B i-Metric
Total Hip System
®
proven
erformanceP
“Closed-Pore”porous coating
potentially sealsthe proximal
femur.(Biomet)
“Open-Pore”porous coatingallowsdebrismigration.(Competition)
AUTHOR
Evans and DeLeeMauerhan, et al.McLaughlinMulti-Center Study
Mallory, et al.RothmanCapello, McClainHeekin, et al.Woolson, MaloneyKim, et al.Kim, et al.Smith, HarrisEngh
REFERENCE
Submitted for PublicationJ. Arthroplasty, 1997JBJSBiomet Clinical Report 1994
AAOS 1995Orthopedics 1994Trans. Int’l Symposium 1992JBJS 1993J. Arthroplasty, 1992Orthop. Trans, 1992–93Orthop. Trans, 1992–93CORR, 1995Presentation, 1992
HIP IMPLANT SYSTEM
Bi-Metric® (Biomet)Integral® (Biomet)Taperloc™ (Biomet)Taperloc,™ Mallory-Head,®
Bi-Metric,® Integral® (Biomet)Mallory-Head® (Biomet)Taperloc™ (Biomet)Omnifit® (Stryker)PCA® (Howmedica)Harris/Galante™ (Zimmer)PCA® (Howmedica)AML® (DePuy)Harris/Galante™ (Zimmer)AML® (DePuy)
YEARS FOLLOWED
5–13 years5–8 years8 –12.5 years5 years
6.5 years avg. follow-up7 years2–6 years5–7 years3.5 years avg. follow-up2–7 years2–7 years4.5 years avg. follow-up7.5 years avg. follow-up
OSTEOLYSIS
0.0% 0.0% 6.0% 0.4%
0.0% 3.0% 44.7% 18.0% 22.0% 37.0% 55.8% 31.0% 28.0%
P erformance proven
Plasma Spray Porous CoatingBiomet’s proprietary plasma spray application is unique in comparison to other techniques due to the fact that
only the titanium powder used to create the coating is heated, not the substrate of the implant – which can result in aweakened material. The porosity of the plasma spray porous surface allows for bone to penetrate and create a mechanical bondwith the porous coating. The pore size of Biomet’s plasma-sprayed components falls within a 100 to 1,000 micron range, thusproviding the optimal pore range for implant fixation and stability. The closed pore plasma porous spray has demonstratedexcellent clinical results since 1984.
B i-Metric
Total Hip
System
A Member of the
Alliance Family®
PP
he Bi-MetricTotal Hip System,
since its introduction,continues to provide a high
degree of versatility and unsurpassed clinicalperformance. After a 5 to 13 year follow-up onthe Bi-Metric Porous Primary,1 Evansand DeLee found:
• 0% Distal Thigh Pain • 0% Distal Osteolysis • 6% Stress Shielding • 100% Survivorship
The Bi-Metric is an essential part of theAlliance family which comprises the largestassembly of hip implants utilizing one simple,accurate and reproducible set of instruments.
Designed to off-load stresses gradually fromproximal to distal, all Bi-Metric componentsincorporate Biomet’s proven bi-planar tapergeometry. The stem geometry promotesincreased proximal off-loading and fills agreater portion of the metaphysis than acylindrical, parallel sided implant. This greatlyreduces the chances of proximal boneresorption and distal hypertrophy.
1Evans, JA, DeLee, JC, Outcome of a Tapered, Titanium, Proximal Load-Bearing, Non-Cemented Femoral THA Component: Submitted for Publication.
T
i-Metric
Primary
The Bi-Metric’s 3˚ tapered stem design (A) provides forexcellent proximal fit and minimizes removal of distalcortical bone as compared to a straight, cylindrical design (B).
Modular Head Options(Available in 22, 26, 28 and 32mm Diameters)
CoCrZirconia / CoCr
29mm Neck(-5)
31mm Neck(-3)
34mm Neck(STD)
CoCr
37mm Neck(+3)
40mm Neck(+6)
43mm Neck(+9)
46mm Neck(+12)
Zirconia/CoCr
Zirconia
28mm Neck(-6)
(A) (B)
B ■ Bi-Planar Stem TaperProvides enhanced proximal stressoff-loading and initial implant stability.
■ Titanium Alloy Plasma SprayPorous CoatingThe coating’s “closed pore” design actsas a potential barrier to the migrationof particulate debris and providesrotational stability and provenlong-term fixation.
