We bring quality to light. Data sheet The goniometer system DMS 903 is ideal for photometric and colorimetric analysis of emissive, transmissive, reflective, and transflective displays and modules up to a size of 72" diagonal. The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different viewing angles and variable electrical driving. The measurement system is perfectly suitable for all applications in research and development, as well as quality control. The comprehensive range of illumination devices and temperature-control chambers makes the DMS 903 an ideal system for determining the characteristics of displays under various environmental and illumination conditions. Product highlights ì Measurement of emissive, transmissive, reflective and transflective displays up to 72" diagonal size ì 6 motorized axes for directional and lateral scanning, adjustment of height and light source inclination ì Comprising electrical driving, illumination, computer and software ì Fast photometer for analysis of temporal luminance variations ì DMSControl and ViPer software for automated measurement sequences and evaluation DMS 903 Goniometer System for complete display characterization

DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.Data sheet

The goniometer system DMS 903 is ideal for photometric and colorimetric analysis of emissive, transmissive, reflective, and transflective displays and modules up to a size of 72" diagonal.

The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different viewing angles and variable electrical driving. The measurement system is perfectly suitable for all applications in research and development, as well as quality control.

The comprehensive range of illumination devices and temperature-control chambers makes the DMS 903 an ideal system for determining the characteristics of displays under various environmental and illumination conditions.

Product highlights

ì Measurement of emissive, transmissive, reflective and transflective displays up to 72" diagonal size

ì 6 motorized axes for directional and lateral scanning, adjustment of height and light source inclination

ì Comprising electrical driving, illumination, computer and software

ì Fast photometer for analysis of temporal luminance variations

ì DMSControl and ViPer software for automated measurement sequences and evaluation

DMS 903Goniometer System

for complete display characterization

Page 2: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

2

Options

ì Spectrometer for radiometric, photometric and colorimetric analysis

ì Directional light sources for measurement of reflectance factor and in-plane BRDF

ì Forced air temperature control of the display under test

ì Interface for various video test-pattern generators

Measurements

ì Angular scan: variation of luminance, contrast and chromaticity with viewing direction for evaluation of directional uniformities and BRDF

ì Area scan: scanning of display surface area for lateral uniformity evaluation

ì EOC: variation of luminance, contrast and chromaticity with electrical driving for evaluation of EOTF-characteristics (e.g. gamma)

ì Transients: luminance versus time for e.g. response time evaluation

ì Reflectance characteristics under diffuse and directional illumination

ì Customer specific measurements defined with DCI scripting sequences

System components

The following components and accessories are included in the DMS 903 standard scope of delivery, if not explicitly marked as optional components or accessories.

Temperature control:ì HCS-4 (optional): air-forced temperature control

device, temperature range -30 °C to +85 °C, with integrated diffuse reflective illumination; optional extension: internal motorized stage for xy-scanning

Light sources and illumination systems:ì HEL-4: light source with halogenic incandescent lampì LC-8 (optional): xenon light source for higher light

outputì DHS: diffusing hemisphere for evaluation of contrast

of reflective displaysì DHS-HC (optional): diffusing hemisphere for

evaluating the contrast of reflective displays; illumination geometry similar to HCS-3

ì PID (optional): collimated parallel beam illuminationì VADIS (optional): variable aperture diffuse illumination

source for e.g. scattering characterization of surfaces (BRDF), measurement of reflectance factor

Light measuring devices:ì PMT 3: fast photometer with photomultiplier tube,

bandwidth 100 kHz, 500 kHz or 2 MHzì CAS 140CT (optional): model VIS; 360 to 830 nm;

1024 pixel back-illuminated CCD detector; 2.2 nm spectral resolution (100 μm slit); 0.5 nm/pixel data point interval (CAS140CT-151)

Page 3: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

3

DUT driving:ì ASG: arbitrary signal generator with 1 or 4 analog

output channels (depending on basic system variant); standard and user defined signal waveforms; for direct addressing, passive matrix driving and control of TFT arrays (gate and data signals), etc.; power amplifiers are available for different voltage and current ranges

ì Source meter (optional): programmable current or voltage source with digital multimeter, in particular for current driven display devices like OLED displays; can be combined with a multiplexer (for driving and measurement of displays with multiple segments and matrix displays)

ì DCI (optional): device control interface for scripting and communication with various external devices, e.g. video pattern generators, module drivers, power supplies, etc. via RS232, GPIB, USB, or TCP/IP

