Upload
gen
View
54
Download
0
Tags:
Embed Size (px)
DESCRIPTION
CALICE pixel Laser testing update. Black cylinder. Laser MIP equivalent calibration update Si detector coupled to low noise CA + differentiator (no shaper) Amplifier Gain measured ~ 7mV/MIP (c.a. 3 times lower than nominal value, confirmed with manufacturer) - PowerPoint PPT Presentation
Citation preview
G.Villani Aug. 07 1
CALICE pixel Laser testing update
Laser MIP equivalent calibration update• Si detector coupled to low noise CA + differentiator (no shaper)• Amplifier Gain measured ~ 7mV/MIP (c.a. 3 times lower than nominal value, confirmed with manufacturer) • Amount of stray light reduced by using a small dark cylinder around the sensor
Black cylinder
G.Villani Aug. 07 2
CALICE pixel Laser testing
A250CF cal - Qvs V -
-2
0
2
4
6
8
10
12
14
16
18
0.498 3.82 7.24 10.53 13.58 16.78
V(mV)
Q(f
C)
Q(fC)=((2.6566+/-0.3)/3.25+V/3.25-1)*3.2A250CF calibration using injected charge through capacitor and pulse
generator
MIP charge signalnoise signal
G.Villani Aug. 07 3
CALICE pixel Laser testing
A250CF cal -Vout (laser E)
V = 1.1429*(%/25+1)2 - 3.4731*(%/25+1) + 14.656 +/-0.0272*(%/25+1)^2-0.0088*(%/25+1)+0.41
0
5
10
15
20
25
30
0 25 50 75 100
% Laser intensity
Vo
(mV
)
Laser output measurement - Vout vs
% intensity.
Laser output 0% Laser output 100%
G.Villani Aug. 07 4
CALICE pixel Laser testing
Q vs % laser intensity
-2
0
2
4
6
8
10
12
14
16
0 25 50 75 100
% Laser intensity
Q(f
C)
MIP eqv vs. % Laser intensity
-0.5
0
0.5
1
1.5
2
2.5
0 25 50 75 100
% Laser intensity
MIP
eqv
Q generated vs % laser intensity – MIP- eqv generated vs % laser intensity with EMI and
amplifier noise quadrature-subtracted
MIP-eqv calibration accurate to +/-15%
MIP ‘noise’Q in S3590-A
G.Villani Aug. 07 5
CALICE pixel Laser testing
Next step:• Refined calibration using non coated sensor and thru-hole sensor housing(back scattered light, non uniformity of thickness)• Study of laser temporal profile (for deconvolution of sensor response)• Comparison of Laser signal with radiation source signal• Calibration using different MIP ranges • SW control from DAQ
ND filters for different MIP ranges
Sensor housing
G.Villani Aug. 07 6
CALICE pixel Deep P-Well simulation
3D simulation of final layout with all around DPW ∆Q ~ 10% @ 2.2*105 pointsc.a. 40 hits points simulated
S1
0
100
200
300
400
500
600
700
800
1
3
5
S1 S2 S3 S4 S5 S6
G.Villani Aug. 07 7
ConclusionsConclusions
• Laser sub-MIP calibration accomplished: further refinement and verifications ongoing
• Full pixel simulation with final layout ongoing
CALICE pixel summary