GCS Training Manual
GCS - TM - 0245 – 2006/12/24
VIDAS / mini VIDASSystem
VIDAS Connection
VIDAS Connection
GCS - TM - 0245 - 2006/12/24
Summary
1. Environment overviewChapter 1.1 Define what a connection isChapter 1.2 System positioning / Use situationsChapter 1.3 General system operationChapter 1.4 Needs to have a functional connectionChapter 1.5 Customer requirement <-> limits and constraints
2. System overviewChapter 2.1 Network basicsChapter 2.2 Principles of BCI Net connectionChapter 2.3 BCI Net format and protocol Chapter 2.4 Principle of BCI RS232 connection Chapter 2.5 BCI RS232 format and protocolChapter 2.6 hardware / Software characteristics
3. Install / UninstallChapter 3.1 Install prerequisitesChapter 3.2 Delivery contentChapter 3.3 BCI Net software (un)installationChapter 3.4 BCI RS232 software (un)installationChapter 3.5 How to check/save the installation/settingsChapter 3.6 How to switch between BCI RS232 and BCI Net
VIDAS Connection
GCS - TM - 0245 - 2006/12/24
Summary
4. BCI Net SoftwareChapter 4.1 Overview of BCI Net softwareChapter 4.2 Send and receive data with VIDAS PC threw
BCI Net softwareChapter 4.3 Installation of BCI Simulator software Chapter 4.4 BCI Simulator overviewChapter 4.5 Simulate Upload and Download operation
5. BCI RS232 SoftwareChapter 5.1 Overview of BCI RS232 softwareChapter 5.2 Send and receive data with VIDAS PC threw
BCI RS232 softwareChapter 5.3 Installation of BCI Simulator software Chapter 5.4 Simulate LIS operation (Download/Upload)Chapter 5.5 Send a result from mini VIDAS
6. Support end usersChapter 6.1 Know the product and toolsChapter 6.2 Identify the different levels to be trainedChapter 6.3 Train end users on user levelChapter 6.4 Train end users how to manage and
maintain the systemChapter 6.5 Identify useful information for LIS vendors
7. Appendices
GCS - TM - 0245 – 2006/10/23 - lesson 1 - p.1
VIDAS Connection
Lesson 1
Environment overview
Content
• 1.1 Define what a connection is• 1.2 System positioning / Use situations• 1.3 General system operation• 1.4 Needs to have a functional connection• 1.5 Adaptation of the customer requirement / limits and constraints
Content
• 1.1 Define what a connection is• 1.2 System positioning / Use situations• 1.3 General system operation• 1.4 Needs to have a functional connection• 1.5 Adaptation of the customer requirement / limits and constraints
TM 0245 12/26/2006 4 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.1
VIDAS Connection Lesson 1 :Environment overview
Chapter 1.1
Define what a connection is
Objectives
• Define the Terms used of a connection :• LIS, BCI, UCI, UPLOAD, DOWNLOAD, Network administrator• Know the types of connection• Know the partners (LIS vendors, Customer, Network administrators)
Objectives
• Define the Terms used of a connection :• LIS, BCI, UCI, UPLOAD, DOWNLOAD, Network administrator• Know the types of connection• Know the partners (LIS vendors, Customer, Network administrators)
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.2
LIS Network administratorBCI UCIUpload DownloadAnalysis request ResultTechnical validation Biological validation
Definition of the terms used on a connection
TM 0245 12/26/2006 5 sur 232
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.3
Definition of the terms used on a connection
BCI SoftwareBCI Software : Software allowing to establish : Software allowing to establish the communication between the bioMérieux the communication between the bioMérieux
Instrument (AI) and the Customer’s Instrument (AI) and the Customer’s Laboratory information System (LIS)Laboratory information System (LIS)
AIAILISLIS
ResultRequest
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.4
Types of a connection
LIS computer
RESULTS
RESULTS
= Upload= UploadUnidirectional (UCI)
TM 0245 12/26/2006 6 sur 232
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.5
Types of a connection
REQUESTS
REQUESTS
= Upload + Download= Upload + DownloadBidirectional (BCI)
LIS computerRESULTS
RESULTS
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.6
Types of a connection
Serial Connection :
Network Connection :
TM 0245 12/26/2006 7 sur 232
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.7
Partners involved in a connection
3 partners always involved :CustomerCustomer LIS CompanybioMérieux
Other possible partners :Hospital information serviceNetwork administrator
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.8
Partners involved in a connection
Role / Responsibility :Customer• Depends on the country : Could be the project leader or
only the client• Setup the LIS software for each parameter
Customer LIS Company• Develop the connection driver• Validate the connection between BCI and LIS
bioMérieux• Provide the BCI software technical specification +
Specificities of each automate• Could verify + validate the connection VIDAS/BCI/LIS
TM 0245 12/26/2006 8 sur 232
Chap 1.1 Define what a connection is
GCS - TM - 0245 – 2006/12/24 - chap 1.1 - p.9
Partners involved in a connection
Role / Responsibility :
Hospital information service / Network administrator• Provide a functional architecture (cable,
switch, router, rj45 plug)
• Provide technical information to put in place a connection (IP Address, Host name, FTP Port value)
TM 0245 12/26/2006 9 sur 232
GCS - TM - 0245 – 2006/12/24- chap 1.2 - p.1
VIDAS Connection Lesson 1 :Environment overview
Chapter 1.2
System positioning / Use situations
Objectives
• Understand the place of the connection in the customer workflow• Understand the aim /advantages of a connection (Transversal)• Know the MKTG strategy and competitors for the connection
Objectives
• Understand the place of the connection in the customer workflow• Understand the aim /advantages of a connection (Transversal)• Know the MKTG strategy and competitors for the connection
Chap 1.2 System positioning / Use situations
GCS - TM - 0245 – 2006/12/24- chap 1.2 - p.2
Place of the connection in the customer workflow
Virology
Serology
Immunology
Supply chain departmentLinen room
Archival storage
Transport
Technical department
Clinic pharmacyOperating
blockDiagnostic
imagingLaboratory
Administrative department
Human resourcePatientHuman
resource department
Activity management
Purchase department
TM 0245 12/26/2006 10 sur 232
Chap 1.2 System positioning / Use situations
GCS - TM - 0245 – 2006/12/24- chap 1.2 - p.3
Place of the connection in the customer workflow
CollectionReception Pre-analytic
RunTechnical validationBiological validation
Chap 1.2 System positioning / Use situations
GCS - TM - 0245 – 2006/12/24- chap 1.2 - p.4
Place of the connection in the customer workflow
BCISoftware
AISoftware
bioMérieux
Lab Information System (LIS)
Connection Driver
BiologicalBiologicalValidationValidation
TechnicalTechnicalValidationValidation
ResultRequest
Front OfficeFront Office
Customer site
RS
Net
TM 0245 12/26/2006 11 sur 232
Chap 1.2 System positioning / Use situations
GCS - TM - 0245 – 2006/12/24- chap 1.2 - p.5
Aim /advantages of a connection
(GBEA) : French protocol for good execution of analysis
Required by ⇒ customerGoal :
• Decrease workload• Decrease work time• Increase security by decreasing manual data input
(I.e. human errors)• Traceability (barcode)• Facilities to manage Patient data• Recommended by the GBEA
Required by ⇒ bioMérieux Goal : Fix the customer
Chap 1.2 System positioning / Use situations
GCS - TM - 0245 – 2006/12/24- chap 1.2 - p.6
MKTG strategy and competitors for the connection
To be defined
TM 0245 12/26/2006 12 sur 232
Chap 1.2 System positioning / Use situations
GCS - TM - 0245 – 2006/12/24- chap 1.2 - p.7
MKTG strategy and competitors for the connection
Competitors :Immulite 2000 Elecsys 2010Access 2Vitros ECI : RS232 + TCP/IP (DRM)Liaison : RS232 serial interfaceModular : RS232 serial interface (DRM)CentaurArchitectAxsymAIA 600 IIAIA 21Cobas Core : RS232 serial interface (DRM)
TM 0245 12/26/2006 13 sur 232
GCS - TM - 0245 – 2006/12/24- chap 1.3 - p.1
VIDAS Connection Lesson 1 :Environment overview
Chapter 1.3
General system operation
Objectives
• Understand the demo / explanation
Objectives
• Understand the demo / explanation
Chap 1.3 General system operation
GCS - TM - 0245 – 2006/12/24- chap 1.3 - p.2
Demo / Explanation
ResultRequest
LIS
AI
Request Result
BCI RS232
AI
Request ResultLIS
TM 0245 12/26/2006 14 sur 232
Chap 1.3 General system operation
GCS - TM - 0245 – 2006/12/24- chap 1.3 - p.3
Demo / Explanation
UCI RS232 Result
LIS
TM 0245 12/26/2006 15 sur 232
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.1
VIDAS Connection Lesson 1 :Environment overview
Chapter 1.4
Needs to have a functional connection
Objectives
• Know what it needs for the driver development• Know what it needs to put in place a connection• Know the role of each partners
Objectives
• Know what it needs for the driver development• Know what it needs to put in place a connection• Know the role of each partners
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.2
Driver Development
BCI SOFT.
bioMérieuxbioMérieuxrepresentativerepresentative
LIS CompanyLIS Company
CONNECTION DRIVER
«Request» specs Available information
«Result» specs Built-in information
Technical Suggestions Specific development
Protocols
Chart
Connectiondriver
TM 0245 12/26/2006 16 sur 232
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.3
AI and BCI features
Lab. organization knowledge
Customer needs knowledge
Host (LIS) features
Host parameters knowledge
Programming knowledge
Work together in the lab.
BCI SOFT.
bioMérieuxbioMérieuxrepresentativerepresentative
LIS CompanyLIS Company
CONNECTION DRIVER
CustomerCustomer
Needs for a connection
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.4
Needs for a connection
AI and BCI features :Specifications of protocol, messagesWhat we can send and receive
Lab. organization knowledge :Type of barcode usedWhere’s technical validationWhat is the list of test performed on AIRerun & confirmationIdentification Patient / SampleDilution managementExternal QCData management
TM 0245 12/26/2006 17 sur 232
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.5
Needs for a connection
Customer needs knowledge :AI & BCI features in concordance with Lab.
Organization knowledge
LIS Features :Specifications : What the LIS can send and
integrate, Why no more ?
We must know all of these points to be an actor in the connection installation / configuration
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.7
Service from LIS company
Installing the communication cable between AI & LIS. If hospital : hospital computer department, to anticipate cable installation prior to the connection !
Setting up of connection protocol
Setting up host computer parameters for :Dispatching requests to VIDAS (BI-DI)
Integrating results from VIDAS
Training of users about LIS use
WHAT ?
TM 0245 12/26/2006 18 sur 232
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.8
Service from LIS company
COST ?
Global offer with the instrument
Depending of the LIS company
Depending of connection typeUni or Bi-directional
One or more connection point (BCI Net)
Ask for an ESTIMATE
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.9
Service from bioMérieux Rep.
BCI software Installation / Setup
Barcode reader Installation (optional)
Good communication verification (simulator)
Communication validation with LIS company ?
Real time trial (daily routine work)
Training of VIDAS/BCI users
BCI Configuration Backup & printout
WHAT ?
TM 0245 12/26/2006 19 sur 232
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.10
WHAT PRICE ?
Free of charge for the customer ?
Must be valorized !!!
Then only the barcode reader may represent a cost for the customer ...
Service from bioMérieux Rep.
Commercial decision
Chap 1.4 Limits and constraints due the environment
GCS - TM - 0245 – 2006/12/24- chap 1.4 - p.11
WHY ?
