Abstract—We describe a finite element (FE) model for simulating wheel-rolling tissue deformations using a rolling finite element model (RFEM). A wheeled probe performing rolling tissue indentation has proven to be a promising approach for compensating for the loss of haptic and tactile feedback experienced during robotic-assisted minimally invasive surgery (MIS) [1-6]. A sound understanding of wheel- tissue rolling interaction dynamics will facilitate the evaluation of signals from rolling indentation. In this paper we model the dynamic interactions between a wheeled probe and a soft tissue sample using the ABAQUS finite element analysis software package. The aim of this work is to more precisely locate abnormalities within soft tissue organs using RFEM and hence aid surgeons improving diagnostic ability. The soft tissue is modeled as a nonlinear hyperelastic material with geometrical nonlinearity. The proposed RFEM was validated on a silicone phantom and a porcine kidney sample. The results show that the proposed method can predict the wheel-tissue interaction forces of rolling indentation with good accuracy and can also identify the location and depth of simulated tumors accurately. Index Terms–-Soft tissue abnormality localization, Minimally Invasive Surgery, Haptic Feedback, Finite Element Modelling, Rolling Indentation, Tumour detection, Soft tissue organ diagnosis I. INTRODUCTION INIMALLY Invasive Surgery (MIS) is performed through small incisions, resulting in reduced trauma, recovery times and hospitalization costs [14]. However, the loss of tactile feedback is a significant drawback with MIS, preventing the surgeon from directly palpating specific tissue regions [10-13]. This sensory reduction is only partially compensated for by employing highly magnified stereo vision during MIS. A wheeled probe for the rapid identification of mechanical properties of soft tissue during MIS was proposed in [1-6]. A sound understanding of wheel-tissue rolling interaction dynamics Manuscript received May 28, 2010. K. Sangpradit is with the Department of Mechanical Engineering, Kings College London, UK (e-mail: kiattisak.sangpradit @kcl.ac.uk). H. Liu is with the Department of Mechanical Engineering, Kings College London, UK (e-mail: [email protected]). P. Dasgupta is with the Department of Urology, King’s College London, Guy’s Hospital, London, UK ([email protected]) K. Alhoefer is with the Department of Mechanical Engineering, Kings College London, UK (e-mail: [email protected]). L. D. Seneviratne is with the Mechanical Engineering Department, King’s College London, Strand, London, WC2R 2LS, UK and Faculty of Engineering, Khalifa University of Science, Technology & Research, Abu Dhabi, UAE (e-mail: [email protected]). will facilitate the evaluation of signals from rolling indentation. This paper presents a Rolling Finite Element Model (RFEM) to simulate soft tissue rolling indentation, in order to predict soft tissue behavior and identify abnormal tissue regions, such as tumors buried within a soft tissue organ. II. BACKGROUND Biological soft tissue is highly nonlinear, viscoelasic, anisotropic, heterogeneous, and nearly incompressible with a porous internal structure. Their mechanical properties also depend on environmental factors such as pH value, temperature and tissue health [7]. The stress-strain relationship of soft tissue is time and strain rate dependent [8, 9, 15, 17, 19] and there is a significant variation in mechanical characteristics of tissues of different pathological states [14, 17, 18, 19]. Hence, employing a mechanical probe to estimate tissue properties provides a potentially useful means of identifying malignant tissue. Many researchers have used FE techniques to simulate soft tissue behavior and for soft tissue parameter identification [20-23]. Tillier et al. applied hyper-elastic constitutive equations, using ANSYS FE software, to identify tissue properties from experimental measurements [20]. The method was applied to lamb kidney and the human uterus, showing good agreement with experimental data. Liu et al. [21] applied a nonlinear hyper-elastic, 8-chain network (Arruda-Boyce model) constitutive equation to model soft tissue undergoing large indentations by using ABAQUS software, and identified two material parameters, the initial modulus and locking stretch. Kerdok et al. used the Arruda- Boyce model with ABAQUS to identify the properties of breast tissue [22]. Zhang et al [23] used a FE model to simulate large tissue deformations. Tonuk et al [24] used FE analysis to simulate indentation behaviour of residual limb tissue. Picinbono et al [25] proposed a FE model for soft tissue deformation based on nonlinear elasticity and anisotropic behavior. Zhong et al [26] presented a methodology for predicting the deformations of soft objects, based on the analogy between heat conduction and elastic deformations. Szekely et al [27] developed a full-scale FE simulation of elastic tissue deformation in complex systems such as the human abdomen. Schwartz et al [28] extended the linear elastic tensor-mass method to simulate biological soft tissue, for planning surgical treatment of liver cancer. Duysak et al [29] used a spring-mass system to simulate facial soft tissue deformation during lower jaw bone realignment. Finite Element Modelling of Soft Tissue Rolling Indentation Kiattisak Sangpradit, Hongbin Liu, Prokar Dasgupta, Kaspar Althoefer and Lakmal D. Seneviratne M

