Innova® E-SeriesAutomated Endoscope Reprocessors With CMS - Channel Monitoring System

Page 2: Automated Endoscope Reprocessors - SciCan Medical · Automated Endoscope Reprocessors ... The Innova E4 accepts two flexible, immersible endoscopes, ... Installation and technical

An innovative approach to endoscope reprocessing

INNOVA E3 with CMSSTAND-ALONE

Water quality you can rely on.Innova utilizes a thermal disinfection process to provide high quality rinse water. DI water is heated in an isolated boiler, cooled, then released into the chamber, meeting endoscope manufacturers’ requirements.

Single use chemistry.The Innova is a complete washer-disinfector with an accurate dosing system that assures the correct chemical concentration and temperature for the washing and disinfection phases. This eliminates the need for daily testing of disinfectant potency.

Bar coding technology. Critical information, such as endoscope serial numbers or the operator's identification is scanned and recorded. This information appears on the print-out and is stored in the cycle data.

Superior ventilating system.All Innova AERs are equipped with high-efficiency venting systems which minimises staff exposure to chemical fumes.

LEAK TESTING

is performed at the beginning of every cycle.

COLD WATER RINSE

Remove debris and avoid blood coagulation.

WASH CYCLE

Fluid flows through the channels of the endoscopes and washes the exterior surface as well.

DISINFECTION STAGE

High level disinfection is achieved in 6 minutes only.

INNOVA E-Series

Page 3: Automated Endoscope Reprocessors - SciCan Medical · Automated Endoscope Reprocessors ... The Innova E4 accepts two flexible, immersible endoscopes, ... Installation and technical

INNOVA E4 with CMSPASS-THROUGH

An innovative approach to endoscope reprocessing

INNOVA E5 with CMSPASS-THROUGH

DISINFECTION STAGE

High level disinfection is achieved in 6 minutes only.

WARM RINSE

with deionized water to remove the disinfectant chemical from the channels and on the surface.

THE COOLING CYCLE

takes only a few minutes.

WARM AIR DRYING CYCLE

ensures that the channels are free of moisture and the endoscopes are ready for immediate use or for overnight storage.

INNOVA E-Series

INNOVA E4 with CMSPASS-THROUGH

Page 4: Automated Endoscope Reprocessors - SciCan Medical · Automated Endoscope Reprocessors ... The Innova E4 accepts two flexible, immersible endoscopes, ... Installation and technical

INNOVA E-Series

CMS – Channel Monitoring SystemThe INNOVA CMS – Channel Monitoring System is an individual channel check option available for the INNOVA E3, E4 and E5. The system checks all channels of an endoscope for any restrictions. Obstructions or any other causes which narrow a channel will be detected and the operator will be warned in time.

The CMS is a modular system allowing to monitor up to six channels per endoscope.

The INNOVA CMS offers the highest quality and validatable process security, and improve: operators and patients safety

handling for the reprocessing staff

cost savings for the institution

the quality and life time of the endoscope in comparison to conventional methods of reprocessing.

When an endoscope is first used, the “scope setup” cycle is run. This characterises the baseline flow for this particular endoscope.

Every time this endoscope is processed thereafter, the CMS compares the pressure readings in each channel to those in the baseline.

The pressures are carefully controlled and the measurements are recorded in the database. If a channel is blocked or if an adapter is loose or disconnected, the location of the problem is specified.

Scope Setup Comparison Control1 2 3

Page 5: Automated Endoscope Reprocessors - SciCan Medical · Automated Endoscope Reprocessors ... The Innova E4 accepts two flexible, immersible endoscopes, ... Installation and technical

INNOVA E-SeriesINNOVA E-Series

The Innova E3 is designed for use in clinics and hospitals. It has a capacity for two endoscopes, has a 36-minute cycle time and can be installed under a counter or free-standing. An optional base is available, providing an ergonomic working height for the load carrier and space for the chemical bottles. The Innova E3's bar code scanner easily records scope serial numbers as well as, staff identification. This information, together with cycle data is recorded on the server.

Compact, extremely reliable and cost-effective, the E3 is the optimal solution to answer your endoscope reprocessing requirements.

Features:Two-scope capacity

Single manual hinged door

Single-use chemistry

Superior drying system

Washing and disinfection in about 36 minutes

Endoscope bar coding identification

Print-out after every cycle1

Thermal disinfection program for the chamber and the entire washing system

Internet based remote maintenance with live support (optional)

INNOVA E3 with CMS

Technical Specifications*

*Specifications subject to change without notice.

1 Optional printer required

DescriptionSingle door, front loader model with manual hinged door

Capacity2 endoscopes

Channel Monitoring System (CMS)Yes

Channels per endoscope6

Dimensions W x H x DFree standing unit 1000 x 1440 x 800 mm

Under counter unit 1200 x 860 x 680 mm

Washing chamber 540 x 510 x 535 mm

Material used for panels and washing chamberStainless steel AISI 304

Electric connections208V, 3 ph, 60Hz, 16 amp

Compressed air for drying (with sterile air filter)DN 10, 2 Bar

Wash Supply Cold and RO-Water 3/4", DN 20

DrainageDN 50

DryingYes

Exhaust connection DN 70

Integrated leak test Yes

Able to use glutaraldehyde as disinfectantYes

Able to use peracetic acid as disinfectantYes

CMS – Channel Monitoring System

TWO SCOPE STAND-ALONE

Page 6: Automated Endoscope Reprocessors - SciCan Medical · Automated Endoscope Reprocessors ... The Innova E4 accepts two flexible, immersible endoscopes, ... Installation and technical