■ Forged TitaniumEnsures biocompatibility and fatiguestrength; titanium displays a lowermodulus of elasticity for enhancedload transfer.
■ Available in both collaredand collarless designs.
■ Cemented option also available.
■ PMMA distal centralizer used to preventvarus or valgus stem positioning.
The Bi-Metric Porous Primary Hip Prosthesis is marketedfor noncemented use in skeletally mature patientsundergoing primary hip replacement surgery as a resultof noninflammatory degenerative joint disease.
Performance rovenP
Bi-Metric Head/Neck Components
i-Metric
Head/Neck
34mmResection
45mmResection
55mmResection
■ Bi-Planar Stem TaperProvides enhanced proximal stressoff-loading and initial implant stability.
■ Interlok® SurfaceInterlok finish provides enhancedcement-to-implant bonding.
■ Three ProximalResection LevelsProvide increased flexibilityto conform to and matchthe patient’s anatomy.• 34mm Resection Level – Primary resection level.• 45mm Resection Level – Bone deficiency at the level of the lesser trochanter.• 55mm Resection Level – Bone deficiency below the lesser trochanter.
■ PMMA Distal Stem Positioner(150mm and 200mm Lengths)Centralizes the stem in the distal canalto help prevent varus and valgusstem placement.
■ Available LengthsThe implant is offered in 150mm and200mm straight stems and in a 250mmbowed option for unmatched flexibility.
The Bi-Metric Head/Neck Prosthesis is marketedfor cemented application.
B
Performance rovenP200mmLength
250mmLength
150mmLength
i-Metric
Revision
Revision stems have a distal anterior bowto anatomically fit the femur.
B ■ Proximal Bi-Planar Stem TaperProvides enhanced proximal stressoff-loading and initial implant stability.
■ Titanium Alloy Plasma SprayPorous CoatingThe coating’s “closed pore” designacts as a potential barrier to the migrationof particulate debris and providesexceptional rotational stability and provenlong-term fixation.
■ Forged TitaniumEnsures biocompatibility and fatiguestrength; titanium displays a lowermodulus of elasticity for enhancedload transfer.
■ CollarProvides additional support and stresstransfer in the calcar region and aids inrotational stability.
■ Available LengthsThe implant is offered in 200mm,250mm, and 300mm options to addressexpanded indications.
Revision Stems are marketed for usewith bone cement in the United States.
Performance rovenP200mmLength
250mmLength
300mmLength
Rasp in-linewith femoralshaft
1Step Femoral Head
Resection
Once the hip is dislocated, the femoralneck cut can be made either by using thefemoral broach as a template (Fig. 1)or by using the femoral resectiontemplate (Fig. 2).
Fig. 1
Cut-linealong trialhead/broachjunction
Trial headcenter, onfemoral headcenter, in-linewith the tip ofthe greatertrochanter
2Step Reaming the
Distal Femur
A starter drill bit (Fig. 3) or a hollowchisel (Fig. 4) can be used to open thefemoral canal. Once the canal has beenlocated, begin reaming with the smallesttapered reamer. Enlarge the canal in asequential fashion until cortical “chatter”is encountered (Fig. 5).
Fig. 3 Fig. 4
Biomet, as the manufacturer of this device, does notpractice medicine and does not recommend anyparticular surgical technique for use on a specificpatient. The surgeon who performs any implantprocedure is responsible for determining and utilizingthe appropriate techniques for implanting theprosthesis in each individual patient. Biomet is notresponsible for selection of the appropriate surgicaltechnique to be utilized for an individual patient. Fig. 5
Fig. 2
i-Metric
Primary
Surgical
Technique
B
Broaching
the Proximal
Femur
Begin broaching with the broach that is 2or 3mm smaller than the last size reamer.It is important that the broach is orientedso that the medial/lateral axis of thebroach is parallel with the anatomicmedial/lateral axis of the femoral neck(Fig. 6). A sequentially larger broach isused until ideal or templated size isreached (Fig. 7). With the proper sizebroach in place, the calcar can be planedflush by using the calcar trimmer *(Fig. 8).
3Step
4StepTrial
Reduction
With the final broach still in place,provisional head/necks are selected todetermine the appropriate neck length andto restore the lateral offset (Fig. 9). A trialreduction is carried out to ensure thatproper leg length and joint stability havebeen achieved (Fig. 10).
