Endoscopes Inspection Report – Sale: 1505 Date: 12 / 05 / 2015
(Condition Reports for other items at the end of the Scopes Report)
These reports are compiled by our Medical Engineering Team & are for guidance only; all
lots are sold as seen, as per our Conditions of Sale. We recommend that wherever possible
you inspect lots prior to purchase.
Lot Description Condition
100 FUJINON
EG-530FP
Videoscope
1000x8.9mm
Insertion Tube – Very good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – No processor to test with
Distal End – Good condition
Image Quality – No processor to test with
Working Channel (s) – Clear
Forceps Elevation Mechanism – N/A
Leak Test – Pass
Overall Condition – Good physical condition, unable to fully test scope’s
working condition
101 FUJINON
EC-530FL
Videoscope
1600x12.5mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – No processor to test with
Distal End – Good
Image Quality – No processor to test with
Working Channel (s) – Clear
Forceps Elevation Mechanism – N/A
Leak Test – Pass
Overall Condition – Good physical condition, unable to fully test scopes
working condition
102 FUJINON
EG-530FP
Videoscope
1000x8.9mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – No processor to test with
Distal End – Good condition
Image Quality – No processor to test with
Working Channel (s) – Clear
Forceps Elevation Mechanism – N/A
Leak Test – Pass
Overall Condition – Good physical condition, unable to fully test scopes
working condition
103 OLYMPUS
No
Identification
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good but only manual
Videoscope
1000x11.6mm
brightness control is working
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End –Fair condition
Image Quality – Satisfactory
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Satisfactory overall condition
104 Olympus
HYF-XP
Direct view
200x3.23mm
Insertion Tube –Satisfactory condition, large dent
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Satisfactory optics
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Satisfactory image (more than 50 fibres missing)
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition –Satisfactory physical condition, large dent and most
of the fibres missing from image
105 OLYMPUS
GIF-Q230
Videoscope
1000x11.87m
Insertion Tube –Satisfactory condition with few minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good focusing
Remote Control Switches – Fully funcional
Distal End – Good condition
Image Quality –Good image quality
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
106 OLYMPUS
CYF- V2
Videoscope
250x5.61mm
Insertion Tube – Satisfactory condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – No processor to test with
Distal End – Good condition
Image Quality – No processor to test with
Working Channel (s) – Clear
Leak Test –Pass
Overall Condition – Good overall condition
107 OLYMPUS
CYF- V2
Videoscope
270x5.66mm
Insertion Tube – Good condition with minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – No processor to test with
Distal End – Good condition
Image Quality – No processor to test with
Working Channel (s) – Clear
Leak Test –Pass
Overall Condition – Good overall condition
108 OLYMPUS
CYF- V2
Videoscope
250x5.61mm
Insertion Tube – Good condition with minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – No processor to test with
Distal End – Good condition
Image Quality – No processor to test with
Working Channel (s) – Clear
Leak Test –Pass
Overall Condition – Good overall condition
109 PENTAX
EC-380IL
Videoscope
1400x12.10mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End – Satisfactory condition
Image Quality – Reasonable image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Satisfactory physical condition
110 OLYMPUS
OSF-2
Direct View
500x12.20mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Poor movement
Light Guide Cable & Light Transmission – Good light guide cable
Objective Lens and Focusing – Poor focussing
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Poor image
Working Channel (s) – Clear
Leak Test –Leaking
Overall Condition – Poor overall condition, dark image is seen and the
objective lens has been come off the probe.
111 OLYMPUS
CYF-3
Direct View
270x5.8mm
Insertion Tube – Satisfactory condition with few minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good focussing
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with few fibres missing
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Average overall condition
112 PENTAX
F1-10BS
Direct View
Insertion Tube – Good condition with few minor dents
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – No light guide cable
Objective Lens and Focusing – Satisfactory optic
550x3.65mm
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Reasonable image (25+ fibres missing)
Working Channel (s) – N/A
Leak Test – Leaking
Overall Condition – Reasonable overall condition, 25+ fibres missing
113 OLYMPUS
ENF-P3
Direct View
200x4.0mm
Insertion Tube – Good condition with few minor dents
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good focussing
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with 5+ fibres missing
Working Channel (s) – N/A
Leak Test – Pass
Overall Condition – Good overall condition
114 OLYMPUS
CF-240AL
Videoscope
1600x12.80mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good focussing
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Good image quality with dim colour and slightly grey
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Satisfactory overall condition
115 OLYMPUS
GIF-XQ240
Videoscope
1000x10.10mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – Good focussing
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Good image but the image colour is slightly grey
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Satisfactory overall condition
116 PENTAX
EB-1830T2
Videoscope
500x6.60mm
Insertion Tube – Good condition with few minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
117 OLYMPUS
No
Identification
Direct View
500x4.40mm
Insertion Tube – Satisfactory condition with few dents
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Reasonable optics
Remote Control Switches – N/A
Distal End – Good
Image Quality – Reasonable image with few pixels missing
Working Channel (s) – N/A
Leak Test – Leaking
Overall Condition – Satisfactory overall condition
118 OLYMPUS
CYF-200
Videoscope
270x5.60mm
Insertion Tube – Good condition with few minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Reasonable optics
Remote Control Switches – Fully functional
Distal End – Good
Image Quality – Reasonable image
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Satisfactory overall condition
119 OLYMPUS
CF-240DL
Videoscope
1600x12.10mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Good overall condition
120 OLYMPUS
BF-P260F
Videoscope
450x4.53mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
121 OLYMPUS
BF-200
Videoscope
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good but very dim light
output
Objective Lens and Focusing – Satisfactory optics
Remote Control Switches – Fully functional
Distal End – Good
450x6.4mm Image Quality – Satisfactory image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
122 Fujinon
NAP-F
Direct View
420x4.65mm
Insertion Tube – Poor condition with half of the insertion tube’s outer
insulation is ripped, exposing inner metal insulation
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Light guide cable is in poor
condition with its outer insulation worn and torn. Good light output
Objective Lens and Focusing – Good focussing
Remote Control Switches – N/A
Distal End – Satisfactory condition
Image Quality – Good image with few pixels missing
Working Channel (s) – N/A
Leak Test – N/A
Overall Condition – Poor overall condition
123 PENTAX
FNL-10P2
Direct View
260x3.95mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good light output
Objective Lens and Focusing – Reasonable optics and focussing
Remote Control Switches – N/A
Distal End – Good
Image Quality – Reasonable image
Working Channel (s) – N/A
Leak Test – Pass
Overall Condition – Good overall condition
124 OLYMPUS
LF-TP
Direct View
450x5.05mm
Insertion Tube – Good condition with few minor dents
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – No light guide cable
Objective Lens and Focusing – Satisfactory optics and focussing
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Reasonable image with few pixels missing
Working Channel (s) – Clear
Leak Test –Pass
Overall Condition – Good overall condition
125 OLYMPUS
TJF-240
Videoscope
1200x12.75mm
Insertion Tube – Good condition with the distal end cap torn and worn
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good Light output but auto
brightness control is not working
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End – Poor condition
Image Quality – Good image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
126 Richard Wolf
No
Identification
Direct View
270x3.66mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good Light output
Objective Lens and Focusing – Good optics and focussing
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with more than 100 pixels missing
Working Channel (s) – N/A
Leak Test – Leaking
Overall Condition – Good overall condition
127 PENTAX
ECY-15P2
Direct View
300x5.60mm
Insertion Tube – Good Condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good light output
Objective Lens and Focusing – Good optics and focussing
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with few pixels missing
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
128 Fujinon
EG-450WR5
Videoscope
1000x10.05mm
Insertion Tube – Good Condition
Bending mechanism & Angulation Control – Good movement and lock
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
129 Fujinon
EC-450WL5
Videoscope
1600x12.55mm
Insertion Tube – Good Condition
Bending mechanism & Angulation Control – Good movement and lock
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
130 Fujinon
No
Identification
Videoscope
Insertion Tube – Good Condition
Bending mechanism & Angulation Control – Good movement and lock
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
1600x13.82mm
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Good overall condition
131 Fujinon
No
identification
Videoscope
1600x12.55mm
Insertion Tube – Good Condition
Bending mechanism & Angulation Control – Good movement but locks
are not functional
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Good overall condition
132 Fujinon
EG-450WR5
Videoscope
