70
© Siemens Healthcare GmbH, 2017 Computed Tomography SOMATOM go.Up Clinical Cases Not for distribution / use in the U.S.

Computed Tomography SOMATOM go.Up Clinical Cases

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Computed Tomography SOMATOM go.Up Clinical Cases

© Siemens Healthcare GmbH, 2017

Computed TomographySOMATOM go.Up

Clinical CasesNot for distribution / use in the U.S.

Page 2: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 2 | © Siemens Healthcare GmbH, 2017

Lung Imaging

Not for distribution / use in the U.S.

Page 3: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 3 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Lung imagingwithout contrast media application

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 9 sScan length: 400 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 130 kV / 101 mAsCTDIvol: 6.58 mGyDLP: 247 mGy cm1 mm slice thickness

• Excellent visualization of lung tissue• Balanced parameter utilization for fast

scanning• Multiple lung nodules visualized

coronal oblique thick-slice MIP

Page 4: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 4 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyCinematic VRT performed with syngo.viaCinematic VRT is recommended for communication, education, and publication purposes and not intended for diagnostic reading.Not for distribution / use in the U.S.

Lung imagingwithout contrast media application

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 9 sScan length: 400 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 130 kV / 101 mAsCTDIvol: 6.58 mGyDLP: 247 mGy cm

• Excellent visualization of lung tissue• High image quality

Page 5: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 5 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Low-dose lung imagingwith Tin Filter

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 13 sScan length: 398 mmRotation time: 0.8 sPitch factor: 1Scan parameters: Sn110 kV / 100 mAsCTDIvol: 1.28 mGyDLP: 47.7 mGy cm1 mm slice thickness

• Excellent visualization of lung tissue• Outstanding low-dose protocol• Improved image quality for air-to-tissue

enabled by Tin Filter technology

coronal and sagittal MPR

Page 6: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 6 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Low-dose lung imagingwith Tin Filter

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 13 sScan length: 398 mmRotation time: 0.8 sPitch factor: 1Scan parameters: Sn110 kV / 100 mAsCTDIvol: 1.28 mGyDLP: 47.7 mGy cm1 mm slice thickness

• Low-dose protocol• Improved image quality for air-to-soft-tissue

enabled by Tin Filter technology• Lung nodules delineated sharply

thick-slice MIP

Page 7: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 7 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Low-dose lung imagingwith Tin Filter

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 8 sScan length: 344 mmRotation time: 0.8 sPitch factor: 1.4Scan parameters: Sn110 kV / 128 mAsCTDIvol: 1.02 mGyDLP: 32.7 mGy cm2 mm slice thickness

• Excellent visualization of lung nodules• Low radiation dose applied• Obese patient

2 mm slice thicknesscoronal MPR

Page 8: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 8 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Low-dose lung imagingwith Tin Filter

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 9 sScan length: 396 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: Sn110 kV / 75 mAsCTDIvol: 0.44 mGyDLP: 16.4 mGy cm2 mm slice thickness

• Outstanding performance enabledby Tin Filter, ideal for lung screening purposes

thick-slice MIP coronal MPR

Page 9: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 9 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Lung imagingwith i.v. contrast media application

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 6 sScan length: 285 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 110 kV / 156 mAsCTDIvol: 4.78 mGyDLP: 125 mGy cm0.8 mm slice thickness

• Rule-out of pulmonary embolism• Severe scoliosis• Obese patient• Test bolus application for CTA planning

oblique MPR topogram

Page 10: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 10 | © Siemens Healthcare GmbH, 2017

Pelvic Imaging

Not for distribution / use in the U.S. Page 10 | © Siemens Healthcare GmbH, 2017

Page 11: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 11 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Pelvic imagingwith i.v. contrast media application

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 13 sScan length: 321 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 9.78 mGyDLP: 290 mGy cm3 mm slice thickness

• Follow-up examination• High image quality even in challenging

body regions like pelvic area

sagittal MPR

Page 12: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 12 | © Siemens Healthcare GmbH, 2017