* U.S. Pat. No. 4,306,550
Fig. 7
Fig. 9
Fig. 10
Fig. 8
Fig. 6
5aStep Inserting
a Porous
Component
The stem corresponding to the size of thefinal broach is threaded onto the steminserter/extractor and impacted into a fullyseated position (Fig. 11).
If desired, another trial reduction can beaccomplished after implantation of thestem and prior to impacting the modularhead onto the stem. Provisional heads inseven neck lengths allow trial reductionsto be performed, using the actual femoralcomponent, to again ensure proper leglength and stability.
After fully seating the femoral component,the appropriate modular head is impactedonto the femoral neck. The hip is nowready to be reduced.
5bStep
Fig. 11
Fig. 12
Inserting
an Interlok
Component
The reaming and broaching are accom-plished in the same fashion as a porouscomponent. The implant should be under-sized by a minimum of 2mm to allow foradequate cement thickness. Example: Reamand broach to 11mm. Implant a size9mm interlok stem. A distal cementrestrictor is placed in the canal to allow a2cm cement column below the tip of the stem(Fig. 12). Cement is injected into the canal ina retrograde fashion and pressurized. Slideappropriate sized distal centralizer ontostem. Example: A 9mm size stem acceptsa 9mm centralizer sleeve. The stem isinserted to a fully seated position andextraneous cement is removed. Once cementhardening has been achieved, a trial reduc-tion can be completed and the correctmodular head chosen for reconstitution ofleg length, lateral offset and stability.
Inserting
a Revision
Component
The revision components are an extendedversion of the primary stems utilizing thesame design criteria. The revision compo-nent uses the same rasp as the primarystem devices.
Distal femoral reaming for the 200mm,250mm and 300mm femoral componentsshould be accomplished with flexiblereamers over a guide wire to avoid distalcanal perforations (Fig. 13). The flexiblereamers should be used to over-ream thedistal medullary canal by approximately1mm to accommodate the distal anteriorbow in the prosthetic stems.
Once femoral preparation is completed,trial long stem components are availableif desired. The trial long stem is identicalto the final implant, minus theporous coating.
Final component insertion can then becarried out. The appropriate size stem isthen seated into the femoral canal.Example: Final broach size is 17mm.Implant a size 17mm revision compo-nent. (Fig. 14). Trial reduction can thenbe accomplished to determine properneck length.
5c
Fig. 14
Fig. 13
Step
1Step Preparing
the Proximal
Femur
Determination of Resection LevelThe proximal femoral cuts can be fashionedby utilizing the trial prosthesis as a guide(Fig. 1). The head/neck prosthesis comes in34, 45, and 55mm calcar resection levels.In general, the 34mm calcar is utilized insituations where there has been little, if any,bone destruction in the proximal femur anda platform can be maintained at approxi-mately 1cm above the lesser trochanter. The45mm design is utilized for bony defectsdown to the lower level of the lessertrochanter and the 55mm prosthesis fordefects below the lesser trochanter (Fig. 2).
Fig. 1
Fig. 2
i-Metric
Head/Neck
Surgical
Technique
B
2StepFemoral
Canal
Preparation
Determination of Stem LengthThe head/neck prosthesis comes in150mm, 200mm and 250mm lengths.Femoral preparation is accomplished byone of three methods after determiningstem length. Thorough debridement of thefemur and accurate preparation of theplatform is necessary for the prosthesis.For 150mm stems, tapered reamers areused to prepare the diaphyseal region(Fig. 3). Tapered reaming continuessequentially until cortical chatter and thereamer is at least 2.0cm below the finalimplant level. When using a 200mmstraight stem, cylindrical reamers should beutilized in the same manner as the taperedreamers to prepare the femoral canal.Flexible reamers are recommended whenusing a 250mm bowed implant due to theanterior bow of the femur. Note: Thedepth markings on the shank of thereamers DO NOT indicate leg lengthmeasurements when used with thehead/neck replacement components.Broaching of the proximal femur is notgenerally necessary, or even possible,when dealing with the compromised femur.If the proximal femur is not entirelycompromised, a broach can be used tofurther shape the proximal envelope(Fig. 4). Broaching is initiated in asequential fashion and continues until thecorresponding size of the last reameris reached.
