Upload
others
View
2
Download
0
Embed Size (px)
Citation preview
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 1 of 36
Final report on
APMP.M.P-S3
Results of the supplementary comparison in gas media in the
range 1.77 MPa to 6.8 MPa.
Prepared by
N. Owen1 and M. Bergoglio
2
Date: April 2012
1 National Measurement Institute, Australia (NMIA), Port Melbourne laboratory, Unit 1 -153
Bertie St Port Melbourne, Australia, 3207
2 Istituto Nazionale di Ricerca Metrologica (INRIM), Strada delle Cacce 91, 10135 Torino,
Italy
Abstract
This report provides the results of the supplementary bilateral comparison
APMP.M.P-S3 between NMI Australia, (NMIA) and INRIM, Italy executed from
November 2010 to November 2011. NMIA acted as the pilot laboratory and provided
a Ruska G series piston gauge as the transfer standard, with a nominal area 16.8 mm²
requiring a nominal load of 12 kg to balance a pressure of 7.0 MPa. The purpose of
this supplementary bilateral comparison was to provide data for linking NMIA to
CCM.P-K1.c in the pressure range greater than 4 MPa to extend the range covered by
Key Comparison APMP.M.P-K1c. The original protocol called for the pressure range
from 80 kPa to 6.80 MPa. Both participants agreed to limit this comparison range to
1.77 MPa to 6.50 MPa.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 2 of 36
TABLE OF CONTENTS
1. INTRODUCTION 3
2. MEASUREMENT SCHEDULE 3
3. TRANSFER STANDARD 3
3.1 Description of the transfer standard and calibration procedure 3
3.2 Stability of the transfer standard during comparison 3
4. PARTICIPANTS STANDARDS 4
4.1 About the participant standards 4
4.2 Results obtained by participants 4
4.3 Estimation of uncertainty 5
5. ANALYSIS OF RESULTS 5
5.1 Relative deviation to reference value 6
5.2 Linking to CCM 6
6. DISCUSSION 7
7. CONCLUSION 8
8. ACKLOWLEDGEMENTS 8
9. REFERENCES 8
APPENDIX A: Figure 1 Stability of the transfer standard 9
Figure 2 INRIM and NMIA bilateral comparison bi 9
Figure 3a to 3e Linking of NMIA to CCM.P-K1c 10
APPENDIX B: Table 2 Participant reference standards 13
Table 3 Effective area of the TS as determined by each participant 14
Table 4 Relative difference of bilateral comparison 14
Table 5 Linking of APMP.M.P-S3 to CCM.P-K1c 15
APPENDIX C: Protocol for 7 MPa Pneumatic Pressure Comparison, APMP.M.P-S3 16
APPENDIX D: Data from the participating laboratories 33
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 3 of 36
1. Introduction
INRIM Italy (previously IMGC-CNR) piloted CCM.P-K1.c to establish a comparison
in gas media and gauge mode in the pressure range 80 kPa to 7 MPa completed in
May 1999 and reported results [1]. This original comparison utilised two piston
gauges of different ranges as the transfer standard. NMIA (previously NML)
participated in APMP.M.P-K1c over the range 0.4 MPa to 4.0 MPa completed in
September 2001 with results reported [2]. The main purpose of this current
comparison APMP.M.P-S3, was to confirm NMIA’s measurement capability in the
pressure range from 4 MPa to 7 MPa and establish linking to CCM.P-K1.c.
2. Measurement Schedule
This comparison was conducted according to the schedule noted in Table 1.
Table 1 Schedule of Activity
Activity Period
Preliminary examination by NMIA Nov 2010/Jan 2011
Measurement by INRIM May 2011
Measurement by NMIA Oct/Nov 2011
3. Transfer Standard
3.1 Description of the transfer standard and calibration procedure
The Transfer Standard (TS) was a Ruska model 2468-730 G series tungsten carbide
piston gauge with nominal area 16.8 mm². This piston gauge was manufactured with
a clearance to suit operation at a full scale pressure of 7.0 MPa (nominal 12 kg load).
The piston gauge was mounted in a Ruska 2465 base and supplied with one Ruska
500 g bell and two standard sets of Ruska ring masses. The ring masses were not
weighed as part of this comparison and mass values were supplied by the pilot
laboratory.
Laboratory environmental conditions and TS temperature were measured by each
participant using their own equipment.
The difference between the reference level of the TS and the laboratories reference
standards were minimised. Each laboratory measured the level differences and made
appropriate corrections (refer Table 2).
3.2 Stability of the Transfer Standard during Comparison
In September 2007 the TS was calibrated traceable to NMIA primary reference
standards over the range 0.81 MPa to 7.0 MPa. In 2010 and 2011 the calibration was
repeated using calibration points that better matched the protocol of this comparison.
Common points from these 3 successive calibrations have been analysed and are
presented in Figure 1 and have been used to assess the TS stability.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 4 of 36
The results in 2007 and 2010 did not coincide precisely with the comparison test
points of CCM.P-K1.c hence the effective areas have been adjusted by up to 0.2 MPa
using equation (1) where Am is the measured effective area at pressure Pm and Ac is the
adjusted effective area at the comparison pressure Pc, adjusted using the distortion
coefficient λ declared by the pilot laboratory for the TS of 1.45 x 10-6
. The effective
area adjustment was less than 5 10-6
mm² (or a relative adjustment of 0.3 10-6
) in
each case.
mcmc PPAA 1 (1)
An estimation of stability of the TS for the duration of the comparison can be
obtained by considering the maximum difference in area at each test pressure for
successive calibrations. Between 2007 and 2010 the maximum difference occurred at
3 MPa and amounts to a relative change of 4.4 10-6
. Between 2010 and 2011 (the
duration of the comparison) the maximum difference occurred at 6.5 MPa and
amounts to a relative change of 4.2 10-6
. When compared to previous assessments
of this type of piston gauge [1,2] this is a very consistent stability. For the purpose of
this report the TS was considered to have an estimated relative stability over the
comparison period of 4 10-6
.
4 Participants Standards
4.1 About the participant’s standards
Table 2 provides the details of the primary standards used by each participant. The
NMIA primary standard is a nominal 25 mm diameter dimensionally characterised
piston cylinder unit for which both effective area and distortion coefficient were
calculated by finite element analysis (FEA) modelling. This primary standard is an
oil operated piston gauge mounted in a Harwood 500 kg load frame with a full scale
range of 10 MPa. To interface with the gas operated TS an oil – gas interface with an
integrated oil level pointer was used. Oil level can be monitored and controlled to
within ±0.1 mm or 1 Pa.
The INRIM pressure standard is a Ruska V series piston gauge traceable via a set of
gas operated piston gauges to an INRIM mercury column primary reference standard
and dimensional measurements of piston and cylinder.
The results in this comparison are considered mutually independent and uncorrelated
with those obtained by other laboratories participating in the CCM.P-K1.c
comparison.
4.2 Results obtained by participants
The comparison protocol called for test pressures at 79.4, 137.8, 196.0, 1767.0,
2935.7, 4104.4, 5273.1, 6441.8, 6792.4 in kPa. The participants agreed that the lower
3 pressures would not be collected as these would require a different reference
standard and were less than 3% of the full scale range of the TS. Also due to the set
of weights and test time limitations the test pressures at nominally 6.4 and 6.8 MPa
were replaced with a single test pressure at 6.5 MPa. These revised pressures were
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 5 of 36
applied in increasing then decreasing order. Three cycles were conducted giving 30
experimental determinations of the effective area of the TS with 6 values at each of 5
nominal pressures. The average effective area A’p in mm² at 20 °C (as defined in
Appendix C equation 2) at each nominal test pressure is given in Table 3.
