MAK 3003 Sampling Stand Alone
Service Manual
SA 121107
Contents
C - 1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
Contents
Contents Page Issue date
1 Safety precautions _______________________________ 1-1General safety instructions __________________________ 1-1Installation instructions _____________________________ 1-2Wiring instructions_________________________________ 1-2Servicing and repair instructions______________________ 1-2
2 MAK 3003 Sampling Stand Alone system ____________ 2-12.1 Block diagram ____________________________________ 2-12.2 Pipe installation system ____________________________ 2-22.3 Wiring diagrams __________________________________ 2-4
3 Compact Controller TIGER MAK Type 6942-10 ________ 3-13.1 Technical data____________________________________ 3-13.2 Compact Controller dimensions ______________________ 3-23.3 Blade-type fuses and LED displays on TIGER 3003
(on motherboard) _________________________________ 3-33.4 CompactFlash card________________________________ 3-43.5 Exchanging the motherboard/CPU board_______________ 3-53.6 Compact Controller wiring___________________________ 3-63.6.1 Power supply_____________________________________ 3-73.6.2 Outputs _________________________________________ 3-73.6.3 Inputs __________________________________________ 3-7
4 Temperature Probe with Sensor Series B Type 6703-11 4-14.1 Technical data____________________________________ 4-14.2 Dimensions ______________________________________ 4-14.3 Assembly/installation ______________________________ 4-2
5 Dual Sampler Controller, Type 6970-30 ______________ 5-15.1 Technical data____________________________________ 5-15.2 Dimensions ______________________________________ 5-2
6 Quad Sampler Type 6900-20 _______________________ 6-16.1 Technical data____________________________________ 6-16.2 Dimensions ______________________________________ 6-2
7 TAG Reader / Writer Type 6900-101 _________________ 7-17.1 Technical data____________________________________ 7-17.2 Dimensions ______________________________________ 7-2
8 GPS Receiver (RS 232) Type 6722-18 ________________ 8-18.1 Technical data____________________________________ 8-18.2 Connection ______________________________________ 8-28.3 Installation and wiring ______________________________ 8-2
9 GPRS Module, Type 6942-100 ______________________ 9-19.1 Magnetic antenna _________________________________ 9-29.2 SIM card specifications for the 3003 system ____________ 9-2
10 Flow Level Meter Type 6826-x _____________________ 10-110.1 Technical data___________________________________ 10-110.2 Dimensions/installation ____________________________ 10-310.3 Wiring _________________________________________ 10-510.4 Electric terminal assignment________________________ 10-5
Contents
C - 2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
All rights reserved. Subject to change without notice.Any duplication, editing and circulation of this document,whether in whole or in part, is strictly subject to written consentfrom BARTEC BENKE.
Copyright © 2013 byBARTEC BENKE GmbHSchulstraße 30,D-94239 Gotteszell
Document: SA 121107 Valid as of: 45.12Revision: 26.07.2013Authors: M. Nagl/M. Meindl
Safety precautions
1-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
1 Safety precautions
General safety instructionsThe equipment operator is responsible for complying with all regulationsapplicable to storage, transport and handling.
All regulations and provisions shall remain in full force when operating theequipment with MAK units.
The units were built in compliance with the applicable regulations and left thefactory in perfect condition. The units must be installed and serviced by dulyskilled and qualified personnel.
Make sure that the information and operating conditions specified by themanufacturer are observed.
Follow the instructions on operating and servicing the units.
If you notice any damage or deterioration on parts of the equipment, or ifsafe operation cannot be guaranteed for other reasons, do not start upthe equipment or - if it is already running - switch it off immediately.Notify your service centre.
You should also contact our service technicians if you notice malfunctionsor defects during operation or if you have any doubts as to whether theunits are working properly.
The MAK units are not a replacement for the safety installations on themilk tanker or for the safety systems put in place by the customer.
