Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison ISBN: 978-0-471-67600-3

Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Embed Size (px)

Citation preview

Page 1: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Chapter 6. Biopotential AmplifiersMichael R. Neuman

Medical Instrumentation Application and Design, 4th Edition

John G. Webster, Univ. of Wisconsin, Madison

ISBN: 978-0-471-67600-3

Page 2: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.1 Rough sketch of the dipole field of the heart when the R wave is maximal. The dipole consists of the points of equal positive and negative charge separated from one another and denoted by the dipole moment vector M.

Page 3: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

fig_06_02

Figure 6.2 Relationships between the two lead vectors a1 and a2 and the cardiac vector M. The component of M in the direction of a1 is given by the dot product of these two vectors and denoted on the figure by va1. Lead vector a2 is perpendicular to the cardiac vector, so no voltage component is seen in this lead.

Page 4: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.3 Cardiologists use a standard notation such that the direction of the lead vector for lead I is 0°, that of lead II is 60°, and that of lead III is 120°. An example of a cardiac vector at 30° with its scalar components seen for each lead is shown.

Page 5: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure E6.1 (a) aVR (b) VR (c) Simplified circuit of (a)

Page 6: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.4 Connection of electrodes to the body to obtain Wilson’s central terminal

Page 7: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.5 (a), (b), (c) Connections of electrodes for the three augmented limb leads, (d) Vector diagram showing standard and augmented lead-vector directions in the frontal plane.

Page 8: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.6 (a) Positions of precordial leads on the chest wall, (b) Directions of precordial lead vectors in the transverse plane.

Page 9: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

fig_06_07

Figure 6.7 Block diagram of an electrocardiograph

Page 10: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Table 6.1

Page 11: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Table 6.1

Page 12: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Table 6.1

Page 13: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.8 Effect of a voltage transient on an ECG recorded on an electrocardiograph in which the transient causes the amplifier to saturate, and a finite period of time is required for the charge to bleed off enough to bring the ECG back into the amplifier's active region of operation. This is followed by a first-order recovery of the system.

Page 14: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.9 (a) 60 Hz power-line interference, (b) Electromyographic interference on the ECG.

Page 15: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

fig_06_10

Figure 6.10 A mechanism of electric-field pickup of an electrocardiograph resulting from the power line. Coupling capacitance between the hot side of the power line and lead wires causes current to flow through skin–electrode impedances on its way to ground.

Page 16: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.11 Current flows from the power line through the body and ground impedance, thus creating a common-mode voltage everywhere on the body. Zin is not only resistive but, as a result of RF bypass capacitors at the amplifier input, has a reactive component as well.

Page 17: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.12 Magnetic-field pickup by the electrocardiograph (a) Lead wires for lead I make a closed loop (shaded area) when patient and electrocardiograph are considered in the circuit. The change in magnetic field passing through this area induces a current in the loop, (b) This effect can be minimized by twisting the lead wires together and keeping them close to the body in order to subtend a much smaller area.

Page 18: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.13 A voltage-protection scheme at the input of an electrocardiograph to protect the machine from high-voltage transients. Circuit elements connected across limb leads on left-hand side are voltage-limiting devices.

Page 19: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.14 Voltage-limiting devices (a) Current–voltage characteristics of a voltage-limiting device, (b) Parallel silicon-diode voltage-limiting circuit, (c) Back-to-back silicon Zener-diode voltage-limiting circuit, (d) Gas-discharge tube (neon light) voltage-limiting circuit element.

Page 20: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

fig_06_15

Figure 6.15 Driven-right-leg circuit for minimizing common-mode interference. The circuit derives common-mode voltage from a pair of averaging resistors connected to v3 and v4 in Figure 3.5. The right leg is not grounded but is connected to output of the auxiliary op amp.

Page 21: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure E6.3 Equivalent circuit of driven-right-leg system of Figure 6.15.

Page 22: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.16 Voltage and frequency ranges of some common biopotential signals; dc potentials include intracellular voltages as well as voltages measured from several points on the body. EOG is the electro-oculogram, EEG is the electroencephalogram, ECG is the electrocardiogram, EMG is the electromyogram, and AAP is the axon action potential. (From J. M. R. Delgado, "Electrodes for Extracellular Recording and Stimulation," in Physical Techniques in Biological Research, edited by W. L. Nastuk, New York: Academic Press, 1964)

Page 23: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.17 (a) Basic arrangement for negative-input-capacitance amplifier. Basic amplifier is on the right-hand side; equivalent source with lumped series resistance Rs and shunt capacitance Cs is on the left, (b) Equivalent circuit of basic negative-input-capacitance amplifier.

Page 24: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.18 This ECG amplifier has a gain of 25 in the dc-coupled stages. The high-pass filter feeds a noninverting-amplifier stage that has a gain of 32. The total gain is 25 X 32 = 800. When mA 776 op amps were used, the circuit was found to have a CMRR of 86 dB at 100 Hz and a noise level of 40 mV peak to peak at the output. The frequency response was 0.04 to 150 Hz for ±3 dB and was flat over 4 to 40 Hz. A single op amp chip, the LM 324, that contains four individual op amps could also be used in this circuit reducing the total parts count.

Page 25: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.19 Timing diagram for beat-to-beat cardiotachometer.

Page 26: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

fig_06_20

Figure 6.20 The various waveforms for the EMG integrator.

Page 27: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.21 Signal-averaging technique for improving the SNR in signals that are repetitive or respond to a known stimulus.

Page 28: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.22 Typical fetal ECG obtained from the maternal abdomen. F represents fetal QRS complexes; M represents maternal QRS complexes. Maternal ECG and fetal ECG (recorded directly from the fetus) are included for comparison. (From "Monitoring of Intrapartum Phenomena," by J. F. Roux, M. R. Neuman, and R. C. Goodlin, in CRC Critical Reviews in Bioengineering, 2, pp. 119-158, January 1975, © CRC Press. Used by permission of CRC Press, Inc.)

Page 29: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.23 Block diagram of a scheme for isolating fetal ECG from an abdominal signal that contains both fetal and maternal ECGs. (From "Monitoring of Intrapartum Phenomena," by J. F. Roux, M. R. Neuman, and R. C. Goodlin, in CRC Critical Reviews in Bioengineering, 2, pp. 119–158, January 1975, © CRC Press. Used by permission of CRC Press, Inc.)

Page 30: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.24 Block diagram of a cardiac monitor.

Page 31: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

Figure 6.25 Block diagram of a system used with cardiac monitors to detect increased electrode impedance, lead wire failure, or electrode fall-off.

Page 32: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

The ECG shown is distorted as a result of an instrumentation problem. Discuss possible causes of this distortion, and suggest means of correcting the problem.

Page 33: Chapter 6. Biopotential Amplifiers Michael R. Neuman Medical Instrumentation Application and Design, 4th Edition John G. Webster, Univ. of Wisconsin, Madison

The figure shows ECGs from simultaneous leads I and II. Sketch the vector loop for this QRS complex in the frontal plane.