Upload
dongoc
View
220
Download
2
Embed Size (px)
Citation preview
1
Advances in biosensors: principle, architecture and applications 1
2
Veeradasan Perumal, Uda Hashim 3
4
Institute of Nano Electronic Engineering (INEE), University Malaysia Perlis (UniMAP), 5
Perlis, Malaysia 6
7
8
Correspondence to: 9
Veeradasan Perumal, 10
Biomedical Nano Diagnostics Research Group, 11
Institute of Nano Electronic Engineering (INEE), 12
University Malaysia Perlis (UniMAP), 13
01000 Kangar Perlis, Malaysia 14
E-mail: [email protected] 15
Tel: 04-9798580 16
Fax’s: 04-9798578 17
18
19
20
Summary 21
The ability to detect pathogenic and physiologically relevant molecules in the body with high 22
sensitivity and specificity offers a powerful opportunity in early diagnosis and treatment of 23
diseases. Early detection and diagnosis can be used to greatly reduce the cost of patient care 24
associated with advanced stages of many diseases. However, despite their widespread clinical 25
use, these techniques have a number of potential limitations. For example, a number of 26
diagnostic devices have slow response times and are burdensome to patients. Furthermore, 27
these assays are expensive and cost the health care industry billions of dollars every year. 28
Therefore, there is a need to develop more efficient and reliable sensing and detection 29
technologies. A biosensor is commonly defined as an analytical device that uses a biological 30
2
recognition system to target molecules or macromolecules. Biosensors can be coupled to a 1
physiochemical transducer that converts this recognition into a detectable output signal. 2
Typically biosensors are comprised of three components: (1) the detector, which identifies the 3
stimulus; (2) the transducer, which converts this stimulus to a useful output; and (3) the signal 4
processing system, which involves amplification and display of the output in an appropriate 5
format. The goal of this combination is to utilize the high sensitivity and selectivity of 6
biological sensing for analytical purposes in various fields of research and technology. Here 7
we review some of the main advances in this field over the past few years, explore the 8
application prospects, and discuss the issues, approaches, and challenges, with the aim of 9
stimulating a broader interest in developing biosensors and improving their applications in 10
medical diagnosis. 11
12
Key words: transducer; bioreceptor; enzyme; antibody; DNA; pathology, diagnosis 13
14
15
INTRODUCTION 16
17
Over the past decade, many important technological advances have provided us with the tools 18
and materials needed to construct biosensor devices. Since the first invention of Clark Oxygen 19
Electrode sensor, there are many improvements in sensitivity, selectivity, and multiplexing 20
capacity of modern biosensor. Before the various types of biosensor technologies and 21
application are discussed, it is first important to understand and define ‘biosensor’. A 22
biosensor according to IUPAC recommendations 1999, a biosensor is an independently 23
integrated receptor transducer device, which is capable of providing selective quantitative or 24
semi-quantitative analytical information using a biological recognition element (Thevenot et 25
al. 1999). Essentially it is an analytical device, which incorporating a biological or biological 26
derived recognition element to detect specific bio-analyte integrated with a transducer to 27
convert a biological signal into an electrical signal (Lowe 2007). The purpose of a biosensor 28
is to provide rapid, real-time, accurate and reliable information about the analyte of 29
interrogation. Ideally, it is a device that is capable of responding continuously, reversibly, and 30
does not perturb the sample. Biosensors have been envisioned to play a significant analytical 31
3
role in medicine, agriculture, food safety, homeland security, bioprocessing, environmental 1
and industrial monitoring (Luong et al. 2008). A biosensor consists of three main elements, a 2
bioreceptor, a transducer and a signal processing system (David et al. 2008). A biological 3
recognition element or bioreceptor is generally consists of an immobilized biocomponent that 4
is able to detect the specific target analyte (Kahn & Plaxco 2010). This biocomponents are 5
mainly composed of antibodies, nucleic acids, enzyme, cell and etc. Transducer in the other 6
hand is a converter. The reaction between the analyte and bioreceptor bring about a chemical 7
changes such as the production of a new chemical, release of heat, flow of electrons and 8
changes in pH or mass. The biochemical signal is converted into an electrical signal by the 9
transducer. Some of the commonly used transducers have been discussed in 3.0. Eventually, 10
signal processing where the electrical signal is amplified and sent to a microelectronics and 11
data processor. A measurable signal is produced, such as digital display, print-out or optical 12
changes. Figure 1 shows a schematic diagram of the typical components in a biosensor. There 13
is a need for simple, rapid and reagentless method for specific determination, both qualitative 14
and quantitative, of various compounds in various applications (Shantilatha et al. 2003). 15
Hence, it is paramount to have fast and accurate chemical intelligence which particularly 16
conspicuous in human health care. 17
18
Fig. 1.0. Biosensor operation. 19
20
21
22
23
4
CLASSIFICATION OF BIOSENSOR 1
2
Biosensors can be classified either by the type of biological signaling mechanism they utilize 3
or by the type of signal transduction they employ. Figure 2.0 shows the different categories of 4
biosensor. 5
6
7
8
Fig. 2.0. Different categories of biosensor. 9
10
11
BASED ON BIOLOGICAL SIGNAL 12
13
The bioreceptor or biological recognition element is the significant distinguishing feature of a 14
biosensor. The bioreceptor compromises the recognition system of a sensor towards the target 15
analyte. Essentially it is crucial for a bioreceptor to be selective and sensitive towards the 16
specific target analyte to prevent the interference by other substance from sample matrix 17
(Lowe 2007). Generally biosensors can be classified by the type of biological signaling 18
mechanism they utilize. The biological signaling used by biosensors can be divided into five 19
major mechanisms (Fig. 3.0). Here, we will discuss each of these mechanisms in detail and 20
their application: 21
22
5
1
Fig. 3.0. Methods of biosensing with various biological signal mechanism: (a) antibody/antigen; (b) 2
enzyme catalyze; (c) nucleic acid; (d) cell-based; (e) biomimetic. 3
4
Enzyme based sensor 5
Enzyme based biosensor are the earliest biosensor among all the other biosensor, these 6
biosensor is first introduced by Clark and Lyons in 1962 an amperometric enzyme electrode 7
for glucose sensor which use ‘soluble’ enzyme electrode (Shantilatha et al. 2003). Since the 8
first biosensor, enzyme based biosensor has face a massive growth in usage for various 9
application till present. Enzymes are very efficient biocatalysts, which have the ability to 10
specifically recognize their substrates and to catalyze their transformation. These unique 11
properties make the enzymes powerful tools to develop analytical devices (Leca-Bouvier and 12
Blum 2010). Enzyme-based biosensors associate intimately a biocatalyst-containing sensing 13
layer with a transducer. Enzyme based biosensor working principal is based on catalytic 14
6
action and binding capabilities for specific detection (David et al. 2008). The enzyme based 1
biosensors were made of enzyme as bioreceptor which is specific to detect targeted analyte 2
form sample matrix. The lock and key and induced fit hypothesis can apply to explain the 3
mechanism of enzyme action which is highly specific for this type of biosensor. This specific 4
catalytic reaction of the enzyme provides these types of biosensor with the ability to detect 5
much lower limits than with normal binding techniques. This high specificity of enzyme–6
substrate interactions and the usually high turnover rates of biocatalysts are the origin of 7
sensitive and specific enzyme-based biosensor devices (Leca-Bouvier and Blum 2010). 8
Ideally enzyme catalytic action can be influence by several factors such as the concentration 9
of the substrate, temperature, presence of competitive and non-competitive inhibitor and pH 10
(Cass 2007). Essentially the Michaelis-Menten equation can be used to further explain the 11
detection limit of enzyme based biosensor (Parkinson and Pejcic 2005). Glucose oxidase 12
(GOD) and horseradish peroxidase (HRP) are the most widely used enzyme based biosensor 13
that has been reported in literature. However, some recent studies have shown that enzyme 14
based biosensor can be used to detect cholesterol, food safety and environmental monitoring, 15
heavy metals and also Pesticides (Amine et al 2006, Zapp et al. 2011, Nomngongo et al. 2012, 16
Soldatkin et al. 2012, Ju and Kandimalla 2008). Moreover, a recent studies reported the used 17
of enzyme catalytic implication incorporating with nucleic acid biosensor for DNA detection 18
(He et al. 2011, Lin et al. 2011). 19
20
Immunosensors 21
An antibodies based biosensor was applied for the first time to detection in the 1950s, opening 22
the doors to the possibility of immuno-diagnosis (Donahue and Albitar 2010). Since then, 23
there have been vigorous effort made to develop immunosensor which composed of 24
antigen/antibody as bioreceptor as a tool for clinical diagnostics (Conroy et al 2009; Orazio 25
2011). An antibody is ‘Y’ shaped immunoglobin (Ig) that is made up of two heavy chains (H) 26
and two light chains (L). However some of human antibodies form dimeric or pentameric 27
structure by utilizing disulphide bonds and an extra protein called the joining or J- chain 28
(Wood 2006, Pohanka 2009). Each of the chain has a constant and variable part. The variable 29
part is specific to the antigen that is bind with corresponding antigen which is highly specific 30
and selective (Conroy et al 2009, Donahue and Albitar 2010). Hence, an immunosensor which 31
composed of antigen as bioreceptor utilizes the ability of antibody to bind with corresponding 32
antigen which is highly specific, stable, and versatile. The specificity of an antibody towards 33
7
the binding side of its antigen is a function of its amio acids (Fowler et al. 2008). Those days, 1
there are two type of detection method which frequently used in immunosensor which are 2
