g1g10g1g10g
D835V2835A0 (0°)0.000.009.566.2225.4%1.010%0.4
D1950V31950A0 (0°)0.000.0040.520.925.4%1.010%0.4
D2450V22450A0 (0°)0.000.0052.424.025.4%1.010%0.4
D5GHzV25200A0 (0°)0.000.0076.521.628.9%1.110%0.4
D5GHzV25800A0 (0°)0.000.0078.021.928.9%1.110%0.4
`
Head
cSAR3D Verification Test - Head Phantom
This spreadsheet is to be used for the System Verification Test described in the cSAR3D System Handbook
Please follow the Application Note "System Check or Verification" and fill the data in the yellow cells.
Date of Tests
Laboratory
Operator
Ambient Temp.
Notes
Location (Orientation)Antenna location and orientation as described in the cSAR3D System Handbook
Total Uncertainty (k = 2)The cSAR3D measurement uncertainty from the cSAR3D System Handbook.
SAR from cSAR3D CertificateEnter the Measured SAR in the System Validation Table (columns 6 and 7) of the cSAR3D calibration certificate.
ToleranceThe allowable repeatability tolerance, as stated in the IEC 62209-3 draft.
Test PassedCheck whether the deviation from the target is within the total uncertainty. 95% of the measured values should be within the expanded (k = 2) uncertainty and 100% should be within the k = 3 uncertainty.0.4139268516
cSAR3D TypeHead Left
Serial Number
Passed Self Test?Verification Against Numerical TargetsVerification Against Measured Reference Values
DipoleTypeDipoleSerial NumberfrequencyLocationPowerMeasured SARMeasured SARNormalized to 1 WNumerical TargetDeviation from TargetTotal Uncertainty(k = 2)Verified Against Numerical Targets?SAR from cSAR3D CertificateDeviation from Cal. Cert.ToleranceVerified Against Measured Values?
(MHz)(dBm)(W/kg)(W/kg)(W/kg)(dB)(%)(dB)(W/kg)(dB)(%)(dB)
1g10g1g10g1g10g1g10g1g10g1g10g1g10g1g10g
D835V2835Top, C(0)0.000.009.776.4831.5%1.210%0.4
Mouth, F(90)0.000.009.806.4931.5%1.210%0.4
Ear, D(90)0.000.007.805.2831.5%1.210%0.4
Neck, H(0)0.000.009.646.5331.5%1.210%0.4
D1950V31950Top, C(0)0.000.0042.621.731.5%1.210%0.4
Mouth, F(90)0.000.0042.221.531.5%1.210%0.4
Ear, D(90)0.000.0028.515.631.5%1.210%0.4
Neck, H(0)0.000.0041.321.131.5%1.210%0.4
D2450V22450Top, C(0)0.000.0053.024.031.5%1.210%0.4
Mouth, F(90)0.000.0054.824.931.5%1.210%0.4
Ear, D(90)0.000.0032.216.031.5%1.210%0.4
Neck, H(0)0.000.0051.024.031.5%1.210%0.4
D5GHzV25200Top, C(0)0.000.0083.921.634.9%1.310%0.4
Mouth, F(90)0.000.0083.521.934.9%1.310%0.4
Ear, D(90)0.000.0055.718.534.9%1.310%0.4
Neck, H(0)0.000.0082.720.934.9%1.310%0.4
D5GHzV25800Top, C(0)0.000.0080.019.934.9%1.310%0.4
Mouth, F(90)0.000.0083.220.834.9%1.310%0.4
Ear, D(90)0.000.0064.421.234.9%1.310%0.4
Neck, H(0)0.000.0079.119.934.9%1.310%0.4
cSAR3D TypeHead Right
Serial Number
Passed Self Test?Verification Against Absolute TargetsVerification Against Measured Reference Values
DipoleTypeDipoleSerial NumberfrequencyLocationPowerMeasured SARMeasured SARNormalized to 1 WNumerical TargetDeviation from TargetTotal Uncertainty(k = 2)Verified Against Numerical Targets?SAR from cSAR3D CertificateDeviation from Cal. Cert.ToleranceVerified Against Measured Values?