The hardware configuration may be different between two customersAdding a connection may imply a hardware investment or a LIS software updateFinancial survey without bad surprise. Do not forget the cost of the barcode wand / Switch !Any order of a connection requires a prior Estimate
Request to estimate
BCI MUST BE VALORIZED TO THE CUSTOMER
TM 0245 12/26/2006 20 sur 232
GCS - TM - 0245 – 2006/12/24- chap 1.5 - p.1
VIDAS Connection Lesson 1 :Environment overview
Chapter 1.5Customer
requirements
Objectives
• Adapt the customer requirements with respect to the limits and constraints
Objectives
• Adapt the customer requirements with respect to the limits and constraints
Limits and constraints
Chap 1.5 Customer requirements <-> limits and constraints
GCS - TM - 0245 – 2006/12/24- chap 1.5 - p.2
Adapt the customer requirements
Type of connection : mono or BidirectionalSize of each field
Sample ID = Max 12Patient ID = Max 12Format of the Date
Dilution managementQuality Control :
Internal (bioMérieux)External
Technical validationWorking listRerun
TM 0245 12/26/2006 21 sur 232
GCS - TM - 0245 – 2006/12/24 - lesson 2 - p.1
VIDAS Connection
Lesson 2
System overview
Content• 2.1 Network basics • 2.2 Principles of BCI Net connection• 2.3 BCI Net format and protocol• 2.4 Principle of BCI RS232 connection• 2.5 BCI RS232 format and protocol• 2.6 Hardware / Software characteristics
Content• 2.1 Network basics • 2.2 Principles of BCI Net connection• 2.3 BCI Net format and protocol• 2.4 Principle of BCI RS232 connection• 2.5 BCI RS232 format and protocol• 2.6 Hardware / Software characteristics
TM 0245 12/26/2006 22 sur 232
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.1
VIDAS Connection Lesson 2 :System overview
Chapter 2.1
Network basics
Objectives
• Assimilate the network basics
Objectives
• Assimilate the network basics
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.2
Network basics
LISLIS
Network
TM 0245 12/26/2006 23 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.3
Network basics
Summary :1. Principle / Terminology2. Architectures3. Protocols4. IP Addressing5. Network components6. Network command (DOS command)
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.4
Network basics
Principle / TerminologyDefinition : a computer network allows several machines to communicate between them, in order to exchange information, to transfer files or to share resources (printer,etc).
Specific Terms :•Packet : Information in circulation on the network•Format : Language•Logical address : Host Name / IP address
•Protocol : Specification describing the rules to be followed in a data exchange
•Client / Server : Host computer (server) makes its information available to clients
TM 0245 12/26/2006 24 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.5
Network basics
Example - Bus Topology : All devices are connected to a central cable (bus, or backbone)
2 types of architectures : “Post to post” : - same rights to all
- exchange without control“Client-Server” : Host computer makes information available for clients
Post to post
Architectures
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.6
Network basics
Ring Topology All devices are connected to one another in the shape of a closed loop, so that each device is connected directly to two other devices, one on either side of it.
Architectures Post to Post
TM 0245 12/26/2006 25 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.7
Network basics
LIS
A CLIENT sends a request to the server thanks to :• Its login and password• address and port name.
The SERVER receives requests and makes information available for the clients.
CLIENT
CLIENT
SERVER
Architectures Client - Server
AI 1
CLIENT
AI 2
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.8
Network basics
Order----
Central Office
ShipmentOrder
Shop
Customer
Network principle
TM 0245 12/26/2006 26 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.9
Network basics
Network principleOrder
----
Recipient Address
Recipient Address
Order----
Transmitter Address
CLIE
NT
CLIE
NT
SE
RV
ER
Transmitter Address
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.10
Network basics
Network principle
ResultRequest
LIS
Request Result
AI = Analysis Instrument
AI 1
TM 0245 12/26/2006 27 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.11
Network basics
Logical Address
Logical Address
Folder
FormatLIS
Logical Address
Transport 1
Transport 2Transport 2
Transport 1
RequestResult
Result
CLIE
NT
CLIE
NT
SE
RV
ER
Request
Transmitter Address
AI 1
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.12
Network basics
ASTM / XMLLIS
IP Address Host name
BCI Net
Protocol :FTP / TCP/IP
Protocol :FTP / TCP/IP
IP Address Host name
AI
IP Address Host name
Protocol :IIOP / TCP/IP
Protocol:IIOP / TCP/IP
Request
Result
ResultRequest
IIOP
CLIE
NT
SE
RV
ER
SFTP
CLIE
NT
AI 1
TM 0245 12/26/2006 28 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.13
IP Addressing
Order----
Central Office
Shop
CustomerCentral OfficeAddress
ShopAddress
CustomerAddress
Central OfficeAddress
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.14
IP Addressing
LISRequest
Result
ResultRequest
IIOP
CLIE
NT
SE
RV
ER
SFTP
CLIE
NT
LIS @IP& host name
AI @IP& host name
BCI host @IP& host name
AI 1
TM 0245 12/26/2006 29 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.15
IP Addressing
IP address (@IP)Four decimal numbers
Each number coding for eight bytes
Separated by dots
Ex : 156.34.10.76
IP addressing structure
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.16
IP Addressing
1 to 126 Class A Network ID Host portion128 to 191 Class B Network ID Host portion192 to 223 Class C Network ID Host portion224 to 239 Class D For Multiclass240 to 254 Class E For Research
Note : • 0 doesn’t exist• 127 is reserved for local host• 255 is reserved for broad cast
First decimal NumberDN DN DN DN
TM 0245 12/26/2006 30 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.17
IP Addressing
IP Address Class ID Network Host portion18.27.39.50 A 18.0.0.0 0.27.39.50
150.175.225.3 B 150.175.0.0 0.0.225.3199.13.18.29 C 199.13.18.0 0.0.0.29
@IP examples :
Class A : 126 (128-2) networks, 16777214 (16-2) hosts
Class B : 16384 networks, 65534 hosts per network
Class C : 2097152 networks, 254 hosts per network
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.18
IP Addressing
Networks : 0.0.0.0 is used for default network127.0.0.0 is used for loop back function
Host :255.0.0.0 is not used for individual address255.1.1.1 is used for broadcast.
TM 0245 12/26/2006 31 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.19
Subnet Addressing
Why Subnets ?
Local administrators had to request another network number from Internet before a new network could be installed locally.
Internet routing tables were beginning to grow
Two level classfull hierarchyNetwork ID Host portion
Three level subnet hierarchyNetwork ID Subnet portion Host portion
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.20
Subnet Addressing
Subnet Mask : 255.255.255.0
200.1.1.34 200.1.1.35200.1.1.65 200.1.1.66
200.1.1.97
200.1.1.98200.1.1.1
TM 0245 12/26/2006 32 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.21
Addressing mode
Manual mode :The network administrator sets manually each @IP
in each PC. Used for small private networks, or to separate some PCs from the rest of the network.
DHCP mode :The DHCP mode uses a client-server model. The
DHCP server manages the information related to the TCP/IP configuration, and its dynamic transmission to the clients.
Information = @IP and Lease (using period)
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.22
Direct connection
Direct Connection : The “source PC” compares its own @IP to the “recipient PC” @IP, then
sends the packet to the “recipient PC”.
200.0.0.1 200.0.0.2
PC1 PC2
TM 0245 12/26/2006 33 sur 232
Chap 2.1 Different formats and protocols
GCS - TM - 0245 –2006/12/24 - chap 2.1 - p.23
Indirect connection
Indirect Connection : PC1 has first to address the packets (for PC2) to Router 1. Then R1 checks its
routing table, and send the packets to R2. Finally R2 sends the packets to PC2.
PC1 PC2
R1 R2200.0.0.1
200.0.0.50 220.12.23.50220.12.23.2
---------------
TM 0245 12/26/2006 34 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 2.2 - p.1
VIDAS Connection Lesson 2 :System overview
Chapter 2.2
Principles of BCI Net connection
Objectives
• Understand the principles of a BCI Net connection
Objectives
• Understand the principles of a BCI Net connection
Chap 2.2 Principle of BCI Net connection
GCS - TM - 0245 – 2006/12/24 - chap 2.2 - p.2
Principe of operation
“Upload” folder
LIS AI
“Download” folder
Pick up ResultPick up Result Send ResultSend Result
Send RequestSend Request Send RequestSend Request
TM 0245 12/26/2006 35 sur 232
Chap 2.2 Principle of BCI Net connection
GCS - TM - 0245 – 2006/12/24 - chap 2.2 - p.3
Principe of operation
“Upload” folder
LIS AI
“Download” folder
Pick up ResultPick up Result Send ResultSend Result
Send RequestSend Request Send RequestSend Request
FTP on TCP/IP
IIOP on TCP/IP
3 LIS Max 5AI Max
Chap 2.2 Principle of BCI Net connection
GCS - TM - 0245 – 2006/12/24 - chap 2.2 - p.4
Principe of operation
“Upload” folder
LIS AI
“Download” folder
Pick up ResultPick up Result Send ResultSend Result
Send RequestSend Request Send RequestSend Request
FTP on TCP/IP
IIOP on TCP/IP
3 LIS Max 5AI Max
File FORMAT : ASTM/XML File FORMAT : ASTM/XML
File FORMAT : ASTM/XML File FORMAT : ASTM/XML
TM 0245 12/26/2006 36 sur 232
Chap 2.2 Principle of BCI Net connection
GCS - TM - 0245 – 2006/12/24 - chap 2.2 - p.5
Dispatching
LISLIS
VIDASPC02VIDASPC02
Network
VIDASPC01VIDASPC01
VU
04
Chap 2.2 Principle of BCI Net connection
GCS - TM - 0245 – 2006/12/24 - chap 2.2 - p.6
Dispatching
LISLIS
VIDASPC02VIDASPC02
Network
VIDASPC01VIDASPC01
VU
04
2x
TM 0245 12/26/2006 37 sur 232
Chap 2.2 Principle of BCI Net connection
GCS - TM - 0245 – 2006/12/24 - chap 2.2 - p.7
Dispatching
Dispatching mode
Systematic dispatching : All analysis request are sent to all AI - used for a single AI connection
Laboratory dispatching : All analysis requests are sent to an AI, depending on the name of the system indicated by the LIS in the download file - used for connected system using different techniques (VIDAS + VIDIA)
bioMérieux dispatching : All analysis request are sent to AI thanks to a routing table set by the customer in BCI Net software - used for connected systems using the same technique
TM 0245 12/26/2006 38 sur 232
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.1
VIDAS Connection Lesson 2 :System overview
Chapter 2.3
BCI Net format and protocol
Objectives
• Understand the format and protocol used on BCI Net connection
Objectives
• Understand the format and protocol used on BCI Net connection
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.2
Protocols
Protocol : All rules and procedures to be applied to send and receive data through a network.
Slow & Uncertain Fast & Secure
Slow & Secure
+AR
AR
TM 0245 12/26/2006 39 sur 232
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.3
Protocols
File Transfer Protocol is used between LIS and BCI Net to exchange data. BCI Net is the FTP server, and LIS is the FTP client.