optical and electrochemical. However optical detection transduction method has suffered from 3
poor sensitivity when coupled with radioimmunoassay, the short half-life of radioactive 4
agents, concerns of health hazards, and disposal problems. Electrochemical detection 5
overcomes problems associated with other modes of detection of immunoassays and 6
immunosensors. In contrast, electrochemical immunoassays and immunosensors enable fast, 7
simple, and economical detection that is free of these problems (Fowler et al. 2008). 8
However, recent advance in science and technology has created an optical transduction 9
method a new path towards highly sophisticated automated instrument. Hence, Optical and 10
electrochemical detection method are gaining mutual importance for development of 11
immunosensor (Shankaran et al. 2007, Bhatta et al. 2010). According to Ramirez et al. 2009, 12
Immunosensors have been envisioned to play an important role in the improvement of public 13
health by providing applications for which rapid detection, high sensitivity, and specificity are 14
important, in areas such as clinical chemistry, food quality, and environmental monitoring. 15
The development of immunosensor for bacteria and pathogen detection has gained a great 16
deal of attention due to its application in the point of care measurement (POC) (Skottrup et al. 17
2008, Barton et al. 2009, Braiek et al. 2012, Holford et al. 2012). Some recent studies shown 18
that immunosensor is widely explored toward the detection of cancer/tumor. Since the 19
traditional diagnostics method is poor in sensitivity, selectivity and time consuming, 20
immunosensor are become promising tools for cancer detection at early stages of cancer 21
[Ushaa et al. 2011]. 22
23
DNA/Nucleic acid sensor 24
The use of nucleic acids sequence for the specific diagnostics application has developed since 25
the early 1953 and still growing widely (Liu A. et al. 2012). The highly specific affinity 26
binding’s reaction between two single strand DNA (ssDNA) chains to form double stranded 27
DNA (dsDNA) is utilized in nucleic acids based biosensor which appoint the nucleic acids as 28
biological recognition element. This method has promoted the development of DNA based 29
sensor from the traditional method such as coupling of electrophoretic separations and radio 30
iso-tropic which are high cost, hazardous, time consuming and etc. (Parkinson and Pejcic 31
2005). This biosensor working principal is based on recognition of the complementary strand 32
by ssDNA to form stable hydrogen bond between two nucleic acids to become dsDNA. In 33
8
order to achieve this, an immobilized ssDNA is used as probe in bioreceptor which the base 1
sequence is complementary to the target of interest. Exposure of target to the probe which 2
results in hybridization of complementary ssDNA to form dsDNA will result in producing 3
biochemical reaction that allows transducer amplified the signal into electrical one. 4
Subsequently, literature shows that the present of some linker such as thiol or biotin is needed 5
in the effort to immobilize the ssDNA onto the sensing surface (Cagnin et al. 2009, Larzerges 6
et al. 2012). An important property of DNA is that the nucleic acid ligands can be denatured 7
to reverse binding and the regenerated by controlling buffer ion concentration (Parkinson and 8
Pejcic 2005). The nucleic acid biological recognition layer which incorporates with transducer 9
is easily synthesizable, highly specific and reusable after thermal melting o the DNA duplex 10
(Teles and Fonseca 2008). In addition, this biosensor possesses a remarkable specificity to 11
provide analytical tools that can measure the presence of a single molecule species in a 12
complex mixture (Brett 2005). DNA based biosensor has potential application in clinical 13
diagnostic for virus and disease detection (Chua et al. 2011, Lui C. 2009, Thuy et al. 2012). 14
Moreover, Yeh et al. 2012 recently has reported optical biochip for bacteria detection based 15
on DNA hybridization with detection limit of 8.25 ng/ml. However, electrochemical 16
transduction is most abandon method used to studying DNA damage and interaction which 17
reported in literature. The development of electrochemical DNA biosensor has received a 18
great deal of attention lately and this has largely been driven by the need to developed rapid 19
response, high sensitivity, good selectivity and experimental convenience (Liu A. et al. 2012). 20
21
Cell based sensor 22
Cell based sensor are the type of biosensor, which use living cell as the biospecific sensing 23
element and are based on the ability of living cell to detect the intracellular and extracellular 24
microenvironment condition, physiological parameter and produces response through the 25
interaction between stimulus and cell (Wang and Liu 2010). Microorganisms such as bacteria 26
and fungi can be used as biosensors to detect specific molecules or the overall ‘‘state’’ of the 27
surrounding environment (Acha et al. 2010). Furthermore, proteins that are present in cells 28
can also be used as bioreceptors for the detection of specific analyte (Acha et al. 2010). 29
Essentially, living cell based biosensor is a unique biosensor in contrast to other type of 30
biosensor that contains materials that extracted from living things. These types of biosensor 31
has leveraged of pros and cons. The detection limit of this biosensor is mainly determined by 32
the natural environmental conditions in which the cell can stay alive for long period where 33
9
need the control the physical and chemical parameter of environment. However the major 1
limitation with cell based biosensor are the stability of the cell, which depends on various 2
conditions such as the sterilization, lifetime, biocompatibility and etc. Another issue that 3
governs the success of a cell based sensor are depends primarily ion selectivity, in which cell 4
based sensor has poor selectivity of microbial sensor due to the multireceptor behavior of the 5
intact cells (Belkin and Gu 2010). Despite these pitfalls the cell based biosensor still favorable 6
among the researcher due to the advantages over the enzymes based biosensor. The cell based 7
biosensor are less sensitive to inhibition by solutes and are more tolerant of suboptimal pH 8
and temperature values than enzyme based biosensor, though they must not exceed the narrow 9
range in case of the cells dying, a longer lifetime can be expected than with the enzymatic 10
sensors and they are much cheaper because active cells do not need to be isolated (Struss et al. 11
2010). Literature revealed that, cell based sensor have become an emerging tools for medical 12
diagnostics (i.e. such as disease detection), environmental analysis, food quality control, 13
chemical-pharmaceutical industry and drugs detection (Veiseh et al. 2007, Banerjee and 14
Bhunia 2009, Banerjee and Bhunia 2010, Wang, T et al. 2010, Melamed et al. 2012, Bohrn et 15
al. 2012, Liu, Q et al. 2012 ). Similarly, Shinde et al. 2012 conclude that cell based biosensor 16
is envisioned to be an emerging frontier in the area of nano-diagnotics due to their attractive 17
characteristic. 18
19
Biomimetic sensor 20
A biomimetic biosensor is an artificial or synthetic sensor that mimics the function of a 21
natural biosensor. These can include aptasensors, where aptasensors use aptamers as the 22
biocomponent (David et al. 2008). Aptamers were reported for the first time in the early 23
1990s where described as artificial nucleic acid ligands. Aptamers were thus chemically 24
related to nucleic acid probes, but behaved more like antibody and showing surprising 25
versatility compared to other bio-recognition components (Schneider et al 2010, Brys et al. 26
2007). Aptamer are synthetic strands of nucleic acid that can be designed to recognize amino 27
acids, oligosaccharides, peptides, and proteins. An aptamer has few advantages over antibody 28
based biosensor such as high binding efficiency, avoiding the use of animal (i.e reduced 29
ethical problem), smaller and less complex, and etc. However, common challenge facing 30
aptasensor is that they inherent the properties of nucleic acids such as structural pleomorphic 31
and chemical simplicity which reduced the assay efficiency and also increase its production 32
cost. Subsequently, some effort has been directed towards characterization and optimization 33
10
of aptamer to overcome this limitation. Aptamer properties such as their high specificity, 1
small size, modification and immobilization versatility, regenerability or conformational 2
change induced by the target binding have been successfully exploited to optimize a variety of 3
bio-sensing formats (Schneider et al 2010). Aptamer based biosensor has been widely used in 4
various application. Recently sufficient progress has been made in biomimetics sensor and 5
aptasensor for clinical application (Vallet-Regi and Arcos 2008). This including clinical 6
diagnostics to detect pathogen, virus and infectious disease (Strehlitz et al 2008, Toress-7
Chavolla and Alocilja 2009, Wang Y. Et al. 2012, Weng et al. 2012). 8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
11
BASED ON TRANSDUCTION 1
Biosensor are normally categorized according to the transduction method they employ (figure 2
2.0). The transducer is component of biosensor, which has important role in the signal 3
detection process. Hence transducer can be defined as a device that convert a wide range of 4
physical, chemical or biological effect into an electrical signal with high sensitivity and 5
minimum disturbance to the measurand (Lowe 2007). There are various number of transducer 6
methods has developed over the decade; however recent literature review has highlighted the 7
most common methods that available (figure 2.0). This group can be further divided into two 8
general categories; labeled and label free type, which label free type biosensor growing 9
further recently (Parkinson and Pejcic 2005). 10
11
12
Fig. 4.0: Methods of biosensing with various transducer mechanisms: (a) Optical; (b) 1
Electrochemical; (c) Calorimetric - thermal; (d) Piezoelectric- mass sensitive. 2
3
Electrochemical biosensor 4
Electrochemical sensor in which an electrode is used as the transduction element, represent an 5
important subclass of sensor. According to IUPAC recommendation 1999, an electrochemical 6
biosensor is self-contained integrated device, which is capable of providing specific 7