1000x9.65mm
Insertion Tube – Good Condition
Bending mechanism & Angulation Control – Good movement and lock
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
133 Fujinon
No
Identification
Videoscope
1000x9.90mm
Insertion Tube – Good Condition with few minor dents
Bending mechanism & Angulation Control – Good movement and lock
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
134 Fujinon
EG-450WR5
Videoscope
1000x10.0mm
Insertion Tube – Good condition with few minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
135 Olympus
BF-P60
Insertion Tube –Good condition with minor dent
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good
Direct view
520x4.95mm
Objective Lens and Focusing – Satisfactory optics
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with one pixel out
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition –Good physical condition
136 Fujinon
No
Identification
Videoscope
1600x12.70mm
Insertion Tube – Good condition with few minor dents
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good light transmission
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Satisfactory condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
137 Richard Wolf
No
Identification
Videoscope
250x5.00mm
Insertion Tube – Good condition with
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Not tested
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
138 Olympus
PCF-240S
Videoscope
900x12.30mm
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition –Good physical condition
139 Olympus
GIF-2T20
Direct view
1000x12.65mm
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with two pixels out
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition –Good physical condition
140 Olympus
TJF-240
Videoscope
1200x12.82mm
Insertion Tube –Good condition with minor dents and missing distal end
cap
Bending mechanism & Angulation Control – Satisfactory movement and
locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Satisfactory optics
Remote Control Switches – Fully functional
Distal End – Poor condition with missing end cap
Image Quality – Good but dull coloured image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition –Good physical condition
141 PENTAX
FCY-15P2
Videoscope
(No case)
300x5.10m
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – N/A
Distal End – Satisfactory condition
Image Quality – Good image with 2 pixels missing
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Good physical condition
212 Olympus
CHF-P60
Direct view
270x5.60mm
Insertion Tube –Good condition with minor dent
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Satisfactory optics with few stains on the
lens
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with 15+ pixels out
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition – Satisfactory physical condition
213 Olympus
BF-B3
Direct view
470x6.40
Insertion Tube – Poor condition with large dents and the outer insulation
is ripped off exposing the inner metal insulation
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Cable has few dents but good
light transmission
Objective Lens and Focusing – Poor optics with stains on the lens
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Poor image with lots of pixels missing
Working Channel (s) – Clear
Leak Test – N/A
Overall Condition – Poor physical condition
214 Olympus
CHF-P60 Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement
Direct view
270x5.0mm
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with 100+ pixels missing
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition –Good physical condition
215 PENTAX
FB-15X
Direct view
510x4.95mm
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Good image with two pixels out
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good physical condition
216 Olympus
CYF-240
Videoscope
240x5.60mm
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional except for one
Distal End – Good condition
Image Quality – Satisfactory image
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good physical condition
217 Olympus
LF-V
Videoscope
510x4.30mm
Insertion Tube – Good condition
Bending mechanism & Angulation Control – Good movement
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Good condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition –Good physical condition
218 Olympus
CF-1T200L
Videoscope
1600x14.65mm
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – Good optics
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Satisfactory image
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition –Good physical condition
222 PENTAX
ED-3410
Videoscope
520x12.20mm
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Satisfactory condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Leaking
Overall Condition –Good physical condition
223 PENTAX
EG-3840TK
ED-3440TK
ED-3440TK
Videoscope
1000x12.05mm
1200x12.96mm
1200x12.40mm
Insertion Tube –Good condition, Both ED-3440TK are missing its distal
end cap
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good except for one (EG-
3840TK) with dim light
Objective Lens and Focusing – No processor to test with
Remote Control Switches – Not tested
Distal End – Satisfactory condition
Image Quality – Not tested
Working Channel (s) – Clear
Leak Test – Leaking (EG-3840TK), Pass
Overall Condition – Reasonable physical condition
224 GYRUS ACMI
DUR-8
Direct View
1000x12.05mm
Insertion Tube –Good condition
Bending mechanism & Angulation Control – Poor movement
Light Guide Cable & Light Transmission – No light guide cable
Objective Lens and Focusing – Below average
Remote Control Switches – N/A
Distal End – Good condition
Image Quality – Poor
Working Channel (s) – Clear
Leak Test – N/A
Overall Condition – Good physical condition
254 Pentax
ED-3410
ES-3840
EG-3440
Videoscope
1200x12.36mm
600x12.54mm
1000x12.55mm
Insertion Tube – Good Condition
Bending mechanism & Angulation Control – Good movement and locks
Light Guide Cable & Light Transmission – Good light output
Objective Lens and Focusing – Good optics and focussing
Remote Control Switches – Fully functional
Distal End – Good condition
Image Quality – Image is slightly yellow
Working Channel (s) – Clear
Leak Test – Pass
Overall Condition – Good overall condition
Condition Report – Sale: 1505 Date: 12 / 05 / 2015
These reports are compiled by our Medical Engineering Team & are for guidance only; all
lots are sold as seen, as per our Conditions of Sale. We recommend that wherever possible
you inspect lots prior to purchase.
Lot
Number
Description Report
1 GE T2000 Treadmill
with GE Case Stress
Test Unit
The unit powers up successfully. The Case software can be
accessed. Treadmill powers up ok. Acquisition module not
included therefore unable to test the system and connection cable
RS232 to the treadmill and ECG unit is not included. No further
test carried out. Few keys are missing from the keyboard. Unit is
in satisfactory physical condition.
5 Olympus Stack with
Olympus Lucera CV-
260 Camera
Controller, CLV-U40
Light Source and
OVC 200 Training
Scope
The processor and light source are able to power up with its
function keys and indicators in working order. The light source is
able to produce a good light output. The brightness adjustment is
in working order. Monitor is not included. Camera head included
and was tested on another processor and monitor and seems to be
working. The stack is in good physical condition.
6 Sonosite iLook
Portable Ultrasound
with Probe, Power
Supply and User
Guide
Unit powers up ok. Display produces a good image. Function
buttons are in working condition. Linear probe is in good physical
condition. No further tests carried out. Sonosite power supply
included. Overall unit is in reasonable physical condition with
wear and tear around buttons.
7 Sonosite iLook
Portable Ultrasound
with Probe, Power
Supply and User
Guide
Unit powers up ok. Display produces a good image. Function
buttons are in working condition. Linear probe is in good physical
condition. No further tests carried out. Sonosite power supply
included. Overall unit is in good physical condition.
9 Datex Ohmeda S/5
Avance Anaesthetic
Machine
The unit powers up and completes the system self test
successfully. Display panel, function keys and rotary control knob
are all in good working order. All ventilation modes which
include VCV, PCV, SIMV/PSV, PSV Pro, SIMV-PC & PCV-VG
are functional and the back up battery is also working. Overall the
unit is reasonable physical and cosmetic condition (the surface
layer on some of the front panel buttons have been worn out
slightly). Bellows, CO2 canister and medical gas hoses are in
pace. The unit does not include a gas module. O2 sensor needs
replacing. SN: ANBH00184. Software version: 06.10.
10 Datex Ohmeda S/5
Patient Monitors x 2
with Some Leads
Both units are able to power up and pass through its initial self-
test successfully. Both of the unit’s display and function buttons
are working in order. One unit has only gas module. Second unit
has multi-parameter and gas module. ECG and SpO2 leads
included and were tested using ECG simulator and the values
were within limit. NIBP not tested but seem to be working.
Pressure, temperature and gas not tested. Both units are in fair
physical condition with broken parts.
12 Olympus CLV-S20
Lightsource
CLV S20 is able to power up and produce a good light output. All
the function buttons and indicators are functioning. Its spare bulb
is in working order. The brightness adjustment for manual mode
is working. Overall unit is in fair physical condition.
14 Olympus CLH-250
Light Source - 28128,
Olympus OTV-S5
Camera Control -
28510, Olympus
OEV142 Trinitron
Monitor, Olympus
OTV-S5 Camera
MH999E - 28129,
522383
The processor, light source and the monitor powers up ok. Light
source failed to produce a light output. The camera head was
tested using the processor and a good image was obtained on the
monitor. No cable or connections included. All units are in
satisfactory physical condition.
15 GE Marquette Dash
3000 Patient Monitor
x 2
Both units are able to power up and pass through its initial self-
test successfully. Both of the unit’s display and function buttons
are working in order. ECG, NIBP, Temperature, Pressure and
SpO2 leads are not included and therefore not tested. CO2
measurement cable in included. Both units are in fair physical
condition with scuffs and marks.
16 Olympus CLK-4 Light
Source
CLK-4 is able to power up however failed to produce light output.
Unit’s air pump is functioning. Overall unit is in fair physical
condition.