Software:ì DMSControl: administration, configuration and control

of the instrument, definition and management of measurement sequences

ì Visual performance evaluation (ViPer): evaluation and representation of measurement results (on- and off-line), data exchange with other software (e.g. spreadsheet applications) and report generation (e.g. PDF documents)

ì Report generator (optional): user configurable data export from DMSControl to spreadsheet software

Footprint

Scope of delivery

The basic system comprises the following components:

ì DMS 903 goniometer mechanismì Microscope viewing opticsì Photometer PMT 3/100ì GRU with ASG-1 (one channel) or ASG-4

(4 channel) and power amplifier PA 3/60 (DMS 903-1 or DMS 903-4)

ì HEL-4 light source for reflective and transmissive illumination

ì PC with tableì Software package: DMSControl, ViPerì Light shielding box

Technical dataDMS 903

Axis Scan range Resolution

Inclination Theta 0 … 80° 0.1°

Rotation Phi 0 … 360° 0.1°

Reflective illumination 0 … 70° 0.1°

x translation -300 … +300 mm 0.1 mm

y translation -225 … +225 mm 0.1 mm

z translation 0 … 280 mm 0.1 mm

DUT dimensions

Maximum size (L x W)1300 mm x 1600 mm (72" diagonal for a 16:9 panel) for center measurement; 650 mm x 1100 mm (50" diagonal for a 16:9 panel) for full x-y scanning range

Maximum thickness 280 mm

Dimensions (Length x Width x Height)1

Mechanics of DMS 903 1900 mm x 1500 mm x 1860 mm

Electronics rack 750 mm x 600 mm x 1250 mm

Mass1

Mechanism 700 kg

Electronics rack 120 kg

Facility requirements

Mains voltage 200 V – 240 VAC (230 V nominal), 50 Hz, 1 phase

Max. mains current / fuse 30 A

Power consumption Max. 1500 W (incl. PC, CCD and 2 x HEL-4)

1 All sizes and masses approximate

DMS 903 mechanism

Electronics rack Thermo-

stream

PCtable

2000

mm

(mac

hine

& d

oor)

5000 mm (incl. opration & maintainance space)

4000 mm

3000

mm

Page 4: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

Instrument Systems GmbHNeumarkter Str. 83 81673 Munich, Germany Phone: +49 89/45 49 43-0 Fax: +49 89/45 49 43-11 E-mail: [email protected] www.instrumentsystems.com

We bring quality to light.

Ambient operating conditions

Ambient temperature 23 °C (±5 °C)

Humidity 0 – 60 % rel. humidity

Photometer PMT 3

Bandwidth 100 kHz (standard), 500 kHz or 2 MHz (optional)

Measurement spot diameter 0.5, 1, 2, 3, 5, 7, 10 mm

Spectrometer CAS 140CT

Spectral range2 360 … 830 nm (usable spectral range depending on software used)

Spectral resolution 2.2 nm

Data point interval3 0.5 nm

Measurement spot diameter 0.5, 1, 2, 3, 5, 7, 10 mm

2 DMSControl wavelength range: 380 … 800 nm3 DMSControl data point interval: 1 nm

Order information

Order number Description

Basic system

DMS-903-1

Model DMS 903-1 comprising: Goniometer with 6 motorized axes; microscope viewing optics; fast photometer PMT 3/100; 1 x HEL-4 light source; transmissive illumination; GRU with ASG-1 (1 channel) and power amplifier PA 3/60; light shielding enclosure; PC with table; DMSControl and ViPer software

DMS-903-4

Model DMS 903-4 comprising: Goniometer with 6 motorized axes; microscope viewing optics; fast photometer PMT 3/100; 1 x HEL-4 light source; transmissive illumination; GRU with ASG-4 (4 channels) and power amplifiers PA 3/60; light shielding enclosure; PC with table; DMSControl and ViPer software

Options - Temperature control

DMS-3-50 HCS-4-32 temperature chamber with ThermoStream control unit

DMS-3-52 HCS-4-32m motorized positioning stage for HCS-4-32 temperature chamber

Options - Light sources and illumination systems

DMS-5-10 HEL-4 light source with 150 W tungsten halogen lamp

DMS-5-12 LC-8 xenon light source for higher light output

DMS-5-21 PID-903 parallel illumination device (collimated beam)

DMS-5-31 VADIS-903 variable aperture diffuse illumination source for reflective measurement

DMS-5-42 DHS-HC9 diffusing hemisphere for evaluating the contrast of reflective displays; illumination geometry similar to HCS-3 source

Options – Light measuring devices

DMS-1-10CCD spectrometer set for DMS systems; incl. CAS 140CT VIS model (CAS140CT-151) 360 nm – 830 nm (usable spectral range depending on software used)

Options – DUT driving

DMS-2-10 Keithley 2400 series source meter set for DMS, including cabling and interface

Options – Software

DMS-SW-35 DCI device control interface for external devices (for DMSControl)

DMS-SW-40 Report generator for data export to Excel spread sheetsd

s_d

ms9

03_e

n_V

2.3

Page 5: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.Data sheet

The DMS 803 offers optimum convenience for angle-dependent electrooptical characterization of small to mid-size displays of any kind (emissive, transmissive, reflective or transflective displays).

