CELL-DYN® 1700 (77) Index ABBOTT ADD INDEX INSTRUMENT SERVICE ADVISORY PRODUCT: CELL-DYN® 1700 (77) DATE: 19-MAY-98 ISA # SUBJECT EFFECTIVITY DATE 77-017 New Hard Drive 26-MAY-98 77-016 Lyse Reagent for Japan 31-MAR-98 77-015 Change in Keyboard Cable 20-OCT-97 77-014 CPU Clock IC Change 23-APR-97 77-013 Noise Reduction Fans 26-FEB-97 77-012 CELL-DYN® External Computer Set-Up 05-NOV-96 77-011 CELL-DYN® 1700 Service Disk 26-FEB-97 77-010 CELL-DYN® 1700 Revision D Software 23-AUG-96 77-009 CANCELLED CANCELLED 77-008 Cubitainer Case Changes 24-APR-96 77-007A Signal Processor Module Ajustment OBSOLETE 77-007 Signal Processor Module Adjustment OBSOLETE 77-006 New Vacuum Pump Configuration 05-DEC-95 77-005 New 10mL Syringe OBSOLETE 77-004 New Optical Sample Probe Assembly 05-DEC-95 77-003 Platelet LRI 11-OCT-95 77-002 New I/O Board 11-OCT-95 77-001 New 2.5mL Syringe 01-JUL-95 PENDING - ISA index number has been reserved for a future ISA. CANCELLED - ISA index number is cancelled. INCORPORATED - ISA was incorporated into another document or manual. OBSOLETE - ISA no longer applies. COMPLETE - ISA is complete. END OF DOCUMENT

ABBOTT INDEX INSTRUMENT SERVICE · PDF file31/03/1998 · ISA# 77-014 - CPU Clock IC Change Page 5 ABBOTT ADD INSTRUMENT SERVICE ADVISORY SUBJECT: CPU Clock IC Change ISA#: 77

Embed Size (px)

Citation preview

CELL-DYN® 1700 (77) Index Page 1

ABBOTTADD

INDEXINSTRUMENT SERVICE ADVISORY

PRODUCT:

CELL-DYN® 1700 (77)DATE:19-MAY-98

ISA # SUBJECT EFFECTIVITY DATE

77-017 New Hard Drive 26-MAY-9877-016 Lyse Reagent for Japan 31-MAR-9877-015 Change in Keyboard Cable 20-OCT-9777-014 CPU Clock IC Change 23-APR-9777-013 Noise Reduction Fans 26-FEB-9777-012 CELL-DYN® External Computer Set-Up 05-NOV-9677-011 CELL-DYN® 1700 Service Disk 26-FEB-9777-010 CELL-DYN® 1700 Revision D Software 23-AUG-9677-009 CANCELLED CANCELLED77-008 Cubitainer Case Changes 24-APR-9677-007A Signal Processor Module Ajustment OBSOLETE77-007 Signal Processor Module Adjustment OBSOLETE77-006 New Vacuum Pump Configuration 05-DEC-9577-005 New 10mL Syringe OBSOLETE77-004 New Optical Sample Probe Assembly 05-DEC-9577-003 Platelet LRI 11-OCT-9577-002 New I/O Board 11-OCT-9577-001 New 2.5mL Syringe 01-JUL-95

PENDING - ISA index number has been reserved for a future ISA.CANCELLED - ISA index number is cancelled.INCORPORATED - ISA was incorporated into another document or manual.OBSOLETE - ISA no longer applies.COMPLETE - ISA is complete.

END OF DOCUMENT

ISA# 77-015 - Change in Keyboard Cable Page 3

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:Change in Keyboard Cable

ISA#:77-015

ORIGINATOR: Laurie A. Wilmot

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C. Stout 20-OCT-97

EFFECTIVITY DATE:20-OCT-97

CELL-DYN is a registered trademark of Abbott Laboratories.

The supplier of the keyboard assembly for the CD1700 analyzers has ceased manufacturing the "curly" retractable cable. The new cable is a straight cable that is approximately six feet or just under two meters in length. This new configuration of the keyboard replaces the current keyboard used on the CD1700, however the part number for the keyboard will remain the same for ordering purposes.