Abdominal andWhole-Body Imaging

Not for distribution / use in the U.S. Page 12 | © Siemens Healthcare GmbH, 2017

Page 13: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 13 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Abdominal imagingwithout contrast media application

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 13 sScan length: 318 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 10.31 mGyDLP: 303 mGy cm3 mm slice thickness

• Obese patient• Excellent dose efficiency enabled by

Stellar detector technology• Impaired left kidney

sagittal and coronal oblique MPR

Page 14: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 14 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Abdominal imagingwith oral contrast media application 1/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 22 sScan length: 513 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 8.92 mGyDLP: 436 mGy cm3 mm slice thickness

• Advanced tumor stage• Rule-out of severe stenosis of the bowel

sagittal and coronal oblique MPR

Page 15: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 15 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Abdominal imagingwith oral contrast media application 2/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 22 sScan length: 513 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 8.92 mGyDLP: 436 mGy cm3 mm slice thickness

• Advanced tumor stage• Rule-out of severe stenosis of the bowel

Page 16: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 16 | © Siemens Healthcare GmbH, 2017

Abdominal imagingwith i.v. and rectal contrast media application 1/3

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 21 sScan length: 494 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.95 mGyDLP: 373 mGy cm3 mm slice thickness

• Excellent dose efficiency enabled byStellar detector technology

• Evaluation of bowel, e.g. rule-out ofperforation

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Page 17: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 17 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Abdominal imagingwith i.v. and rectal contrast media application 2/3

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 21 sScan length: 494 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.95 mGyDLP: 373 mGy cm1 mm slice thickness

• Fine visualization even of small bonystructures in routine scanning

• Automatic MPR orientation thanksto Recon&GO

Page 18: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 18 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Abdominal imagingwith i.v. and rectal contrast media application 3/3

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 21 sScan length: 494 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.95 mGyDLP: 373 mGy cm3 mm slice thickness

• High resolution data in routinefor volume scanning

Page 19: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 19 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Abdominal imagingwith i.v. contrast media application

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 23 sScan length: 526 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.31 mGyDLP: 366 mGy cm3 mm slice thickness

• Follow-up scan in case of double-pigtailcatheter placement

sagittal and coronal oblique MPRthick-slice MIP

demonstrating catheter placement

Page 20: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 20 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Long scan range scanwith i.v. contrast media application

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 17 sScan length: 729 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 130 kV / 184 mAsCTDIvol: 8.3 mGyDLP: 585 mGy cm3 mm slice thickness

• Staging examination with delayed phase• Accidental finding of aortic aneurysm• Easy postprocessing available with Vessel

Extension tools at CT View&GO

sagittal and coronal oblique MPR

Page 21: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 21 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 1/2

SOMATOM go.Up

Collimation: 32 x 0.7 mm (each)Scan time: 16 s (thorax)

21 s (abdomen)Scan length: 392 mm (thorax)

502 mm (abdomen)Rotation time: 0.8 s (each)Pitch factor: 0.8 (each)Scan parameters: 130 kV / 58 mAs (thorax)

130 kV / 98 mAs (abdomenCTDIvol: 5.37 mGy (thorax)

10.5 mGy (abdomen)DLP: 197 mGy cm (thorax)

502 mGy cm (abdomen)

• Follow-up after partial liver resection• Multi-step CT protocol

3 mm MPR

Page 22: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 22 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 2/2

SOMATOM go.Up

Collimation: 32 x 0.7 mm (each)Scan time: 16 s (thorax)Scan length: 392 mm (thorax)Rotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 58 mAs (thorax)CTDIvol: 5.37 mGy (thorax)DLP: 197 mGy cm (thorax)

• VRT based on thoracic spiral

Page 23: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 23 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 1/3

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 14 sScan length: 345 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 54 mAsCTDIvol: 3.1 mGyDLP: 100 mGy cm

• Multiple infiltrates and lesions with cavitiesclearly visualized

Page 24: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 24 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 2/3