It is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different viewing angles and variable electrical driving. The measurement system is perfectly suitable for all applications in research and development, as well as quality control.

The comprehensive range of illumination devices and temperature-control chambers makes the DMS 803 an ideal system for determining the characteristics of displays for mobile applications (e.g. smartphones or tablet PCs).

Product highlights

ì 6 motorized axes fordirectional and lateralscanning, adjustment ofheight and light sourceinclination

ì Comprising electrical driving,illumination, computer andsoftware

ì Fast photometer for analysisof temporal luminancevariations

ì DMSControl software forautomated measurementsequences

ì ViPer software for analysis,evaluation and visualizationof measurement data

DMS 803Goniometer System

for complete display characterization

Page 6: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

2

Options

ì Spectrometer for radiometric, photometric and colorimetric analysis

ì Directional light sources for measurement of reflectance factor and in-plane BRDF

ì Forced air temperature control of the display under test

ì Interface for various video test-pattern generators

Measurements

ì Angular scan: variation of luminance, contrast and chromaticity with viewing direction for evaluation of directional uniformities and BRDF

ì Area scan: scanning of display surface area for lateral uniformity evaluation

ì EOC: variation of luminance, contrast and chromaticity with electrical driving for evaluation of EOTF-characteristics (e.g. gamma)

ì Transients: luminance versus time for e.g. response time evaluation

ì Reflectance characteristics under diffuse and directional illumination

ì Customer specific measurements defined with DCI scripting sequences

System components

The following components and accessories are included in the DMS 803 standard scope of delivery, if not explicitly marked as optional components or accessories.

Temperature control:ì HCS-3 (optional): forced air temperature control

device, temperature range -35 °C to +85 °C, with integrated diffuse reflective illumination

ì HCS-3A (optional): modified version of HCS-3 improved for easier handling, temperature range -40 °C to +85 °C, adjustable DUT fixture

Light sources and illumination systems:ì HEL-4: light source with halogenic incandescent lampì LC-8 (optional): xenon light source for higher light

outputì DHS: diffusing hemisphere for evaluation of contrast

of reflective displaysì DHS-HC (optional): diffusing hemisphere for

evaluating the contrast of reflective displays; illumination geometry similar to HCS-3

ì PID (optional): collimated parallel beam illuminationì VADIS (optional): variable aperture diffuse illumination

source for e.g. scattering characterization of surfaces (BRDF), measurement of reflectance factor

ì PLS (optional): LED point-light source for measurement of reflectance characteristics including BRDF

ì SDR (optional): integrating sphere for measurement of diffuse reflectance factor, specular component excluded or included (d/8, SCE or SCI)

Light measuring devices:ì PMT 3: fast photometer with photomultiplier tube,

bandwidth 100 kHz, 500 kHz or 2 MHzì CAS 140CT (optional): Model VIS; 360 to 830 nm;

1024 pixel back-illuminated CCD detector; 2.2 nm spectral resolution (100 μm slit); 0.5 nm/pixel data point interval (CAS140CT-151)

Page 7: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

3

DUT driving:ì ASG: arbitrary signal generator with 1 or 4 analog

output channels (depending on basic system variant);standard and user defined signal waveforms; fordirect addressing, passive matrix driving and controlof TFT arrays (gate and data signals), etc.