BARTEC BENKE GmbH and its agents shall only be liable in case ofpremeditation or gross negligence. The extent of liability of such cases shallbe limited to the value of the order placed with BARTEC BENKE GmbH.BARTEC BENKE shall not be liable for damage arising from failure toobserve the safety precautions, operating instructions or operatingconditions. Consequential damages are excluded from the scope of liability.
Exclusion of liability
Safety precautions
1-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
Installation instructions
Install the units as necessary in order to meet the ambient temperaturespecifications. Use enclosures, heating or cooling systems to protectthem if necessary.
The installation location should be as vibration-free as possible. Usesturdy retaining fixtures to protect all the components from vibrations.
The printer must be installed in a permanently clean and dry location.
The power supply to the system must be stopped during weldingwork or assisted starting on the vehicle.
Take measures to keep the units clean during installation, especially theprinter (e.g. protect from metal cuttings, etc.).
Do not blank off unassigned cable glands.
Remove the transportation lock from the printer before putting it intooperation.
All built-in solenoid valves must be shielded from interference(suppressor diode).
Wiring instructions
The wiring must be undertaken by trained professionals.
The installation must be carried out in accordance with the relevantnational regulations.
The unit must be wired with exclusive reference to, and in strictcompliance with, the servicing instructions.
The connection cables must be fitted in such a way as to eliminate anysagging.An allowance of approx. 15 - 20 cm is advisable for attachment with clipsor cable ties.Extra care is needed when fitting the cables near the radiator, in theengine compartment, in parts of the truck frame and in the delivery cabin.
Under no circumstances should there be any vulnerable points where thecables are kinked or susceptible to abrasion.
Fit the cable ends with ferrules.
Tighten the screws on non-assigned terminals.
Servicing and repair instructions
Servicing and repairs must be carried out by trained professionals only.
Switch the units off before servicing and repair work, and take measuresto prevent them from being switched back on while servicing and repairwork is in progress.
MAK 3003 Sampling Stand Alone system
2-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
2 MAK 3003 Sampling StandAlone system
2.1 Block diagram
MAK 3003 Sampling Stand Alone system
2-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
2.2 Pipe installation system
Flow direction
MAK 3003 Sampling Stand Alone system
2-3
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
3003
Flo
wd
irection
MAK 3003 Sampling Stand Alone system
2-4
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
2.3 Wiring diagrams
MAK 3003 Sampling Stand Alone system
2-5
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
MAK 3003 Sampling Stand Alone system
2-6
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
MAK 3003 Sampling Stand Alone system
2-7
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
Flo
wLevel+
Milk
Sensor
MAK 3003 Sampling Stand Alone system
2-8
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
Compact Controller TIGER MAK Type 6942-10
3-1
MAK 3003 Sampling Stand
3 Compact Controller TIGERMAK Type 6942-10Order no.: 302428
3.1 TechDevice-specific data
Nominal operating temperature
Electrical data
Operating voltage
CPU
Processor
Built-in memory
Removable memory
Real-time clock
Display/keypad
Display
Keypad
Measuring system interface
Power inputs
Voltage inputs
Temperature sensors
Pulse transmitter input
PWM output
N"
B:
Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
nical data
23 ± 2 °C
DC 24 V (9 - 36V) on-board power supply (load-dump protection,stabilised)
Power PCTM
MPC8270 (266MHz)
FLASH: 256 MB FLASH (program memory)
SDRAM: 128 MB SDRAM (working memory)