18 ConMed Hyfrecator
2000 ESU
Unit power up ok, the numerical display and RF power selector
knob is working fine. The monopolar and bipolar modes was
tested using an ESU analyser and all of the RF powers measured
were the set values. The unit is in good physical condition but
come with no mobile stand.
21 Eschmannn TD 830
Electrosurgical Unit
Unit power up ok, the numerical display and RF power selector
keys are working fine. The unit was tested using an ESU analyser.
The bipolar RF outputs were well within the set values. However,
the monopolar mode CUT has no RF power output. COAG output
is working fine. The unit is in average physical condition and no
monopolar and bipolar footswitch included.
27 Carl Zeiss 150 FC
Colposcope
Lamp is powering up and produces a good light output with its
brightness adjustment in working order. Reasonable optic system.
Magnification is working. The objective lens is f=300. Eyepiece
included. The mechanical movements and locks are in working
order. The physical condition is average.
Eyepiece: 10x
36 Eschmann TD850
Electrosurgical Unit
Unit power up ok, the numerical display and RF power selector
knob is working fine. The monopolar mode was tested using an
ESU analyser and some of the RF power measured were slightly
low than the set values for CUT; COAG was ok. The bipolar RF
outputs were within the set values. The unit is in fair physical
condition and monopolay and bipolar footswitches are included.
44 Camlab ALC 4232
Centrifuge
ALC 4232 centrifuge power up ok, with its function buttons in
working order. The lid lock is functioning. Timer and speed
controls are working. Unit completed a cycle without any errors.
Rotor and test tube holders in fair condition. Overall lot is in
satisfactory physical condition with some scuffs and marks.
48 Penlon Prima SP
Anaesthetic Machine
with Penlon AV900
Touch Screen
Ventilator
Unit powers up ok. The pressure gauges and flow meters are
working fine. O2 flush is functional. Basal flow and anti-hypoxic
feature is working in order. The AV900 ventilator powers up and
passed the initial self-test successfully. The touch screen and
display are working. The re-breathing circuit and absorber are all
intact. The ventilator is working. All the set parameters and alarm
system is functional. Oxygen sensor is not included. The medical
gas hose is included for the O2, Air and N2O. AGSS hose
included. Overall the lot is in satisfactory physical condition with
scuffs and marks.
Sotware version : 3.22
49 Datex Ohmeda Light
Patient Monitor x 2
Both units are able to power up and pass through its initial self-
test successfully. Power adaptor not included but one was used to
test. Both of the unit’s function buttons are working in order.
LCD of both units are defective however one unit can still display
some information whereas the second one is fully blank. SRAM
error shown on one of the unit. ECG for one of the unit was tested
using test leads and ECG simulator and the values were within
limit and the recorder is working. No leads or accessories
included. Both units are in good physical condition with scuffs
and marks.
52 Mitek VAPR 3 ESU Unit power up with error message ‘Faulty Accessories’ The
footswitch is included but hand piece.
53 Pulse Oximeter Ohmeda Biox 3740 x3 does not have any mains adapters or finger
probe; unable to test these units. The Nellcor N-300 power up ok
display and function keys working but no probe – in fair physical
condition. Datex Ohmeda 3800 – power up but the LCD remains
bright and unable to adjust contrast.
66 HP Pagewriter XLS
ECG Recorder with
ECG Leads
The unit power up ok and the initial screen is displayed but the
system failed to proceed to the function screen menu. The lot
comes with ECG leads. Unit not tested. fair physical condition.
67 Datex Ohmeda Excel
210SE Anaesthetic
Machine 7800
Ventilator
Unit powers up ok. The pressure gauges and flow meters are
working fine. O2 flush is functional. Basal flow and anti-hypoxic
feature is working. The 7800 ventilator powers up and passed the
initial self-test successfully. The re-breathing circuit and absorber
are all intact. The ventilator is functional. All the set parameters
and alarm system is working. The medical gas hose is included
for the O2, Air and N2O. AGSS hose included. Overall the lot is
in satisfactory physical condition with scuffs and marks.
68 Olympus Stack
Trolley with Sony
Trinitron Monitor
UHI-2 Insufflator,
CLV 530 Light Source
OTV-56 Camera
Controller, MH317
Footswitch, Sony VCR
and Sony Colour
Video Printer, OTV
5X2 Control
The camera processor and light source are able to power up with
its function keys and indicators are working in order. Camera
head is not included therefore unable to fully test. The light
source is able to produce a good light output. The brightness
adjustment is in working order. The monitor powers up. The
insufflation unit, color video printer and video cassette recorder
power up but not tested. Footswitch for insufflator included. The
stack system is in good physical condition.
70 Medtronic Lifepack
12 Defibrillator with
ECG and Patient
Lead
Defibrillator powers up and is able to pass through its initial self-
test successfully. Function buttons, selector knob and display are
in working order. AED cable and ECG lead included and was
used to test with Defibrillator analyser. Unit showed to be
functional in AED, Pacer and manual modes when tested. Unit is
able to charge and deliver shock and output was within limit. User
test passed. Audio voice prompts and recorder was functional. AC
adaptor included. Good physical condition.
73 Philips Intellivue
MP70 Monitor with
Monitor powers up ok with its touch screen and display in
working order. ECG was tested with test leads and ECG simulator
M3105A Module x 2 and is within limit. SpO2, NBP, Temperature and pressure not
tested. No leads included. Satisfactory physical condition.
YOM: 2005
74 Philips Intellivue MP
70 Monitor with
M3105A Module x 2
Monitor powers up ok with its touch screen and display in
working order. ECG was tested with test leads and ECG simulator
and is within limit. SpO2, NBP, Temperature and pressure not
tested. No leads included. Satisfactory physical condition with
some cracks.
YOM: 2005
76 EX Medispin
Centrifuge
IEC Medispin centrifuge powers up ok ALC 4232 centrifuge
power up ok. The lid lock is functioning. Unit completed a cycle
without any errors. Rotor and holders are in fair condition.
Overall lot is in good physical condition.
86 GE Trasport Pro
Patient Monitor and
GE Dash 4000 Patient
Monitor
One of the units does not have the power adaptor therefore unable
to test. Second unit powers up ok with its display and function
buttons in working order. No leads included therefore unable to
test. Both units are in reasonable physical condition.
152 Olympus 7° Rigid
Scope
Image is blurry (Optic not clear)
153 MEA Applicator Both devices appear to intact, no physical damage found. Device
not tested.
175 Dyonics 30° Rigid
Scope x 2
In good physical condition but one rigid scope has a crack in the
optics and the image from the other rigid scope is not clear. The
focus controls are working.
180 Storz Laparoscopic
Instruments x 4
In satisfactory physical condition. The handle of one of the
forceps has been damaged.
182 Olympus OTV-S4
Camera, Video
Adaptor x 2 (ARTF-2,
ARTL08E)
The camera head is in fair physical condition. No processor
therefore unable to test the image quality.
183 Olympus OTV-S5 x 2
with Camera Head
and Vista Camera
Head
All three camera heads are in fair physical condition. No
processor therefore unable to test the image quality.
232 Datex Ohmeda Cardio
Cap/5 Monitors x 4 -
10364, 15162, 13644,
All four units are able to power up with its display and function
buttons in working order except for one with damaged LCD. 2
units have Back-up battery failure error. No leads are included but
10361 one was used to test. ECG was tested using ECG simulator and a
good waveform can be obtained on all of the units. SpO2 was
working in all units except for one. All units are in poor physical
condition with corroding paint work and broken parts.
249 Datascope Passport 2
Patient Monitor with
Gas Module x 4
All 4 units are able to power up. All units pass through its initial
self-test successfully except for one in which display does not
come on and an alarm sounds. Three of the unit’s display and
function buttons are working in order. All units have gas module
but not tested. No leads or accessories included therefore not fully
tested. All units are in reasonable physical condition.
253 Sonosite Scanner ilook The unit power up ok, the LCD and function soft keys are
working fine. Good image obtained with the linear probe. A slight
damage to the corner of the probe is found but doesn’t seem to
affect the image. A selection button is missing. Sonosite charger
is included. Unit is in good physical condition.
254 Stack System to
Include Sony Monitor
Pentax EPM - 3500
Light Source, Sony
VCR, Sony Colour
Video Printer, Pentax
05-A13 White Balance
x 2, Pentax EG-3440
Video Colonoscope,
Pentax EG 3840
Videoscope, Pentaxed
341- Videoscope
All units on the stack are able to power up. The function keys and
indicators are working on the EPM 3500 video processor except
for few keys which are worn out and stiff. Videoscope (x3)
included and is mentioned under scope report lot 254. The light
source is able to produce a good light output and the air pump is
functional. The Sony monitor is working fine. The videocassette
recorder and printer power up. Keyboard included. The stack
system is in good physical condition.