LIS
Request Result
Protocol :FTP / TCP/IP
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.4
Protocols
LIS
LIS open a session
LIS Identifies : Login + Password
LIS “puts” request file
BCI Net checks Login, PWD & request file *
BCI Net’s acknowledges
LIS closes session
LIS “gets” result file
LIS identifies : Login + Password
Request Result
BCI Net sends the request file to the appropriate AI AI puts the result file
BCI Net checks the result file *
TM 0245 12/26/2006 40 sur 232
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.5
Protocols
IIOP protocol is used between BCI Net - AI to exchange data. BCI Net is the IIOP server and each AI is an IIOP client
Request
Result
AI 1
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.6
Protocols
Open a session / AI connected to BCI Net
BCI Net send the request to the appropriate AI
BCI Net records a Transaction event
BCI Net stores result file
BCI Net records a Transaction event
BCI Net checks the result file
AI sends result to BCI Net
Request
Result
AI 1
TM 0245 12/26/2006 41 sur 232
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.7
BCI Net Format
Communication : FTP protocol on BCI Net
LIS
FTPaccountName_YYMMDD_HHMMSS_s.xml
AIname_YYMMDD_HHMMSS_s.xml
AIname_YYMMDD_HHMMSS_s.xml.OK.FTPaccountName
Confir m
at ion
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.8
BCI Net Format
Message structure : A message consists in a collection of FIELDS, containing various DATA, for example :
DATA :
M
Field 1 Field 2
D C
Field 3
P T
Field 4 Field 5
Field 1 ManufacturerField 2 Delivery addressField 3
Field 5Field 4
ContentsPriceTaxes
TM 0245 12/26/2006 42 sur 232
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.9
BCI Net Format
FIELDS :
mprRequest
MessageType
PatientID
123 SMITH
PatientName
456 DIG
SampleID
Assay
I require a DIG analysis for patient 123, Mr SMITH, sample 456
I require Message typePatient 123 Patient IDMr SMITH
DIG analysisSample 456
Patient nameSample IDAssay
mpr (RS232) / Request (Net)123SMITH456DIG
CONTENTS :
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.10
BCI Net Format
On BCI RS232 :mtmpr|pi123|pnSMITH|si|ci456|rtDIG|Equivalenton BCI Net :
<request><PatInfo>
<PI>123</ PI><PN>SMITH</ PN>
</PatInfo><TestOrder>
<Specimen><CI>456</CI>
</Specimen><Test>
<RTInfo><RT>DIG</RT>
</RTInfo></Test>
</TestOrder></request>
TM 0245 12/26/2006 43 sur 232
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.11
BCI Net Format
rslResult
MessageType Patient
ID
123 SMITH
PatientName
456
SampleID
DIG
Assay
I send result for patient 123, Mr SMITH, sample 456 on DIG
LongAssayName
Digoxin
Cmpl.Time
Cmpl.Date
Qualit.Result
Quant.Result
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.12
BCI Net Format
BCI RS232 : mtrsl|pi123|pnSMITH|si|ci456|rtDIG|rnDigoxine|tt10:11|td03/03/05|ql|qn19 ng/ml|
BCI Net : <requestresult><PatInfo>
<PI>123</PI><PN>SMITH</PN>
</PatInfo><TestOrder>
<Specimen><SpecI>456</SpecI>
</Specimen><Test>
<RTInfo><RT>DIG</RT><RN>Digoxin</RN>
</RTInfo></Test>
</TestOrder><Result>
<Value>19</Value><Unit>ng/ml</Unit><operator name>vidas</operator name><ResultStatus>F</ResultStatus>
</Result></requestresult>
TM 0245 12/26/2006 44 sur 232
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.13
BCI Net Format
• A XML file is structured using markups :<markup>markup value</markup>
• Each system is based on a XML diagram :ASTM.xsd
(in d:\bmx\bcinet\lib\resources\xml\ folder)
• Some markup values are preformatted :<sex>F,M,U</sex>
Chap 2.3 BCI Net format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.3 - p.14
BCI Net Format
XML Advantages :
Tree structure
Easy to read
Easy to create
Easy to update
Universal
TM 0245 12/26/2006 45 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 2.4 - p.1
VIDAS Connection Lesson 2 :System overview
Chapter 2.4
Principles of BCI RS232 connection
Objectives
• Understand the principles of a BCI RS232 connection
Objectives
• Understand the principles of a BCI RS232 connection
Chap 2.4 Principle of BCI RS232 connection
GCS - TM - 0245 – 2006/12/24 - chap 2.4 - p.2
Principe of operation
“Upload” folder
LIS AI
“Download” folder
Send RequestSend Request Send RequestSend Request
Send ResultSend ResultSend ResultSend Result
TM 0245 12/26/2006 46 sur 232
Chap 2.4 Principle of BCI RS232 connection
GCS - TM - 0245 – 2006/12/24 - chap 2.4 - p.3
Principe of operation
“Upload” folder
LIS AI
“Download” folder
Send RequestSend Request Send RequestSend Request
bioMérieux alternate or
literal
5 LIS Max 1AI Max
Send ResultSend ResultSend ResultSend Result
bioMérieux alternate or
literal
Chap 2.4 Principle of BCI RS232 connection
GCS - TM - 0245 – 2006/12/24 - chap 2.4 - p.4
Principe of operation
“Upload” folder
LIS AI
“Download” folder
Send RequestSend Request Send RequestSend Request
3 LIS Max 1AI Max
File FORMAT : ASCII File FORMAT : ASCII
File FORMAT : ASCII File FORMAT : ASCII
Send ResultSend ResultSend ResultSend Result
bioMérieux alternate or
literal
bioMérieux alternate or
literal
TM 0245 12/26/2006 47 sur 232
Chap 2.4 Principle of BCI RS232 connection
GCS - TM - 0245 – 2006/12/24 - chap 2.4 - p.5
Request
LISLIS
RS232 CableOr thanks a special device
VIDASPC01VIDASPC01
Chap 2.4 Principle of BCI RS232 connection
GCS - TM - 0245 – 2006/12/24 - chap 2.4 - p.6
Result
LISLIS
VIDASPC01VIDASPC01
RS232 CableOr thanks a special device
TM 0245 12/26/2006 48 sur 232
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.1
VIDAS Connection Lesson 2 :System overview
Chapter 2.5
BCI RS232 format and protocol
Objectives
• Understand the format and protocol used on BCI RS232 connection
Objectives
• Understand the format and protocol used on BCI RS232 connection
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.2
Protocols
Protocol : All rules and procedures to be applied to send and receive data through a network.
Slow & Uncertain Fast & Secure
Slow & Secure
+AR
AR
TM 0245 12/26/2006 49 sur 232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.3
Serial Communication Protocol
LISLIS
<< YES >>
<< GOODBYE >>
<< OK >>
<< HELLO >>
<< MY DATA >>
BCI RS232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.4
Each session STARTS with «ENQ» character : <ENQ> (Enquiry)
<ACK> (Acknowledge) ⇔ OKafter <ENQ> : ready to receive dataafter <GS>xx : successful data transmission
<NAK> (No Acknowledge) ⇔ BADafter <ENQ> : communication impossible
- transmission errorafter <GS>xx : wrong checksum
Each session ENDS with an «EOT» character :<EOT> (End of Transmission)
Serial Communication Protocol
A tN oS ∩W e E n R qS ∪
TM 0245 12/26/2006 50 sur 232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.5
bioMérieux Alternate protocol<ENQ><ACK><STX> ...
<GS>xx<ETX><ACK><EOT>
Mode «One by One»<ENQ>..frame1..<EOT> <ENQ>..frame2..<EOT>
Different possibilitiesAlternate «One by One»Alternate «Continuous»Literal «One by One»Literal «Continuous»
Serial Communication Protocol
bioMérieux Literal protocol
<ENQ><ACK><STX> ... <GS>xx<ACK><ETX><EOT>
Mode «Continuous»
<ENQ>..frame1.. <ETX><ETX>..frame2..<EOT>
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.6
Serial Communication Protocol
ResultsAlternate «One by One»Literal «One by One»
Requests
Alternate «One by One»Alternate «Continuous»Literal «One by One»Literal «Continuous»
TM 0245 12/26/2006 51 sur 232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.7
Serial Communication Protocol
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.8
<ENQ>
<ENQ>
}<ENQ> retryinterval (sec)
<ENQ>
}<ENQ
>
}}
<EOT>
<ENQ>retry limit (nbr)
Serial Communication Protocol
<ENQ> retryinterval (sec)
<ENQ> retryinterval (sec)
<ENQ> retryinterval (sec)
TM 0245 12/26/2006 52 sur 232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.9
<ENQ>
}}}
<EOT>
Checksumretry limit
(nbr)
Serial Communication Protocol
Checksum retryinterval (sec)
Checksum retryinterval (sec)
Checksum retryinterval (sec)
<STX> ... <GS>xx
<STX> ... <GS>xx
<STX> ... <GS>xx
<ACK>
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.10
<ENQ>
}
<EOT>
Serial Communication Protocol
Checksum retryinterval (sec)
<STX> ... <GS>yy
<STX> ... <GS>xx
<ACK>
<NAK>
<ACK>
TM 0245 12/26/2006 53 sur 232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.11
}
<EOT>
Serial Communication Protocol
<ENQ>
Inter recorddelay (sec)
<STX>
<RS> ... <GS>xx
<ACK>
<ACK>
<RS> …
<ETX>
<ENQ>
{Last Master Delay (sec)
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.12
Serial Communication Protocol
<ACK>
} Inter messagedelay (sec)
<EOT>
<ACK>
<ENQ>
<STX> ... <GS>xx<ETX>
<STX> ... <GS>yy<ETX>
<ACK>
LIS >> AI only<ENQ> <STX>…frame1…<ETX>
<STX>…frame2…<ETX><EOT>
TM 0245 12/26/2006 54 sur 232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.13
<EOT>{Host
ResponseTimeout
Serial Communication Protocol
<ACK>
<ENQ>
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.14
Serial Communication Protocol
<ENQ>
<EOT>
<ACK>
<ENQ>
<STX> ... <GS>xx<ETX>
<ACK>
Priority given tothe LIS
TM 0245 12/26/2006 55 sur 232
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.15
Adding Carriage Return and Line Feed at different places :
After <STX> : <STX> <CR> <LF>After <ETX> : <ETX> <CR> <LF>After <RS> : <RS> <CR> <LF>After <GS> : <GS>xx <CR> <LF>
Serial Communication Protocol
Using different separators Default : Field | Date / Time :Keep protocols homogeneous : NEVER CHANGE
THEMRk : | will become a ù on the screen (Vidas text)
Chap 2.5 BCI RS232 format and protocol
GCS - TM - 0245 –2006/12/24 - chap 2.5 - p.16
Differences betweenVIDAS and mini VIDAS
VIDAS Mini VIDASProtocol : bioMérieux literal
or alternate
Stop bits : 1 or 2
Character <RS> : located at the beginning of the message (after <STX>) and all 80 characters if the lenghtof the message exceeds 80 characters.
