© 1997 Waters Corporation Fast and Easy Solid-Phase Extraction with a New Reversed-Phase Sorbent Dorothy J. Phillips, Laura L. Bean, Edouard S.P. Bouvier, Mark Capparella, Yung-Fong Cheng, Pamela C. Iraneta and Uwe D. Neue chemistry in the discovery programs has resulted in an increase in the number of drug candidates in the preclinical phase. Drug metabolism groups now have many more candidates to screen and need to work even faster to shorten the drug discovery cycle. Hence, there is a critical need for more efficient methods to extract and analyze samples from bio- logical fluids. The analytical technique that is stepping up to the challenge is LC/MS/MS which can analyse samples at a rate of one sample every 2–5 minutes. Sample preparation, required for the analysis of most drugs in The Challenge Over the past five years, pharmaceutical companies have made a fundamental change in the way they conduct their drug discovery programs. Their preferred discipline now is combinatorial chem- istry. Combinatorial chemistry integrates such tools as novel synthesis, reaction equipment and advanced information systems to accelerate the generation and optimisation of pharmaceutical candi- dates (1). The ultimate vision for a combinatorial or high throughput chem- istry laboratory is the synthesis of millions of individual, pure compounds each day (1). The implementation of combinatorial biological fluids, is now the bottleneck and needs to be automated. But liquid-liquid extraction and protein precipitation methods do not lend them- selves to automation. The method of choice is solid-phase extraction (SPE). Using robotics and miniaturisation has enabled GlaxoWellcome to process 50,000 samples a day (2). Many other pharmaceutical companies are struggling to increase sample throughput with vacuum manifolds and robotic sample processors. 1 Table of Contents Solid-Phase Extraction ...........................1-7 Improving Method Development ...........8-12 Evaluation of Imitation μBondapak Columns........................13-15 How to Reach Us...................................15 Papers Presented at HPLC ‘97 ............16-17 What’s New .........................................18 Application Notes................................19-22 Androgens: Testosterone Esters Morphine and Its Metabolites Editor: Uwe D. Neue, Ph. D. ISSN # 1084-0540 Waters Oasis HLB extraction cartridge — the most important advance in solid-phase extraction in 20 years. Volume VI Issue 4, 1997

Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

Embed Size (px)

Citation preview

Page 1: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

© 1997 Waters Corporation

Fast and Easy Solid-Phase Extraction with a New Reversed-Phase SorbentDorothy J. Phillips, Laura L. Bean, Edouard S.P. Bouvier, Mark Capparella, Yung-Fong Cheng, Pamela C. Iraneta and Uwe D. Neue

chemistry in the discovery programs has

resulted in an increase in the number of

drug candidates in the preclinical phase.

Drug metabolism groups now have

many more candidates to screen and

need to work even faster to shorten the

drug discovery cycle. Hence, there is a

critical need for more efficient methods

to extract and analyze samples from bio-

logical fluids. The analytical technique

that is stepping up to the challenge is

LC/MS/MS which can analyse samples

at a rate of one sample every 2–5

minutes. Sample preparation, required

for the analysis of most drugs in

The ChallengeOver the past five years, pharmaceutical

companies have made a fundamental

change in the way they conduct their

drug discovery programs. Their preferred

discipline now is combinatorial chem-

istry. Combinatorial chemistry integrates

such tools as novel synthesis, reaction

equipment and advanced information

systems to accelerate the generation and

optimisation of pharmaceutical candi-

dates (1). The ultimate vision for a

combinatorial or high throughput chem-

istry laboratory is the synthesis of millions

of individual, pure compounds each day

(1). The implementation of combinatorial

biological fluids, is now the bottleneck

and needs to be automated. But

liquid-liquid extraction and protein

precipitation methods do not lend them-

selves to automation. The method of

choice is solid-phase extraction (SPE).

Using robotics and miniaturisation has

enabled GlaxoWellcome to process

50,000 samples a day (2). Many

other pharmaceutical companies are

struggling to increase sample throughput

with vacuum manifolds and robotic

sample processors.