SRAM: 4 MB (fail-safe data memory with battery backup)
CompactFlash: all sizes
With battery backup
LC graphic display, 120 x 89 mm, monochrome, transflective
320x240-pixel resolution,
adjustable LED backlighting
Resistive touch
3 x DC 24 V, 0/4 - 20 mA
2 x DC 24 V, 0 - 4V 1 x DC 12 V, 0 - 4V
2 x Pt 100, 4-conductor
24V, stabilised, max. 1kHz
24V, stabilised, max. 400Hz
Fingers only" control
Compact Controller TIGER MAK Type 6942-10
3-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
Analogue output (opt) 0/4 - 20 mA, RB max. 500 Ω, 10-bit, isolated
Outputs 16 x positive switching 1 A, 24 V, stabilised (3 A total)
Digital inputs 8 x optoisolated, input resistance 2.6 kΩ, bipolar, solid state
Pulse counter 3-channel max. 100Imp/s, input optoisolated 5 kΩ pull-up
Interfaces
Printer interface RS 232, RxD, TxD, CTS, RTS 115 K, isolated
Fieldbus interface P-NET RS485, 76800 baud, isolated
Other interfaces RS485, Ethernet 100 Mbit (on CPU), RS232 (excl. handshake)
Ambient conditions
Operating temperature - 20 … + 50 °C
Storage temperature - 25 … + 60 °C
Climatic category ISF in acc. with DIN 40050
Protection class IP 65 in acc. with DIN 40040
Mechanical data
Dimensions See scale drawing
Material Die-cast aluminium, blue coated
Front film PES
Weight 65 N (6.5 kg)
3.2 Compact Controller dimensions
Compact Controller TIGER MAK Type 6942-10
3-3
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
3.3 Blade-type fuses and LED displayson TIGER 3003 (on motherboard)
Standard blade-type motor vehicle fuses are fitted here with the followingratings:
Fuse no. Rating Voltage Circuit/commentsSI 4 20 A 24V_S Backup fuse for selective circuits SI18, SI12, SI13, SI14, SI15, SI25, SI5SI 5 1 A V24V_S 4-20mA inputs; voltage inputsSI 12 5 A +UB_1-4 Supply outputs 1-4; terminals 14-21SI 13 5 A +UB_5-8 Supply outputs 5-8; terminals 22-29SI 14 5 A +UB_9-12 Supply outputs 9-12; terminals 76-83SI 15 5 A +UB_13-16 Supply outputs 13-16; terminals 84-91SI 18 3 A 24V.COM RS 485, GPS, printer, P-netSI 20 1 A V24_GPRS Supply for GPRS modemSI 25 3 A PWM out Not in use
LED displays StatusLED 4 5.5 supply for CPU, motherboard, electronic measuring equipment Lights upLED 13 Supply voltage 24 V_O/12 V_I analogue inputs Lights upLED 16 µ Controller power supply "busy" FlashesLED 18 USB communication, digital IN, analogue IO FlashesLED 21 µ Controller k_mif "busy" (PT100, 4-20 mA, U_IN,…) Lights upLED 31 µ Controller pulse counter "busy" Lights upLED 73 USB hub for GPRS, USB socket, P-net Lights upLED 77 USB communication, P-net, digital outputs FlashesLED 78 µ Controller k_pio "busy" (P-net, digital outputs) FlashesLED CPU Top Supply voltage Lights upLED CPU Bottom Busy Flashes
Compact Controller TIGER MAK Type 6942-10
3-4
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
3.4 CompactFlash card
Direction of CF
Make sure the equipment is disconnected before changing the CompactFlash card!
Compact Controller TIGER MAK Type 6942-10
3-5
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
3.5 Exchanging the motherboard/CPUboardOrder no.: U891176942111A
The following points must be noted when replacing the motherboard:
The configuration can be transferred with the Compact Flash card. Theconfiguration must first be saved on the Compact Flash card via theService menu, then the configuration can be loaded from theCompactFlash card onto the new motherboard. Only if the calibrationswitch is closed will the calibration parameters not be loaded and willthen have to be adjusted.
To change the motherboard, the display must be unplugged (flip stripat end of ribbon cable), the terminal blocks disconnected, and the wiresdisconnected from the shield bar. The entire motherboard can now beunscrewed at all four corners (M6x12) and taken out.
Apply heat transfer compound supplied with new motherboard and fit.
If it is necessary to remove the cover- in order, for example, to check that theribbon cable is in the correct position for the display - it is essential toremove the Compact Flash card first.