Finite-Element Modeling of Soft Tissue Rolling Indentation

Embed Size (px)

Citation preview

Abstract—We describe a finite element (FE) model for simulating wheel-rolling tissue deformations using a rolling finite element model (RFEM). A wheeled probe performing rolling tissue indentation has proven to be a promising approach for compensating for the loss of haptic and tactile feedback experienced during robotic-assisted minimally invasive surgery (MIS) [1-6]. A sound understanding of wheel-tissue rolling interaction dynamics will facilitate the evaluation of signals from rolling indentation. In this paper we model the dynamic interactions between a wheeled probe and a soft tissue sample using the ABAQUS finite element analysis software package. The aim of this work is to more precisely locate abnormalities within soft tissue organs using RFEM and hence aid surgeons improving diagnostic ability. The soft tissue is modeled as a nonlinear hyperelastic material with geometrical nonlinearity. The proposed RFEM was validated on a silicone phantom and a porcine kidney sample. The results show that the proposed method can predict the wheel-tissue interaction forces of rolling indentation with good accuracy and can also identify the location and depth of simulated tumors accurately.

Index Terms–-Soft tissue abnormality localization, Minimally

Invasive Surgery, Haptic Feedback, Finite Element Modelling, Rolling Indentation, Tumour detection, Soft tissue organ diagnosis

I. INTRODUCTION

INIMALLY Invasive Surgery (MIS) is performed through small incisions, resulting in reduced trauma, recovery times and hospitalization costs [14].

However, the loss of tactile feedback is a significant drawback with MIS, preventing the surgeon from directly palpating specific tissue regions [10-13]. This sensory reduction is only partially compensated for by employing highly magnified stereo vision during MIS. A wheeled probe for the rapid identification of mechanical properties of soft tissue during MIS was proposed in [1-6]. A sound understanding of wheel-tissue rolling interaction dynamics

Manuscript received May 28, 2010. K. Sangpradit is with the Department of Mechanical Engineering, Kings

College London, UK (e-mail: kiattisak.sangpradit @kcl.ac.uk). H. Liu is with the Department of Mechanical Engineering, Kings College

London, UK (e-mail: [email protected]). P. Dasgupta is with the Department of Urology, King’s College London,

Guy’s Hospital, London, UK ([email protected]) K. Alhoefer is with the Department of Mechanical Engineering, Kings

College London, UK (e-mail: [email protected]). L. D. Seneviratne is with the Mechanical Engineering Department,

King’s College London, Strand, London, WC2R 2LS, UK and Faculty of Engineering, Khalifa University of Science, Technology & Research, Abu Dhabi, UAE (e-mail: [email protected]).

will facilitate the evaluation of signals from rolling indentation. This paper presents a Rolling Finite Element Model (RFEM) to simulate soft tissue rolling indentation, in order to predict soft tissue behavior and identify abnormal tissue regions, such as tumors buried within a soft tissue organ.