1

Table of Contents

Solid-Phase Extraction ...........................1-7

Improving Method Development...........8-12

Evaluation of Imitation µBondapak™ Columns........................13-15

How to Reach Us...................................15

Papers Presented at HPLC ‘97............16-17

What’s New .........................................18

Application Notes................................19-22

Androgens: Testosterone Esters

Morphine and Its Metabolites

Editor: Uwe D. Neue, Ph. D.

ISSN # 1084-0540

Waters Oasis™ HLB extraction cartridge — the most important advance in solid-phase extraction in 20 years.

Volume VIIssue 4, 1997

Page 2: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

2© 1997 Waters Corporation

Traditional ProblemsDevelopment and implementation of

solid-phase extraction methods can

be tedious. It is a challenge to devise

methods that deliver high recoveries

reproducibly. For at least three reasons,

traditional SPE products, whether used

with manual vacuum manifolds or robotic

systems, often do not achieve this goal.

First, the silica-based reversed-phase

sorbents in these traditional SPE products

weakly retain the more polar drugs and

metabolites. Such weak retention causes

sample breakthrough yielding unpre-

dictably low recoveries.

Second, optimising a method for

reproducible recoveries can be time-

consuming because traditional sorbents

contain a heterogeneous population of

surface silanol groups which have a

strong and variable retentive effect on

basic analytes.

Finally, perhaps the major contributor

to irreproducibly low recoveries is the

inability of the traditional silica- and

polymer-based reversed-phase sorbents

to be wetted with water. If these sorbents

are allowed to run dry at any time

before the sample loading step in an

SPE protocol, only a fraction of the

sorbent’s available surface will be wetted

by an aqueous sample. Adsorption

capacity for drugs and metabolites will

thereby be reduced, resulting in low,

variable recoveries (3,4).

A Unique SolutionWaters new Oasis™ HLB extraction car-

tridges overcome these limitations of

traditional silica- and polymer-based

reversed-phase SPE products (3,4).

Oasis™ HLB extraction cartridges contain

a novel polymeric sorbent (patent

pending) that makes it fast and easy to

develop reproducible extraction methods

yielding high recoveries. The simple

reversed-phase extraction mechanism of

this unique new sorbent offers stronger,

more predictable retention of polar

drugs and metabolites, including basic

compounds.

But most important, Oasis™ HLB extrac-

tion cartridges eliminate the major cause

of variability in solid-phase extraction.

Unlike traditional hydrophobic reversed-

phase sorbents, Oasis™ HLB sorbent is

fully wettable with water. This means

there is no longer a need to monitor

vacuum levels, program robotic adjust-

ments, or twist stopcocks to keep

sorbent beds from drying out during the

critical cartridge conditioning, equilibra-

tion, and sample loading steps.

Oasis™ HLB sorbent is a macroporous

copolymer (poly(divinylbenzene-co-N-

vinylpyrrolidone) composed of both

lipophilic and hydrophilic monomer

units. This hydrophilic-lipophilic balance

gives Oasis™ HLB sorbent its unique

reversed-phase extraction performance

advantages (and its name):

• High and reproducible recoveries for

acidic, basic and neutral drugs with

a wide range of polarities.

• Simple and rapid methods

development for drugs and

metabolites.

• Easily automated methods for high

sample throughput.

• Streamlined manual or automated

analyses since the sorbent can run

dry during the process without

affecting the reproducibility of results.

Bouvier et al. discussed the theory and

advantages of the Oasis™ HLB sorbent’s

water wettability in earlier papers

(3, 4). In this paper, we report the

extraction of drugs from serum using

Oasis™ HLB extraction cartridges. A

simple SPE method gave reproducibly

high recoveries for acidic and basic

drugs of varying polarity.

