13
OJIYAS M3X0.5 THREAD PLUG GAUGE Procedure: 33K6-4-203-1 Order Number: PH-WO-10 ID Number: 43719232 Serial Number : 437 Asset Number: 005856 Calibration Date: February Temperature: 20.1°C Relative Humidity: 50 Thread Plug Accuracy Function Nominal As Found As Left Result PD (Go Plug) Top 2.6750 mm 2.6750 mm 2.6764 mm 2.6764 mm 2.6801 mm PASS PD (Go Plug) Middle 2.6750 mm 2.6750 mm 2.6764 mm 2.6764 mm 2.6801 mm PASS PD (Go Plug) Bottom 2.6750 mm 2.6750 mm 2.6764 mm 2.6764 mm 2.6801 mm PASS PD (No Go Plug) Top 2.7750 mm 2.7699 mm 2.7742 mm 2.7742 mm 2.7750 mm PASS PD (No Go Plug) Middle 2.7750 mm 2.7699 mm 2.7704 mm 2.7704 mm 2.7750 mm PASS PD (No Go Plug) Bottom 2.7750 mm 2.7699 mm 2.7704 mm 2.7704 mm 2.7750 mm PASS Function As Found As Left Major Diameter (GO Plug) 3.0077 mm 3.0077 mm 0.000601 ajor Diameter (NO GO Plug 2.9826 mm 2.9826 mm 0.000601 Lower Limit Upper Limit Uncertaint y

43719232

Embed Size (px)

DESCRIPTION

dsfsd

Citation preview

OJIYAS M3X0.5 THREAD PLUG GAUGEProcedure: ID Number: Asset Number: Temperature: 33K6-4-203-1 43719232 005856 20.1C Order Number: Serial Number : Calibration Date: Relative Humidity: PH-WO-1082173808 43719232 February 28, 2013 50.0%

Thread Plug Accuracy Function PD (Go Plug) Top PD (Go Plug) Middle PD (Go Plug) Bottom PD (No Go Plug) Top PD (No Go Plug) Middle PD (No Go Plug) Bottom Function Major Diameter (GO Plug) Major Diameter (NO GO Plug) Nominal 2.6750 mm 2.6750 mm 2.6750 mm 2.7750 mm 2.7750 mm 2.7750 mm Lower Limit 2.6750 mm 2.6750 mm 2.6750 mm 2.7699 mm 2.7699 mm 2.7699 mm As Found 2.6764 mm 2.6764 mm 2.6764 mm 2.7742 mm 2.7704 mm 2.7704 mm As Left 2.6764 mm 2.6764 mm 2.6764 mm 2.7742 mm 2.7704 mm 2.7704 mm Upper Limit 2.6801 mm 2.6801 mm 2.6801 mm 2.7750 mm 2.7750 mm 2.7750 mm Uncertainty 0.000601 0.000601 Result PASS PASS PASS PASS PASS PASS Uncertainty 0.000601 0.000601 0.000601 0.000601 0.000601 0.000601

As Found 3.0077 mm 2.9826 mm

As Left 3.0077 mm 2.9826 mm

Page 1 of 12

Outside Diameter (GO) Nominal #REF! Outside Diameter (NO GO) Nominal #REF!

Test 1 3.0077

Test 2 3.0077

Test 3 3.0077

Test 4 3.0077

Test 5 3.0077

Avg. 3.0077

Test 1 2.9826

Test 2 2.9826

Test 3 2.9826

Test 4 2.9826

Test 5 2.9826

Avg. 2.9826

Pitch Diameter (GO) Nominal 2.6750 2.6750 2.6750 Pitch Diameter (NO GO) Nominal 2.7750 2.7750 2.7750

Test 1 2.6764 2.6764 2.6764 Test 1 2.7742 2.7704 2.7704

Test 2 2.6764 2.6764 2.6764 Test 2 2.7742 2.7704 2.7704

Test 3 2.6764 2.6764 2.6764 Test 3 2.7742 2.7704 2.7704

Test 4 2.6764 2.6764 2.6764 Test 4 2.7742 2.7704 2.7704

Test 5 2.6764 2.6764 2.6764 Test 5 2.7742 2.7704 2.7704

Avg. 2.6764 2.6764 2.6764 Avg. 2.77419 2.77038 2.77038

NOTE THE WIRE USED::::

0.01155

mm Go Component TYPE A Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000 Value Distribution Type Divisor Divisor Value Adjusted Component Squared Value

Long Term Stability

0.0

Normal

1.0

1.000

0.0000

0.00

TYPE B Uncertainty as reported Uncertainty as reported 0.0001016 0.0000737 Rectangular Rectangular 2.0 2.0 2.000 2.000 0.000051 0.000037 0.000 0.000

