PD-0100-0073 Revision A Air A5, Air A20 LABORATORY ZERO-AIR GENERATORS Installation, Operation & Maintenance Manual

Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

PD-0100-0073 Revision A

Air A5, Air A20 LABORATORY ZERO-AIR

GENERATORS

Installation, Operation & Maintenance Manual

Page 2: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 2 of 22

Zero Air 5 and Zero Air20 Laboratory Zero-Air Generators Installation,

Operation & Maintenance Manual Covering the following laboratory zero-air generator models:

Model Proton OnSite Part Number

Zero Air 5, 120 VAC 50/60 Hz 54-0113-0009

Zero Air 5, 230 VAC 50/60 Hz 54-0113-0010

Zero Air 20, 120 VAC 50/60 Hz 54-0113--0011

Zero Air 20, 230 VAC 50/60 Hz 54-0113-0012

Proton OnSite 10 Technology Drive

Wallingford, CT 06492 203-949-8697 203-678-2000

[email protected] www.protononsite.com

© 2014 Proton Energy Systems, Inc. d/b/a Proton OnSite

All rights not expressly granted herein are reserved.

Information in this document is provided in connection with Proton Energy Systems Inc. d/b/a Proton OnSite (“Proton”) products and services. No license, express or implied, by estoppel or otherwise, to any intellectual property rights is granted by this document. Except as provided in Proton Terms and Conditions of Sale for such products and services, Proton assumes no liability whatsoever, and Proton disclaims any express or implied warranty, whether written or oral, relating to sale and/or use of Proton products or services including liability or warranties relating to fitness for a particular purpose, merchantability, or infringement of any patent, copyright or other intellectual property right. Proton products are not intended for use in medical, life saving, or life sustaining applications.

Proton may make changes to specifications and product descriptions at any time, without notice.

This document as well as any software, processes or procedures described in it is furnished under license and may only be used or copied in accordance with the terms of the license. The information in this document is furnished for informational use only, is subject to change without notice, and should not be construed as a commitment by Proton. Proton assumes no responsibility or liability for any errors or inaccuracies that may appear in this document or any software that may be provided in association with this document.

Except as permitted by such license, no part of this document may be reproduced, stored in a retrieval system, or transmitted in any form or by any means without the express written consent of Proton.

® Proton, Proton OnSite, Proton Energy Systems, the Proton design, StableFlow, StableFlow Hydrogen Control System and design, HOGEN, and FuelGen are trademarks of Proton Energy Systems, Inc.

Any other brands and/or names used herein are the property of their respective owners.

Page 3: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22

Disclaimer Proton OnSite has written this manual to be an easy to use guide for the Zero Air 5 and Zero Air 20 Zero-Air Generators. All statements, technical information, and recommendations in this manual and in any guides or related documents are believed reliable, but the accuracy and completeness thereof are not guaranteed or warranted. They are not intended to be, nor should they be understood to be, representations or warranties concerning the products described.

This zero-air gas generator has been sold subject to the limited warranties set forth in the warranty statement. Further, Proton reserves the right to make changes in the specifications of the products described in this manual at any time without notice and without obligation to notify any person of such changes.

This laboratory zero-air generator is designed for operation only by trained personnel familiar with the use of similar equipment and with safety requirements for the use of zero-air and other industrial gases. Before operating this laboratory zero-air generator, make sure to read and understand this information provided herein. If you have questions, please contact Proton OnSite Systems or your generator supplier.

Page 4: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 4 of 22

LIMITED WARRANTY

ZERO-AIR GAS GENERATOR SYSTEMS

LIMITED WARRANTY: PROTON ENERGY SYSTEMS INC. (“PROTON”) warrants that the ITEMS LISTED BELOW shall be free from defects in material and workmanship for the stated period of time commencing from date of shipment or as stated below.

ZERO-AIR GAS GENERATOR SYSTEMS: One (1) year from shipment

Repair or replacement parts for Zero-Gas Generator System: Ninety (90) days from shipment

EXCLUDED FROM THIS LIMITED WARRANTY: The following shall be excluded from the Limited Warranty:

• Parts and items considered consumable in normal operations, including those parts and items supplied with the Zero-Air Gas Generator System (“System”) for maintenance.

• Any System and its parts that are not installed, operated, and maintained in accordance with the unit’s manual supplied with the System.

• Damages due to accident, abuse, acts of God, acts of terrorism, misuse or negligence, or which result, in whole or in part, from improper or unauthorized use or repair of the System, or use of the System in a manner for which it was not designed, or by causes external to the System such as, but not limited to, power or air conditioning failure or voltage irregularities.