255 Olympus Evis TK-E2
Procedure Cart with
Olympus CV-240
Camera Control Unit,
Olmypus CLV-U40
Light Source and
Olympus DEV191H
LCD Monitor -
507601, 2840, 2841,
2839 and Olympus
MH-977 'Pigtail'
Connector
The processor and light source are able to power up with its
function keys and indicators in working order. The light source is
able to produce a good light output. The brightness adjustment is
in working order. LCD Monitor included and is able to power up.
Camera head not included therefore unable to fully test. Keyboard
included. Overall unit is in good physical condition.
256 Drager Babylog 8000
Ventilator - 1224,
The ventilator is able to power up and complete its initial self-test
successfully. The unit’s function buttons and display are in
13048 working order. All of the ventilators modes can be selected and
was tested with medical air compressor and test lung; the
ventilator was functional at the time of test. Exhalation valve is
not included. Flow sensor cable included but no flow sensor. A
test flow sensor was used for testing. Unable to calibrate O2
(100% oxygen was not being used). Air and O2 medical gas hoses
included. The ventilator is in satisfactory physical condition.
Vent modes: CPAP, IPPV/IMV, SIMV and SIPPV
Version: 5.01
Hours of operation: 51613
258 Maquet Betastar
Electric Operating
Table (No Remote)
The operating table’s column stainless steel cover with some
damages, the castors are not moving freely. The cushions are in
average physical condition with few wear & tear; and back rest
mechanical adjustment is working fine. No remote control to test
the motorised functions.
261 Olympus WM-30
Workstation with
Sony Monitor,
Olympus EU-M30
Endoscope Ultrasound
Centre, Sony Digital
Still
The monitor, processor and Sony digital still recorder all power
up ok. No camera head included therefore unable to test. Overall
stack is in satisfactory physical condition.
263 B Braun Dialog Plus
Dialysis Machine x 3 -
23156, 22030, 23431
All units are able to power up, with their function buttons and
touchscreen displays functional. All menus accessible. The units
are not fully tested. BP not tested. All of the unit’s double blood
pumps are intact, as are the heparin pumps. Units also include one
air detectors which again are also intact. Does not include water
and drainage hoses. Overall the dialysis machines are in average
physical condition.
Version: 8.28 (x3)
Operation hours: 29167, 28630, 31786
266 Drager Incubator
8000SC
Unit powers up ok with its numerical display and function button
in working order. Inop alarm appears. No further tests were
carried out. The height adjustment is in working order. Good
physical condition with some scratches.
269 Vallylab Force FX
Diathermy (No Foot
The unit powers up and passes through its initial self-test
successfully. The numerical display and function keys are
Switch) working. Monopolar (Cut & Coag) and bipolar outputs were
tested using Electrosurgical analyser and measured values were
within the set limit. Monopolar, bipolar footswitches and hand
control leads are not included. Unit is in good physical condition
with few scratch marks on the casing and fainted paint work.
270 Erbe ICC 200
Diathermy with Erbe
APC 300 Argon
Diathermey on Trolley
- 107104, 107105
The unit powers up and passes through its initial self-test
successfully. The numerical display and function keys are
working. Unable to test the monopolar and bipolar outputs.
Monopolar footswitches included and seem to be working.
Bipolar footswitch and hand control leads are not included. APC
300 powers up ok. Not fully tested. Overall unit is in satisfactory
physical condition.
271 Olympus WM-60
Workstation To
Include Sony Monitor,
Storz Telecam DX Pal
Processor, Circon
MV9083 Automatic
Video Light Source,
Linvatec D3000
Shaver
The monitor and processor powers up ok. Not fully tested due to
no camera head available. The light source failed to power up.
The D3000 shaver powers up but no footswitch included. Unit not
tested. Overall stack is in satisfactory physical condition.
276 Datex Ohmeda
Cardiocap/5 Monitor
on Stand with Fukuda
Denshi Module Racks
x 4
The unit is able to power up with its display and function buttons
in working order. No leads included but one was used to test.
ECG was tested using ECG simulator and the values were within
limit. SpO2 was tested and is working. NIBP and pressure not
tested. Four module racks included. Average physical condition.
293 Fujifilm Dry Pix 4000 The unit power up and system self-test completed ok. The
warming up phase completed. The menu is accessible and
functioning ok. A test print was done and good printout obtained.
Unit appears to be working fine. Good physical condition.
296 Drager Babylog 8000
Ventilator - 1226, 1107
The ventilator is able to power up and complete its initial self-test
successfully. The unit’s function buttons and display are in
working order. All of the ventilators modes can be selected and
was tested with medical air compressor and test lung; the
ventilator was functional at the time of test. Exhalation valve is
not included. Flow sensor cable included but no flow sensor. A
test flow sensor was used for testing purpose. Unable to calibrate
O2 (100% oxygen was not being used). Air and O2 medical gas
hoses included. The ventilator is in satisfactory physical
condition.
Vent modes: CPAP, IPPV/IMV, SIMV, SIPPV and PSV
Version: 5.01
Hours of operation: 15908
314 HP Pagewriter XLS
ECG Recorder with
Leads
The unit power up ok and the initial screen is displayed but the
system failed to proceed to the function screen menu. The lot
comes with ECG leads but not tested. Fair physical condition.
317 Dinamap Pro 300 V2
Patient Monitor and
Criticare 506DX
Patient Monitor with
Leads
Both unit powers up ok with its display and function button in
working order. NIBP and SpO2 of Vital care 506DX is working.
Dinamap has its NIBP working whereas SpO2 not tested as there
is no probe included. Both are in fair physical condition with
some wear and tear around the buttons.
320 Datex Ohmeda ADU
Carestation
Anaesthetic Machine
with Isoflurane
Vaporiser
Unit powers up ok. The pressure gauges are working fine. O2
flush is working. The ventilator powers up and passed the initial
self-test successfully. The re-breathing circuit is disconnected and
absorber is missing. The ventilator is working. All the set
parameters and alarm system is functional. Back up battery needs
charging. The medical gas hose is included for the O2, Air and
N2O. AGSS hose included. Patient monitor power up and
completes its initial self-test successfully. Multi parameter
module, NIBP module and gas module included but not tested. No
leads included. Overall the lot is in fair physical condition with
scuffs and marks.
328 Mac 5000 GE
Marquette ECG
Machine with Leads -
100982
Mains power indicator is on when connected and the screen
comes on. Function buttons are working. ECG waveforms can be
acquired using the included leads and the analyser. The recorder is
working and the print is clear. Back-up battery needs charging.
Unit is in satisfactory physical condition with scuffs and marks
and few cracks on the casing.
332 Philips Heartstart XL
Defib with Power
Cables x 2
Both defibrillators power up and are able to pass through its initial
self-test successfully. Function buttons, selector knob and display
are in working order. Units include AED cable (x2) and are
working fine. Units showed to be functional in both AED and
manual modes when tested with defibrillator analyser. Both are
able to charge and deliver shock within the limit. ECG part of the
unit not tested. Both units have pacer mode but not tested. Audio
voice prompts and printer were functional. Battery and ECG leads
not included. Both units are in fair physical condition.
356 GE Mac 1200ST ECG
Machine (No Leads)
The unit power up ok but the initial screen cannot be displayed.
The display keeps flashing and the system failed to proceed to the
function screen menu. Unit not tested. Fair physical condition.
364 Stryker Stack trolley The processor 688i and camera head produces good image quality
on the monitor but if the camera head’s cable is move side to side
interferences can be seem on the image. This indicating broken
wires or loose contact. The Sony colour printer feed the paper
through but not printing the image. The Arthroscope Pump power
up and self-test ok. Display and function keys working fine; not
fully tested. In good physical condition. The CORE Powered
Instrument Driver power up ok and the display and touch screen
are working fine but not fully tested.
365 Valleylab Force 30
Diathermy with
Footswitches - 303041,
Valleylab Optimumm
Smoke Evacuator -
KJ01739
The unit powers up and passes through its initial self-test
successfully. The numerical display and function keys are
working. Monopolar (Cut & Coag) and bipolar outputs were
tested using Electrosurgical analyser and measured values were
within the set limit. Monopolar footswitches included and are
working fine. Bipolar footswitch and hand control leads are not
included. Smoke evacuator powers up ok. The filter needs
replacing. All the selection knobs and indicators are working. Not
fully tested. Overall unit is in good physical condition with few
scratch marks.