quantitative or semi-quantitative analytical information using a biological recognition element 8
(biochemical receptor) which is retained in direct spatial contact with an electrochemical 9
transduction element (Thevenot et al. 1999, 2001). Electrochemical biosensors measure the 10
current produced from oxidation and reduction reactions. This current produced can be 11
correlated to either the concentration of the electroactive species present or its rate of 12
production/consumption. The resulting electrical signal is related to the recognition process 13
by target and analyte, and is proportional to the analyte concentration. A review by Wang et 14
al. (2008) has revealed that electrochemical immunosensors are gradually increasing in 15
popularity in clinical analysis and this is partly due to improved sensor design. Similarly, 16
Maria et al. (2008) revealed that the electrochemical immunosensor is a promising alternative 17
compared to existing laboratory method. Wang (2006) has reported the use of 18
electrochemical-based device for point-of-care cancer diagnostics which has brought the 19
electrochemical biosensor to new level in clinical diagnostics. Subsequently, all this report 20
has reveal the electrochemical biosensor has advantages such as fast, simple, low cost, high 21
sensitivity, and relatively simple instrumentation (Grieshaber et al. 2008, Mono et al. 2012). 22
Depending upon the nature of electrochemical changes detection during a biorecognition 23
event, electrochemical biosensors fall into one of four categories: amperometric, 24
potentiometric, impedance and conductometric. 25
26
Amperometric 27
Amperometric sensor are based on the measurement of current as a function of time resulting 28
from the oxidation and reduction of an electroactive species in a biochemical reaction that 29
mainly depends on the concentration of an analyte with a fixed potential. There are three type 30
of electrode that usually employed in a amperometric sensor; The working electrode which 31
usually made of gold (Au), carbon (c), platinum (Pt), A reference electrode usually silver or 32
13
silver chloride (Ag/AgCl) which has a fixed potential that controlled the potential of the 1
working electrode and the third electrode which called the counter or auxiliary where included 2
sometime to help measure the current flow (Wang J. 1996, Wang Y. et al. 2008). As the 3
certain molecule are oxidized or reduced (redox reactions) at inert metal electrodes, electrons 4
are transferred from the analyte to the working electrode or to the analyte from the electrode 5
(Banica 2012). The direction of flow of electrons depends upon the properties of the analyte 6
which can be controlled by the electric potential applied to the working electrode. If the 7
working electrode is driven to a positive potential an oxidation reaction occurs, and the 8
current flow depends on the concentration of the electroactive species (analyte) diffusing to 9
the surface of the working electrode. Similarly, if the working electrode is driven to a 10
negative potential then a reduction reaction occurs (Parkinson and Pejcic 2005). A third 11
electrode called the counter electrode is often used to help measure the current flow. There are 12
three generation of biosensor : first generation biosensor where the normal product of the 13
reaction diffuses to the transducer and causes the electrical response, second generation 14
biosensors which involve specific “mediators” between the reaction and the transducer in 15
order to generate improved response, and third generation biosensors where the reaction itself 16
causes the response and no product or mediator diffusion is directly involved (Saxena and 17
Malhotra 2003, Wang Y. et al. 2008). The amperometric transduction can be integrated with 18
Enzyme, Nucleic acids and Immunosensor biological recognition element for various 19
application such as glucose detection, disease, environmental monitoring and etc. (Ivnitski et 20
al. 1998, Ivnitski et al. 2003, Amine et al. 2006, Wang J. et al. 2006, Wang Y. et al. 2008, 21
Fang et al. 2011). However, there are limitation using this biosensor where, the presence of 22
electroactive interference in sample matrix can cause the transducer generate the false current 23
reading (Rogers and Mascini 1999). There been various method identified to overcome this 24
limitation such as diluting the sample, coating the electrode with various polymer, change the 25
medium of analyte, adding mediator and etc. (Jelen et al 2002, Belluzo et al. 2007, Shinde et 26
al. 2012). 27
28
Potentiometric 29
Potentiometric biosensors rely on the use of ion-selective electrode and ion-sensitive field 30
effect transistor for obtaining the analytical information. In such sensor, the biological 31
recognition element converts the recognition process into a potential signal to provide an 32
analytical signal. Potentiometric transduction is first reported in the year 1969 where the 33
14
enzyme based sensor is used for urea detection (Schelfer and Shcubert 1992, Saxena and 1
Malhotra 2003). Potentiometric sensor consists of two electrode where working (or Indicator) 2
electrode to develops a variable potential from recognition process and reference electrode 3
(usually Ag/AgCl) which require to provide a constant half-cell potential. The working 4
principle of the potentiometric transduction is depends on the potential difference between the 5
indicator electrode and reference electrode which accumulate during recognition process in an 6
electrochemical cell when zero or negligible amount of current flow through the electrode. 7
The electrical potential difference or electromotive force (EMF) between two electrodes is 8
measured using a high impedance voltmeter. The working electrode is made of permselective 9
ion-conductive membrane which sometimes called an ion-selective electrode (ISE). The 10
measurement of potential response of an potentiometric device is governed by the Nernst 11
equation in which the logarithm of the concentration of the substance being measured is 12
proportional to the potential difference (Grieshaber et al. 2008, Wang Y. et al. 2008). 13
According to Bakker and Pretsch 2010 recent progress enabled development and application 14
of potentiometric sensors with limits of detection (LODs) in the range 10-8 to 10-11 M. These 15
LODs relate to total sample concentrations and are defined according to a definition unique to 16
potentiometric sensors. The ISEs has the largest number of application in clinical chemistry 17
particularly the determination of the biologically relevant electrolytes in physiological fluids, 18
as billions of measurement are performed each year all over the world (Makarychev-19
Mikhailov et al. 2008). This comes as no surprise considering that potentiometric transduction 20
mechanism is very attractive for the operation of biosensor due to its selectivity, simplicity, 21
rapid, low cost and maintenance free measurement. However, the device is still less sensitive 22
and often slower than the amperometric counterpart (Makarychev-Mikhailov et al. 2008). The 23
field effect transistor have been used for the fabrication of integrated biosensor and there are 24
many device derived from this principle such as ion-selective field effect transistor (ISFET), 25
electrolyte –insulator semiconductor (EIS), light addressable potentiometric sensor (LAPS), 26
CHEMFET, BioFET and ENFET (Saxena and Malhotra 2003). Essentially, the LAPS device 27
is an example optical/electrochemical hybrid technique which consists of n-type silicon doped 28
with phosphorus and an insulating layer that takes advantage of the photovoltaic effect in 29
present of light source as a light emitting diode (LED), flashes rapidly (Pohanka and Skladal 30
2008). A review by Koncki 2007 has revealed the application of potentiometric based 31
biosensor in biomedical analysis. The author concludes that bioaffinity-based biosensor using 32
potentiometric transduction is not reliable. However, LAPS devices has proved to be highly 33
successful for immunoassay and overcome the previous less sensitivity problem encounter by 34
15
potentiometric transduction (Wang et al. 2010, Jia et al. 2012). Recently, Jia et al. 2012 has 1
developed graphene oxide modified light addressable potentiometric sensor for ssDNA with 2
detection limit (LOC) of 1 pM to 10 nM. 3
4
Conductometric 5
In this method the analytical information is obtained by measuring of electrolyte conductivity, 6
which varies with the changes in the ionic species concentration. In other word, 7
conductometric based transduction provides information about the ability of an electrolyte 8
solutions to conduct an electric current between electrodes. The conductometric device made 9
of two electrodes which separated by certain distance or medium such as nanowire, etc. Most 10
of the work that is reported in the literature on conductometric sensor associated with 11
enzymes where the ionic strength, and thus the conductivity, of a solution between two 12
electrodes changes as a result of an enzymatic reaction. The AC supply is used to apply across 13
the electrode for conductivity measurement. Thus, the ionic composition changes and 14
provides a conductance which measure using an Ohmmeter. However, some recent studies 15
has shown the conductometric transduction employed in micro/nano electronic device such as 16
FET to directly monitor the changes in conductance of an electrode as a result of the 17
hybridization of DNA, complementary antibody-antigen pair, etc.(Arya et al. 2007, Hnaiein et 18
al. 2008, Tang et al. 2011). The major advantages of conductometric device are non-19
requirement of reference electrode, inexpensive, possibility of miniaturization and direct 20
electrical response (Grleshaber et al. 2008, Lee et al. 2012). Unfortunately, the 21
conductometric transduction measurement is an additive property, hence less sensitive 22
compared to the other electrochemical methods and strongly dependence of the response upon 23
buffer capacity (Wang 1996). A review by Arora 2012 revealed that, conductometric based 24
transducer have envisioned for detection of food borne pathogens. However, recent literature 25
shown that conductometric based transduction has received a great deal of attention and 26
overcome the previous less sensitivity problem. Subsequently, some research has been 27