REMEDY: BUYER’S sole and exclusive remedy in the event of defect, and the liability of PROTON hereunder is limited to the adjustment, repair, or replacement of the defective item or part with a similar item or part free of defect.

Such adjustments, repairs, or replacements will be made at PROTON’S Wallingford, Connecticut, plant or, for Systems only, at the site of the System, if BUYER so elects. All costs for shipping equipment or parts shall be on the account of the BUYER whether to or from the point of manufacture. Labor costs associated with travel, expenses, and subsistence costs for field services shall be on the account of the BUYER.

VOIDING OF THE LIMITED WARRANTY: This Limited Warranty is immediately void upon:

• THE OPENING OR DISASSEMBLY OF THE SYSTEM CABINET (OR ANY PART THEREIN), OR

• THE SALE, ASSIGNMENT OR ANY OTHER TRANSFER OF TITLE BY BUYER OF THE ITEMS OR PARTS OTHERWISE COVERED UNDER THIS LIMITED WARRANTY

WAIVER OF ALL OTHER WARRANTIES: THE LIMITED WARRANTY PROVIDED HEREUNDER AND THE RIGHTS AND REMEDIES OF THE BUYER HEREUNDER ARE IN LIEU OF, AND BUYER EXPRESSLY WAIVES, ALL OTHER WARRANTIES, GUARANTEES, OBLIGATIONS, LIABILITIES, OR REMEDIES, EXPRESSED OR IMPLIED, ARISING BY LAW OR OTHERWISE, INCLUDING WITHOUT LIMITATION IMPLIED WARRANTIES OF MERCHANTABILITY AND NON-INFRINGEMENT, IMPLIED WARRANTIES ARISING FORM THE COURSE OF DEALING OR USAGE OF TRADE AND IMPLIED WARRANTIES OF SUITABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

LIMITATION OF LIABILITY: THE REMEDIES PROVIDED IN THIS LIMITED WARRANTY ARE EXCLUSIVE AND PROTON SHALL IN NO WAY BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND WHATSOEVER INCLUDING WITHOUT LIMITATION LOSS OF USE, REVENUE OR PROFIT.

Page 5: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 5 of 22

Air A5 and Air A20 Laboratory Zero-Air Generators

INSTALLATION, OPERATION & MAINTENANCE MANUAL

Prepared by

PROTON ONSITE®

CONTENTS

1 Introduction ..................................................................................................................8

General Information ................................................................................................................... 9

Product Specifications............................................................................................................... 10

2 Safety .......................................................................................................................... 12

Using Zero-Air Generators ....................................................................................................... 12

Safety Devices ............................................................................................................................ 12

Technical Service Contact Information .................................................................................... 12

3 Placement ................................................................................................................... 13

Providing Suitable Location ..................................................................................................... 13

Unpacking the Generator ......................................................................................................... 13

Positioning the Generator ......................................................................................................... 14

4 Installation Procedure ................................................................................................ 15

4.1 Mechanical Setup ................................................................................................................ 15

4.2 Electrical Setup ................................................................................................................... 16

4.3 Before Using the Generator ................................................................................................ 16

4.4 Shutdown ............................................................................................................................. 17

5 Maintenance ............................................................................................................... 18

Cleaning the Generator............................................................................................................. 18

6 Troubleshooting .......................................................................................................... 19

7 APPENDIX A—FORMS ............................................................................................ 20

Customer Registration Form .................................................................................................... 20

Page 6: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 6 of 22

TABLE OF FIGURES FIGURE 1 INSTALLATION CLEARANCES .............................................................................................................. 14 FIGURE 2: ZERO AIR 5/ZERO AIR 20 INLET/OUTLET, RIGHT SIDE ........................................................................ 15 FIGURE 3: POWER SWITCH, LEFT SIDE .............................................................................................................. 17

LIST OF TABLES TABLE 1: TECHNICAL SPECIFICATIONS FOR THE ZERO AIR 5 AND ZERO AIR 20 GENERATORS.................................. 10 TABLE 2: GENERATOR GAS INLET/OUTLET FITTINGS .......................................................................................... 11

Page 7: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 7 of 22

This Installation, Operation & Maintenance Manual of the Air Generator uses the following safety symbols and conventions to alert you to information intended to help you operate your laboratory air generators correctly and safely.

Manual Conventions

—Indicates a potentially hazardous situation, which, if not avoided, could result in death or serious injury. The reader is in a situation which could cause bodily injury. Do not proceed beyond a Warning symbol until the indicated conditions are fully understood and met.