Fig. 3
Fig. 4
Taperedreaming150mmlength
Cylindricalreaming200mmlength
Flexiblereaming250mmlength
3Step Head/Neck
Trial
Insertion and
Reduction
Trial InsertionThe trial prosthesis should be very stablewhen it is driven in place. Once the trial isin position, mark with methylene bluewhere the keel comes into contact withthe medial calcar (Fig. 5).
After marking the location for the keel,remove the trial stem and prepare agroove or recess in this same location onthe medial bone using a power bur(Fig. 6). After the groove has beenfashioned, reinsert the trial stem, makingsure it seats firmly on the horizontalfemoral platform.
Trial ReductionWith the trial calcar prosthesis in place(Fig. 7), a trial reduction can be per-formed to evaluate proper leg length andlateral offset. Note: A trial reductioncannot be performed with thebroach. This will not provide accu-rate leg length and offset measure-ments. Modular trial prosthesesare available for 200mm and250mm stems, while one-piecetrials are used for 150mm stems.
Fig. 6
Fig. 5
Fig. 7
4StepHead/Neck
Component
Insertion
A distal cement restrictor is placed in thecanal to allow a 2.0cm cement columnbelow the tip of the stem (for 150mmlength stems). Cement is injected into thecanal in a retrograde fashion and pressur-ized. Note: When using the Bi-Metrichead/neck implants, one mustdownsize the implant to allow for anadequate cement mantle. Example:Ream and broach to 13mm. Implantan 11mm head/neck stem. The stemcan be inserted with the forked inserter(Fig. 8). Place medial keel into medialcalcar slot. Extraneous cement is thenremoved. Once cement hardening hasbeen achieved, a trial reduction can becompleted to access the correct modularhead, proper leg length, lateral offsetand stability. The appropriate modularhead can then be impacted onto thestem (Fig. 9).
Fig. 9
Fig. 8
i-Metric
Parts ListingB Bi-Metric Porous Prosthesis
162310162251162311162252162312162253162313162254162314162255162315
Collarless
Part No.
Collared
Part No.Stem Diameter
StemLength mm
7.0mm 8.0mm 9.0mm10.0mm11.0mm12.0mm13.0mm14.0mm15.0mm16.0mm17.0mm
115120125130135140145150155160165
162370162626162371162627162372162628162373162629162374162630162375
162301162302162303162304162305162306
Bi-Metric Interlok® Prosthesis
Part No.
Collarless
Part No.
CollaredStem Diameter
StemLength mm
7.0mm 9.0mm11.0mm13.0mm15.0mm17.0mm
115125135145155165
162387162388162389162390162391162392
PMMA Centering Sleeves
162907162909162911162913162915162917
Part No. Size
7.0mm 9.0mm11.0mm13.0mm15.0mm17.0mm
162350162351162352162353162354162355162356162357162358162359
162340162341162342162343162344162345162346162347162348162349
Porous Bi-Metric Collared Long Stem Prosthesis
Part No.
Stem Diameter
9.0mm Stem Bowed Right 9.0mm Stem Bowed Left11.0mm Stem Bowed Right11.0mm Stem Bowed Left13.0mm Stem Bowed Right13.0mm Stem Bowed Left15.0mm Stem Bowed Right15.0mm Stem Bowed Left17.0mm Stem Bowed Right17.0mm Stem Bowed Left
31-47364031-47364131-47364231-47364331-47364431-47364531-47364631-47364731-47364831-473649
31-47365031-47365131-47365231-47365331-47365431-47365531-47365631-47365731-47365831-473659
200mmLength
Part No.
200mmLength
Provisional
Part No. Part No. Part No.
