Upload
mer
View
214
Download
0
Embed Size (px)
Citation preview
ESTRO 31 S537
of the radioligand was estimated using three ROI delineation methods
varying in complexity and execution time. ROIs were drawn on a)
antero-posterior compressed PET images (AP), b) subsamples of the
organs (S), and c) a 3D-volume covering the whole-organ (W).
Residence times for each organ were calculated from the time course
of the radioligand. The OLINDA/EXM software package was used to
obtain dosimetry estimates.
Results: Measured time activity curves (TACs) showed a close
agreement with the simulated TACs when S and W ROI delineation
methods were used. When the AP method was used, measured TACs
displayed higher activity than simulated TACs in the organs where the
ROI enclosed the whole-organ (heart, lungs, liver and kidneys). The
radiation-absorbed dose (AD) from the critical organ (liver), the
effective dose equivalent (EDE) and the effective dose (ED) estimates
using S and W methods were comparable to the simulated values
(theor.) while the AP method yielded higher AD, EDE and ED estimates
(table 1).
Conclusions: ED estimations are sensitive to the delineation method
applied. 3D methods (S and W) provided the most accurate dosimetry
estimates while the AP method yielded higher dosimetry estimates
because of overlapping between ROIs. The S method showed the best
trade-off between accuracy and practical implementation.
EP-1412
DOSIMETRIC CHARACTERIZATION OF A CONE-BEAM CT FOR
RADIOTHERAPY
T. Martínez Jurado1
, D. Rodríguez Latorre1
, N. Jornet1
, A. Ruiz1
, A.
Latorre Musoll1
, P. Carrasco de Fez1
, T. Eudaldo1
, M. Ribas Morales1
1
Hospital de la Santa Creu i Sant Pau, Servei de Radiofísica i
Radioprotecció, Barcelona, Spain
Purpose/Objective: Many of the currently available linear
accelerators have a gantry mounted kV cone-beam CT (CBCT) with a
wide beam. The dosimetric characterization is not straightforward as
standard CT dosimetric indices and dose measurement equipment are
designed for narrow collimation widths. The aim of this work was to
characterize our CBCT for CT dose considerations by measuring the
dose profile in a long phantom and to quantify the differences in the
measurement of CTDIw index when using different detectors and
phantoms.
Materials and Methods: Measurements were performed on a Clinac
2100C/D (Varian Medical Systems) equipped with a kV imaging OBI
system v1.5 in the pre-defined pelvis mode: 125 kV, 680 mAs and 20.6
cm nominal beam width (y1=y2=10.3 cm).We used a 15 cm long CTDI
standard body phantom and a 45 cm cylindrical phantom made by
joining three CTDI standard body phantoms. Measurements were
performed using two detectors: a solid-state detector RTI CT Dose
Profiler and a CT pencil ion chamber 100 mm long PTW Freiburg
M30009-0186. We measured dose profiles along the couch axis at the
centre and periphery of the 45 cm long phantom point by point with
the solid state detector; we used a spatial resolution of 1 cm and
recorded the dose of a complete axial scan in every point. The couch
was moved over a range of 46 cm. The phantom always covered the
whole beam width at the isocenter. We also made dose measurements
at the centre and the periphery of the phantom, first using the pencil
ion chamber and the 15 cm long phantom, and then using the 45 cm
long phantom. To calculate CTDIw and CTDIw,∞, and determine CTDIw
efficiency we applied IEC 2009 and recent IAEA Human Health Reports
Nº 5 definitions.
Results: Measures performed using a 15 cm long phantom
underestimated CTDI100 by 0.78 in the centre and by 0.95 at the
periphery compared to measures made in a 45 cm long phantom. (see
Table 1). CTDIw was underestimated by 0.91. Our determined CTDIw
were larger than the CTDIw data by Varian. Although there was a
significant difference, the CTDIw values obtained are compatible with
those reported by other authors. By means of the dose profile we
calculated the CTDIw, which was similar to the CTDIw measured with
the pencil ion chamber and the 45 cm long phantom. We also
calculated the CTDIw,∞ defined as the weighted CTDI, considering
the whole dose profile, and the CTDIw efficiency given by the ratio
between the CTDIw and the CTDIw,∞.
Conclusions: CT dosimetric indices are underestimated when
dosimetric parameters and dose measurement equipment designed for
narrow collimation widths are applied to CBCT as the phantom and
the detector lengths are smaller than the whole beam. Therefore,
only a part of the primary radiation and a part of the scattered
radiation are measured. Measuring the dose profile with a solid
detector in a 45 cm long phantom is an accurate method to
characterize the properties of the CBCT. However, a special phantom
is required and about 2h of treatment unit per profile are needed.