Compact Controller TIGER MAK Type 6942-10
3-6
3.6 Compact Controller wiring
vp3003
10
Note: Connections to external sources of voltage must be isolated. Positive and negative
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
connections are required at each output.
Compact Controller TIGER MAK Type 6942-10
3-7
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
3.6.1 Power supply
The solenoid valves, all the measuring sensors and the ticket printer aresupplied by the Controller (see wiring diagram, pages 3-6).
3.6.2 OutputsThe outputs are user-configurable. The following assignment isrecommended:
phy. log. Inv. Function1 1 n Start pump2 n3 n4 n5 n6 n7 n8 n9 n10 n11 n12 n13 n14 n15 n
3.6.3 InputsThe inputs are user-configurable. The following assignment isrecommended:
phy. log. inv. Idle ID Function1 1 n L MS1 Milk sensor2 2 n L Start CIP3 n L4 n L5 n L6 n L7 n L8 n L
Shielded 2.5 mm² cable (16 A fuse, earth and positive branch) is required forthe 24 V power supply, including corresponding switch.
Compact Controller TIGER MAK Type 6942-10
3-8
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
Temperature Probe with Sensor Series B Type 6703-11
4-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
4 Temperature Probe withSensor Series B Type 6703-11Order no.: 216272
4.1 Technical dataDevice-specific data
Measuring element Pt 100 DIN IEC 751 Class A
Electrical data
Operating voltage DC 24 V, 50 mA
Connection 5 m cable (wire end ferrules)
Milk sensor alternating voltage 12 V AC, approx. 1.5 kHz
Milk sensor switching threshold Approx. 500
Milk sensor output Positive switching (DC +24 V) impulse current 20 mA
Ambient conditions
Operating temperature - 20 ... + 60 °C
Sensor element temperature range - 20 ... + 100 °C
Climatic category / protection class JUC / IP 65 (encapsulated)
Mechanical data
Housing specification V2A 1.4301
Weight 0.5 kg
Fitting DN 20 clamp
Installation position Optional, preferably vertical upright for sensor function
4.2 Dimensions
Temperature Probe with Sensor Series B Type 6703-11
4-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
4.3 Assembly/installationFollow the instructions below for installation in the milk pipeline by means ofwelded nipple type 6701-00-021:
Weld the nipple into the milk pipeline, mask joints and clean.
Apply a light coating of grease to O-ring D12.0 x 3.0 (order no. U220751)(attached to temperature probe), insert it in the welded nipple and, usinga fingertip, fit it carefully in the O-ring groove at the front end. Adjust anduse fingertip to check that it is seated evenly.
Fit clamp-type flat gasket DN 20 EPDM on tip of probe and insert probecarefully in welded nipple as far as it will go, making slight turns to theleft/right to ease it in.
Lock the probe in position with the two-hinged clamp, tightening the wingbolt by hand only.
Dual Sampler Controller, Type 6970-30
5-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
5 Dual Sampler Controller,Type 6970-30Order no.: 303360
The Dual Sampler Controller is used as a control unit for a sample collectionsystem. The high-performance unit, which is equipped with an RS232 P-Netconverter and a microprocessor-controlled data processing unit, has arobust die-cast aluminium housing and is designed for heavy-duty use ontankers.