—Indicates a potentially hazardous situation, which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. This could result in equipment damage or loss of data. Do not proceed beyond a Caution symbol until the indicated conditions are fully understood and met.

—Notes contain helpful suggestions or references.

Page 8: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 8 of 22

1 Introduction

Congratulations on your purchase of a Air A5 or Air A20 zero-air generator, the most capable family of laboratory zero-air generator products available today. Zero Air purifiers are intended to reduce the hydrocarbon content of compressed air used for laboratory instrumentation. Our zero-air generators provide high purity-grade, zero-air on demand and provide the following benefits:

• Up to 20 Lpm (liters per minute—depending on model)

• Up to 145 psig delivery pressure Proton OnSite has provided these instructions to guide in the installation, operation, and maintenance of the zero-air generator. These instructions provide technical product information, installation requirements, and detailed mechanical interface specifications. Important safety information is also included in this manual. Please take time to familiarize yourself with the system and the manual. This manual attempts to answer most of the frequently asked questions with regards to the operation of the unit. However, should you have any questions, the Proton OnSite technical staff stands ready to answer them and support the successful deployment of this equipment. Please call (203) 949-8697 and ask for field service technical support or email [email protected]. Please have the serial number and configuration number of your unit available.

IT IS THE CUSTOMER’S RESPONSIBILITY TO CONSULT WITH THE LOCAL AUTHORITY HAVING JURISDICTION (AHJ) REGARDING LOCAL CODE REQUIREMENTS FOR INSTALLATION AND OPERATION OF THIS EQUIPMENT.

Proton Onsite can offer a full range of installation services if you are not comfortable with installation of the zero-air generator or purifier.

There are no user serviceable parts inside the laboratory zero-air generator. Any interference with the internal parts voids the manufacturer’s warranty.

Page 9: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 9 of 22

General Information

The Air A5 and Air A20 are passive devices that reduce hydrocarbon content of an air-feed to less than 1 ppmv (parts per million by volume).

The ability to meet hydrocarbon specifications diminishes as the rated flow-rate is exceeded. This manual covers the installing, operating, and maintaining the Zero-Air Generator. The manual assumes the user is familiar with pneumatic circuit components, and is aware of appropriate safety precautions and use of compressed air and zero-air gases.

Detailed technical data are found in Product Specifications (Pg. 10), below. Dimensions for the Air A5 or Air A20 cabinet, or enclosure, are found in Table 1: Technical Specifications for the Zero Air 5 and Zero Air 20 Generators

Required fittings are displayed in Table 2 (Pg. 11).

Page 10: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 10 of 22

Product Specifications

Requirement Zero Air 5 Zero Air 20

Product Flow Rate 5 SLPM 20 SLPM

Purity < 1 ppm total hydrocarbons as methane

Delivery Pressure (Output Pressure), bar (psi) 0–10 bar (0–145 psi)

Input Voltage 120 or 230 VAC 50/60Hz, 8A, factory configured

Air Input Requirements Particulate ≤ 0.1 mg/m3 (≤ 0.01 μm)

Oil vapors ≤ 0.01 mg / m3 Moisture ≤ 3°C (37°F) dew point in pressure

0–10 bar (0–145 psi)

Enclosure Characteristics

External Dimensions (HxDxW) 39.4 x 30.5 x 20.3 cm (15.5 x 12 x 8 inches)

Weight 24 lbs / 10.9 kg

Environmental Considerations

Location–Min/Max Ambient Operating Temperature 5–40°C (40–104°F)

Location–Desired Ambient Operating Temperature1 20–25°C (68–77°F)

Location–Max Ambient Humidity ≤ 80% (non-condensing or HR)

Maximum Level or Altitude 2000 m (6580 ft) above sea level

Location–Variation from Level +/- 3º maximum

Location–Clearance Recommendation 0.30 m (12 in) Minimum on Left and Right Sides

Display Catalytic Bed Temperature Display

Environment Indoor, non-hazardous, non-classified, non-explosive

Safety

Enclosure Protection Rating IP20

Noise <54 dB A

Mechanical Ports Accept ¼” OD tubing

Expected yearly operator usage 6000 Hrs

Technologies Used catalytic reactor

Applications For GC FID, THC, NOX, BTX and SOX analyzers

Table 1: Technical Specifications for the Zero Air 5 and Zero Air 20 Generators

Page 11: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 11 of 22

Substance Flow Direction Fitting Mating Information Fitting Location

Purified Air Product Out ¼” Push to Connect 1/4"-OD Tubing Right Side

Air Inlet In ¼” Push to Connect 1/4"-OD Tubing Right Side

AC Power-In In Cord Set Cord Set Left Side Table 2: Generator Gas Inlet/Outlet Fittings

Page 12: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 12 of 22

2 Safety

The safety guidelines below may not cover all situations. If there are concerns or questions, please contact Proton OnSite Technical Service or check with local authorities.