The protocol requires the selected test pressures be within ± 1 kPa when tested at each
laboratory. Due to the large difference in gravity value between Australia and Italy
the test pressure deviated by between 1 kPa and 4 kPa across the test pressure range
while loading the transfer standard with the same masses at each laboratory. Both
laboratories agreed this was the preferred approach.
4.3 Estimation of Uncertainty
In this report all uncertainty calculations are performed and reported as relative
standard uncertainties with k = 1.
Each laboratory supplied the relative standard uncertainty of the test pressure (up), the
relative standard uncertainty associated with test temperature (uT) and the relative
standard deviation of the effective area, s(A’p) at each test pressure. An estimate of
the type A relative standard uncertainty (uA) associated with the average effective area
can be found from (2).
n
Asu
p
A
'
(2)
Where: n is the number of results collected at each test pressure.
'
pAs is the relative standard deviation of area at each test pressure
The combined relative standard uncertainty for effective area at each test pressure is
given by equation (3) as described in the attached protocol in Appendix C.
222
PTAc uuuu (3)
The combined relative standard uncertainty uc at each nominal test pressure for each
laboratory is given in Table 3.
5 Analysis of Results
The analysis is conducted in two parts. The first part considers the bilateral
comparison between INRIM and NMIA. The second part considers the linking of
NMIA via the CCM key comparison CCM.P-K1c reported October 2000.
All charts in this report have the same vertical scale of ± 30 × 10-6
relative units to
permit convenient comparison.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 6 of 36
In this analysis the degrees of equivalence described in the bilateral comparison is
referred to using the term bi and the degrees of equivalence described in the linking
analysis is referred to using the term di.
5.1 Relative difference to APMP.M.P-S3 reference value
The average effective area A’p and combined relative standard uncertainty uc(A
’p)
determined by each participant are provide in Table 3. Using procedure A of [3] the
following value were calculated for the INRIM and NMIA data and are presented in
Table 4 and shown in Fig 2;
1. weighted mean comparison reference value Aref
2. relative standard uncertainty in this value u(Aref)
3. degrees of equivalence bi
4. relative standard uncertainty u(bi)
Where for 2 participants:
NMIApcINRIMpc
NMIApcNMIApINRIMpcINRIMp
refAuAu
AuAAuAA
'2'2
'2''2'
11 (4)
NMIApcINRIMpcref AuAuAu '2'22
111 (5)
refINRIMpi AAb ' (6)
refcINRIMpci AuAubu 2'22 (7)
The degrees of freedom in this case is 1. In accordance with [3] using equation (8)
regard the consistency check as failing if
05.0Pr 22
obs (8)
For this work no results failed this test.
The results in Fig 2 are shown as offset from the nominal pressure for convenience
only.
5.2 Linking to CCM
This bilateral comparison APMP.M.P-S3 was conducted using a different artefact to
that used during the original key comparison CCM.P-K1c. To link the results of
NMIA obtained in the bilateral comparison APMP.MP-S3, to the KCRV of
CCMP-K1c, the method described in [4] was used with INRIM (formerly IMGC-
CNR) as the linking laboratory. It was necessary to transform the data collected for
APMP.M.P-S3 via a set of estimated r values to express this data in the terms of the
original CCM.P-K1c data. These were determined as the ratio of the effective area
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 7 of 36
determined by the linking laboratory for CCM.P-K1c and APMP.M.P-S3 at each test
pressure as given by equation (9) at each test pressure.
INRIMp
IMGCp
A
Ar
'
'
(9)
These r values are then multiplied by each NMIA and INRIM effective area
determined during the APMP.M.P-S3 comparison to give a transformed effective area
A’pT suitable to compare with CCM.P-K1c. An estimate of the degrees of
equivalence and associated standard uncertainty was determined using equations (10)
to (18) of [4] with an assumed correlation coefficient ρl of 0.8. The calculations were
performed on both NMIA and INRIM data to compare the INRIM results with the
reported values in [1] for IMGC. It is noted that using the method described in [4],
the values determined for relative standard uncertainty of degrees of equivalence is
smaller in this current evaluation than the original key comparison or u2
c(A’pINRIM) <
u2
c(A’pIMGC). This is related to the application of the more recent evaluation method of
[4].
The results for transformed effective area A’pT, degree of equivalence di, and relative
standard uncertainty of degree of equivalence u(di) are tabulated in Table 5 for each
laboratory.
The charts of Fig 3a to Fig 3e provide the original CCM.P-K1c data with the results
of APMP.M.P-S3 added.
6 Discussion
(a) Due to operational constraints a maximum test pressure at 6.5 MPa was
substituted for test pressures at 6.4 MPa and 6.8 MPa. The area data collected
at 6.5 MPa was adjusted using equation (1) to area values at 6.44 MPa to
allow direct comparison with CCM.P-K1c. This adjustment did not exceed
1.5 x 10-6
mm².
(b) To compensate for using a different artefact in this bilateral comparison,
APMP.M.P-S3 from that used in the original comparison CCM.P-K1c, a
scaling factor r has been applied to the APMP data. The validity of this
scaling factor was checked by comparing the INRIM data from APMP.M.P-S3
with the IMGC data from CCM.P-K1c. From Figure 3a to 3e it can be seen
that the INRIM and IMGC relative difference (di) values are nominally
identical as expected.
(c) The pressure standards at INRIM have been periodically checked to assure
reliable conformance with the CCM.P-K1c.
(d) Using the method described in [3] for the bilateral comparison, the values of
u(bi), given in Table 4 represent the relative standard uncertainty associated
with the bilateral comparison. As expected the uncertainty of the bilateral
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 8 of 36
comparison is smaller than u(di) of Table 5 calculated for linking NMIA with
the KCRV of CCM.P-K1c.
7 Conclusion
As a participant in the original CCM.P-K1c, IMGC (now called INRIM) provided a
linking laboratory for NMIA to examine measurement capability in the gas operating
pressure range up to nominally 7 MPa.
As shown in Table 4 the relative difference to each comparison value for INRIM and
NMIA are smaller than the relative standard uncertainty estimated for each difference.
These comparison results are considered satisfactory.
Despite using a different transfer standard in this comparison it was possible to select
pressure test points that matched the pressure point of CCM.P-K1c to within 4 kPa,
thereby testing was conducted under as near as possible the same conditions in
APMP.M.P-S3 as in CCM.P-K1c.
By making reference to [1] and [4] it was possible to link the NMIA results to the
original CCM.P-K1c KCRV. As shown in Table 5 the relative difference to the
CCM.P-K1c KCRV for each comparison value for NMIA is smaller than the relative
standard uncertainty estimated for each difference. These comparison results are
considered satisfactory.
8 Acknowledgments
The authors would like to thank Gianfranco Molinar for his encouragement and
support to undertake this comparison and for his participation in the data collection
(quite possibly his last bench work before fully retiring) and data analysis for this
report.
9 References
[1] Results of the CCM pressure key comparison (Phase B) in gas media and
gauge mode from 80 kPa to 7 MPa. Molinar, Legras, Jajer, Ooiwa, Schmidt.
(CCM.P-K1c), IMGC-CNR Technical Report 42, October 2000.
[2] Results of the APMP pressure key comparison in gas media and gauge mode
from 0.4 MPa to 4.0 MPa. Bandyopadhyay, Woo, Fitzgerald, Man, Ooiwa,
Jescheck, Jian, Fatt, Chan, Moore, Tawil. Final report of APMP Key
Comparison (APMP.M.P-K1c).
[3] The evaluation of key comparison data. M. G. Cox. Metrologia, 2002, 39,
589-595.