Protocol : bioMérieux alternate
Stop bits : 1
Character <RS> located before each field of the message
Reset of memory results by switching off the machine
TM 0245 12/26/2006 56 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.1
VIDAS Connection Lesson 2 :System overview
Chapter 2.6
Hardware / Software characteristics
Objectives
• Know hardware characteristics :• Cable• Network components• Barcode reader
• Know software characteristics :• Specifications• Architecture• Backup• VIDAS software specificities
Objectives
• Know hardware characteristics :• Cable• Network components• Barcode reader
• Know software characteristics :• Specifications• Architecture• Backup• VIDAS software specificities
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.2
Cable / barcode reader
Cable : RJ45
Barcode reader : HHP
The barcode reader is used to scan the barcode printed out in the sample tube. It is used to decrease human error
TM 0245 12/26/2006 57 sur 232
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.3
Network component
Network card
RJ45 connector
Repetitor
Modem
Router
Switch
Firewall
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.4
VIDAS PC Software
Simple demographics = ci, piComplete demographics = ci, pi, pn, pb, so, Industrial demographics = ci, pi
TM 0245 12/26/2006 58 sur 232
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.5
VIDAS PC Software
The analysis request come from LIS arrive in the Job list area
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.6
VIDAS PC Software
Depending on the configuration, to send a result to the LIS :
Validate the result. Automatically the result is sent to LISResend a result threw the Resent icon
TM 0245 12/26/2006 59 sur 232
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.7
VIDAS PC Software
Specificities of VIDAS PC SoftwareIndustrial demographicsDilution managementExternal Quality ControlSample in 24 Hours2 Patients with the same Patient ID but with a different demographyInterpretation BCI RS232 / BCI Net
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.8
VIDAS PC Backup
BCI Net : The configuration and the logs are saved during a VIDAS PC data backup
BCI RS232 : The configuration can be saved in the Field Service folder. After that, the configuration will be saved during a VIDAS PC backup data
TM 0245 12/26/2006 60 sur 232
Chap 2.6 Hardware / Software characteristics
GCS - TM - 0245 – 2006/12/24 - chap 2.6 - p.9
mini VIDAS Software
Barcode Wand
Sound Menu
Misc Items
LIS User Options
LIS Upload Enabled
Warn about Missing IDs
LIS auto validation
LIS auto validationValidate LIS Results
If LIS auto validation is not selected, you have to validate each result to send them to LIS
TM 0245 12/26/2006 61 sur 232
GCS - TM - 0245 – 2006/12/24 - lesson 3 - p.1
VIDAS Connection
Lesson 3
Install / Uninstall
Content
• 3.1 Install prerequisites• 3.2 Delivery content• 3.3 BCI Net software (un)installation• 3.4 BCI RS232 software (un)installation• 3.5 How to check/save the installation/settings• 3.6 How to switch between BCI RS232 and BCI Net
Content
• 3.1 Install prerequisites• 3.2 Delivery content• 3.3 BCI Net software (un)installation• 3.4 BCI RS232 software (un)installation• 3.5 How to check/save the installation/settings• 3.6 How to switch between BCI RS232 and BCI Net
TM 0245 12/26/2006 62 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 3.1 - p.1
VIDAS Connection Lesson 3 :Install / Uninstall
Chapter 3.1
Install prerequisites
Objectives
• Know the install prerequisites
Objectives
• Know the install prerequisites
Chap 3.1 Install prerequisites
GCS - TM - 0245 – 2006/12/24 - chap 3.1 - p.2
Prerequisite document
TM 0245 12/26/2006 63 sur 232
Chap 3.1 Install prerequisites
GCS - TM - 0245 – 2006/12/24 - chap 3.1 - p.3
Prerequisite document
Chap 3.1 Install prerequisites
GCS - TM - 0245 – 2006/12/24 - chap 3.1 - p.4
Prerequisite document
Read the prerequisite before to go to the customer siteFor BCI Net, Fix an appointment with the network administrator and the customerTry to homogenize the parameter with the LIS vendors (FTP account name, uploading period, FTP port value, etc …)Validate this document with the sales people
TM 0245 12/26/2006 64 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 3.2 - p.1
VIDAS Connection Lesson 3 :Install / Uninstall
Chapter 3.2
Delivery content
Objectives
• Check and know the delivery content
Objectives
• Check and know the delivery content
Chap 3.2 Delivery Content
GCS - TM - 0245 – 2006/12/24 - chap 3.2 - p.2
VIDAS BCI Content
VIDAS PC / BCI Net :VIDAS PC VU04BCI Net User Manual Ref :99605Service Manual Connectivity Ref :99604
VIDAS PC / BCI RS232 :VIDAS PC VU02/VU04VIDAS PC User Manual Ref :93498/93553Service Manual Connectivity Ref :99604
BCI NET Pack Simu Ref 99603 :BCI Net Software V1.1.0BCI Simulator software
TM 0245 12/26/2006 65 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.1
VIDAS Connection Lesson 3 :Install / Uninstall
Chapter 3.3BCI Net software (un)installation
Objectives
• Perform and verify the physical / BCI Net Software (un)installation• Understand the users rights to access and set software• Know the conditions in order to start settings• Know the location of configuration data• Know the limits and constraints for each setting• Know the default setting• Understand the impact for each modification
Objectives
• Perform and verify the physical / BCI Net Software (un)installation• Understand the users rights to access and set software• Know the conditions in order to start settings• Know the location of configuration data• Know the limits and constraints for each setting• Know the default setting• Understand the impact for each modification
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.2
Installation schedule
1. Install configuration2. Activate BCI Net software3. Checking4. Troubleshoot installation
TM 0245 12/26/2006 66 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.3
Installation configuration
LISLIS
VIDASPC02VIDASPC02
Network
VIDASPC01VIDASPC01
VU
04
VU
04
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.4
Software Update / Installation
New PC installationwith V4.0.0*
Install/Update VIDAS PC V4.0.0 software
New VIDAS Customer Current VIDAS Customer
Software updateto V4.0.0
VIDAS Text VIDAS PC
* Depends on the country
TM 0245 12/26/2006 67 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.5
Software Update / Installation
Install V4.0.0 software
V4.0.0 software update
• Follow the procedure Ref :?????
• Follow the procedure Ref :?????
Improvements on V4.0.0 VIDAS PC software• Windows® XPe• PTC Updates thanks to new PDF barcode reader
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.6
Network configuration
Network configuration VIDAS PC
Hostname
VIDASPCØ?
TM 0245 12/26/2006 68 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.7
Network configuration
IP address
2ØØ.Ø.Ø.?
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.8
Network checking
Network configuration checking
• To check, on CMD window, use « ping » command
TM 0245 12/26/2006 69 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.9
Installation configuration
Physical installation
Install*/Update VIDAS PC V4.0.0 software
Network configuration VIDAS PC & VIDIA
Network configuration checking
Configuration installed
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.10
Installation schedule
1. Install configuration2. Activate BCI Net software3. Checking4. Troubleshoot installation
TM 0245 12/26/2006 70 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.11
BCI Net activation
On VIDAS PC V4.0.0 software
From Windows® Explorer, run « InstallBCINet » script file
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.12
BCI Net activation
BCI Net starts automaticallyBCI Server starts automaticallyAuto-start feature = {
TM 0245 12/26/2006 71 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.13
Installation schedule
1. Install configuration2. Activate BCI Net software3. Checking4. Troubleshoot installation
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.14
Checking
The BCI Net icon appears on the desktop
After login, check that the servers are started
TM 0245 12/26/2006 72 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.15
Installation schedule
1. Install configuration2. Activate BCI Net software3. Checking4. Troubleshoot installation
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.16
Troubleshoot Installation
We need your feed back !!!!
TM 0245 12/26/2006 73 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.17
Installation recap
Already doneInstall BCI Net software
To be doneSetup VIDIA UC network
To be done*VIDAS PC software V4.0.0 Install or Update
Already doneVIDIA Software InstallingTo be doneSetup VIDASPC network
To be doneActivate BCI Net on VIDAS PC
Already createdBCI_ADMIN & BCI_ROUTINE groups
Windows XPRJ45 available
PC hosting BCI Net
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.18
AS/SE
Customer
bmx_adminDynamic passworddb2bmx
admintobeset
vidasvidas
GROUPSUSERpassword
BCI_ADMINBCI_ROUTINE
BCI Net
Temp_admin
User rights to access and manage software
TM 0245 12/26/2006 74 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.19
bmx_adminDynamic passworddb2bmx
adminTobeset
vidasvidas
GROUPSUSERpassword
VIDAS_ADMINVIDAS_ROUTINE
VIDAS PC
AS/SE
CustomerTemp_adminDynamic password
User rights to access and manage software
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.20
Conditions in order to start settings
To modify VIDAS PC settings :For Windows : have to belong to Administrators
groups (db2bmx, bmx_admin)For VIDAS PC Software : have to belong to
VIDAS_ADMIN group (db2bmx, bmx_admin, admin)
To modify BCI Net software settings :Have to belong to BCI_ROUTINE and
BCI_ADMIN groups To modify Windows settings :
Have to belong to Administrators group
TM 0245 12/26/2006 75 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.21
Location of configuration data
For VIDAS PC settings :On the VIDAS PC Configuration menu
For Windows : Network settings, Printer settings, Computer name
For BCI Net software settings :On the BCI Net software configuration menu. Stop the server before to modify the configuration
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.22
PC hosting BCI Net
IIOP PortVPC=18001
VIDAS PC settings
TM 0245 12/26/2006 76 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.23
BCI Net software settings
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.24
Uploading period : Delay (in sec.) for LIS to get the upload packet. If not, Warning appears on BCI Net softwarePort number : Discuss with the LIS vendors to know the value. The LIS is connected through this FTP portPending file storage : Pending directory is used to store the message awaiting to be sent to a disconnected AI. Delay before transaction are moved to “failed” directoryPending file scanning : Delay for BCI net to scan the “pending” directory, and if BCI Net find a transaction oldest that x mn, BCI Net moves it to “failed” directoryLog file storage : Delay before “transactions” and “operations” are moved to historicLog file Scanning : delay between two readings of files storage
Limits and constraint for each setting
TM 0245 12/26/2006 77 sur 232
Chap 3.3 BCI Net software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.3 - p.25
Uploading period : NAPort number : 21Pending file storage : 60mnPending file scanning : 10mnLog file storage : 7200mn (5 days)Log file Scanning :1440mn (1 day)Dispatching mode : Systematic dispatching
After modify the FTP account management, Think to restart servers
Default setting
TM 0245 12/26/2006 78 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.1
VIDAS Connection Lesson 3 :Install / Uninstall
Chapter 3.4BCI RS232 software (un)installation
Objectives
• Perform and verify the physical / BCI RS232 Software (un)installation• Understand the users rights to access and set software• Know the conditions in order to start settings• Know the location of configuration data• Know the limits and constraints for each setting• Know the default setting• Understand the impact for each modification
Objectives
• Perform and verify the physical / BCI RS232 Software (un)installation• Understand the users rights to access and set software• Know the conditions in order to start settings• Know the location of configuration data• Know the limits and constraints for each setting• Know the default setting• Understand the impact for each modification
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.2
Installation schedule
1. Install configuration2. Activate BCI RS232 software3. Checking4. Troubleshoot installation
BCI RS232
TM 0245 12/26/2006 79 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.3
Installation configuration
LISLIS
Serial Cable
VU
04BCI RS232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.4
Physical installation
DB9s2 : RxD
3 : TxD
5 : GndNull-modem cable
Ref 451 1251A
DB9s2 : RxD
3 : TxD
5 : Gnd
On VIDAS PC, connect the serial cable to the Com3On mini VIDAS, connect the serial cable to the “Computer” plug
TM 0245 12/26/2006 80 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.5
Software Update / Installation
Install/Update VIDAS PC V4.0.0 software
BCI RS232 is already installed, need to be activated
Install/Update mini VIDAS V5.2.0 software
UCI RS232 is already installed and activated
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.6
Installation configuration
Physical installation
Software installation is done
Configuration installed
TM 0245 12/26/2006 81 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.7
Installation schedule
1. Install configuration2. Activate RS232 software3. Checking4. Troubleshoot installation
BCI RS232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.8
BCI RS232 activation
On VIDAS PC V4.0.0 software
From Windows® Explorer, run « InstallBCINT » script file
TM 0245 12/26/2006 82 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.9
BCI RS232 activation
Restart the VIDAS PC computer
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.10
Installation schedule
1. Install configuration2. Activate BCI RS232 software3. Checking4. Troubleshoot installation
BCI RS232
TM 0245 12/26/2006 83 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.11
Checking
The BCI RS232 icon appears on the desktop
After login, check that the communication is started
(not at the first time)
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.12
Installation schedule
1. Install configuration2. Activate BCI RS232 software3. Checking4. Troubleshoot installation
BCI RS232
TM 0245 12/26/2006 84 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.13
Troubleshoot Installation
We need your feed back !!!!
