9
Christopher B. Arena 1 Bioelectromechanical Systems Lab, Virginia Tech-Wake Forest School of Biomedical Engineering and Sciences, Virginia Tech, 330 Kelly Hall (MC0298), Stanger Street, Blacksburg, VA 24061 e-mail: [email protected] Roop L. Mahajan Institute for Critical Technology and Applied Science (ICTAS), Virginia Tech Department of Mechanical Engineering, Virginia Tech Department of Engineering Science and Mechanics, Virginia Tech, 410 H Kelly Hall (MC0298), Stanger Street, Blacksburg, VA 24061 Marissa Nichole Rylander Tissue Engineering Nanotechnology and Cancer Research Lab, Virginia Tech-Wake Forest School of Biomedical Engineering and Sciences, Virginia Tech Department of Mechanical Engineering, Virginia Tech, 335 Kelly Hall (MC0298), Stanger Street, Blacksburg, VA 24061 Rafael V. Davalos Bioelectromechanical Systems Lab, Virginia Tech-Wake Forest School of Biomedical Engineering and Sciences, Virginia Tech, 329 Kelly Hall (MC0298), Stanger Street, Blacksburg, VA 24061 An Experimental and Numerical Investigation of Phase Change Electrodes for Therapeutic Irreversible Electroporation Irreversible electroporation (IRE) is a new technology for ablating aberrant tissue that utilizes pulsed electric fields (PEFs) to kill cells by destabilizing their plasma membrane. When treatments are planned correctly, the pulse parameters and location of the electro- des for delivering the pulses are selected to permit destruction of the target tissue without causing thermal damage to the surrounding structures. This allows for the treatment of surgically inoperable masses that are located near major blood vessels and nerves. In select cases of high-dose IRE, where a large ablation volume is desired without increasing the number of electrode insertions, it can become challenging to design a pulse protocol that is inherently nonthermal. To solve this problem we have developed a new electrosur- gical device that requires no external equipment or protocol modifications. The design incorporates a phase change material (PCM) into the electrode core that melts during treatment and absorbs heat out of the surrounding tissue. Here, this idea is reduced to practice by testing hollow electrodes filled with gallium on tissue phantoms and monitor- ing temperature in real time. Additionally, the experimental data generated are used to validate a numerical model of the heat transfer problem, which is then applied to investi- gate the cooling performance of other classes of PCMs. The results indicate that metallic PCMs, such as gallium, are better suited than organics or salt hydrates for thermal man- agement, because their comparatively higher thermal conductivity aids in heat dissipation. However, the melting point of the metallic PCM must be properly adjusted to ensure that the phase transition is not completed before the end of treatment. When translated clini- cally, phase change electrodes have the potential to continue to allow IRE to be performed safely near critical structures, even in high-dose cases. [DOI: 10.1115/1.4025334] Keywords: nonthermal ablation, irreversible electroporation, electrochemotherapy, phase change materials, latent heat storage, Joule heating, thermal damage, radio fre- quency ablation Introduction Reversible electroporation has been used over the past two dec- ades to facilitate the transport of molecules across cell membranes without directly compromising cell viability [13]. Additionally, irreversible electroporation (IRE) has emerged as an effective focal ablation technique that disrupts cell membranes beyond the point of recovery [4]. When performed clinically, these proce- dures involve placing electrodes into, or around, a target tissue and applying a series of short, but intense, pulsed electric fields (PEFs). Oftentimes, patient specific treatment plans are employed to guide procedures by merging medical imaging with algorithms for determining the electric field distribution in the tissue [57]. The electric field dictates treatment outcomes by increasing a cell’s transmembrane potential to levels where it becomes ener- getically favorable for the membrane to shift to a state of enhanced permeability [8]. Successful cancer treatments have been designed that combine reversible electroporation with the delivery of chemotherapeutic agents [9] or plasmid DNA [10] in the form of electrochemotherapy (ECT) and electrogenetherapy (EGT), respectively. The same is true for IRE, without the requirement for adjuvant molecules [11]. PEF protocols for tissue electroporation are designed to limit Joule heating. That is, the number of pulses, pulse duration, pulse amplitude, and pulse repetition rate are chosen in combinations that can induce electroporation without raising the tissue tempera- ture to levels capable of causing thermal damage. ECT and IRE typically involve multiple, high voltage (1000 V) pulses with durations on the order of hundreds of microseconds, whereas EGT protocols also include lower voltage (100 V) pulses with dura- tions on the order of tens of milliseconds to facilitate electropho- resis of DNA [12]. Because the mechanism of action is nonthermal, treatments can be performed in close proximity to major blood vessels without concern for heat sink effects that can protect tumors from thermally mediated therapies. Additionally, the nonthermal mode of ablation in IRE spares extracellular ma- trix components, including major nerve [13] and blood vessel [14] architecture, which promotes rapid recovery following treatment [15] and allows for the treatment of several surgically inoperable tumors [6,7]. Situations can arise in which it is challenging to design an elec- troporation protocol that simultaneously covers the targeted tissue with a sufficient electric field and avoids unwanted thermal effects. For instance, hallmarks of thermal damage have been seen adjacent to the electrodes in select cases of high-dose IRE, where the applied voltage was raised (3000 V) to ablate a large volume in a single treatment [16]. This is presumably due to the steep electric potential gradient that occurs in the vicinity of the elec- trode edges. Similarly, when treating subcutaneous lesions, there 1 Corresponding author. Contributed by the Bioengineering Division of ASME for publication in the JOURNAL OF BIOMECHANICAL ENGINEERING. Manuscript received April 10, 2013; final manuscript received July 30, 2013; accepted manuscript posted September 6, 2013; published online October 3, 2013. Assoc. Editor: Ram Devireddy. Journal of Biomechanical Engineering NOVEMBER 2013, Vol. 135 / 111009-1 Copyright V C 2013 by ASME Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

Christopher B. Arena1

Bioelectromechanical Systems Lab,

Virginia Tech-Wake Forest School of Biomedical

Engineering and Sciences,

Virginia Tech, 330 Kelly Hall (MC0298),

Stanger Street,

Blacksburg, VA 24061

e-mail: [email protected]

