Agreement Between Corneal Topography with the Dual-Scheimpflug Placido Corneal Topographer, Scheimpflug Corneal Topographer, and Scanning-Slit Topography in Keratoconus: Implications for Assessing Progression Pravin K Vaddavalli MD, Shyamsunder T, Parthasarthy K, Preetam Kumar, Shrikant D, Preeji MS, Reekham S, Faiza S Cornea Service, Contact lens service LV Prasad Eye Institute Hyderabad, India None of the authors has any financial interests or conflict of interest to disclose

Agreement Between Corneal Topography with the Dual- Scheimpflug Placido Corneal Topographer, Scheimpflug Corneal Topographer, and Scanning-Slit Topography

Embed Size (px)

Citation preview

Agreement Between Corneal Topography with the Dual-Scheimpflug Placido Corneal Topographer, Scheimpflug Corneal Topographer, and Scanning-Slit Topography in

Keratoconus: Implications for Assessing Progression

Pravin K Vaddavalli MD, Shyamsunder T, Parthasarthy K, Preetam Kumar, Shrikant D, Preeji MS, Reekham S, Faiza S

Cornea Service, Contact lens service

LV Prasad Eye Institute

Hyderabad, India

None of the authors has any financial interests or conflict of interest to disclose

Purpose

• To assess the repeatability of the Orbscan, Pentacam and Galilei topography systems in normal and keratoconus eyes

• To assess the agreement between these three topography machines in assessing normal and keratoconus eyes

PENTACAM GALILEIORBSCAN

Methods• 96 right eyes on normal patients• 45 clinically diagnosed keratoconus• 2 scans each on Orbscan, Pentacam and Galilei• Intersession variability, instrument agreement

ORBSCAN PENTACAM GALILEI

NORMALS NORMALS NORMALS

KERATOCONUS KERATOCONUS KERATOCONUS

INTE

RSES

SIO

N R

EPEA

TABI

LITY

(M

AX/S

TEEP

KER

ATO

MET

RY)

NORMALS NORMALS NORMALS

KERATOCONUS

PENTACAM vs GALILEI

KERATOCONUS

ORBSCAN vs PENTACAM ORBSCAN vs GALILEI

KERATOCONUS

INTE

R IN

STRU

MEN

T RE

PEAT

ABIL

ITY

(MAX

/STE

EP K

ERAT

OM

ETRY

)

SUMMARY

• Normals – Pentacam had least intersession variability

• Keratoconus eyes – all topographers showed greater

intersession variation compared to normals, Galilei was best

• Between instruments, Pentacam and Galilei had the highest

agreement in both normals and keratoconus eyes