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.14
Installation recap
Already doneInstall BCI RS232 software
To be done*VIDAS PC software V4.0.0 Install or Update
To be doneSetup VIDASPC network
To be doneActivate BCI RS232 on VIDAS PC
Already createdBCI_ADMIN & BCI_ROUTINE groups
Windows XP/NTCOM available
PC hosting BCI RS232
Already donemini VIDAS (UCI installation)
To be donemini VIDAS software setup
TM 0245 12/26/2006 85 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.15
AS/SE
Customer
bmx_adminDynamic passworddb2bmx
admintobeset
vidasvidas
GROUPSUSERpassword
BCI_ADMINBCI_ROUTINE
BCI Net
Temp_admin
User rights to access and manage software
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.16
bmx_adminDynamic passworddb2bmx
adminTobeset
vidasvidas
GROUPSUSERpassword
VIDAS_ADMINVIDAS_ROUTINE
VIDAS PC
AS/SE
CustomerTemp_adminDynamic password
User rights to access and manage software
TM 0245 12/26/2006 86 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.17
Conditions in order to start settings
To modify VIDAS PC settings :For Windows : have to belong to Administrators
groups (db2bmx, bmx_admin)For VIDAS PC Software : have to belong to
VIDAS_ADMIN group (db2bmx, bmx_admin, admin)
To modify BCI RS232 software settings :Have to belong to BCI_ROUTINE and BCI_ADMIN
groups
To modify Windows settings :Have to belong to Administrators group
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.18
Location of configuration data
For VIDAS PC settings :On the VIDAS PC Configuration menu
For Windows : Network settings, Printer settings, Computer name
For BCI RS232 software settings :On the BCI RS232 software configuration menu. Stop the communication before to modify the configuration
TM 0245 12/26/2006 87 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.19
BCI RS232 software settings
Link name = Link name set on VIDAS PC
= LIS setup
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.20
Link name = Link name set on BCI
VIDAS PC settings
TM 0245 12/26/2006 88 sur 232
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.21
VIDAS PC settings
Check if the LIS is able to manage them
Explain to the customer and to the LIS vendors the impacts to unselect them
Upload option :Patient namePatient Date of BirthPatient SexSample originUnit associated with qnDilutionCalibration expiredInstrument IDSerial numberTechnologist
Download option :Patient namePatient Date of BirthPatient SexSample originDilution
Chap 3.4 BCI RS232 software (un)installation
GCS - TM - 0245 – 2006/12/24 - chap 3.4 - p.22
Check this configuration with the LIS vendors.Not necessary to set a high Baud rate
BCI Software ConfigurationDefault setting
TM 0245 12/26/2006 89 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 3.5 - p.1
VIDAS Connection Lesson 3 :Install / Uninstall
Chapter 3.5How to check/save the installation/settings
Objectives
• Verify and save the installation and settings
Objectives
• Verify and save the installation and settings
Chap 3.5 How to check/save the installation/settings
GCS - TM - 0245 – 2006/12/24 - chap 3.5 - p.2
Checking installation/settings
After installation, Restart the computer and run BCI software
Check the connection between VIDAS PC and BCI, then check the connection between BCI and LIS
Check that the settings are the same between BCI Software and the LIS (Option field on VIDAS PC Software)
TM 0245 12/26/2006 90 sur 232
Chap 3.5 How to check/save the installation/settings
GCS - TM - 0245 – 2006/12/24 - chap 3.5 - p.3
Save installation / checking
From BCI RS232 software :Stop the connectionCheck that you are belong to BCI_ADMIN groupSave the configuration on D:\bmx\vidas\fieldservice
For BCI Net software :Perform a data backupDuring the Backup, the BCI Net folders are copied to the CD : bcinet_backup.zip
Chap 3.5 How to check/save the installation/settings
GCS - TM - 0245 – 2006/12/24 - chap 3.5 - p.4
Restore configuration
From BCI RS232 software :Stop the connectionCheck that you are belong to BCI_ADMIN groupOpen the configuration from the “fieldservice” folder or from the data backup
For BCI Net software :Insert a data backupFrom windows explorer, copy the “BciRoot\config”
folder located on the CD : bcinet_backup.zipOverwrite the folder “config” on d:\bmx\bcinet\BciRoot\config
TM 0245 12/26/2006 91 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 3.6 - p.1
VIDAS Connection Lesson 3 :Install / Uninstall
Chapter 3.6How to switch between BCI RS232 and BCI Net
Objectives
• Perform and verify the switch between VIDAS BCI RS232 and BCI Net
Objectives
• Perform and verify the switch between VIDAS BCI RS232 and BCI Net
Chap 3.6 How to switch between BCI RS232 and BCI Net
GCS - TM - 0245 – 2006/12/24 - chap 3.6 - p.2
Switch BCI RS232 to BCI Net
Log you as db2bmx userFrom Windows Explorer, run “UninstallBCINT.wsf” file located on d:\bmx\system\gcs
Restart the computerLog you as db2bmx userFrom Windows Explorer, run InstallBCINeT.wsf file located on d:\bmx\system\gcs
Restart the computer, check that the BCI Net software is launched automatically and set VIDAS PC and BCI Net software
TM 0245 12/26/2006 92 sur 232
Chap 3.6 How to switch between BCI RS232 and BCI Net
GCS - TM - 0245 – 2006/12/24 - chap 3.6 - p.3
Switch BCI Net to BCI RS232
Log you as db2bmx userFrom Windows Explorer, run UninstallBCINet.wsf file located on d:\bmx\system\gcs
Restart the computerLog you as db2bmx userFrom Windows Explorer, run InstallBCINT.wsf file located on d:\bmx\system\gcs
Restart the computer, check that the BCI RS232 software can be launched manually and set VIDAS PC and BCI RS232 software
TM 0245 12/26/2006 93 sur 232
GCS - TM - 0245 – 2006/12/24 - lesson 4 - p.1
VIDAS Connection
Lesson 4
BCI Net Software
Content
• 4.1 Overview of BCI Net Software• 4.2 Send / Receive data with VIDAS PC threw BCI Net Software• 4.3 Installation of BCI Simulator Software• 4.4 BCI Simulator Overview• 4.5 Simulate download and upload operation
Content
• 4.1 Overview of BCI Net Software• 4.2 Send / Receive data with VIDAS PC threw BCI Net Software• 4.3 Installation of BCI Simulator Software• 4.4 BCI Simulator Overview• 4.5 Simulate download and upload operation
TM 0245 12/26/2006 94 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.1
VIDAS Connection Lesson 4 :BCI Net Software
Chapter 4.1Overview of BCI Net software
Objectives
• Know the different menus of the BCI Net software
Objectives
• Know the different menus of the BCI Net software
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.2
Global overview
AlarmsMenus
Status bar
Action Bar
TM 0245 12/26/2006 95 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.3
Global overview
LIS area
Max 3 LIS
AI area
Max 5 AI
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.4
Global overview
Connection Device / BCI is not established
Connection Device / BCI is established and works well (remains gray for LIS)
Warning alarm : minor problem
Error alarm : Major problem, COMMUNICATION STOPPED
Think to check the work zone status !!!
TM 0245 12/26/2006 96 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.5
Global overview
Alarms
OK, all is fine
Warning (not blocking)
Blocking alarm
Transaction alarms :Ie : New analysis system features Dispatch configuration
Operation alarms :Ie : LIS connection Incorrect FTP login
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.6
Global overview
Supervision of connection : shows the connection status between AI to LIS
Configuration of the dispatcher : used when several AI connected, using the same technique
Journal of transaction : used to display the content of the selected exchanged message
Journal of operation : used to display history of operations on BCI Net (Connect – Disconnect)
FTP accounts management, software settings
TM 0245 12/26/2006 97 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.7
Global overview
Start the FTP Server (LIS – BCI) and BMX Server (AI – BCI)
Stop the FTP Server (LIS – BCI) and BMX Server (AI – BCI)
Test the correct process directory BCI to LIS
Stop the Servers ⇔ Stop the communicationbetween all AI connected to BCI and BCI to LIS
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.8
Global overview
Clock Software version
Login name : Click on the to change login
Check that the clocks between BCI Net software, AI software and the LIS are the same.
Minimize the window(Ie : to switch to VIDAS PC Software)
Close the BCI Net softwareBe careful : Stops Servers
TM 0245 12/26/2006 98 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.9
How to run BCI Net Software
BCI Net software :
• Starts automatically depends on setting
• Starts manually
Start or Restart manually BCI Net software :
• Double click on BCI Net icon
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.10
How to run BCI Net Software
BCI Net software starts
Enter “Login”and “Password”
TM 0245 12/26/2006 99 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.11
How to create a FTP Account
FTP Account = LISOnly access by users from
BCI_ADMIN group
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.12
How to create a FTP Account
To modify BCI Net configuration,Stop servers
Stop the Servers Stop the communication between all AI
connected to BCI and BCI to LIS
TM 0245 12/26/2006 100 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.13
How to create a FTP Account
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.14
How to create a FTP Account
AI Software
LIS computer
FTP port (21)
IIOP port (VPC = 18001VIDIA = 19005
Remote = 24000)
TM 0245 12/26/2006 101 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.15
How to create a FTP Account
LIS will be connected thanks to the “login” and “password” set in configuration Menu
Uploading period : Delay (in sec.) for LIS to get the upload packet
Log files :
• storage : delay before “transactions” and “operations” are moved to historic
• scanning : delay between two readings of files storage
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.16
How to create a FTP Account
Port number : LIS connected through this port number
Pending files :• storage : Delay (in min.) for the storage of message
awaiting to be sent to a disconnected AI (pending directory). After this delay, an alarm is send to the user, and the message is redirected into the Failed directory.
• scanning : Delay (in min.) between 2 readings of the pending directory.
Dispatching mode : see p.18
TM 0245 12/26/2006 102 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.17
How to create a FTP Account
Save
Start servers
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.18
How to dispatch Analysis Request
Systematic dispatching : All analysis request are sent to all AI - used for a single AI connection
Laboratory dispatching : All analysis requests are sent to an AI, depending on the name of the system indicated by the LIS in the download file - used for connected system using different techniques (VIDAS + VIDIA)
bioMérieux dispatching : All analysis request are sent to AI thanks to a routing table set by the customer in BCI Net software - used for connected systems using the same technique
TM 0245 12/26/2006 103 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.19
How to dispatch Analysis Request
+ Laboratorydispatching=
+ bioMérieuxdispatching=
Systematicdispatching=
RECAP :
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.20
How to dispatch Analysis Request
SaveSame dispatching mode for all connected LIS !
TM 0245 12/26/2006 104 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.21
Systematic Dispatching mode
LISLIS BCI Net sends requests (LIS to connected AI) without
checking the format and the content of the analysis request
BCI Net postal box
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.22
bioMérieux Dispatching mode
TM 0245 12/26/2006 105 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.23
bioMérieux Dispatching mode
Each time it connects, AI sends its activated assays to BCI Net software if bioMérieux dispatching selected.
Alarm appears only if a new assay
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.24
bioMérieux Dispatching mode
Select Routine or Rerun mode for each Assay
TM 0245 12/26/2006 106 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.25
Laboratory Dispatching mode
LISLIS BCI Net sends analysis requests to AI, according to the instrument name indicated
inside the file by the LIS.
The LIS can interpret result files thanks to the instrument
name indicated inside the analysis result file.
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.26
How to acknowledge an Error
TM 0245 12/26/2006 107 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.27
How to acknowledge an Error
Click on OK
After acknowledgement, selected message disappears
Export in HTML
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.28
How to acknowledge an Error
Acknowledge alarm/error
Resend selected request to AI
Display warning/error message and corrective action to be done
Export selected warning/error messages into HTML format to save or print out
TM 0245 12/26/2006 108 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.29
How to search Analysis Request
Search could be done by customer or
FSE
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.30
How to search Analysis Request
Transaction menu is refreshed on a daily basis
TM 0245 12/26/2006 109 sur 232
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.31
How to search Analysis Request
Select directions
Select time
Transaction menu is refreshed on a daily basis
Chap 4.1 Overview of BCI Net Software
GCS - TM - 0245 – 2006/12/24 - chap 4.1 - p.32
How to search Analysis Request
Keyword : * means …..