THE AIR A5 AND AIR A20 ARE LABORATORY INSTRUMENTS DESIGNED FOR OPERATION BY TRAINED PERSONNEL.

ALL PERSONNEL HANDLING, USING OR MAINTAINING THIS GENERATOR MUST EMPLOY SAFE WORKING PRACTICES AND OBSERVE ALL RELEVANT LOCAL HEALTH AND SAFETY REGULATIONS.

Using Zero-Air Generators

Unit relies on a ventilation fan to ensure adequate cooling—do not block vents. The generator must be operated in an area with sufficient air circulation. The power switch and electrical disconnect (removable cord set) are located on the left side of the generator.

Safety Devices

MAXIMUM TEMPERATURE: The purifier is fitted with an internal catalyzer, the temperature of which is constantly maintained at 600°C. A thermal switch is attached to the catalyzer bed housing and will trip if the exterior of the catalyzer exceeds 104 °C (220 °F).

Technical Service Contact Information

Proton OnSite Technical Service can be contacted as follows:

Proton OnSite 10 Technology Drive Wallingford, CT 06492 203-949-8697 203-678-2000 [email protected] www.protononsite.com

Page 13: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 13 of 22

3 Placement

Providing Suitable Location

To avoid risks of damage to its electronic components, install the purifier in an environment consistent with the specification called out in Table 1. The ambient temperature in the purifier installation area must be between 5°C and 40°C.

Unpacking the Generator

The purifier is delivered in cardboard packaging. Remove all packaging, taking care not to damage the purifier’s panels. If possible, store the packaging to ensure adequate protection when moved in the future.

Unpack the generator and retain the packing materials in case it is necessary to ship it in the future.

The Air A5 and Air A20 each weigh approximately 10.9 kg (24 lbs). Please use the following instructions to unpack your Air5/Air20 zero-air generator: 1. Remove the Air5/Air20 zero-air generator from the box carefully by lifting it straight up

while grasping underneath the generator from the front and back.

2. Keep the generator in an upright position and place the generator carefully in the designated location for it.

• THE PURIFIER MUST ALWAYS OPERATE WITH ITS PROTECTIVE HOUSING INSTALLED.

• NEVER REMOVE THE COVER WHILE THE PURIFER IS OPERATING OR FOR 24 HOURS AFTER SHUTDOWN FOR SERVICE.

• WHEN OPERATION HAS CEASED. THE CATALYZER BED WILL REMAIN HOT AND PRESENTS A BURN HAZARD.

Page 14: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 14 of 22

Positioning the Generator

The zero-air generator is designed for indoor, laboratory installation. AC power that meets the specifications outlined in Product Specifications (Section 0 on Page 10) is the only utility required. Verify that the generator is located so that it complies with local safety regulations. Following instructions to place your zero-air generator:

1. Place the generator in a location that meets the technical specifications from Product Specifications (Section 0 on Page 10). Some specifications are highlighted below. We recommend that the generator be placed as close to the gas application as possible.

2. Locate the unit, keeping in mind the continuing need for access to controls, gas outlets, and power and control wiring.

3. Install on a level site (+/- 3° required).

4. Keep the front of unit clear with at least 30 cm (12 in) of space on either side of the generator for adequate ventilation air flow. The power switch and electrical disconnect (removable cord set) are located on the left side of the unit and must be accessible at all times during operation of the generator. Position the unit so that it is possible to operate the disconnecting device.

Figure 1 Installation Clearances

Page 15: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 15 of 22

4 Installation Procedure

4.1 Mechanical Setup

Air Inlet and Outlet: The air inlet distribution lines must be made using tubing rated to the operating pressure and the pipeline diameters must be sized according to the acceptable pressure drop values. The max acceptable inlet pressure is 10 bar (145 psi).

This section provides a detailed description of the mechanical setup requirements. 1. Locate the filters, marked “INLET” and “OUTLET” at the right of the purifier (See

Figure 2). 2. Make the appropriate pneumatic connection between the purifier and the lines.

Figure 2: Zero Air 5/Zero Air 20 Inlet/Outlet, Right Side

Air Inlet Filter

(Coalescing) Air Outlet Filter

(Particulate)

Page 16: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 16 of 22

4.2 Electrical Setup

This section provides a detailed description of the mechanical setup requirements. 1. Determine the electrical supply outlet that serves the generator.