[4] Proposal for linking the results of CIPM and RMO key comparisons. Elster,
Link, Woger. Metrolgia, 2003, 40, 189-194.
[5] ISO/IEC Guide 98-3. Uncertainty of measurement – Part 3: Guide to the
expression of uncertainty in measurement (GUM:1995).
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 9 of 36
APPENDIX A
Figure 1 Stability of the transfer standard
-30
-20
-10
0
10
20
30
0 1000 2000 3000 4000 5000 6000 7000
Test Pressure in kPa
Rela
tive C
han
ge in
Eff
ecti
ve A
rea X
10-6
2007
2010
2011
Figure 2 Relative deviation to reference value
APMP.M.P-S3 bilateral comparison between INRIM and NMIA
-30
-20
-10
0
10
20
30
0 1000 2000 3000 4000 5000 6000 7000
Pressure in kPa
bi x 1
0-6
INRIM
NMIA
Note: Results offset on the pressure scale to aid viewing.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 10 of 36
Figure 3a Linking of NMIA to CCM.P-K1c KCRV 1767 kPa
APMP.M.P-S3 Nominal Pressure 1767 kPa
Degree of Equivalence: di with uncertainty bars for a coverage factor k = 1
-30
-20
-10
0
10
20
30
Laboratory
di x 1
0-6
IMGC
BNM-LNE
PTB
NIST
NRLM
INRIM
NMIA
CCM.P-K1.c APMP.M.P-S3
Figure 3b Linking of NMIA to CCM.P-K1c KCRV 2936 kPa
APMP.M.P-S3 Nominal Pressure 2963 kPa
Degree of Equivalence: di with uncertainty bars for a coverage factor k = 1
-30
-20
-10
0
10
20
30
Laboratory
di x
10
-6
IMGC
BNM-LNE
PTB
NIST
NRLM
INRIM
NMIA
CCM.P-K1.c APMP.M.P-S3
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 11 of 36
Figure 3c Linking of NMIA to CCM.P-K1c KCRV 4104 kPa
APMP.M.P-S3 Nominal Pressure 4104 kPa
Degree of Equivalence: di with uncertainty bars for a coverage factor k = 1
-30
-20
-10
0
10
20
30
Laboratory
di x
10
-6
IMGC
BNM-LNE
PTB
NIST
NRLM
INRIM
NMIA
CCM.P-K1.c APMP.M.P-S3
Figure 3d Linking of NMIA to CCM.P-K1c KCRV 5273 kPa
APMP.M.P-S3 Nominal Pressure 5273 kPa
Degree of Equivalence: di with uncertainty bars for a coverage factor k = 1
-30
-20
-10
0
10
20
30
Laboratory
di x
10
-6
IMGC
BNM-LNE
PTB
NIST
NRLM
INRIM
NMIA
CCM.P-K1.c APMP.M.P-S3
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 12 of 36
Figure 3e Linking of NMIA to CCM.P-K1c KCRV 6442 kPa
APMP.M.P-S3 Nominal Pressure 6442 kPa
Degree of Equivalence: di with uncertainty bars for a coverage factor k = 1
-30
-20
-10
0
10
20
30
Laboratory
di x
10
-6
IMGC
BNM-LNE
PTB
NIST
NRLM
INRIM
NMIA
CCM.P-K1.c APMP.M.P-S3
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 13 of 36
APPENDIX B
Table 2 Participants Refernce Standards
Details NMI Australia INRIM
Manufacturer Harwood/DHI/NMI via oil/gas interface
Ruska_
Measurement range in MPa 1.2 MPa to 10 MPa Used from 1.8 MPa to 6.5 MPa
Material of piston Tungsten carbide Tungsten carbide
Material of cylinder Tungsten carbide Tungsten carbide
Operation mode, free-deformation or controlled-clearance Free-deformation Free-deformation
Zero-pressure effective area (A0) at reference temperature in mm2 490.2589 8,385652
Relative uncertainty of A0 in 10-6
7.5 9.6
Pressure distortion coefficient ( ) in MPa-1
1.12x10-6
2x10-6
Uncertainty of in MPa-1
0.09 x10-6
0.2 x10-6
Relative uncertainty of mass pieces in 10-6
3 2.2
Linear thermal expansion coefficient of piston and cylinder ( p+ c) in °C-1
9.1x10-6
9.1x10-6
Reference temperature (t0) in °C 20 20
Local gravity (g) in m/s2 9.799499 9.805328
Relative uncertainty of g in 10-6
0.5 1
Height difference between laboratory standard (LS) and TS (h, positive if LS is higher than TS) in mm
-0.3 +1.3
Uncertainty of h in mm 0.3 (includes oil/gas interface)
1
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 14 of 36
Table 3 Effective area of the TS as determined by each participant
Laboratory INRIM NMIA
Nominal Pressure Average effective area
(A’p)
Combined Relative Standard
Uncertainty
(uc(A’p))
Average effective area
(A’p)
Combined Relative Standard
Uncertainty
(uc(A’p))
MPa mm² 10-6
mm² 10-6
1.8
2.9
4.1
5.3
6.5
16.788858
16.788893
16.788939
16.788986
16.789023
10.1
10.2
10.0
10.0
10.0
16.789020
16.788967
16.788982
16.789035
16.789129
10.2
9.4
9.4
9.5
9.5
Table 4 Relative difference to APMP.M.P-S3 reference value Aref
Comparison Reference Value INRIM NMIA
Nominal Pressure Effective area
(Aref)
Relative standard
uncertainty in
effective area
(u(Aref))
Degrees of
equivalence
(bi)
Relative standard
uncertainty
(u(bi))
Degrees of
equivalence
(bi)
Relative standard
uncertainty
(u(bi))
MPa mm² 10-6
10-6
10-6
10-6
10-6
1.8
2.9
4.1
5.3
6.5
16.78894
16.78893
16.78896
16.78901
16.78908
7.2
6.9
6.8
6.9
6.9
- 4.8
- 2.4
- 1.4
- 1.5
- 3.3
7.1
7.5
7.3
7.3
7.3
+ 4.9
+ 2.0
+ 1.2
+ 1.4
+ 3.0
7.3
6.4
6.4
6.5
6.5
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 15 of 36
Table 5 Linking APMP.M.P-S3 to CCM.P-K1c KCRV
CCM.P-K1.c (Table 5) INRIM to KCRV NMIA to KCRV
Nominal
Pressure
Ref Value
(A’pT)
Relative
standard
uncertainty
(u(A’pT))
Transformed
effective area
(A’pT)
Degree of
equivalence
(di)
Relative
standard
uncertainty
(u(di))
Transformed
effective area
(A’pT)
Degree of
equivalence
(di)
Relative
standard
uncertainty
(u(di))
MPa mm² 10-6
mm² 10-6
10-6
10-6
10-6
1.8
2.9
4.1
5.3
6.5
8.38859
8.38863
8.38868
8.38872
8.38877
7.2
7.2
7.2
7.2
7.2
8.388538
8.388619
8.388668
8.388683
8.388747
- 5.7
- 1.6
- 1.3
- 5.0
- 3.1
11.2
11.3
11.1
11.1
11.1
8.388619
8.388656
8.388690
8.388708
8.388800
+ 3.9
+ 2.9
+ 1.2
- 2.2
+ 3.2
11.3
10.1
10.1
10.1
10.1
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 16 of 36
Appendix C:
ASIA-PACIFIC METROLOGY PROGRAMME
7 MPa PNEUMATIC PRESSURE INTERLABORATORY COMPARISON
Comparison Identifier: APMP.M.P-S3
Technical Protocol
Edition 1.0
25 February, 2011
Pilot Institute:
Pressure Group
Melbourne Physical Metrology
National Measurement Institute Australia, NMIA
Contents:
Page
1. General information 2
2. Participating institutes 2
3. Time schedule 3
4. Transfer standard 3
5. Transportation 5
6. Measurements 5
7. Reporting of results 8
8. Preparation of the report 11
References 11
A1. Report for transportation (Arrival) 13
A2. Report for transportation (Departure) 13
A3. State record of the transfer standard 13
A4. Details of the participating institute’s standard 13
A5. Details of the measurement conditions 15
A6. Results in individual cycles 16
A7. Summary of all cycles 17
A8. Calculated result of A’0 and ’ 18
B1. Technical data of the transfer standard 19
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 17 of 36
1. General information
The National Measurement Institute, Australia (NMIA), has been agreed by
the Asia-Pacific Metrology Programme (APMP) Technical Committee for Mass and
Related Quantities (TCM) to coordinate a bilateral comparison between NMIA and
INRiM Italy for medium pressure. The comparison is identified as APMP.M.P-K1x
by the Bureau International des Poids et Mesures (BIPM) and APMP.