II. BACKGROUND Biological soft tissue is highly nonlinear, viscoelasic,

anisotropic, heterogeneous, and nearly incompressible with a porous internal structure. Their mechanical properties also depend on environmental factors such as pH value, temperature and tissue health [7]. The stress-strain relationship of soft tissue is time and strain rate dependent [8, 9, 15, 17, 19] and there is a significant variation in mechanical characteristics of tissues of different pathological states [14, 17, 18, 19]. Hence, employing a mechanical probe to estimate tissue properties provides a potentially useful means of identifying malignant tissue. Many researchers have used FE techniques to simulate soft tissue behavior and for soft tissue parameter identification [20-23]. Tillier et al. applied hyper-elastic constitutive equations, using ANSYS FE software, to identify tissue properties from experimental measurements [20]. The method was applied to lamb kidney and the human uterus, showing good agreement with experimental data. Liu et al. [21] applied a nonlinear hyper-elastic, 8-chain network (Arruda-Boyce model) constitutive equation to model soft tissue undergoing large indentations by using ABAQUS software, and identified two material parameters, the initial modulus and locking stretch. Kerdok et al. used the Arruda-Boyce model with ABAQUS to identify the properties of breast tissue [22]. Zhang et al [23] used a FE model to simulate large tissue deformations. Tonuk et al [24] used FE analysis to simulate indentation behaviour of residual limb tissue. Picinbono et al [25] proposed a FE model for soft tissue deformation based on nonlinear elasticity and anisotropic behavior. Zhong et al [26] presented a methodology for predicting the deformations of soft objects, based on the analogy between heat conduction and elastic deformations. Szekely et al [27] developed a full-scale FE simulation of elastic tissue deformation in complex systems such as the human abdomen. Schwartz et al [28] extended the linear elastic tensor-mass method to simulate biological soft tissue, for planning surgical treatment of liver cancer. Duysak et al [29] used a spring-mass system to simulate facial soft tissue deformation during lower jaw bone realignment.

Finite Element Modelling of Soft Tissue Rolling Indentation

Kiattisak Sangpradit, Hongbin Liu, Prokar Dasgupta, Kaspar Althoefer and Lakmal D. Seneviratne

M

Rolling indentation gives continuous measurements of the tissue mechanical response, while covering a large tissue area in a relatively short time [1, 3, 5]. However, a systematic methodology is required to accurately interpret signals from rolling indentation. This paper presents a rolling finite element model (RFEM) for soft tissue, to predict tool - tissue interaction forces during rolling indentation, and hence identify and locate tissue abnormalities.

III. EXPERIMENTAL ESTIMATIONS OF TISSUE PARAMETERS

In this paper the hyperelastic, Arruda-Boyce equations are used to model tissue behavior and the material properties are assumed to be non-linear, incompressible and isotropic. Hyperelastic materials can be described using a strain energy function, U (ε), which is defined as the strain energy density at a given location in the material. Equation (1) gives the Arruda-Boyce strain energy function [30]:

( )+−= ∑+

−ii

ii

m

i ICU 31

5

122λ

μ

( )⎥⎦

⎤⎢⎣

⎡−

−el

el JjD

ln2

11 2 (1)

With

,1050

11,201,

21

321 === CCC

673750519,

700019

54 == CC (2)

where: U = The strain energy density μ = The initial shear modulus λm = The locking stretch Jel = The elastic volume ratio Ci = Material constants defined in equation (2) D = Is a measure of compressibility and is zero for fully incompressible materials I1 = The first stain invariant of the deviatoric strain, given in terms of the principal stretches λ1, λ2 and λ3:

23

22

211 λλλ ++=I (3)

The material stress is obtained by differentiating equation

(1) with respect to stretch, λu : nu

U σλ

=∂∂

, where;

σn = The normal stress. λu = The stretch in the loading direction, under uniaxial loading, and

λ1 = λu and λ2= λ3 =

uλ1 (4)

The true stress is given by:

un λσσ ..0 = (5)

Differentiating equation (1) gives:

( ) ( ) ( )⎢⎣⎡ +−+−+−= −−− 36

424

212

0 17511

51

uum

uum

uu λλλ

λλλ

λλμσ

( ) ( )⎥⎦⎤−+− −− 510

848

6 67375519

87519

uum

uum

λλλ

λλλ

(6)