directed toward the improvement of performance and sensitivity (Kannan et al. 2012). 28
29
Electrical impedance spectroscopy 30
Electrical impedance spectroscopy (EIS) based transduction method is not typically used 31
electrochemical detection; however this method has only recently become popular tools for 32
16
bioreceptor transduction. These method are known to be similar to electrochemical detection 1
an conductivity detection that scan the detection volume with an electrical frequency sweep, 2
in the range of 10khz and 10mhz (McGuinness and Verdonk 2009). Essentially, impedance 3
spectroscopy has major advantages on lower concentration detection. A recent study shows 4
the EIS based transduction has been employed in tumor growth detection in 100-600 fg/ml 5
with the sensitivity/detection limit of 100 fg/ml (Onur and Kemal 2011). The EIS type 6
measurement is suitable for real time monitoring since it is able to provide a label free or 7
reagentless detection (Parkinson and Pejcic 2005, Pejcic et al. 2006). In EIS measurement, the 8
sample is placed on the sensing device such as nanogap, and a controlled alternating voltage 9
is applied to the electrode, and the current flows through the sample are monitored. The 10
electrical impedance resulting from the sample is calculated as the ratio of voltage over 11
current. The resulting electrical impedance measurement has both a magnitude and a phase, a 12
complex number. For any time-varying voltage applied, the resulting current can be in phase 13
with the applied voltage (resistive behavior), or out of phase with it (capacitive behavior). The 14
EIS consists of 3 electrode system, a potentiostats and a frequency response analyzer (FRA). 15
The three electrode which are working electrode that provides the measurement of current, 16
counter electrode that provides current to cell and reference electrode for voltage 17
measurement. The potentiostats function as a high input impedance provider and to maintain 18
the voltage across the electrode (Barsoukov and Macdonald 2005). Eventually, the FRA is 19
incorporated to supply the excitation waveform and provide a very convenient, high precision, 20
wide band method of measuring the impedance (Barsoukov and Macdonald 2005). A recent 21
review by Yang and Bashir 2008 discussed progress and application of impedance biosensor 22
for foodborne pathogenic bacteria detection. Impedance spectroscopy has been widely used 23
by many research groups to detect cancer/tumor cell, virus, bacteria and pathogen (Bayoudh 24
et al. 2008, Diouani et al. 2008, Kukol et al. 2008, Hong and Jang 2012, Thi et al 2012, Ohno 25
et al. 2012). This biosensor can become a powerful tool in the imminent future for clinical 26
diagnostics (Pejcic and Marco 2006). 27
28
Optical based biosensor 29
Over the past decades, optical biosensor had made rapid advances and has been applied in a 30
number of important area including food safety, security, lifescience, environmental 31
monitoring and medical (Tsoka et al. 1998, Vedrine et al. 2003, Johansson et al 2008, Bhatta 32
et al. 2010a, Bhatta et al 2012, Md Muslim et al 2012). In medical, optical transducer have 33
17
been established in both routine medical diagnostic and for medical research application 1
(Caygill et al. 2010). The word “optrode” is a combination of the words “optical” and 2
“electrode” is sometimes used to define optic based device (Biran et al. 2008). This method of 3
transduction has employed in many class of biosensor due to the many different types of 4
spectroscopy such as absorption, fluorescence, phosphorescence, Raman, SERS, refraction 5
and dispersion spectroscopy (Abdulhalim et al. 2007). These transduction methods are 6
capable of measure different properties of target/analyte. Optical based biosensor able to 7
provide a label free, real time and parallel detection (Francia et al. 2005, Fan et al. 2008, 8
Bhatta et al. 2010b). The surface plasmon resonance or fluorescence which integrated with 9
optical fiber is most popular method available for optical based biosensing (Caygill et al 10
2010). It appears that sensor based on optical fiber principal has gaining the research interest 11
and being employed in biosensor studies for. 12
13
Surface plasmon resonance 14
Surface plasmon resonance (SPR) biosensor use surface plasmon waves (electromagnetic 15
wave) to detect changes when the target analyte interact with biorecognition element on the 16
sensor. In principle, when the SPR biosensor is exposed to any changes, it will induce 17
changes in the refractive index which used to measure or observed the reaction. The SPR 18
transducer are incorporate with biomolecule /biorecognition element which recognize and 19
able to interact with specific analyte (Mol et al. 2010). Hence when target analyte interact 20
with the immobilized biomolecule on the sensor surface, it produces a change in the refractive 21
index at the sensor surface (Mol et al. 2010). This, changes produce a variation in the 22
propagation constant of the surface plasmon wave and this variation is measure to produce 23
reading. A spectrophotometer is used to measure the absorption spectrum of sample. There 24
been various biorecognition element have been incorporates with SPR biosensor such as 25
protein, antibody-antigen, nucleic acids and enzyme (Endo et al. 2005, Mannelli et al 2006, 26
Nguyen et al. 2007, Nakamura et al. 2008, Park et al. 2009). An important feature of SPR 27
biosensor is that it is able to provide label-free sensing without radioactive and fluorescence 28
which makes it highly attractive for real time monitoring (Fan et al. 2008, Endo et al. 2008). 29
In addition, the SPR based transduction can be used to and interaction without exhibit any 30
special properties of fluorescence or characteristic absorption and scattering bands (Homola et 31
al. 2008). However, some reports suggest that these method has suffer from specificity due to 32
non-specific interaction with biorecognition element which wrongly correlated by these 33
18
biosensor (Homola et al. 2008). The SPR based transductions are not suitable for studying 1
small analytes. Because the SPR measures the mass of material binding to the sensor surface, 2
very small analytes (Mr
<1000) give very small responses (Merve 2001) . The recent 3
improvements in signal to noise ratio have made it possible to measure binding of such small 4
analytes (Merve 2001). SPR biosensors can effectively detect binding by molecules as small 5
as about 2 kDa, but smaller molecules generate insufficient changes in bound mass and so 6
cannot be directly measured adequately ( Mitchell and Wu 2010). To date, SPR has been 7
widely used in fundamental biological studies, health science research, drug discover, clinical 8
diagnosis and environmental and agriculture monitoring (Shankaran et al 2007). Several 9
articles have appeared in the literature reviewing the application of SPR based biosensor in 10
pathogen and disease detection (Skottrup et al. 2008, Caygil et al 2010). Recently, Springer et 11
al. 2010 demonstrated that the SPR based biosensor is able to detect short sequence of nucleic 12
acids (20 bases) characteristic for Escherichia coli down to femtomole level in 4 minute. 13
14
Chemiluminescence 15
Chemiluminescence can be describe as method of energy produce from chemical reaction 16
which produce a emission of light or usually describe as Luminescence (Li D. 2008). 17
Essentially, when a chemical reaction occur, the atom or molecule relaxes from excited state 18
to its ground state which then a luminescence is produce as side product of the reaction. 19
Hence, chemiluminescence can be used to detect specific biochemical reactions which occur 20
and this property has contributed for chemiluminescence based biosensor development. In the 21
chemiluminescence biosensor, the reaction between analyte and the immobilized biomolecule 22
which has been marked with chemiluminescence species will end in generating light as result 23
of biochemical reaction. This emitted light can be detected using a Photo Multiplier Tube 24
(PMT). A review by Dodeigne et al. 2000 and Zhang et al. 2005 has shown that 25
chemiluminescence is an emerging tool for diagnostics with extremely high sensitivity along 26
with the simple instrumentation, fast dynamic response properties, and wide calibration range. 27
Not with standings, chemiluminescence based transduction has been widely applied and 28
embraced for immunosensing and nucleic acid hybridization (Atias et al. 2009, Ding et al 29
2008, Holford et al. 2012, Guo et al. 2013). Similarly, a number of papers have shown that 30
chemiluminescence applications in clinical, pharmaceutical, environmental and food analysis 31
(Kricka 2003, Gámiz-gracia et al. 2009, Lara et al. 2010, Liu M. et al. 2010). These methods 32
were also incorporated with immunosensor and optical fiber for detection of dengue virus in 33
19
human (Atias et al. 2009). More importantly, it was revealed that this type of transduction has 1
detection limit of 5.5 × 10−13 M (Ding et al. 2008). However, chemiluminescence transduction 2
has few drawbacks such as less quantitative accuracy due to short lifetime and not suitable for 3
real time monitoring (Parkinson and Pejcic 2005, Zhang et al. 2005, Mathews et al. 2009). 4
5
Fluorescence 6
The term Luminescence as describe above is the product of atoms or molecule which relaxes 7
from excited state to its ground state. The various types of luminescence differ from the 8
source of energy to obtain the excited state. This energy can be supplied by electromagnetic 9
radiation (photoluminescence also termed as fluorescence or phosphorescence), by heat 10
(pyroluminescence), by frictional forces (triboluminescence), by electron impact 11
(cathodoluminescence) or by crystallization (crystalloluminescence) (Dodeigne et al. 2000). 12
Hence, the fluorescence requires external light source (short-wavelength light) to initiate the 13
electronic transitions in an atoms or molecule which then produce luminescence (Longer 14
wavelength light). Eventually, fluorescence based biosensor has incorporate with 15
flourochrome molecules which used to produce light during the biorecognition event (Daly 16
and McGrath 2003). Since most of the biological sensing element and most analyte does not 17
possess intrinsic spectral properties, the biorecognition event is transduced to optical signal by 18
coupling fluorescence an optically responsive reagents to the sensing elements (Biran et al. 19
