CHEMGENE HLD4L DISINFECTANT
Rebreather Cleaning Instructions & Safety Data Sheet
Chemgene HLD4L Disinfectant has been chosen for its ability to destroy a wide variety of bacteria, viruses and fungi including Legionella, Weil's disease,
Tuberculosis and HIV. If used properly Chemgene HLD4L also has a very low hazard risk and is odour free.
The 1 litre Chemgene HLD4L is supplied in a handy reservoir (squeeze and measure) bottle. Simply remove the lid and squeeze the container until the liquid
fills the reservoir, use the scale on each side of the reservoir to measure out the desired amount. The concentrated solution can then be diluted with cold potable water for pouring, spraying or soaking the equipment.
Recommended dilution ratios for cleaning a rebreather:
Situation Component Dilution Ratio
Water:Chemgene Suggested
Cleaning Method
Switching User All 10:1 Soak for 10 mins
Cleaning vomit, blood or excessive
mucus
All 10:1 Soak for 10 mins
End of days diving Mouthpiece
(including hoses) 100:1 Rinse through
After 6 hours of
use All 100:1 Rinse through
Prior to long term storage
All 100:1 Rinse through
General surface cleaning
All 100:1 Wipe with soaked cloth
IMPORTANT: After disinfecting always rinse off with plenty of cold potable
water until all the disinfectant has been washed away and allow to dry naturally.