ì Source meter (optional): programmable current orvoltage source with digital multimeter, in particularfor current driven display devices like OLED displays;can be combined with a multiplexer (for driving andmeasurement of displays with multiple segments andmatrix displays)

ì DCI (optional): device control interface for scriptingand communication with various external devices,e.g. video pattern generators, module drivers, powersupplies, etc. via RS232, GPIB, USB, or TCP/IP

Software:ì DMSControl: administration, configuration and control

of the instrument, definition and management ofmeasurement sequences

ì Visual performance evaluation (ViPer): evaluationand representation of measurement results (on- andoff-line), data exchange with other software (e.g.spreadsheet applications) and report generation (e.g.PDF documents)

ì Report generator (optional): user configurable dataexport from DMSControl to spreadsheet software

Footprint

Scope of delivery

The basic system comprises the following components:

ì DMS 803 goniometer mechanismì Measurement microscope with viewfinderì Photometer PMT 3/100ì GRU with ASG-1 or ASG-4 and power amplifier

PA 3/60 (DMS 803-1 or DMS 803-4)ì DHSì 2 x HEL-4 light source for reflective and transmissive

illuminationì PC with tableì Software package: DMSControl, ViPerì Light shielding box

Technical dataDMS 805

Axis Scan range Min. step size

Inclination Theta 0 … 90°1 0.01°

Rotation Phi 0 … 360° 0.01°

Reflective illumination 0 … 90°1 0.01°

x translation -50 … +50 mm 0.01 mm

y translation -50 … +50 mm 0.01 mm

z translation 0 … 250 mm 0.01 mm

Sphere of confusion (all axes) 100 µm

DUT dimensions

Maximum diagonal600 mm (23" diagonal) for center measurement; 459 mm (18" diagonal) for full x-y scanning range

Maximum thickness 250 mm

Dimensions2

DMS mechanism (L x W x H) 1430 mm x 980 mm x 1900 mm (incl. light shielding box)

Light shielding box 1430 mm x 980 mm x 1050 mm

Electronics rack 700 mm x 560 mm x 1250 mm (or 1050 mm depending on accessories)

Mass2

Mechanism 320 kg

Electronics rack 105 kg

Light shielding box 85 kg

Facility requirements

Mains voltage 200 V – 240 VAC (230 V nominal), 50 Hz, 1 phase

Max. mains current / fuse 30 A

Power consumption Appr. 880 W (DMS incl. PC, electronics rack, spectroradiometer and 2 x HEL-4 in operation)

1 70° for all automated measurements, 90° for operator control2 All sizes and masses approximate

Electronics rack

1800

mm

(mac

hine

& d

oor)

DMS 803 mechanism

3000 mm

PCtable

2800

mm

Thermo-stream

4000 mm (incl. opration & maintainance space)

Page 8: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

Instrument Systems GmbHNeumarkter Str. 83 81673 Munich, Germany Phone: +49 89/45 49 43-0 Fax: +49 89/45 49 43-11 E-mail: [email protected]

We bring quality to light.

Ambient operating conditions

Ambient temperature 23 °C (±5 °C)

Humidity 0 – 60 % rel. humidity

Photometer PMT 3

Bandwidth 100 kHz (standard), 500 kHz or 2 MHz (optional)

Measurement spot diameter 0.5, 1, 3, 8 mm

Spectrometer CAS 140CT

Spectral range3 360 … 830 nm (usable spectral range depending on software used)

Spectral resolution 2.2 nm

Data point interval4 0.5 nm

Measurement spot diameter 0.5, 1, 3, 8 mm

3 DMSControl software range: 380 … 800 nm4 DMSControl data point interval: 1 nm

Order informationOrder number Description

Basic system

DMS-803-1

Model DMS 803-1 comprising: Goniometer with 6 motorized axes; microscope viewing optics; fast photometer PMT 3/100; 2 x HEL-4 light sources; transmissive illumination; DHS diffuse-reflective source; GRU with ASG-1 (1 channel) and power amplifier PA 3/60; light shielding box; PC with table; DMSControl and ViPer software

DMS-803-4

Model DMS 803-4 comprising: Goniometer with 6 motorized axes; microscope viewing optics; fast photometer PMT 3/100; 2 x HEL-4 light sources; transmissive illumination; DHS diffuse-reflective source; GRU with ASG-4 (4 channels) and power amplifiers PA 3/60; light shielding box; PC with table; DMSControl and ViPer software

Options - Temperature control

DMS-3-10 HCS-3 temperature chamber with ThermoStream control unit

DMS-3-20 HCS-3A temperature chamber with ThermoStream control unit

Options - Light sources and illumination systems

DMS-5-10 HEL-4 light source with 150 W tungsten halogen lamp

DMS-5-12 LC-8 xenon light source for higher light output

DMS-5-20 PID-800 parallel illumination device (collimated beam)

DMS-5-30 VADIS-800 variable aperture diffuse illumination source for reflective measurement

DMS-5-41 DHS-HC diffusing hemisphere for evaluating the contrast of reflective displays; illumination geometry similar to HCS-3 source