2008). For example, the nucleic acid or antibodies is used to tag with flourochrome and 20
convert the hybridization interaction between two complementary DNA stands into an optical 21
signal (Ramanathan et al. 2005, Berdat et al. 2006, Schultz et al. 2008). In fact, fluorescence 22
technology has been widely applied and embraced for immunochemical sensing in the 23
medical field (Hong and Kang 2006, Moghissi et al. 2008). Fluorescence based biosensing 24
were optimized widely in environmental monitoring as reported by Védrine et al. 2003. 25
Nonetheless a recent article has been appear in the literature on the use of flourescene 26
marking DNA biochip for analysis of DNA- carcinogen adducts (Grubor et al. 2004). The 27
major drawbacks of fluorescence technology are additional complexity of time-resolved 28
instrumentation, in either the time or frequency domains or both and not suitable for real time 29
monitoring (Thompson 2008, Abdulhalim et al. 2007). 30
31
20
Optical fiber 1
Optical fiber or which sometimes called optrodes, has received considerable interest for 2
developing as biosensor particularly in lower detection limit (LOC) sensing application. 3
Optrodes based optical fiber biosensors are composed of few major components: a light 4
source; biorecognition element which immobilized; an optical fiber which responsible to 5
transmit the light and act as the substrate; and a detector (e.g., spectrophotometer) where the 6
output light signals is measure. Hence, when the target analyte interacts with biorecognition 7
elements at the surface of the fiber, a biochemical reaction takes place resulting in the changes 8
of optical properties. These changes can be collated to the analyte concentration. The light 9
source is transmitted through the fiber optic, where the biorecognition event takes place. The 10
same or different fiber is used to guide the output light to the detector. The current trends of 11
bio-sensing using optical fiber are gaining researcher interest due to its major advantages such 12
as miniaturized sensor with higher performances, fiber optical sensors with small size, high 13
sensitivity, fast response, high selectivity, and low detection limits (Lee B. et al. 2009, 14
Zhang,L. et al. 2011). Optical fiber based biosensor also have several advantages compared to 15
electrochemical and other biosensing such as higher sensitivity, safer, free from 16
electromagnetic interference, no reference electrode needed and suitable for real time 17
monitoring (Biran et al. 2008). In addition this type of transduction is flexible hence suitable 18
for remote sensing, single molecule detection and reagentless (Biran et al. 2008). However, 19
there are several drawbacks such as poor stability of biorecognition elements, sensitive to 20
ambient light and etc. (Biran et al. 2008). Most of the work that is reported in the literature on 21
optical fiber biosensor applications has employed the optic fiber with other optical method 22
(Atias et al. 2009, Jang et al. 2009, Huang et al. 2009, Cecchini et al. 2012). Furthermore, 23
these methods have been used for various biomolecule detection such as antigen and nucleic 24
acid (Brogan et al. 2005, Leung et al. 2007, Amin et al. 2012). Thus, explained the versatility 25
of optical fiber biosensing. The optical fiber based biosensing has widely being used pathogen 26
and virus detection (Atias et al. 2009, Huang et al. 2009, Caygill et al 2010). Leung et al. 27
2008 has reported label free detection of DNA hybridization for pathogen detection using 28
fiber optic biosensor. 29
30
Piezoelectric based biosensor 31
Piezoelectricity can be explain as a linear interaction between mechanical and electrical 32
systems in non-centric crystal or similar structure which first discover by Curie brothers in 33
21
1880 (Katzir 2006, Tichy et al. 2010). Essentially, the piezoelectric based biosensor working 1
on the principal that an oscillating crystal resonates at a natural resonance frequency (Steinem 2
and Janshoff 2007, Tichy et al. 2010). The fundamental elements in a biosensor are transducer 3
and biorecognition element. Hence, in piezoelectric biosensor the transducer is made of 4
piezoelectric material (e.g., quartz) and the biosensing material that coated on the 5
piezoelectric material which vibrate at the natural frequency. The frequency is control by the 6
external electrical signal which produces a certain value of current, when the target analyte is 7
exposed to the sensing material the attachment/ reaction will cause the frequency shift which 8
will produce changes in current reading that can be collated to the mass of the analyte of 9
interest. There are two main types of piezoelectric sensors: bulk wave (BW) and surface 10
acoustic wave (SAW). However, literature shows piezoelectric sensors are not receive much 11
attention and inferior compared to electrochemical and optical based biosensing. 12
13
Bulk wave (BW) and Surface acoustic wave (SAW) 14
The bulk wave, quartz crystal microbalance and surface acoustic wave transducer is 15
fundamentally based on the piezoelectric effect. The unique properties of piezoelectric 16
material are utilized in this type of sensing. Quartz is the most commonly used piezoelectric 17
since it is cheap, can be processed to yield single crystal and can withstand chemical, thermal 18
and mechanical stress; however, there is report that lithium niobate and lithium tantalate can 19
aslo be used (Tichy et al. 2010). A recent review has shown that this method is very appealing 20
when integrate with Microelectromechanical systems (MEMS) for biosensing application 21
(Nicu et al. 2005). In addition the review states that this type of transduction is suitable for 22
sensitive, portable and real time biosensing (Nicu et al. 2005). Piezoelectric transducer has 23
been widely applied and embraced for immunosensing application (Vaughan et al. 2007, 24
Serra et al. 2008, Tothill 2009). Some report suggests that the piezoelectric transducer is 25
suitable for DNA and protein detection with detection limit of 1 ng/cm2 (Nirschl et al. 2009). 26
Several article have appear in the literature reporting the use of piezoelectric sensor in various 27
application such as cholera toxin diagnostic detection, hepatitis B, hepatitis C, food borne 28
pathogen detection and etc. (Skládal et al. 2004, Yao et al 2008, Serra et al. 2008, Chen et al. 29
2008, Chen et al. 2010). More importantly, it was revealed that piezoelectric is very sensitive 30
method, noting that a detection limit of 8.6 pg/l was obtained for hepatitis B virus DNA and 31
25ng/mL for cholera toxin detection (Yao et al. 2008, Chen et al. 2010). The advantages 32
using this type of transduction is the real time monitoring, label free detection and simplicity 33
22
of use (Dell’Atti et al. 2006, Chen et al. 2010). However, there are some drawbacks need to 1
overcome such as specificity, sensitivity as well as interference reduction (Glynn and 2
O’Connor 2010). In addition, this type of transducer method involves format and calibration 3
requirement (Leca-Bouvier and Blum 2010). A recent review by Kim et al. 2011 has review 4
the principle and application of nano diagnostic for nanobiosensor. The review also 5
concluded, that application range of the quartz crystal has been gradually expanded, new 6
measuring techniques that use the quartz crystal as a transducer for chemical sensors and 7
biosensors have been also developed (Kim et al 2011). 8
9
Calorimetric based biosensor 10
First enzyme based biosensors introduce by Clark and Lyons in 1962 has inspired and attracts 11
researcher interest in developing calorimetric based transduction. In fact, almost all chemical 12
and biological reaction involves exchange of heat (Xie et al. 1999). Thus, the general idea of, 13
generation and absorption of heat resulting in all biochemical reaction has contributed to the 14
birth of calorimetric based biosensing device. Initially, the calorimetric transduction has been 15
employed for enzyme based sensor, and has subsequently been applied in cell and 16
immunosensor (Xie et al. 1999, Ahmad et al. 2010). The principal of calorimetric measured 17
the changes in temperature in the reaction between biorecognition element and a suitable 18
analyte. This change in temperature can be correlated to the amount of reactants consumed or 19
products formed (Xie et al. 1999). A review by Yakovleva et al has discuss that among all 20
various concept of thermal detection, enzyme thermistor (ET) has been research widely for 21
various application. The major advantages of this type of thermal detection are the stability, 22
increase sensitivity and possibility of miniaturation (Ahmad et al. 2010). In addition 23
calorimetric based biosensor can be easily miniaturised and integrated with microfluidic for 24
increased sensitivity (Zhang and Tadigadapa 2004). In the calorimetric device, the heat 25
change is measured using either a thermistor (usually metal oxide) or thermopile (usually 26
ceramic semiconductor). A review by Cooper 2003 has shown that this method is very 27
appealing for label free screening of biomolecule interaction. Some recent studies have shown 28
that this technique capable of rapidly detecting the DNA hybridization (Watterson et al. 2002, 29
Buurma and Haq 2008, Paul et al. 2010, Xi et al. 2010). Recently, calorimetric method also 30
has been used in food industry and environmental monitoring (Maskow et al. 2012; Kirchner 31
et al 2012). 32
23
1
CONCLUSION AND OUTLOOK 2
3
The past decade has seen great advancements in the field of biosensor along many fronts. This 4
dynamic tool has been applied in many area of life science research, health care, 5
environmental, food and military application. Biosensor technology has received heightened 6
interest over the past decade, since it is a promising candidate for lower detection limit with 7
rapid analysis time at relatively low cost. However, the review shows that there is a lot of 8
studies have been undertaken using indirect measurement with simple clean buffer solution 9
instead direct measurement for in situ real-sample monitoring which is more vital. 10
Technological advances have provided us with the tools and materials needed to construct 11
biochip which integrated with microfluidic system, probe, sampler, detector, amplifier and 12
logic circuitry. This biochip is a promising candidate for label free, reagentless, real time 13