Accuracy of Standard Resolution

0.000508 0.000025

Rectangular Rectangular

1.732 1.732

1.732 1.732

0.0003

0.000000

0.0000 0.000000 Sum 0.0000 U = Sqrt(Sum) = 0.000 Expanded U = (for k=2) 0.001

NO GO Component TYPE A Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000 Value Distribution Type Divisor Divisor Value Adjusted Component Squared Value

Long Term Stability

0.0

Normal

1.0

1.000

0.0000

0.00

TYPE B Uncertainty as reported Uncertainty as reported 0.0001016 0.0000737 Rectangular Rectangular 2.0 2.0 2.000 2.000 0.000051 0.000037 0.000 0.000

Accuracy of Standard Resolution

0.000508 0.000025

Rectangular Rectangular

1.732 1.732

1.732 1.732

0.0003

0.000000

0.0000 0.000000 Sum 0.0000 U = Sqrt(Sum) = 0.000 Expanded U = (for k=2) 0.001

unit mm

test 1 3.0077

test 2 3.0077

test 3 3.0077

test 4 3.0077

test 5 3.0077

test times 5

Notes

Need to generate at least 5 easurement for one particular range. Need long term stability of CDI system. If not available, use 1 year stability specifications.

Uncertainty reported in your Calibration Certificate of the Thread Wire Gauge Uncertainty reported in your Calibration Certificate of the Super Micrometer

Accuracy of STD Resolution of STD 1/4 read

Use 0.000601mm

unit mm

test 1 2.9826

test 2 2.9826

test 3 2.9826

test 4 2.9826

test 5 2.9826

test times 5

Notes

Need to generate at least 5 easurement for one particular range. Need long term stability of CDI system. If not available, use 1 year stability specifications.

Uncertainty reported in your Calibration Certificate of the Thread Wire Gauge Uncertainty reported in your Calibration Certificate of the Super Micrometer

Accuracy of STD Resolution of UUT

Use 0.000601mm

average 3.00770

variance 0.0000000

var square-root 0.0000000

average 2.98260

variance 0.0000000

var square-root 0.0000000

GO Component TYPE A Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000 Value Distribution Type Divisor Divisor Value Adjusted Component Squared Value

Long Term Stability TYPE B Uncertainty as reported Uncertainty as reported Accuracy of Standard Resolution

0.0

Normal

1.0

1.000

0.0000

0.00

0.0001016 0.0000737

Rectangular Rectangular

2.0 2.0

2.000 2.000

0.000051 0.000037

0.000 0.000

0.000508 0.000025

Rectangular Rectangular

1.732 1.732

1.732 1.732

0.0003

0.000000 0.000000 0.0000 0.000 0.001

0.0000 Sum U = Sqrt(Sum) = Expanded U = (for k=2)

NO GO Component TYPE A Repeatability 0.000 Normal 1.0 1.000 0.0000 0.000 Value Distribution Type Divisor Divisor Value Adjusted Component Squared Value

Long Term Stability TYPE B Uncertainty as reported Uncertainty as reported

0.0

Normal

1.0

1.000

0.0000

0.00

0.0001016 0.0000737

Rectangular Rectangular

2.0 2.0

2.000 2.000

0.000051 0.000037

0.000 0.000

Accuracy of Standard Resolution

0.000508 0.000025

Rectangular Rectangular

1.732 1.732

1.732 1.732

0.0003

0.000000 0.000000 0.0000 0.000 0.001

0.0000 Sum U = Sqrt(Sum) = Expanded U = (for k=2)

unit mm

test 1 2.67640

test 2 2.67640

test 3 2.67640

test 4 2.67640

test 5 2.67640

Notes

Need to generate at least 5 easurement for one particular range. Need long term stability of CDI system. If not available, use 1 year stability specifications.

Uncertainty reported in your Calibration Certificate of the Thread Wire Gauge Uncertainty reported in your Calibration Certificate of the Super Micrometer

Accuracy of STD Resolution of UUT

Use 0.000601mm

unit mm

test 1 2.77419

test 2 2.77419

test 3 2.77419

test 4 2.77419

test 5 2.77419

Notes

Need to generate at least 5 easurement for one particular range. Need long term stability of CDI system. If not available, use 1 year stability specifications.

Uncertainty reported in your Calibration Certificate of the Thread Wire Gauge Uncertainty reported in your Calibration Certificate of the Super Micrometer

Accuracy of STD Resolution of UUT

Use 0.000601mm

test times 5

average 2.67640

variance var square-root 0.0000000 0.0000000

test times 5

average 2.77419

variance var square-root 0.0000000 0.0000000