Roop L. MahajanInstitute for Critical Technology and

Applied Science (ICTAS),

Virginia Tech Department of

Mechanical Engineering,

Virginia Tech Department of Engineering

Science and Mechanics,

Virginia Tech, 410 H Kelly Hall (MC0298),

Stanger Street,

Blacksburg, VA 24061

Marissa Nichole RylanderTissue Engineering Nanotechnology and

Cancer Research Lab,

Virginia Tech-Wake Forest

School of Biomedical Engineering and Sciences,

Virginia Tech Department of

Mechanical Engineering,

Virginia Tech, 335 Kelly Hall (MC0298),

Stanger Street,

Blacksburg, VA 24061

Rafael V. DavalosBioelectromechanical Systems Lab,

Virginia Tech-Wake Forest School of

Biomedical Engineering and Sciences,

Virginia Tech, 329 Kelly Hall (MC0298),

Stanger Street,

Blacksburg, VA 24061

An Experimental and NumericalInvestigation of Phase ChangeElectrodes for TherapeuticIrreversible ElectroporationIrreversible electroporation (IRE) is a new technology for ablating aberrant tissue thatutilizes pulsed electric fields (PEFs) to kill cells by destabilizing their plasma membrane.When treatments are planned correctly, the pulse parameters and location of the electro-des for delivering the pulses are selected to permit destruction of the target tissue withoutcausing thermal damage to the surrounding structures. This allows for the treatment ofsurgically inoperable masses that are located near major blood vessels and nerves. Inselect cases of high-dose IRE, where a large ablation volume is desired without increasingthe number of electrode insertions, it can become challenging to design a pulse protocolthat is inherently nonthermal. To solve this problem we have developed a new electrosur-gical device that requires no external equipment or protocol modifications. The designincorporates a phase change material (PCM) into the electrode core that melts duringtreatment and absorbs heat out of the surrounding tissue. Here, this idea is reduced topractice by testing hollow electrodes filled with gallium on tissue phantoms and monitor-ing temperature in real time. Additionally, the experimental data generated are used tovalidate a numerical model of the heat transfer problem, which is then applied to investi-gate the cooling performance of other classes of PCMs. The results indicate that metallicPCMs, such as gallium, are better suited than organics or salt hydrates for thermal man-agement, because their comparatively higher thermal conductivity aids in heat dissipation.However, the melting point of the metallic PCM must be properly adjusted to ensure thatthe phase transition is not completed before the end of treatment. When translated clini-cally, phase change electrodes have the potential to continue to allow IRE to be performedsafely near critical structures, even in high-dose cases. [DOI: 10.1115/1.4025334]

Keywords: nonthermal ablation, irreversible electroporation, electrochemotherapy,phase change materials, latent heat storage, Joule heating, thermal damage, radio fre-quency ablation

Introduction

Reversible electroporation has been used over the past two dec-ades to facilitate the transport of molecules across cell membraneswithout directly compromising cell viability [1–3]. Additionally,irreversible electroporation (IRE) has emerged as an effectivefocal ablation technique that disrupts cell membranes beyond thepoint of recovery [4]. When performed clinically, these proce-dures involve placing electrodes into, or around, a target tissueand applying a series of short, but intense, pulsed electric fields(PEFs). Oftentimes, patient specific treatment plans are employedto guide procedures by merging medical imaging with algorithmsfor determining the electric field distribution in the tissue [5–7].The electric field dictates treatment outcomes by increasing acell’s transmembrane potential to levels where it becomes ener-getically favorable for the membrane to shift to a state ofenhanced permeability [8]. Successful cancer treatments havebeen designed that combine reversible electroporation with thedelivery of chemotherapeutic agents [9] or plasmid DNA [10] inthe form of electrochemotherapy (ECT) and electrogenetherapy(EGT), respectively. The same is true for IRE, without therequirement for adjuvant molecules [11].

PEF protocols for tissue electroporation are designed to limitJoule heating. That is, the number of pulses, pulse duration, pulseamplitude, and pulse repetition rate are chosen in combinationsthat can induce electroporation without raising the tissue tempera-ture to levels capable of causing thermal damage. ECT and IREtypically involve multiple, high voltage (�1000 V) pulses withdurations on the order of hundreds of microseconds, whereas EGTprotocols also include lower voltage (�100 V) pulses with dura-tions on the order of tens of milliseconds to facilitate electropho-resis of DNA [12]. Because the mechanism of action isnonthermal, treatments can be performed in close proximity tomajor blood vessels without concern for heat sink effects that canprotect tumors from thermally mediated therapies. Additionally,the nonthermal mode of ablation in IRE spares extracellular ma-trix components, including major nerve [13] and blood vessel [14]architecture, which promotes rapid recovery following treatment[15] and allows for the treatment of several surgically inoperabletumors [6,7].

Situations can arise in which it is challenging to design an elec-troporation protocol that simultaneously covers the targeted tissuewith a sufficient electric field and avoids unwanted thermaleffects. For instance, hallmarks of thermal damage have been seenadjacent to the electrodes in select cases of high-dose IRE, wherethe applied voltage was raised (�3000 V) to ablate a large volumein a single treatment [16]. This is presumably due to the steepelectric potential gradient that occurs in the vicinity of the elec-trode edges. Similarly, when treating subcutaneous lesions, there

1Corresponding author.Contributed by the Bioengineering Division of ASME for publication in the

JOURNAL OF BIOMECHANICAL ENGINEERING. Manuscript received April 10, 2013; finalmanuscript received July 30, 2013; accepted manuscript posted September 6, 2013;published online October 3, 2013. Assoc. Editor: Ram Devireddy.

Journal of Biomechanical Engineering NOVEMBER 2013, Vol. 135 / 111009-1Copyright VC 2013 by ASME

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 2: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

is a high probability for thermal damage in the overlying skinlayers [17]. The skin acts as a resistive barrier that effectivelyshields the underlying tissue from experiencing electroporation.As a result, the skin is exposed to a comparatively higher electricfield, and the applied voltage must be increased in this case aswell to ensure proper treatment of the underlying tissue.

In our previous work, we theoretically showed that phasechange materials (PCM) could be incorporated into electrodedesigns to reduce the probability for thermal damage adjacent toelectrode boundaries during large IRE ablations of the liver [18].The specific PCM investigated was disodium hydrogen phosphatedodecahydrate, which is a salt hydrate that stores a large amountof latent heat upon melting during treatment narrowly above phys-iologic temperature. To the best of our knowledge, this was thefirst time a numerical model of the phase change problem wasdeveloped to study electroporation devices. Up to this point,PCMs have been primarily used in the biomedical field in storagedevices to protect temperature sensitive substances during trans-port [19], in clothing to protect the human body from extremetemperatures [20], and in contrast agents for image enhancement[21,22]. Outside the biomedical field, they are slowly becomingcommonplace in industrial applications of thermal protection ofelectronic devices [23,24], green energy storage [25], and vehicleand building temperature regulation [26].

Here, we expand upon our previous theoretical results by exper-imentally investigating the use of PCMs in electroporation devi-ces. In this first reduction to practice, electrodes were fabricatedwith a gallium core and pulses were applied to ultrasound gel,which served as a tissue phantom [27]. For comparison, solidmetal electrodes were also fabricated and tested. Temperature wasmonitored in real time, and numerical models were constructed tomatch the data generated for each electrode type. The modelswere identical to those presented in [18], with the exception thatno blood perfusion or metabolic processes were present in the tis-sue phantom. A parametric study was conducted on the experi-mentally validated models to elucidate the effect of varyingelectrode radius, electrode spacing, PCM melting point, and PCMthermal conductivity on temperature mitigation. A major findingis that with proper tuning of the PCM melting point, metallicPCMs possessing a comparatively higher thermal conductivitythan salt hydrates are better suited as an electrode core material.