monitoring, miniaturization and low cost application. For medical application, this cost 14
advantage will allow the development of extremely low cost, disposable biochips that can be 15
used for in-home medical diagnostics of diseases without the need of sending samples to a 16
laboratory for analysis which time consuming. 17
18
19
ACKNOWLEDGEMENT 20
21
Author Veeradasan Perumal thankfully acknowledges collaborators at the Institute of Nano 22
Electronic Engineering (INEE) at University Malaysia Perlis (UniMAP). This work was 23
supported by INEE at (UniMAP), through the Nano Technology project. The views expressed 24
in this publication are those of the authors and do not necessarily reflect the official view of 25
the funding agencies on the subject. The appreciation also goes to all the team members in the 26
Institute of Nanoelectronic Engineering especially in the Biomedical Nano Diagnostics 27
Research Group. 28
29
24
REFERENCES 1
Abdulhalim I, Zourob M, Lakhtakia A. Overview of Optical Biosensing Techniques. In 2 Marks RS, Lowe CR, Cullen DC, Weetall HH, Karube I (ed): Handbook of Biosensors 3 and Biochips. Wiley, Weinheim 2007. 4
Acha V, Andrews T, Huang Q, Sardar DK, Hornsby PJ. Tissue-Based Biosensors. In 5 Zourob M (ed): Recognition receptors in biosensors. Springer, New York 2010, pp. 365-6 382. 7
Ahmad LM, Towe B, Wolf A, Mertens F, Lerchner J. Binding event measurement using a 8 chip calorimeter coupled to magnetic beads. Sens Actuators B Chem. 145: 239–245, 9 2010. 10
Amin R, Kulkarni A, Kim T, Park SH. DNA thin film coated optical fiber biosensor. Curr 11 Appl Phys. 12: 841–845, 2012. 12
Amine A, Mohammadi H, Bourais I, Palleschi G. Enzyme inhibition-based biosensors for 13 food safety and environmental monitoring. Biosens Bioelectron. 21: 1405–1423, 2006. 14
Arora P, Sindhu A, Dilbaghi N, Chaudhury A. Biosensors as innovative tools for the 15 detection of food borne pathogens. Biosens Bioelectron. 28: 1–12, 2011. 16
Arya SK, Singh SP, Malhotra, BD. Electrochemical Techniques in Biosensors. In Marks 17 RS, Lowe CR, Cullen DC, Weetall HH, Karube I (ed): Handbook of Biosensors and 18 Biochips. Wiley, Weinheim 2007. 19
Atias D, Liebes Y, Chalifa-Caspi V, Bremand L, Lobel L, Marks RS, Dussart, P. 20 Chemiluminescent optical fiber immunosensor for the detection of IgM antibody to 21 dengue virus in humans. Sens Actuators B Chem. 140: 206–215, 2009. 22
Bakker E, Pretsch E. Potentiometric sensors for trace-level analysis. Trends Analyt Chem. 23 24: 199–207, 2006. 24
Banerjee P, Bhunia AK. Mammalian cell-based biosensors for pathogens and toxins. Trends 25 Biotechnol. 27: 179–88, 2009. 26
Banerjee P, Bhunia AK. Cell-based biosensor for rapid screening of pathogens and toxins. 27 Biosens Bioelectron. 26: 99–106, 2010. 28
Banica FG. Chemical Sensors and Biosensors: Fundamentals and Applications. Wiley, 29 Chichester 2012. 30
Barsoukov E, Macdonald J. Impedance spectroscopy: theory, experiment, and applications. 31 Wiley, Hoboken 2005. 32
Barton AC, Collyer SD, Davis F, Garifallou GZ, Tsekenis G, Tully E, O’Kennedy R, 33 Gibson T, Millner PA, Higson SPJ. Labeless AC impedimetric antibody-based sensors 34 with pgml(-1) sensitivities for point-of-care biomedical applications. Biosens 35 Bioelectron. 24: 1090–1095, 2009. 36
Bayoudh S, Othmane A, Ponsonnet L, Ben H. Electrical detection and characterization of 37 bacterial adhesion using electrochemical impedance spectroscopy-based flow chamber. 38 Colloid Surfaces A. 318: 291–300, 2008. 39
Belkin S, Gu MB. Whole Cell Sensing Systems I: Reporter Cells and Devices. Springer, 40 Hiedelberg 2010. 41
25
Belluzo MS, Ribone MÉ, Lagier CM: Assembling Amperometric Biosensors for Clinical 1 Diagnostics. Sensors. 8: 1366–1399, 2007. 2
Berdat D, Marin A, Herrera F, Gijs MAM. DNA biosensor using fluorescence microscopy 3 and impedance spectroscopy. Sens Actuators B Chem. 118: 53–59, 2006. 4
Bhatta D, Stadden E, Hashem E, Sparrow IJG, Emmerson GD. Label-free monitoring of 5 antibody-antigen interactions using optical microchip biosensors. J Immunol Method. 6 362: 121–126, 2010b. 7
Bhatta D, Stadden E, Hashem E, Sparrow IJG, Emmerson GD. Multi-purpose optical 8 biosensors for real-time detection of bacteria, viruses and toxins. Sens Actuators B 9 Chem. 149: 233–238, 2010a. 10
Bhatta D, Villalba MM, Johnson CL, Emmerson GD, Ferris NP, King DP, Lowe CR. Rapid 11 Detection of Foot-and-Mouth Disease Virus with Optical Microchip Sensors. Procedia 12 Chem. 6: 2–10, 2012. 13
Biran I, Yu X, Walt DR. Optrode-based fiber optic biosensors (bio-optrode). In Ligler F, 14 Taitt C (ed): Optical biosensors: today and tomorrow. Elsevier, Amsterdam 2008, pp.3-15 82 16
Bohrn U, Stütz E, Fuchs K, Fleischer M, Schöning MJ, Wagner P. Monitoring of irritant 17 gas using a whole-cell-based sensor system. Sens Actuators B Chem. 175: 208–217, 18 2012. 19
Braiek M, Rokbani KB, Chrouda A, Bakhrouf BA, Maaref A, Jaffrezic-Renault N. An 20 Electrochemical Immunosensor for Detection of Staphylococcus aureus Bacteria Based 21 on Immobilization of Antibodies on Self-Assembled Monolayers-Functionalized Gold 22 Electrode. Biosensors. 2: 417–426, 2012 23
Brett AMO. DNA based biosensors. In Gorton L (ed): Comprehensive Analytical 24 Chemistry XLIV: Biosensors and Modern Biospecific Analytical Techniques. Elsevier, 25 Amsterdam 2005, pp. 179-208. 26
Brogan KL, Walt DR. Optical fiber-based sensors: application to chemical biology. Curr 27 Opin Chem Biol. 9: 494–500, 2005. 28
Brys E, Tombelli S, Minunni M, Mascini M, Turner APF. Approaches to Allergy Detection 29 Using Aptasensors. In Knopf GK, Bassi AS (ed): Smart Biosensor Technology. CRC 30 Press, Boca Raton 2007, pp. 539-566. 31
Buurma NJ, Haq I. Calorimetric and spectroscopic studies of Hoechst 33258: self-32 association and binding to non-cognate DNA. J Mol Biol. 381: 607–621, 2008. 33
Cagnin S, Caraballo M, Guiducci C, Martini P, Ross M, Santaana M, Danley D, West T, 34 Lanfranchi G. Overview of electrochemical DNA biosensors: new approaches to detect 35 the expression of life. Sensor. 9: 3122–3148, 2009. 36
Cass T. Enzymology. In Marks RS, Lowe CR, Cullen DC, Weetall HH, Karube I (ed): 37 Handbook of Biosensors and Biochips. Wiley, Weinheim 2007. 38
Caygill RL, Blair GE, Millner PA. A review on viral biosensors to detect human pathogens. 39 Anal Chim Acta. 681: 8–15, 2010. 40
Cecchini F, Manzano M, Mandabi Y, Perelman E, Marks RS. Chemiluminescent DNA 41 optical fibre sensor for Brettanomyces bruxellensis detection. J biotechnol. 157: 25–30, 42 2012 43
26
Chen H, Hu QY, Yue-Zheng, Jiang JH, Shen GL, Yu RQ. Construction of supported lipid 1 membrane modified piezoelectric biosensor for sensitive assay of cholera toxin based on 2 surface-agglutination of ganglioside-bearing liposomes. Anal. Chim. Acta. 657: 204–3 209, 2010. 4
Chen SH, Wu VCH, Chuang YC, Lin CS. Using oligonucleotide-functionalized Au 5 nanoparticles to rapidly detect foodborne pathogens on a piezoelectric biosensor. J 6 Microbiol Meth. 73: 7–17, 2008. 7
Chua A, Yean CY, Ravichandran M, Lim B, Lalitha P. A rapid DNA biosensor for the 8 molecular diagnosis of infectious disease. Biosens Bioelectron. 26: 3825–3831, 2011. 9
Conroy PJ, Hearty S, Leonard P, O’Kennedy RJ. Antibody production, design and use for 10 biosensor-based applications. Semin Cell Dev Biol. 20: 10–26, 2009. 11
Cooper MA. Label-free screening of bio-molecular interactions. Anal Bioanal Chem. 377: 12 834–842, 2003. 13
Daly CJ, McGrath JC. Fluorescent ligands, antibodies, and proteins for the study of 14 receptors. Pharmacol Therapeut. 100: 101–118, 2003. 15
Dell’Atti D, Tombelli S, Minunni M, Mascini M. Detection of clinically relevant point 16 mutations by a novel piezoelectric biosensor. Biosens Bioelectron. 21: 1876–1879, 2006. 17
Ding C, Zhong H, Zhang S. Ultrasensitive flow injection chemiluminescence detection of 18 DNA hybridization using nanoCuS tags. Biosens Bioelectron. 23: 1314–1318, 2008. 19
Diouani MF, Helali S, Hafaid I, Hassen WM, Snoussi MA, Ghram A. Miniaturized 20 biosensor for avian influenza virus detection. Mat Sci Eng C. 28: 580–583, 2008. 21
Dodeigne C, Thunus L, Lejeune R. Chemiluminescence as diagnostic tool. A review. 22 Talanta. 51: 415–439, 2000. 23
Donahue AC, Albitar M. Antibodies in Biosensing. In Zourob M (ed): Recognition 24 receptors in biosensors. Springer, New York 2010, pp. 221-248. 25
Endo T, Yamamura S, Kerman K, Tamiya E. Label-free cell-based assay using localized 26 surface plasmon resonance biosensor. Anal Chim Acta. 614: 182–189, 2008. 27
Endo T, Yamamura S, Nagatani N, Morita Y, Takamura Y, Tamiya E. Localized surface 28 plasmon resonance based optical biosensor using surface modified nanoparticle layer for 29 label-free monitoring of antigent-antibody reaction. Sci Technol Adv Mat. 6: 491-500, 30 2005. 31
Fan X, White IM, Shopova SI, Zhu H, Suter JD, Sun Y. Sensitive optical biosensors for 32 unlabeled targets: a review. Anal Chim Acta. 620: 8–26, 2008. 33
Fang C, He J, Chen Z. A disposable amperometric biosensor for determining total 34 cholesterol in whole blood. Sens Actuators B Chem. 155: 545–550, 2011. 35
Fowler JM, Wong DKY, Halsall HB, Heineman WR. Recent developments in 36 electrochemical immunoassays and immunosensors. In Zhang X, Ju H, Wang J (ed): 37 Electrochemical Sensors, Biosensors and Their Biomedical Applications. Elsevier, San 38 Diego 2008, pp. 115-140. 39
Francia GD, Ferrara VL, Manzo S, Chiavarini S. Towards a label-free optical porous silicon 40 DNA sensor. Biosens Bioelectron. 21: 661–665, 2005. 41
27
Gámiz-gracia L, García-campa AM, Huertas-pérez JF, Lara FJ. Chemiluminescence 1 detection in liquid chromatography�: Applications to clinical , pharmaceutical , 2 environmental and food analysis — A review. Anal Chim Acta. 640: 7–28, 2009. 3
Glynn B, O’Connor L. Nucleic Acid Diagnostic Biosensors. In Zourob M (ed): Recognition 4 receptors in biosensors. Springer, New York 2010, pp. 343-364. 5
Guo Z, Yang F, Zhang L, Zheng X. Electrogenerated chemiluminescence energy transfer 6 for the label-free detection of DNA. Sens Actuators B Chem. 177: 316–321, 2013. 7
Grieshaber D, MacKenzie R, Voros J, Reimhult E. Electrochemical Biosensors - Sensor 8 Principles and Architectures. Sensors. 8: 1400–1458, 2008. 9
Grubor NM, Shinar R, Jankowiak R, Porter MD, Small GJ. Novel biosensor chip for 10 simultaneous detection of DNA-carcinogen adducts with low-temperature fluorescence. 11 Biosens Bioelectron. 19: 547–556, 2004. 12