Experimental

Reagents and Materials Salicylic acid, benzoic acid, propra-

nolol, dipropyl phthalate and

butylparaben were purchased from

Aldrich Chemical Company

(Milwaukee, WI). Procainamide, raniti-

dine, acetaminophen, sulfanilamide,

propranolol, doxepin hydrochloride,

betamethasone valerate, minocycline,

tetracycline and demeclocycline were

obtained from Sigma Chemical

Company (St. Louis, MO). Acetonitrile,

methanol, potassium phosphate, sodium

phosphate, sodium chloride, glacial

acetic acid and phosphoric acid were

obtained from J. T. Baker (Phillipsburg,

NJ). Porcine serum was obtained from

Equitech-Bio (Ingram, TX). Oasis™ HLB

extraction cartridges, 1 cc/30 mg

were obtained from Waters Corporation

(Milford, MA). Bond Elut® C18,

1 cc/100 mg solid-phase extraction

(SPE) cartridges were purchased from

Varian Sample Preparation Products

(Harbor City, CA).

Solid-Phase Extraction of Drugs in SerumSample pretreatment for the tetracy-

clines, salicylic acid and benzoic acid

involved adding 20 µL of phosphoric

acid to 1 mL of spiked porcine serum

before sample loading. Figure 1 outlines

the general solid-phase extraction proce-

dure developed on the Oasis™ HLB

extraction cartridges (3). The cartridges

were placed on a 20–position Waters

Extraction Manifold attached to a

Waters vacuum pump. The cartridges

were conditioned with 1 mL methanol

and equilibrated with 1 mL water; the

specific flow rate was not important for

these steps. Then a 1 mL serum sample

containing the drug(s) was added at a

flow rate of < 2 mL/min (vacuum level

~ 3-4” Hg). The cartridges were then

washed with 5% methanol in water to

remove polar interferences. The analyte

was eluted at a flow rate of < 2 mL/min

with methanol.

Figure 1: General SPE method developedfor analysis of drugs in biofluids usingOasis™ HLB extraction cartridges.

Condition1 mL methanol

Load1 mL spiked serum

Equilibrate1 mL water

Evaporate and Reconstitute40 ˚C /200 µL mobile phase

Wash1 mL 5% methanol in water

Elute1 mL methanol

Page 3: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

3© 1997 Waters Corporation

Finally, the methanol eluate was

evaporated to dryness at 40 ˚C under

a nitrogen stream, and the sample was

reconstituted in mobile phase at the

desired concentration. Flow through the

Oasis™ HLB extraction cartridges did not

have to be monitored because the car-

tridges can run dry without loss of

recovery or reproducibility.

In each study, a pair of porcine serum

samples were prepared, each containing

a different drug level, and then each

sample was extracted in replicates of

six, using an appropriate internal stan-

dard as indicated. For the tetracyclines

study, the first serum sample contained

tetracycline and minocycline each at

2.5 µg/mL; in the second sample, the

concentrations of each were 0.5 µg/mL.

Demeclocycline was added to both

serum samples as an internal standard at

5.0 µg/mL. Similarly, for the acids

experiments, salicyclic and benzoic

acids were added to a pair of porcine

serum samples at respective drug levels

of 1 µg/mL and 5 µg/mL. In this case,

the internal standard, nalidixic acid at an

effective level of 5 µg/mL, was added

to the analyte-containing fraction from

each extracted sample after its evapora-

tion and reconstitution in mobile phase.

The same protocol above was repeated

using the Bond Elut® C18 cartridges, with

only one critical difference: with these

silica-based C18 cartridges, the entire

SPE protocol was monitored closely to

prevent cartridge drying, especially

during the conditioning, equilibration,

and sample loading steps.

Batch Reproducibility StudyPhosphate-buffered saline was prepared

containing the following amounts of

anhydrous salts per liter of water:

a) 200 mg potassium chloride,

b) 8000 mg sodium chloride,

c) 200 mg potassium monobasic

phosphate and

d) 1150 mg sodium dibasic phosphate.

The solution was adjusted to pH 7.0

with 10% phosphoric acid.

Phosphate-buffered saline was spiked

with two drug mixtures. One drug

mixture contained procainamide, raniti-

dine, and acetaminophen; the other mix

had propranolol, doxepin, dipropyl

phthalate and betamethasone valerate.