367 Valleylab Force 30
Diathermey with
Footswitches - 303043
The unit powers up and passes through its initial self-test
successfully. The numerical display and function keys are
working. Monopolar (Cut & Coag) and bipolar outputs were
tested using Electrosurgical analyser and measured values were
within the set limit. Monopolar footswitches included with slight
damage to the cable but are working fine. Bipolar footswitch and
hand control leads are not included. Overall unit is in satisfactory
physical condition.
370 Spacelabs Eclipse Plus
ECG Recorder (No
Leads) - CP1057
Mains power indicator is on when connected and the screen
comes on. Function buttons are working. ECG leads not included
therefore no further tests were carried out. Unit is in reasonable
physical condition with scuffs and marks and few cracks on the
casing. Unit has ‘For spare parts only’ label.
371 Drager Babylog 8000
Plus Ventilator -
ARJL - 0055
The ventilator is able to power up and complete its initial self-test
successfully. The unit’s function buttons and display are in
working order. All of the ventilators modes can be selected and
was tested with medical air compressor and test lung; the
ventilator was functional at the time of test. Exhalation valve is
not included. Flow sensor cable and flow sensor not included. A
test flow sensor was used for testing purpose. Unable to calibrate
O2 (100% oxygen was not being used). Air and O2 medical gas
hoses included. The ventilator is in satisfactory physical condition
with broken front panel cover.
Vent modes: CPAP, IPPV/IMV, SIMV, SIPPV and PSV
Version: 5.01
Hours of operation: 27890
377 Evis Olympus
Procedures Trolley
TK-E1 with Olympus
CV-240 Camera
Controller and
Olympus CLV-U20
Light Source -
2234013, C010963,
C004529
The camera head processor and light source are able to power up
with its function keys and indicators in working order. Camera
head is not included therefore unable to fully test. The light
source is able to produce a good light output. The brightness
adjustment is in working order. The monitor not included. The
procedure trolley is in satisfactory physical condition.
378 Datex Ohmeda S/5
Avance Anaesthetic
Machine (Rear Wheel
Broken)
The unit powers up and completes the system self test
successfully. Display panel, function keys and rotary control knob
are all in good working order. All ventilation modes which
include VCV, PCV, SIMV/PSV, PSV Pro & SIMV-PC are
functional. The back up battery is flat. Overall the unit is
reasonable physical and cosmetic condition (one of the wheels is
jammed). Bellows, CO2 canister and medical gas hoses are in
pace. The gas module (M-CAiO) is also included.
SN: ANBH00154. Software version: 06.10.
379 Datex Ohmeda S/5
Avance Anaesthetic
Machine
The unit powers up and completes the system self test
successfully. Display panel, function keys and rotary control knob
are all in good working order. All ventilation modes which
include VCV, PCV, SIMV/PSV, PSV Pro & SIMV-PC are
functional and the back up battery is also working. The system a
leak but overall the unit is reasonable physical and cosmetic
condition. Bellows, CO2 canister and medical gas hoses are in
pace. The unit does not include a gas module.
SN: ANBH001846. Software version: 06.10.
382 Drager Babytherm
8000 Incubator 24582
(Incomplete)
Unit powers up with its numerical display in working order. Error
70 is displayed. No further tests were carried out. Not a complete
set. Poor physical condition with missing button covers.
383 Drager Babylog 8000
Ventilator - 0034
The ventilator is able to power up and complete its initial self-test
successfully. The unit’s function buttons and display are in
working order. All of the ventilators modes can be selected and
was tested with medical air compressor and test lung; the
ventilator was functional at the time of test. Exhalation valve is
not included. Flow sensor cable and flow sensor not included. A
test flow sensor was used for testing purpose. Unable to calibrate
O2 (100% oxygen was not being used). Air and O2 medical gas
hoses included. The ventilator is in satisfactory physical
condition.
Vent modes: CPAP, IPPV/IMV, SIMV, SIPPV and PSV
Version: 5.01
Hours of operation: 27025
385 Olympus TI-1900
Imaging Trolley with
Sony Monitor - 2843,
2842
Monitor powers up ok. Three stripes of colored lines appear at the
top of the screen. No further tests carried out. The stack and the
monitor are in reasonable physical condition.
392 Spacelabs Eclipse Plus
ECG Recorder (No
Leads) - CP1663
The unit powers up ok. The display screen and keyboard is
working fine. ECG leads not included therefore no further tests
were carried out. Good print out was produced. The unit works on
backup battery pack as well as mains connection. Satisfactory
physical condition.
393 Dinamap Pro 400
Vital Signs Monitor -
4373 (with Fingerclip)
Unit failed to power up therefore unable to fully test. SpO2 probe
included. Fair physical condition with broken printer lid.
405 Storz Tricam Camera
Head - RHE632182-H
The camera head is in good physical condition. No processor
therefore unable to test the image quality.
409 Olympus OTV-S6
Camera Head -
7003366
The camera head is in fair physical condition. No processor
therefore unable to test the image quality.
409 Olympus OTV-S6
Camera Head -
7003366
The camera head is working but the image is not very clear
(showing noise in the image). Focus control is operational.
416 GE 2000 Treadmill
with GE 2000 Stress
Unit
The unit powers up successfully. Unable to boot up due to no
operating system found. Treadmill powers up ok. Acquisition
module and connection cable RS232 to the treadmill and ECG
unit is not included. No further test carried out. Unit is in
satisfactory physical condition.
425 Assorted Philips HP
and Agilent M3046A
Patient Monitors with
Multiparameter
Moduels x 5
All 5 units are able to power up successfully. All the touch
buttons and display are in working order. 5 multi-parameter
modules included. MEASSRV unsupported error message
displays on two of the units and does not recognize the modules.
Three of the units were tested with simulator and test leads and
the ECG was within limit. NIBP and SpO2 are working. No leads
or accessories are included. All units are in satisfactory physical
condition with a few worn out buttons and minor scratches on
display.
436 Olympus CLE-4E
Lightsource - 2211
CLE-4E is able to power up and produce a good light output. The
brightness adjustment is working. Air pump is functional. Overall
unit is in fair physical condition.
439 Datex Ohmeda
Cardiocap/5 Critical
Care Monitor
3 of the monitors power up and self-test completed. The displays
are working fine except for one that the screen failed to come on.
The parameters were not tested. The units are in poor cosmetic
condition with flaking body paint and corrosion on the surface.
440 Datex Ohmeda
Cardiocap/5 Critical
Care Monitor
All 4 monitors power up and self-test completed. The displays are
working fine except for one with pixel problem, with line on the
screen. The parameters were not tested. The units are in poor
cosmetic condition with flaking body paint and corrosion on the
surface.
446 Olympus Stack
System
Unit power up ok, Xenon lamp is working fine. Brightness
control, auto and manual mode fully functional. The backup
halogen lamp is also working. Unit is in good condition. The
processor power up ok and colour bar is displayed on the monitor.
No camera head to check image quality. Both Sony colour printer
comes on but with error. Stack system is in fair physical
condition.
451 Stryker 1088i HD
Camera Head -
05G032304 and
Controller - 015213
Both the video processor and the camera head are fully functional
and in good physical condition. The system produces good quality
images.
455 Olympus Maintenance
Unit MU-1 x2
Both units power up ok, the pumps are working fine. Fair physical
condition.
468 Datex Ohmeda
monitor
The Light monitor has no mains adapter and the printer assembly
is missing. The front indicator cover is also missing. Not tested.
The Cardiocap II was tested with patient simulator for ECG and
SpO2 and all the parameter s and values were within limit. The
Start/Stop soft key membrane has worn out. Fair physical
condition.
471 Storz Telecam SL The video processor powers up but a camera head is not included
to perform function check.
478 Eschmann Little
Sister 3 Steam
Autoclave.
The unit power ok, the display and cycle selector switch are
working. Only 134 degrees C with and without drying are
activated. The door micro-switch and water level sensor are
working. The chamber and door seal is in fair condition. Test
cycle completed with no error. The temperature and pressure
maintained during sterilization phase. The printer is working but
needs new ribbon. Cycle count: 990. Unit is in fair physical
condition.
482 Eschmann TD
Electrosurgical Unit
Unit power up ok, the numerical display and RF power selector
keys are working fine. The ESU was tested using an ESU
analyser. The bipolar RF outputs were well within the set values.