EP-1413
ACCEPTANCE TESTS PROCEDURE VERSUS CLINICAL IMAGING QUALITY
ASSURANCE FOR KV-CBCT IMAGE SYSTEM
J. Galhardas1
, R. Malveiro1
, M.E.R. Poli1
1
University Hospital Santa Maria, Medical Physics Unit, Lisboa,
Portugal
Purpose/Objective: The aim of this study is to compare the imaging
acceptance tests procedure (ATP) from manufacturer to the imaging
quality assurance (QA) for clinical imaging setup (preset) using
imaging quality parameters (defined on the clinical presets) to deliver
a kV cone-beam CT (CBCT) adequate for clinical use.
Materials and Methods: The kV-CBCT images were acquired using the
Elekta Synergy S linac with integrated X-ray Volume Imaging (XVI)
system.Tests were performed in a Catphan 600 phantom and a TOR
18FG (Leeds) phantom X-ray imaging with 1 mm Cu filter.
The ATP was performed according to the manufacturer instructions
and for the clinical imaging QA, the tests were: (1) 3D Uniformity, (2)
3D Low contrast visibility, (3) 3D Spatial Resolution, (4) 3D Transverse
vertical and (5) horizontal scale, (6) 3D Sagittal geometric, (7) 2D Low
contrast visibility, (8) 2D Spatial resolution and (9) Imaging doses
using CBCT dose index (CBDI).
For both, ATP and clinical imaging QA, the material and equipment
setup used to perform both set of tests were the same and all the
tests were performed under the same conditions (e.g. heat units
between 4% and 7%).
The differences between the ATP and clinical QA tests relied on the
imaging quality parameters such as detector resolution,
S538
reconstruction algorithm, acquisition angle and kV tha
the presets. The ATP required specific presets and th
the presets available for clinical routine.
Imaging doses were estimated for each preset using
TM30009 PTW and a cylinder phantom with 16 cm
Delivery time, reconstruction time and image s
evaluated.
Results: Table 1 shows the results for both ATP and
different presets. Using the same tolerances from ATP
the kV-CBCT QA fail for all the clinical presets. In ord
in these tests, the imaging quality parameters of the
were changed. By increasing the detector resol
acquisition angle, the quality of CBCT image can be im
impact was the increase of dose, delivery time, reco
and image size on disk, up to 70%, 66%, 1090% and 334%
Test Parameter
ATP
Preset
Head
and
Neck
S20
Chest
M20
Pelvis
M20
Prosta
M15
ATP QA QA QA QA
1 uniformity P F F F F
2 contrast P F F F F
3 resolution P F F F F
4 scale P P P P P
5 scale P P P P P
6 scale P P P P P
7 contrast P P P P P
8 resolution P P P P P
Table 1 - Results for ATP and clinical QA tests, where
is Fail.
Conclusions: The tolerances from ATP could not be
the clinical tests and should not be used as reference
QA. Performing an imaging QA for clinical presets is an
to define references for quality control tests and to im
images based on clinical routine.
One can note that image system used in radiotherapy
pass the acceptance tests but fail the intent for its pur
EP-1414
THE SECONDARY EXPOSED DOSE PRODUCED BY THE V
GUIDED RADIATION THERAPY SYSTEMS
T. Isobe1
, K. Takada1
, E. Sato1
, K. Shida1
, D. Kobayashi
Mori1
, H. Sakurai1
, T. Sakae1
1
University of Tsukuba, Graduate School of Comprehen
Sciences, Tsukuba, Japan
2
Tsukuba University Hospital, Department of Radiology
Japan
Purpose/Objective: Image-Guided radiation thera
reduce the setup error in radiation therapy and the c
increasing. On the other hand, various systems to achi
produce an excessive exposed dose. It is anxious abo
the carcinogenic risk by this excessive exposed dose. I
compared the secondary exposed dose associated wi
systems.