Transaction menu is refreshed on a daily basis
TM 0245 12/26/2006 110 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 4.2 - p.1
VIDAS Connection Lesson 4 :BCI Net Software
Chapter 4.2Send / Receive data with VIDAS PC
threw BCI Net software
Objectives
• Know how to send and receive data with VIDAS PC threw BCI Net software
Objectives
• Know how to send and receive data with VIDAS PC threw BCI Net software
Chap 4.2 Send / Receive data with VIDAS PC threw BCI Net software
GCS - TM - 0245 – 2006/12/24 - chap 4.2 - p.2
Function : Send a result from VIDAS PC
Network
VIDASPC01VIDASPC01
VU
04
LISLIS
TM 0245 12/26/2006 111 sur 232
Chap 4.2 Send / Receive data with VIDAS PC threw BCI Net software
GCS - TM - 0245 – 2006/12/24 - chap 4.2 - p.3
Function : Send a result from VIDAS PC
Send automatically a result : depends on the VIDAS PC Configuration menu
(re)Send manually a VIDAS PC result via the result menu
Chap 4.2 Send / Receive data with VIDAS PC threw BCI Net software
GCS - TM - 0245 – 2006/12/24 - chap 4.2 - p.4
Function : Send a result from VIDAS PC
Check in the transaction menu that the result has been correctly sent from VIDAS PC to BCI
AI to BCI
TM 0245 12/26/2006 112 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.1
VIDAS Connection Lesson 4 :BCI Net Software
Chapter 4.3Install BCI Simulator software
Objectives
• Install, set and check BCI simulators software
Objectives
• Install, set and check BCI simulators software
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.2
BCI Simulator software installation
“BCI Simulators” is used with both BCI RS232 and BCI Net, and includes two simulators :
• an LIS simulator to send analysis requests to BCI software
• an AI simulator to send analysis results to BCI software
TM 0245 12/26/2006 113 sur 232
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.3
BCI Simulator software installation
Technical specifications
PC Pentium III – 128 Mo RAMSerial port / Network port
Hardware
Compliant with : BCI RS232 R2.1.0 (or higher) and BCI Net R1.1.0
Software
Login : bciPassword : bci
User & Pwd
Supported by :Windows® XP SP1, 2000 SP2 and NT4
Operating system
Appearance set to Windows Classic
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.4
BCI Simulator software installation
Insert CD IMM13 and go to “BCI Simulator”Run “j2re-1_4_1_07-windows-i586-i.exe” program located in “BCI SIMULATORS R2.0.0\java”Accept Java License
Select
TM 0245 12/26/2006 114 sur 232
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.5
BCI Simulator software installation
Run “install.exe” program located in “BCI SIMULATORS R2.0.0\simulators”
Select the installation folder of BCI Simulators
Rk : Zip labels may be in French
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.6
BCI Simulator software installation
Select the file “environment_***.reg” (*** depends on O.S.) located in\BCI SIMULATORS\simulators.
Double-click on it to import the new values
Close the current Windows® session and open a new one or restart the PC
Run “init.bat” file located in C:\BCI SIMULATORS* \bin\ (*=default)
All necessary « Upload » and « Download » folders are created automatically
TM 0245 12/26/2006 115 sur 232
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.7
BCI Simulator Setting
Simulator.BCIServer = BCI Net hostname (or @IP)
Simulator.BCIPort = FTP port set in BCI Net software
Simulator.User = login set in BCI Net software
Simulator.Pwd = password set in BCI Net software
The ftp.properties file gives to BCI Simulators software the necessary parameters to simulate LIS connection properly.
To modify it :Right click on the file nameEdit with Wordpad®
Location : \BCI SIMULATORS\lib\properties\simulator\
To simulate a LIS :
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.8
BCI Simulator Setting
To simulate a AI :
JAVA_HOME"="C:\\Program Files\\Java\\j2re1.4.1_07
NAMESERVICE_HOST"="localhost"
NAMESERVICE_PORT"="29000"
Java Path
Set Port value to :• 29000, 19005 or 18001 for BCI Net• 14000 for BCI RS232
Replace localhost by hostname or @IP of PC hosting BCI Net
or BCI RS232
TM 0245 12/26/2006 116 sur 232
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.9
BCI Simulator launching
Double click on “run.bat” file to launch BCI Simulators software, and log as “bci”
pwd : “bci”
Chap 4.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 4.3 - p.10
Troubleshooting Installation
PB to connect to BCI Net software :Modify “run.bat” file to stop the firewall service
PB to access to serial port :Modify “run.bat” file to write the complete way to access to java software
Run.bat :net stop sharedaccess@echo offecho Compatible with CosNAming SUN JDK1.4.1 and CosNaming
Visibroker 5.0echo Variable NAMESERVICE_HOST : %NAMESERVICE_HOST%echo Variable NAMESERVICE_PORT : %NAMESERVICE_PORT%Set SIMULATOR_ROOT=..@echo on"C:\Program Files\Java\j2re1.4.1_07\bin\javaw" -classpath …..pausenet start sharedaccess
TM 0245 12/26/2006 117 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.1
VIDAS Connection Lesson 4 :BCI Net Software
Chapter 4.4BCI Simulator overview
Objectives
Know the different menus of the BCI Simulator software
Objectives
Know the different menus of the BCI Simulator software
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.2
BCI Simulator Software Overview
11/ LIS BCI RS232
2
2/ LIS BCI Net
3
3/ AI Simu
Consultation Zone of XML
message Consultation Zone of error message
Edition Zone of XML message
RS 232 <> FTP ⇒ change of environment
TM 0245 12/26/2006 118 sur 232
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.3
BCI Simulator Software Overview
New message
Load message / ScriptSave
Save as …
Send … to BCI• Results• Requests
Refresh directory / Reload FTP config
Upload results from BCI to LIS
Connect / Disconnect AI
Send AI properties
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.4
LIS BCI Net
Add NodeRemove Node
XML Editor
TM 0245 12/26/2006 119 sur 232
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.5
LIS BCI Net
The function Add node / Remove Node allows too to create an analysis request to test system
Add node / Remove Node function :This function allows to add and remove Node through right click.In Red color, the mandatory fields appear.For some of fields, a choice is showed. For example, The field Sex (F,M,U)
The Analysis Request created are saved in the « Lis_WorkList » directory
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.6
LIS BCI Net
Text Editor
To allow to modify rapidly a File.The Function Copy/Paste is running
in the screen.
TM 0245 12/26/2006 120 sur 232
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.7
LIS BCI Net
View of Upload directory content
bmx_000000_000000_0.xml
bmx_000000_000000_0.xml.ok.LS1
Result management
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.8
LIS BCI Net
The FTP account setting is recorded in the « ftp.properties » file.
FTPaccount
TM 0245 12/26/2006 121 sur 232
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.9
AI BCI Net
Simulate AI with BCI Net :• XML Editor• TXT Editor
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.10
AI BCI Net
The Analysis system account setting is done in the « Instrument.properties » file.
Analysis systemconnection
TM 0245 12/26/2006 122 sur 232
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.11
AI BCI Net
Instrument.properties (Located : \BCI SIMULATORS\lib\properties\simulator)
AI Name
Test Name
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.12
AI BCI Net
Special case : bmx_simulator• Without modify the LIS configuration, use bmx_simulator type in order to simulate analysis receipt or sending result.
• The simulator is managed by BCI Net as long as it is connected.
TM 0245 12/26/2006 123 sur 232
Chap 4.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 4.4 - p.13
AI BCI Net
Modify Environment variables and restart the PC
NAMESERVICE_HOST"="localhost"
NAMESERVICE_PORT"="29000"
Set Port value to :• 29000, 19005 or 18001 for BCI Net• 14000 for BCI RS232
Replace localhost by hostname or @IP of PC hosting BCI Net
or BCI RS232
TM 0245 12/26/2006 124 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.1
VIDAS Connection Lesson 4 :BCI Net Software
Chapter 4.5Simulate download and upload operation
Objectives
• Know how to simulate LIS operation• Know how to simulate AI operation
Objectives
• Know how to simulate LIS operation• Know how to simulate AI operation
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.2
Simulate LIS BCI Net Operation
On BCI Simulators software :
• Modify « ftp.properties »
• Refresh “FTP account”
On BCI Net software :
• Create FTP account
• Start FTP server
TM 0245 12/26/2006 125 sur 232
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.3
Simulate LIS Operation
Check that LIS is recognized by BCI Net
The FTP server must be started on BCI Net !!!
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.4
How to create an Analysis Request
New
The analysis request file must be saved as“ftp account name_YYMMDD_HHMMSS_s.xml”
Modify by “Add” or “Remove” node
Save as…
Reminder
XML Editor
TM 0245 12/26/2006 126 sur 232
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.5
How to create an Analysis Request
Root node
Child node
Parent node
Value of node
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.6
How to create an Analysis Request
To set a value (new or predefined)
To add in the same positionTo add information below the nodeTo suppress node
Add nodeInsert node
Remove nodeSet value
TM 0245 12/26/2006 127 sur 232
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.7
How to create an Analysis Request
Save as…
New
In Text Editor
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.8
How to save an Analysis Request
Save as…
Select astm_xml
Modify directly in the file
TM 0245 12/26/2006 128 sur 232
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.9
How to save an Analysis Request
Rules of the format :Analysis request :
FTP account name_YYMMDD_HHMMSS_s.xmlFTP account name = FTP account name set in BCI Net soft s = sequence (0 to 99)
Analysis result :AI name_YYMMDD_HHMMSS_s.xml
Ai name = name of the automate set in VIDAS PC Windows (computer name)
s = sequence (0 to 99)
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.10
How to send an Analysis Request
Load
Send
TM 0245 12/26/2006 129 sur 232
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.11
How to check the reception
From BCI Net software, the Transaction Journal allows you to check transactions LIS ↔ BCI Net
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.12
How to check the download result
From BCI Net send a “dummy message” to LIS
Download function :
TM 0245 12/26/2006 130 sur 232
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.13
How to check the download of the result
From BCI Simulators, upload dummy message
bmx_000000_000000_0.xmlbmx_000000_000000_0.xml.ok.LS1
The file bmx_000000_000000_0.xml.ok.FTP account
is created by BCI Net after uploading
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.14
How to simulate a AI
Modify Environment variables and restart the PCModify “Instrument.properties” fileStart BCI Simulator softwareCheck on BCI Software that the server are startedConnect the AI simulatedCheck on BCI Software that the AI is connected
TM 0245 12/26/2006 131 sur 232
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.15
How to send an Analysis Result
Chap 4.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.16
How to check the reception
TM 0245 12/26/2006 132 sur 232
GCS - TM - 0245 – 2006/12/24 - lesson 5 - p.1
VIDAS Connection
Lesson 5BCI RS232 Software
Content
• 5.1 Overview of BCI RS232 Software• 5.2 Send / Receive data with VIDAS PC threw BCI RS232 Software• 5.3 Installation of BCI Simulator Software• 5.4 Simulate LIS operation (Download/Upload)• 5.5 Send a result from mini VIDAS
Content
• 5.1 Overview of BCI RS232 Software• 5.2 Send / Receive data with VIDAS PC threw BCI RS232 Software• 5.3 Installation of BCI Simulator Software• 5.4 Simulate LIS operation (Download/Upload)• 5.5 Send a result from mini VIDAS
TM 0245 12/26/2006 133 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.1
VIDAS Connection Lesson 5 :BCI RS232 Software
Chapter 5.1Overview of BCI RS232 software
Objectives
• Know the different menus of the BCI RS232 software
Objectives
• Know the different menus of the BCI RS232 software
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.2
Supervision Menu
TM 0245 12/26/2006 134 sur 232
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.3
Transaction Logs Menu
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.4
Link Operations Menu
By default, the History option is selected.