However, the monopolar mode has no RF power output. The unit
is in average physical condition and monopolar and bipolar
footswitches are included. The build-in suction unit and light
source are working fine.
483 Stack Trolley with
Sony Trinitron
Monitor, Storz
Hysteroflater, Downs
Camera Controller
and 1457 Camera
head and Downs Light
Source
Light source and camera head processor powers up with its
function buttons and numerical display in working order. Sony
monitor powers up however an image is failed to obtain due to
faulty connections. No further tests were carried out. Camera head
not tested. Hysteroflator powers up. Overall the stack is in
reasonable physical condition.
490 Olympus CLV-S30
Light Source
Unit power up ok, Xenon lamp is working fine. Brightness
control, auto and manual mode fully functional. The backup
halogen lamp is also working. Unit is in good condition.
491 Eschmann TD850
Electrosurgical Unit
Unit power up ok, the numerical display and RF power selector
knob is working fine. The monopolar mode was tested using an
ESU analyser and some of the RF power measured were slightly
low than the set values for cut; COAG was ok. The bipolar RF
outputs were within the set values. The unit is in fair physical
condition and monopolay and bipolar footswitches are included.
492 Philips Heartstart XL
Defib x 2 with Patient
Both defibrillators power up and is able to pass through its initial
self-test successfully. Function buttons, selector knob and display
Lead x 2 are in working order. Units do not include ECG leads but one was
used to test. Units showed to be functional in both AED and
manual modes when tested with ECG simulator. Both are able to
charge and deliver shock but not fully tested. Audio voice
prompts and printer were functional. Battery not included. Both
units are in good physical condition.
500 Valleylab Force 2
Electrosurgical Unit
The ESU power up ok and the numerical displays and output
selector keys are working fine. The monopolar mode was tested
with an ESU analyser but no RF outputs were recorder on the
analyser. The unit is in fair physical condition but no RF output.
503 HP M3046A Viridia
M4 Patient Monitor,
HP Module Rack with
Modules, HP
Omnicare 24C
Monitor
M3046A unit powers up ok. No modules included but one was
used for testing. No error messages were displayed. ECG was
within limit. NIBP seems to be working but not fully tested. The
display and the touch buttons are good working order. Unit is in
fair physical condition with cracked case and scratches on the
display.
HP Omnicare unit powers up ok. Module rack and modules (x8)
included. ECG was tested with test leads and simulator and was
within limit. NIBP and SpO2 are working. No leads or accessories
included. Recorder included but has broken paper door. Overall in
satisfactory physical condition.
505 Doctors Friend DF-
750 Suction Unit x 1,
SAM 12 Suction Unit
x 1 and Weyer Mini-
vac ACU Paed Suction
Unit x 1 - 6672, 16744,
80021108
Weyer mini-vac and DF 750 suction units powers up ok.
Manometer and suction out is working but not verified. Suction
regulator is functional. DF 750 is relatively new (with box,
surgical connecting tube and manual) and is in good physical
condition. SAM 12 unit has its top plug cut off therefore not
tested. Both mini-vac and SAM 12 are in satisfactory physical
condition.
506 Assistina 301 Plus
Units x 3 - 19931,
19811, 12516
All three units are in satisfactory physical conditions. Unable to
test functionally. One unit is not complete.
515 Penlon Prima SP
Anaesthetic Machine
with AV900 Ventilator
and Sevoflurane
Vaporisers x 2
Unit powers up ok. The pressure gauges and flow meters are
working fine. O2 flush is functional however a major leak is
detected. Basal flow and anti-hypoxic feature is working. The
AV900 ventilator powers up and passed the initial self-test
successfully. The touch screen and display are working. The re-
breathing circuit and absorber are all intact. The ventilator is
functional. All the set parameters and alarm system is working.
The medical gas hose is included for the O2, Air and N2O. AGSS
hose included. Overall the lot is in fair physical condition with
scuffs and marks.
Software version: 3.32
522 Acuson Sequoia C512
Ultrasound with
Probe x 2 (7V3C,
3V2C)
Unit powers up ok. Display, control panel and function buttons
are in working order. Both probes are in good condition with
minor drop out to the crystals. One of the probe produced good
grey scale and fair resolution with good caliper measurements and
second one showed to have average grey scale and poor
resolution, when tested with phantom. The unit seemed functional
in D, PW, CW and M modes. ECG leads not included and
therefore not tested. Sony video cassette recorder and Sony video
graphic printer powers up ok. Overall unit is in reasonable
physical condition.
532 Datex Ohmeda ADU
Carestation
Anaesthetic Machine
with Isoflurane
Vaporiser
Unit powers up ok. The pressure gauges are working fine. O2
flush is working. The ventilator powers up and passed the initial
self-test successfully. The re-breathing circuit is disconnected and
absorber is missing. The ventilator is working. All the set
parameters and alarm system is functional. The medical gas hose
is included for the O2, Air and N2O. AGSS hose included. Patient
monitor power up and completed its initial self-test. Its display
and function buttons are in working order. SpO2 was tested and is
working. Multi parameter module, NIBP module and gas module
included but not tested. SpO2 probe, NIBP trunk cable and ECG
trunk cable included. Overall the lot is in fair physical condition
with scuffs and marks.
535 Eschmann Little
Sister 2 Steam
Autoclave.
The unit power ok, but the overheat indicator is on. No further test
was conducted. Cycle count: 16979. Unit is in poor physical
condition.
536 B Braun Dialog Plus
Dialysis Machine x 4 -
35209, 21220, 23115,
23152
All units are able to power up, with their function buttons and
touchscreen displays functional. All menus accessible. The units
are not fully tested. BP not tested. All of the unit’s double blood
pumps are intact except for one with missing pump, as are the
heparin pumps. Units also include one air detectors which again
are also intact. Does not include water and drainage hoses.
Overall the dialysis machines are in average physical condition.
Version: 8.28 (x4)
Operation hours: 35136, 35209, 28877, 31762
539 Drager Phototherapy
Lamp 800
The unit power up ok, the fan and lamp are working fine.
Operating hours: 33882hrs. Average physical condition.
540 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg
adjustment). Castors, brake and mattress in good condition. The
bed is not working on battery.
541 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg
adjustment). Castors, brake and mattress in good condition. The
bed is not working on battery.
542 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg
adjustment). Castors, brake and mattress in good condition. The
bed is not working on battery.
545 Braun SCT 2
Tournequet x3
All three tourniquets were tested with air pressure and the gauges
and regulator are working fine. Units are in fair physical
condition.
553 Philips Ambulatory
ECG Recorders x 7 -
3839, 3833, 3823,
3820, 3832, Innervator
NS252 Stimulator and
Nonin 2120 Monitor -
1191, 505080
Nonin 2120 power up ok. Error E17 is displayed. No further
tested conducted. ECG recorders M2601B (x7) are in fair physical
conditions. Nerve simulator is in poor condition with worn out
buttons and broken parts.
556 Dinamap Pro 400/
Criticare Poet Plus
8100 monitor
The Dinamap power up ok, display and function keys working.
NIBP tested ok. However the unit is in poor cosmetic condition
with mold on the NIBP tubing and SpO2 probe. The Criticare
monitor power up ok and in better condition, not fully tested.
558 ASP Tournequet SCT
2
Unit is in below average physical condition, the regular and
manometers seem to be ok. Unit not tested with medical air.
563 Agilent Pagewriter
100 ECG Machine on
Trolley with Leads
Unit powers up ok and all the LEDs are lit up. Function buttons
are working. ECG waveforms can be acquired using the included
leads and the analyser. The recorder is working and the print is
clear. Unit is in satisfactory physical condition with scuffs and
marks.
577 Drager Phototherapy
Lamp x2
Both units power up ok, the fan and lamp are working fine.
Operating hours: 07271 & 28331hrs. Fair physical condition.
579 Satelec X Mind
System Dentak X Ray
system on mobile
stand
Unit power up and all the indicators, selector keys and timer
numerical display are working fine. The unit was tested and x ray
was present at the output. The power ON/OFF switch is slightly
loose but functional and there two missing castors on the mobile
stand. Unit is in fair condition.
580 Smith & Nephew
Stack System, 350
CCD Camera
Processor with
Camera Head x 3
(One Incomplete) ,
Xenon XL Light
Source, Sony Colour
Video Printer and
Sony Trinitron CRT
Monitor
All units within the stack system are in good working order. The
video processor and camera head produces good quality images
and all 3 camera heads are functional (one of the camera heads do
not include the lens/adaptor). The light source is fully functional
and the printer also appears to be working (out of paper). Overall
the stack system is in reasonable physical condition.