DescriptionDouble-door pass-through model with manual hinged doors

Capacity2 endoscopes

Channel Monitoring System (CMS)Yes

Channels per endoscope6

Dimensions W x H x DMachine 1000 x 2000 x 750 mm

Washing chamber 500 x 625 x 650 mm

Material used for panels and washing chamberStainless steel AISI 304

Electric connections208V, 3 ph, 60Hz, 16 amp

Compressed air for drying (with sterile air filter)DN 10, 2 Bar

Wash Supply Cold and RO-Water 3/4", DN 20

DrainageDN 70

DryingYes

Exhaust connection DN 100

Integrated leak test Yes

Able to use glutaraldehyde as disinfectantYes

Able to use peracetic acid as disinfectantYes

The Innova E4 is designed for use in Central Sterile Services Departments. It has two doors, one on the “clean” side and one on the “dirty” side.

Endoscopes are loaded on the dirty side, where the control panel is installed and removed on the clean side, where the printer1 is located. The Innova E4’s software ensures that the user cannot open both doors at the same time.

The Innova E4 accepts two flexible, immersible endoscopes, and completes a cycle in about 37 minutes. Extended chemical suction hoses allow the chemical supply bottles (enzymatic detergent and high-level disinfectant) to be located at a distance from the machine.

*Specifications subject to change without notice.

Features:Two-scope capacity

Two doors, with control panel on dirty side and printer1 on clean side

Single-use chemistry

Superior drying system

Washing and disinfection in about 37 minutes

Endoscope bar coding identification

Print-out after every cycle1

Thermal disinfection program for the chamber and the entire washing system

Internet based remote maintenance with live support (optional)

INNOVA E-Series

INNOVA E4 with CMS

Technical Specifications*

1 Optional printer required

TWO SCOPE PASS-THROUGH

*Specifications subject to change without notice.

1 Optional printer required

Page 7: Automated Endoscope Reprocessors - SciCan Medical · Automated Endoscope Reprocessors ... The Innova E4 accepts two flexible, immersible endoscopes, ... Installation and technical

DescriptionDouble-door pass-through model with automatic sliding doors

Capacity2-4 endoscopes

Channel Monitoring System (CMS)Yes

Channels per endoscope6

Dimensions W x H x DMachine 1400 x 2000 x 840 mm

Washing chamber 650 x 625 x 680 mm

Material used for panels and washing chamberStainless steel AISI 304

Electric connections208V, 3 ph, 60Hz, 16A

Compressed air for drying (with sterile air filter)DN 10, 2 Bar

Wash Supply Cold and RO-Water 3/4", DN 20

DrainageDN 70

DryingYes

Exhaust connection DN 150

Integrated leak test Yes

Able to use glutaraldehyde as disinfectantYes

Able to use peracetic acid as disinfectantYes

While the Innova E4 with hinged doors is designed for smaller capacities, the Innova E5 with sliding doors is ideal for volumes of up to four endoscopes. When placed in a line, the pass-through models form a perfect, individual clean room concept.

The transport trolley is the ideal supplement to Innova E5 units. The trolley aids loading and unloading and also ensures safe and comfortable transportation of the reprocessed items.

The Innova E5 offers unique, unparalleled flexibility. For hospitals expecting future procedural growth, the E5 can initially be installed in the two scope configuration and expanded to four scope configuration when required.

Features:Up to four scope capacity

Two doors, with control panel on dirty side and printer1 on clean side

Single-use chemistry

Superior drying system

Washing and disinfection in about 37 minutes

Endoscope bar coding identification

Print-out after every cycle

Thermal disinfection program for the chamber and the entire washing system

Internet based remote maintenance with live support (optional)

INNOVA E-SeriesINNOVA E-Series

INNOVA E5 with CMS

Technical Specifications*

*Specifications subject to change without notice.

1 Optional printer required

FOUR SCOPE PASS-THROUGH FOUR SCOPE PASS-THROUGH

Page 8: Automated Endoscope Reprocessors - SciCan Medical · Automated Endoscope Reprocessors ... The Innova E4 accepts two flexible, immersible endoscopes, ... Installation and technical

Hygiene Management System

www.scican.com

FOLLOW US ON:Distributed by:

SciCan Ltd.

1440 Don Mills Road, Toronto, Ontario M3B 3P9Tel: 416.445.1600 Fax: 416.445.2727Toll free: 1.800.667.7733 [email protected]

Manufactured by:

BHT Hygienetechnik GmbHMesserschmittstrasse 11D - 86368 GersthofenTel: +49 (0)821 27893-0Fax: +49 (0)821 27893-79

Technical Services

Installation and technical service is performed by skilled technicians, who have been trained in the design and operation of the Innova. If you require a service contract, we can tailor one to meet your explicit needs. An extensive inventory of parts is also at your disposal.

Preventative maintenance and repair

Customized service contracts

In-Service training provided by a clinical specialist

The Hygiene Management System (HMS) module independently documents the automated cleaning and disinfection process, and correlates this information to the barcode of each specific endoscope. All essential process data are stored on the hospital network. This includes the details of the pressure measurements from each endoscope channel.

NetworkWhen multiple Innova machines are networked, endoscope calibration data is centralized on the server and available to all machines. A single report can be generated to show an endoscope’s reprocessing history, regardless of the machines in which it has been processed.