Materials and Methods: In this study, we measured
exposed dose produced by various IGRT systems
dosimeters and the human phantom. The glass dosime
because it is possible to measure at several points
phantom. The system used in this study was the
accelerator (Varian Medical Systems, Palo Alto, CA)
First, we verified the precision of glass dosimeter.
measured value of glass dosimeters with calibr
chamber using the X-ray simulator (Toshiba Medical S
In consideration of the X ray energy generated from
simultaneous irradiation of these detectors was perfo
X ray energies (40 kV-120 kV), and measured values w
Next, the secondary exposed dose produced by vario
(Orthogonal On-board Image, Cone beam CT, Ex
measured. We placed the glass dosimeters on cent
(anterior, left side, right side, posterior) of a phantom
were performed 3 times respectively. The parameters
at are defined in
he QA tests used
an ion chamber
m of diameter.
size were also
d clinical QA for
P, tests 1 to 3 of
der to get a pass
clinical presets
lution and the
mproved but the
onstruction time
% respectively.
ate
ATP
tolerance
≤1.5%
≤1.5%
≥10 lp/cm
±1.0mm
±1.0mm
±1.0mm
≥12 discs
≥1.4 lp/mm
e P is Pass and F
achieved during
e for the clinical
n important step
mprove kV-CBCT
can successfully
rpose.
VARIOUS IMAGE-
2
, K. Suzuki2
, Y.
nsive Human
y, Tsukuba,
py (IGRT) can
clinical utility is
eve an IGRT will
out elevation of
n this study, we
ith various IGRT
d the secondary
with the glass
eter was chosen
s in the human
e Trilogy linear
We related the
ated ionization
Systems, Japan).
m IGRT systems,
ormed in various
were compared.
us IGRT systems
xactrac®) were
er and surfaces
m. Measurements
of imaging were
applied the clinical settings. Dose evaluation
kerma.
Results: As a result of comparing glass dosim
chamber, the glass dosimeters have underest
maximum compared to the ionization cha
produced by each IGRT systems became t
mGy, OBI=11.01 mGy, and Exactrac =1.34 m
dose, it became the highest by CBCT=18.88 m
about secondary exposed dose associated
documented. The result of present study (29
18.88 mGy at the center) was consistent
(Table 1) If it is performed daily CBCT
fractions), we can estimate that it become
dose of 1.1 Gy at the center and 0.7 Gy at
most of secondary exposed dose associated
CBCT, it is necessary to take into account
parameters of CBCT to reduce the patient do
Table 1 Comparison of the secondary expos
CBCT
absorbe
surface
present study 29.75
Wen N, et al.
(Phys Med Biol. 52, 2007 ) 30~60
Song WY, et al.
(Med Phys. 35, 2008) n.e.
Amer A, et al.
(BJR. 80, 2007) 21~34
Islam MK, et al.
(Med Phys. 33, 2006) 23
n.e. : not estimated
Conclusions: The secondary exposed dose
IGRT systems was highest in CBCT, and it be
of OBI, and Exactrac. In the case of per
prostate IMRT, it was suggested that the seco
1.1 Gy at the center and 0.7 Gy at the surfac
EP-1415
COMPUTATIONAL LYMPH NODE MODELS IN RE
RADIONUCLIDE THERAPY DOSIMETRY
S. Lamart1
, M.B. Wayson2
, W.E. Bolch2
, C. Lee
1
National Cancer Institute National Institutes
Cancer Epidemiology and Genetics, Rockville
2
University of Florida, J. Crayton Pruitt Fami
Biomedical Engineering, Gainesville FL, USA
Purpose/Objective: Explicit radiation dose c
the lymphatic node system in nuclear medic
using surrogate tissues since its computatio
available in traditional stylized phantoms
existing clinical dosimetry tools such as OL
develop a new lymph node model within a ne
human phantoms, called hybrid phantoms,
examples of its application to radionucl
Materials and Methods: The lymph node mod
ESTRO 31
was performed by the air
meters with an ionization
timated about 9.6% at the
mber. The surface dose
he order of CBCT=29.75
mGy. Also in the central
mGy. (Fig. 1) The reports
with CBCT were well-
9.75 mGy at the surface,
with the other reports.
for prostate IMRT (39
es the secondary exposed
the surface. Because the
d with IGRT depends on
t the frequency and the
se to healthy tissue.
sed dose associated with
ed dose (mGy)
center
18.88
n.e.
16
21
16
associated with various
ecame lower in the order
rforming daily CBCT for
ondary exposed dose were
ce of the human phantom.
EFERENCE PHANTOMS FOR
e1
s of Health, Division of
MD, USA
ily Department of
calculation accounting for
cine has been performed
onal model has not been
which are the basis of
LINDA. This work aims to
ew class of computational
and to show illustrative
lide therapy dosimetry.
dels were developed for a