Cellvizio Lab Handbook
Citation preview
www.cellviziolab.com
IN VIVO MOLECULAR IMAGINGWITH CELLULAR RESOLUTION
IN SERVICEPRESENTATION
www.cellviziolab.com
Disclaimer
• This presentation is not a substitute for the Instructions For Use provided with Cellvizio, including but not limited to the Cellvizio User Guide, the Quick Guide, the Fast System Setup and the Probe Disinfection User Guide.
• Always refer to the last revisions of the Instructions For Use provided with Cellvizio.
www.cellviziolab.com
Agenda
System description Before the Procedure Probe Handling
After the Procedure Troubleshooting Contact Us
www.cellviziolab.com
System Description: pCLE
ProFlex™ Microprobes•Designed for different applications•High Resolution: up to 1.4 µm•Thin Diameter: down to 300 µm
Laser Scanning Unit•Confocal Microscope•488 or 660 nm excitation beam•High Sensitivity(APD)•Handy, turn-key, easy-to-use
ImageCell™ Software•Real-Time image processing•Quantification features•Framerate up to 200 fps•LSU control
Foot Pedal•ImageCell™ remote control
www.cellviziolab.com
High technology•Simultaneous injection and collection•Patented technology•High sensitivity
Each fiber acts as a point source and a point detector or pinhole
«»
CONFOCALFLUORESCENCEDETECTIONIMAGER
System Description: pCLE
www.cellviziolab.com
OS+
OS++
Optical SectioningOS+
Working DistanceWD
WD+ WD++
Confocal Microprobes TURN TOOPTICALBIOPSY
Tailored by Application
www.cellviziolab.com
Model ApplicationsTip
Diameter (mm)
Lateral Resolution
(µm)
Optical Sectioning
(µm)
Working Distance
(µm)
Max Field Of view
(µm)
Deep brain imaging, designed for permanent implantation on freely moving mice
Part of the NeuroPak™ solution
0,350,47 3,3 15 0 325
Brain, deep brain in mice, other organs at depth if low invasiveness is mandatory 0,3 3,3 15 0 300
Surface Imaging
Brain, deep brain in rats, other organs at depth if low invasiveness is mandatory 0,65 3,3 15 0 600
General applicability, can be used to check fluorescence sensitivity in most targets 1,5 3,3 15 0 600
Vessels, angiogenesis, cell fate, cell morphology, utility depends on cell layer thickness and invasiveness
2,6 1,4 10 60 240
Hi-ResImaging
Vessels, angiogenesis, cell fate, cell morphology, utility depends on cell layer thickness and invasiveness
4,2 1,4 10 30 240
ImagingVessels, angiogenesis, cell fate, cell morphology, utility depends on cell layer thickness and invasiveness
4,2 1,4 10 100 240
Blood flow through the vessel (without penetration) image deeper cell layers of tumor, organ or tissue
1,8 3,5 70100/170
at488/660
600
Depth Imaging
Cavities, eye 0,94 3,5 30
50/70at
488/660 325
SSeries
MSeries
S-300*
S-0650*
S-1500
UltraMiniO
MiniO/30
MiniO/100
Z-1800
Mini-Z
ZSeries
CerboFlex™CerboFlex™ J
> Made of thousands fiber-optics
Confocal Microprobes
www.cellviziolab.com8
Preparing Cellvizio®
Connect the M
icroprobe
Lock the co
nnector Clockwise
Cellvizio®
Imaging
Microprobe Warm
-Up
15 minutes Calib
ration
Using th
e Quantik
it™
Launch ImageCell™
Select a Storage Dire
ctory
Clean the M
icroprobe
Using th
e Cletop®-S
Turn ON th
e System
www.cellviziolab.com9
• Turn On computer• Turn On LSU- On/Off switch at the rear of the LSU- press On/Off on front face- look for 1st green light
Log into Cellvizio accountlogin: Cellvizio
password: cellvizio2004
TURN ON THE SYSTEM
www.cellviziolab.com10
login : Administratorpassword : admin
• Click on the ImageCell™ Software icon on the desktop or in the Dock
• Log into the Cellvizio user account of your choice (by default: administrator)
ImageCell™ 3.8.3
LAUNCH THE SOFTWARE
www.cellviziolab.com11
Hold the Cletop-S system in one hand with the blue ratchet facing upwards.
Press down the blue ratchet to uncover the cleaning ribbon.
Hold the ratchet down while positioning the Confocal Microprobe’s connector (black proximal part) on the cleaning ribbon.
Perform a ¼ turn (90°) of the connector while keeping it in contact with the cleaning ribbon.
Slide the connector optical interface along the cleaning ribbon while maintaining contact.
1
3
2
5
4
Do not use the Cletop-S cleaning system to clean
the probe’s distal end
Clean the Confocal Microprobe’s connector
www.cellviziolab.com12
1. Remove yellow cap from LSU. Then turn the LSU connector wheel counterclockwise until it blocks.
2. Hold the Microprobe with the flat side and the arrow on the label facing upwards.
3. Insert the Confocal Miniprobe™ connector in the female portion on the LSU until it blocks.
4. Hold the connector in place, and turn the wheel clockwise until it clicks. Keep turning until a 2nd click.
5. Turn the connector wheel for another 45° (1/8 turn) clockwise after the 2nd click.
6. Second green light appears on LSU
15
2
Always check that the probe has been cleaned and is not damaged
Connect the Microprobe to the LSU
1
23
4
www.cellviziolab.com13
• Once the Confocal Microprobe is connected, click on start laser to initiate the warm-up.