Methods

Experimental Setup. To fabricate the PCM electrodes, roundbrass tubing (4.0 mm diameter, 0.4 mm thickness, K&S PrecisionMetals, Chicago, IL, USA) was machined to length (2.46 cm), andhigh voltage wire was soldered to one end (22 AWG, Daburn,Dover, NJ, USA) to simultaneously plug the tubing and form con-nections to the pulse generator (ECM 830, BTX Harvard Appara-tus, Holliston, MA, USA). Then, solid gallium (99.99% pure,GalliumSouce, Scotts Valley, CA, USA) was liquefied above itsmelting point (29.8 �C) [28] in de-ionized water and injected intothe remaining open end of the hollow tubing. The gallium wasallowed to re-solidify in the electrode core at room temperature[Fig. 1(a)]. Solid metal electrodes were fabricated in a similar pro-cess using brass rods of equivalent dimensions.

Electrodes were spaced 0.43 cm edge-to-edge by drilling holesinto a printed circuit board. The printed circuit board also pro-vided a convenient port for a fiber optic temperature probe (Lux-tron m3300, LumaSense, Santa Clara, CA, USA), which waslocated �2.0 mm in the x and y direction relative to the center ofthe leftmost electrode. Real-time temperature measurements weretaken every 0.25 s to validate the numerical model describedbelow. The entire setup was placed on the surface of a 100 mlglass beaker that was filled with ultrasound transmission gel(Mojility agent, MOtion1, Nashville, TN, USA), such that theelectrodes extended �1.93 cm into the ultrasound gel in the zdirection. As shown by the output of the numerical model (seeFig. 3), the height (7.05 cm) and diameter (5.03 cm) of the beakerwere large enough to avoid edge effects. Etheridge et al. havecharacterized the thermal properties of ultrasound gel and, in theprocess, shown that it is a viable tissue phantom material [27].The phosphorescent sensor at the end of the temperature probeextended �1.80 cm in the z direction from the surface of the ultra-sound gel. All experiments were conducted inside an incubator setto 25 �C. This ensured that temperatures reached during treatmentsurpassed the melting point of gallium.

The pulse protocol consisted of ninety-nine 2 ms pulses deliv-ered at a rate of 1 pulse per second. The applied voltage was set to500 V, which was the maximum voltage at which the pulse gener-ator could operate in low voltage mode. This produced a higherquality rectangular waveform and lowered the likelihood of arcing

Fig. 1 (a) Image of the experimental setup used to monitor temperature during electroporationwith phase change electrodes. The electrodes and fiber optic temperature probe are spacedusing a printed circuit board, inserted into the tissue phantom, and placed inside an incubator(white arrow). The insert displays the hollow brass electrodes with the inclusion of a galliumcore (black arrow). (b) Representative mesh of the finite element model used to simulate theelectroporation protocol with 79,198 elements.

111009-2 / Vol. 135, NOVEMBER 2013 Transactions of the ASME

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 3: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

in the ultrasound gel. At this voltage-to-distance ratio, the rela-tively long 2 ms pulses that are characteristic of EGT procedureswere required to generate temperatures capable of initiating thegallium phase transition. A total of six beakers filled with ultra-sound gel were placed in the incubator 6 h prior to treatment inorder to allow time for equilibration with the surrounding environ-ment. Treatments with the solid electrodes (n¼ 3) and PCM elec-trodes (n¼ 3) were conducted at random, and a 10 min delayperiod was included between trials to allow the electrodes toreturn to their baseline temperature.

Simulation of the Experimental Setup. A 3D finite elementmodel was developed using COMSOL Multiphysics 4.2 a (Stock-holm, Sweden) to simulate the electric field and temperature dis-tributions throughout the experimental procedure (Fig. 1(b)). Thecore material of the electrodes was varied between the proposedgallium PCM and a solid metal equivalent to the brass shell. Theuse of the glass beaker to contain the ultrasound gel was replacedwith insulative boundary conditions to reduce the total number ofmesh elements and speed computation time. Specifically, a finer,free tetrahedral mesh was used, which resulted in 79,198 elementsand less than 0.5% difference in temperature calculations at thelocation of the fiber optic probe upon successive refinements.Replicating the 150 s experimental treatments (99 s of pulsing plusan additional 51 s to observe cooling) required approximately 20min of computational time. This relatively long time is due to thesmall time step (0.05 s) and low tolerance (0.001 relative, 0.0001absolute) that was chosen in order to accurately resolve the phasetransition in the PCM electrodes without oscillations in the pre-dicted temperature. All simulations were performed on a Dell Pre-cision T3500 Workstation (3.20 GHz, 12 GB Ram).

To account for the solid (s)–liquid (l) phase transition in thePCM domain, the effective heat capacity method was implementedby including the latent heat of fusion k (80.2 kJ/kg) [28] in the vol-ume average of the specific heat capacity at constant pressure

cp;eq ¼ Hðcp;s þ kPÞ þ ð1�HÞðcp;l þ kPÞ (1)

where H is the volume fraction of solid material defined by 1– flc2hs [(T� TM),TR] and TM is the melting temperature over thetransition range TR. This Heaviside function approximates the log-ical step H¼ 1 for T< TM and H¼ 0 for T> TM by smoothing thetransition within the interval TM – TR<T< TMþ TR using a con-tinuous second derivative. Smoothing is required for the strictsolver to handle step changes [29,30]. The term P is a normalizedpulse around the transition temperature defined by differentiatingflc2hs [(T�TM),TR] with respect to temperature. The temperature(T) distribution was solved using a modified heat conductionequation that includes Joule heating (QJ) due to the application ofPEFs

qcp

� �eq

@T

@t¼ r � ðkeqrTÞ þ QJ (2)

where qeq and keq are volume averages of the density and thermalconductivity, respectively

qeq ¼ Hqs þ ð1�HÞql (3)

keq ¼ Hks þ ð1�HÞkl (4)

Outside the PCM domain, the equations for equivalent specificheat capacity [Eq. (1)], density [Eq. (3)], and thermal conductivity[Eq. (4)] were replaced with constant values to represent the brassshell and tissue phantom domains. The thermal conductivity andspecific heat of the ultrasound gel (Table 1) were taken frompiecewise fitting equations [27] at 25 �C to two significant figures.The density of the ultrasound gel (Table 1) was taken to be the

manufacture’s value reported in [27]. In all domains, the Jouleheating term

QJ ¼ r Ej j2 (5)

was obtained by first solving for the spatial distribution of electricpotential (U) under the electroquasistatic approximation:

r � r -rUð Þ½ � ¼ 0 (6)

where r is the electrical conductivity, and the electric field (E) isdefined as -rU. This describes the stationary field present duringthe constant voltage pulses. The electrical conductivity of theultrasound gel (Table 1) was measured at 25 �C using a conductiv-ity meter (B-173, Horiba, Kyoto, Japan). The Joule heating termwas scaled by the duty cycle (pulse duration/period betweenpulses¼ 2� 10�3) to ensure that the total energy depositionmatched the experimental protocol (2 ms pulse once every sec-ond). Using the duty cycle approach, as opposed to simulatingeach individual pulse, eases the computational burden and allevi-ates the need for an adaptive time stepping to resolve microsecondand millisecond order pulses within a treatment lasting severalseconds. Additionally, it has been validated as providing an accu-rate prediction of the temperature distribution [31].