The locking stretch, λm, can be obtained from the limiting chain stretch, λlim, which is the stretch at which the stress starts to increase without limit, the value of, λm, can be obtained from [31]:

⎥⎦

⎤⎢⎣

⎡+=

lim

2lim

231

λλλm

(7)

A uniaxial compression test is conducted on rubber and

silicone material and porcine kidney, using an INSTRON 5543 machine, Fig.1, to experimentally estimate the material properties. A silicone cube of 15x15x15 mm3, a rubber cylinder (hard nodule) of 10 mm diameter and 10 mm length and a kidney cube of 15x15x15 mm3, was compressed by 6 mm at a speed of 1 mm/sec.

Fig.1 Uniaxial compression tests using the INSTRON 5543 machine, (a) Silicone tissue phantom (b) Rubber used to simulate tumours (c) Porcine kidney.

The Arruda-Boyce model requires two parameters, μ and λm, to describe nonlinear, hyperelastic material behavior. These parameters for rubber, silicone and porcine kidney are estimated by comparing the uniaxial compression test results, Fig 1(a), 1(b), and 1(c), with the Arruda-Boyce strain energy function, using Matlab, and the results are lists in Table I.

TABLE I

PROPERTIES OF THE TEST MATERIALS

μ ,shear Modulus

(kPa)

λm,

Locking stretch

Mass Density (kg/m3)

Rubber 73.40 1.01 1000

Silicone (RTV6166 gel) 4.98 1.05 980

Porcine kidney 1.58 1.05 850

IV. ROLLING INDENTATION EXPERIMENTS

A. Silicone Phantom Tests The rolling indentation tests are performed using a

wheeled probe attached to the distal tip of a Mitsubishi RV-6SL 6-DOF robot manipulator. An ATI NANO17 Force/Torque sensor was mounted between the wheeled probe and the manipulator distal tip to measure the three force components (Fx, Fy, Fz) imparted by the tissue onto the wheel, Fig 3.

Rolling indentation, experiments are performed on a silicone phantom and a porcine kidney sample. The silicone phantom is made from RTV6166 gel [16], and embedded with rubber nodules to simulate tumors. The tissue phantom is a 30 (height) ×150 (length) ×50(width) mm3 block, containing three embedded nodules (cylinders of diameter 10mm x length 10 mm) located in the tissue phantom as indicated in Fig.2.

Fig.2 The silicone phantom with three rubber nodules.

Fig. 3 The experiments on the silicone phantom. During the experiments, a wheeled indenter (8mm

diameter and 8mm length) is used to perform rolling indentation on the tissue phantom, at a constant indention depth and a constant speed of 45mm/s, covering the top surface from left to right, Fig.3. During the tests, the force exerted by the phantom onto the wheel is recorded. Four tests, with indentation depths of 3mm, 4mm, 5mm and 6mm, were performed, with each test repeated 3 times. The averaged test results are shown in Fig 11.

B. Porcine Kidney Tests Experiments are conducted on an Ex-vivo porcine kidney

sample (weight= 0.2 kg, lab temperature =15.1C, lab humidity =30%) with a rubber cylinder embedded within the kidney sample, at a depth of 10 mm from the top surface, Fig. 4.

Fig.4 A rubber cylinder embedded within the porcine sample

Fig.5 The probe path during rolling indentation.

The wheeled probe was then rolled over the kidney top

surface at a speed of 45 mm/s, Fig.5, while keeping the indentation depth approximately constant at 4 mm; the tests were repeated three times. During the tests, a plastic cling film of negligible thickness was used to cover the kidney sample to prevent tissue damage and loss of water from the sample.

V. ROLLING INDENTATION FINITE ELEMENT SIMULATION

A. FE Modelling of the Silicone Phantom The silicone phantom is modeled using ABAQUS, as a

planar rectangle (dimensions of 30 mm by 150 mm), with three holes corresponding to the nodule locations, Fig. 6. The FE mesh consists of 4884 four-node bilinear plane stress quadrilateral elements (ABAQUS element CPS4R).