250mmLength
250mmLength
Provisional
300mmLength
Description
162360162361162362162363162364162365162366162367162368162369
31-16233031-16233131-16233231-16233331-16233431-16233531-16233631-16233731-16233831-16233931-16234031-162341
Bi-Metric CoCr Head/Neck ProvisionalProximal – 200 and 250mm
Part No.Stem
DiameterResection
Level 11.0mm 11.0mm 11.0mm 13.0mm 13.0mm 13.0mm 15.0mm 15.0mm 15.0mm 17.0mm 17.0mm 17.0mm
344555344555344555344555
11.0mm 13.0mm 15.0mm 17.0mm 11.0mm 13.0mm 15.0mm 17.0mm
31-16234231-16234331-16234431-16234531-16234631-16234731-16234831-162349
Bi-Metric CoCr Head/Neck ProvisionalDistal – 200 and 250mm
Part No.Stem
DiameterLength
mm
200200200200250250250250
12-16258012-16258112-16258212-16258312-16258412-16258512-16257412-162575
12-16258612-16258712-16258812-16258912-16259012-16259112-16257612-162577
12-16259212-16259312-16259412-16259512-16259612-16259712-16257812-162579
250 Rt250 Lt250 Rt250 Lt250 Rt250 Lt250 Rt250 Lt
250 Rt250 Lt250 Rt250 Lt250 Rt250 Lt250 Rt250 Lt
250 Rt250 Lt250 Rt250 Lt250 Rt250 Lt250 Rt250 Lt
Lengthmm
Bi-Metric CoCr Head/Neck Prosthesis250mm
Part No. StemDiameter
ResectionLevel
9.0mm 9.0mm 11.0mm 11.0mm 13.0mm 13.0mm 15.0mm 15.0mm
9.0mm 9.0mm 11.0mm 11.0mm 13.0mm 13.0mm 15.0mm 15.0mm
9.0mm 9.0mm 11.0mm 11.0mm 13.0mm 13.0mm 15.0mm 15.0mm
3434343434343434
4545454545454545
5555555555555555
12-16238012-16238112-16238212-16238312-16238412-16238512-16238612-16238712-162388
Bi-Metric CoCr Head/Neck Prosthesis150mm
Part No.Stem
DiameterLength
mmResection
Level
9.0mm 11.0mm 13.0mm 9.0mm 11.0mm 13.0mm 9.0mm 11.0mm 13.0mm
150150150150150150150150150
343434454545555555
StemDiameter
12-16248012-16248112-16248212-16248912-16248312-16248412-16248512-16249012-16248612-16248712-16248812-162491
Bi-Metric CoCr Head/Neck Prosthesis200mm
Part No.Length
mmResection
Level 9.0mm 11.0mm 13.0mm 15.0mm 9.0mm 11.0mm 13.0mm 15.0mm 9.0mm 11.0mm 13.0mm 15.0mm
200200200200200200200200200200200200
343434344545454555555555
31-16238031-16238131-16238231-16238931-16238331-16238431-16238531-16239031-16238631-16238731-16238831-162391
Bi-Metric CoCr Head/Neck Provisional150mm
Part No.Stem
DiameterLength
mmResection
Level 9.0mm 11.0mm 13.0mm 15.0mm 9.0mm 11.0mm 13.0mm 15.0mm 9.0mm 11.0mm 13.0mm 15.0mm
150150150150150150150150150150150150
343434344545454555555555
P.O. Box 587, Warsaw, IN 46581-0587 • 219.267.6639 • ©1998 Biomet, Inc. All Rights Reservedweb site: http://www.biomet.com • eMail: [email protected]
Form No. Y-BMT-567/073198/M
Alliance Instrumentation
CASE #595551 – Insertion/Extraction
CASE #595552 – Alliance Tapered Reamers
CASE #595553 – Alliance Broaches
Bi-Metric X-Ray Templates
162396 Bi-Metric Primary12-162397 Bi-Metric Head/Neck162395 Bi-Metric Revision
CASE # 595551 – Insertion/Extraction
31-473622 Resection Guide466365 Drill Bit31-473625 Osteotomy Guide31-473679 Moore Hollow Chisel31-473590 Femoral Inserter with Fork31-473639 Femoral Inserter with Neck Control31-473621 Slap Hammer with Thread Tip31-476946 Universal Head Driver424456 Universal Head Extractor
CASE #595552 – Alliance Tapered Reamers
31-473200 7mm31-473201 8mm31-473202 9mm31-473203 10mm31-473204 11mm31-473205 12mm31-473206 13mm31-473207 14mm
CASE #595553 – Alliance Broaches
31-473611 7 x 115mm31-473670 8 x 120mm31-473612 9 x 125mm31-473671 10 x 130mm31-473613 11 x 135mm31-473672 12 x 140mm31-473614 13 x 145mm31-473673 14 x 150mm31-473615 15 x 155mm31-473674 16 x 160mm31-473616 17 x 165mm31-473675 18 x 170mm31-473617 19 x 175mm
31-473610 Broach Handle31-473593 Broach Extractor31-473701 Calcar Trimmer
31-473208 15mm31-473209 16mm31-473210 17mm31-473211 18mm31-473212 19mm31-473213 20mm
31-483630 T-Handle
B i-Metric
Total Hip System
®
proven
erformanceP