This image cannot currently be displayed.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 14 sScan length: 345 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 54 mAsCTDIvol: 3.1 mGyDLP: 100 mGy cm1 mm slice thickness

• Multiple infiltrates and lesions• High resolution data in routine for

volume scanning

Page 25: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 25 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 3/3

coronal oblique MPR

arterial phase derived from thorax spiral

corresponding portal-venous phase from abdominal spiral

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 21 sScan length: 490 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 6.3 mGyDLP: 295 mGy cm3 mm slice thickness

• Thrombosis of the portal vein

Page 26: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 26 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 1/4

Topogram

SOMATOM go.Up

Collimation: 32 x 0.7 mm (each)Scan time: 16 s (thorax); 21 s (abdomen)Scan length: 390 mm (thorax); 484 mm (abdomen)Rotation time: 0.8 s (each)Pitch factor: 0.8 (each)Scan parameters: 130 kV / 54 mAs (thorax)

130 kV / 98 mAs (abdomenCTDIvol: 4.3 mGy (thorax); 6.77 mGy (abdomen)DLP: 157 mGy cm (thorax); 311 mGy cm (abdomen)3 mm slice thickness

• Follow-up scan / staging• Resection of right lung• Spondylolisthesis lumbar spine with dorsal

fixateur interne and fusion

Page 27: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 27 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 2/4

coronal MPR3 mm

coronal MPR1 mm

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 16 sScan length: 390 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 54 mAs (thorax)CTDIvol: 4.3 mGyDLP: 157 mGy cm

• Resection of right lung• Scar tissue left lung well delineated• High image quality even in challenging body

regions like lung apex

Page 28: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 28 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Two-step thorax and abdomen imagingwith i.v. contrast media application 3/4

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 21 sScan length: 484 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 6.77 mGyDLP: 311 mGy cm1 mm slice thickness

• Spondylolisthesis lumbar spine with dorsalfixateur intern and fusion

• Defect left ala os ilii• Hemangioma thoracic vertebra

Page 29: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 29 | © Siemens Healthcare GmbH, 2017

Two-step thorax and abdomen imagingwith i.v. contrast media application 4/4

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 21 sScan length: 484 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 6.77 mGyDLP: 311 mGy cm

Courtesy of Erlangen University Hospital, Erlangen, GermanyCinematic VRT performed with syngo.viaCinematic VRT is recommended for communication, education, and publication purposes and not intended for diagnostic reading.Not for distribution / use in the U.S.

Page 30: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 30 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Multi-phase abdomen imagingwith i.v. contrast media application 1/3

SOMATOM go.Up

Collimation: 32 x 0.7 mm (each)Scan time: 10 s (native) 5.2 s (corticomedullar);

22 s (parenchymal)Scan length: 240 mm (native);

242 mm (corticomed.); 570 mm (parench.)Rotation time: 0.8 s (each)Pitch factor: 0.8 (native), 1.5 (corticomed.); 0.8 (parench.)Scan parameters: 130 kV/ 98 mAs (native)

130 kV / 261 mAs (corticomed.)130 kV / 98 mAs (parench.)

CTDIvol: 12.14 mGy (native); 12.4 mGy (corticomed.); 11 mGy (parench.)DLP: 262 mGy cm (native); 271 mGy cm (corticomed.); 529 mGy cm (parench)3 mm slice thickness

• Unclear mass of the right kidney

Page 31: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 31 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Multi-phase abdomen imagingwith i.v. contrast media application 2/3

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 22 s (parenchymal)Scan length: 570 mm (parench.)Rotation time: 0.8 sPitch factor: 0.8 (parench.)Scan parameters: 130 kV / 98 mAs (parench.)CTDIvol: 11 mGy (parench.)DLP: 529 mGy cm (parench.)Slice thickness: 3 mm

• Unclear mass of the right kidney

Page 32: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 32 | © Siemens Healthcare GmbH, 2017

Multi-phase abdomen imagingwith i.v. contrast media application 3/3

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 5.2 s (corticomedullar)Scan length: 242 mm (corticomed.)Rotation time: 0.8 sPitch factor: 1.5 (corticomed.)Scan parameters: 130 kV / 261 mAsCTDIvol: 12.4 mGy (corticomed.)DLP: 271 mGy cm (corticomed.)