5.1 Technical dataDevice-specific data
Nominal operating temperature 23 2 °C
Electrical data
Voltage rating for electronics DC 24 V 5%, fine-wire fuse 3 A
Operating voltage I/O and stepper motors On-board power supply DC min. 23 V max. 30 V,fine-wire fuse 6.3 A
Power consumption of electronics 165 mA, no-load
Power consumption I/O and steppermotors
100 mA, excl. stepper motors, no-load
Required power supply DC 24 V 5%, > 200 W
Interface 1
System bus interface RS485 P-Net 76800 bit/sec
Interface 2
RS 232 RxD, TxD, CTS, RTS, 9.6 Kbaud, isolated
Output for auxiliary power DC 24 V, max. 150 mA, fuse 300 mA
Interface 3
RS 232 RxD, TxD, CTS, RTS, 9.6 Kbaud, isolated
Output for auxiliary power DC 24 V, max. 150 mA, fuse 300 mA
Outputs
Number of channels 2
Switch High-side solid-state (MOSFET)
Voltage DC 24 V on-board power supply
Current Max. 1 A per output
RON Approx. 200 m
Digital inputs
Number 2, electrically isolated with optocoupler (3750 Vrms)
Input impedance Approx. 3 k
Voltage DC 24 V on-board power supply
Switch Bipolar, solid-state
Dual Sampler Controller, Type 6970-30
5-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
Milk sensor input
Pt100 Pt100 4-conductor, -20 … +100 °C, 0.05% resolution
Output for auxiliary power DC 24 V, max. 100 mA, fuse 400 mA
Flow level meter input
Voltage input (FLM fill level) 0 - 2.4 V = 100 - 0 %, 0.5 % resolution
Power input (FLM flow) 4 - 20 mA = 0 - 100%, 0.5% resolution
Output for auxiliary power DC 24 V, max. 250 mA, fuse 400 mA
Sampler control tank/sample
Power consumption 2 x 2.7 A
Maximum stepping frequency 10 kHz
Number of samples 2
Ambient conditions
Operating temperature - 20 … + 50 °C
Storage temperature - 20 … + 60 °C
Protection class IP 65 in acc. with DIN 40040
Mechanical data
Dimensions See scale drawing
Material Die-cast aluminium, blue coated
Weight 25 N (2.5 kg)
5.2 Dimensions
Quad Sampler Type 6900-20
6-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
6 Quad Sampler Type 6900-20Order no.: 279874
The Quad Sampler Type 6900-20 can take up to four samplessimultaneously online, directly from the intake line.
6.1 Technical dataDevice-specific data
Nominal operating temperature 23 ± 2 °C
Electrical data - sampler control
Auxiliary power 24 … 36 V
Power consumption Max. 4 x 4 A
Drive from two Sampler
Controllers, e.g. type 6970-30
Stepping frequency Max. 10 kHz
Number of samplers 4
Electrical isolation No
Cable length Max. 2.5 m
Connection 2 cables, 10 x 0.5-core, shielded
Ambient conditions
Operating temperature - 10 … + 50 °C
Storage temperature - 20 … + 70 °C
Protection class IP 65
Mechanical data
Dimensions See scale drawing
Materials which come into
contact with milk1.4301, silicone hose
Weight 8.3 kg
Accessories:
3" sampler pipe, type 6900-00-007 Order no.: 235920TAG Reader, type 6900-101 Order no.: 240074
Quad Sampler Type 6900-20
6-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
6.2 Dimensions
TAG Reader / Writer Type 6900-101
7-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
7 TAG Reader / Writer Type6900-101Order no.: 240074
The Tag Reader/Writer enables continuous bottle identification during thecollection of samples.
7.1 Technical dataDevice-specific data
Nominal operating temperature 23 ± 2 °C
Electrical data - sampler control
Auxiliary power DC 24 V
Power consumption 2 x 130 mA
Supply from two Sampler
Controllers, e.g. type 6970-30
Interface RS 232, 4 x RxD, TxD
Baud rate 9600 bit/sec
Electrical isolation No
Cable length Max. 2.5 m
Connection 2 cables, 7 x 0.25-core, shielded
Certificate ISO 15693
Readers/writers 4
Read-write distance Max. 30 mm
Ambient conditions
Operating temperature - 10 … + 50 °C
Storage temperature - 20 … + 70 °C
Protection class IP 65
Mechanical data
Dimensions See scale drawing
Materials which come intocontact with milk
PVDF, 1.4301
Weight 1.9 kg
TAG Reader / Writer Type 6900-101
7-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
7.2 Dimensions
GPS Receiver (RS 232) Type 6722-18
8-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
8 GPS Receiver (RS 232) Type6722-18Order no.: 241920
Automatic identification of suppliers and customers by means of the locatingfacility in conjunction with MAK 3003. Exact positioning via satellite.