Fig. 6 2D FE model for silicone.

Fig.7 2D FE model for rubber nodules

Fig.8 Assembled 2D finite element model for silicone phantom.

The three rubber nodules are modeled as 10 diameter 2D-circles, Fig. 7, using 112 four-node bilinear plane stress quadrilateral elements (ABAQUS element CPS4R). The three nodules are then assembled into the holes in the silicone model, Fig. 8, positioned at depths of 10 mm, 15 mm and 20 mm from the top surface. The indenter was modeled as a 8mm diameter 2D-circle with 25 elements of the two-node 2D linear rigid link (ABAQUS element R2D2). Both the ABAQUS elements used in this paper (CPS4R and R2D2) use reduced integration and hourglass control, assuming non-linear, isotropic, incompressible, and hyperelastic material behavior.

Fig. 9 ABAQUS simulation of rolling indentation on the silicone phantom. In the FE model, the contact between the rubber and

silicone is defined as “rough” and that between the indenter and silicone phantom as “frictionless” (ABAQUS contact). The lower surface of the silicone phantom is grounded and the experimental tests were repeated using FE simulations. Each rolling indentation FE simulation test requires approximately 10 minutes of computational time, on a 2.8 GHz Pentium (R)D machine with 3.5GB of RAM.

B. FE modelling of the porcine kidney sample

Fig.10 The porcine kidney rolling finite element simulation

The porcine kidney is modeled as a two-dimensional planar entity using the ABAQUS finite element software, with all dimensions obtained from measurements on the test sample. The kidney FE mesh consists of 2021 ABAQUS CPS4R elements, Fig.10. The rubber cylinders were modeled as described in the previous section, using 32 ABAQUS CPS4R elements. The simulated tumor is positioned at a depth of 10 mm from the surface, inside the kidney model, Fig. 10. The indenter is modeled as described in the previous section.

The porcine kidney experimental tests are repeated using FE simulations. Each FE simulation test requires approximately 8 minutes of computational time on a 2.8 GHz Pentium(R) D machine with 3.5GB of RAM.

VI. EXPERIMENT RESULTS AND ANALYSIS

A. Silicone phantom The experimental and FE model simulation results for the

silicone phantom are shown in Fig.11. As seen in Fig.11 and Table II, the RFEM results are in good agreement with the corresponding experimental data; the RMS errors range from 0.0289 to 0.1449, with the errors increasing with tumor depth.

Ø 10 mm

A1 A2

Fig.11 Silicone phantom results: Experimental (dashed line) and FE model (solid line). **Rd, = rolling indentation depth,

TABLE II

THE RMS ERRORS BETWEEN FINITE ELEMENT AND ROLLING INDENTATION

TESTS

B. Porcine kidney Fig.12 shows the rolling-tissue interaction forces from

porcine kidney experiments and RFEM simulations. The FE model predictions are in good agreement with the experimental results, with an RMS error of 0.05. Both results show an increased stiffness (force peak) at the vicinity of the embedded tumor. These results show the ability of RFEM to simulate a complex material like biological soft tissue.

0 5 10 15 20 25 30 35 40 45 50

0.4

0.5

0.6

0.7

0.8

0.9

1

Distance along rolling kidney path (mm)

Forc

e (N

)

Fig.12 Porcine kidney (dashed line) and FE model (solid line) results.

VII. LOCALIZING ABNORMALITIES WITHIN SOFT TISSUE

It has been shown that the wheel-tissue rolling interaction signals can be used to rapidly detect the presence of abnormalities within soft tissue during MIS [1-6]. However, localizing the abnormalities requires a detailed understanding of the underlying physics. Hence it is informative to study the relationship between the locations

of abnormalities and the tissue force-displacement characteristics, during rolling interactions. This section investigates the variation in reaction forces with respect to the tumor depth using FE simulations.