END OF DOCUMENT

ISA# 77-015 - Change in Keyboard Cable Page 4

ISA# 77-014 - CPU Clock IC Change Page 5

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:CPU Clock IC Change

ISA#:77-014

ORIGINATOR: Laurie A. Wilmot

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C Stout 4/29/97

EFFECTIVITY DATE:04/23/97

CELL-DYN is a registered trademark of Abbott Laboratories.

The purpose of this ISA is to inform the field that there has been a change to the CPU PCB (P/N 1700022) on the CELL-DYN® 1700. Some of you are aware that the CD 1700 clock looses time as the CPU ages. This change will improve the operation of the CPU real time clock and consequently improve the clock's reliability. The part number for the new CPU PCB is 9360210.

ISA# 77-014 - CPU Clock IC Change Page 6END OF DOCUMENT

ISA# 77-013 - Noise Reduction Fans Page 7

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:Noise Reduction Fans

ISA#:77-013

ORIGINATOR: Jamie G. Besley

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John Stout

EFFECTIVITY DATE:26-FEB-97

CELL-DYN is a registered trademark of Abbott Laboratories.

The CELL-DYN® 1700 115v Air Fan Assembly (P/N 9360179) will be replaced by a 88 CFM Low Noise Air Fan Assembly (P/N 2604019). This assembly will help lower the acoustic noise generated by the analyzer. This new fan assembly will be installed on new analyzers manufactured after December, 1996. You will need to order a set of replacement screws (P/N 2801316, SCR Pan PH 8-32, .5) and nuts (P/N 2802252, Nut 8-32 STL) when replacing the new fan assembly with an older fan assembly. The screw and nut changes are due to the replacement fans' increased

ISA# 77-013 - Noise Reduction Fans Page 8depth. No other modifications will be needed with this part change.

END OF DOCUMENT

ISA# 77-012 - CELL-DYN® External Computer Set-Up Page 9

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:CELL-DYN® External Computer Set-Up

ISA#:77-012

ORIGINATOR: Jamie G. Besley/Steve Fortenberry

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C Stout 11-5-96

EFFECTIVITY DATE:05-NOV-96

CELL-DYN is a registered trademark of Abbott Laboratories. COMPAQ is registered trademark of Compaq Computer Corporation.

WINDOWS is registered trademark of Microsoft Corporation.

This procedure is for the connection of the CELL-DYN® 3000 series and the 1700 analyzer to the Compaq® Contura laptop computer used by Field Service. This will allow the FSR to connect to the external computer connection on the back of the Data Station for verification of correct information transmission from the CELL-DYN to the customers' LIS (Lab Information System).

ISA# 77-012 - CELL-DYN® External Computer Set-Up Page 10Tools Needed:

1. Compaq Contura laptop computer with Microsoft Windows® software

2. 9 pin Null Modem Cable with female connections on both ends (the cable can be purchased from any computer or electronics store)

Connection:

3000 Series

Connect one end of the nine pin cable to the External Computer connection (Com 1) of the Data Station. This connection is labelled on the old style (one unit) data stations and is on the same board as the graphics printer connection. On the new style (PC-type) data station the connection is labelled as "Com 1", and is also on the same board as the graphics printer labelled as "LPT1". The other end of the cable should be connected to the Compaq Contura computer port labelled "I0I0I". This concluded the hardware connections required.

1700

Connect one end of the nine pin cable to the I/O port above the graphics printer connection on the same card. This board is the closest to the front of the instrument. The other end of the cable should be connected to the Compaq Contura computer port labelled "I0I0I". This concluded the hardware connections required.

ISA# 77-012 - CELL-DYN® External Computer Set-Up Page 11Laptop Software Set-Up

The software connection with the analyzer is done with the Windows "Terminal" program which is a standard program in Windows. This program is found in the Accessories sub window. The CELL-DYN® computer set-up information will be needed to set-up the connection to the Compaq® computer. To get the computer set-up information you will go to the set-up screen from the main menu. In set-up you will need to write down the following information: Baud Rate, Data Bits, Stop Bits, Parity, and the CTS state (not enabled or enable).