598 GE Dash 4000 Patient
Monitor x 2
One of the monitors power up successfully and the display
panel/function keys are in working order. The LCD is missing a
few lines of pixels and the front frame has been damaged. The
other monitor powers up but fails to complete the start up process
or reach the home screen. The casing has few small cracks and the
front panel buttons have been worn out. No accessories included
with units.
607 Penlon Sigma
Isoflorane Vapouriser
Vapouriser is in very good condition and calibrated 24/04/15.
Calibration certificate attached.
608 Penlon Sigma
Isoflorane Vapouriser
Vapouriser is in very good condition and calibrated 24/04/15.
Calibration certificate attached.
609 Datex Ohmeda
Critical care monitors
Only one of the monitor was setup and power up ok. The lot
consists of x4 module racks (no modules), x4 LCD monitor with
only one control with DVI cable; mains adapter included, bags
containing NIBP hose & cuff, SpO2 finger probes and ECG
cables. Lot is in fair physical condition.
618 Drager Infant
Incubator 8000 IC
Power up and self-test ok. The control for air temperature mode,
oxygen monitor and humidity are working. The air mode was
tested and the set temperature was maintained accordingly and the
external thermometer reading was within limit. Humidity control
is also functional and glass water bottles are fitted at the back.
The oxygen sensor completes calibration process. The canopy,
mattress and incubator frame is in good physical condition. The
motorised height adjustment is fully functional. Oxygen hose
attached.
619 Drager Infant
Incubator 8000 IC
Power up and self-test ok. The control for air temperature mode,
oxygen monitor and humidity are working. The air mode was
tested and the set temperature was maintained accordingly and the
external thermometer reading was within limit. Humidity control
is also functional and glass water bottles are fitted at the back.
The oxygen sensor completes calibration process. The canopy,
mattress and incubator frame is in good physical condition. The
motorised height adjustment is fully functional. Oxygen hose
attached.
622 Dinamap Pro
300/300V2/100 &
compact TS
All four NIBP monitor power up ok, display and functions keys
are working; a few of the soft keys has worn out but functional.
The NIBP was tested on all the units and was working fine. The
Compact TS unit SpO2 is working ok but the finger probe spring
is damaged. Lot is in fair physical condition.
625 Keeler Cyromatic
Cryounit with Foot
Pumps and
Connection Leads
Powers up and the display panel is functional. Unable to carry out
any further checks. The unit is in good physical and cosmetic
condition. Footswitch and 2 probes/handpieces are included. The
footswitch connector is slightly damaged.
642 Zeiss ENT Operating
Microscope
The light source power up ok, bulb is working. Good optical
system with eye pieces and objective lens. The motorised zoom &
focus control and joystick is damaged. Average physical
condition.
643 HGM Elite
Ophthalmic Laser
System with Slit Lamp
The Slit lamp power supply, bulb, fixation light and brightness
control are in good working order. All the slit lamp movement
and lock were functional. The motorised table height adjustment
is working fine. The Elite laser system power up ok the display
and laser parameter selector are working. The unit was tested on
thermal paper and there was indication of laser present at the
output. However the laser output power was not verified. Safety
goggles x3, Laser Key, footswitch and carry case included. Unit is
in good physical condition..
644 Leica CM 1100-
Kryostat on KE
Standard Supertrans
Demtruk Trolley
Leica CM 1100 powers up ok. Its control unit, hand wheel and its
lever, coarse feed wheel and the chamber lid is in good condition.
The clamping lever, knife holder, waste tray and section waste
tray are included. No further tests were carried out. Temperature
and defrost time can be adjusted. The compressor is slightly
noisy. Overall lot is in good physical condition.
646 GE Capnostat CO2
Sensors
Lot is in good physical condition with adult x3 & neonatal x2
away adapters. Sensor not tested.
649 GE E-Prestn
Multiparameter
Modules x 2 - 016985,
E-PRESTN (x2) modules are in satisfactory physical condition
with scuffs and marks. SpO2, temperature, pressure, NBP and
ECG parameters were not tested.
016987 YOM: 2007 (x2)
672 Olympus Stack system
with processor CV-
230 & Light source
CLV-U40
The light source power up ok, the xenon lamp is on and
brightness, air pump and auto/manual working fine. The processor
was tested with a camera head but no scope to determine the
picture quality. The Sony monitor is not working properly, the
video channel fail to switch over or change. Average overall
physical condition.
689 Lifepak 20 & 12 with
charger and leads
All Lifepak 20 power up and self-test completed successfully. The
LCD and all function buttons are in good working order. The
defibrillator was tested using a defibrillator analyser and the ECG
traces and delivered energy was within the set limit. Both AED
and manual defibrillation modes are working and so is the
monitor. The voice prompt is also working fine. Lot is in good
overall condition. Backup battery is not working and there is no
pacer option or paddles. The service light is on.
All Lifepak 12 power up and self-test completed successfully. The
LCD and all function buttons are in good working order. The
manual defibrillator was tested using a defibrillator analyser and
the ECG traces and delivered energy was within the set limit. The
monitor is also working fine. Lot is in good overall condition. No
battery packs or paddles included.
The two Medtronic Battery Support System 2 power up ok and
self-test completed. Units are in good condition but not fully
tested. The lot also come with AED therapy cables x8 and ECG
cable (3 wires) x7
704 Philips Heartstart
AED
All AED power up and self-test completed successfully. The LCD
and function buttons are in good working order. The AED was
tested using a defibrillator analyser and the ECG traces and
delivered energy was within the set limit. The voice prompt is
also working fine. Lot is in good overall condition. Battery packs
are not included.
705 Philips Heartstart
AED
All AED power up and self-test completed successfully. The LCD
and function buttons are in good working order. The AED was
tested using a defibrillator analyser and the ECG traces and
delivered energy was within the set limit. The voice prompt is
also working fine. One has a poor casing but the rest are in good
overall condition. Battery packs are not included.
706 Philips Heartstart
AED
All AED power up and self-test completed successfully. The LCD
and function buttons are in good working order. The AED was
tested using a defibrillator analyser and the ECG traces and
delivered energy was within the set limit. The voice prompt is
also working fine. One has a poor casing but the rest are in good
overall condition. Battery packs are not included.
724 Sonosite Scanner ilook The unit power up ok, the LCD and function soft keys are
working fine. Good image obtained with the linear probe. The
modes, CPD & 2D are functional. Charger is included. Unit is in
good physical condition. A lock screw for the unit tilt is missing.
2 battery packs included.
726 Critikon Dinamap
Vital Signs Monitor
8100T
All power up & display self-test ok but follow with continuous
audible alarm. Unit not working correctly. Fair physical
condition
732 Drager Infant
Incubator 8000 SC
Power up and self-test ok. The Control for Skin and air modes is
working. The skin probe is not included there this mode was not
fully tested. The air mode was tested and the set temperature was
maintained accordingly and the external thermometer reading was
within limit. Humidity control is also functional. The canopy,
mattress and incubator frame is in good physical condition. The
motorised height adjustment is fully functional.
736 Tournequet Unit is in average physical condition. The timer comes on; the
connection for the medical air is missing. The unit was not fully
tested with air pressure. The regulator and manometers appears to
ok.
744 Sonosite Ultrasound
system 180 Plus
The unit power up ok, the LCD and keyboard is working but there
are some of the soft keys not responding. Good image obtained
with the convex and linear probe. The modes, D, CPD, M, & 2D
are functional. Charger is included. Unit is in good physical
condition.
754 GE Dinamap Pro 400
x4
All power up ok, display and function keys are working. BP was
tested and working; SpO2 function not fully tested. Lot comes
with NIBP hose and cuff, most of the cuff Velcro has worn out.
Units are in fair physical condition
755 GE Marquette 2000
Treadmill with GE
Case Stress Best Unit
with Suntech Tango
The unit powers up successfully. Unable to boot up due to no
operating system found. Treadmill powers up ok. Acquisition
module and connection cable RS232 to the treadmill and ECG
unit is not included. No further test carried out. Unit is in
BP Monitor satisfactory physical condition.
780 Eschmann Little
Sister 2000 Steam
Autoclave.
The unit power ok, the display and cycle selector switch are
working. The door micro-switch and manual water fill are
working. The chamber and door seal is in fair condition. Test
cycle completed with no error. The temperature and pressure
maintained during sterilization phase. Unit is in fair physical
condition.