He Y, Zhang S, Zhang X, Baloda M, Gurung AS, Xu H, Zhang X, Liu G. Ultrasensitive 13 nucleic acid biosensor based on enzyme-gold nanoparticle dual label and lateral flow 14 strip biosensor. Biosens Bioelectron. 26: 2018–2024, 2011. 15
Hnaiein M, Hassen WM, Abdelghani A, Fournier-Wirth C, Coste J, Bessueille F, Leonard 16 D, Jaffrezic-Renault N. A conductometric immunosensor based on functionalized 17 magnetite nanoparticles for E. coli detection. Electrochem Commun. 10: 1152–1154, 18 2008. 19
Holford TRJ, Davis F, Higson SPJ. Recent trends in antibody based sensors. Biosens 20 Bioelectron. 34: 12–24, 2012. 21
Homola J, Yee SS, Myszka D. Surface plasmon resonance biosensors. In Ligler F, Taitt C 22 (ed): Optical biosensors: today and tomorrow. Elsevier, Amsterdam 2008, pp. 185-242. 23
Hong B, Kang KA. Biocompatible, nanogold-particle fluorescence enhancer for 24 fluorophore mediated, optical immunosensor. Biosens Bioelectron. 21: 1333–1338, 25 2006. 26
Hong J, Lan K, Jang L. Chemical Electrical characteristics analysis of various cancer cells 27 using a microfluidic device based on single-cell impedance measurement. Sens 28 Actuators B Chem. 173: 927–934, 2012. 29
Huang JC, Chang YF, Chen KH, Su LC, Lee CW, Chen CC, Chen YMA, Chou C. 30 Detection of severe acute respiratory syndrome (SARS) coronavirus nucleocapsid 31 protein in human serum using a localized surface plasmon coupled fluorescence fiber-32 optic biosensor. Biosens Bioelectron. 25: 320–325, 2009. 33
Ivnitski D, Wolf T, Solomon B. An amperometric biosensor for real-time analysis of 34 molecular recognition. Bioelectroch Bioener. 45: 27–32, 1998. 35
Ivnitski D, Sitdikov R, Ivnitski N. Non-invasive electrochemical hand-held biosensor as 36 diagnostic indicator of dental diseases. Electrochem. Commun. 5: 225–229, 2003. 37
Jang HS, Park KN, Kang CD, Kim JP, Sim SJ, Lee KS. Optical fiber SPR biosensor with 38 sandwich assay for the detection of prostate specific antigen. Opt Commun. 282: 2827–39 2830, 2009. 40
Jelen F, Yosypchuk B, Kourilová A, Novotný L, Palecek E. Label-free determination of 41 picogram quantities of DNA by stripping voltammetry with solid copper amalgam or 42 mercury electrodes in the presence of copper. Anal chem. 74: 4788-4793, 2002. 43
28
Jia Y, Yin XB, Zhang J, Zhou S, Song M, Xing KL. Graphene oxide modified light 1 addressable potentiometric sensor and its application for ssDNA monitoring. Analyst. 2 137: 5866–5873, 2012. 3
Johansson A, Kromer K, Sroka R, Stepp H. Clinical optical diagnostics – Status and 4 perspectives. Med Laser Appl. 23: 155–174, 2008. 5
Ju H, Kandimalla VB. Biosensors for pesticides. In Zhang X, Ju H, Wang J (ed): 6 Electrochemical Sensors, Biosensors and Their Biomedical Applications. Elsevier, San 7 Diego 2008, pp. 32-50. 8
Kahn K, Plaxco KW. Principles of Biomolecular Recognition. In Zourob M (ed): 9 Recognition receptors in biosensors. Springer, New York 2010, pp. 3-46. 10
Kannan B, Williams DE, Laslau C, Travas-Sejdic J. A highly sensitive, label-free gene 11 sensor based on a single conducting polymer nanowire. Biosens Bioelectron. 35: 258–12 264, 2012. 13
Katzir S. The Beginnings of Piezoelectricity: A Study in Mundane Physics. Springer, 14 Dordrecht 2006. 15
Kim JM, Chang SM, Muramatsu H, Isao K. The principles and applications of nano-16 diagnosis system for a nano-biosensor. Korean J Chem Eng. 28: 987–1008, 2011. 17
Kirchner P, Oberländer J, Friedrich P, Berger J, Rysstad G, Keusgen M, Schöning MJ. 18 Realisation of a calorimetric gas sensor on polyimide foil for applications in aseptic food 19 industry. Sens. Actuators B Chem. 170: 60–66, 2012. 20
Koncki R. Recent developments in potentiometric biosensors for biomedical analysis. Anal 21 Chim Acta. 599: 7–15, 2007. 22
Kricka L. Clinical applications of chemiluminescence. Anal Chim Acta. 500: 279–286, 23 2003. 24
Kukol A, Li P, Estrela P, Ko-ferrigno P, Migliorato P. Label-free electrical detection of 25 DNA hybridization for the example of influenza virus gene sequences. Anal Biochem. 26 374: 143–153, 2008. 27
Lara FJ, García-campa AM, Aaron J. Analytical applications of photoinduced 28 chemiluminescence in flow systems — A review. Anal Chim Acta. 679: 17–30, 2010. 29
Lazerges M, Perrot H, Rabehagasoa N, Compère C. Thiol- and Biotin-Labeled Probes for 30 Oligonucleotide Quartz Crystal Microbalance Biosensors of Microalga Alexandrium 31 Minutum. Biosensors. 2: 245–254, 2012. 32
Leca-Bouvier BD, Blum LJ. Enzyme for Biosensing Applications. In Zourob M (ed): 33 Recognition receptors in biosensors. Springer, New York 2010, pp. 177-220. 34
Lee B, Roh S, Park J. Current status of micro- and nano-structured optical fiber sensors. Opt 35 Fiber Technol. 15: 209–221, 2009. 36
Lee I, Luo X, Huang J, Cui XT, Yun M. Detection of Cardiac Biomarkers Using Single 37 Polyaniline Nanowire-Based Conductometric Biosensors. Biosensors. 2: 205–220, 2012. 38
Leung A, Shankar PM, Mutharasan R. A review of fiber-optic biosensors. Sens Actuators B 39 Chem. 125: 688–703, 2007. 40
Leung A, Shankar PM, Mutharasan R. Label-free detection of DNA hybridization using 41 gold-coated tapered fiber optic biosensors (TFOBS) in a flow cell at 1310nm and 42 1550nm. Sens Actuators B Chem. 131: 640–645, 2008. 43
29
Li D. Chemiluminescence. In Li D (ed): Encyclopedia of Microfluidics and Nanofluidics. 1 Springer, New York 2008, pp. 260. 2
Lin L, Liu Q, Wang L, Liu A, Weng S, Lei Y, Chen W, Lin X, Chen Y. Enzyme-amplified 3 electrochemical biosensor for detection of PML-RARα fusion gene based on hairpin 4 LNA probe. Biosens. Bioelectrons. 28: 277–83, 2011. 5
Liu A, Wang K, Weng S, Lei Y, Lin L, Chen W, Lin X, Chen. (2012). Development of 6 electrochemical DNA biosensors. Trends. Anal. Chem. 37: 101–111, 2012. 7
Liu CC. Electrochemical Based Biosensors. Biosensors. 2: 269–272, 2012. 8
Liu M, Lin Z, Lin J. A review on applications of chemiluminescence detection in food 9 analysis. Anal Chim Acta. 670: 1–10, 2010 10
Liu Q, Hu N, Zhang F, Wang H, Ye W, Wang P. Neurosecretory cell-based biosensor: 11 monitoring secretion of adrenal chromaffin cells by local extracellular acidification using 12 light-addressable potentiometric sensor. Biosens Bioelectron. 35: 421–424, 2012. 13
Luong JHT, Male KB, Glennon JD. Biosensor technology: Technology push versus market 14 pull. Biotechnol. Adv. 26: 492–500, 2008. 15
Makarychev-Mikhailov, S., Shvarev, A., & Bakker, E.: New trends in ion-selective 16 electrodes. In Zhang X, Ju H, Wang J (ed): Electrochemical Sensors, Biosensors and 17 Their Biomedical Applications. Elsevier, San Diego 2008, pp. 71-109. 18
Mannelli I, Courtois V, Lecaruyer P, Roger G, Millot MC, Goossens M, Canva M. Surface 19 plasmon resonance imaging (SPRI) system and real-time monitoring of DNA biochip for 20 human genetic mutation diagnosis of DNA amplified samples. Sens Actuators B Chem. 21 119: 583–591, 2006. 22
Maskow T, Wolf K, Kunze W, Enders S, Harms H. Rapid analysis of bacterial 23 contamination of tap water using isothermal calorimetry. Thermochim Acta. 543: 273–24 280, 2012. 25
Mathews ST, Plaisance EP, Kim T. Imaging Systems for Westerns: Chemiluminescence vs. 26 Infrared Detection. In Kurien BT, Scofield RH (ed): Methods in Molecular Biology, 27 Protein Blotting and Detection. Humana Press, Totowa 2009, pp. 499-513 28
McGuinness RP, Verdonk E. Electrical impedance technology applied to cell-based assays. 29 In Cooper MA (ed): Label Free Biosensors: Techniques and Applications. Cambridge 30 University Press, New York 2009, pp. 251-278 31
Md Muslim NZ, Ahmad M, Heng LY, Saad B. Optical biosensor test strip for the screening 32 and direct determination of l-glutamate in food samples. Sens Actuators B Chem. 161: 33 493–497, 2012. 34
Melamed S, Elad T, Belkin S. Microbial sensor cell arrays. Curr Opin Biotechnol. 23: 2–8, 35 2012. 36
Merwe PAVD. Surface plasmon resonance. In Harding S, Chowdhry PZ (ed): Protein-37 Ligand interactions: hydrodynamics and calorimetry. Oxford University Press, Bath 38 2001, pp. 137-170. 39
Mitchell J, Wu Y. Surface Plasmon Resonance Biosensors for Highly Sensitive Detection 40 of Small Biomolecules. In Serra PA (ed): Biosensors. InTech, Rijeka 2010, pp. 151-168. 41
Moghissi K, Stringer MR, Dixon K. Fluorescence photodiagnosis in clinical practice. 42 Photodiagn Photodyn. 5: 235–237, 2008. 43
30
Mol NJD, Fischer MJE. Surface Plasmon Resonance: A General Introduction. In Mol NJD, 1 Fischer MJE (ed): Surface Plasmon Resonance Methods and Protocols. Springer, New 2 York 2010, pp. 1-14. 3
Mono R, Stred M, Sturd E. Application of Electrochemical Biosensors in Clinical 4 Diagnosis. J Clin Lab Anal. 34: 22–34, 2012. 5
Morrison DWG, Dokmeci MRT, Demirci U, Hossein AK. Clinical Applications of Micro- 6 and Nanoscale Biosensors. In Gonsalves K, Halberstadt C, Laurencin CT, Nair (ed): 7 Biomedical Nanostructures. Wiley, Hoboken 2008, pp. 439-460. 8
Nakamura H, Mogi Y, Akimoto T, Naemura K, Kato T, Yano K, Karube I. An enzyme-9 chromogenic surface plasmon resonance biosensor probe for hydrogen peroxide 10 determination using a modified Trinder’s reagent. Biosens Bioelectron. 24: 455–460, 11 2008. 12
Nguyen B, Tanious FA, Wilson WD. Biosensor-surface plasmon resonance: quantitative 13 analysis of small molecule-nucleic acid interactions. Methods. 42: 150–161, 2007. 14
Nicu L, Guirardel M, Chambosse FDR, Rougerie P, Hinh S, Trevisiol E, Francois JM, 15 Majoral JP, Caminade AM, Cattan E, Bergaud C. Resonating piezoelectric membranes 16 for microelectromechanically based bioassay: detection of streptavidin gold 17 nanoparticles interaction with biotinylated DNA. Sens Actuators B Chem. 110: 125-136, 18 2005. 19
Nirschl M, Blüher A, Erler C, Katzschner B, Vikholm-Lundin I, Auer S, Vörös J, Pompe 20 W, Schreiter M, Mertig M. Film bulk acoustic resonators for DNA and protein detection 21 and investigation of in vitro bacterial S-layer formation. Sens Actuators A Phys. 156: 22 180–184, 2009. 23
Nomngongo PN, Catherine Ngila J, Msagati TAM, Gumbi BP, Iwuoha EI. Determination 24 of selected persistent organic pollutants in wastewater from landfill leachates, using an 25 amperometric biosensor. Phys Chem Earth Pt A/B/C. 50-52: 252–261, 2012. 26
Ohno R, Ohnuki H, Wang H, Yokoyama T, Endo H, Tsuya , Izumi M. Electrochemical 27 impedance spectroscopy biosensor with interdigitated electrode for detection of human 28 immunoglobulin A. Biosens Bioelectron. 40: 422–426, 2012. 29
Onur Z, Kemal M. A novel , ultra sensible biosensor built by layer-by-layer covalent 30 attachment of a receptor for diagnosis of tumor growth. Anal Chim Acta. 706: 343–348, 31 2011. 32
Orazio PD. Biosensors in clinical chemistry — 2011 update. Clin Chim Acta. 412: 1749–33 1761, 2011. 34