8.1 Technical dataElectrical data
Auxiliary power DC 6 ... 40 V
Typical power consumption 50 mA at DC 24 V
Receiver aerial Inbuilt
Interfaces RS 232
Electrical connection 5m of cable, open wire ends
Ambient conditions
Operating temperature - 30 ... + 80°C
Protection class IP 67 (1 m immersion depth for 30 min)
Mechanical data
Dimensions/mounting holes
Mounting bolts 3 x M4, max. thread depth of 8.0 mm not to be exceeded(risk of breakage)
Weight 330 g incl. 5 m of cable
GPS Receiver (RS 232) Type 6722-18
8-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
8.2 ConnectionLength of cable supplied: 5 m, 130 mm of insulation stripped from cable end.
Assignment:
Signal Colour
+DC 24 V RedGND/Remote Black/yellowRxD BlueTxD White
8.3 Installation and wiringInstallation:
The aerial is built into the receiver therefore the receiver must be fitted ontop of the vehicle (e.g. on the rear cabin) and must be as horizontal aspossible. The unit must never be covered by metal parts and must be able to"see" as well as possible in all directions on all sides.
Wiring:
See wiring diagram for Compact Controller, pages 3-6.
GPRS Module, Type 6942-100
9-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
9 GPRS Module, Type 6942-100Order no.: 290337
Front view Rear view
Inserting the SIM card
1. Open the cover.
2. Open the slot and insert the SIM card.
Power cableconnection
RS 232 interface(15-pin D-sub)
Aerial socket
The modem must be affixed with plastic bolts.
GPRS Module, Type 6942-100
9-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
9.1 Magnetic antennaOrder no.: 232484
Ø 60mm / H: 13 mmAntenna length: 5 m
Fit the antenna in a location with good reception. The magnetic antenna maynot be used in potentially explosive atmospheres.
First enter PIN in configuration then insert SIM card while the system isswitched off.
9.2 SIM card specifications for the3003 systemThe following connections must be set up by the provider in order toguarantee efficient online support.
Service Port DescriptionSSH 22 Connection from 3003 system to officeSSH 22 Connection from office to 3003 systemHTTP 80 Connection from office to 3003 systemHTTPS 443 Connection from 3003 system to officePing Connection from 3003 system to office
Additional options for office link-up via FTL and FTP (optional):
Service Port DescriptionFTP *1)(passive)
21 / all portsabove 1023
Office link-up for Tiger MAK / PETRO andPetro 3003
Note:The IMEI number of the modem must be activated and supported by thenetwork operator/provider.
*1)
FTP connections can also be protected via SSL. This is subject to theprovider allowing FTP via SSL. Communications are then no longer in plaintext but in SSL encryption form.
