Upload
cirnatu-dana
View
240
Download
0
Embed Size (px)
Citation preview
8/12/2019 2013_BestPractices for HLD(1)
1/53
Best Practices forHigh-Level Disinfection
(HLD)
8/12/2019 2013_BestPractices for HLD(1)
2/53
Objectives
At the end of this program, participants will be able to:
explain the Spaulding classification system for thereprocessing of reusable medical devices,
understand the importance of effective cleaning anddecontamination prior to HLD,
identify commonly used agents for HLD,
discuss SGNA recommended steps for the HLD of
flexible endoscopes.
8/12/2019 2013_BestPractices for HLD(1)
3/53
Spaulding Classification System
In 1968, Dr. Earle Spaulding devised a rationalapproach to disinfection and sterilization that isstill in use today. He believed that instrumentsand equipment should be reprocessed accordingto the nature of the item and the level of riskassociated with their intended use.
This is referred to as Spaulding's classification system
and it has been refined and retained over the years,because it is so clear and logical. The three (3) categorieshe described were critical, semi-critical and non-critical.
8/12/2019 2013_BestPractices for HLD(1)
4/53
Spaulding Classification System
Critical items are medical devices that enter sterile tissueor the vascular system. These items should be sterilewhen used. Examples include, but are not limited to:surgical instruments, cutting endoscopic accessories that
break the mucosal barrier, endoscopes used in sterile bodycavities, cardiac, vascular or urinary catheters, implants,needles and ultrasound probes used in the sterile bodycavities.
http://www.google.com/url?sa=i&rct=j&q=cardiac+catheters&source=images&cd=&cad=rja&docid=tBMvuXVzzK4jdM&tbnid=U-WbhaHRxUGt8M:&ved=0CAUQjRw&url=http://www.fda.gov/ohrms/dockets/ac/04/briefing/4018B1_01.htm&ei=BJVRUaG9IYT30gGppYHQBw&bvm=bv.44158598,d.dmg&psig=AFQjCNGepJuUF8ma6nfI05X34oKXwsIs3w&ust=13643874458933378/12/2019 2013_BestPractices for HLD(1)
5/53
Spaulding Classification System
Semi-critical items are medical devices that come intocontact with non-intact skin or mucous membranes.These items should be high level disinfected when used.Examples include, but are not limited to: vaginal andrectal probes, anesthesia equipment, laryngoscopes,
bronchoscopes and gastrointestinal endoscopes (andaccessories).
8/12/2019 2013_BestPractices for HLD(1)
6/53
Spaulding Classification System
Non-critical items are medical devices that come intocontact with only intact skin. These items should receiveintermediate level disinfection, low-level disinfection orcleaning. Intact skin acts is considered an effective
barrier to most organisms. Examples of non-critical iteminclude, but are not limited to: tourniquets and blood
pressure cuffs, linens, bed pans and stethoscopes.
8/12/2019 2013_BestPractices for HLD(1)
7/53
The terminology adopted by the CDC and widely used,describes disinfectants in terms of their activity as set out
below. This program will focus on high-level disinfection(HLD).
High-level disinfectantsare chemical sterilants, which whenused for a shorter exposure period than would be requiredfor sterilization, kill all microorganisms with the exceptionof high numbers of bacterial spores.
Intermediate-level disinfectantsmay kill mycobacteria,vegetative bacteria, most viruses, and most fungi but do notnecessarily kill bacterial spores
Low-level disinfectantsmay kill most vegetative bacteria,some fungi, and some viruses.
Three Levels of Disinfection
http://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asphttp://www.health.qld.gov.au/EndoscopeReprocessing/glossary.asp8/12/2019 2013_BestPractices for HLD(1)
8/53
Pre-cleaning must occur at point of use in order to keep
blood and other organic material from drying. Blood and
body fluids, as well as saline, are highly corrosive and can
damage instruments. Dried blood and debris is difficult, if
not impossible to clean during decontamination, which cancause disinfection or sterilization not to be achieved.
