16
EHRA educational review and preparatory course for accreditation examination Invasive Cardiac Electrophysiology examination preparatory course Basic concepts, techniques and safety issues of arrhythmia ablation Prof. Dr. M. Antz Oldenburg Heart Center, Germany European Heart House Sophia-Antipolis, France 25th of February, 2011: 9:00 - 10:30 Matthias Antz

Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

EHRA educational review and preparatory course for accreditation examination Invasive Cardiac Electrophysiology examination preparatory course

Basic concepts techniques and safety

issues of arrhythmia ablation

Prof Dr M Antz Oldenburg Heart Center Germany

European Heart House Sophia-Antipolis France

25th of February 2011 900 - 1030

Matthias Antz

Power How RF ablation works

bull RFC is passed through the patient circuit The tissue around the probe tip is heated by the electric current

bull The rise in tissue temperature causes a lesion and interrupts transmission of myocardial signals

bull The tip does not get hot itself

bull The temperature rise takes place directly in the tissue and is measured in the tip by a

Thermocouple exactly in the

center of a full metal tip

for best thermal conductivity

or

Thermistor compatible with older

generator technologies

Tissue Heating

Electrode tip

heated from

the tissue

Heat lost to

circulating blood

Zone of resistive

heating

Zone of conductive

heating

Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

Current flow

The current follows two pathways blood and tissue each with different impedance (R)

bull RBlood is smaller than RTissue most current ist lost to blood bull Good wall contact more contact area more current

enters the tissue

Temperature controlled ablationEffects of low and high blood flow

Low blood flow High blood flow target temperature reached target temperature reached with 15 W small lesion with 30 W large lesion

Conventional Electrodes 4 mm vs 8 mm Tip

18 W

60deg 60deg

35 W

4 mm Tip 8 mm Tip

Electrode Electrode plt005 vs

Blood Flow 350 mlmin 4mm Tip

Otomo et al JCE 1998947-54

Experimental Experience

1) High blood flow allows higher power delivery to

the tissue and this results in deeper lesions

2) Larger electrodes provide greater electrode

cooling allowing higher power delivery and this

results in deeper lesions

3) Small electrodes have a higher recording

resolution

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 2: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Power How RF ablation works

bull RFC is passed through the patient circuit The tissue around the probe tip is heated by the electric current

bull The rise in tissue temperature causes a lesion and interrupts transmission of myocardial signals

bull The tip does not get hot itself

bull The temperature rise takes place directly in the tissue and is measured in the tip by a

Thermocouple exactly in the

center of a full metal tip

for best thermal conductivity

or

Thermistor compatible with older

generator technologies

Tissue Heating

Electrode tip

heated from

the tissue

Heat lost to

circulating blood

Zone of resistive

heating

Zone of conductive

heating

Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

Current flow

The current follows two pathways blood and tissue each with different impedance (R)

bull RBlood is smaller than RTissue most current ist lost to blood bull Good wall contact more contact area more current

enters the tissue

Temperature controlled ablationEffects of low and high blood flow

Low blood flow High blood flow target temperature reached target temperature reached with 15 W small lesion with 30 W large lesion

Conventional Electrodes 4 mm vs 8 mm Tip

18 W

60deg 60deg

35 W

4 mm Tip 8 mm Tip

Electrode Electrode plt005 vs

Blood Flow 350 mlmin 4mm Tip

Otomo et al JCE 1998947-54

Experimental Experience

1) High blood flow allows higher power delivery to

the tissue and this results in deeper lesions

2) Larger electrodes provide greater electrode

cooling allowing higher power delivery and this

results in deeper lesions

3) Small electrodes have a higher recording

resolution

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 3: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Tissue Heating

Electrode tip

heated from

the tissue

Heat lost to

circulating blood

Zone of resistive

heating

Zone of conductive

heating

Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

Current flow

The current follows two pathways blood and tissue each with different impedance (R)

bull RBlood is smaller than RTissue most current ist lost to blood bull Good wall contact more contact area more current

enters the tissue

Temperature controlled ablationEffects of low and high blood flow

Low blood flow High blood flow target temperature reached target temperature reached with 15 W small lesion with 30 W large lesion