Parkinson G, Pejcic B. Using Biosensors to Detect Emerging Infectious Diseases, prepared 35 for The Australian Biosecurity Cooperative Research Centre: pp.1-80, 2005. Retrieved 36 from: http://www.abcrc.org.au/pages/project.aspx?projectid=75. 37
Park TJ, Hyun MS, Lee HJ, Lee SY, Ko S. A self-assembled fusion protein-based surface 38 plasmon resonance biosensor for rapid diagnosis of severe acute respiratory syndrome. 39 Talanta. 79: 295–301, 2009. 40
Paul P, Hossain M, Yadav RC, Kumar GS. Biophysical studies on the base specificity and 41 energetics of the DNA interaction of photoactive dye thionine: spectroscopic and 42 calorimetric approach. Biophy Chem. 148: 93–103, 2010. 43
31
Pejcic B, Marco RD. Impedance spectroscopy: Over 35 years of electrochemical sensor 1 optimization. Electrochim Acta. 51: 6217–6229, 2006. 2
Pejcic B, Marco RD, Parkinson G. The role of biosensors in the detection of emerging 3 infectious diseases. Analyst. 131: 1079–1090, 2006. 4
Pohanka M. Monoclonal and polyclonal antibodies production-preparation of potent 5 biorecognition element. J Appl Biomed. 7: 115–121, 2009. 6
Pohanka M, Skládal P. Electrochemical biosensors – principles and applications. J Appl 7 Biomed. 6: 57–64, 2008. 8
Ramanathan A, Pape L, Schwartz DC. High-density polymerase-mediated incorporation of 9 fluorochrome-labeled nucleotides. Anal biochem. 337: 1–11, 2005. 10
Ramírez NB, Salgado AM, Valdman B. The Evolution and Development of Immunosensors 11 for Health and Environmental Monitoring: Problems and Persperctives. Brazilian J 12 Chem Eng. 26: 227–249, 2009. 13
Rogers KR, Mascini M. Biosensors for field analytical monitoring. Field Anal Chem 14 Technol. 2: 317-331, 1999. 15
Lowe RS. Overview of Biosensor and Bioarray Technologies. In Marks RS, Lowe CR, 16 Cullen DC, Weetall HH, Karube I (ed): Handbook of Biosensors and Biochips. Wiley, 17 Weinheim 2007. 18
Saxena V, Malhotra BD. Electrochemical Biosensors. In Malhotra BD, Turner APF (ed): 19 Advances In Biosensors: Perspectives In Biosensor. JAI Press, Stamford 2003, pp. 63-20 100. 21
Schelfer F, Schubert F. Biosensors�: Techniques and Instrumentation in Analytical 22 Chemistry. Elsevier, Amsterdam 1992. 23
Schneider HJ, Lim S, Strongin RM. Biomimetic Synthetic Receptors as Molecular 24 Recognition Elements. In Zourob M (ed): Recognition receptors in biosensors. Springer, 25 New York 2010, pp. 777-818. 26
Schultz E, Galland R, Du Bouëtiez D, Flahaut T, Planat-Chrétien A, Lesbre F, Hoang A, 27 Volland H, Perraut F. A novel fluorescence-based array biosensor: principle and 28 application to DNA hybridization assays. Biosens Bioelectron. 23: 987–994, 2008. 29
Serra B, Gamella M, Reviejo AJ, Pingarrón JM. Lectin-modified piezoelectric biosensors 30 for bacteria recognition and quantification. Anal Bioanal Chem. 391: 1853–1860, 2008. 31
Shankaran D, Gobi K, Miura N. Recent advancements in surface plasmon resonance 32 immunosensors for detection of small molecules of biomedical, food and environmental 33 interest. Sens Actuators B Chem. 121: 158–177, 2007. 34
Shantilatha R, Varma S, Mitra CK. Designing a simple biosensor. In Biosensors: 35 Perspectives. In Malhotra BD, Turner APF (ed): Advances In Biosensors: Perspectives 36 In Biosensor. JAI Press, Stamford 2003, pp. 1-36. 37
Shinde SB, Fernandes CB, Patravale VB. Recent trends in in-vitro nanodiagnostics for 38 detection of pathogens. J Control Release. 159: 164–180, 2012. 39
Skládal P, Riccardi CDS, Yamanaka H, Da Costa PI. Piezoelectric biosensors for real-time 40 monitoring of hybridization and detection of hepatitis C virus. J Virol Methods. 117: 41 145–151, 2004. 42
32
Skottrup PD, Nicolaisen M, Justesen AF. Towards on-site pathogen detection using 1 antibody-based sensors. Biosens Bioelectron. 24: 339–348, 2008. 2
Soldatkin OO, Kucherenko IS, Pyeshkova VM, Kukla AL, Jaffrezic-Renault N, El’skaya 3 AV, Dzyadevych SV, Soldatkin AP. Novel conductometric biosensor based on three-4 enzyme system for selective determination of heavy metal ions. Bioelectrochemistry. 83: 5 25–30, 2012. 6
Springer T, Piliarik M, Homola J. Surface plasmon resonance sensor with dispersionless 7 microfluidics for direct detection of nucleic acids at the low femtomole level. Sens 8 Actuators B Chem. 145: 588–591, 2010. 9
Steinem C, Janshoff A. Springer Series on Chemical Sensors and Biosensors: Piezoelectric 10 Sensors. Springer, Heidelberg 2007. 11
Strehlitz B, Nikolaus N, Stoltenburg R. Protein Detection with Aptamer Biosensors. 12 Sensors. 8: 4296–4307, 2008. 13
Struss AK, Pasini P, Daunert S. Biosensing Systems Based on Genetically Engineered 14 Whole Cells. In Zourob M (ed): Recognition receptors in biosensors. Springer, New 15 York 2010, pp. 565-598. 16
Tang J, Huang J, Su B, Chen H, Tang D. Sandwich-type conductometric immunoassay of 17 alpha-fetoprotein in human serum using carbon nanoparticles as labels. Biochem Eng J. 18 53: 223–228, 2011. 19
Teles F, Fonseca L. Trends in DNA biosensors. Talanta. 77: 606–623, 2008. 20
Thevenot DR, Toth K, Durst RA, Wilson GS. Electrochemical biosensors�: recommended 21 definitions and classification. Biosens Bioelectron. 16: 121–131, 2001. 22
Thevenot DR, Toth K, Durst, RA, Wilson GS. Electrochemical Biosensors�: 23 Recommended defnitions and classification. Pure Appl Chem. 71: 2333–2348, 1999. 24
Thi B, Nguyen T, En A, Peh K, Yue C, Chee L, Fink K, Chow VTK, Ng MML, Toh CS. 25 Bioelectrochemistry Electrochemical impedance spectroscopy characterization of 26 nanoporous alumina dengue virus biosensor. Bioelectrochem. 88: 15–21, 2012. 27
Thompson RB. Fluorescence lifetime biosensing: Entering the mainstream. In Ligler F, 28 Taitt C (ed): Optical biosensors: today and tomorrow. Elsevier, Amsterdam 2008, pp. 29 287-316. 30
Thuy NT, Tam PD, Tuan MA, Le AT, Tam LT, Thu VV, Hieu NV, Chien ND. Detection of 31 pathogenic microorganisms using biosensor based on multi-walled carbon nanotubes 32 dispersed in DNA solution. Curr Appl Phys. 12: 1553–1560, 2012. 33
Tichý J, Erhart J, Kittinger E, Prívratská J. Fundamentals of Piezoelectric Sensorics: 34 Mechanical, Dielectric, and Thermodynamical Properties of Piezoelectric Materials. 35 Springer, Heidelberg 2010. 36
Torres-Chavolla E, Alocilja EC. Aptasensors for detection of microbial and viral pathogens. 37 Biosens Bioelectron. 24: 3175–3182, 2009. 38
Tothill IE. Biosensors for cancer markers diagnosis. Semin Cell Dev Biol. 20: 55–62, 2009. 39
Tsoka S, Gill A, Brookman JL, Hoare M. Rapid monitoring of virus-like particles using an 40 optical biosensor: a feasibility study. J Biotechnol. 63: 147–153, 1998. 41
Ushaa SM, Nagar NI, Rao GM. Design and analysis of nanowire sensor array for prostate 42 cancer detection. Int J Nano Biomater. 3: 239–255, 2011. 43
33
Vallet-Regí M, Arcos DA. Biomimetic Nanoceramics in Clinical Use: From Materials to 1 Applications. RSC publishing, Cambridge 2008. 2
Vaughan RD, Guilbault GG. Piezoelectric Immunosensors. Piezoelectric Sensor. 5: 237–3 280, 2007. 4
Védrine C, Leclerc JC, Durrieu C, Tran-Minh C. Optical whole-cell biosensor using 5 Chlorella vulgaris designed for monitoring herbicides. Biosens Bioelectron. 18: 457–6 463, 2003. 7
Veiseh M, Veiseh O, Martin MC, Bertozzi C, Zhang M. Single-cell-based sensors and 8 synchrotron FTIR spectroscopy: a hybrid system towards bacterial detection. Biosens 9 Bioelectron. 23: 253–260, 2007. 10
Wang J. Electrochemical biosensors: towards point-of-care cancer diagnostics. Biosens 11 Bioelectron. 21: 1887–1892, 2006. 12
Wang J. Electrochemical transduction. In Taylor RF, Schultz JS (ed): Handbook of 13 Chemical and Biological Sensors. IOP Publishing Ltd, Bristol 1996, pp. 119-133. 14
Wang P, Liu Q. Cell-based biosensors: Principles and Applications. Artech House, 15 Norwood 2010. 16
Wang TH, Hui GH, Deng SP. A novel sweet taste cell-based sensor. Biosens Bioelectron. 17 26: 929–934, 2010. 18
Wang YX, Ye ZZ, Si CY, Ying YB. Application of Aptamer Based Biosensors for 19 Detection of Pathogenic Microorganisms. Chinese J Anal Chem. 40: 634–642, 2012. 20
Wang Y, Zhang Z, Jain V, Yi J, Mueller S, Sokolov J, Liu Z, Levon K, Rigas B, 21 Rafailovich MH. Potentiometric sensors based on surface molecular imprinting: 22 Detection of cancer biomarkers and viruses. Sens Actuators B Chem. 146: 381–387, 23 2010. 24
Wang Y, Xu H, Zhang J, Li G. Electrochemical Sensors for Clinic Analysis. Sensors. 8: 25 2043–2081, 2008. 26
Watterson J, Piunno PAE, Krull UJ. Practical physical aspects of interfacial nucleic acid 27 oligomer hybridisation for biosensor design. Anal Chim Acta. 469: 115–127, 2002. 28
Weng CH, Huang CJ., Lee GB. Screening of Aptamers on Microfluidic Systems for 29 Clinical Applications. Sensors. 12: 9514–9529, 2012. 30
Wood, P. Understanding Immunology. Pearson Education Limited, Dorchester 2006. 31
Xi H, Kumar S, Dosen-Micovic L, Arya DP. Calorimetric and spectroscopic studies of 32 aminoglycoside binding to AT-rich DNA triple helices. Biochimie, 92: 514–529, 2010. 33
Xie B, Ramanathan K, Danielsson B. Principles of enzyme thermistor systems: applications 34 to biomedical and other measurements. Adv Biochem Eng/Biotechnol. 64: 1–33, 1999. 35
Yakovleva M, Bhand S, Danielsson B. The enzyme thermistor–a realistic biosensor concept 36 A critical review. Anal Chim Acta. In Press, 2012. doi:10.1016/j.aca.2012.12.004 37
Yang L, Bashir R. Electrical / electrochemical impedance for rapid detection of foodborne 38 pathogenic bacteria. Biotechnol Adv. 26: 135–150, 2008. 39
Yao C, Zhu T, Tang J, Wu R, Chen Q, Chen M, Zhang B, Huang J, Fu W. Hybridization 40 assay of hepatitis B virus by QCM peptide nucleic acid biosensor. Biosens. Bioelectrons. 41 23: 879–885, 2008. 42
34
Yeh CH, Chang YH, Lin HP, Chang TC, Lin YC. A newly developed optical biochip for 1 bacteria detection based on DNA hybridization. Sens Actuators B Chem. 161: 1168–2 1175, 2012. 3
Zapp E, Brondani D, Vieira IC, Scheeren CW, Dupont J, Barbosa AMJ, Ferreira VS. 4 Biomonitoring of methomyl pesticide by laccase inhibition on sensor containing 5 platinum nanoparticles in ionic liquid phase supported in montmorillonite. Sens 6 Actuators B Chem. 155: 331–339, 2011. 7
Zhang L, Lou J, Tong L. Micro/nanofiber optical sensors. Photonic Sensor. 1: 31–42, 2011. 8
Zhang Y, Tadigadapa S. Calorimetric biosensors with integrated microfluidic channels. 9 Biosens Bioelectron. 19: 1733–1743, 2004. 10
Zhang, Z., Zhang, S., & Zhang, X.: Recent developments and applications of 11 chemiluminescence sensors. Anal Chim Acta. 541: 37–46, 2005. 12
13