Electric potential boundary conditions of 486 V (output aver-age) and 0 V were applied along the outer surfaces of the ener-gized and grounded electrode shells, respectively, and the initialvoltage was set to 0 V in all domains. Additionally, all outer boun-daries were treated as electrically insulating (� n � J¼ 0). In termsof the thermal initial and boundary conditions, the initial tempera-ture T0 was set to 25.9 �C (baseline average) in all domains and inthe exterior surroundings Text, and the bottom and vertical bounda-ries of the tissue phantom domain were set as insulation [�n(�krT)¼ 0]. Treating the upper surface of the tissue phantom andelectrodes as the upside of a horizontal plate, and the sides of theelectrodes as vertical surfaces, the temperature was calculated usingbuoyancy induced convection [�n (�krT)¼ h(Text�T)]. For eachcase, the average heat transfer coefficient (h) was determined from

h ¼ kNu

L(7)

where Nu is the Nusselt number and L is a characteristic lengthscale. The length scale for the vertical case is the height of theexposed electrode surface, and the length scale for each horizontalcase is the respective ratio of area/perimeter. The average Nussletnumber for each case can be found in [33], and all material prop-erties are evaluated at [(Tþ Text)/2].

A parametric study was performed on the transition range ofthe gallium to optimize the fit with the experimental results. TR

was varied (0.05, 0.1, 0.5, 1, and 1.5 �C), and an overall regressioncoefficient was determined according to [27]

R2 ¼ 1�

X150

t¼0

Texp � Tnum

� �2

X150

t¼0

Texp � Texp

� �2

266664

377775

(8)

Table 1 PCM, electrode, and tissue phantom properties from[27],a [28],b and [32]c

Property Gallium (s/l) Brass Ultrasound gel

k (W m�1 K�1) 40.6/28.1b 109 0.52c

cp (J kg�1 K�1) 376.8/343.3a 377 3800c

q (kg m�3) 5904/6093a 8520 1020c

r (S m�1) 3.7 � 106 a 1.7 � 107 0.13

Journal of Biomechanical Engineering NOVEMBER 2013, Vol. 135 / 111009-3

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 4: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

where Texp and Tnum are the experimental temperatures and nu-merical temperatures, respectively, for a given time at the locationof the fiber optic probe. Texp is the average experimental tempera-ture over all times at the location of the fiber optic probe. Experi-mental and numerical temperatures were taken at 1 s intervals for150 s.

Effects of Varying Electrode Properties. The experimentallyvalidated model was used to investigate a clinical IRE scenariowith varying electrode properties. A parametric study was per-formed on electrode radius (0.5, 0.75, 1.0, and 1.25 mm), andPCM melting point (TM� T0¼ 1, 4, 7, and 10 �C). The range ofapplicable electrode radii spans from commercially available IREprobes (NanoKnife Single Electrode, AngioDynamics, Queens-bury, NY, USA) at the lower end to biopsy needles at the upperend [34]. While the variation in PCM melting point may seemidealized, it is physically possible to raise/lower the melting pointof gallium through alloying processes [35]. To maximize the vol-ume of PCM inside the electrode core, the shell thickness wasreduced to 0.25 mm, which is similar to certain hollow dispensingneedles (Howard Electronic Instruments Inc., El Dorado, KS,USA). An IRE protocol reported in [16] was chosen here to inves-tigate whether PCMs could mitigate temperature increase duringthese type of large liver ablations. Specifically, the applied voltagewas increased to 3000 V across an electrode spacing of 1.5 cm(edge-to-edge). The edge-to-edge spacing was held constant forall electrode radii to ensure equal voltage-to-distance ratios(2000 V/cm). The pulse duration and the total time of pulsingwere reduced to 90 ls and 70 s, respectively, and the electricalconductivity of the ultrasound gel was maintained at 0.13 S/m,which matches that of liver tissue at 25 �C [36]. Following the ini-tial parametric study, simulations on reduced electrode spacings(1.5, 1.0, and 0.5 cm) more comparable to the experimental setupwere also performed to test the limit of PCM effectiveness andevaluate changes in the spatial temperature distribution. In each ofthese cases, the applied voltage was varied to maintain a constantvoltage-to-distance ratio of 2000 V/cm.

The parametric study on electrode radius, spacing, and meltingpoint was repeated for PCM thermal conductivities one and twoorders of magnitude lower than gallium in both the solid and liquidphases. This simulation was designed to conceptualize alternateclasses of PCMs, including salt hydrates (keq/10) and organics (keq/100). All other thermal properties were held constant. This ensuredan equivalent heat of fusion per unit volume and heat capacity perunit volume between all trials. Interestingly, if the thermal conduc-tivity is also held constant, varying the PCM density, heat of fusion,and heat capacity does not affect the results as long as the productsof heat of fusion with density and heat capacity with density remainthe same. This is because the product (qcp)eq on the left-hand sideof Eq. (2) is unchanged. Salt hydrates and organics with similarproducts to gallium are commercially available [37].

Results

Experimental Results and Numerical Model Validation. Acomparison between the experimentally measured temperaturesand the numerically predicted temperatures is shown in Fig. 2.The experimental data are presented as the average of n¼ 3 trialsby the dotted lines, and sample standard deviations are given at 5 sintervals. In the case of the solid electrodes, the temperature at thelocation of the fiber optic probe increased over 11 �C above thebaseline during treatment. The temperature increase was limitedto approximately 4 �C in the case of the PCM electrodes, which isover a twofold reduction in peak temperature when compared tothe solid electrodes. Despite this significant reduction in tempera-ture, the gallium core remained fixed within the brass shell, whichsuggests that the phase transition process is in its early stages.

In the case of the solid electrodes, the numerical model (solidlines) accurately predicts the experimentally measured temperatures

within a single standard deviation at a majority of time points byusing analytically derived boundary conditions and material prop-erties taken from the literature (Fig. 2). The only unknown vari-able for the PCM electrodes was the transition range of gallium.The results of the parametric study performed on this variableindicate that the highest correlation between measured and pre-dicted temperatures occurs when TR was set to 0.5 �C (R2¼ 0.74),which results in a 1 �C transition interval (TM 6 TR). Lower valuesof TR, or more abrupt transitions, overestimated the measuredtemperature at the onset of the phase transition, whereas highervalues of TR, or more gradual transitions, underestimated themeasured temperature at the onset of the phase transition. The op-posite was true during the return to baseline following treatment.As in the case of the solid electrodes, the predicted temperaturesproduced by the optimized gallium transition range were in agree-ment with measured values from the PCM electrodes within a sin-gle standard deviation at a majority of time points. A plot of thevolume fraction of solid gallium within one of the electrodes (Fig.2, insert) indicates that there was preferential heating/melting onthe deep-medial portion of the electrodes. It also suggests that thephase transition in these regions is complete, while a majority ofthe superficial gallium remains solidified ensuring containmentwithin the brass shell.