Solid-phase extraction was done with six

replicates for each drug mixture on

each of three batches of Oasis™ HLB

sorbent. The SPE procedure in Figure 1

was used. The internal standard (10 µL

of either sulfanilamide or butyl paraben

at 1 mg/mL, respectively) was added

after reconstitution of the appropriate

eluate fraction.

HPLC Analysis of DrugsThe analyses used various Waters HPLC

instruments equipped with Waters

717plus Autosamplers and the

Millennium® 2010 Chromatography

Manager (Waters Corporation, Milford,

MA), either a Waters 625 LC System or

a Waters 616 Multi Solvent Delivery

System, and either a Waters 486

Tunable Absorbance Detector or a

Waters 996 Photodiode Array Detector.

A Spark Holland Mistral Column Oven

(Euramark, Mount Prospect, IL) was used

to control column temperature.

For analysis of tetracycline and minocy-

cline, a Waters SymmetryShield™ RP8

column (3.0 x 150 mm) coupled with a

Sentry™ SymmetryShield™ RP8 guard

column was used at room temperature.

The mobile phase consisted of 0.1%

TFA in water/methanol/acetonitrile

91/2/7 (v/v/v). Flow rate was

0.9 mL/min; detection was at 350 nm.

The injection amount was 20 µL of

reconstituted porcine serum extract.

Salicylic and benzoic acid were ana-

lysed on a Waters SymmetryShield™ RP8

column (3.0 x 150 mm) coupled with a

Sentry™ SymmetryShield™ RP8 guard

column; the column temperature was

40 ˚C. The mobile phase was 1.25%

glacial acetic acid/acetonitrile

77:23 (v/v). Flow rate was

0.6 mL/min; detection was at 236 nm.

The injection volume was 10 µL.

Procainamide, ranitidine and aceta-

minophen were analysed on the

Waters SymmetryShield™ RP8 column

(3.9 x 150 mm) coupled with a Sentry™

SymmetryShield™ RP8 guard column. The

temperature was controlled to 30 ˚C.

The mobile phase was 20 mM potas-

sium phosphate, pH 3/methanol

97:3 (v/v). Flow rate was 1.0 mL/min;

detection was at 220 nm. The injection

volume was 50 µL.

Doxepin, propranolol and betametha-

sone valerate were analysed on a

Waters Symmetry® C18 column

(3.9 x 150 mm) with a Sentry™

Symmetry® C18 guard column. The

mobile phase was 20 mM potassium

phosphate buffer, pH 7/methanol

35:65 (v/v). Flow rate was 1.0

mL/min; detection was at 254 nm.

The injection volume was 20 µL.

ResultsThe simple, rugged, and highly repro-

ducible solid-phase extraction (SPE)

method that was developed for the

Oasis™ HLB extraction cartridges is out-

lined in Figure 1 (3). Due to the simplicity

of the interaction of analytes with the

Oasis™ HLB sorbent, it was possible to

develop a general SPE method for

extraction of many kinds of drugs and

their metabolites from serum and plasma.

This SPE method is applicable to a wide

range of analytes; it can be modified or

fine tuned to remove specific interfer-

ences or to extract compounds at the

extremes of polarity (very hydrophilic or

hydrophobic drugs).

TetracyclinesOne of the causes for low and variable

recovery on the traditional silica-based

reversed-phase cartridges is the strong

retention of basic compounds by ionic

interaction with deprotonated silanols. In

addition, metals present in the silica

packings can also retain drugs, such as

tetracyclines, which are metal chelators.

Higher and more consistent recoveries of

such compounds would be expected

from the Oasis™ HLB extraction car-

tridges which do not contain either

silanols or metals. Indeed, when the

amphoteric tetracyclines were extracted

from serum using Oasis™ HLB extraction

cartridges and the simple general SPE

method, recovery and reproducibility

were found to be significantly better than

the results obtained when C18-silica SPE

cartridges were used (see Table 1) (5).

