53
Cryogenics & Cryomodules Part 2: Keeping Cold J. G. Weisend II Deputy Head of Accelerator Projects www.europeanspallationsource.se 11 September 2015

Cryogenics & Cryomodules Part 2: Keeping Cold

Embed Size (px)

DESCRIPTION

Overview of Cryogenic Safety Describe Oxygen Deficiency Hazards and their mitigations Describe the hazards and mitigations associated with the excess pressure in cryogenic systems Present general safety guidelines Warning: This is NOT a complete safety class Additional information may be found in monographs, bulletins of the Compressed Gas Association and in the ES&H manuals of the various national labs (Fermilab, JLAB, SLAC etc) See Safety Reference list

Citation preview

Page 1: Cryogenics & Cryomodules Part 2: Keeping Cold

Cryogenics & CryomodulesPart 2: Keeping Cold

J. G. Weisend IIDeputy Head of Accelerator Projects

www.europeanspallationsource.se11 September 2015

Page 2: Cryogenics & Cryomodules Part 2: Keeping Cold

2

Overview of Cryogenic Safety

• Describe Oxygen Deficiency Hazards and their mitigations

• Describe the hazards and mitigations associated with the excess pressure in cryogenic systems

• Present general safety guidelines• Warning: This is NOT a complete safety class• Additional information may be found in monographs,

bulletins of the Compressed Gas Association and in the ES&H manuals of the various national labs (Fermilab, JLAB, SLAC etc) See Safety Reference list

Page 3: Cryogenics & Cryomodules Part 2: Keeping Cold

3

Some Basic Safety Issues

• Cryogenic fluids and their boil off gases are extremely cold and can cause burns (frost bite) and damage to eyes– Always wear safety glasses and protective gloves when handling cryogenic fluids or when

touching things that may have contacted cryogenic fluids• Gases from cryogenic systems can displace oxygen in an area leading to

potentially lethal conditions: Oxygen Deficiency Hazards (ODH)• Large volume expansion between cryogenic liquids and 300 K gases can result in

high pressures if not relieved.• Be aware that if cryogenic fluids come in contact with materials that are aren’t

designed for that temperature it can lead to sudden failure of the material• If surfaces cooled to cryogenic temperatures are exposed to room air the air will

condense on the surface and due to its higher boiling point the condensate will be oxygen rich. – this can result in a significant fire hazard– Always insulate surfaces to keep them above cryogenic temperatures

• Safety considerations should be designed in from the beginning. Retrofits are time consuming & expensive

Page 4: Cryogenics & Cryomodules Part 2: Keeping Cold

4

Volume Changes for Cryogenic Fluidsfrom normal boiling point to 300 K & 1 atm

Page 5: Cryogenics & Cryomodules Part 2: Keeping Cold

5

Oxygen Deficiency Hazards

• Gases used in cryogenic systems such as He, N2, Ar, H2 can displace oxygen in an area causing the area to be unsafe for human life– Any oxygen concentration less that 19.5 % is considered oxygen deficient

(OSHA)• There are several aspects to this problem

– Large volume changes from cryogenic liquids to room temperatures gases– Even small amounts of liquid can be a hazard is the if released into a

small enough volume e.g. small rooms, elevators or cars– Little or no warning of the hazard at sufficiently low O2 concentrations– Consequences can easily be fatal

• This is not just a problem in large cryogenic installations– It can easily be a problem in small labs and university settings – in fact,

complacency in smaller settings may be an added risk factor

Page 6: Cryogenics & Cryomodules Part 2: Keeping Cold

6

Approximate time of useful consciousness for a seated subject at sea level vs % O2

At low enough concentrations you can be unconscious in less that a minute with NO warning

This is one of the things that makes ODH so dangerous & frequently results in multiple fatalities

Page 7: Cryogenics & Cryomodules Part 2: Keeping Cold

7

ODH Safety Basics

• Understand the problem• Determine level of risk

– For each use of cryogenic liquids or inert gases a formal written analysis of the risk ODH posed should be done. The details of this may vary from institution to institution and may be driven by regulatory requirements.