On the Compaq Contura you will need to open the Accessories window by double clicking on it, and double clicking on the "Terminal" icon. The Terminal program needs to be set-up for the same settings that the analyzer is set at. Under the settings menu, click on "Communications.." menu; this will bring up the set-up screen. In the "communications" screen, set the Baud Rate, Data Bits, Stop Bits, and Parity to the values you wrote down from the CELL-DYN computer set-up screen. If the CELL-DYN is set for CTS enable, set the Flow Control to Hardware. If the CELL-DYN is set for CTS not enabled, set the Flow Control to Xon/Xoff. Set the connection field to Com 1:. You can click on the "OK" key to activate these settings.

The Terminal program will need to be told where it is going to save the information you are about to download. To activate the computer for downloading you will need to click on the "Transfers" menu and click on "Receive Text File...". You will need to type in a file name with the extension ".TXT" (example "CD3000.TXT") and click on the "OK" key. The Compaq is now ready to receive all data. After samples have been sent to the Compaq, you will be able to see this information on the screen as well as having it sent to the file you have set up. After completion of data transfers, you will need to click on the "Stop" key at the bottom of the screen to end the connection.

Hard Copy of Information

ISA# 77-012 - CELL-DYN® External Computer Set-Up Page 12In some instances you may need a hard copy of the information received from the CELL-DYN for evaluation or faxing to Santa Clara. The Received Text file can be printed by any word processor that can load an ASCII text file by changing from your word processor file type to an ASCII file type.

Now you can compare your results to the examples in the back of the Interface Specifications guide or present the printout to whom ever needs it.

END OF DOCUMENT

ISA# 77-011 - CELL-DYN® 1700 Service Disk Page 13

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:CELL-DYN® 1700 Service Disk

ISA#:77-011

ORIGINATOR: Jamie G. Besley

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John Stout

EFFECTIVITY DATE:26-FEB-97

CELL-DYN is a registered trademark of Abbott Laboratories.

MS-DOS is a registered trademark of Microsoft Corporation.

A Service Disk for the CELL-DYN® 1700 Analyzers has been released. This Service Disk (P/N 9380474) is similar to the CELL-DYN 3000 series analyzers. The Service Disk will have the MS-DOS® and boot command files, which will enable Field Service and the Customers to format and reinstall the installation software (P/N #9380468) on the analyzer hard drive. It will also enable the user to create a DOS partition on a new hard drive (fdisk.com). The Service Disk will be included in the CD 1700 Accessory Kit (P/N 03H54-01).

ISA# 77-011 - CELL-DYN® 1700 Service Disk Page 14

Installing a New Hard Drive:

1. After installing a new hard drive, place the CD 1700 Service Disk in the floppy drive and power the analyzer on. During the Boot-Up process press the DEL key when prompted and proceed to the Standard CMOS Set-Up. Verify the hard drive settings are correct (See Note Below). Also verify the Advanced CMOS Set-Up. Scroll to the System Boot-Up sequence and verify it's set at A: , C: . Use the Pg Up or Pg Dn keys to change the setting if necessary. Exit CMOS Set-Up, save your current settings and let the system reboot.

NOTE: The Hard Drive Set-Up information has changed. New Hard Drives currently shipped are formatted at the new settings, please note the following changes and add them to your Service Manual (page 5-14).

TYPE CYCL HEAD WP LZ SECT SIZE

47 1024 16 0 1025 63 504MB

2. At the A> type FDISK , and press the Enter key. Choose option [1] (Create DOS partition or Logical DOS Drive) and press the Enter key.

3. At the Create DOS Partition or Logical DOS Drive Menu, choose option [1] (Primary DOS Partition), and press the Enter key. Enter [Y] to use the maximum size for the hard drive DOS partition when prompted.

4. An active DOS partition will be created on the hard drive. At the restart prompt, press the

ISA# 77-011 - CELL-DYN® 1700 Service Disk Page 15Enter key to restart the system, leaving the Service Disk in the floppy drive.