• Unclear mass of the right kidney

Courtesy of Erlangen University Hospital, Erlangen, GermanyCinematic VRT performed with syngo.viaCinematic VRT is recommended for communication, education, and publication purposes and not intended for diagnostic readingNot for distribution / use in the U.S.

Page 33: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 33 | © Siemens Healthcare GmbH, 2017

Thorax imagingwithout contrast media application

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

1 mm MPR thick slice MIP

inline spine labelinginline rib unfolding

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 13 sScan length: 398 mmRotation time: 0.8 sPitch factor: 1Scan parameters: 110 kV / 100 mAsCTDIvol: 1.28 mGyDLP: 47.7 mGy cm

• Rule-out of fractures• Inline rib unfolding and labeling

of the vertebra

Page 34: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 34 | © Siemens Healthcare GmbH, 2017

Two-step thorax and abdomen imagingwith i.v. contrast media application 1/2

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

topogram

coronal 1.5 mm MPR

coronal 5 mm MPR

SOMATOM go.Up

Collimation: 32 x 0.7 mm (each)Scan time: 15 s (thorax); 22 s (abdomen)Scan length: 362 mm (thorax); 519 mm (abdomen)Rotation time: 0.8 s (each)Pitch factor: 0.8 (each)Scan parameters: 130 kV / 54 mAs (thorax)

130 kV / 98 mAs (abdomen)

CTDIvol: 6.34 mGy (thorax); 7.09 mGy (abdomen)DLP: 214 mGy cm (thorax); 380 mGy cm (abdomen)

• Obese patient• BMI 37

Page 35: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 35 | © Siemens Healthcare GmbH, 2017

Two-step thorax and abdomen imagingwith i.v. contrast media application 2/2

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

axial sagittal and coronal 3 mm MPR

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 22 sScan length: 519 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.09 mGyDLP: 380 mGy cm

• Obese patient• BMI 37

Page 36: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 36 | © Siemens Healthcare GmbH, 2017

Two-step thorax and abdomen imagingwith i.v. contrast media application 1/3

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

axial. sagittal and coronal 1 mm MPR

oblique 5 mm MPR

oblique 10 mm MIP

SOMATOM go.Up

Collimation: 32 x 0.7 mm (each)Scan time: 17 s (thorax); 24 s (abdomen)Scan length: 396 mm (thorax); 557 mm (abdomen)Rotation time: 0.8 s (each)Pitch factor: 0.8 (each)Scan parameters: 130 kV / 54 mAs (thorax);

130 kV / 98 mAs (abdomen)CTDIvol: 4.19 mGy (thorax); 8.06 mGy (abdomen)DLP: 156 mGy cm (thorax); 429 mGy cm (abdomen)

• Excellent dose efficiency enabled byStellar technology

Page 37: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 37 | © Siemens Healthcare GmbH, 2017

Two-step thorax and abdomen imagingwith i.v. contrast media application 2/3

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

coronal 1 mm MPR coronal 3 mm MPR

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 24 sScan length: 557 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 8.06 mGyDLP: 429 mGy cm

• Excellent dose efficiency enabled byStellar technology

Page 38: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 38 | © Siemens Healthcare GmbH, 2017

Two-step thorax and abdomen imagingwith i.v. contrast media application 3/3

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

axial 1 mm MPR axial 3 mm MPR

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 24 sScan length: 557 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 8.06 mGyDLP: 429 mGy cm

• Excellent dose efficiency enabled byStellar detector technology

Page 39: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 39 | © Siemens Healthcare GmbH, 2017