TM 0245 12/26/2006 135 sur 232
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.5
Link Operations Menu
The real time function allows to check the
connection between AI and LIS
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.6
BCI Configuration Menu
TM 0245 12/26/2006 136 sur 232
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.7
How to run BCI RS232 Software
Start or Restart manually BCI RS232 software :
• Double click on BCI RS232 icon
BCI RS232 software starts
Enter “Login”and “Password”
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.8
How to search Analysis Request
TM 0245 12/26/2006 137 sur 232
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.9
How to acknowledge an Error
Chap 5.1 Overview of BCI RS232 Software
GCS - TM - 0245 – 2006/12/24 - chap 5.1 - p.10
How to acknowledge an Error
TM 0245 12/26/2006 138 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 5.2 - p.1
VIDAS Connection Lesson 5 :BCI RS232 Software
Chapter 5.2Send / Receive data with VIDAS PC
threw BCI RS232 software
Objectives
• Know how to send and receive data with VIDAS PC threw BCI RS232 software
Objectives
• Know how to send and receive data with VIDAS PC threw BCI RS232 software
Chap 5.2 Send / Receive data with VIDAS PC threw BCI RS232 software
GCS - TM - 0245 – 2006/12/24 - chap 5.2 - p.2
Function : Receive an analysis request
RS232Cable
VIDASPC01VIDASPC01
LISLIS
TM 0245 12/26/2006 139 sur 232
Chap 5.2 Send / Receive data with VIDAS PC threw BCI RS232 software
GCS - TM - 0245 – 2006/12/24 - chap 5.2 - p.3
Function : Receive an analysis request
If the test is not calibrated, automatically the VIDAS PC software add Cx + Sx
Chap 5.2 Send / Receive data with VIDAS PC threw BCI RS232 software
GCS - TM - 0245 – 2006/12/24 - chap 5.2 - p.4
Function : Send a result from VIDAS PC
RS232Cable
VIDASPC01VIDASPC01
LISLIS
TM 0245 12/26/2006 140 sur 232
Chap 5.2 Send / Receive data with VIDAS PC threw BCI RS232 software
GCS - TM - 0245 – 2006/12/24 - chap 5.2 - p.5
Function : Send a result from VIDAS PC
Send automatically a result : depends on the VIDAS PC Configuration menu
(re)Send manually a VIDAS PC result via the result menu
Chap 5.2 Send / Receive data with VIDAS PC threw BCI RS232 software
GCS - TM - 0245 – 2006/12/24 - chap 5.2 - p.6
Function : Send a result from VIDAS PC
Check in the link operation menu of BCI RS232 software that the result has been correctly sent from BCI to LIS, and so from VIDAS PC to LIS
TM 0245 12/26/2006 141 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.1
VIDAS Connection Lesson 5 :BCI RS232 Software
Chapter 5.3Install BCI Simulator software
Objectives
• Install, set and check BCI simulators software
Objectives
• Install, set and check BCI simulators software
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.2
BCI Simulator software installation
“BCI Simulators” is used with both BCI RS232 and BCI Net, and includes two simulators :
• an LIS simulator to send analysis requests to BCI software
• an AI simulator to send analysis results to BCI software
TM 0245 12/26/2006 142 sur 232
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.3
BCI Simulator software installation
Technical specifications
PC Pentium III – 128 Mo RAMSerial port / Network port
Hardware
Compliant with : BCI RS232 R2.1.0 (or higher) and BCI Net R1.1.0
Software
Login : bciPassword : bci
User & Pwd
Supported by :Windows® XP SP1, 2000 SP2 and NT4
Operating system
Appearance set to Windows Classic
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.4
BCI Simulator software installation
Insert CD IMM13 and go to “BCI Simulator”Run “j2re-1_4_1_07-windows-i586-i.exe” program located in “BCI SIMULATORS R2.0.0\java”Accept Java License
Select
TM 0245 12/26/2006 143 sur 232
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.5
BCI Simulator software installation
Run “install.exe” program located in “BCI SIMULATORS R2.0.0\simulators”
Select the installation folder of BCI Simulators
Rk : Zip labels may be in French
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.6
BCI Simulator software installation
Select the file “environment_***.reg” (*** depends on O.S.) located in\BCI SIMULATORS\simulators.
Double-click on it to import the new values
Close the current Windows® session and open a new one or restart the PC
Run “init.bat” file located in C:\BCI SIMULATORS* \bin\ (*=default)
All necessary « Upload » and « Download » folders are created automatically
TM 0245 12/26/2006 144 sur 232
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.7
BCI Simulator Setting
Simulator.BCIServer = BCI Net hostname (or @IP)
Simulator.BCIPort = FTP port set in BCI Net software
Simulator.User = login set in BCI Net software
Simulator.Pwd = password set in BCI Net software
The ftp.properties file gives to BCI Simulators software the necessary parameters to simulate LIS connection properly.
To modify it :Right click on the file nameEdit with Wordpad®
Location : \BCI SIMULATORS\lib\properties\simulator\
To simulate a LIS :
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.8
BCI Simulator Setting
To simulate a AI :
JAVA_HOME"="C:\\Program Files\\Java\\j2re1.4.1_07
NAMESERVICE_HOST"="localhost"
NAMESERVICE_PORT"="29000"
Java Path
Set Port value to :• 29000, 19005 or 18001 for BCI Net• 14000 for BCI RS232
Replace localhost by hostname or @IP of PC hosting BCI Net
or BCI RS232
TM 0245 12/26/2006 145 sur 232
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.9
BCI Simulator launching
Double click on “run.bat” file to launch BCI Simulators software, and log as “bci”
pwd : “bci”
Chap 5.3 Install BCI Simulator software
GCS - TM - 0245 – 2006/12/24 - chap 5.3 - p.10
Troubleshooting Installation
PB to connect to BCI Net software :Modify “run.bat” file to stop the firewall service
PB to access to serial port :Modify “run.bat” file to write the complete way to access to java software
Run.bat :net stop sharedaccess@echo offecho Compatible with CosNAming SUN JDK1.4.1 and CosNaming
Visibroker 5.0echo Variable NAMESERVICE_HOST : %NAMESERVICE_HOST%echo Variable NAMESERVICE_PORT : %NAMESERVICE_PORT%Set SIMULATOR_ROOT=..@echo on"C:\Program Files\Java\j2re1.4.1_07\bin\javaw" -classpath …..pausenet start sharedaccess
TM 0245 12/26/2006 146 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.1
VIDAS Connection Lesson 5 :BCI RS232 Software
Chapter 5.4BCI Simulator overview
Objectives
Know the different menus of the BCI Simulator software
Objectives
Know the different menus of the BCI Simulator software
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.2
BCI Simulator Software Overview
11/ LIS BCI RS232
2
2/ LIS BCI Net
3
3/ AI Simu
Consultation Zone of XML
message Consultation Zone of error message
Edition Zone of XML message
RS 232 <> FTP ⇒ change of environment
TM 0245 12/26/2006 147 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.3
BCI Simulator Software Overview
New message
Load message / ScriptSave
Save as …
Send … to BCI• Results• Requests
Refresh directory / Reload FTP config
Upload results from BCI to LIS
Connect / Disconnect AI
Send AI properties
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.4
LIS BCI RS232
LIS BCI RS232 is used to simulate a LIS to :Send an analysis request to VIDASCheck that the Download process runs correctlyReceive a result from VIDASCheck that the Upload process runs correctly
TM 0245 12/26/2006 148 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.5
LIS BCI RS232 Configuration
Set serial connection
Local PC Serial port !
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.6
LIS BCI RS232 Status
Make sure that the Status is “READY”
TM 0245 12/26/2006 149 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.7
LIS BCI RS232 : How to create an Analysis Request
Right click to create
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.8
LIS BCI RS232 : How to create an Analysis Request
Enter all information, then click on “OK”
TM 0245 12/26/2006 150 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.9
LIS BCI RS232 : How to create an Analysis Request
Save into “lis_worklist\bioMerieux” directory
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.10
LIS BCI RS232 : How to send an Analysis Request
Send the message to BCI RS232
TM 0245 12/26/2006 151 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.11
LIS BCI RS232 : How to update an Analysis Request
Right click to update, duplicate or delete
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.12
LIS BCI RS232 : How to load and send an Analysis Request
To allow to modify rapidly a File.The Function Copy/Paste is running
in the screen.
TM 0245 12/26/2006 152 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.13
LIS BCI RS232
The Analysis Request created are saved in the «Lis_WorkList» directoryThe name of the file can be what you want, no rules.The sample separator is a mandatory fieldYou can modify the option field which appears on the selection
To do it, modify the content of these files
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.14
AI BCI RS232
AI Simulator is used to simulate a AI to :Receive an analysis request from LISCheck that the Download process runs correctlySend a result to LIS threw BCI RS232Check that the Upload process runs correctly
LIS
TM 0245 12/26/2006 153 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.15
AI BCI RS232
Connect the network cable between the PC hosted BCI RS232 and the PC used to simulate a AIModify Environment variables and restart the PC
NAMESERVICE_HOST"="localhost"
NAMESERVICE_PORT"="29000"
Set Port value to :• 29000, 19005 or 18001 for BCI Net• 14000 for BCI RS232
Replace localhost by hostname or @IP of
PC hosting BCI RS232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.16
AI BCI RS232
Instrument.properties (Located : \BCI SIMULATORS\lib\properties\simulator)
AI Name
Test Name
TM 0245 12/26/2006 154 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.17
AI BCI RS232
Special case : bmx_simulator• Without modify the LIS configuration, use bmx_simulator type in order to simulate analysis receipt or sending result.
• The simulator is managed by BCI Net as long as it is connected.
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.18
AI BCI RS232
TM 0245 12/26/2006 155 sur 232
Chap 5.4 BCI Simulator overview
GCS - TM - 0245 – 2006/12/24 - chap 5.4 - p.19
AI BCI RS232
TM 0245 12/26/2006 156 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 4.5 - p.1
VIDAS Connection Lesson 5 :BCI RS232 Software
Chapter 5.5Simulate download and upload operation
Objectives
Know how to simulate download and upload operation
Objectives
Know how to simulate download and upload operation
Chap 5.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 5.5 - p.2
Simulate LIS BCI RS232 Operation
On BCI Simulators software :
• Set the LIS BCI RS232 configuration
• Start the communication
On BCI RS232 software :
• Set the link configuration
• Start the communication
On VIDAS PC software :
• Set the link configuration
TM 0245 12/26/2006 157 sur 232
Chap 5.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 5.5 - p.3
Simulate LIS OperationSend a request
On BCI Simulators software :
• Create an analysis request
• Save and send it
On BCI RS232 software :
• Select “real time” in Link Operation
• Check the communication
On VIDAS PC software :
• Check that the analysis request is correctly received
Chap 5.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 5.5 - p.4
Simulate LIS OperationReceive a result
On VIDAS PC software :
• Send a result to the LIS
On BCI RS232 software :
• Select “real time” in Link Operation
• Check the communication
On the PC hosted the BCI Simulators software :
• Check that the result is received in the “LIS Result\bioMerieux” folder
TM 0245 12/26/2006 158 sur 232
Chap 5.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 5.5 - p.5
How to simulate a AI
Modify Environment variables on the PC hosted the BCI Simulator software and restart the PC :
NameServiceHost = Host name of the PC hosted BCI RS232NameServiceport = 18001 (If VIDAS PC)
Modify “Instrument.properties” file :The link name = link name set in BCI RS232 softwareFormat = bioMerieux
Start BCI Simulator software
Chap 5.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 5.5 - p.6
How to simulate a AI
Check that the communication is started on BCI RS232 software
Connect the AI simulated
Set in real time the link operation in BCI RS232 software
TM 0245 12/26/2006 159 sur 232
Chap 5.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 5.5 - p.7
How to send an Analysis Result
Create an analysis request, save and send it
Chap 5.5 Simulate download and upload operation
GCS - TM - 0245 – 2006/12/24 - chap 5.5 - p.8
How to check the reception
1st Step
2nd Step
TM 0245 12/26/2006 160 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 5.6 - p.1
VIDAS Connection Lesson 5 :BCI RS232 Software
Chapter 5.6Send a result from mini VIDAS
Objectives
• Know how to send a result from mini VIDAS
Objectives
• Know how to send a result from mini VIDAS
Chap 5.6 Send a result from mini VIDAS
GCS - TM - 0245 – 2006/12/24 - chap 5.6 - p.2
Send a result from mini VIDAS
If the LIS auto validation option is selected, it means that automatically all result will be sent to LIS. In this case, The technical validation is done i the LIS software
LIS Upload Enabled
Warn about Missing IDs
LIS auto validation
TM 0245 12/26/2006 161 sur 232
Chap 5.6 Send a result from mini VIDAS
GCS - TM - 0245 – 2006/12/24 - chap 5.6 - p.3
Send a result from mini VIDAS
The result can be sent or resent manually to LIS
Validate LIS Results
Reprint Results
Print Assay List
B 10:20 12 DecTXGA 12:20 12 DecTXGB 16:20 12 DecTXG
Status Screen
Results Menu
Utility Menu
Master Lot Menu
Start Section
TM 0245 12/26/2006 162 sur 232
GCS - TM - 0245 – 2006/12/24 - lesson 6 - p.1
VIDAS Connection
Lesson 6
Support end users
Content
• 6.1 Know the product and tools• 6.2 Identify the different levels to be trained• 6.3 Train end users on routine use• 6.4 Train end users how to manage and maintain the system• 6.5 Identify useful information for LIS vendors / End users
Content
• 6.1 Know the product and tools• 6.2 Identify the different levels to be trained• 6.3 Train end users on routine use• 6.4 Train end users how to manage and maintain the system• 6.5 Identify useful information for LIS vendors / End users
TM 0245 12/26/2006 163 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 6.1 - p.1
VIDAS Connection Lesson 6 :Support end users
Chapter 6.1Know the product and tools
Objectives
• Know the product and tools :• Manuals• Reference• Product list• Teaching scenario
• Know the support tools from GCS
Objectives
• Know the product and tools :• Manuals• Reference• Product list• Teaching scenario
• Know the support tools from GCS
Chap 6.1 Know the product and tools
GCS - TM - 0245 – 2006/12/24 - chap 6.1 - p.2
Products and tools
Manuals :BCI RS232 User Manual : Ref 38091BCI Net User Manual : Ref 99605Connectivity Service Manual : Ref 99604
Software :BCI Net Pack simu : Ref 99603BCI RS232 / BCI Net already installed in the VIDAS/VIDIA Master
Teaching Scenario on GCS portal
TM 0245 12/26/2006 164 sur 232
Chap 6.1 Know the product and tools
GCS - TM - 0245 – 2006/12/24 - chap 6.1 - p.3
Support tools from GCS
Global overview of the Global Customer Servicehttps://gcs.biomerieux.fr/tc/Welcoming
booklet/GCS Training Information booklet 2006.pdf.