DMS-5-50 SDR integrating sphere for diffuse reflectance

DMS-5-60 PLS LED point-light source illumination for measurement of BRDF

Options – Light measuring devices

DMS-1-10CCD spectrometer set for DMS systems; incl. CAS 140CT VIS model (CAS140CT-151) 360 nm – 830 nm (usable spectral range depending on software used)

Options – DUT driving

DMS-2-10 Keithley 2400 series source meter set for DMS, including cabling and interface

Options – Software

DMS-SW-35 DCI device control interface for external devices (for DMSControl)

DMS-SW-40 Report generator for data export to Excel spread sheets

ds_

dm

s803

_en_

V2.5

Page 9: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.Data sheet

The DMS 505 is based on a 6-axis goniometer with 3 motorized axes. The measurement system permits angle-dependent analysis of all electro-optical characteristics of displays, including luminance, contrast, and color properties of displays at different viewing angles and variable electrical driving.

The DMS 505 is ideal for determining the characteristics of displays for mobile applications (smartphones, tablet PCs) with the comprehensive range of accessories including illumination devices and temperature-control chambers.

Product highlights

ì 3 motorized axes for directional scanning and motorized height adjustment

ì Comprising electrical driving, illumination, computer and software

ì Fast photometer for analysis of temporal luminance variations

ì DMSControl software for automated measurement sequences

ì ViPer software for analysis, evaluation and visualization of measurement data

DMS 505Goniometer System

for complete display characterization

Page 10: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

2

Options

ì Spectrometer for radiometric, photometric and colorimetric analysis

ì Directional light sources for measurement of reflectance factor and in-plane BRDF

ì Forced air temperature control of the display under test

ì Interface for various video test-pattern generators

Measurements

ì Angular scan: variation of luminance, contrast and chromaticity with viewing direction for evaluation of directional uniformities and BRDF

ì EOC: variation of luminance, contrast and chromaticity with electrical driving for evaluation of EOTF-characteristics (e.g. gamma)

ì Transients: luminance versus time for e.g. response time evaluation

ì Reflectance characteristics under diffuse and directional illumination

ì Customer specific measurements defined with DCI scripting sequences

System components

The following components and accessories are included in the DMS 505 standard scope of delivery, if not explicitly marked as optional components or accessories.

Temperature control:ì HCS-3 (optional): forced air temperature control

device, temperature range -35 °C to +85 °C, with integrated diffuse reflective illumination

ì HCS-3A (optional): modified version of HCS-3 improved for easier handling, temperature range -40 °C to +85 °C, adjustable DUT fixture

Light sources and illumination systems:ì HEL-4: light source with halogenic incandescent lampì LC-8 (optional): xenon light source for higher light

outputì DHS: diffusing hemisphere for evaluation of contrast

of reflective displaysì DHS-HC (optional): diffusing hemisphere for

evaluating the contrast of reflective displays; illumination geometry similar to HCS-3

ì PID (optional): collimated parallel beam illuminationì VADIS (optional): variable aperture diffuse illumination

source for e.g. scattering characterization of surfaces (BRDF), measurement of reflectance factor

ì PLS (optional): LED point-light source for measurement of reflectance characteristics including BRDF

Light measuring devices:ì PMT 3: fast photometer with photomultiplier tube,

bandwidth 100 kHz, 500 kHz or 2 MHzì CAS 140CT (optional): Model VIS; 360 to 830 nm;

1024 pixel back-illuminated CCD detector; 2.2 nm spectral resolution (100 μm slit); 0.5 nm/pixel data point interval (CAS140CT-151)

Page 11: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

3

DUT driving:ì ASG: arbitrary signal generator with 1 or 4 analog

output channels (depending on basic system variant); standard and user defined signal waveforms; for direct addressing, passive matrix driving and control of TFT arrays (gate and data signals), etc.

ì Source meter (optional): programmable current or voltage source with digital multimeter, in particular for current driven display devices like OLED displays; can be combined with a multiplexer (for driving and measurement of displays with multiple segments and matrix displays)

ì DCI (optional): device control interface for scripting and communication with various external devices, e.g. video pattern generators, module drivers, power supplies, etc. via RS232, GPIB, USB, or TCP/IP

Software:ì DMSControl: administration, configuration and control

of the instrument, definition and management of measurement sequences

ì Visual performance evaluation (ViPer): evaluation and representation of measurement results (on- and off-line), data exchange with other software (e.g. spreadsheet applications) and report generation (e.g. PDF documents)

ì Report generator (optional): user configurable data export from DMSControl to spreadsheet software

