Upload
nguyenmien
View
222
Download
3
Embed Size (px)
Citation preview
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2016 Waters Corporation
METHODS Conditions
System: ACQUITY UPLC H-Class with Column Manager
Column: ACQUITY UPLC BEH C18, 1.7 µm, 2.1 x 50 mm
Mobile phase A: 0.1% (v/v) Formic acid in Water
Mobile phase B: 0.1% (v/v) Formic acid in Acetonitrile
Column Temperature: 30 C
Injection volume: 2 uL
Flow rate: 0.8 mL/min
Isocratic:60% A: 40%B
ACQUITY PDA Detector
Wavelength range: 210-400 nm
Resolution: 3.6 nm
Selected wavelengths: 228 nm, 4.8 nm resolution
Time Constant: Normal
Sampling rate: 20 pts/s
ACQUITY ELSD Detector Isocratic Solvent Manager
Gas: 25 psi Solvent: 0.1% (v/v) formic acid in methanol
Data rate: 10 pps Flow rate: 0.3 mL/min
Nebulizer Mode: Cooling
Nebulizer Temperature: 55 ˚C
Sample Preparation:
Glimepiride and related compounds B and C were purchased from the USP.
All standards were dissolved in 55:45 methanol: water and sonicated.
The drug substance glimepiride was obtained from an outside source. Acid
hydrolysis was conducted at 40 ˚C for 0-7 days . The concentration of acid
was 0.1M HCl in the degradation reaction.
INTRODUCTION
Forced degradation studies are typically performed using
HPLC and UV detectors to understand the degradation
pathway of pharmaceuticals and to insure all impurities
are accounted. In these studies, performing mass
balance or the conservation of mass is crucial. Multiple
orthogonal detectors based on different principles can be
used to measure or identify compounds with different
chemical or physical properties. We will evaluate mass
balance using a triple detection system consisting of a
PDA, ELSD and a mass detector. Relative response ratios
will then be used to perform mass balance. The
degradation path way will then be confirmed using of a
mass detector, specifically through the identification of
impurities and their by-products.
USE OF A TRIPLE DETECTION (UV-ELSD-MS) SYSTEM FOR MASS BALANCE IN FORCED DEGRADATION OF PHARMACEUTICALS Paula Hong and Patricia R. McConville
Waters Corporation, 34 Maple Street, Milford, MA 01757
References
1. Chapter <621> CHROMATOGRAPHY United States Pharmacopeia and National Formulary (USP 37-NF
32 S1) Baltimore, MD: United Book Press, Inc.; 2014. p. 6376-85.
2. Mark AN, Andreas K, Patrick JJ. Role of Mass Balance in pharmaceutical stress testing. Pharmaceutical
Stress Testing: CRC Press; 2011. p. 233-53.
CONCLUSIONS Triple detection system in combination with
Empower 3 FR2 provides various tools to assist in
mass balance, including:
Determination of relative response factors by
using the ratio of UV peak and the log of ELSD
peak responses
The ability to input relative response values into
Empower 3 FR2 to determine corrected area
values for impurities for mass balance
determinations
Figure 1. ACQUITY UPLC H-Class system with triple detection including
ACQUITY PDA, ELSD and QDa detectors. The triple detection system
includes an isocratic solvent manager (ISM) which provides make-up
solvent to the QDa detector and houses the splitter required for the ELSD
and the QDa. After the PDA detector, the flow is split to the ELSD detector
and QDa. The composition and flow rate of the make-up solvent impact the
split ratio to the ELSD and the QDa.
Pre-configured Splitter
in the Isocratic Solvent
ACQUITY UPLC H-Class
System with Column Managers and Triple
Detection (UV-ELSD-PDA)
Flow diagram
Wavelength Separation Conditions
RRF Rel Compound B
RRF Rel Compound C
228 Isocratic 1.36 1.10
Table 1. Relative Response Factors for Related compound B and C using the
ratio of the slope of the API/slope of the impurity. The value for related
compound B is outside of 0.8-1.2 range and, therefore, should be applied, as
specified by the USP Chapter <621>.1
AU
0.00
0.05
0.10
0.15
Minutes
0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50 5.00
LS
U
0.00
10.00
20.00
Minutes
1.00 2.00 3.00 4.00 5.00
Inte
nsity
0.0
5.0x105
1.0x106
1.5x106
2.0x106
Minutes
1.00 2.00 3.00 4.00 5.00
PDA @ 228 nm
ELSD
MS XIC
ESI+
Figure 2. Separation of standards of active pharmaceutical ingredient (API),
related compound B and related compound C under isocratic conditions. The
overlay of standards at 250 µg/mL for the API and 10 µg/mL for related
compounds B and C shows the differences in relative response among the
detectors. The UV and ELSD give similar relative response for the three
compounds. In the mass detector related compound C has a greater peak
area than related compound B.