Surgical instruments should be wiped as needed with
sterile surgical sponges moistened with sterile water duringthe procedure to remove gross soil. Instruments with
lumens should be irrigated with sterile water.
Cleaning and Decontamination
8/12/2019 2013_BestPractices for HLD(1)
9/53
Flexible endoscopesused on the sterile field should be
pre-cleaned to the following recommended steps:
1. Wipe external surfaces with a lint-free cloth saturated
with sterile water.2. Alternate suctioning the channels with sterile water
and air.
3. Hand the endoscope and accessories to the circulatoras soon as possible, so he/she may pre-clean (follow
the steps on the next slide).
Cleaning and Decontamination
8/12/2019 2013_BestPractices for HLD(1)
10/53
Flexible endoscopesused in GI/Endoscopy Centers
should be pre-cleaned in the Procedure Room wearing
appropriate PPE and using the following recommended
steps:
1. External surface of insertion tube should be cleanedwith a soft cloth or sponge and an enzymatic detergent.
2. Internal suction/biopsy channels cleaned by suctioning
copious amounts of enzymatic detergent and air.
3. Air/water channels flushed with enzymatic solution,
then flushed using low-pressure compressed air or a
syringe if air is not available.
Cleaning and Decontamination
8/12/2019 2013_BestPractices for HLD(1)
11/53
4. Complex design components or channels should be
flushed or purged with water and/or enzymatic
detergent solution per the MFGs IFU.
5. The tip of the endoscope should be inspected for
damage to any surface and any working part, and forcleanliness.
6. The video protective cap (if applicable) should be
attached after removing the endoscope from the light
source and suction.
7. Remove all detachable parts and immerse in enzymatic
detergent solution until transport.
Cleaning and Decontamination
8/12/2019 2013_BestPractices for HLD(1)
12/53
Cleaning and Decontamination
Flexible endoscopes are someof the most complex devices
for health care workers to
reprocess, due to their unique
design and multiple reprocessingsteps.
Strict adherence to the scope
MFGs validated instructions
for use (IFU) is required.
8/12/2019 2013_BestPractices for HLD(1)
13/53
After pre-cleaning at point of use, contaminated itemsshould be immediately transported to the decontaminationarea before any remaining organic material dries on thesurface, box lock, crevices or channels of the instruments.Surgical instruments should be covered with a wet towel
or treated with an instrument cleaner prior to transport.
Contaminated items can expose health care workers andcan contaminate the environment during transport. A
sealed container should be used to avoid contaminatingthe environment or exposure to health care workers. PerOSHA, the transport container must be labeled to indicate
biohazardous contents.
Transport
8/12/2019 2013_BestPractices for HLD(1)
14/53
Meticulous cleaning and decontamination must proceed
HLD or sterilization. Medical device manufacturers are
required to provide healthcare facilities with validated
reprocessing instructions for use (IFU).
It is important to have and follow each device MFGs IFU
to ensure worker and patient safety. This is especially true
for complex devices with multiple pieces that have joints,
crevices, lumens, ports and channels. Proper PPE isrequired for all cleaning and decontamination steps.
Cleaning and Decontamination
8/12/2019 2013_BestPractices for HLD(1)
15/53
Pasteurization is a heat-automated HLD process that uses
time and heat (i.e. 160-170F/21.7-25C) for 30 minutes
for HLD of heat-sensitive semi-critical devices. Medical
washer/pasteurizers have wash, rinse and pasteurization
cycles. Some pasteurizers offer quality assurance datarecorders that document the temperature and cycle time.
Heat HLD
8/12/2019 2013_BestPractices for HLD(1)
16/53
For chemical HLD, health care facilities must purchase
FDA cleared HLD products as listed on the FDA website.
Products selected should be compatible and efficacious
with the materials or items to be disinfected.