Conventional Electrodes 4 mm vs 8 mm Tip

18 W

60deg 60deg

35 W

4 mm Tip 8 mm Tip

Electrode Electrode plt005 vs

Blood Flow 350 mlmin 4mm Tip

Otomo et al JCE 1998947-54

Experimental Experience

1) High blood flow allows higher power delivery to

the tissue and this results in deeper lesions

2) Larger electrodes provide greater electrode

cooling allowing higher power delivery and this

results in deeper lesions

3) Small electrodes have a higher recording

resolution

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 4: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Current flow

The current follows two pathways blood and tissue each with different impedance (R)

bull RBlood is smaller than RTissue most current ist lost to blood bull Good wall contact more contact area more current

enters the tissue

Temperature controlled ablationEffects of low and high blood flow

Low blood flow High blood flow target temperature reached target temperature reached with 15 W small lesion with 30 W large lesion

Conventional Electrodes 4 mm vs 8 mm Tip

18 W

60deg 60deg

35 W

4 mm Tip 8 mm Tip

Electrode Electrode plt005 vs

Blood Flow 350 mlmin 4mm Tip

Otomo et al JCE 1998947-54

Experimental Experience

1) High blood flow allows higher power delivery to

the tissue and this results in deeper lesions

2) Larger electrodes provide greater electrode

cooling allowing higher power delivery and this

results in deeper lesions

3) Small electrodes have a higher recording

resolution

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 5: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Temperature controlled ablationEffects of low and high blood flow

Low blood flow High blood flow target temperature reached target temperature reached with 15 W small lesion with 30 W large lesion

Conventional Electrodes 4 mm vs 8 mm Tip

18 W

60deg 60deg

35 W

4 mm Tip 8 mm Tip

Electrode Electrode plt005 vs

Blood Flow 350 mlmin 4mm Tip

Otomo et al JCE 1998947-54

Experimental Experience

1) High blood flow allows higher power delivery to

the tissue and this results in deeper lesions

2) Larger electrodes provide greater electrode

cooling allowing higher power delivery and this

results in deeper lesions

3) Small electrodes have a higher recording

resolution

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 6: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Conventional Electrodes 4 mm vs 8 mm Tip

18 W

60deg 60deg

35 W

4 mm Tip 8 mm Tip

Electrode Electrode plt005 vs

Blood Flow 350 mlmin 4mm Tip

Otomo et al JCE 1998947-54

Experimental Experience

1) High blood flow allows higher power delivery to

the tissue and this results in deeper lesions

2) Larger electrodes provide greater electrode

cooling allowing higher power delivery and this

results in deeper lesions

3) Small electrodes have a higher recording

resolution

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 7: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Experimental Experience

1) High blood flow allows higher power delivery to

the tissue and this results in deeper lesions

2) Larger electrodes provide greater electrode

cooling allowing higher power delivery and this

results in deeper lesions

3) Small electrodes have a higher recording

resolution

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 8: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Different electrodes

4mm 8mm irrigated

Recording resolution high low high

Risk of thrombus high high low

Large lesions possible no yes yes

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 9: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Irrigated tip ablation ndash Lesion size

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 10: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Irrigated Tip Ablation ndash Lesion shape

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 11: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Ablations using irrigated tip (Oldenburg)

bull Irrigation flow - mapping 2 mlmin - ablation 17 mlmin (lt30W) - ablation 30 mlmin (ge30W)

bull Power control - CSMCV 10 W (increase in 5 W steps) - other parts 30 W (max 40 W caveat 50W)

bull Temperature limit 43degC bull Duration of ablation 120 sec bull Cave volume overload due to irrigation fluid

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 12: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Effect of Irrigation Flow on Ablation (50W 60s)

10mlmin 17mlmin 30mlmin 60mlmin

Electrode Temp 59 6degC 47 5degC 39 3degC 34 2degC

85 33 0 0

46 13 0 0

Irrigation Flow

35 mm Irrigated Tip Electrode (7F)

Thrombus

Impedance Rise

plt005 between flow rates

Matsudaira Nakagawa et al NASPE 1999

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 13: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