(MHz)(dBm)(W/kg)(W/kg)(W/kg)(dB)(%)(dB)(W/kg)(dB)(%)(dB)
1g10g1g10g1g10g1g10g1g10g1g10g1g10g1g10g
D835V2835Top, C(0)0.000.009.776.4831.5%1.210%0.4
Mouth, F(90)0.000.009.806.4931.5%1.210%0.4
Ear, D(90)0.000.007.805.2831.5%1.210%0.4
Neck, H(0)0.000.009.646.5331.5%1.210%0.4
D1950V31950Top, C(0)0.000.0042.621.731.5%1.210%0.4
Mouth, F(90)0.000.0042.221.531.5%1.210%0.4
Ear, D(90)0.000.0028.515.631.5%1.210%0.4
Neck, H(0)0.000.0041.321.131.5%1.210%0.4
D2450V22450Top, C(0)0.000.0053.024.031.5%1.210%0.4
Mouth, F(90)0.000.0054.824.931.5%1.210%0.4
Ear, D(90)0.000.0032.216.031.5%1.210%0.4
Neck, H(0)0.000.0051.024.031.5%1.210%0.4
D5GHzV25200Top, C(0)0.000.0083.921.634.9%1.310%0.4
Mouth, F(90)0.000.0083.521.934.9%1.310%0.4
Ear, D(90)0.000.0055.718.534.9%1.310%0.4
Neck, H(0)0.000.0082.720.934.9%1.310%0.4
D5GHzV25800Top, C(0)0.000.0080.019.934.9%1.310%0.4
Mouth, F(90)0.000.0083.220.834.9%1.310%0.4
Ear, D(90)0.000.0064.421.234.9%1.310%0.4
Neck, H(0)0.000.0079.119.934.9%1.310%0.4
Quad
cSAR3D Verification Test - Quad Phantom
This spreadsheet is to be used for the System Verification Test described in the cSAR3D System Handbook
Please follow the Application Note "System Check or Verification" and fill the data in the yellow cells.
Date of Tests
Laboratory
Operator
Ambient Temp.
Notes
Location (Orientation)Antenna location and orientation as described in the cSAR3D System Handbook
Total Uncertainty (k = 2)The cSAR3D measurement uncertainty from the cSAR3D System Handbook. The tolerance in dB takes into consideration the verification dipole uncertainty.
SAR from cSAR3D CertificateEnter the Measured SAR in the System Validation Table (columns 6 and 7) of the cSAR3D calibration certificate.
ToleranceThe allowable repeatability tolerance, as stated in the IEC 62209-3 draft.
Test PassedCheck whether the deviation from the target is within the total uncertainty. 95% of the measured values should be within the expanded (k = 2) uncertainty and 100% should be within the k = 3 uncertainty.0.4139268516
cSAR3D TypeQuad MSL Area
Serial Number
Passed Self Test?Verification Against Numerical TargetsVerification Against Measured Reference Values
DipoleTypeDipoleSerial NumberfrequencyLocationPowerMeasured SARMeasured SARNormalized to 1 WNumerical TargetDeviation from TargetTotal Uncertainty(k = 2)Verified Against Numerical Targets?SAR from cSAR3D CertificateDeviation from Cal. Cert.ToleranceVerified Against Measured Values?
(MHz)(dBm)(W/kg)(W/kg)(W/kg)(dB)(%)(dB)(W/kg)(dB)(%)(dB)
1g10g1g10g1g10g1g10g1g10g1g10g1g10g1g10g
D835V2835center0.000.009.716.3824.3%0.910%0.4
D1950V31950center0.000.0040.321.024.3%0.910%0.4
D2450V22450center0.000.0051.223.724.3%0.910%0.4
D5GHzV25200center0.000.0071.820.128.8%1.110%0.4
D5GHzV25800center0.000.0074.120.528.8%1.110%0.4
cSAR3D TypeQuad HSL Area
Serial Number
Passed Self Test?Verification Against Absolute TargetsVerification Against Measured Reference Values
DipoleTypeDipoleSerial NumberfrequencyLocation (Orientation)PowerMeasured SARMeasured SARNormalized to 1 WNumerical TargetDeviation from TargetTotal Uncertainty(k = 2)Verified Against Numerical Targets?SAR from cSAR3D CertificateDeviation from Cal. Cert.ToleranceVerified Against Measured Values?