The use of incompatible chemicals can damage the
surfaces of the instrument, causing corrosion, scratches
and other surface irregularities. Such damage can be a
challenge for cleaning, HLD, interfere with properfunction, and reduce the life and cosmetic appearance
of the device.
Chemical HLD
8/12/2019 2013_BestPractices for HLD(1)
17/53
FDA-cleared Chemical sterilants and HLDs
Device Type # of Registered
Products
Chemical
Sterilization
High Level
Disinfection
Glutaraldehyde 17 X X
Hydrogen Peroxide 4 X X
Ortho-phthaldehyde
(OPA)
3 X
Peracetic Acid 2 X X
Sodium Hypochlorite 1 X
Chemical Vapor
w/Formaldehyde
2 X
Hydrogen Peroxide GasPlasma
1 X
Hydrogen Peroxide
without Plasma
1 X
Ozone Gas 1 X
Source: www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/ReprocessingofSingle-UseDevices/ucm133514.htm
8/12/2019 2013_BestPractices for HLD(1)
18/53
HLD requires appropriate temperature, contact time, and
length of use following solution activation. MFGs IFU
should be followed when preparing disinfectant solutions,
calculating expirations dates, and labeling solution soaking
containers.
Chemical HLD
8/12/2019 2013_BestPractices for HLD(1)
19/53
Glutaraldehydehas been widely used for a long time in
health care facilities as a HLD for reusable medical devices.
Most solutions are acidic and must be activated to become
sporicidal. There are a variety of brand names available in
a variety of concentrations, with and without surfactants.
Chemical HLD
HLD and Steriliant
requires no mixing or
activation
8/12/2019 2013_BestPractices for HLD(1)
20/53
Ortho-phthaladehyde (OPA) has demonstrated superior
mycobactericidal activity compared to glutaraldehyde and
requires no mixing or activation. OPA has been shown to
last longer before reaching its MEC and the concentration
of the active ingredient does not decrease with age alone.
Chemical HLD
8/12/2019 2013_BestPractices for HLD(1)
21/53
Other solutions FDA-cleared for HLD include hydrogen
peroxide, peracetic acid and sodium hypochlorite in a
variety of concentrations and combinations. The FDA
website has a listing of manufacturers, active ingredients
and contact conditions for each cleared solution.
Chemical HLD
8/12/2019 2013_BestPractices for HLD(1)
22/53
Because most HLDs are reused, they must be tested and
recorded prior to each use to assure that they remain
above their MEC. If the test strip fails, the HLD solution
should not be used, even if its within the reuse life.
Chemical HLD
8/12/2019 2013_BestPractices for HLD(1)
23/53
Reprocessing of Flexible Endoscopes
Chemical HLD is recognized as the standard for the
reprocessing of flexible gastrointestinal endoscopes by
SGNA, ASGE, ACG, AGA, APIC and AST. Also, the
CDC and The Joint Commission recognize HLD as
appropriate for gastrointestinal endoscopes.
8/12/2019 2013_BestPractices for HLD(1)
24/53
Standards of Infection Control in
Reprocessing of Flexible
Gastrointestinal Endoscopes
Are you following the updated 2012 SGNA Standards?
8/12/2019 2013_BestPractices for HLD(1)
25/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
5. Manual HLD:
1. Completely immerse the endoscope and all
removable parts in a basin of HLD.a. The basin must be of a size to accommodate the
endoscope without undue coiling, and must have a
tight-fitting lid to contain the chemical vapors.
b. To prevent damage, the endoscope should be notbe soaked with other sharp instruments.
8/12/2019 2013_BestPractices for HLD(1)
26/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
5. Manual HLD
2. Flush disinfectant into all channels of the
endoscope until it can be seen exiting the oppositeend of each channel. Take care that all channels are
filled with the chemical, and that no air pockets
remain within the channels.