Imaging inPresence of Implants

Not for distribution / use in the U.S. Page 39 | © Siemens Healthcare GmbH, 2017

Page 40: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 40 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Improving image quality with iMARTumor staging 1/2

without iMAR with iMAR

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 22 sScan length: 524 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.74 mGyDLP: 386 mGy cm

• Staging examination• Non-diagnostic images in case of metal

artifacts in the small pelvis area• iMAR allows evaluation of pelvic organs,

e.g. iliac lymph node stages

Page 41: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 41 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Improving image quality with iMARTumor staging 2/2

SOMATOM go.Up

Collimation: 32 x 0.7 mm (each)Scan time: 16 s (thorax); 22 s (abdomen)Scan length: 389 mm (thorax); 524 mm (abdomen)Rotation time: 0.8 s (each)Pitch factor: 0.8 (each)Scan parameters: 130 kV / 54 mAs (thorax)

130 kv / 98 mAs (abdomen)CTDIvol: 4.73 mGy (thorax); 7.74 mGy (abdomen)DLP: 173 mGy cm

• Staging examination• Advanced metastatic disease

Page 42: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 42 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Improving image quality with iMARVisualization of urinary system 1/2

without iMAR

with iMAR

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 36 sScan length: 829 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.74 mGyDLP: 386 mGy cm

• Evaluation of the urinary system• Hip implants causing severe beam

hardening artifacts• Long scan range

Page 43: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 43 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Improving image quality with iMARVisualization of urinary system

3 mm axial obliquewith iMAR

3 mm coronal obliquewith iMAR

thick-slice MIP obliquewith iMAR

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 36 sScan length: 829 mmRotation time: 0.8 sPitch factor: 0.8Scan parameters: 130 kV / 98 mAsCTDIvol: 7.74 mGyDLP: 386 mGy cm

• Evaluation of the urinary system• Hip implants causing severe beam

hardening artifacts• Long scan range

Page 44: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 44 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Improving image quality with iMARTumor recurrence

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 24 sScan length: 7712 mmRotation time: 0.8 sPitch factor: 1.1Scan parameters: 130 kV / 135 mAsCTDIvol: 9.85 mGyDLP: 736 mGy cm3 mm slice thickness

• Re-staging examination• Instrumentation with fixateur intern• Tumor recurrence with infiltration

of the spinal canal

Page 45: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 45 | © Siemens Healthcare GmbH, 2017

CT Angiography

Not for distribution / use in the U.S. Page 45 | © Siemens Healthcare GmbH, 2017

Page 46: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 46 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Thoracic aorta CT angiographyat 80 kV

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 10 sScan length: 452 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 80 kV / 375 mAsCTDIvol: 3.45 mGyDLP: 147 mGy cm

• CT angiography at low kV and high mAs• High Power 80 for improved iodine contrast• VRT inline results with CT View&GO• Potential reduction of required contrast

media dosage

Page 47: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 47 | © Siemens Healthcare GmbH, 2017

CT angiography of the whole aorta

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Not for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 27 sScan length: 607 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters:130 kV

• Aortic dissection• Perfusion deficency left

kidney; perfused falselumen, thrombus

• Get the timing right withthe new timeline

3mm MPR shown

Page 48: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 48 | © Siemens Healthcare GmbH, 2017

Thoracic aorta CT angiography

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 10 sScan length: 452 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 130 kV / 98 mAsCTDIvol: 6.88 mGyDLP: 230 mGy cm

• Cinematic rendering with syngo.via forvisualization of double aortic arch

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Cinematic VRT performed with syngo.viaCinematic VRT is recommended for communication, education, and publication purposes and not intended for diagnostic reading.Not for distribution / use in the U.S.