The objective of the project is to compare the performance of pneumatic
pressure standards in participating institutes, in the pressure range 80 kPa to 7 MPa in
gauge mode. The results of this comparison will be essential supporting evidence for
medium pressure CMCs1,2,3
. Both participating institutes have the opportunity to get
results in the comparison at a level of uncertainty appropriate for them4. The results of
this comparison will also be linked to the corresponding CCM key comparison,
CCM.P-K1c5. As this protocol is based on CCM.P-K1c additional information can be
obtained from that protocol.
Pilot institute:
The National Measurement Institute, Australia (NMIA)
Program Coordinator:
Mr Neville Owen
Officer in Charge, Melbourne Physical Metrology
National Measurement Institute, Australia
Unit 1- 153 Bertie Street Port Melbourne, 3207 Victoria, Australia
Tel:+61-03-9644 4907, Fax:+61-03-9644 4888,
E-mail: [email protected]
Document:
This document has been prepared by Mr Neville Owen, NMIA and approved
by the participant (INRiM).
2. Participating institutes
The number of participating institute is 2 including the pilot institute. The
institutes along with their contact persons are listed as follows:
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 18 of 36
List of participating institutes
Participating Institutes Contact Persons
Country: Italy
Acronym: INRiM
Institute:
Istituto Nazionale di Ricerca Metrologica
Address:
Strada delle Cacce 91
10135 Torino (Italy)
Name: Dr.ssa Mercede Bergoglio
Tel: 39-011-3919 920
Fax: 39-011-3919 926
E-mail: [email protected]
Country: Australia
Acronym: NMIA
Institute:
National Measurement Institute Australia
Address:
Unit 1 – 153 Bertie Street Port Melbourne, Vic
3207 Australia
Name: Mr. Neville Owen
Tel: 03 9644 4907
Fax: 03 9644 4900
E-mail: [email protected]
3. Time schedule
Time schedule for this comparison has been developed as follows:
Period of measurement Country Institute Carnet
Dec. to Feb. 2010 Australia NMIA ○
May 2011 Italy INRiM ○
June 2011 Australia NMIA ○
Initial testing of the transfer standard commenced in Dec 2010. The bilateral-
comparison program is scheduled to commence in Feb 2011, and to finish in June
2011. The artifact will travel under ATA CARNET.
4. Transfer standard
In this 7 MPa bilateral comparison, a piston-cylinder assembly of nominally
16.8 mm2 effective area with serial number G137 is circulated as the transfer standard
(TS) and includes a pneumatic pressure balance equipped with a mass loading bell
and a mass set manufactured by Ruska Instrument Corporation, a division of General
Electric Company. Temperature and piston position are monitored using a PC and
Excel spreadsheet via an Agilent logger and sensors supplied with the TS.
The TS stability will be established by the pilot laboratory using calibration
history and by repeat measurements upon return of the TS to Australia. The
characteristics of the pressure balance and the effect on the measured pressure by
environmental condition will be evaluated at NMIA during the initial investigation.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 19 of 36
The details of the initial TS evaluation by the pilot institute and all relevant technical
data of TS are given in Appendix B1.
Some general information concerning the pressure balance is given in the
operation and maintenance manuals6, which can be downloaded from the
manufacturer’s website and will be enclosed in a transfer package.
Reference level
The reference level of the TS is to be taken as the base of the piston. The
value to be used in this comparison is given in Appendix B1. The piston is secured in
the cylinder by a retaining circlip near the base of the piston. For the purpose of this
comparison it will be assumed this circlip occupies the same volume as the gap in
which it is secured thereby contributing no additional volume to the piston.
Connecting port:
The connection type of the pressure balance is ¼” Swagelok. Each
participant has to prepare the necessary pipe and fittings for the connection.
Package of transfer standard: The serial number of instrument platform, piston-cylinder assembly, mass
loading bell used for the TS is as follows:
Type Model Serial number
Piston-cylinder assembly 2468-730 G137
Instrument platform 2465-754 50667
Mass set 2465-799 50664
50666
A single box is used for carrying the TS and logger. A second box is used for
the weights and accessories.
5. Transportation
To prevent the transfer package from any damage, all effort should be made by
each participant to handled equipment with care, i.e., only by qualified metrology
personnel. To expedite this comparison a representative from the pilot institute will
assist with unpacking and set up of the TS at INRiM
ATA carnet:
For the transportation of this comparison, ATA CARNET will be used. ATA
CARNET is valid for a year. Upon each movement of the package the person
organizing the transit must ensure that the carnet is presented to customs on leaving
the country, and upon its arrival in the country of destination. When the package is
sent unaccompanied, the carnet must be included with the other forwarding
documents so that the handling agent can obtain customs clearance.
6. Measurements
Measurements should be done after an appropriate acclimatization time (at
least 1 day after receipt), and the results be written on the forms annexed. The TS
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 20 of 36
should be handled and the piston-cylinder assembly mounted in accordance with the
instructions given in the Ruska Piston Pressure Gauge Users Manual.
Working fluid:
The TS pressure balance should be operated with high purity nitrogen as a
pressure transmitting medium.
Measurements for determining the environmental condition:
The measurements needed for determining the environmental condition, for
example, atmospheric pressure, ambient temperature and relative humidity, will be
assumed to be available at each institute. The institutes should operate their pressure
standards at their normal operating temperature.
Metrological characterization of the transfer standard
The results of a metrological characterization of TS by the pilot institute are
presented in Appendix B1. They should help participants to verify that the TS
operates normally. In the case of any anomaly or significant deviation from the results
of the pilot institute it should be contacted.
The magnetization of the piston and cylinder will be checked at the pilot
institute during the comparison. If the magnetization at their surfaces is higher than 2
x 10-4
Tesla, the parts will be demagnetized. Participating institutes may check
magnetization using a non magnetizing test procedure but should inform the pilot
institute before taking any action.
Installation of the transfer standard:
Gloves should be worn when handling the piston, the carrying bell and the
masses of TS.
1) The TS is set on the place where it is going to be calibrated. TS is recommended
to be located close to the institute’s reference standard to keep the pressure line
between the two instruments as short as possible. It is also recommended to adjust
the height position of TS to minimize the height difference between the reference
level of TS defined in Appendix B1 and the reference level of the participant’s
institute standard. (See “Height difference and head correction” )
2) It is recommended to install a shut off valve between the pressure balance and the
participant’s pressure calibration system to check the TS for leaks.