5. At the A:> type FORMAT C: /S , and press the Enter key. Type [Y} after the destruction warning and press the Enter key. After successfully completing this step, press the Enter key after the Volume Label option is displayed (No need to enter a Volume Label). The hard drive is now ready to accept the installation disks.

6. With the Service Disk still installed, at the A>, type INSTALL and press the Enter key. The Service Disk will install all the Boot and DOS files on the hard drive. When this sequence is finished, power the analyzer down and remove the Service Disk. After 30 seconds, power on the analyzer.

7. After 30 seconds, power on the analyzer, when the Boot-Up sequence has completed the C> is displayed. Install the CD 1700 Installation Disk (P/N 9380468) in the floppy drive and type A:CS or A:NCS (CS for a closed mode analyzer) or (NCS for a non-closed mode analyzer) and press the Enter key. After successfully completing this step, power down the analyzer, remove the Installation Disk, and restart the analyzer (after 30 seconds).

* NOTE: Any Data or QC Log creation errors during the software installation will be corrected during the initial Boot-Up sequence after software installation has completed.

8. Place the Installation Disk is a safe location, reinstall all previous copied data (if available), and perform the Total Service Call Procedure to verify proper instrument operation.

** You may also use the Service Disk to erase an existing hard drive and reload the software is data corruption is suspected.

ISA# 77-011 - CELL-DYN® 1700 Service Disk Page 16END OF DOCUMENT

ISA# 77-010 - CELL-DYN® 1700 Revision D Software Page 17

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:CELL-DYN® 1700 Revision D Software

ISA#:77-010

ORIGINATOR: Jamie G. Besley

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C Stout 8/23/96

EFFECTIVITY DATE:23-AUG-96

CELL-DYN is a registered trademark of Abbott Laboratories.

The CELL-DYN® 1700 Revision D Software has been released for distribution in China only. The revisionincludes a PCT (plateletcrit) and PDW (platelet distribution width) print option for the graphics and ticket printer Setup screens.

END OF DOCUMENT

ISA# 77-008 - Cubitainer Case Changes Page 18

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:Cubitainer Case Changes

ISA#:77-008

ORIGINATOR: Jamie G. Besley

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C. Stout 4/24/96

EFFECTIVITY DATE:24-APR-96

CELL-DYN is a registered trademark of Abbott Laboratories.

I. DISTRIBUTION:International and USA

II. PURPOSE:The design of the 20L and 10L Reagent Cubitainer Case has been changed. The plastic carrying handle has been replaced by side perforated handles for easier handling. The plastic spout cover

ISA# 77-008 - Cubitainer Case Changes Page 19has been replaced by a tamper proof perforated circle above the spout. The internal bladder and cap will remain unchanged.

END OF DOCUMENT

ISA# 77-006 - New Vacuum Pump Configuration Page 20

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:New Vacuum Pump Configuration

ISA#:77-006

ORIGINATOR: Mark A. Allen

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C Stout

EFFECTIVITY DATE:05-DEC-95

CELL-DYN is a registered trademark of Abbott Laboratories.

In Accordance with ECO # 16800, the Vacuum Pump configuration has been changed to add an assembly which provides an in-line filter to add condensation protection for the pump. All new vacuum pumps will have this assembly and can be ordered using the current catalog number.

Removal and Replacement Procedure

1. The tubing assembly for the modified vacuum pump assembly is shown below:

ISA# 77-006 - New Vacuum Pump Configuration Page 21

1. 50 1. 5 0 1. 5 0

1. 5 0

3108201

9130548 9130548 3106532

3108001

31065328270040501

3108173

*P/N 9130548 50 ft Roll Tubing

2. Disconnect the following tubing from the vacuum pump:

a. Remove the tubing that comes from the vacuum accumulator to the vacuum pump inlet port.

b. Remove the coiled tubing assembly that connects the vacuum pump outlet port to the fitting on the bottom of the instrument chassis.

3. Disconnect the vacuum pump power cord "P2".

4. Unscrew and remove the vacuum pump assembly.

5. Check the tubing coming from the accumulator. If the tubing is deformed, cut 1/2 inch off the

ISA# 77-006 - New Vacuum Pump Configuration Page 22end as this will assure a better connection to the "T" fitting on the tubing assembly (as shown above).