RESULTS AND DISCUSSION Multi-detection of API and Related Compounds
API Rel Cmpd B
Rel Cmpd C
Figure 4. ELSD calibration curves for glimepiride related compound C. The
ELS detector has a quadratic fit to the calibration curve (left) for peak area
vs. the amount. If the values are converted to the logarithmic functions
(inset), the calibration curve fit is linear (right). The R2 value for this curve
is 0.999140.
Determination of Relative Response Factors
Calibration curve ELSD stack
Project Name: MassBalance Glimep 14Oct15Reported by User: System
Report Method: Calibration curve ELSD stack Date Printed:
24851 2/26/2016Report Method ID: 24851
9:49:19 AM US/Eastern
Calibration Plot
Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.999140; Equation Y = 1.67e+000 X + 2.14e+000
1.50
2.00
2.50
3.00
3.50
4.00
4.50
5.00
5.50
6.00
6.50
7.00
LogAmount
0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40
Point Information ' Peak: GlimepirideB' table contains no data.
1
2
3
4
5
6
7
8
9
10
Name Level X Value Response Calc. Value % Deviation Manual Ignore
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
1.000000
1.000000
1.000000
1.698970
1.698970
1.698970
2.096910
2.096910
2.096910
2.397940
3.838315
3.805043
3.820003
4.921575
4.942371
4.964010
5.629349
5.648075
5.678314
6.122215
1.018281
0.998353
1.007314
1.667098
1.679553
1.692515
2.091018
2.102234
2.120345
2.386219
1.828
-0.165
0.731
-1.876
-1.143
-0.380
-0.281
0.254
1.118
-0.489
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
Peak: Glimepiride C
Calibration curve ELSD stack
Project Name: MassBalance Glimep 14Oct15Reported by User: System
Report Method: Calibration curve ELSD stack Date Printed:
24851 2/26/2016Report Method ID: 24851
9:50:31 AM US/Eastern
Calibration Plot
Name: Glimepiride C; Fit Type: Quadratic (2nd Order); R^2: 0.996621; Equation Y = 1.62e+001 X̂ 2 + 1.65e+003 X - 2.00e+004
-2.0x105
0.0
2.0x105
4.0x105
6.0x105
8.0x105
1.0x106
1.2x106
1.4x106
1.6x106
Amount
0.00 20.00 40.00 60.00 80.00 100.00 120.00 140.00 160.00 180.00 200.00 220.00 240.00
Point Information ' Peak: GlimepirideB' table contains no data.
1
2
3
4
5
6
7
8
9
10
Name Level X Value Response Calc. Value % Deviation Manual Ignore
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
Glimepiride C
10.000000
10.000000
10.000000
50.000000
50.000000
50.000000
125.000000
125.000000
125.000000
250.000000
6891.517641
6383.265905
6606.982471
83478.507006
87573.059729
92047.180381
425941.030337
444707.880031
476775.136351
1324998.044518
14.265562
14.025072
14.131038
43.764007
45.086203
46.510456
122.439136
125.740811
131.244015
241.342736
42.656
40.251
41.310
-12.472
-9.828
-6.979
-2.049
0.593
4.995
-3.463
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
No
Peak: Glimepiride CELSD calibration curve
Quadratic fit
ELSD calibration curve
Linear fit
Calibration curve
Project Name: MassBalance Glimep 14Oct15Reported by User: System
Report Method: Calibration curve Date Printed:
23805 12/28/2015Report Method ID: 23805
8:36:43 PM US/Eastern
Calibration Plot
Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.996948; Equation Y = 1.84e+000 X + 1.38e+000
Name: GlimepirideB; Fit Type: Linear (1st Order); R^2: 0.999125; Equation Y = 1.69e+000 X + 2.25e+000
Name: GlimepirideB; Fit Type: Linear (1st Order); R^2: 0.999125; Equation Y = 1.69e+000 X + 2.25e+000
Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.996948; Equation Y = 1.84e+000 X + 1.38e+000
Name: Glimepiride C; Fit Type: Linear (1st Order); R^2: 0.999140; Equation Y = 1.67e+000 X + 2.14e+000
Name: Glimepiride; Fit Type: Linear (1st Order); R^2: 0.996948; Equation Y = 1.84e+000 X + 1.38e+000
LogA
rea
1.00
2.00
3.00
4.00
5.00
6.00
7.00
ppm
0.00 0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40
1
2
3
4
5
Name Y-axis X-axis Fit Type R^2 Equation
GlimepirideB
Glimepiride C
Glimepiride
Glimepiride C
Glimepiride
LogArea
LogArea
LogArea
LogAmount
LogAmount
LogAmount
Linear (1st Order)
Linear (1st Order)
Linear (1st Order)
0.999125
0.999140
0.996948
Y = 1.69e+000 X + 2.25e+000
Y = 1.67e+000 X + 2.14e+000
Y = 1.84e+000 X + 1.38e+000
Figure 3. Overlay of glimepiride related compound C standards (10-250 µg/
mL) in PDA and ELSD. The UV detector produces a linear response for
standards. Evaluating the peak areas in the ELSD, a non-linear or logarithmic
response is observed. For example, at 10 µg/mL the response in the ELSD
(pink trace) is significantly lower than that observed in the PDA.