Solutions from the validated numerical model for the electricfield and temperature distributions at the end of treatment areshown in Fig. 3. There was no electrical conductivity dependenceon temperature included in the model, and the electric field distri-butions for both scenarios are identical [Fig. 3(a)]. As a generalrule of thumb, any regions of tissue exposed to an electric fieldthreshold of 500 V/cm or greater are subject to IRE [38,39]. Acomparison of the volume of tissue subject to temperature changesindicative of thermal damage versus the volume of tissue subjectto IRE is presented following the parametric study on electrodespacing (see Table 2). For the solid electrodes [Fig. 3(b)], a maxi-mum temperature of 47.9 �C occurred at the center of the tissuedomain between the electrodes. This was a 22 �C increase abovethe baseline temperature. By incorporating a PCM core, the maxi-mum temperature reached between the electrodes was only41.2 �C. Therefore, the change in temperature was reduced byapproximately 7 �C. This reduction would likely become morepronounced during longer protocols, or shorter protocols at highervoltage, until the gallium completely enters the liquid state.

Effects of Varying Electrode Properties. Figure 4 shows tem-perature profiles resulting from the parametric study on electroderadius and metallic PCM melting temperature during the clinicalIRE scenario (1.5 cm electrode spacing). The temperature data

Fig. 2 Comparison of the measured temperatures (dottedlines) with those predicted by the numerical model (solid lines)for solid electrodes (black) and phase change electrodes (gray).The insert displays the volume fraction of solid material pre-dicted by the numerical model at the end of the 99 s pulse proto-col. Due to symmetry, only a single electrode is shown with themedial portion on the left.

111009-4 / Vol. 135, NOVEMBER 2013 Transactions of the ASME

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 5: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

was evaluated at the medial electrode–tissue interface 1.80 cmbelow the surface of the ultrasound gel (same depth as experimen-tal recordings). In all cases, including the solid electrodes (dottedlines), as electrode radius increased, the peak temperature at theend of treatment was reduced. This is due to the fact that largerelectrodes have a diminished electric potential gradient along theirboundaries. However, solid electrodes generally produced highertemperatures than PCM electrodes of equivalent radius. Oneexception occurred for large PCM electrodes with a TM thatexceeds the core temperature during treatment [Fig. 4(d) whereTM� T0¼ 10 �C]. In this case there was no phase transition andthe heat capacity per unit volume was greater for the solidelectrode.

The temperature profiles also illustrate the tradeoff betweenchoosing a TM that is low enough to begin absorbing heat at thestart of treatment and a TM that is high enough to avoid completemelting before the end of treatment. For example, in Fig. 4(c)when TM was 1 �C above the ambient temperature, the center ofthe core at a depth of 1.80 cm was completely melted by 35 s(indicated by the vertical arrow), and the temperature increased ata faster rate for the remainder of the treatment due to a decreasedheat capacity per unit volume. For larger electrodes capable ofholding a greater volume of PCM, the transition time is extended,and a TM of 1 �C above the baseline is advantageous [Fig. 4(d)].

Figure 5 shows temperature profiles at identical points of inter-est from the parametric study when repeated for an organic PCM(keq/100). Similar trends were observed as in the metallic case,with the main exception being the rate of temperature increaseduring the phase transition. Due to the increased thermal resist-ance, the temperature increased at a faster rate throughout thephase transition. However, this was balanced by the fact that ittakes longer for the organic to melt. For example, in Fig. 5(c)when TM was 1 �C above ambient conditions, it took 68 s for thecenter of the core to melt at a depth of 1.80 cm (indicated by thevertical arrow). This occurred nearly in unison with the end oftreatment at 70 s. As a result, the organic PCM maintained a finalpeak temperature that was lower than that of the metallic PCM ofequivalent radius and melting point. It is important to note that thesalt hydrate PCM (keq/10) resulted in the highest temperaturesduring treatment for all parameter combinations (data not shown).

Figure 6 summarizes the results of Figs. 4 and 5 by compilingthe data for temperature increase by the end of treatment. As awhole, the metallic PCM provides the most reduction in tempera-ture at any given radius if the melting point is tuned correctly.However, as mentioned, if melting is completed relatively earlyon, it becomes possible for the organic PCM to provide betterresults given the increased thermal resistance and extended transi-tion zone. This special case is highlighted by the negative bar inFig. 6(c).

Differences in phase transition kinetics for various thermal con-ductivities are shown in Fig. 7. The electrode radius is 1 mm andthe PCM melts at 1 �C above baseline. The metallic PCM wascharacterized by smooth transition at the center of the core at adepth of 1.80 cm [Fig. 7(a)]. In contrast, the organic PCMremained solid at that position until 66.65 s and then meltedabruptly over the next 1.8 s. The salt hydrate PCM presented cer-tain features of both the metallic PCM and organic PCM, giving itthe longest overall transition time of 38 s. Due to the high thermalresistance of the organic PCM, it took the longest for heat to con-duct inwards from the treated tissue to the electrode core. As aresult, a ridge of solid material remained in the core at the end oftreatment [Fig. 7(b)]. The melting front in the metallic or salthydrate core surpassed or was approaching the air–tissue inter-face, respectively.

Table 2 compares the volume of tissue experiencing an increasein temperature capable of causing thermal damage (VTD) to thevolume of tissue subject to IRE (VIRE) when reduced electrodespacings are used. The reported values are for an electrode radiusof 1 mm and PCM melting points of 1 and 4 �C above baseline.During in vivo IRE, when temperatures go above 43 �C, traces ofthermal coagulation as well as heat shock proteins are seen [40].This is equivalent to a temperature rise of 6 �C (T70 – T0) abovephysiologic baseline (37 �C), which is taken here to represent thethreshold for thermal damage. As mentioned, the IRE threshold isbased on a value of 500 V/cm [38,39]. Spatially, as the electrodespacing decreases from 1.5 to 0.5 cm, the location of the peaktemperature within the tissue domain shifts from the electrode–tis-sue interface to the center between the electrodes. Therefore, thePCM is less effective at removing heat as the spacing decreases.However, at a spacing of 0.5 cm, there is no need for the use of a

Fig. 3 Surface contour plots from the numerical model showing the (a) electric field distribu-tion and temperature distribution for the (b) solid electrodes and (c) phase change electrodes.All plots are shown at the end of the 99 s pulse protocol. In the x-y plane, the cross section istaken at a depth of 1.80 cm from the surface of the ultrasound gel. In the x-z plane the cross sec-tion is taken through the center of the electrodes (y 5 0).

Journal of Biomechanical Engineering NOVEMBER 2013, Vol. 135 / 111009-5

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 6: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

Fig. 4 Temperature profiles resulting from the parametric study on electrode radius (0.5, 0.75,1.0, and 1.25 mm) and PCM melting point above the baseline temperature (1, 4, 7, and 10 �C) forgallium. Temperatures are plotted at the medial electrode–tissue interface at a depth of 1.80 cmfrom the surface of the ultrasound gel. Dotted lines represent solid electrodes for comparison.In (c) the vertical arrow indicates the end of melting and the horizontal arrow highlights the tem-perature at the end of treatment.

Fig. 5 Temperature profiles resulting from the parametric study on electrode radius (0.5, 0.75,1.0, and 1.25 mm) and PCM melting point above the baseline temperature (1, 4, 7, and 10 �C)when keq/100. Temperatures are plotted at the medial electrode–tissue interface at a depth of1.80 cm from the surface of the ultrasound gel. Dotted lines represent solid electrodes for com-parison. In (c) the vertical arrow indicates the end of melting and the horizontal arrow highlightsthe temperature at the end of treatment.