Footprint

Scope of delivery

The basic system comprises the following components:

ì DMS 505 goniometer mechanismì Measurement microscope with viewfinderì Photometer PMT 3/100ì GRU with ASG-1 or ASG-4 and power amplifier

PA 3/60 (DMS 505-1 or DMS 505-4)ì DHSì 2 x HEL-4 light source for reflective and transmissive

illuminationì PC with tableì Software package: DMSControl, ViPerì Light shielding box

Technical dataDMS 505

Axis Scan range Min. step size

Inclination Theta (motorized) 0 … 90°1 0.01°

Rotation Phi (motorized) 0 … 360° 0.01°

Reflective illumination (manual) 0 … 90°1 0.01°

x translation (manual) -10 … +10 mm 0.01 mm

y translation (manual) -10 … +10 mm 0.01 mm

z translation (motorized) 0 … 114 mm 0.01 mm

Sphere of confusion (all axes) 100 µm

DUT dimensions

Maximum diagonal600 mm (23" diagonal) for center measurement; 459 mm (18" diagonal) for full x-y scanning range

Maximum thickness 114 mm

Dimensions2

DMS mechanism (L x H x W) 1430 mm x 980 mm x appr. 1650 mm (H 1900 mm with light shielding box)

Light shielding box (L x H x W) 1430 mm x 980 mm x 1050 mm

Electronics rack (L x H x W) 700 mm x 560 mm x 1250 mm (max.)

Mass2

Mechanism 320 kg

Electronics rack 105 kg

Light shielding box 85 kg

Facility requirements

Mains voltage 200 V – 240 VAC (230 V nominal), 50 Hz, 1 phase

Max. mains current / fuse 30 A

Power consumption Appr. 880 W (DMS incl. PC, electronics rack, spectroradiometer and 2 x HEL-4 in operation)

1 70° for all automated measurements, 90° for operator control2 All sizes and masses approximate

Electronics rack

1800

mm

(mac

hine

& d

oor)

DMS 505 mechanism

3000 mm

PCtable

2800

mm

Thermo-stream

4000 mm (incl. opration & maintainance space)

Page 12: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

Instrument Systems GmbHNeumarkter Str. 83 81673 Munich, Germany Phone: +49 89/45 49 43-0 Fax: +49 89/45 49 43-11 E-mail: [email protected] www.instrumentsystems.com

We bring quality to light.

Ambient operating conditions

Ambient temperature 23 °C (±5 °C)

Humidity 0 – 60 % rel. humidity

Photometer PMT 3

Bandwidth 100 kHz (standard), 500 kHz or 2 MHz (optional)

Measurement spot diameter 0.5, 1, 3, 8 mm

Spectrometer CAS 140CT

Spectral range3 360 … 830 nm (usable spectral range depending on software used)

Spectral resolution 2.2 nm

Data point interval4 0.5 nm

Measurement spot diameter 0.5, 1, 3, 8 mm

3 DMSControl software range: 380 … 800 nm4 DMSControl data point interval: 1 nm

Order informationOrder number Description

Basic system

DMS-505-1

Model DMS 505-1 comprising: Goniometer with 3 motorized and 3 manual axes; microscope viewing optics; fast photometer PMT 3/100; 2 x HEL-4 light sources; transmissive illumination; DHS diffuse-reflective source; GRU with ASG-1 (1 channel) and power amplifier PA 3/60; light shielding box; PC with table; DMSControl and ViPer software

DMS-505-4

Model DMS 505-4 comprising: Goniometer with 3 motorized and 3 manual axes; microscope viewing optics; fast photometer PMT 3/100; 2 x HEL-4 light sources; transmissive illumination; DHS diffuse-reflective source; GRU with ASG-4 (4 channels) and power amplifiers PA 3/60; light shielding box; PC with table; DMSControl and ViPer software

Options - Temperature control

DMS-3-10 HCS-3 temperature chamber with ThermoStream control unit

DMS-3-20 HCS-3A temperature chamber with ThermoStream control unit

Options - Light sources and illumination systems

DMS-5-10 HEL-4 light source with 150 W tungsten halogen lamp

DMS-5-12 LC-8 xenon light source for higher light output

DMS-5-20 PID-800 parallel illumination device (collimated beam)

DMS-5-30 VADIS-800 variable aperture diffuse illumination source for reflective measurement

DMS-5-41 DHS-HC diffusing hemisphere for evaluating the contrast of reflective displays; illumination geometry similar to HCS-3 source