AU
0.00
0.20
0.40
0.60
0.80
1.00
1.20
1.40
1.60
1.80
Minutes
0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00
PDA 250 µg/mL 125 µg/mL 50 µg/mL 10 µg/mL
LS
U
0.00
100.00
200.00
300.00
400.00
500.00
600.00
700.00
800.00
900.00
Minutes
0.30 0.40 0.50 0.60 0.70 0.80 0.90 1.00
250 µg/mL 125 µg/mL
50 µg/mL
ELSD
Table 3. Mass balance determinations for forced degradation of glimepiride.
The calculations were performed using RRF determined with the ELSD
method. The RRF were entered into Empower 3 FR 2 for corrected values of
the related impurities. All mass balance values were within 2%.
n = 3
Reference
T=0
1 day
3 days
5 days
7 days
Amount
238.4 236.1 239.0 242.7 239.1 244.1
% Recovery
99.0 100.2 101.8 100.3 102.4
AU
0.00
0.20
0.40
374.
3
432.
2
513.
2
AU
0.00
0.20
0.40
AU
0.00
0.10
0.20
0.30
0.40
0.50
Minutes
0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 4.50
t=0
t= 3 days
t= 5 days
Figure 5. UV chromatograms of forced degradation of glimepiride drug
substance with base mass labels. The drug substance was exposed to acidic
hydrolysis conditions at 40 C over a period of days. Over the course of the
study the two impurity peaks ( related compound C and B) increased in peak
area.
API
Rel Cmpd B
Rel Cmpd C
Standard
RRF Rel Compound B
RRF Rel Compound C
50 1.25 1.10
125 1.24 1.09
Table 2. Relative Response Factors for related compound B and C using the
ratio of the UV peak area to the log of the ELSD peak area. RRF can be
calculated using the response of the UV detector to a mass concentration
dependent detector.2 This assumes a linear relationship for both detectors.
To convert the ELSD calibration response to a linear function, the log of
both x and y values can be used. Thereby, using the log of the ESLD peak
area, we can calculate RRF factors for both impurities. These values have
good correlation with those obtained using the slopes of the calibration
curves in the UV.
Mass Balance for Forced Degradation Studies
𝑅𝑅𝐹 = [𝑆𝑙𝑜𝑝𝑒 𝑜𝑓 𝐼𝑚𝑝𝑢𝑟𝑖𝑡𝑦]
[𝑆𝑙𝑜𝑝𝑒 𝑜𝑓 𝐴𝑃𝐼]
𝑅𝑅𝐹 = 𝑈𝑉 𝐴𝑟𝑒𝑎𝐼𝑚𝑝𝑢𝑟𝑖𝑡𝑦
𝐶𝑜𝑛𝑐 𝐷𝑒𝑡𝑒𝑐𝑡𝑜𝑟 𝑟𝑒𝑠𝑝𝑜𝑛𝑠𝑒 𝑖𝑚𝑝𝑢𝑟𝑖𝑡𝑦/
𝑈𝑉 𝐴𝑟𝑒𝑎𝐴𝑃𝐼𝐶𝑜𝑛𝑐 𝐷𝑒𝑡𝑒𝑐𝑡𝑜𝑟 𝑟𝑒𝑠𝑝𝑜𝑛𝑠𝑒 𝐴𝑃𝐼
𝑅𝑅𝐹 = 𝑈𝑉𝐴𝑟𝑒𝑎𝐼𝑚𝑝𝑢𝑟𝑖𝑡𝑦
𝐸𝐿𝑆𝐷 log(𝐴𝑟𝑒𝑎 𝑖𝑚𝑝𝑢𝑟𝑖𝑡𝑦 )/
𝑈𝑉 𝐴𝑟𝑒𝑎𝐴𝑃𝐼𝐸𝐿𝑆𝐷 log (𝐴𝑟𝑒𝑎 𝐴𝑃𝐼)
=
10 µg/mL