111009-6 / Vol. 135, NOVEMBER 2013 Transactions of the ASME

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 7: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

PCM core, as the solid electrodes alone do not cause thermal dam-age. The benefit of PCM inclusion is seen clearly for a spacing of1.5 cm. Gallium that melts 4 �C above baseline reduces the VTD/VIRE

ratio from 0.1097 in the solid case to 0.0207, indicating that non-thermal IRE is the primary mechanism of cell death. For an elec-trode with a radius of 1.25 mm (not in Table 2), a gallium corethat melts 1 �C above baseline eliminates thermal damage alto-gether by the standards described here. This specific case isemphasized again in the discussion.

Discussion

The results presented above demonstrate that electrodes con-taining a PCM core can be used to mitigate temperature develop-ment during therapeutic electroporation. The choice of gallium asthe PCM was made for several reasons. First, its phase is easily

manipulated slightly above room temperature, which makes it alogical selection for experimental validation. Second, the thermalconductivity of metallic PCMs are considerably higher than otherclasses of PCMs, such as salt hydrates and organics. This enhan-ces heat dissipation to the external environment and keeps temper-atures closer to the melting point during the phase transition.Lastly, as alluded to earlier, gallium can be alloyed with bismuthto raise its melting point above physiologic temperature [35]. Thiswill be required for the treatment of deep-seated tumors. Fortumors treated noninvasively with plate electrodes, pure galliummay be acceptable since the equilibrium temperature between thetarget tissue and air is expected to be below 30 �C.

Because metallic PCMs have a low thermal resistance and heatuniformly, they melt relatively quickly compared to other classesof PCMs. For shorter ECT and EGT protocols, or when usingelectrodes with a larger radius, metallic PCMs with a meltingpoint close to ambient conditions are ideal. However, for longerIRE pulse protocols, such as those requiring 90 pulses to achievecontiguous ablations when the electrodes are spaced 2 cm apart[41], or when using electrodes with a smaller radius, the meltingpoint should be raised to ensure that the phase transition processdoes not subside before the end of treatment. This relationshipbetween the optimal melting point and electrode radius can beseen in Fig. 6(a). As the electrode radius increases, the optimalmelting point for mitigating peak temperature shifts to lower val-ues. In industrial applications demanding multiple melting–freezingcycles, the re-solidification period also plays into the choice ofPCM [42]. For electroporation-based therapies, this is less of aconcern as electrodes are typically single use.

According to the parametric study, the greatest reduction inpeak temperature at the end of treatment, when compared to solidelectrodes, was 5.6 �C [Fig. 6(a)]. This was obtained with a1.25 mm radius electrode containing a metallic PCM that melts1 �C above ambient conditions (T70, solid – T70, gallium¼ 34.8 �C– 29.2 �C¼ 5.6 �C). IRE protocols that produce tissue tempera-tures below 42 �C are associated only with nonthermal pathologicfeatures, including dehydrated cells and hemorrhagic infiltrate[40,41]. As mentioned, when temperature goes above 43 �C,traces of thermal coagulation as well as heat shock proteins areseen [40]. Therefore, even a modest reduction of 5.6 �C couldimprove the safety of high-dose IRE protocols performed nearthermally sensitive structures. If our results are projected from aphysiologic baseline temperature (37 �C) as opposed to the incu-bator temperature (25.9 �C), any thermal damage that occurswhen solid electrodes are used is likely to be eliminated by incor-porating a PCM core (T70, solid – T70, gallium¼ 45.9 �C –40.3 �C¼ 5.6 �C).

Specific calculations for thermal damage based on the simula-tions presented herein were not made due to the fact that

Fig. 6 Comparison of the temperature change during treat-ment for (a) gallium core electrodes and (b) when keq/100 (b). (c)The difference between (b) and (a) to emphasize the most effec-tive PCM for each parameter combination. White bars representsolid electrodes for comparison.

Fig. 7 (a) Melting profile for different PCM cores (gallium, keq/10, and keq/100) witha melting point 1 �C above baseline housed within a 1 mm radius electrode. Thedata is taken at the center of the core at a depth of 1.80 cm from the surface ofthe ultrasound gel. (b) Cross section of the volume fraction solid material throughthe center of the electrodes at the end of treatment.

Journal of Biomechanical Engineering NOVEMBER 2013, Vol. 135 / 111009-7

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 8: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

experiments were conducted below physiologic levels and the em-phasis was placed on elucidating the effects of varying electrodeproperties. However, relationships between electrode propertiesand the increase in temperature above a baseline are still expectedto translate clinically when trying to avoid reaching temperaturesabove 43 �C. Volume changes and fluid flow of the PCM meltwere not considered as part of the experimental validation, as amajority of gallium remained solidified. Future work is requiredto investigate the extent to which these phenomena influence sim-ulations during which the phase transition is completed. The effectof buoyancy induced convection is expected to be small due to thesmall volume of PCM between the electrodes, which are acting asfins [43]. Presumably, convection will have a greater influence onPCMs that transition from liquids to gases around body tempera-ture, such as low boiling point perfluorocarbons [21,22] that mayserve to further mitigate temperature increase. Additionally, whilegallium contracts upon melting, perfluorocarbons expand upon va-porization, which must be taken into consideration during thedesign and manufacturing processes.

It was shown in our previous theoretical study [18] that thegreatest difference in tissue temperature between solid and PCMcore electrodes occurs along the electrode–tissue interface, whichis coincidentally where the highest probability for thermal damageoccurs during large ablations [16,40,41]. Therefore, a point alongthis interface was also selected here for a majority of the paramet-ric study analysis. In the experimental setup, the fiber optic tem-perature probe was placed in the port on the printed circuit boardthat fell closest to this interface to try and depict the maximumdifference. Placement of the probe between the electrodes andplacement of multiple probes was avoided, as they have beenshown to distort the electric field distribution [39] and likely theaccompanying temperature distribution. However, validation ofthe numerical model at multiple points of interest should be per-formed in future work. Additionally, it was assumed that the alter-nate classes of PCMs had the same optimized transition intervalas gallium (1 �C), and more experiments on organics and salthydrates are required to better characterize their coolingperformance.

It is clear that additional benchtop testing is required to addressthe limitations posed above. The promising results presentedhere also warrant in vivo experiments on living tissue. With thesestudies it would be possible to perform gross pathologic and histo-pathologic evaluations for comparison with temperature measure-ments and thermal damage calculations. In living tissue, asopposed to phantoms, electroporation increases the tissue electri-cal conductivity [44], and subsequently heating. This dynamicfeature should be incorporated into future models, as it may affectthermal damage calculations [45] and phase change kinetics.