Fig.13. Silicone phantom FE simulations with varying tumour locations: (a)The hard nodule is placed at eight locations and (b) The RFEM of the silicone phantom

The tumor was placed at eight different locations, along

the mid-plane of the block, at depths of 8, 10, 12, 15, 18, 20, 22 and 24 mm, Fig. 13(a). The rolling indention was performed at four different depths; 3, 4, 5 and 6 mm. The rolling length is 60mm; a tumor is embedded at the mid-plane of the phantom block.

Rolling Indentation depth, Rd

3mm 4mm 5mm 6mm

RMS Error 0.0289 0.0613 0.1050 0.1449

Fig.14. (a) RFEM results for 4 different rolling indentation depths, with 8mm tumor depth. (b) Peak forces for eight different tumor depths with 6mm rolling indentation depth.

It is noted that the horizontal location of the tumor can be

determined by using the peak of the rolling force profile. The peak force increases with increased rolling indentation depth, Fig. 14 (a), and the peak force decreases with increased tumor depth, Fig 14 (b). In this section we investigate whether the magnitude of the indentation force can be correlated to the tumor depth for localization.

Fig.15. (a-c) RFEM results of 3 mm indentation depth with various tumor depths.

Fig.15 shows RFEM results at an indentation depth of 3

mm, for various tumor depths. The results indicate that there is a measurable increase in stiffness for tumor depths of 8 mm, 10 mm and 12 mm, Fig. 15 (b). However, deeper tumors (depths of 15-24 mm), with a 3mm indentation depth, does not appear to give a distinguishable variation in stiffness, Fig. 15 (c).

Fig. 16 Peak forces for eight different tumor depths with different rolling indentation depths.

For a given rolling indentation depth, the peak force values

appear to approach a constant asymptote, with increasing tumor depth, Fig. 16. Hence the rolling indentation method has the potential to locate the depth of tumors, with the maximum identifiable tumor depth depending on the indentation depth. Only three tumor depths are identifiable with an indentation depth of 3 mm (8, 10 and 12mm), seven tumor depths (15mm, 18mm, 20mm and 22mm) are identifiable with an indentation depth of 6mm.

From the RFEM curves in Figure 16, an empirical equation can be generated by curve fitting, to relate peak force to tumor depth, for this particular set of experimental data, Table III.

TABLE III EMPIRICAL EQUATIONS FOR TUMOUR DEPTH: dN =SIMULATION TUMOR

DEPTH, F = PEAK FORCE

Rolling indentation depth Empirical equation

3mm

Nd = 246.38F2 - 502.75F + 262.91

4mm

Nd = 36.057F2 - 119.44F + 106.71

5mm

Nd = 36.206F2 - 159.17F + 182.57

6mm

Nd = 12.983F2 - 78.552F + 126.92

To investigate the ability of the empirical equations in

Table III to identify tumor depths, the experiments with the silicone phantom (section IV-A) were carried out. The predicted depths and the prediction errors are shown in Table IV. It is noted that the predicted tumor depths are accurate, with errors ranging from 0.19 mm to 0.75 mm.

TABLE IV

USE OF EMPIRICAL EQUATIONS TO IDENTIFY TUMOUR DEPTH (NI-NOT

IDENTIFIABLE).

True simulated

tumour depth

Rd=3mm Rd=4mm Rd=5mm Rd=6mm

Predicted depth

Error (mm)

Predicted depth

Error (mm)

Predicted depth

Error (mm)

Predicteddepth

Error (mm)

A1(10mm) 10.19 0.19 10.75 0.75 9.69 0.31 9.68 0.32

A2(15mm) NI NI 15.43 0.43 15.56 0.56 14.46 0.54

A3(20mm) NI NI NI NI NI NI 19.70 0.30

VIII. CONCLUSIONS

This paper presents a rolling finite element model (RFEM) to simulate the rolling tissue indentation technique, which can cover a large tissue surface in a relatively short time. It is concluded that a rolling finite element model shows good correlation with experimental data. Following the abnormality localization, it shows good understanding on the relationship between the locations of abnormalities and tissue reaction forces during rolling interaction. Hence, it has a potential to predict tumour location. Future work will be carried out to develop generalised algorithms to identify the mechanical properties of tumours as well as tumour depths, using real-time RFEM.