– One technique used by many US laboratories (Fermilab, Jlab, SLAC, BNL) is the calculation of a ODH Fatality Rate. The size of this rate is then tied to a ODH class and each class is linked to specific required mitigations

• Apply mitigations to reduce the risk• Have a plan to respond to emergencies• ALL users of cryogenic fluids no matter how small should analyze

their risk and consider mitigations– At a minimum, everyone should be trained to understand the hazard

Page 8: Cryogenics & Cryomodules Part 2: Keeping Cold

8

ODH Mitigations

• Best solution: Eliminate the hazard by design choices– Reduce inventory of cryogenic fluids & compressed gases– Don’t conduct cryogenic activities in small spaces– Don’t use LN2 underground

• Training– Everyone working in a possible ODH area should be made aware of the

hazard and know what to do in the event of an incident or alarm• This includes periodic workers such as security staff, custodial staff and

contractors• Visitors should be escorted

• Signs– Notify people of the hazard and proper response– Indicate that only trained people are authorized to be there

Page 9: Cryogenics & Cryomodules Part 2: Keeping Cold

9

ODH Mitigations

• Ventilation systems to increase air exchange and reduce the possibility of an oxygen deficient atmosphere forming– Warning If this approach is taken, the ventilation system must now be treated as

a safety system with appropriate controls and redundancies• What happens during maintenance or equipment failure?• How do you know ventilation system is working?

• ODH Monitors and Alarms– A very common and effective mitigation. Commercial devices exist.– Indicates when a hazard exists– Very valuable in showing if a area has become dangerous during off hours– Alarms generally set to trip at 19.5% Oxygen – Alarms should include lights & horn as well as an indicator at entrance to area– Alarms should register in a remote center (control room or fire dept) as well– As a safety system it requires appropriate controls & backups (UPS, redundancy

etc)– In some cases personal monitors will add additional safety

Page 10: Cryogenics & Cryomodules Part 2: Keeping Cold

10

Response to ODH Alarms & Emergencies

• In the event of an alarm or other indication of a hazard immediately leave the area

• Do not reenter the area unless properly trained and equipped (e.g. supplementary air tanks)– Don’t just run in to see what the problem is

• Only properly trained and equipped professionals should attempt a rescue in an ODH situation

• Response to alarms should be agreed upon in advance, documented and be part of training

Page 11: Cryogenics & Cryomodules Part 2: Keeping Cold

11

A Common Safety Issue in Cryostat Design is Over Pressurization

• This hazard results from the large volume expansion that occurs when a cryogenic liquid is converted to 300 K gas

• As a result large pressure rises are possible• This could easily result in material failures, and

explosive bursting

Page 12: Cryogenics & Cryomodules Part 2: Keeping Cold

12

The Solution is to Always Install Pressure Relief Devices• Redundant devices should be used. Typically relief valve + burst disc• The device should be set to open at or below the MAWP of the vessel• Valves should be sized for worst case scenarios• What is the worst case?

– Typically this is taken as a catastrophic failure of the vacuum space – thus admitting air into the cryostat.

– The air condenses onto the cold surface and in doing so deposits heat into the liquid cryogen causing rapid boiling and expansion

– This effect ( ~ 0.6 - 4 W/cm2) is far more significant than the increased heat leak from 300 K due to the spoiling of the vacuum

– Note that 0.6 W/cm2 assumes the presence of MLI on cold surfaces• This is an important secondary use of MLI

• Valves should be tested & certified before installation• Ensure that there are no trapped volumes. Remember – valves may leak

or be operated incorrectly and cryogenic systems may warm up unexpectedly (vacuum spaces need relief valves as well)

• Never store cryogenic liquids in a container without a relief system

Page 13: Cryogenics & Cryomodules Part 2: Keeping Cold

13

Relief Devices

• Do not put shut off valves between system and relief valves

• Ask What If ? questions

Page 14: Cryogenics & Cryomodules Part 2: Keeping Cold

14

General Cryogenic Safety Guidelines

• Perform a formal hazard analysis on any cryogenic system or process prior to the start of work and design. Identify the hazards and how you will mitigate them. Ask “What If” questions. Keep in mind that equipment breaks, cryogenic fluids can turn rapidly into gas, valves leak or are improperly operated and vacuums can fail. This analysis should be done regardless of the size or complexity of the cryogenic system.

• Design safety into equipment and processes from the beginning. Adding safety features at the end of the design process can be expensive, time consuming and may result in hazards being missed. Note that it is always better to remove a hazard via engineering design than to try to mitigate the hazard.