8/12/2019 2013_BestPractices for HLD(1)
27/53
SGNA Standards of Infection Control in
Reprocessing of Flexible Endoscopes
5A. Manual HLD:
a. Complete microbial destruction cannot occur
unless all surfaces are in complete contact with thechemical.
b. Since internal contact cannot be visually confirmed
because of scope design, purging until a steady flow
of solution observed is necessary.
8/12/2019 2013_BestPractices for HLD(1)
28/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
5A. Manual HLD:
3. Cover the soaking basin with a tight-fitting lid
to minimize chemical vapor exposure.
Note that:
a. Exposure to chemical vapors may present a
health hazard.b. The reprocessing area should have engineering
controls to ensure good air quality.
8/12/2019 2013_BestPractices for HLD(1)
29/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
5A. Manual HLD:
4. Soak the endoscope in the HLD solution for
the time/temperature required to achieve HLD.Use a timer to verify soaking time.
5. Purge all channels completely with air before
removing the endoscope from the HLD solution.Note that purging the channels preserves the
concentration and volume of the chemical, and
prevents exposure from dripping and spilling.
8/12/2019 2013_BestPractices for HLD(1)
30/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
5A. Manual HLD:
6. RINSE(same as after Manual Cleaning)
a. Thoroughly rinse the endoscope and all removableparts with clean water to remove residual debris and
detergent.
b. Purge water from all channels using forced air.
Dry the exterior of the endoscope with a soft, lint-free cloth to prevent dilution of the liquid chemical
germicide used in subsequent steps.
8/12/2019 2013_BestPractices for HLD(1)
31/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
6. Drying:
a. Purge all channels with air until dry.
Note that:1) Bacteria such asPseudomonas aeruginosa have beenidentified in both tap and filtered water, and may multiplyin a moist environment.
2) Avoid the use of excessively high air pressure whichcan damage the internal channels of flexible endoscopes.
8/12/2019 2013_BestPractices for HLD(1)
32/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
6. Drying:
b. Flush all channels, including accessory channels,with alcohol until the alcohol can be seen exiting the
opposite end of each channel.1) 70% isopropyl alcohol is used to assist in drying theinterior channel surfaces.
2) It must be properly stored in a closed container
between uses, because when exposed to air, it rapidlyevaporates, and if less than recommended % level,cannot be relied upon to assist in the drying process.
3) Alcohol flushes should be used even when sterile
water is used for rinsing.
8/12/2019 2013_BestPractices for HLD(1)
33/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
6. Drying:
c. Purge all channels with air. Note that alcohol
mixes with the remaining water on the channel
surfaces and acts to encourage evaporation of the
residual water as air flows through the channel.
d. Remove all channel adapters.
8/12/2019 2013_BestPractices for HLD(1)
34/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
6. Drying:
e. Dry the exterior of the endoscope with a soft,
clean lint-free towel.f. Thoroughly rinse and dry all removable parts.
Do not attach removable parts (e.g. valves, etc.)
to the endoscope during storage as this can trap
liquid inside.
8/12/2019 2013_BestPractices for HLD(1)
35/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
7. Storage:
Hang the endoscope in a vertical position to facilitate
drying (with caps, valves, and other detachable componentsremoved, per MFGs IFU).
a. The storage area should be clean, well
ventilated and dust free.
b. Correct storage will prevent damage.c. The interval of storage before use has
limited investigations and warrants further
data.
8/12/2019 2013_BestPractices for HLD(1)
36/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
5B. Automated HLD:
Automated Endoscope Reprocessors (AERs)
standardize the disinfection process and decreasepersonnel exposure to HLDs.
NOTE:
It is necessary to followall steps for the manual
cleaning prior to using
an AER.
8/12/2019 2013_BestPractices for HLD(1)
37/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
5B. Automated HLD:
An AER should have the following features:
a. Circulate fluids through all endoscope channels at an equalpressure without trapping air. Channel flow sensors providean added measure of compliance.
b. Detergent and disinfectant cycles should be followed bythorough rinse cycles and forced air to remove all used
solutions.
c. Disinfectant should not be diluted with any fluids.