ACKNOWLEDGMENT Support for this work has been provided by Rajamangala

University of technology (RMUTT), Thailand.

REFERENCES [1] Liu, H, Noonan, D. P., Challacombe, B. J., Dasgupta, P., Seneviratne,

L. D., Althoefer, K, Rolling Mechanical Imaging for Tissue Abnormality Localization During Minimally Invasive Surgery, IEEE Transactions on Biomedical Engineering, 2010 Feb;57(2): 404-14.

[2] K. Sangpradit, H. Liu, L. Seneviratne, K. Althoefer, “Tissue Identification using Inverse Finite Element Analysis of Rolling Indentation” IEEE Int. Conf. on Robotics and Automation, (ICRA), Kobe, Japan, 2009.

[3] H. Liu, D. Noonan, K. Althoefer, L. Seneviratne, “The Rolling Approach for Soft Tissue Modelling and Mechanical Imaging during Robot-Assisted Minimally Invasive Surgery”, Proc. IEEE Int. Conf. Robot. Autom., 2008.

[4] H Liu, P Puangmali, D Zbyszewski, O Elhage, P Dasgupta, J S Dai, L Seneviratne, and K Althoefer, An indentation depth–force sensing wheeled probe for abnormality identification during minimally invasive surgery Proceedings of the Institution of Mechanical Engineers, Part H, 2010.

[5] D. Noonan, H. Liu, Y. Zweiri, K. Althoefer, L. Seneviratne, “A Dual-Function Wheeled Probe for Tissue Viscoelastic Property Identification during Minimally Invasive Surgery”, in Proc. IEEE Int. Conf. Robot. Autom., 2008, pp.2629-2634.

[6] H.Liu, J.Li, QI.Poon, L.D.Seneviratne, K.Althoefer, "Miniaturized Force Indentation-depth Sensor for Tissue abnormality identification”, IEEE Int. Conf. on Robotics & Automation 2010.

[7] H. Liu, D. Noonan, Y. Zweiri, K. Althoefer, L. Seneviratne, The Development of Nonlinear Viscoelastic Model for the Application of Soft Tissue Identification, accepted by IEEE IROS 2007, Scheduled Code: TuA7.5, 30 October, 2007, San Diego, USA

[8] Talal M. Al-jaafreh, Yahya H. Zweiri, Lakmal D. Seneviratne, and Kaspar Althoefer, A New Soft Tissue Indentation Model for Estimating `Force-Displacement' Characteristics using Circular Indenters. Accepted for publication in the Proceedings of Institution of Mechanical Engineers, Part H, Journal of Engineering in Medicine, Vol. 222, No. 5, pp805-815, 2008.

[9] T. M. Al-jaafreh, L.D Seneviratne, Y.H Zweiri and K Althoefer, Modelling soft tissue-mechatronic tool interactions during indentation. International Journal of Modelling, Identification and Control, Vol. 5, No.4, 2008.

[10] B. Challacombe, K. Althoefer, H. Liu, D. Zbyszewski, P. Puangmali, D. Noonan, L. Seneviratne, D. Stoianovici, “Emerging Robotics”, unconditionally accepted for publication New Technologies in Urology, Springer publishing house 2009.

[11] P. Puangmali, K. Althoefer, L. D. Seneviratne, D. Murphy, and P. Dasgupta, “State-of-the-Art in Force and Tactile Sensing for Minimally Invasive Surgery”, IEEE Sensors Journal, vol. 8, no. 4, pp. 371-381, 2008.

[12] P Puangmali, K Althoefer, L Seneviratne, Optical Fiber Sensor for Soft Tissue Investigation during Minimally Invasive Surgery, IEEE ICRA pp. 2934-2939, 2008.

[13] Umesh Vivekananda, Alastair Henderson, Declan G. Murphy, Kaspar Althoefer, Lakmal Seneviratne and Prokar Dasgupta. The Science Behind Haptics in Robotic Urological Surgery, B J U International, 104 , 4 3 3 – 4 3 6, 2009.