Conclusion

The incorporation of PCMs into biomedical devices offers alargely unexplored opportunity to enhance therapeutic outcomes.Here we show their potential to safely expand the parameter spacewhen applying nonthermal electroporation to treat aberrant cellmasses (i.e., treat large masses with a single electrode insertionwithout exceeding a critical thermal limit). In future work there is

room to optimize electrode designs for improved heat transfer intothe PCM domain, especially in the case of organics. This couldinvolve added PCM compartmentalization with fins. Outside thefield of electroporation, PCMs may also prove useful for thermalablation modalities, either to prevent tissue vaporization at the siteof treatment or reduce the risk of distant grounding pad burns.Furthermore, PCMs present many unique solutions for improvingdiagnostic imaging and aiding in the storage and preservation ofbiological materials. Due to this wide variety of applications thatcan benefit from the inclusion of PCMs, it is likely that their usagewill increase over the next decade.

Acknowledgment

This work was supported by the NSF (CBET-0933335 and CA-REER CBET-1055913) and the ICTAS Multiscale Bio-Engineered Devices and Systems (MBEDS) center. We thankJames P. Lane and Jason D. Schroedl for setting up the COMSOLserver, and Paulo A. Garcia and Katherine J. Prokop for reviewingthe manuscript.

References[1] Okino, M., and Mohri, H., 1987, “Effects of a High-Voltage Electrical Impulse

and an Anticancer Drug on in Vivo Growing Tumors,” Jpn. J. Cancer Res.,78(12), pp. 1319–1321.

[2] Titomirov, A. V., Sukharev, S., and Kistanova, E., 1991, “In Vivo Electropora-tion and Stable Transformation of Skin Cells of Newborn Mice by PlasmidDNA,” Biochim. Biophys. Acta, 1088(1), pp. 131–134.

[3] Neumann, E., Schaeferridder, M., Wang, Y., and Hofschneider, P. H., 1982,“Gene Transfer into Mouse Lyoma Cells by Electroporation in High ElectricFields,” Embo. J., 1(7), pp. 841–845.

[4] Davalos, R. V., Mir, L. M., and Rubinsky, B., 2005, “Tissue Ablation With Irre-versible Electroporation,” Ann. Biomed. Eng., 33(2), pp. 223–231.

[5] Pavliha, D., Kos, B., Zupanic, A., Marcan, M., Sersa, G., and Miklavcic, D.,2012, “Patient-Specific Treatment Planning of Electrochemotherapy: ProcedureDesign and Possible Pitfalls,” Bioelectrochemistry, 87, pp. 265–273.

[6] Garcia, P. A., Pancotto, T., Rossmeisl, J. H., Henao-Guerrero, N., Gustafson, N.R., Daniel, G. B., Robertson, J. L., Ellis, T. L., and Davalos, R. V., 2011, “Non-Thermal Irreversible Electroporation (N-TIRE) and Adjuvant FractionatedRadiotherapeutic Multimodal Therapy for Intracranial Malignant Glioma in aCanine Patient,” Technol. Cancer Res. Treat., 10(1), pp. 73–83.

[7] Neal, R. E. 2nd, Rossmeisl, J. H. Jr., Garcia, P. A., Lanz, O. I., Henao-Guerrero,N., and Davalos, R. V., 2011, “Successful Treatment of a Large Soft TissueSarcoma With Irreversible Electroporation,” J. Clin. Oncol. 29(13), pp.e372–e377.

[8] Weaver, J. C., and Chizmadzhev, Y. A., 1996, “Theory of Electroporation: AReview,” Bioelectrochem. Bioenergy, 41(2), pp. 135–160.

[9] Mir, L. M., Belehradek, M., Domenge, C., Orlowski, S., Poddevin, B., Belehra-dek, J. Jr., Schwaab, G., Luboinski, B., and Paoletti, C., 1991,“Electrochemotherapy, A New Antitumor Treatment: First Clinical Trial,” C.R. Acad. Sci. III, 313(13), pp. 613–618.

[10] Daud, A. I., DeConti, R. C., Andrews, S., Urbas, P., Riker, A. I., Sondak, V. K.,Munster, P. N., Sullivan, D. M., Ugen, K. E., Messina, J. L., and Heller, R.,2008, “Phase I Trial of Interleukin-12 Plasmid Electroporation in Patients WithMetastatic Melanoma,” J. Clin. Oncol., 26(36), pp. 5896–5903.

[11] Thomson, K. R., Cheung, W., Ellis, S. J., Park, D., Kavnoudias, H., Loader-Oliver, D., Roberts, S., Evans, P., Ball, C., and Haydon, A., 2011,“Investigation of the Safety of Irreversible Electroporation in Humans,” J.Vasc. Interv. Radiol., 22(5), pp. 611–621.

[12] Gehl, J., 2003, “Electroporation: Theory and Methods, Perspectives for DrugDelivery, Gene Therapy and Research,” Acta Physiol. Scand., 177(4), pp.437–447.

[13] Li, W., Fan, Q. Y., Ji, Z. W., Qiu, X. C., and Li, Z., 2011, “The Effects of Irre-versible Electroporation (IRE) on Nerves,” Plos One, 6(4), p. e18831.

[14] Maor, E., Ivorra, A., Leor, J., and Rubinsky, B., 2007, “The Effect of Irreversi-ble Electroporation on Blood Vessels,” Technol. Cancer Res. Treat., 6(4), pp.307–312.

[15] Rubinsky, B., Onik, G., and Mikus, P., 2007, “Irreversible Electroporation: ANew Ablation Modality—Clinical Implications,” Technol. Cancer Res. Treat.,6(1), pp. 37–48.

[16] Appelbaum, L., Ben-David, E., Sosna, J., Nissenbaum, Y., and Goldberg, S. N.,2012, “US Findings After Irreversible Electroporation Ablation: Radiologic-Pathologic Correlation,” Radiology, 262(1), pp. 117–125.

[17] Becker, S. M., and Kuznetsov, A. V., 2007, “Thermal Damage Reduction Asso-ciated With in Vivo Skin Electroporation: A Numerical Investigation JustifyingAggressive Pre-Cooling,” Int. J. Heat Mass Transfer, 50, pp. 105–116.

[18] Arena, C. B., Mahajan, R. L., Rylander, M. N., and Davalos, R. V., 2012,“Towards the Development of Latent Heat Storage Electrodes forElectroporation-Based Therapies,” Appl. Phys. Lett., 101(8), p. 083902.

[19] Mondieig, D., Rajabalee, F., Laprie, A., Oonk, H. A. J., Calvet, T., and Cuevas-Diarte, M. A., 2003, “Protection of Temperature Sensitive Biomedical Products

Table 2 VTD/VIRE ratio for 1 mm radius electrodes with varyingcore materials and spacings

Melting point TM�T0¼ 1 �C TM�T0¼ 4 �C

Electrode spacing 0.5 cm 1.0 cm 1.5 cm 0.5 cm 1.0 cm 1.5 cm

PCM (keq) 0 0 0.0221 0 0 0.0207PCM (keq/10) 0 0 0.0237 0 0 0.0238PCM (keq/100) 0 0 0.0211 0 0 0.0442Solid 0 0.0002 0.1097 0 0.0002 0.1097

111009-8 / Vol. 135, NOVEMBER 2013 Transactions of the ASME

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms

Page 9: Biomedical Engineering | Biomedical Engineering and Mechanics - … › davalos › pdf › Arena_2013_JBME.pdf · 2016-02-04 · Bioelectromechanical Systems Lab, Virginia Tech-Wake

Using Molecular Alloys As Phase Change Material,” Transfus Apher Sci.,28(2), pp. 143–148.