3) Before mounting the assembly the piston and cylinder should be cleaned
according to the usual practice in the institute. The piston-cylinder assembly
should be installed in the mounting post carefully in accordance with the
instructions given in the Ruska Piston Pressure Gauge Users Manual. The
verticality of the piston and cylinder should be adjusted in the participant’s
manner. After the adjustment, ensure that the platform is firmly mounted on a
surface. Also check the verticality during the course of the measurement.
Applying pressure and leak check:
After the installation, TS will be pressurized using the participant’s standard
up to 7 MPa. Any leak in the calibration system should be checked and fixed.
To check the performance of the TS, the piston fall rate shall be measured at
7 MPa. Wait a minimum of 10 minutes after generating the pressure in the TS
measurement system prior to starting the piston fall rate measurements in order to
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 21 of 36
stabilize the TS temperature. It is recommended to set a shut off valve between the
pressure balance and the participant’s pressure calibration system. The valve should
be closed when the piston fall rate is measured because even a minimal leak in the
pressure generating system can significantly disturb the results. The target fall rates
are given in Appendix B1.
Height difference and head correction:
The pressure generated by a pressure standard at the reference level of the TS,
p', is represented by the following equation:
p' = ps + (ρf − ρa)·g·h (1)
where, ps is the pressure generated by the institute’s pressure standard at its reference
level; (ρf − ρa)·g·h, is the head correction, with ρf the density of the working fluid, ρa
the air density, g the local acceleration due to gravity, and h the height difference
between the reference levels of the two intercompared standards (institute standard
and TS). h is positive if the level of the institute’s standard is higher.
To minimize uncertainties in pressure measurement, the height difference
between the reference levels should be kept as low as possible. Also, it is
recommended to measure the height difference accurately.
The piston can be rotated by a motor in the platform instrument or manually.
The rotation direction is clockwise. The recommended rotation speed is not critical
but should not exceed 60 rpm when the mass carrier bell is loaded. Do not use the
rotation speed control function more than necessary since the motor is a major source
of heat.
The piston free rotation time is given in Appendix B1.
Reference temperature:
The reference temperature of the comparison is 20 °C. If measurements are
performed at a temperature deviating from 20 °C, the effective area of TS should be
referred to 20 °C using the piston-cylinder thermal expansion coefficient given in
Appendix B1.
Measurement method:
The measurements shall include three cycles each at the nominal pressures:
79.4, 137.8, 196.0, 1767.0, 2935.7, 4104.4, 5273.1, 6441.8, 6792.4 in kPa generated
in increasing then decreasing order. The generated pressures (p' in equation (1))
should not deviate from these nominal values by more than 0.1 kPa up to 1 MPa then
by more than 1 kPa up to 7 MPa. At least 15 minutes should be allowed between two
consequent measurements at 7 MPa.
The time between a pressure level change and the acquisition of the data
corresponding to the equilibrium of the participating institute’s standard and TS
should be not shorter than 5 minutes. It may be necessary to dismantle the TS for
cleaning during testing if the spin time appears to reduce or if stalling occurs at low
rotation speeds. If this is required the TS should be dry wiped only using a lint free
tissue and a spin test conducted to establish normal operation in accordance with
Appendix B1. The TS must be left to stabilize at least an hour after this dry cleaning.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 22 of 36
If additional cleaning is required involving solvent or water the TS must be left over
night before resuming testing.
The results of the measurement cycle should be recorded on the measurement
results sheet as shown in Appendix A6. It is desirable to complete either a half or full
cycle in one day.
7. Reporting of results
The pilot institute will collate the results including:
(i) Details of the participating institute’s standard (Appendix A4)
(ii) Details of the measurement conditions (Appendix A5)
(iii) Results in individual cycles [1/3, 2/3, 3/3] (Appendix A6)
(iv) Summary of all cycles (Appendix A7)
(v) Calculated result of A’0 and ’ (Appendix A8)
(vi) Uncertainty budget (using the participant’s usual format).
The participating institute may add other information they think useful.
(i) Participating institute’s standard:
The pilot institute will collate the information related to the institute standard
against which the TS was calibrated, including the origin of its traceability to the SI,
as listed in the table reported in Appendix A4. If more than one standard is used, it
must be clearly indicated the standard used in respect of the established nominal
pressure points.
The type of the piston-cylinder assembly used as institute standard, the
characteristics of its effective area and the estimation of the uncertainty, as well as any
other useful information, will be added to the table.
(ii) Measurement conditions:
The parameters used for the measurements will be reported using Appendix
A5. Those are local gravity, height difference of the reference levels between the
participating institute’s standard and the TS. The instruments for measuring
environmental condition, as well as any other useful information, will be added to the
table.
(iii and iv) Measurement and calculation results:
Calculate the applied pressure with the associated standard uncertainty [k=1]
at the reference level of the TS. Any influence quantity for the institute system must
be taken into account and included in the appropriate uncertainty estimation. The
correction by the height difference of the reference levels between the participating
institute’s standard and the TS should be considered. Record these data in the cell on
the sheet for the measurement results in Appendix A6 as follows.
Example: Meas.
No.
Nom.
Pres.
[MPa]
Local
Time
Ambi
Temp.
[°C]
Atmo
R.H.
[%]
Atmo
Pres.
[kPa]
Temp.
t
[°C]
Temp.
t’
[°C]
Pressure
p’
[MPa]
E. A.
A’p
[mm2]
1 50 9:00 20.0 50.0 100.0 20.0 20.0
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 23 of 36
- Ambient temperature, relative humidity and atmospheric pressure are measured
using the participant’s own devices.
- Temperature of piston-cylinder assembly is displayed on PC in Excel.
- p’ is the pressure measured with the institute standard at the local gravity g and the
local air density a and calculated at the reference level of the TS using equation (1).
The participating institutes will report their results, which are measured and
calculated values at the 9 nominal pressures specified, each with an uncertainty in the
measurement (coverage factor k = 1). The date(s) on which practical work was
undertaken is reported in the sheet for the measurement results. The uncertainties shall
be estimated and combined following the Guide to the Expression of Uncertainty in
Measurement (GUM).
To help the pilot institute to get the data of each institute in a consistent form,
each institute should use Appendix A6 to send their data (and Ms-Excel file provided
additionally). Appendix A6 is essentially a data sheet reporting the data obtained at
each comparison point, for each cycle of the planned comparison. In current case the
pilot institute expects 3 data sheets from each participant.
Effective area calculation:
The effective area of TS determined from a particular measurement (A’p)
referred to 20 °C can be calculated with the equation
'
0
''
c
'
p
'
'
a'
'
p1
1
ttααp
ρ
ρgm
Ai
i
i
, (2)
where
m’i are true masses of the piston, the mass loading bell and the mass pieces
placed on the mass loading bell of TS;
’i are densities of the parts with masses m’i;
a is air density;
g is local gravity acceleration;
p' is pressure generated by the institute’s standard at the TS reference level;
’p and ’c are thermal expansion coefficients of the piston and cylinder
materials, respectively;
t' is temperature of TS;
t’0 is reference temperature, t’0 = 20 °C.
The values of p’, a and t’ as well as the masses of the participating institute
are to be calculated or measured by the institute. All other parameters can be
found in Appendix B1.
In Appendix A7, each institute should report whether equation (2) or an
alternative equation was used for A’p determination. The formula to calculate p’
should be given.
(v) Calculated result of A’0 and ’:
In Appendix A8, the zero-pressure effective area of TS (A’0) and its pressure
distortion coefficient ( ’) which satisfy equation
A’p = A’0(1+ ’p) (3)
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 24 of 36
and are based on the results of all 60 measurements, the combined standard
uncertainties of A’0 and ’ as well as a description of how they were calculated should
be included.