• Review, Review, Review. Everyone, even experts, miss things or make mistakes. Having the safety of your cryogenic system reviewed by other people, be they other members of your team, external consultants or formal review committees is key to improving the likelihood of a safe system.

• Whenever working with cryogenic liquids or inert gases, no matter how small the amount, always consider the possibility that an Oxygen Deficiency Hazard (ODH) may exist. Either prove by analysis that such a hazard doesn’t exist or apply appropriate design changes or mitigations to eliminate or reduce the hazard. ODH issues are particularly dangerous due to the large volume of gas evolved by even small amounts of cryogenic liquid and by the fact that in low enough oxygen concentrations, the first physiological symptom can be sudden unconsciousness, followed rapidly by coma and death.

• Take advantage of existing regulations and industry codes and standards to develop a safe cryogenic system.• Never disable or remove safety devices, relief valves or operate systems outside their operating parameters.• Only use materials at cryogenic temperatures that have been proven to work at those temperatures. Keep in mind during the hazard analysis that

materials that nominally operate at room temperature (such as the outer walls of vacuum vessels) could reach cryogenic temperatures in certain failure modes.

• Ensure that everyone working with or around cryogenic systems, even casual or periodic users, have the appropriate level of training in cryogenic and ODH safety.

• Always use required Personal Protective Equipment and follow required operating procedures. Taking shortcuts frequently leads to accidents.• Be aware that surfaces exposed to air and colder than 77 K will start to condense the air and that the condensate will be oxygen rich (due to the

higher boiling point of oxygen). This can lead to an enhanced flammability hazard. Insulate surfaces to prevent air condensation.• Working with hydrogen, oxygen and LNG presents additional flammability hazards not present with inert cryogenic fluids and thus requires

additional specialized knowledge. Make sure that such expertise is present in the designing, reviewing and operating of these systems.• Take advantage of existing information & expertise. There are significant resources available on cryogenic safety. These include books, pamphlets,

research papers, codes & standards, websites and conferences. Cryogenic safety standards from other institutions such as Fermilab may also provide valuable information. The use of experts from other institutions as external reviewers or consultants may be beneficial.

Page 15: Cryogenics & Cryomodules Part 2: Keeping Cold

15

Cryostats and Cryomodules

• What is a cryostat?– A device or system for maintaining objects at cryogenic temperatures.

• Cryostats that contain SRF systems are frequently known as a cryomodules. Design approaches are the same

• Cryostat design involves many subtopics:– Development of requirements– Materials selection– Thermal insulation – Support systems– Safety – Instrumentation

• One of the best ways to learn about cryostat design is through examples• There are many different types of cryostats with differing requirements

– The basic principles of cryostat design remain the same– Before we can do anything else we have to define our requirements

Page 16: Cryogenics & Cryomodules Part 2: Keeping Cold

16

Cryostat Requirements

• Maximum allowable heat leak at various temperature levels– This may be driven by the number of cryostats to be built as well as by the

impact of large dynamic heat loads (SCRF or target cryostats)• Alignment and vibration requirements

– Impact of thermal cycles– Need to adjust alignment when cold or under vacuum?– Alignment tolerances can be quite tight (TESLA: +/- 0.5 mm for cavities and

+/- 0.3 mm for SC magnets)• Number of feed throughs for power, instrumentation, cryogenic fluid

flows, external manipulators• Safety requirements (relief valves/burst discs)• Size and weight

– Particularly important in space systems• Instrumentation required

– Difference between prototype and mass production

Page 17: Cryogenics & Cryomodules Part 2: Keeping Cold

17

Cryostat Requirements

• Ease of access to cryostat components• Existing code requirements (e.g. TUV or ASME)• Need, if any, for optical windows Presence of ionizing radiation or magnetic fields

Magnetic shielding is a important part of SRF cryomodules Expected cryostat life time Will this be a one of a kind device or something to be

mass produced? Schedule and Cost

• This should be considered from the beginning

All Design is Compromise

Page 18: Cryogenics & Cryomodules Part 2: Keeping Cold

18

Cryostat Materials

• Materials for cryostats must function properly at cryogenic temperatures• Material properties change greatly as a function of temperature between

300 K and cryogenic temperatures – such variations must be taken into account in the cryostat design

• Typical materials for cryostats include:– Austenitic stainless steels– Aluminum– Copper– G-10 (Fiber reinforced plastic)– Glass – Quartz (for viewing ports)

• As will be seen, the location of materials used in a cryostat is optimized to reduce heat leak.