DMS-5-60 PLS LED point-light source illumination for measurement of BRDF

Options – Light measuring devices

DMS-1-10CCD spectrometer set for DMS systems; incl. CAS 140CT VIS model (CAS140CT-151) 360 nm – 830 nm (usable spectral range depending on software used)

Options – DUT driving

DMS-2-10 Keithley 2400 series source meter set for DMS, including cabling and interface

Options – Software

DMS-SW-35 DCI device control interface for external devices (for DMSControl)

DMS-SW-40 Report generator for data export to Excel spread sheetsd

s_d

ms5

05_e

n_V

2.4

Page 13: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

DMS 201Goniometer System

with manual positioning

The DMS 201 is a cost-effective goniometer for analyzing the electro-optical characteristics of displays. The measurement system is based on a manual positioning device with 2 axes (optionally 3 axes) and is used to determine luminance, contrast, and color properties at variable electrical driving and various viewing angles.

The DMS 201 is an ideal system for applications that do not require automated angular scans.

Product highlights

ì 2 manual axes for angular positioning and adjustment

ì Measurement and evaluation of emissive, transmissive, reflective, and transflective displays

ì For evaluation of all photometric and colorimetric characteristics versus angle of inclination, electrical driving, time and temperature

Data sheet

The DMS 201 is a cost-effective goniometer for analyzing the electro-optical characteristics of displays. The measurement system is based on a manual positioning device with 2 axes (optionally 3 axes) and is used to determine luminance, contrast, and color properties at variable electrical driving and various viewing angles.

The DMS 201 is an ideal system for applications that do not require automated angular scans.

Page 14: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

2

Options

ì Fast photometer for analysis of temporal luminance variations (flicker evaluation, gray-to-gray transition times, etc.)

ì Spectrometer for radiometric, photometric and colorimetric analysis

ì Directional light sources for measurement of reflectance factor and in-plane BRDF

ì Forced air temperature control of the display under testì DMSControl software for automated measurement

sequencesì ViPer software for analysis, evaluation and

visualization of measurement dataì Software interface for various video test-pattern

generators

Measurements and evaluations

ì Angular scan: variation of luminance, contrast and chromaticity with viewing direction for evaluation of directional uniformities and BRDF

ì EOC: variation of luminance, contrast and chromaticity with electrical driving for evaluation of EOTF-characteristics (e.g. threshold-, saturation-voltage, gamma)

ì Transients: luminance versus time for e.g. response time evaluation

ì Reflectance characteristics under diffuse and directional illumination

ì Comprehensive software package for measurement analysis (optional)

ì Customer specific measurements defined with DCI scripting sequences

Standard components

ì Table top versionì Manual inclination stage for microscope, with scaleì Manual adjustment of height (z axis), appr. 300 mm

stroke ì Measurement microscope with adjustable spot sizes

(depending on sensor system)ì Threads for mounting the adapter for HCS chamber

and for DUT fixturesì Integrated transmissive illumination

Optional components

The following optional components and accessories are available for the DMS 201:

Goniometer:ì Fixture for diffuse reflective illumination (DHS and

SDR) and cover of heat-cool stage (HCS)ì Manual inclination stage of reflective illumination (PID,

VADIS, PLS), with scaleì Manual rotary stage with scale

Temperature control:ì HCS-3: forced air temperature control device,

temperature range -35 °C to +85 °C, with integrated diffuse reflective illumination

ì HCS-3A: modified version of HCS-3 improved for easier handling, temperature range -40 °C to +85 °C, adjustable DUT fixture

Light sources and illumination systems:ì HEL-4: light source with halogenic incandescent lampì LC-8: xenon light source for higher light outputì DHS: diffusing hemisphere for evaluation of contrast

of reflective displaysì DHS-HC: diffusing hemisphere for evaluating the

contrast of reflective displays; illumination geometry similar to HCS-3

ì PID: collimated parallel beam illuminationì VADIS: variable aperture diffuse illumination source

for e.g. scattering characterization of surfaces (BRDF), measurement of reflectance factor

ì PLS: LED point-light source for measurement of reflectance characteristics including BRDF

ì SDR: integrating sphere for measurement of diffuse reflectance factor, specular component excluded or included (d/8, SCE or SCI)

Light measuring devices:ì PMT 3: fast photometer with photomultiplier tube,

bandwidth 100 kHz, 500 kHz or 2 MHzì CAS 140CT: Model VIS; 360 to 830 nm; 1024 pixel

back-illuminated CCD detector; 2.2 nm spectral resolution (100 μm slit); 0.5 nm/pixel data point interval (CAS140CT-151)

Page 15: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

We bring quality to light.