Any additional information which in opinion of the participating institute is
important for the interpretation of the comparison results is welcome.
(vi) Uncertainty budget:
A list of the principal components of the uncertainty and any necessary
advice on how uncertainties are estimated should be reported. This is based on the
principles laid out in the Guide to the Expression of Uncertainty in Measurement
(GUM)4. In addition to the principal components of the uncertainty, common to all of
the participants, individual institutes may add any others they consider appropriate.
Uncertainties are evaluated at a level of one standard uncertainty (coverage factor k =
1).
Submission:
Reports (i) – (vi) described above are expected to be prepared as a WORD
file with the tables from Appendixes A6 and A7 being additionally provided in an
EXCEL file. The reports should be sent to the pilot institute by e-mail to the address
Mr Neville Owen
Officer in Charge, Melbourne Physical Metrology
National Measurement Institute, Australia
Unit 1- 153 Bertie Street Port Melbourne, 3207 Victoria, Australia
Tel:+61-03-9644 4907, Fax:+61-03-9644 4888,
E-mail: [email protected]
8. Preparation of the report
The pilot institute is responsible for the preparation of a report on the
comparison. The analysis of data, based on results of participants, will be done by the
pilot institute. In principle, the pilot institute will treat all data and reports according
to the guidelines described in the reference 1.
References:
1.Guidelines for CIPM key comparisons, (Appendix F to the "Mutual recognition of
national measurements standards and of measurement certificates issued by national
metrology institutes" (MRA)), 1999. revised 2003
(http://www.bipm.fr/pdf/guidelines.pdf)
2.Formalities required for the CCM key comparisons (2nd
revised draft), 2001.
3.Entering the details and results of RMO key and supplementary comparisons into the
BIPM key comparison database, 2000.
4.ISO/IEC 2008 Guide 98-3 Guide to the Expression of Uncertainty in Measurement
(GUM: 1995) (Geneva: International Organization for Standardization).
5.G. Molinar et al, IMGC-CNR Technical Report 42 October 2000. Draft B–Results of
the CCM Pressure key comparison (Phase B) in gas media and gauge mode from 80
kPa to 7 MPa.
6.General Electric Company, Ruska Instrument Division, Gas Lubricated Piston
Pressure Gauge Model 2465-754 Users Manual (Release 2465-1D01, revisions G,
December 3, 1992).
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 25 of 36
Appendix A1. Report for transportation (Arrival)
Advise the pilot institute by email of the date and condition upon arrival.
Appendix A2. Report for transportation (Departure)
Advise the pilot institute by email of the date of departure.
Appendix A3. State record of the transfer standard
Advise the pilot institute by email of any fault or failure of the TS during testing.
Appendix A4. Details of the participating institute’s standard
Institute:
Details of the pressure balances used for the comparison:
Manufacturer Model Description
Base
Piston-cylinder Operation mode#1
:
Pressure range [MPa]:
Weights Total mass [kg]:
Typical relative uncertainty of mass
pieces (k=1) [ppm]:
Thermometer
#1 for example: Simple or controlled-clearance, etc
Details of the piston-cylinder:
Material Linear thermal expansion
coefficient ( ) [°C-1
]
Piston
Cylinder
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 26 of 36
Details of the Effective area of the piston-cylinder:
Value Unc. [k=1] Traceability#2
Cali.Date (if applicable)
Zero-pressure effective area
[mm2] at ref. temp., A0,tr
#3
(Ref. temp.: tr)
(tr : [°C])
Pressure distortion
coefficient #3
[MPa-1
]
Pressure distortion
coefficient #3
[MPa-2
] (if applicable)
#2 for example: from PTB, Certificate number 1234, or mercury manometer, etc.
#3 Ap = A0 (1 + p + p2)
Piston rotation:
Method: ( ) By hand or ( ) By Motor [Please check one]
Rotational Speed: [rpm]
Details of the traceability to SI units (for primary standard):
Traceability of the participating institute’s standard to SI units should be explained.
Details of how A0 and ( ) and their uncertainties were determined should be
reported.
Comments:
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 27 of 36
Appendix A5. Details of the measurement conditions
Institute:
Local gravity:
g [m/s2] u(g) [m/s
2] (k=1)
Height difference of the reference levels between the participating institute’s
standard and the transfer standard: (Positive: if the level of the institute’s standard is higher).
h [mm] u(h) [mm] (k=1)
Instruments for measuring environmental condition:
Uncertainty[k=1] Comments
Atmo. Temp
Atmo. RH
Atmo.
Pressure
Method for achieving pressure equilibrium:
It should be reported how TS was connected to the institute standard and how their
pressure equilibrium was achieved and controlled.
Comments:
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 28 of 36
Appendix A6. Results in individual cycles
Institute:
Date:
Cycle Number: ( )/ 3
Meas.
No.
Nom.
Pres.
[MPa] Local
Time
Ambi
Temp.
[°C]
Atmo
R.H.
[%]
Atmo
Pres.
[kPa]
Temp.
t
[°C]
Temp.
t’
[°C]
Pressure
p’
[MPa]
E. A.
A’p
[mm2]
1 79.4
2 137.8
3 196.0
4 1767.0.
5 2935.7
6 4104.4
7 5273.1
8 6441.8
9 6792.4
10 6792.4
11 6441.8
12 5273.1
13 4104.4
14 2935.7
15 1767.0
16 196.0
17 137.8
18 79.4
t is temperature of the participating institute’s standard;
t’ is temperature of the TS; p’ is the pressure measured with the institute standard at the local gravity g and the local air density a
and calculated at the reference level of the TS.
A’p is effective area of TS at the reference temperature 20 °C.
The formula to calculate p’ must be reported.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 29 of 36
Appendix A7. Summary of all cycles
Institute:
Period: dd mm 20yy - dd mm 20yy
: Nom.
Pres.
[MPa]
Typical min.
adjusted mass
[mg] 1)
Average of A’p,
< A’p > [mm2] 2) Rel. standard
deviation of
< A’p > [10-6] 3)
Rel. standard
uncertainty of
p’ [10-6] 4)
Standard
uncertainty of
t’ [°C] 5)
Rel. standard
uncertainty of
< A’p > [10-6] 6)
79.4
137.8
196.0
1767.0.
2935.7
4104.4
5273.1
6441.8
6792.4
1) the smallest mass adjusted on the piston of the TS to reach the equilibrium between
it and the participating institute’s standard, if the classical fall rate method to control
the equilibrium is used. If another method is applied, the typical uncertainty of the
pressure difference between the reference standard and TS (in MPa) due to this
method should be given. In this case the heading of this column shall be appropriately
changed;
2) average of the values measured at the same nominal pressure;
3) standard deviation of the mean value;
4) type B uncertainty including the uncertainty of the pressure at the reference level of
TS, which includes uncertainty of pressure generated by the institute’s standard, of the
height difference between the institute standard and TS, of the density of the pressure
transmitting medium, etc.;
5) type B uncertainty of the temperature measurement on TS;
6) combined uncertainty of the mean values in 2);
In addition, a list of the main uncertainty sources and their contributions to
<A’p> for pressures 80 kPa and 7 MPa must be presented in a separate sheet.
All the uncertainties should be expressed as the standard ones.
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 30 of 36
Appendix A8. Calculated result of A’0 and ’
Institute:
Period: dd mm 20yy - dd mm 20yy
:
Calculated result of A’0 and ’:
Value Unc. [k=1]
Zero-pressure effective area [mm2]
at ref. temp., A’0,t’0 #
(t’0 = 20 °C)
Pressure distortion coefficient ’ # [MPa
-
1]
# A’p = A’0(1+ ’p)
Details of how A’0 and ’ and their uncertainties were determined should be reported.