Page 49: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 49 | © Siemens Healthcare GmbH, 2017

Peripheral run-off CT angiography

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Not for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 32 sScan length: 934 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 80 kV

• Fine visualization ofperipheral vessels

• Post processing angio toolsat hand thanks to CTView&GO

1 mm axial MPR and oblique VRT shown

Page 50: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 50 | © Siemens Healthcare GmbH, 2017

Head Imaging

Not for distribution / use in the U.S. Page 50 | © Siemens Healthcare GmbH, 2017

Page 51: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 51 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Brain imagingwithout contrast media application 1/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 16 sScan length: 192 mmRotation time: 1 sPitch factor: 0.55Scan parameters: 130 kV / 104 mAsCTDIvol: 47.26 mGyDLP: 789 mGy cm

• Excellent low-contrast performance fordifferentiation of grey and white matter

• High density of detector channels for furtherimprovements in low contrast performance

axial

5 mm MPR reconstruction

Page 52: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 52 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Brain imagingwithout contrast media application 2/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 16 sScan length: 192 mmRotation time: 1 sPitch factor: 0.55Scan parameters: 130 kV / 104 mAsCTDIvol: 47.26 mGyDLP: 789 mGy cm

• High image quality also in the posterior fossa

axial

3 mm MPR reconstruction

Page 53: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 53 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Brain imagingwithout contrast media application 1/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 13 sScan length: 180 mmRotation time: 1 sPitch factor: 0.55Scan parameters: 130 kV / 104 mAsCTDIvol: 44.4 mGyDLP: 691 mGy cm3 mm slice thickness

• High contrast of brain structures• Excellent gray/white matter differentiation• Follow-up of stroke patient

Page 54: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 54 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Brain imagingwithout contrast media application 2/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 13 sScan length: 180 mmRotation time: 1 sPitch factor: 0.55Scan parameters: 130 kV / 104 mAsCTDIvol: 44.4 mGyDLP: 691 mGy cm3 mm slice thickness

• High resolution also of fine bony structuresfrom standard brain scan data

Page 55: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 55 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Skull base imaging

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 8 sScan length: 117 mmRotation time: 1 sPitch factor: 0.55Scan parameters: 130 kV / 58 mAsCTDIvol: 33.1 mGyDLP: 306 mGy cm

• 0.6 mm reconstructions due toStellar detector technology

Page 56: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 56 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Low-dose head imagingof the sinus

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 11 sScan length: 161 mmRotation time: 1 sPitch factor: 0.55Scan parameters: Sn110 kV / 28 mAsCTDIvol: 1.34 mGyDLP: 18.28 mGy cm

• Tin Filter technology for exceptional low-doseperformance with high contrast

• Detailed visualization of fine bone structureswith minimum slice thickness of 0.6 mm

0.6 mm reconstruction 0.8 mm reconstruction

Page 57: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 57 | © Siemens Healthcare GmbH, 2017

Sinus imagingwith Tin Filter

Courtesy of Centro Hospitalar de São João, Porto, PortugalNot for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 10 sScan length: 143 mmRotation time: 1.0 sPitch factor: 0.55Scan parameters: Sn 110 kV / 94 mAsCTDIvol: 2.58 mGyDLP: 30 mGy cm

• Low-dose protocol• Tin Filter technology for exceptionaly low-dose

performance with high contrast• Detailed visualization of bone structures and

swelling of soft tissue

1 mm slice thickness MPR 3 mm slice thickness MPR

Page 58: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 58 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Brain imagingwith i.v. contrast media application 1/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 15 sScan length: 214 mmRotation time: 1 sPitch factor: 0.55Scan parameters: 130 kV / 104 mAsCTDIvol: 41.2 mGyDLP: 779 mGy cm

• Visualization of aneurysm

axial and sagittal thin-slice MIP

Page 59: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 59 | © Siemens Healthcare GmbH, 2017

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

Brain imagingwith i.v. contrast media application 2/2

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 15 sScan length: 214 mmRotation time: 1 sPitch factor: 0.55Scan parameters: 130 kV / 104 mAsCTDIvol: 41.2 mGyDLP: 779 mGy cm

• Visualization of aneurysm

3 mm MPR

Page 60: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 60 | © Siemens Healthcare GmbH, 2017

Brain imagingwith i.v. contrast media application

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Not for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 11 sScan length: 159 mmScan parameters: 130 kV

• Sequence mode• Tilt functionality available for

thorough head neuro evaluation• Native and contrast media

enhanced scans shown

Page 61: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 61 | © Siemens Healthcare GmbH, 2017

Computed TomographyPage 61 | © Siemens Healthcare GmbH, 2017

Musculosceletal Imaging

Not for distribution / use in the U.S.