FAQ / GCS Info : see PortalMovex Code :
Software code : BCI RS232 C0913BCI Net C0035
TM 0245 12/26/2006 165 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 6.2 - p.1
VIDAS Connection Lesson 6 :Support end users
Chapter 6.2Identify the different levels to be trained
Objectives
• Identify the different levels to be trained
Objectives
• Identify the different levels to be trained
Chap 6.2 Identify the different levels to be trained
GCS - TM - 0245 – 2006/12/24 - chap 6.2 - p.2
Different levels required on BCI
Two different levels :
BCI_ROUNTINE group :Know the BCI Software principleKnow how to stop and start BCI SoftwareKnow the routine use
BCI_ADMIN group :Know and understand the BCI_ADMIN group functionsKnow how to check a transactionKnow how to set the connection on BCI SoftwareKnow how to solve main problems on the connection
TM 0245 12/26/2006 166 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 6.2 - p.1
VIDAS Connection Lesson 6 :Support end users
Chapter 6.3Train end users on routine use
Objectives
• Understand the teaching scenario• Explain the whole system in connection mode :
• Show the connection in routine use• Explain the use of BCI software• Explain the use of VIDAS / mini VIDAS Software
• Confirm the knowledge of the customer
Objectives
• Understand the teaching scenario• Explain the whole system in connection mode :
• Show the connection in routine use• Explain the use of BCI software• Explain the use of VIDAS / mini VIDAS Software
• Confirm the knowledge of the customer
Chap 6.3 Train end users on user level
GCS - TM - 0245 – 2006/12/24 - chap 6.3 - p.2
Tools provided from GCS
Teaching scenario
Program
Material used
Method
Timing
See in the GCS Portal
Translate these information to your local market / Local rules
TM 0245 12/26/2006 167 sur 232
Chap 6.3 Train end users on user level
GCS - TM - 0245 – 2006/12/24 - chap 6.3 - p.3
Demonstration
Show the connection in routine useExplain the use of BCI softwareExplain the use of VIDAS / mini VIDAS Software
TM 0245 12/26/2006 168 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 6.4 - p.1
VIDAS Connection Lesson 6 :Support end users
Chapter 6.4Train end users how to manage
and maintain the system
Objectives
• Understand the teaching scenario• Explain the system setup in connection mode :
• Explain the setup of BCI software• Explain the setup of VIDAS / mini VIDAS Software
• Explain how to solve basic connection problems• Confirm the knowledge of the customer
Objectives
• Understand the teaching scenario• Explain the system setup in connection mode :
• Explain the setup of BCI software• Explain the setup of VIDAS / mini VIDAS Software
• Explain how to solve basic connection problems• Confirm the knowledge of the customer
Chap 6.4 Train end users how to manage and maintain the system
GCS - TM - 0245 – 2006/12/24 - chap 6.4 - p.2
Teaching scenario
Teaching scenario
Program
Material used
Method
Timing
See in the GCS Portal
Translate these information to your local market / Local rules
TM 0245 12/26/2006 169 sur 232
Chap 6.4 Train end users how to manage and maintain the system
GCS - TM - 0245 – 2006/12/24 - chap 6.4 - p.3
Demonstration
Explain the system setup in connection mode (BCI / VIDAS / mini VIDAS)
Explain how to solve basics connection problems
TM 0245 12/26/2006 170 sur 232
GCS - TM - 0245 – 2006/12/24 - chap 6.5 - p.1
VIDAS Connection Lesson 6 :Support end users
Chapter 6.5Identify useful information for LIS
Objectives
• Identify useful information for LIS
Objectives
• Identify useful information for LIS
Chap 6.5 Identify useful information for LIS
GCS - TM - 0245 – 2006/12/24 - chap 6.5 - p.2
Information for LIS
For each reagent :Test nameUnitInterpretationThresholdTest Value
For specific reagents :HIV5HBSAG
For each modification of these parameters
TM 0245 12/26/2006 171 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #07 Saving : Always use “Save as”, never “Save”
Steered Work :
Run “BCI Simulators” Click on “new Worklist as xml schema”
Create a Worklist with :
o ID patient o Patient name o Birth date o Sex o Location o Specimen Id o Test o Dilution
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
Report :
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 172 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #08 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Systematic dispatching mode - 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_0.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_0.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_0.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_0.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 173 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #08 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 174 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #09 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Systematic dispatching mode - 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_1.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_1.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_1.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_1.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 175 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #09 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 176 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #10 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Systematic dispatching mode - 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_2.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_2.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_2.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_2.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 177 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #10 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 178 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #11 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Systematic dispatching mode - 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_3.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_3.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_3.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_3.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 179 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #11 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 180 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #12 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Systematic dispatching mode - 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_4.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_4.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_4.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_4.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 181 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #12 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 182 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #13 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- bioMérieux dispatching mode o HBE : Rerun o HCG Rerun + Routine o FER : Routine o RBG : Rerun + Routine
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_5.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_5.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_5.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_5.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 183 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #13 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 184 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #14 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- bioMérieux dispatching mode o HBE : Rerun o HCG Rerun + Routine o FER : Routine o RBG : Rerun + Routine
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_6.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_6.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_6.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_6.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 185 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #14 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 186 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #15 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- bioMérieux dispatching mode o HBE : Rerun o HCG Rerun + Routine o FER : Routine o RBG : Rerun + Routine
- 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_7.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_7.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_7.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_7.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 187 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #15 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 188 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #16 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Laboratory dispatching mode - 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_8.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_8.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_8.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_8.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 189 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #16 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 190 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #17 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Laboratory dispatching mode - 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_9.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_9.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_9.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_9.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 191 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #17 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 192 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #18 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Laboratory dispatching mode - 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_10.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_10.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_10.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_10.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 193 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #18 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 194 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #19 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Laboratory dispatching mode - 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_11.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_11.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_11.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_11.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 195 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #19 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 196 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #20 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Laboratory dispatching mode - 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_12.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_12.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_12.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_12.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 197 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #20 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 198 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #21 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Laboratory dispatching mode - 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_13.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Result Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_result\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_13.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_13.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_13.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 199 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #21 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 200 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #22 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI Net software :
- Laboratory dispatching mode - 2 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “LIS_061224_000000_14.xml” from your CD ROM, folder CD IMM07\Training documents\Frames\Result Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_result\astm_xml\. (Windows Explorer)
Rename the frame : “LIS_061224_000000_14.xml” to “Your-FTPAccount_YYMMDD_HHMMSS_14.xml”
Run “BCI Simulators”
Load “Your-FTPAccount_YYMMDD_HHMMSS_14.xml” frame from “LIS Simulator
FTP” and “XML Worklist management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI NET (Transactions Menu)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 201 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #22 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received successfully by BCI Net software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 202 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #30 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW30.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW30.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 203 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #30 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 204 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #31 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW31.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW31.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 205 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #31 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 206 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #32 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW32.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW32.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 207 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #32 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 208 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #33 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW33.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW33.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 209 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #33 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 210 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #34 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW34.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW34.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 211 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #34 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 212 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #35 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW35.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW35.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 213 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #35 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 214 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #36 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW36.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW36.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 215 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #36 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 216 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #37 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW37.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW37.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 217 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #37 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 218 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #38 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW38.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW38.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 219 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #38 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 220 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #39 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW39.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW39.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 221 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #39 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 222 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #40 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW40.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW40.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 223 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #40 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 224 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #41 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW41.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW41.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 225 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #41 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 226 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #42 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW42.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Requests Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_worklist\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW42.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 227 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #42 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 228 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #43 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW43.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Result Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_result\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW43.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 229 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #43 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 230 sur 232
VIDAS Connection Training Course – IMM 07 Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #44 Saving : Always use “Save as”, never “Save”
Configuration : VIDAS PC Software :
- Complete demographic - Automatic validation - Make sure HBE, HCG, FER, RBG are activated.
BCI RS232 Software :
- 1 VIDAS PC connected - 1 LIS set and connected
Steered Work :
Insert the CD Rom CD IMM07 on the PC used to simulate LIS
Copy the frame “SW44.txt” from your CD ROM, folder CD IMM07\Training documents\Frames\Result Frames\3-Troubleshooting into your hard disk folder \BCI SIMULATORS\lis_result\bioMérieux\. (Windows Explorer)
Run “BCI Simulators”
Load “SW44.txt” frame from “LIS Simulator RS232” and “bioMérieux Worklist
management”,
Send this analysis request from “BCI Simulators”.
Check its reception by BCI RS232 (Real Time)
Check that analysis request appears on the Job List menu of VIDAS PC
Clear VIDAS PC work list
TM 0245 12/26/2006 231 sur 232
VIDAS Connection Training Course – IMM 07
Steered Header : No change unless indicated Work Frames : Only use the frames as indicated, follow naming rules #44 Saving : Always use “Save as”, never “Save”
Analysis request send successfully from BCI Simulators : Yes No
Analysis request received by BCI RS232 software : Yes No
Analysis request appears on VIDAS PC Software (Job list) : Yes No
Pop up appears on VIDAS PC software : Yes No
Error message logged on Event Viewer of VIDAS PC : Yes No
Message send by VIDAS PC to LIS : Yes No
If Yes : ________________________________ Status X : Yes No
Symptom and correction :
Remarks :
TM 0245 12/26/2006 232 sur 232