8/12/2019 2013_BestPractices for HLD(1)
38/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
An AERshould have the following features:
d. Machine should be self-disinfecting.
e. No residual water should remain in hoses and reservoirs.f. Cycles for alcohol flushing and forced air drying are desirable.
g. Should also feature a self-contained or external water
filtration system.
In addition, a method to automatically store or print data
verification of cycle completion, is desirable.
8/12/2019 2013_BestPractices for HLD(1)
39/53
SGNA Standards of Infection Control
in Reprocessing of Flexible Endoscopes
To use an AER:
1. Follow steps for manual cleaning of endoscope.
2. Prepare the AER according to the MFGs guidelines.3. Place the endoscope in the AER and attach all channeladapters according to the MFGs IFU.
a. The elevator channel of a duodenoscope has a very small
lumen. Since most AERs cannot generate pressure required toforce fluid through the lumen, a 2-5 ml syringe must be usedto manually reprocess (all steps) the elevator channel unlessthe AER is validated to perfuse this channel.
8/12/2019 2013_BestPractices for HLD(1)
40/53
To use an AER:
4. Place valves and other removable parts into thesoaking basin of the AER. Unless the AER has adedicated space for accessories, reprocess theseitems separately.
5. If the AER has a cycle that uses enzymaticdetergent, it should be a product that is compatiblewith the AER and the endoscope.
6. Set the AER for the appropriate time andtemperature depending on the chemical used.
7. Start the AER and allow it to complete allcycles or phases. Note that if cycles or phasesare interrupted, HLD cannot be ensured andthe full cycle must be repeated.
8/12/2019 2013_BestPractices for HLD(1)
41/53
To use an AER:
8. If AER does not include a final alcohol rinse,
this step should be done manually followed bypurging all the channels with air until dry.
The ERCP elevator and elevator channel must
be manually perfused and dried per MFGs
instructions.9. Drying and storage procedures are the same as
described in the manual disinfection section.
To get a free copy of the complete SGNA
document, go to: www.sgna.org
8/12/2019 2013_BestPractices for HLD(1)
42/53
Healthcare facilities are responsible for providing a safe
work and patient care environment. Patients, visitors, and
health care workers should be protected from injuries or
illnesses caused by hazardous chemicals.
When handling HLDs, personnel should wear protective
apparel that may include, but is not limited to:
100% nitrile rubber or 100% butyl rubber gloves when
handling glutaraldehyde. PVC gloves should not beworn because they absorb glutaraldehyde.
Protective eye wear, face mask, and impervious gown.
Chemical HLD Safety
8/12/2019 2013_BestPractices for HLD(1)
43/53
Glutaraldehydeshould only be used in well ventilated
areas or in freestanding or vented chemical fume hoods.Vapor generated from glutaraldehyde can may aggravate
preexisting respiratory conditions.
AAMIdescribes adequate ventilation as:1. Room large enough to ensure adequate dilution of vapors.
2. 10 air exchanges per hour.
3. Exhaust located at the source of the discharge of vapors.
4. Fresh air return at ceiling level across room from exhaust vents.5. Routine maintenance and surveillance of system.
6. Elimination of cross draft effects.
7. Air must not be recirculated.
Chemical HLD Safety
8/12/2019 2013_BestPractices for HLD(1)
44/53
Glutaraldehydecan be absorbed by inhalation, ingestion
and through the skin. It has a detectable odor at 0.04 parts
per million volume (ppmv) and is irritating to skin and
mucous membranes at 0.3 ppmv.
Vapors are released whenever solutions are disturbed and
the surface tension is broken, such as mixing, adding and
removing equipment, or disposing of a glutaraldehyde
solution can cause a break in the surface tension.Whenever the glutaraldehyde solution is not being
accessed, it should be covered with a tight-fitting lid.