Flow Level Meter Type 6826-x
10-1
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
10 Flow Level Meter Type 6826-xOrder no.: see technical data
10.1 Technical dataDevice-specific data
Liquid to be measured Milk
Fluid temperature 0 °C … +100 °C
Pressure Max. 10 bar
Electrical data
Auxiliary power (voltage, electricity) 24 V DC ± 10 % max. 6 W
Volume flow rate
NW 50 Max. 500 l/min
NW 65 Max. 1000 l/min
NW 76 Max. 1400l/min
Direction of flowAs indicated on metal type plate(+ direction)
SensorsMagnetic induction flow rate meter with µP-controlled functionmonitoring; conductivity sensor for level measurement
Electrical outputs
Magnetic induction flow rate meter Analogue output 4 … 20 mA
Fill level sensor Analogue output approx. 0.3 … 2.3 V
Electrical connection5 m connection cable with PG-11 gland and internal terminals,open cable ends
Ambient conditions
Operating temperature -10 °C … +50 °C
Maximum cleaning temperature 100 °C
Cleaning method CIP
Climatic category IWA in acc. with DIN 40040
Mechanical data
Dimensions 250 x 180 x 110 mm
Weight Approx. 50 N (= 5 kg)
Flow Level Meter Type 6826-x
10-2
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
Pipe connections
NW 50Connecting flange, joined by seal ring and clamp in acc. with DIN32676 NW 50
NW 65Connecting flange, joined by seal ring and clamp in acc. with ISO2852 NW 2.5"
NW 76Connecting flange, joined by seal ring and clamp in acc. with ISO2852 NW 3"
BuildNoryl red PPO housing with stainless steel measuring tubeTeflon-coated stainless steel measuring tube
InstallationMeasuring tube strictly horizontal, with upstream flow-calmingsection > 3 x NW
Order details
Designation Order number
Horizontal installation (flat):
Flow Level Meter 2", Type 6826-10 U950 17 682610
Flow Level Meter 2.5", Type 6826-20 U950 17 682620
Flow Level Meter 3", Type 6826-30 302063
Installation position
Vertical installation (on end):
Flow Level Meter 2" 90°, Type 6826-11 U950 17 682611
Flow Level Meter 2.5" 90°, Type 6826-21 U950 17 682621
Flow Level Meter 3" 90°, Type 6826-31 304782
Installation position
NB: Note direction of flow (arrow on type plate)
Accessories:
Seal ring clamp 2" U495077
Hinged clamp clip 2" U495080
Seal ring clamp 2.5" U495059
Hinged clamp clip 2.5" U495056
Seal ring clamp 3" U495031
Hinged clamp clip 3" U495065
Flow Level Meter Type 6826-x
10-3
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
10.2 Dimensions/installation
Fitting
The Flow Level Meter can either be laid flat or fitted upright depending onthe design of the unit.
The milk pipe must always be horizontal.
Horizontal (flat) installation: order "Horizontal" version
Vertical (upright) installation: order "Vertical" version
NB: Note direction of flow (arrow on type plate)
Build
1 = Cover 5 = Flat gasket2 = Clamp 6 = Type plate3 = Cable inlet 7 = Hexagon head bolt4 = Connection piece 8 = Seal ring
Flow Level Meter Type 6826-x
10-4
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
Instructions for mounting and dismounting:
1. First remove threaded milk pipe adapter, if still attached to FLM. To dothis, open the clamp fasteners (2) on both sides and remove the pipeconnections from the FLM.
NB:Do not lose the seals and put them back carefully whenreassembling.
2. First screw the threaded milk pipe adapters with the union nuts on bothsides of the pipe connections.
3. Then fit the FLM (not forgetting the flat gaskets (5)) and secure withthe clamp fasteners (2). It might be easier if the pipe supports/anchorsare slightly loosened first. Tighten again at the end.
Dismounting:
1. First take out the FLM by removing the clamp fasteners (2) on bothsides. It might be easier if the pipe supports/anchors are slightlyloosened first.
2. Then, if necessary, remove the threaded milk pipe adapters from theFLM.
NB:Do not lose the seals and put them back carefully whenreassembling.
Flow Level Meter Type 6826-x
10-5
MAK 3003 Sampling Stand Alone, Serviceanleitung Manual, SA 121107 (26.07.2013)
10.3 WiringThe FLM is supplied as standard with 5 m of cable which is connected asshown on the diagram below. Please follow the directions for use of thefollow-on device when wiring.
10.4 Electric terminal assignmentTerminal block Signal Colour 6970-30
3 - Flow 4 - 20 mA Black 483 +Flow 4 - 20 mA Purple 472 - Level 0.3 - 2.5 V Brown 462 + Level 0.3 - 2.5 V Yellow 454 P-NET A White Insulated4 P-NET B Blue Insulated4 P-NET S --- ---1 - 24 V DC (GND) Shield 441 + 24 V DC Pink/grey 43
Please insulate any unneeded conductors individually.
MAK 3003 Sampling Stand Alone, Serviceanleitunganual, SA 121107 (26.07.2013)
***