Page 19: Cryogenics & Cryomodules Part 2: Keeping Cold

19

Thermal Insulation of CryostatsThree Ways to Transfer Heat

• Conduction– Heat transfer through solid material

• Convection– Heat transfer via a moving fluid

• Natural or free convection – motion caused by gravity (i.e. density changes)

• Forced – motion caused by external force such as a pump• Radiation

– Heat transferred by electromagnetic radiation/photons• There is no such thing as a perfect insulator – though we can design systems with very

small heat leaks• All matter above 0 K radiate heat

– Remember we can’t get to 0 K – 3rd Law of Thermodynamics though we can get vanishingly close

• Heat flows from high temperature to low– Heat leaks in, cold doesn’t leak out

Page 20: Cryogenics & Cryomodules Part 2: Keeping Cold

20

Conduction Heat Transfer

• Fundamental Equation – The Fourier Law in one dimension

• If we assume constant cross section we get:

• Reduce conduction heat leak by:– Low conductivity material: make K(T) small– Reduce cross sectional area: make A small– Increase length: make L large– For a given TC make TH smaller: i.e. use intermediate temperature heat

intercepts• You still have heat leak from 300 K to this intermediate temperature but remember

Carnot, It’s more thermodynamically efficient to remove heat at higher temperatures

xTxATKQ

H

C

T

T

dTTKLAQ )(/

Page 21: Cryogenics & Cryomodules Part 2: Keeping Cold

21

Thermal Conductivities of Metals

From Lakeshore Cryotronics

Variation of Thermal conductivity with temperature can be addressed with Thermal Conductivity integralsSee backup slides

Page 22: Cryogenics & Cryomodules Part 2: Keeping Cold

22

Convection Heat Transfer

• Fundamental Equation: Newton’s law of cooling Q = hA(Tsurface – Tfluid)

where h is the heat transfer coefficient and is a function of Re, Pr, geometry etc depending on the situation

• In cryogenics we eliminate convection heat leak in cryogenic systems by “simply” eliminating the fluid – vacuum insulation

• Using vacuum insulation to create vessels capable of storing cryogenic liquids was first done by James Dewar – who liquefied hydrogen– Such vessels are frequently called dewars – though not always, more

later– Thermos bottles are a simple example of this approach

Page 23: Cryogenics & Cryomodules Part 2: Keeping Cold

23

Vacuum Insulation

• How much vacuum is enough?– This of course depends on the heat leak requirements but generally we want to be

below 10-5 torr If we maintain this level or better we can generally neglect the convection heat leak for most applications.

• Cryogenic Engineering, Flynn (1997) has a good discussion of calculating heat leak due to residual gas pressure

• Cryopumping– At cryogenic temperatures almost all common gases condense and freeze onto the

cold surface. Typically, we’ll see that once surface are cooled to ~ 77 K the isolation vacuum will drop to the 10-8 torr or better range if the system is leak tight and doesn’t have significant outgassing

– But don’t just start cooling with everything at room pressure• Heat leak will likely be too high• Safety hazards due to enrichment of LOX on cold surfaces• Large amounts of condensed gases in vacuum space can lead to other problems including rapid

pressure rise upon warming and possible solid conduction• Best practice is to be at least 10-3 torr before cooling, lower pressures are better but there may be

operational tradeoffs

Page 24: Cryogenics & Cryomodules Part 2: Keeping Cold

24

Outgassing and Getters

• All materials outgas into a vacuum. This can raise the pressure in a sealed vacuum space

• Reduce outgassing by:– Minimize amount of polymers, wire insulation, FRP etc – difficult– Keep vacuum surfaces as clean as possible. Remove any oil or

cutting fluid, wear gloves etc.• Getters: materials (such as activated charcoal or silica gel)

inserted into vacuum spaces to remove residual gas at low pressures

• In cryogenic systems, getters may be useful in removing residual gas and passively managing small leaks