Comments:
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 31 of 36
Appendix B1. Technical data of the transfer standard
All uncertainties in this appendix are the standard ones.
Piston-cylinder assembly:
The serial number is G137. The nominal effective area of the assembly is:
A’0,nom = 16.8 mm
2.
Piston-cylinder material properties:
The type of piston-cylinder assembly is simple, free deformation. The
measurement range is up to 7 MPa. The cylinder of the assembly is made of tungsten
carbide, and the piston is made of tungsten carbide with the following linear thermal
expansion coefficient ( ).
Material [°C]
Piston Tungsten carbide 4.55 ·10-6
Cylinder Tungsten carbide 4.54 ·10-6
The thermal expansion coefficient of the piston-cylinder unit can be taken as
’p + ’c = (9.1 0.46) 10-6
°C-1 #
.
Piston mass and density:
True mass [g] Equivalent density [kg/m3]
#
Piston 22.174 99 ± 0.000 31 11 200·(1 5·10-2
)
Piston length:
The total piston length (distance from the piston lower face to the upper
piston cup edge) is 64 ± 0.05 [mm].
Reference level and piston working position:
The reference level of the TS is the base of the piston. The recommended
piston working position is physically about 6 mm above its lowermost (low stop)
position.
Note: Please use the participant’s own devices to measure the piston working
position.
Typical cross-float sensitivity and reproducibility:
The relative cross-float sensitivity is between 5 and 10 mg over the test
pressure range. The relative experimental standard deviations of single values of the
effective areas measured at the pressures specified for the comparison are less than 1
10-6
.
Piston fall rates:
Piston fall rate (vf) measured by the pilot institute at around 20 °C is
p [MPa] vf [mm/min]
6.9 1.8
Draft B APMP.M.P-S3, 4 April 2012, version 1 Page 32 of 36
It should be waited minimum 10 minutes after generating the pressure in the
TS measurement system prior to starting the piston fall rate measurements in order to
stabilize the TS temperature. It is recommended to set a shut off valve between the
pressure balance and the participant’s pressure calibration system. When the valve is
available, it should be closed when the piston fall rate is measured because even a
minimal leak in the pressure generating system can significantly disturb the results.
Piston free rotation time:
When piston rotates freely (without motor) with a load of 20 g applied to the
piston (pressure nominally 24.6 kPa) and a temperature of 20 °C, the rotation speed
descends from 90 rpm to stationary within 1.9 minutes.
Stability of the transfer standard:
Several measurements performed between 2007 and 2010 showed maximum
differences in generated pressure of 5·10-6
.
Mass loading bell, mass and density:
The mass loading bell, serial number 50664, is made of non-magnetic
stainless steel and is one of the mass component items.
Temperature probe:
The temperature of the piston-cylinder assembly (t’) is measured with a platinum
resistance thermometer (PRT) supplied with a logger, PC and data acquisition
software and should be used to determine the temperature of the piston-cylinder unit.
Pressure transmitting medium:
The working fluid is high purity nitrogen.
FINAL COMPARISON REPORT MAR 2013 (APMP REVIEW).DOC 33
Appendix D: Data from the participating laboratories.
INRIM
Table 1 - Laboratory standard and measurement conditions
The uncertainties here are expressed as the standard ones.
Manufacturer Ruska_
Measurement range in MPa Used from 1.8 MPa to 6.5 MPa
Material of piston Tungsten carbide
Material of cylinder Tungsten carbide
Operation mode, free-deformation or controlled-clearance Free-deformation
Zero-pressure effective area (A0) at reference temperature in mm2 8,385652
Relative uncertainty of A0 in 10-6
9.6
Pressure distortion coefficient ( ) in MPa-1
2x10-12
Uncertainty of in MPa-1
2.0 x10-13
Relative uncertainty of mass pieces in 10-6
2.2
Linear thermal expansion coefficient of piston and cylinder ( p+ c) in °C-1
9.1x10-6
Reference temperature (t0) in °C 20
Local gravity (g) in m/s2 9.805328
Relative uncertainty of g in 10-6
1
Height difference between laboratory standard (LS) and TS (h, positive if LS is higher than TS) in mm
+1.3
Uncertainty of h in mm 1
Table 2 - Results in individual cycles
Laboratory name: INRIM
Date (period): 12 May 2011 :
Cycle number: 1
Meas. number
Nominal pressure [MPa]
tamb [°C] R.H. [%]
p atm [kPa]
t [°C] t’ [°C] DUT
p’ [Pa] A’p [mm2]
1 1.8 21.60 49.9 98860 21.03 21.31 1764882.27 16.788872
2 2.9 21.65 51.5 98846 21.19 21.54 2932814.63 16.788913
3 4.1 21.80 51.2 98850 21.30 21.70 4106579.85 16.788918
4 5.3 21.80 51.5 98852 21.34 21.75 5274451.11 16.789013
5 6.4 21.70 51.8 98838 21.37 21.79 6448182.92 16.789040
6 6.5 21.80 52.4 98803 21.41 21.84 6506577.17 16.789039
7 6.5 21.80 52.5 98679 21.37 21.76 6506581.18 16.789044
8 6.4 21.80 53.0 98640 21.42 21.87 6448176.37 16.789051
9 5.3 21.80 53.5 98623 21.46 21.92 5274455.29 16.788980
10 4.1 21.80 51.9 98608 21.55 22.01 4106560.34 16.788956
11 2.9 21.90 52.5 98601 21.54 22.01 2932801.08 16.788928
12 1.8 21.90 52.8 98618 21.52 22.02 1764875.50 16.788838
Table 3 - Results in individual cycles
Laboratory name: INRIM
Date (period): 14 May 2011
Cycle number: 2
Meas. number
Nominal pressure [MPa]
tamb [°C] R.H. [%]
p atm [kPa]
t [°C] t’ [°C] DUT
p’ [Pa] A’p [mm2]
1 1.8 21.00 33.2 99325 20.67 20.97 1764882.99 16.788896
FINAL COMPARISON REPORT MAR 2013 (APMP REVIEW).DOC 34
2 2.9 21.00 32.8 99154 20.88 21.22 2932820.76 16.788906
3 4.1 21.00 32.4 99189 20.85 21.19 4106586.51 16.788944
4 5.3 21.00 32 99220 20.83 21.15 5274488.99 16.788962
5 6.4 21.00 32.2 99297 20.77 21.10 6448207.03 16.789063
6 6.5 21.10 32.2 99309 20.73 21.06 6506611.87 16.789048
7 6.5 21.20 32.3 99073 20.93 21.28 6506611.85 16.789017
8 5.3 21.10 32.8 99076 20.94 21.29 5274468.34 16.789012
9 4.1 21.10 33.1 99107 20.93 21.31 4106582.58 16.788948
10 2.9 21.10 32.8 99119 20.92 21.28 2932818.62 16.788914
11 1.8 21.20 32.8 99138 20.89 21.23 1764882.00 16.788872
Table 4 - Results in individual cycles
Laboratory name: INRIM Date (period): 17 May 2011 : Cycle number: 3
Meas. number
Nominal pressure [MPa]
tamb [°C] R.H. [%]
p atm
[kPa]
t [°C] t’ [°C]
DUT
p’ [Pa] A’p [mm2]
1 1.8 20.90 38.3 99356 20.54 20.78 1764893.19 16.788827
2 2.9 21.00 37.7 99336 20.71 21.01 2932835.98 16.788846
3 4.1 21.00 37.4 99307 20.76 21.08 4106594.89 16.788925
4 5.3 21.10 37.9 99272 20.79 21.10 5274493.25 16.788955
5 6.5 21.00 38.4 99242 20.82 21.14 6506631.75 16.788981
6 6.5 21.10 37.6 99156 20.84 21.17 6506618.98 16.789012
7 5.3 21.10 37.4 99094 20.87 21.24 5274475.90 16.788994
8 4.1 21.10 37.6 99045 20.91 21.28 4106585.22 16.788942
9 2.9 21.1 38.4 98999 20.99 21.44 2932825.47 16.788851
10 1.8 21.1 55.6 98742 20.55 20.76 1764893.16 16.788845
Table 5 - Summary of all cycles
Nominal pressure
(MPa)
Typical min.