2. Ensure that the electrical supply is properly grounded and the voltage corresponds with the generator rating plate.

Do not bypass ground pin. 3. Ensure that there is a good fuse in the power switch fuse drawer (located under the cord

set female fitting). The proper fuse is rated for 8 amps. 4. Connect the supplied cord set to the side of the unit.

5. Plug the generator into the electric outlet.

4.3 Before Using the Generator

1. Switch the power on using the power switch on the left side of the unit (Figure 3: Power Switch).

2. The temperature controller power on and begins heating the catalyzer bed. The temperature of the bed is displayed in degrees Celsius.

3. Allow the unit to reach the operating temperature of 600° C before flowing air through the system.

4. When the set point is reached, allow air to flow through the unit. Adjust the air flow to ensure the flow rate does not exceed the rating of the unit.

5. The catalyzer should take less than 60 minutes to reach operating temperature.

Page 17: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 17 of 22

Figure 3: Power Switch, Left Side

Hydrocarbon content may not be within specs until the catalyzer has established operating temperature.

4.4 Shutdown

Under normal circumstances, no special procedures are required for shutdown of the zero-air generator. Simply use the electronic controls as follow:

1. Shut off the power using the power switch on the side of the generator. 2. Turn off the flow of feed-air to the zero-air generator.

This procedure ceases generation and zero-air flow and removes all power from all system components. To restart, plug in the unit and follow the instructions found on page 16.

Power Inlet Module

Page 18: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 18 of 22

5 Maintenance

The Zero Air generator uses an inlet coalescing filter and a particulate filter on the outlet of the generator. These filters should be replaced at a minimum of 6 month intervals. The Zero Air generator does not include consumable parts inside the enclosure. Scheduled maintenance is limited to the external filters every 6 months and a periodic inspection (general overhaul) recommended every 6,000 hours of operation. During this inspection, the unit is checked for leaks. The cooling fan is also checked to make sure that it's running properly.

The lifetime of the catalyzer depends under what conditions it is used, in particular, the quality of inlet air. Refer to Table 1on page 10.

The Inlet filter kits can be ordered by contacting Proton OnSite Customer Service. The part number for the annual filter kit for the Air A5 and Air A20 is KT-0201-0019

Cleaning the Generator

1. Clean only the outside of the generator. 2. Disconnect the generator power cord before cleaning the generator.

3. Use a damp sponge or soft cloth and a mild soap solution to clean the outside of the generator.

TO AVOID MARRING THE PLASTIC AND POWDER COATED METAL SURFACES, DO NOT USE ABRASIVE CLEANERS ON THE GENERATOR.

Page 19: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 19 of 22

6 Troubleshooting

These purifier models do not have alarms.

In case of malfunction, make sure that the gas flow through the purifier does not exceed the maximum value given in its technical specifications (Section 0, Page 10); and/or check the displayed operating temperature. If the temperature on the display does not increase, please call the factory. Phone and contact info is found on page 2.

Page 20: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 20 of 22

7 APPENDIX A—FORMS

Customer Registration Form

Please fill out the form on the following page, tear it out of the manual, and fax it to Proton OnSite (Fax Number: 1-203-949-8016) to ensure that your Zero Air 5/Zero Air 20 Generator is properly registered. The form may also be sent by mail to the following address:

Customer Service Proton OnSite, Inc. 10 Technology Drive Wallingford, CT 06492 Proper registration is important for product support and automatic notification regarding routine maintenance.

Page 21: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 21 of 22

This page is intentionally left blank.

Page 22: Air A5, Air A20 - Cole-Parmer · Installation, Operation & Maintenance Manual PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 3 of 22 Disclaimer Proton OnSite has written

Air A5 and Air A20 Laboratory Zero-Air Generators

Installation, Operation & Maintenance Manual

PD-0100-0073 Rev. A --UNAUTHORIZED COPIES PROHIBITED-- Page 22 of 22

CUSTOMER REGISTRATION FORM

LCMS Laboratory Zero Air 5 / Zero Air 20 Generator Generator Model Name?

Generator Serial Number?

From what supplier did you purchase the generator?

Name of your company?

Company website address?

The application for your zero-air generator?

Your name?

Your title?

Your phone number with country code?

Your fax number with country code?

Your e-mail address?

The place where the generator is located (city, state, country)

Are you the person responsible for the operation of the zero-air generator?

Do you have any questions at this time that you need Proton OnSite to answer?