Page 25: Cryogenics & Cryomodules Part 2: Keeping Cold

25

Radiation Heat Transfer

• Frequently the largest source of heat leak to cryogenic systems• Fundamental Equation: Stefan-Boltzmann Law – energy emitted from an

ideal black body: Eb = sT4 where s = 5.67x10-8 W/m2K4

– Real world Assumptions:• Emissivity (e) << 1 and independent of wavelength (grey body)

• Two parallel infinite plates: Radiative heat flux (W/m2)

Eq. A

• Frequently in cryogenic systems e1 ~ e2 << 1 then Eq. A becomes:

Eq. B

424

2121

211TTqr

seeee

ee

42412TTqr

se

Page 26: Cryogenics & Cryomodules Part 2: Keeping Cold

26

Radiation Heat Transfer

» Two long concentric cylinders or concentric spheres (1 represents the inner cylinder): the radiative heat flux (W/m2) on the inner cylinder is

Eq. C

» Note as is frequently the case in cryogenics, if the spacing between the cylinders is small compared to the inner radius (i.e. A1 ~ A2 ) Eq. C becomes Eq. A

112

12

1

1

42

4

11

ee

s

AA

TTq

Page 27: Cryogenics & Cryomodules Part 2: Keeping Cold

27

Radiation Heat Transfer

• Looking at Eq. A, How do we reduce the radiation heat transfer?

• We could reduce the emissivity (e)– This is done in some cases;

using either reflective tape or silver plating

– Better below 77 K– It’s also part of MLI systems

(see below)– We have to consider tarnishing– May be labor intensive

Page 28: Cryogenics & Cryomodules Part 2: Keeping Cold

28

Radiation Heat Transfer

• Another way to reduce radiation heat transfer is to install intermediate actively cooled radiation shields that operate at a temperature between 300 K and the lowest cryogenic temperature. This has several advantages.– It greatly reduces the heat load to the lowest temperature level

• Assume parallel plates with e = 0.2 • then from Eq. B q ( 300 K – 4.2 K ) = 46 W/m2 while q (77 – 4.2) = 0.2 W/m2

– It allows heat interception at higher temperatures & thus better Carnot efficiency– Such an actively cooled shield provides a convenient heat intercept for supports,

wires etc to reduce conduction heat leak.• Shields may be cooled by

– Liquid baths ( LN2)– Vapor boil off from stored liquid – common in LHe storage dewars– Cooling flows from refrigeration plants– Conductive cooling via small cryocoolers

Page 29: Cryogenics & Cryomodules Part 2: Keeping Cold

29

Thermal Radiation Shields

• Uncooled thermal radiation shields placed in a vacuum space between the warm & cold surfaces also help reduce the thermal radiation heat leak

• It can be shown (with the grey approximation and equal emissivities) that with N shields thermal radiation heat transfer is given by:

TT LHNq 44

21

se

This is the motivation behind Multilayer Insulation

Page 30: Cryogenics & Cryomodules Part 2: Keeping Cold

30

MultiLayer Insulation

• Also referred to as superinsulation • Used in the vacuum space of many cryostats (10-5 torr or better for best

performance) – can be used instead of or with actively cooled thermal shields or baths

• Consists of highly reflective thin sheets with poor thermal contact between sheets.– Made of aluminized Mylar ( or less frequently Kapton)– May include separate non conducting mesh– May use Mylar aluminized on only one side and crinkled to allow only point

contacts between sheets– Frequently perforated to allow for better pumping

• Can be made up into blankets for ease of installation• Don’t pack MLI too tightly. Doing so will cause increased conduction heat leak

negating the value of the MLI

Page 31: Cryogenics & Cryomodules Part 2: Keeping Cold

31

MultiLayer Insulation

• Great care must be taken with seams, penetrations and ends.– Problems with these can dominate the heat leak

Page 32: Cryogenics & Cryomodules Part 2: Keeping Cold

32

MLI Example from LHC cryostats

“SERIES-PRODUCED HELIUM II CRYOSTATSFOR THE LHC MAGNETS: TECHNICAL CHOICES,INDUSTRIALISATION, COSTS”A. Poncet and V. ParmaAdv. Cryo. Engr. Vol 53

Page 33: Cryogenics & Cryomodules Part 2: Keeping Cold

33

Cryomodule Examples: Elliptical Cavities

• Due to the cavity shapes and the general efficiency of using cylinders as pressure/vacuum vessels, these systems are almost always cylindrical in geometry– Cold Masses are loaded in from the ends– Cryogenic piping connections go out the ends– Common design features with accelerator magnet systems– Can be further divided into “Space Frame” designs” & “Internal Transfer Line” designs

Page 34: Cryogenics & Cryomodules Part 2: Keeping Cold

34

Space Frame Based Designs

• Cold Mass Components are installed into a Space Frame that is inserted into a cylindrical vacuum vessel.