• The warm-up should take at least 15 minutes to be fully completed.
• Follow the calibration procedure as described on the screen, using the corresponding vials of the Quantikit
• The calibration procedure should be completed before each procedure
MAKE SURE QUANTIKIT HAS NOT EXPIRED
Warm-up and calibration
www.cellviziolab.com
1 Calibration: enables real time image processing
2 Warm-up: reduces fibers’ fluorescence background
Optimized image quality
Raw image Processed image
Why a warm-up and a calibration?
www.cellviziolab.comAcquisition
LUT
Toolbox
Toolbox
www.cellviziolab.com
START/STOP
LASER START/STOP
RECORDING
«Freeze» LUT
Imaging control with the foot pedal
www.cellviziolab.comBEST IMAGES = clean probe + good calibration
Calibration done too long ago
Probe has moved in the connector
RECALIBRATE
Bad image with “snow effect”
How to get THE BEST image quality
1 2OR
MAKE SURE QUANTIKITS HAVE NOT EXPIRED
www.cellviziolab.com
Troubleshooting
If Microprobe is dirty
Probe handling: Soft contact with tissue
Clean the distal tip using the cleaning vial + QTip and check if image is clear
If not, unplug the probe and clean the proximal part with Cletop-S
CLEAN DIRTY
CENTERING FAILED
Unplug the probe and plug it again
www.cellviziolab.com
After the Procedure
Turn off the laser Disconnect the probe Put yellow cap on LSU
Clean and rinse the probe
Dry it
Carefully fold the
microprobe
1 2 3
4
Prepared by H. Gharbi | © Mauna Kea Technologies
After the procedure• Catch the animal and anesthetize it• Place it under the stereotaxic frame• Attach the CerboFlex body to the stereotaxic holder• Carefully unlock the CerboFlex screw on the implant’s guide• Slightly remove the CerboFlex from the implant• Clean the CerboFlex tip by keeping the Laser ON for onscreen check• Put the protection cap into the implant’s cannula• Carefully remove the CerboFlex microprobe from the holder• Fold the probe as shown
22
Note 1 : NeuroPak contents & How to pack the Cerboflex
Disconnect the microprobe by turning the Cellvizio’s connector clockwise until the click. Then, release and remove the CerboFlex.
1
Note 1 : NeuroPak contents & How to pack the Cerboflex
Put the distal and proximal protection back over the tip of the CerboFlex.
Note 1 : NeuroPak contents & How to pack the Cerboflex
2
Note 1 : N
euroPak contents & How
to pack the Cerboflex
Curve the fiber into a circle as shown. The diameter should fit the CerboFlex print in the storage box.
3
CAUTION: Do NOT twist or fold the optical fiber bundle of the CerboFlex
Prepared by H. Gharbi | © Mauna Kea Technologies
After the procedure• Catch the animal and anesthetize it• Place it under the stereotaxic frame• Attach the CerboFlex body to the stereotaxic holder• Carefully unlock the CerboFlex screw on the implant’s guide• Slightly remove the CerboFlex from the implant• Clean the CerboFlex tip by keeping the Laser ON for onscreen check• Put the protection cap into the implant’s cannula• Carefully remove the CerboFlex microprobe from the holder• Fold the probe as shown
22
Note 1 : NeuroPak contents & How to pack the Cerboflex
Disconnect the microprobe by turning the Cellvizio’s connector clockwise until the click. Then, release and remove the CerboFlex.
1
Note 1 : NeuroPak contents & How to pack the Cerboflex
Put the distal and proximal protection back over the tip of the CerboFlex.
Note 1 : NeuroPak contents & How to pack the Cerboflex
2
Note 1 : N
euroPak contents & How
to pack the Cerboflex
Curve the fiber into a circle as shown. The diameter should fit the CerboFlex print in the storage box.
3
CAUTION: Do NOT twist or fold the optical fiber bundle of the CerboFlex
5
www.cellviziolab.com
Post-processing with IC Viewer
Double-click on the corresponding movie icon to view it
Icon Description
• Superimpose a micrometric scale on the image
• Measure a distance* from a fixed point of reference
• Take a snapshot and export it
• Select some frames of the entire sequence and export it
Find your recorder sequences in dataImageCell
Process the sequence
* measurement values are indicative and should be considered as such
IC Viewer
1
2
3
www.cellviziolab.com
Quick and easy exportsSelect frames1
Choose export format2
3. Export3
From IC Viewer
From ImageCell
Exporting Cellvizio sequences
www.cellviziolab.com
Always use clean and disinfected probes
Do not use undue force on probes during procedure
Do not cover the top of the LSU nor put the screen on it
Use only recommended cleaning and disinfection procedures.
Read instruction guides completely
! !
General Safety Information
www.cellviziolab.com
Laser labeling and status
Do not stare with naked eyes or optical instruments at the laser beam emitted by the Confocal Miniprobe™
Laser class: Class 2M (resp. Class II) according to IEC 60825-1:2007 (resp. 21 CFR Part 1040.10 and 1040.11)
For detailed laser safety information, please refer to the Cellvizio User Guide
! !
LSU
Back panelFront panelThird green light
Laser Safety Information
ImageCell™