1

ECG

(Electrocardiogram)

Achmad RizalBioSPIN

ARL-EL4703-Instrumentasi Biomedis 2

Cardiac Conduction System (1)

ARL-EL4703-Instrumentasi Biomedis 3

Cardiac Conduction System (2)

ARL-EL4703-Instrumentasi Biomedis 4

The ECG (1)

ARL-EL4703-Instrumentasi Biomedis 5

The ECG (1)

ARL-EL4703-Instrumentasi Biomedis 6

The ECG (1)Electrocardiography1. Standard clinical ECG (12 leads)2. VCG (3 orthogonal leads)3. Monitoring ECG (1 or 2 leads)

Page 2: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 2/92

ARL-EL4703-Instrumentasi Biomedis 7

Modern ECG

Figures above show how theearliest ECGs were recorded byEinthoven at 1900.Vats of salt water provided theelectrical connection to the body.The string galvanometer served asthe measurement instrument forrecording the ECG. ARL-EL4703-Instrumentasi Biomedis 8

Electrocardiograph

ARL-EL4703-Instrumentasi Biomedis 9

Holter Monitor

ARL-EL4703-Instrumentasi Biomedis 10

Dipole Model

ARL-EL4703-Instrumentasi Biomedis 11

Standard Location ?

ARL-EL4703-Instrumentasi Biomedis 12

ECG Lead Configurations• A pair of electrodes, or a combination of several electrodes

through a resistive network that gives an equivalent pair iscalled a lead.

• A vector that connects a lead electrode pair is the lead vector.• If the cardiac vector M is known, the voltage generated at any

lead can be easily computed.

Page 3: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 3/93

ARL-EL4703-Instrumentasi Biomedis 13

Frontal Plane Bipolar LeadsEinthovenTriangle

ARL-EL4703-Instrumentasi Biomedis 14

The Einthoven Triangle

ARL-EL4703-Instrumentasi Biomedis 15

Lead I

ARL-EL4703-Instrumentasi Biomedis 16

Lead II

ARL-EL4703-Instrumentasi Biomedis 17

Lead III

ARL-EL4703-Instrumentasi Biomedis 18

Cardiac Vector Through the Cardiac CycleAs Seen By Leads I, II and III

Page 4: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 4/94

ARL-EL4703-Instrumentasi Biomedis 19

Cardiac VectorThrough theCardiac Cycle AsSeen By Leads I,II and III

ARL-EL4703-Instrumentasi Biomedis 20

Cardiac VectorThrough the CardiacCycle As Seen ByLeads I, II and III

ARL-EL4703-Instrumentasi Biomedis 21

Cardiac Vector Throughthe Cardiac Cycle AsSeen By Leads I, II andIII

ARL-EL4703-Instrumentasi Biomedis 22

UnipolarMeasurements:Wilson Central Terminal

• Three additional leads are used for frontal planemeasurements.

• These are the measurements at the specific electrode, withrespect to a reference electrode

• One commonly used reference electrode is the WilsonCentral Terminal, obtained through a resistive network,combining limb electrodes

• The new set of leads obtained by combining the standardlimb electrodes to the Wilson terminal form the augmentedleads

• These leads provide additional vector views of cardiacdepolarization in the frontal plane. Unlike leads I, II, III, theaugmented leads utilize WCT, a central negative terminal.This virtual "electrode" is calculated by the EKG computerto measure vectors originating roughly at the center of theheart.

Note that the voltage at Wilson’s terminal is zero (why?)