783 Esaote AU4 Idea
Ultrasound with
Probes x 2 (LA14-
10Mhz and LA13-
7.5Mhz)
The unit power up ok, the display and control keypad is working.
Good image obtained with two probes. The modes 2D, PW and M
are functional. No further test carried out. Unit is in good physical
condition.
789 Hill-Rom Birth Bed All motorised positions are working ok (height, back rest and leg
adjustment). Castors, brake and mattress in good condition. The
bed is not working on battery, needs charging. Good overall
condition.
790 B Braun Dialog Plus
Dialysis Machine x 3 -
22023, 22027, 21192
All units are able to power up, with their function buttons and
touchscreen displays functional. All menus accessible. The units
are not fully tested. BP not tested. All of the unit’s double blood
pumps are intact except for one with missing pump. Units also
include one air detectors which again are also intact, as are the
heparin pumps. Does not include water and drainage hoses.
Buttons for one of the unit is slightly worn out. Disinfection
documents attached. Overall the dialysis machines are in average
physical condition.
Version: 8.28 (x3)
Operation hours: 38900, 22847, 32134
794 Stryker X7000
Lightsource, Stryker
1088i HD Camera
Control Unit with
Camera Head,
Stryker Flow Control
Arthroscopy Pump
and Stryker Core
Powered Instrument
Driver, Sony Printer -
508661, 015215,
1088i HD camera processor powers up ok and is able to produce
good image when tested with the camera head and the monitor in
lot 879. The camera control buttons are operational. X7000 light
source is able to power up but lamp is missing therefore unable to
fully test. Instrument driver, arthroscopy pump and Sony video
graphic printer all power up ok. Not fully tested. The touch screen
and function buttons of the instrument driver is working in order.
No footswitches or hand pieces included. The knob for locking
the optic cable on the light source is missing. Monitor not
included. Overall the lot is in good physical condition.
015205, 508665,
015210
812 Datex Ohmeda S/5
Monitor
Only two of the monitor power up. Self-test completed and the
LCD is working fine. But the second one has a green bar on the
left hand side of the display. The complete working was tested
with patient simulator and the ECG and Spo2 was both working
fine. X2 with front panel damage and missing knobs.
816 Dinamap Pro Datex
Cardiocap II Patient
Monitor
The monitor power up ok, the CRT display is working fine. ECG
& SpO2 was tested using a patient simulator and all the reading
and traces displayed were with set limits. There was an error
message ‘memory circuit failure’ flashing on the display. Unit is
in fair physical condition.
830 Eschmann Little
Sister 3 Steam
Autoclave.
The unit power ok, the display and cycle selector switch are
working. Only 134 degrees C with and without drying are
activated. The door micro-switch and water level sensor are
working. The chamber and door seal is in fair condition. Test
cycle completed with no error. The temperature and pressure
maintained during sterilization phase. Cycle count: 2511. Unit is
in fair physical condition with a bit of rust on the door frame.
831 Eschmann Little
Sister 3 Steam
Autoclave.
The unit power ok, the display and cycle selector switch are
working. Only 134 degrees C with and without drying are
activated. The door micro-switch and water level sensor are
working. The chamber and door seal is in fair condition. Test
cycle completed with no error. The temperature and pressure
maintained during sterilization phase. Cycle count: 8264. Unit is
in fair physical condition with a bit of rust on the door frame.
837 Leica Microscope The unit has no eye pieces or objective lens. The power supply is
missing as well. Poor condition.
838 Leica Microscopes x2 The microscopes are in poor condition with the optics dirty. One
of them with a corroded stage. Come with some objective lens
and eye pieces.
847 Datex Cardiocap II
Patient Monitor
The unit failed to power up.
868 Drager Infant
Incubator 8000 SC
Power up and self-test ok. The Control for Skin and air modes is
working. The skin probe is not included there this mode was not
fully tested. The air mode was tested and the set temperature was
maintained accordingly and the external thermometer reading was
within limit. Humidity control is also functional. The canopy,
mattress and incubator frame is in good physical condition. The
motorised height adjustment is fully functional.
873 GE Vingmed System
Five Performance
Ultrasound with 5Mhz
x 1, 3.5Mhz x 1,
2.5Mhz x 1 and
10Mhz Probe x 1,
Footswitch and VCR -
YZ7112
Ultrasound powers up but an error message is displayed and the
system fails to proceed to main screen. No further tests carried
out. Printer and videocassette recorder powers up. Reasonable
physical condition.
876 Operating Table
accessories
Accessories in fair physical condition consisting of Anetic Aid leg
support x2 of different shape, Maquet Beta leg section, spare
cushion & one arm support.
879 Stryker Stack Trolley
with SV-2 HD
Monitor
Monitor is able to power up ok. The screen is slightly scraped at
the middle with few scratches but does not seem to affect the
image when tested with processor in lot 794. The stack and the
monitor are in satisfactory physical condition.
883 Smith & Nephew
Trolley with Sony
LCD Monitor, Storz
Trolley with Sony
LCD monitor
Both Sony LCD monitors power ok, good display and menu
functional. The trolleys are in good physical condition. Sony LCD
monitor Model: MD-214MD (100-240V) & MD1950MD/HD
(24V DC), mains adapter included.
890 Olympus
EndoscopeTrolley
The trolley is in fair physical condition, the pole fitting is missing,
and castors are ok.
899 Olympus CLV-U40
Light Source
The light source power up ok, The Xenon lamp is working,
brightness control is functional and all the light output modes are
working ok. The air pump is functioning fine. Unit is in good
physical condition.
908 Eschmann Little
Sister 2000 Steam
Autoclave.
The unit power ok, the display and cycle selector switch are
working. The door micro-switch and water level sensor are
working. The chamber and door seal is in fair condition. Test
cycle completed with no error. The temperature and pressure
maintained during sterilization phase. Unit is in fair physical
condition.
937 Datex Ohmeda S/5
patient monitor
Lot consists of LCD monitor & control pad x9, Module racks x9
M-NESTPR modules x6, M-PP module x4 & M-PT modules x6.
There is also a S/5 Light Monitor. It power up ok and the display
and function keys are working. S RAM error is displayed. Some
of the setup has been lost, with the SpO2 plug in there was no
response on the screen. Fair & physical condition but lot has no
interconnecting leads or patient cables.
978 Humphrey 350 Lens
Analyser - 303076
Powers up and the display panel/function keys are in good
working order. All system menus are accessible, no further checks
performed. The unit is in good physical and cosmetic condition.
Model: 350. Version: A3.
982 Datex Ohmeda S/5
Monitor with qty of
accessories
The lot consists of the following: Module rack x10, LCD
monitors x3, M-PRESTN module x10, M-NESTPR module x5,
M-CAIO x3, M-COPSv x2 & M-miniC x4. Quantity of ECG
leads, SpO2 finger probe and extension cable, inter connecting
cables, NIBP hose & cuffs. Only one of the patient monitor was
set up and tested and appears to be working fine. The ECG, SpO2
and NIBP were tested and were functional. The lot is in fair
physical condition, not all units were tested.
985 Nidek KM-500
Keratometer and
Datex M-Caio Module
Mains adaptor for the charger is not included. Unable to perform
any function check. Both the keratometer and the M-CAiO
module are in reasonable physical condition.
1003 Storz Stack trolley
with LCD monitor
The trolley power up ok and the LCD is working fine. The trolley
and monitor is in good physical condition. Keyboard and HP
printer included.
1059 Kodak Point of Care
CR140- X-Ray Printer
- 800195
X-ray printer powers up ok. No workstation included therefore
unable to fully test. No x-ray cassettes included. Overall lot is in
fair physical condition.
1076 Siemens VF13-5
Ultrasound Probe -
F13C055335
The probe is in satisfactory physical condition with damages at
the corners and a slight cut to the membrane of the probe head.
Probe is not fully tested as there is no suitable ultrasound unit to
test with.
1117 Stryker Insufflator The unit power up but with error messages – ‘Call Service’,
‘Contamination’ & ‘Device defective’. Unit not working properly.
Cylinder connection missing.
1142 Medelec Thermal
Threshold Tester -
6438, Agilent Docking
Stations x 2 and
Philips Modules x 2 -
11900, 11870, 15409,
Ardo Calypso Breast
Pump, Freewat Elite
Threshold tester power up ok. Not tested. Fair physical condition.
M3015A Philips module is in good physical condition. ECG,
Pressure, temperature, NBP and SpO2 parameters were not tested.
YOM: 2004
Cpap, Assorted
Thermometers x 3 -
573432, 512656,
512657