• Space Frame may also contain thermal shield and MLI systems

• Generally alignment is done warm and system is designed to not require realignment once cold

• Very suitable for systems in which the cryo segmentation is done per module with a parallel transfer line

• Higher cost and heat loads tend to limit this approach to small & medium sized machines ( ~ 50 CMs)

• Examples include: ESS, SNS and Jlab 12 GeV Upgrade

Page 35: Cryogenics & Cryomodules Part 2: Keeping Cold

35

JLab 12 GeV Upgrade Cryomodule

• Based on successful SNS design - a total of 10 CMs were needed• Cavities at 2.1 K, thermal shield at 50 K – no s/c magnets present• Cold Mass tied to space frame via nitronic rods• Space frame rolls into vacuum vessel• Despite extensive SNS experience, design changes (mainly in cold mass)

were required after 1st prototype – value of prototyping

Page 36: Cryogenics & Cryomodules Part 2: Keeping Cold

Similar to CEBAF/SNS cryomodule concept with 4 cavities per cryomodule Common design for medium (6 cells) and high beta (5 cells) cavities

Proton Beam

Jumper connection

Cold to warm transition

Magnetic shielding

Power couplerCavity with Helium tank

2 Phase He pipe

Thermal shielding

Regulation He valve

Intercavities belows

Vacuum valve

Spaceframe support Heat exchanger

Alignement fiducial

36August 2015

ESS Elliptical Cavities & Cryomodule

Page 37: Cryogenics & Cryomodules Part 2: Keeping Cold

37

Internal Transfer Line Based Designs

• Cryogenic distribution piping is contained with cryomodules - limited external transfer lines

• Approach commonly used where infrequent cryo-segmentation is desired

• Lower static heat leak & lower cost• More difficult maintainability (larger number of CMs to thermal

cycle)• Suitable for systems containing large numbers of cryomodules

(ILC, XFEL, LCLS II)• Designs frequently use large diameter distribution lines as the

structural backbone for the cold mass

Page 38: Cryogenics & Cryomodules Part 2: Keeping Cold

38

ILC Cryomodule Design

• Based on extensive TESLA development program• Cavities at 2 K, thermal shields at 5 K and 40/80 K, s/c magnets at 4.5 or 2 K• Design uses large GRP (whose size is dictated by allowable pressure drop) as the

structural backbone for the cold mass• FRP posts connect GRP to vacuum vessel• Alignment is done warm. System is designed to keep alignment when cooled• Design drivers

– ~ 2000 cryomodules required– Low heat leak and low cost– Infrequent cryo-segmentation– Ease of mass production

• Significant amounts of prototyping (FLASH) • 3 generations of design with continuous improvement• Design being used on XFEL and LCLS II (with mods)

Page 39: Cryogenics & Cryomodules Part 2: Keeping Cold

39

3rd Generation ILC Cryomodule

Page 40: Cryogenics & Cryomodules Part 2: Keeping Cold

40

Cryomodules with QWR and HWR Cavities

• Low to medium beta cavities typically used in heavy ion machines IISAC-II, (TRIUMF) ATLAS Upgrade (ANL), RAON & FRIB (MSU)

• Due to cavity geometry, particularly the QWRs, these cryomodules have a rectilinear geometry

• The designs can be further divided into “Top Supported” or “Bottom Supported” Designs

• Cryogenic lines go out either the top or bottom of assembly• So far only used in systems where cryo-segmentation is at

the individual CM level – external transfer line

Page 41: Cryogenics & Cryomodules Part 2: Keeping Cold

41

Top Down Designs

• All components are suspended from a top plate via metallic rods– Length between top plate and connection to cavity mounting beam

can be maximized to reduce heat leak• System is lowered into a simple bath tub style vacuum vessel• Possibility of better vibration response ( longer path between

floor and cavity• Cryogenic and other feedthroughs go through the top plate

though coupler and tuner feedthroughs may be through the side or bottom

• Most common type of HWR/QWR cryomodule

Page 42: Cryogenics & Cryomodules Part 2: Keeping Cold

42

ISSAC - II Cryomodule (TRIUMF)