6. Connect the tubing from the outlet port of the vacuum pump to the fitting on the bottom of the instrument chassis and then secure the vacuum pump assembly.

7. The in-line filter needs to be inspected and replaced at every PM.

END OF DOCUMENT

ISA# 77-004 - New Optical Sample Probe Assembly Page 23

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:New Optical Sample Probe Assembly

ISA#:77-004

ORIGINATOR: Mark A. Allen

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C Stout

EFFECTIVITY DATE:05-DEC-95

CELL-DYN is a registered trademark of Abbott Laboratories.

In Accordance with ECO # 16656; the sample probe assembly has been changed from a microswitch setup to an optical sensor set up. This was done to increase the reliability of the sample probe assembly. This can be ordered using the same catalog number. Also it should be noted that the material for the washblock has been changed from Acrylic to Ultem (ECO #16633). If a system has had the sample arm upgrade then the washblock can be ordered separately (P/N 8921025301). The old style sample probe assembly and wash block can no longer be obtained.

ISA# 77-004 - New Optical Sample Probe Assembly Page 24END OF DOCUMENT

ISA# 77-003 - PLT LRI Problems Page 25

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:PLT LRI Problems

ISA#:77-003

ORIGINATOR: David Bannister

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C Stout

EFFECTIVITY DATE:01-OCT-95

CELL-DYN is a registered trademark of Abbott Laboratories.

It has been found in the Failure Analysis Department that several instruments experiencing PLT LRI flags may be corrected by simply ensuring that the mounting screws for the PRE-AMP cover are all securely fastened to the standoffs. This problem has been noted along with the standoffs being loose, that should be tightened down also.

END OF DOCUMENT

ISA# 77-002 - New I/O Circuit Board Page 26

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:New I/O Circuit Board

ISA#:77-002

ORIGINATOR: Bob Horn

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C Stout

EFFECTIVITY DATE:09-JAN-95

CELL-DYN is a registered trademark of Abbott Laboratories.

Per ECO 16421, the I/O circuit boards (DFI model DIO-500AT) previously used in the Data Station and the 1700 CPU Assembly are no longer available. The new board is a DFI model DIO-550 and functions as a direct replacement. The current part number (8170000701) will not change. For the 1700 this will apply only to the Ticket Printer interface card.

The new I/O circuit board is slightly different in appearance, it can be identified by the DB 9 and DB 25 connectors that are now white. To set the addresses the new board now uses jumpers instead of DIP

ISA# 77-002 - New I/O Circuit Board Page 27switches. Refer to the attached diagram to set the addresses on the new I/O board. Ignore the setup for the Graphic Printer.

ISA# 77-002 - New I/O Circuit Board Page 28Figure 5A

I/O, DIO-550G+Jumper Settings

Graphics Printer COM1, LPT1settings

Ticket Printer

COM2, LPT2settings

ISA# 77-002 - New I/O Circuit Board Page 29

END OF DOCUMENT

ISA# 77-001 - New 2.5 ml Syringe Page 30

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:New 2.5 ml Syringe

ISA#:77-001

ORIGINATOR: Randy Lloyd

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C. Stout

EFFECTIVITY DATE:01-JUL-95

CELL-DYN, IMx, TDx, and ADx are registered trademarks of Abbott Laboratories.

A new 2.5ml syringe has been approved for use on all CELL-DYN® instruments. The new style syringe, manufactured by Abbott, is a double sealed glass barrel type. It is the same design used on the IMx®, TDx®, and ADx® analyzers, except for the flange at the lower end of the barrel. The new Abbott syringe has proven to be more reliable than the commercial syringe previously used. It is also assembled by Abbott making it more cost effective. The new syringe (lyse syringe), mounts exactly the same way as the old syringe. The list number (28561-01), will remain the same.