adjusted mass
(mg) 1)
Average of A’p,
(mm2)
2)
Relative standard
deviation of A’p
mean (10
-6)
3)
u(p’)/p’
(10-6
) 4)
u(t
’)
(°C) 5)
Rel. stan-dard
uncertainty of <A
’p>
(10-6
) 6)
1.8 2 16.788858 1.5 10.0 0.015 10.1
2.9 2.5 16.788893 2.1 9.9 0.015 10.2
4.1 5 16.788939 0.9 9.9 0.015 10.0
5.3 4 16.788986 1.5 9.9 0.015 10.0
6.5 5 16.789024 1.5 9.8 0.015 10.0
FINAL COMPARISON REPORT MAR 2013 (APMP REVIEW).DOC 35
NMIA
Table 1 - Laboratory standard and measurement conditions
The uncertainties here are expressed as the standard ones.
Manufacturer Harwood/DHI/NMI via oil/gas interface
Measurement range in MPa 1.2 MPa to 10 MPa
Material of piston Tungsten carbide
Material of cylinder Tungsten carbide
Operation mode, free-deformation or controlled-clearance Free-deformation
Zero-pressure effective area (A0) at reference temperature in mm2 490.2589
Relative uncertainty of A0 in 10-6
7.5
Pressure distortion coefficient ( ) in MPa-1
1.12x10-6
Uncertainty of in MPa-1
0.09 x10-6
Relative uncertainty of mass pieces in 10-6
3
Linear thermal expansion coefficient of piston and cylinder ( p+ c) in °C-1
9.1x10-6
Reference temperature (t0) in °C 20
Local gravity (g) in m/s2 9.799499
Relative uncertainty of g in 10-6
0.5
Height difference between laboratory standard (LS) and TS (h, positive if LS is higher than TS) in mm
-0.3
Uncertainty of h in mm 0.3 (includes oil/gas interface)
Table 2 - Results in individual cycles
Laboratory name: NMIA
Date (period): 25 Oct 2011 :
Cycle number: 1
Meas. number
Nominal pressure
[MPa]
tamb [°C] R.H. [%]
p atm
[kPa]
t [°C] t’ [°C]
DUT
p’ [Pa] A’p [mm2]
1 1.8 20.39 39.75 101.675 20.24 20.36 1763839.2 16.78888
2 2.9 20.45 39.99 101.670 20.30 20.39 2931104.0 16.78882
3 4.1 20.46 40.49 101.384 20.30 20.40 4104191.8 16.78881
4 5.3 20.52 40.49 101.697 20.35 20.45 5271408.5 16.78883
5 6.4 20.51 41.17 101.700 20.39 20.48 6444456.8 16.78888
6 6.5 20.48 39.66 101.693 20.43 20.49 6502757.4 16.78903
7 6.4 20.49 39.47 101.721 20.47 20.53 6444390.1 16.78904
8 5.3 20.49 39.12 101.737 20.48 20.52 5271360.4 16.78897
9 6.5 20.48 34.78 102.164 20.15 20.35 6502758.8 16.78904
10 4.1 20.51 35.07 102.180 20.23 20.41 4104147.5 16.78898
11 2.9 20.48 35.03 102.195 20.30 20.45 2931074.3 16.78896
12 1.8 20.49 35.4 102.201 20.35 20.47 1763823.6 16.78900
Table 3 - Results in individual cycles
Laboratory name: INRIM
Date (period): 28 Oct 2011
Cycle number: 2
Meas. number
Nominal pressure
[MPa]
tamb [°C] R.H. [%]
p atm [kPa]
t [°C] t’ [°C] DUT
p’ [Pa] A’p [mm2]
1 1.8 20.6 44.06 100.940 20.30 20.32 1763856.2 16.78874
FINAL COMPARISON REPORT MAR 2013 (APMP REVIEW).DOC 36
2 2.9 20.59 43.78 100.914 20.40 20.38 2931098.0 16.78888
3 4.1 20.58 44.8 100.888 20.47 20.43 4104156.9 16.78897
4 5.3 20.57 45.38 100.854 20.53 20.47 5271335.3 16.78908
5 6.4 20.57 46.46 100.729 20.49 20.47 6444364.9 16.78914
6 6.5 20.6 45.87 100.695 20.52 20.49 6502692.3 16.78923
7 6.5 20.62 47.39 100.666 20.53 20.50 6502695.9 16.78922
8 6.4 20.6 47.28 100.624 20.56 20.52 6444328.4 16.78923
9 5.3 20.64 46.63 100.557 20.51 20.51 5271309.3 16.78917
10 4.1 20.64 47.58 100.530 20.52 20.52 4104115.8 16.78914
11 2.9 20.61 49.67 100.494 20.58 20.54 2931055.0 16.78911
12 1.8 20.62 49.91 100.488 20.62 20.56 1763803.6 16.78922
Table 4 - Results in individual cycles
Laboratory name: INRIM Date (period): 31 Oct 2011 : Cycle number: 3
Meas. number
Nominal pressure
[MPa]
tamb [°C] R.H. [%]
p atm [kPa] t [°C] t’ [°C] DUT
p’ [Pa] A’p [mm2]
1 1.8 20.33 35.19 102.511 20.12 20.14 1763827.2 16.78900
2 2.9 20.35 35.64 102.504 20.15 20.18 2931092.2 16.78890
3 4.1 20.33 35.48 102.467 20.21 20.21 4104175.5 16.78889
4 5.3 20.38 35.7 102.460 20.27 20.24 5271354.0 16.78901
5 6.4 20.43 35.93 102.438 20.32 20.27 6444384.9 16.78907
6 6.5 20.36 36.04 102.413 20.37 20.29 6502718.5 16.78914
7 6.5 20.4 35.76 102.410 20.39 20.29 6502728.8 16.78912
8 6.4 20.39 35.85 101.643 20.18 20.14 6444363.0 16.78917
9 5.3 20.38 35.52 101.685 20.27 20.21 5271319.3 16.78915
10 4.1 20.39 35.35 101.688 20.30 20.24 4104128.2 16.78910
11 2.9 20.44 35.29 101.668 20.36 20.27 2931052.9 16.78913
12 1.8 20.46 35.84 101.665 20.40 20.28 1763798.0 16.78928
Table 5 - Summary of all cycles
Nominal pressure
(MPa)
Typical min.
adjusted mass
(mg) 1)
Average of A’p,
(mm2)
2)
Relative standard
deviation of A’p
mean (10
-6)
3)
u(p’)/p’
(10-6
) 4)
u(t
’)
(°C) 5)
Rel. stan-dard
uncertainty of <A
’p>
(10-6
) 6)
1.8 0.62 16.789020 12.10 8.82 0.18 10.24
2.9 0.86 16.788967 7.57 8.76 0.18 9.43
4.1 1.71 16.788982 7.40 8.76 0.18 9.39
5.3 2.23 16.789035 7.55 8.81 0.18 9.46
6.5 3.43 16.789130 5.08 8.84 0.18 9.49