[20] Shim, H., McCullough, E. A.,and Jones, B. W., 2001, “Using Phase ChangeMaterials in Clothing,” Textile Res. J., 71(6), pp. 495–502.

[21] Sheeran, P. S., and Dayton, P. A., 2012, “Phase-Change Contrast Agents forImaging and Therapy,” Curr. Pharm. Design, 18(15), pp. 2152–2165.

[22] Sheeran, P. S., Luois, S. H., Mullin, L. B., Matsunaga, T. O., and Dayton, P. A.,2012, “Design of Ultrasonically-Activatable Nanoparticles Using Low BoilingPoint Perfluorocarbons,” Biomaterials, 33(11), pp. 3262–3269.

[23] Marongiu, M. J., and Clarksean, R., 1997, “Thermal Management of Electron-ics Enclosures Under Unsteady Heating/Cooling Conditions Using PhaseChange Materials (PCM),” Proceedings of the 32nd Intersociety Energy Con-version Engineering Conference (IECEC), pp. 1865–1870.

[24] Ge, H., and Liu, J., 2012, “Phase Change Effect of Low Melting Point Metal foran Automatic Cooling of USB Flash Memory,” Front. Energy, 6(3), pp. 207–209.

[25] Prakash, J., Garg, H. P., and Datta, G., 1985, “A Solar Water Heater With aBuilt-in Latent-Heat Storage,” Energy Convers. Manage., 25(1), pp. 51–56.

[26] Benard, C., Body, Y., and Zanoli, A., 1985, “Experimental Comparison ofLatent and Sensible Heat Thermal Walls,” Solar Energy, 34(6), pp. 475–487.

[27] Etheridge, M. L., Choi, J., Ramadhyani, S., and Bischof, J. C., 2013, “Methodsfor Characterizing Convective Cryoprobe Heat Transfer in Ultrasound GelPhantoms,” ASME J. Biomech. Eng., 135(2), p. 021001.

[28] Wagner, G. H., and Gitzen, W. H., 1952, “Gallium,” J. Chem. Educ., 29(4), p. 162.[29] Krishnan, S., and Garimella, S. V., 2004, “Analysis of a Phase Change Energy

Storage System for Pulsed Power Dissipation,” IEEE Trans. Compon. PackTechnol., 27(1), pp. 191–199.

[30] Dantzig, J. A., 1989, “Modeling Liquid-Solid Phase Changes With Melt Con-vection,” Int. J. Numer. Methods Eng., 28(8), pp. 1769–1785.

[31] Garcia, P. A., Davalos, R. V., and Pearce, J. A., 2012, “A Comparison Betweenthe Pulsed and Duty Cycle Approaches Used to Capture the Thermal Responseof Tissue During Electroporation-Based Therapies,” ASME Summer Bioengin-eering Conference Fajardo, Puerto Rico, p. 80574.

[32] Deri, B., Kotovsky, J., and Spadaccini, C., 2010, “Assessment of Latent HeatReservoirs for Thermal Management of QCW Laser Diodes,” No. LLNL-TR-425903, Lawrence Livermore National Laboratory, Livermore, CA.

[33] Incropera, F. P., and DeWitt, D. P., 1996, Fundamentals of Heat and MassTransfer, Wiley, New York.

[34] Amin, D. V., Lozanne, K., Parry, P. V., Engh, J. A., Seelman, K., and Mintz, A.,2011, “Image-Guided Frameless Stereotactic Needle Biopsy in Awake PatientsWithout the Use of Rigid Head Fixation,” J. Neurosurg., 114(5), pp. 1414–1420.

[35] Predel, B., 1960, “Die Zustandsbilder Gallium-Wismut und Gallium-Quecksilber, Vergleich der Koexistenzkurven mit den Theorien derEntmischung,” Z. Phys. Chem., 24(3–4), pp. 206–216.

[36] Antunes, C. L., Almeida, T. R. O., Raposeiro, N., Goncalves, B., and Almeida,P., 2012, “Using a Tubular Electrode for Radiofrequency Ablation Numericaland Experimental Analysis,” Compel, 31(4), pp. 1077–1086.

[37] Zalba, B., Marin, J. M., Cabeza, L. F., and Mehling, H., 2003, “Review onThermal Energy Storage With Phase Change: Materials, Heat Transfer Analysisand Applications,” Appl. Therm. Eng., 23(3), pp. 251–283.

[38] Garcia, P. A., Rossmeisl, J. H., Neal, R. E., Ellis, T. L., Olson, J. D., Henao-Guerrero, N., Robertson, J., and Davalos, R. V., 2010, “Intracranial NonthermalIrreversible Electroporation: in Vivo Analysis,” J. Membr. Biol., 236(1), pp.127–136.

[39] Arena, C. B., Szot, C. S., Garcia, P. A., Rylander, M. N., and Davalos, R. V.,2012, “A Three-Dimensional In Vitro Tumor Platform for Modeling Therapeu-tic Irreversible Electroporation,” Biophys. J, 103(9), pp. 2033–2042.

[40] Faroja, M., Ahmed, M., Appelbaum, L., Ben-David, E., Moussa, M., Sosna, J.,Nissenbaum, I., and Goldberg, S. N., 2013, “Irreversible Electroporation Abla-tion: Is All the Damage Nonthermal?,” Radiology, 266(2), pp. 462–470.

[41] Ben-David, E., Appelbaum, L., Sosna, J., Nissenbaum, I., and Goldberg, S. N.,2012, “Characterization of Irreversible Electroporation Ablation in in Vivo Por-cine Liver,” Am. J. Roentgenol., 198(1), pp. W62–W68.

[42] Stupar, A., Drofenik, U., and Kolar, J. W., 2010, “Application of Phase ChangeMaterials for Low Duty Cycle High Peak Load Power Supplies,” 6th Interna-tional Conference on Integrated Power Electronics Systems (CIPS), pp. 1–11.

[43] Lacroix, M., 2001, “Contact Melting of a Phase Change Material Inside aHeated Parallelepedic Capsule,” Energy Convers. Manage., 42(1), pp. 35–47.

[44] Davalos, R. V., Otten, D. M., Mir, L. M., and Rubinsky, B., 2004, “ElectricalImpedance Tomography for Imaging Tissue Electroporation,” IEEE Trans.Biomed. Eng., 51(5), pp. 761–767.

[45] Garcia, P. A., Rossmeisl, J. H. Jr., Neal, R. E. 2nd, Ellis, T. L., and Davalos, R.V., 2011, “A Parametric Study Delineating Irreversible Electroporation FromThermal Damage Based on a Minimally Invasive Intracranial Procedure,”Biomed. Eng. Online, 10(1), p. 34.

Journal of Biomechanical Engineering NOVEMBER 2013, Vol. 135 / 111009-9

Downloaded From: http://biomechanical.asmedigitalcollection.asme.org/ on 11/18/2013 Terms of Use: http://asme.org/terms