ARL-EL4703-Instrumentasi Biomedis 23

Wilson’s Central Terminal

ARL-EL4703-Instrumentasi Biomedis 24

Augmented Leads: aVR

21 III

V V aVR −−=

Page 5: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 5/95

ARL-EL4703-Instrumentasi Biomedis 25

Augmented Leads: aVL

21 II

V

V aVL −=

ARL-EL4703-Instrumentasi Biomedis 26

Augmented Leads: aVF

2

I II

V V aVF −=

ARL-EL4703-Instrumentasi Biomedis 27

All Frontal Leads

ARL-EL4703-Instrumentasi Biomedis 28

Frontal Lead

ARL-EL4703-Instrumentasi Biomedis 29

Transversal Lead

ARL-EL4703-Instrumentasi Biomedis 30

12 LeadsECGRecording

Page 6: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 6/96

ARL-EL4703-Instrumentasi Biomedis 31 ARL-EL4703-Instrumentasi Biomedis 32

ECG Intrumentation

• Single Channel ECG

ARL-EL4703-Instrumentasi Biomedis 33

ECG Intrumentation

Electrode Transducer Characteristic :

acquiring signal withamplitude between0.01mV – 3mVanticipate impedance till> 5 M Ω

small current leakages< 1 µ-Aresponse frequency0.05 – 150 Hz

ARL-EL4703-Instrumentasi Biomedis 34

ECG Intrumentation

ECG Amplifier schematicsRight leg driven loop- noise reducing- patient protection- Reduce Common Mode

Voltage

ARL-EL4703-Instrumentasi Biomedis 35

ECG Intrumentation

• Instrumentation Amplifier

ARL-EL4703-Instrumentasi Biomedis 36

ECG Intrumentation

• AD620 single chip instrumentationamplifier

Page 7: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 7/97

ARL-EL4703-Instrumentasi Biomedis 37

ECG IntrumentationHigh Pass Filter Second Amplifier

Low Pass Filter 20 Hz

Clamper

Low Pass Filter 40 Hz

Analog subsystemARL-EL4703-Instrumentasi Biomedis 38

ECG Intrumentation

Digital Subsystem

ARL-EL4703-Instrumentasi Biomedis 39

ECG Intrumentation

Display Unit• Galvanometer • Osciloskop• CRT

• LCD• PC + Monitor • Paper chart• PDA/HP display

ARL-EL4703-Instrumentasi Biomedis 40

Some Problems in ECG

Noise & interferences

ARL-EL4703-Instrumentasi Biomedis 41

Some Problems in ECG

• Souces of interference in an ECG trace – base line wander : patient movement, electrode

reposition, dry electrode, static buildup around patient

– AC noise (50 Hz) : powerline/common modeinterfrence

– intermittent signal : lead connection not tightand secured, dry electrode, damage cable orlead wire

– Other Electrical interference (EI)

Another ECG System

Page 8: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 8/9

Page 9: Bab7-ECG

8/16/2019 Bab7-ECG

http://slidepdf.com/reader/full/bab7-ecg 9/9

ARL-EL4703-Instrumentasi Biomedis 49

Low Power, Low Cost,Wireless ECG(Holter Monitor)

ESDDefibrillation

Protection

ADuC7022

ADF702x*

ESDDefib

Protect

RLElectrode

RAElectrode

RLElectrode

AD623G=10.6

High-PassPole0.5Hz

0.47μF

5pF

681kΩ

499kΩ

5.23kΩ 5.23kΩ

475Ω 10kΩ

108pF

Anti-aliasFilter-3dB, 1592 Hz

AdditionalDigitalFilter

Level ShiftAnti-Alias

TotalGain ~ 220

G =22.05

AD8641

AD8603

Dependent on FrequencyBand Used( WMTS or ISM)

ARL-EL4703-Instrumentasi Biomedis 50

Low Cost, Low Power AED

Filter and

ProtectionNetwork

ADR365

AD7685

HighVoltage

Discharge

Power Control

ADP1712

AD8220/4 AD8606

AD8605

PatientPads

Li IonBattery

ADuC70xxor

Blackfin®

LCD Display

Single poleanti-alias

filter

Digital Filtering

Blanking

ARL-EL4703-Instrumentasi Biomedis 51

Exercise Heart Rate Monitor (Example: treadmill with heartrate monitor)

AD8692AD8692

REF

REF REF

AD8553

AD8691

High PassFilter

ADuC70xxREF

InputProtection

+5V

40 HzLowPass Filter