Chemical HLD Safety
8/12/2019 2013_BestPractices for HLD(1)
45/53
8/12/2019 2013_BestPractices for HLD(1)
46/53
To avoid these glutaraldehyde issues, many health care
facilities have switched to using an OPAfor HLD. The
RapicideOPA/28 features the fastest disinfection time,
twice the reuse period of other OPA brands and guaranteed
materials compatibility.
Chemical HLD
8/12/2019 2013_BestPractices for HLD(1)
47/53
Exposure monitoring is not required; however, OPA is
still a potential irritant of eyes, skin, nose and other tissues
resulting in symptoms such as stinging, excessive tearing,
coughing, and sneezing. Like glutaraldehyde, OPAfixes
proteins, allows for biofilm formation and exposure causesstaining on linen, skin, instruments and AERs.
Chemical HLD Safety
8/12/2019 2013_BestPractices for HLD(1)
48/53
Personnel should receive initial training and competency
validation on procedures, chemicals used, and PPE and
should receive additional training when new equipment,
instruments, supplies, or procedures are introduced.
Employers must provide a written hazard communication
program, hazard evaluation, hazardous materials inventory,
Safety Data Sheets, labels on all containers of hazardous
chemicals, and employee training.
Training and Education
8/12/2019 2013_BestPractices for HLD(1)
49/53
A quality control program should be established in all
areas where HLD is used. Quality control programs
should be documented and should include, but not be
limited to:
Orientation programs Competency assurance
Continuing education
Quality control checks Investigation of adverse events
Monitoring of solution replacement intervals
Quality Control Program
8/12/2019 2013_BestPractices for HLD(1)
50/53
Surfaces of complex instruments require meticulous
cleaning in order to minimize infection control risks
to patients and staff. Inadequate cleaning can potentially
leave residual protein on a surface.
Quality control checks, such as
ATP systems and VERiFIND
Protein Detection Kit, provide
rapid and easy to readcleaning verification.
Quality Control Program
8/12/2019 2013_BestPractices for HLD(1)
51/53
Health care workers responsible for HLD must maintain
the technical skills needed to establish and maintain a safe
practice environment for patients, visitors and peers.
Administrative personnel must ensure competency
validation of personnel participating in decontamination
and HLD of reusable medical devices. The validation of
competencies should include all types of devices the
individual is authorized to reprocess.
CONCLUSION
8/12/2019 2013_BestPractices for HLD(1)
52/53
THANK YOU!
SPSmedical Supply Corp.
Sterilization Products & Services
6789 W. Henrietta Road
Rush, NY 14543 USA
Fax: (585) 359-0167
Ph: (800) 722-1529
website: www.SPSmedical.com
2013, SPSmedical Supply Corp.
Free CE programs are available in our
Sterilization Classroom on the
SPSmedical website.
8/12/2019 2013_BestPractices for HLD(1)
53/53
References & Resources
Association for the Advancement of Medical Instrumentation. (2010). Chemical sterilization and
high level disinfection in health care facilities (ANSI/AAMI ST58:2005/(R) 2010). Arlington, VA.
Association of periOperative Registered Nurses. (2013 Edition). Recommended Practices for
High-Level Disinfection.
Society of Gastroenterology Nurses and Association, Inc. (2007) Guideline for High-Level
Disinfectants & Sterilants for Reprocessing Flexible Gastrointestinal Endoscopes.
Society of Gastroenterology Nurses and Association, Inc. (2012) Standards of Infection Control
in Reprocessing of Flexible Gastrointestinal Endoscopes.
Occupational Health and Safety Administration (OSHA). (2012). Hazardous waste and emergency
response. www.osha.gov
Rutala, W. A., Weber, D. J., & the Healthcare Infection Control Practices Advisory Committee
(HICPAC). (2008). Guideline for Disinfection and Sterilization in Healthcare Facilities.