26 W 36 W

2 mm versus 5 mm Irrigated Electrode

30 W 30 W

Irrigated Tip Irrigated Tip 2 mm 5 mm

Nakagawa et al Circulation 1998 98 458-465

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 14: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

RF Generator 50 V

RF Generator 50 V

RRemoteRRemote

RTissueRTissue

RBloodRBlood

Ablation ElectrodeAblation Electrode

Circuit for RF Ablation

RRemote 30 W 30 WRTissue 199 W 198 WRBlood 103 W 50 W

RRemote 30 W 30 W RTissue 199 W 198 W RBlood 103 W 50 W

2 mm 5 mm2 mm 5 mm perpendicularperpendicular

Blood

Tissue

Nakagawa et al Circulation 1998 98 458-465

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 15: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Summary

bull RFC heats the tissue which then heats the ablation electrode

bull Blood flow and electrode-tissue contact have the largest impact on lesion formation during conventional RFCablation

bull During ablation the electrode temperature power and impedance should be monitored

bull Risk of thrombus formation is increased at high electrode temperatures

bull High electrode temperatures predominantely occur at highpower and can be avoided by using an irrigated tip electrode

Matthias Antz

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469

Page 16: Basic concepts techniques and safety issues of arrhythmia ...assets.escardio.org/.../Antz-talk-Basic-Concepts.pdfInvasive Cardiac Electrophysiology examination preparatory course Basic

Literature

1 Antz M Otomo K Yamanashi WS Nakagawa H Jackman WM Kuck KH Radiofrequency current catheter ablation with the

split tip electrode in the temperature controlled mode Pacing Clin Electrophysiol 2001 24(12)1765-1773

2 Dorwarth U Fiek M Remp T Reithmann C Dugas M Steinbeck G Hoffmann E Radiofrequency Catheter Ablation

Different Cooled and Noncooled Electrode Systems Induce Specific Lesion Geometries and Adverse Effect Profiles PACE

2003261438-45

3 Huang and Wilber Radiofrequency Catheter Ablation of Cardiac Arrhythmias Futura Publishing 2000

4 Matsudaira K Nakagawa H Wittkampf FH Yamanshi WS Imai S Pitha JV Lazzara R Jackmann WM High incidence of

thrombus formation without impedance rise during radiofrequency ablation using electrode temperature control Pacing Clin

Electrophysiol 200326(5)1227-37

5 Nakagawa H Wittkampf FHM Yamanashi WS Pitha JV Imai S Campbell B Arruda M Lazzara R JackmanWM Inverse

Relationship Between Electrode Size and Lesion Size During Radiofrequency Ablation With Active Electrode Cooling

Circulation 199898458-465

6 Otomo K Yamanashi WS Tondo C Antz M Bussey J Pitha JV Arruda M Nakagawa H Wittkampf FH Lazzara R

Jackman WM Why a large tip electrode makes a deeper radiofrequency lesion effects of increase in electrode cooling and

electrode-tissue interface area J Cardiovasc Electrophysiology 19989(1)47-54

7 Petersen HH Chen X Pietersen A Svendsen JH Haunso S Lesion Dimensions During Temperature-Controlled

Radiofrequency Catheter Ablation of Left Ventricular Porcine Myocardium Impact of Ablation Site Electrode Size and

Convective Cooling Circulation 199999319-25

8 Rodriguez LM Nabar A Timmermans C Wellens HJJ Comparison of results of an 8-mm split-tip versus a 4-mm Tipp

ablation catheter to perform radiofrequency ablation of type I atrial flutter Am J Cardiol 200085(1)109-112

9 Rosenbaum R Greenspon AJ Smith M Walinsky P Advances radiofrequency catheter ablation in canine myocardium Am

Heart J 1994127(4)851-857

10 Weiss C Antz M Eick O Eshagzaiy K Meinertz T Willems S Radiofrequency catheter ablation using cooled electrodes

impact of irrigation flow rate and catheter contact pressure on lesion dimensions Pacing Clin Electrophysiol 200225(4)463-

Matthias Antz

469