• Permits adjustment of alignment while system is cold

• 4.5 K cavities & solenoid• Common cavity & isolation

vacuum - thus no MLI and higher heat loads

• 8 Cryomodules now in use at TRIUMF

“Design of a Medium-Beta Cryomodule for IISAC-II Superconducting Heavy Ion Accelerator” G. Stanford et al., Adv. Cryo. Engr. Vol 49A (2004)

Page 43: Cryogenics & Cryomodules Part 2: Keeping Cold

43

ALTAS Upgrade Cryomodule

• Top down design similar to IISAC-II but separate cavity & isolation vacuum – thus clean room assembly needed but higher performance cavities possible

• 7 QWR plus 1 solenoid all at 4.5 K

“Commissioning of the ATLAS Upgrade Cryomodule”, P.N. Ostroumov et al. Proceedings of HIAT09 (2009)

Page 44: Cryogenics & Cryomodules Part 2: Keeping Cold

44

Bottom Supported Design

• Here the components, instead of being hung from the top plate, are supported by FRG posts from a bottom plate.

• Vacuum vessel lid is lowered down on top• This was chosen for FRIB as a result of the alignment issues with ReA3 – but

this approach also results in other advantages– Cryolines can go out the bottom plate allowing for U-tubes in the constricted FRIB

tunnel– Simpler installation of thermal & magnetic shields and MLI– Lower CG and easier transport– Lower risk insertion of couplers through the bottom plate– Lack of top plate means simpler assembly process

• Cons– Unusual design - more prototyping and testing required– High risk vacuum seal – welding is backup solution

Page 45: Cryogenics & Cryomodules Part 2: Keeping Cold

45

FRIB HWR Cryomodule

Courtesy F. Casagrande FRIB

Page 46: Cryogenics & Cryomodules Part 2: Keeping Cold

46

FRIB Prototype CM

Tests verified both Alignment and Heat Leak Performance

Courtesy F. Casagrande FRIB

Page 47: Cryogenics & Cryomodules Part 2: Keeping Cold

47

Example of Bottom Support in a Cylindrical Geometry

Proposed Fermilab Single Spoke ResonatorEach component has its own support post – very “stiff design”

Page 48: Cryogenics & Cryomodules Part 2: Keeping Cold

48

Back Up Slides

Page 49: Cryogenics & Cryomodules Part 2: Keeping Cold

Q = – G ( 2 – 1 )Thermal Conductivity

Integrals The strong temperature dependence of K makes heat transfer

calculations difficult The solution is frequently to use thermal conductivity integrals The heat conduction equation

Is written as:

12 GQ

xTxATKQ

8/26/14

Cryostat Design - J.G. Weisend II

Slide 49

Page 50: Cryogenics & Cryomodules Part 2: Keeping Cold

G is the geometry factor

is the thermal conductivity integral

Thermal ConductivityIntegrals

i = K ( T ) dT

0Ti

2

1

1x

x xAdx

G

dTTKiT

i 0

8/26/14

Cryostat Design - J.G. Weisend II

Slide 50

Page 51: Cryogenics & Cryomodules Part 2: Keeping Cold

Advantages:• Simple• Only end point temperatures are important. (assuming there are no

intermediate heat sinks) The actual temperature distribution is not. • Thermal conductivity integrals have been calculated for many engineering

materials• This is quite useful for heat leak calculations

Thermal ConductivityIntegrals

8/26/14

Cryostat Design - J.G. Weisend II

Slide 51

Page 52: Cryogenics & Cryomodules Part 2: Keeping Cold

Thermal Conductivity Integrals of Metals

From Handbook of Cryogenic Engineering, J. Weisend II

(Ed)

8/26/14

Cryostat Design - J.G. Weisend II

Slide 52

Page 53: Cryogenics & Cryomodules Part 2: Keeping Cold

From Lakeshore Cryotronics

Thermal Conductivity Integrals of Metals & Nonmetals

8/26/14

Cryostat Design - J.G. Weisend II

Slide 53