[14] K.H. Fuchs, “Minimially Invasive Surgery”, Endoscopy, 34: 154-159, 2002.

[15] Fung YC. “A First Course in Continuum Mechanics,” Englewood Clifs, NJ: Prentice Hall, 1994.

[16] A. M. Okamura, C. Simone, and M.D. O’Leary, “Force Modelling for Needle Insertion into Soft Tissue,” IEEE Transactions on Biomedical Engineering, Vol. 51, No. 10, pp. 1707-1716, 2004.

[17] F. Tendick, M.C. Cavusoglu, “Human Machine Interfaces for Minimally Invasive Surgery,” Proc. 19th Intl. Conference IEEE/EMBS, pp 2771-277, 1997.

[18] Southern Surgeons Club, “A prospective Analysis of 1518 Laparoscopic Colecystectomies,” New England Journal of Medicine, Vol. 324, pp 1079-1078, 1991.

[19] S. Phipps, T.H.J. Yang, F.K. Habib, R.L. Reuben and S.A. McNeill, “Measurement of the mechanical characteristics of benign prostatic tissues: A novel method for assessing benign prostatic disease,” Urology 65(5) (2004), 1024–1028.

[20] Y. Tillier, et al, “Finite element modeling for soft tissue surgery base on nonlinear elasticity behavior,” Excerpta Media International congress series 1268, Elsevier, Amsterdam, 2004, pp. 384-389

[21] Yi liu, “A nonlinear finite element model of soft tissue indentation Proceedings of Medical Simulation,” International Symposium - ISMS 2004, Cambridge, MA, June 17-18, 2004, Lecture Notes in Computer Science vol. 3078, Springer-Verlag, pp. 67-76.

[22] Amy E. Kerdok, “Identification of nonlinear constitutive law parameters of breast tissue,” Proceeding of 2005 summer Bioengineering Conference, June 22-26, Vail Cascade Resort & Spa, Vail, Colorado

[23] M. Zhang, Y. P. Zheng, and A. F. T. Mak. “Estimating the effective Young's modulus of soft tissues from indentation tests, nonlinear Finite element analysis of effects of friction and large deformation.” Medical Engineering and Physics., vol.19, pages: 512-517, 1997.

[24] E. Tonuk and M. B. Silver-Thorn. “Nonlinear elastic material property estimation of lower extremity residual limb tissues,” IEEE Transactions on Neural Systems and Rehabilitation Engineering, vol.11, pages: 43-53, 2003.

[25] G. Picinbono, H. Delingette, and N. Ayache. “Nonlinear anisotropic elasticity for real time surgery simulation,” Graphical Models, vol.65, pages: 305-321, 2003.

[26] Y. Zhong, B. Shirinzadeh, G. Alici, and J. Smith. “A new methodology for deformable object simulation,” IEEE International Conference on Robotics and Automation, pages: 1902-1907, Barcelona, Spain, 2005.

[27] G. Szekely, Ch. Brechbuhler, R. Hutter, A. Rhomberg, N. Ironmonger, and P. Schmid. “Modelling of soft tissue deformation for laparoscopic surgery simulation,” Medical Image Analysis, vol.4, pages: 57-66, 2000.

[28] J. Schwartz, M. Denninger, D. Rancourt, C. Moisan, and D. Laurendeau. “Modelling liver tissue properties using a nonlinear viscoelastic model for surgery simulation,” Medical Image Analysis, vol.9, pages:103-112, 2005.

[29] A. Duysak, J.J. Zhang, and V. Ilankovan. “An Efficient modelling and simulation of soft tissue deformation using mass-spring systems,” International Congress Series, vol.1256, pages: 337-342, 2003.

[30] ABAQUSTM ABAQUS version 6.8-1 Manual, 2008. [31] J. S. Bergstrom and M. C. Boyce “Constitutive Modelling of the Large

Strain Time-Dependent Behaviour of Elastomers” J. Mech. Phys. Solids. Vol. 46, pp. 931-954, 1998.