Page 62: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 62 | © Siemens Healthcare GmbH, 2017

Hand imagingwithout contrast media application

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

sagittal and coronal0.8 mm MPR

axial 0.8 mm MPR

VRT

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 7 sScan length: 141 mmRotation time: 1 sPitch factor: 0.8Scan parameters: 130 kV / 56 mAsCTDIvol: 4.97 mGyDLP: 58 mGy cm

• Rule-out of fresh fracture

Page 63: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 63 | © Siemens Healthcare GmbH, 2017

Computed TomographyPage 63 | © Siemens Healthcare GmbH, 2017

Virtual Colonoscopy

Not for distribution / use in the U.S.

Page 64: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 64 | © Siemens Healthcare GmbH, 2017

Virtual colonoscopywith Tin Filter

Courtesy of Erlangen University Hospital, Erlangen, GermanyNot for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 10 sScan length: 430 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: Sn110 kV / 40 mAsCTDIvol: 0.63 mGyDLP: 29 mGy cm

• Tin Filter low-dose protocol• Low-dose protocol• Virtual colonoscopy

Page 65: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 65 | © Siemens Healthcare GmbH, 2017

Computed TomographyPage 65 | © Siemens Healthcare GmbH, 2017

Pediatrics

Not for distribution / use in the U.S.

Page 66: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 66 | © Siemens Healthcare GmbH, 2017

Abdomen imagingwith i.v. contrast media application

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Not for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 9 sScan length: 419 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: 80 kV / 129 mAsCTDIvol: 0.33 mGyDLP: 13 mGy cm3 mm slice thickness

• Low-dose protocol• 2-years-old child• CARE Child protocols for ultra-low dose in

pediatrics

Page 67: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 67 | © Siemens Healthcare GmbH, 2017

Abdomen imagingwith i.v. contrast media application

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Not for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 9 sScan length: 338 mmRotation time: 0.8 sPitch factor: 1.2Scan parameters: 80 kV / 135 mAsCTDIvol: 0.39 mGyDLP: 12 mGy cm3 mm slice thickness

• Low-dose protocol• 5-years-old child• CARE Child protocols for ultra-low dose in

pediatrics

Page 68: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 68 | © Siemens Healthcare GmbH, 2017

Lung imagingwith Tin Filter

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Not for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 7 sScan length: 312 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: Sn 110 kV / 94 mAsCTDIvol: 0.16 mGyDLP: 4.6 mGy cm3 mm slice thickness

• Low-dose protocol• 9-years-old child• CARE Child protocols for ultra-low dose in

pediatrics

Page 69: Computed Tomography SOMATOM go.Up Clinical Cases

Computed TomographyPage 69 | © Siemens Healthcare GmbH, 2017

Lung imagingwith Tin Filter

Courtesy of Centro Hospitalar de São João, Porto, Portugal.Not for distribution / use in the U.S.

SOMATOM go.Up

Collimation: 32 x 0.7 mmScan time: 5 sScan length: 245 mmRotation time: 0.8 sPitch factor: 1.5Scan parameters: Sn110 kV / 94 mAsCTDIvol: 0.2 mGyDLP: 4.35 mGy cm2 mm slice thickness

• Low-dose protocol• 4-years-old child• CARE Child protocols in combination with Tin

Filter for ultra-low dose in pediatrics

Page 70: Computed Tomography SOMATOM go.Up Clinical Cases

Engineering success. Pioneering healthcare. Together.

© Siemens Healthcare GmbH, 2017

Now’s our time

to inspirethe future

of healthcare together

Not for distribution / use in the U.S.