ISA# 77-001 - New 2.5 ml Syringe Page 31

Seal

O-ring

Plunger

Leur lock

Barrel

2.5ml Syringe Assembly (LN 28561-01)

END OF DOCUMENT

ISA# 77-017 - New Hard Drive Page 1

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:New Hard Drive

ISA#:77-017

ORIGINATOR: Randy Lloyd

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:David Bannister 05/18/98

EFFECTIVITY DATE:26-MAY-98

CELL-DYN® is a registered trademark of Abbott Laboratories

A new model hard drive has been approved for use on all DOS based Cell-Dyn hematology instruments that use a hard drive. The following model numbers are included in this update.

Cell-Dyn® 3000CS (68)Cell-Dyn® 3000SL (95)Cell-Dyn® 3500CS (96)Cell-Dyn® 3500SL (97)Cell-Dyn® 3200CS (80, not launched)Cell-Dyn® 3200SL (109)Cell-Dyn® 1700 (77)Cell-Dyn® 1700CS (79)

The new hard drive is the latest version of the Quantum Fireball SE® series 3.5" IDE drive. This new model is being introduced as a direct replacement for all previous models of hard drive which are now being discontinued. The new drive can be identified by the marking on the label (SE21A012 or SE21AP4). Due to common application to all CD3000 series, and CD1700 instruments, all hard drives will now be stocked without having any software installed on them. When a replacement hard drive is needed in the field, the drive will need to be partitioned and formatted by the FSR before the system software can be installed. The catalog number will be (8200570701). The old catalog number of (8200570601) is now obsolete. The new drive is a 2.1 GB capacity drive, but the extra storage capacity does not provide any advantage to the system. The CD1700 will use the same CMOS "user" settings as the previous drive.

ISA# 77-017 - New Hard Drive Page 2

Cell-Dyn 1700 Hard Drive Installation Instructions:

Note: The CD1700 does not have a setup disk to backup nonvol information. If it is possible to access the existing hard drive, it will save time and trouble to print out the calibration and dilution factors, QC file setups, etc for later use before changing the drive.

1. Install the new hard drive using the existing hardware and cables.2. Power up the system and verify the CMOS settings. The new drive can use the same settings as the previous drive. Cyl =

1024, HD = 16, WP = 0, LZ = 1025, Sec = 63.3. Place the CD1700 Service Disk (9380474) into the floppy drive, and reboot the system. The A> prompt will appear.4. At the A> prompt type FDISK <enter> .5. At the first menu choose option [1] <enter>, "Create DOS Partition or Logical DOS Drive" .6. At the next menu choose option [1] <enter>, "Create a Primary DOS Partition" .If you get the message "Primary DOS

partition already exists", press [ESC] to exit. Then go to step 9. 7. Answer [Y], <enter> to use maximum size for the hard drive DOS partition and make it active.8. When partition is complete, exit FDISK by pressing [ESC].9. At the A> prompt type FORMAT C: /S <enter>. Answer [Y] <enter> to the question "are you sure ?". When complete answer

[N] <enter> to the question "format another ?".10. At the A> prompt type INSTALL <enter>. This will install DOS and system files. 11. When complete, remove the service disk and reboot to the C> prompt. Insert the customer's system software disk into the

floppy drive and type A:NCS <enter> or (A:CS <enter> if CS).12. When installation is complete, remove the diskette and power off, then reboot the system.13. If printouts are available, re-enter the data from the printouts. If not the instrument will need to be calibrated from scratch.

End of Procedure

End of Document

ISA# 77-016 - Lyse Reagent for Japan Page 3

ABBOTTADD INSTRUMENT SERVICE ADVISORY

SUBJECT:Lyse Reagent for Japan

ISA#:77-016

ORIGINATOR: Laurie A. Wilmot

PRODUCT: CELL-DYN® 1700 (77)

APPROVED:John C. Stout

EFFECTIVITY DATE:03/31/98

CELL-DYN is a registered trademark of Abbott Laboratories

This ISA is to inform the Field that there is a new Lyse reagent developed specially for use in Japan with the CD1700 series analyzers. This new reagent is formulated for use ONLY in Japan and should NOT be used on any CD1700 analyzer outside the country of Japan. The reagent comes in a one liter bottle with a blue and white label and has a list number of 03H59-01.

DO NOT USE THIS REAGENT OUTSIDE THE COUNTRY OF JAPAN.