(MHz)(dBm)(W/kg)(W/kg)(W/kg)(dB)(%)(dB)(W/kg)(dB)(%)(dB)
1g10g1g10g1g10g1g10g1g10g1g10g1g10g1g10g
D835V2835center0.000.009.566.2224.3%0.910%0.4
D1950V31950center0.000.0040.520.924.3%0.910%0.4
D2450V22450center0.000.0052.424.024.3%0.910%0.4
D5GHzV25200center0.000.0076.521.628.8%1.110%0.4
D5GHzV25800center0.000.0078.021.928.8%1.110%0.4
`
5. PROCEDURE Application Note
Figure 1.7: Running the self test procedure.
Figure 1.8: Self test passed.
cSAR3D Application Note: System Check or Veri�cation, December 2017 viii
5. PROCEDURE Application Note
Figure 1.9: If self test failed, follow the instructions on the screen.
cSAR3D Application Note: System Check or Veri�cation, December 2017 ix
5. PROCEDURE Application Note
5.3 Veri�cation Testing
This section describes the procedure for accurately measuring the SAR using the dipole antenna. The detailed
setup for setting the power is described in other documents. If the POWERSOURCE1 is used as the RF source,
follow Application Note �POWERSOURCE1�. Otherwise, follow section �Veri�cation Measurement Protocol� of
the cSAR3D System Handbook.
Figure 1.10: Prede�ned setup for veri�cation measurements.
1. Select the prede�ned setup called �Dipole Veri�cation, 5 frequencies� (see Figure 1.10). Add or remove
phantoms and frequency bands to the Lilies wheel as desired.
For each dipole antenna, apply the following steps.
2. Set the frequency of the RF source to match the operating frequency of the dipole antenna.
3. Set the power level at 15 dBm or more. Measure the output power using a calibrated power meter (connection
1 of either Figure 1.4 or 1.5, depending on the source type). If using a calibrated POWERSOURCE1, use
the output power indicated on the screen, as power measurement is not needed.
4. Enter the power level in the Excel worksheet.
For each dipole antenna and antenna position, apply the following steps.
5. Connect the dipole antenna at the end of the RF source (see Figure 1.4 or connection 2 of 1.5).
6. Connect the dipole antenna to the appropriate low-loss spacer (see Figure 1.11). The spacer maintains a
�xed distance from the axis of the dipole antenna to the tissue simulating medium (including the 2 mm thick
phantom shell). The total spacing is:
15 mm for frequencies up to 1 GHz
10 mm for frequencies above 1 GHz
cSAR3D Application Note: System Check or Veri�cation, December 2017 x
5. PROCEDURE Application Note
Figure 1.11: Low-loss spacer connected at the dipole feed point.
7. Place the dipole antenna at a location on the cSAR3D phantom. For Flat and Quad phantoms, use the
VDSV1 Veri�cation Dipole Stand for cSAR3D Flat/Quad V1 to hold the dipole antenna at the A0 position.
For Head phantoms use VAHV1 Veri�cation Antenna Holder for cSAR3D Heads V1. If not available, hold
the dipole by hand at one of the four locations of the Head Veri�cation mask.
8. Ensure that the dipole arms are parallel to the phantom surface and oriented at the correct angle. For the
Head phantoms, the red Veri�cation Mask has raised sections to assist in checking that the dipole arms are
parallel and in the correct orientation (see Figure 1.13).
9. Measure the peak spatial-average SAR values at the phantom: SAR1g and SAR10g.
10. Enter the measured SAR values in the Excel worksheet.
If the above test is outside the tolerances for any of the tested con�gurations, check the setup thoroughly and
repeat all above steps for this con�guration. It is recommended to repeat at least 10 times in order to preclude
measurement outliers. Record the average from the 10+ values in the Excel worksheet.
cSAR3D Application Note: System Check or Veri�cation, December 2017 xi
5. PROCEDURE Application Note
Figure 1.12: Veri�cation dipole mounted on the dipole stand.
cSAR3D Application Note: System Check or Veri�cation, December 2017 xii
5. PROCEDURE Application Note
Figure 1.13: Positioning the veri�cation dipole at the Mouth location on the cSAR3D phantom. The dipole arms
shall be aligned parallel with the markings on the mask.
cSAR3D Application Note: System Check or Veri�cation, December 2017 xiii
Application NoteIntroductionObjectiveEquipment RequiredcSAR3D Verification DipolesPower Source
Verification Source LocationsProcedurePreparationSelf Test of cSAR3D sensorsVerification Testing
Recommended