3

DUT driving:ì ASG: arbitrary signal generator with 1 or 4 analog

output channels (depending on basic system variant); standard and user defined signal waveforms; for direct addressing, passive matrix driving and control of TFT arrays (gate and data signals), etc.

ì Source meter: programmable current or voltage source with digital multimeter, in particular for current driven display devices like OLED displays; can be combined with a multiplexer (for driving and measurement of displays with multiple segments and matrix displays)

ì DCI: device control interface for scripting and communication with various external devices, e.g. video pattern generators, module drivers, power supplies, etc. via RS232, GPIB, USB, or TCP/IP

Software:ì DMSControl: administration, configuration and control

of the instrument, definition and management of measurement sequences

ì Visual performance evaluation (ViPer): evaluation and representation of measurement results (on- and off-line), data exchange with other software (e.g. spreadsheet applications) and report generation (e.g. PDF documents)

ì Report generator: user configurable data export from DMSControl to spreadsheet software

Technical data

DMS 201

Axis Scan range

Range of inclination for microscope -90° … 90°

Range of inclination for illumination (optional) -90° … 90°

z stroke (height) 0 … 300 mm

DUT dimensions

Maximum size (L x W)500 mm x 400 mm (23" diagonal for a 16:9 panel) for center measurement; 250 mm x 200 mm (11.5" diagonal for a 16:9 panel) for full manual x-y positioning range

Dimensions1

Size (table top version) (L x W x H) 790 mm x 500 mm x 580 mm

Height of measurement table 140 mm

Mass1

Mechanism (incl. microscope) 35 kg

Ambient operating conditions

Ambient temperature 23 °C (±5 °C)

Humidity 0 – 60 % rel. humidity

1 All sizes and masses approximate

Photometer PMT 3

Bandwidth 100 kHz, 500 kHz or 2 MHz

Measurement spot diameter 0.5, 1, 3, 8 mm

Spectrometer CAS 140CT

Spectral range2 360 … 830 nm (usable spectral range depending on software used)

Spectral resolution 2.2 nm

Data point interval3 0.5 nm

Measurement spot diameter 0.5, 1, 3, 8 mm

2 DMSControl software range: 380 … 800 nm3 DMSControl data point interval: 1 nm

Page 16: DMS 903 Goniometer System - laserchina. · The turnkey solution is based on a fully motorized 6-axis goniometer for determining luminance, contrast, and color properties at different

Instrument Systems GmbHNeumarkter Str. 83 81673 Munich, Germany Phone: +49 89/45 49 43-0 Fax: +49 89/45 49 43-11 E-mail: [email protected] www.instrumentsystems.com

We bring quality to light.

Order information

Order number Description

Basic system

DMS-201-1Model DMS 201-1 comprising: - Goniometer with 2 manual axes - Microscope viewing optics

Options - Temperature control

DMS-3-10 HCS-3 temperature chamber with ThermoStream control unit

DMS-3-20 HCS-3A temperature chamber with ThermoStream control unit

Options - Light sources and illumination systems

DMS-5-10 HEL-4 light source with 150 W tungsten halogen lamp

DMS-5-12 LC-8 xenon light source for higher light output

DMS-5-20 PID-800 parallel illumination device (collimated beam)

DMS-5-30 VADIS-800 variable aperture diffuse illumination source for reflective measurement

DMS-5-40 DHS diffusing hemisphere for reflective displays

DMS-5-41DHS-HC diffusing hemisphere for evaluating the contrast of reflective displays; illumination geometry similar to HCS-3 source

DMS-5-50 SDR integrating sphere for diffuse reflectance

DMS-5-60 PLS LED point-light source illumination for measurement of BRDF

Options – Light measuring devices

DMS-1-10CCD spectrometer set for DMS systems; incl. CAS 140CT VIS model (CAS140CT-151) 360 nm – 830 nm (usable spectral range depending on software used)

DMS-1-20 Fast photometer with PMT 3/100, PMC and DAQ board

Options – DUT driving

DMS-2-10 Keithley 2400 series source meter set for DMS, including cabling and interface

DMS-2-20 GRU display driving and data acquisition unit; with ASG-1 (1 channel) and power amplifier PA 3/60

Options – Software

DMS-SW-10 DMSControl and ViPer software for control, measurement, evaluation and illustration

DMS-SW-35 DCI device control interface for external devices (for DMSControl)

DMS-SW-40 Report generator for data export to Excel spread sheets

ds_

dm

s201

_en_

V2.

2