The structures of the tetracyclines used in

this study are shown in Figure 2. Under

typical SPE conditions, tetracyclines are

amphoteric and can chelate with metals.

The results in Table 2 show that optimum

extraction of the tetracyclines is achieved

if the spiked serum is pretreated with

acid before loading on the cartridges.

Page 4: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

4© 1997 Waters Corporation

The tetracyclines are known to bind to

serum proteins; the literature reports up

to 75% retention of tetracyclines by

serum proteins (6, 7). The three acids,

2% phosphoric acid, 1% TFA and

63 mM citric acid, were equally effec-

tive and resulted in essentially 100%

recovery of minocycline and tetracycline.

All three acids were able to disrupt the

drug-protein interaction; phosphoric acid

was selected for reasons of safety and

convenience.

The chromatograms in Figure 3 of the

blank serum (A) and the extracted drug-

spiked sample (B) show that no

endogenous peaks interfere with the

HPLC analysis of the drugs or the internal

standard. Choosing an absorption

maximum wavelength of 270 or

350 nm allows detection of the tetracy-

clines where few endogenous

compounds absorb. The excellent peak

shapes on the SymmetryShield™ RP8

column also facilitated quantitation with

high accuracy and precision. Extractions

using the Oasis™ HLB extraction car-

tridges with our general method gave

less than 1.5% RSDs and greater than

95% recoveries for two levels of minocy-

cline and tetracycline; the data for

both the 2.5 and 0.5 µg/mL levels

are in Table 1.

With the method outlined in Figure 1,

results obtained using Oasis™ HLB

extraction cartridges were compared to

those using the silica-based Bond Elut®

C18 cartridges. Table 1 lists 40.7% and

67.4% recoveries from the Bond Elut®

C18 cartridges versus 94.8% and 104%

recoveries from the Oasis™ HLB extrac-

tion cartridges for minocycline and

tetracycline, respectively. The reason(s)

for the inferior recoveries obtained by

SPE with the C18 cartridges are not

clearly understood. Interactions of tetra-

cyclines with the surface of silica-based

reversed-phase sorbents are typically

manifested in chromatograms by tailing

and distorted peaks. In the literature, this

behavior has been attributed to silanol

interaction and/or complexing with

metal contaminants. These complications

clearly do not occur with the unique

polymer sorbent in Oasis™ HLB

extraction cartridges.

Table 1: Comparison of SPE results for tetracyclines.

Minocylcline 98.7% 102% 105%

Tetracycline 97.2% 106% 101%

2% 1% 63 mMPhosphoric Acid Trifluoracetic Acid Citric Acid

Table 2: Recovery of tetracyclines as a function of the type of acid used for sample pretreatment. The SPE method using Oasis™ HLB extraction cartridges is described in the text.

Oasis™ HLB Extraction Cartridge C18 Cartridge

Concentration Recovery RSD Recovery RSD (µg/mL) (%) (%) (%) (%)

Minocycline 2.5 94.8 1.4 40.7 0.820.5 101 0.54

Tetracycline 2.5 104 0.55 76.4 0.440.5 97.8 1.4

COOH

COOH

OH

CH2CH3

H3C

COOH

N

O

N

Benzoic acidMinocycline

Tetracycline Salicylic acid

Nalidixic acid (I.S.)

H HOH

OHOH OHO O

NMe2 NMe2

NMe2

NMe2

CONH2

OH

CONH2

OH

CONH2

HHO

OH O OHOH

O

CH3

Demeclocycline (I.S.)

H HOH

OH O OHOH

O

CI

Figure 2: Structures of drugs and internal standards.

Page 5: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

5© 1997 Waters Corporation

Polar Acidic DrugsThe Oasis™ HLB extraction cartridges

retain drugs and their metabolites by the

same reversed-phase mechanism as do

the traditional silica-based reversed-

phase cartridges. Acidic and neutral

drugs do not interact secondarily with the

silanols on the silica-based sorbents.

Therefore, these compounds should

interact with the Bond Elut® C18 and the

Oasis™ HLB sorbents in a similar manner.

Both cartridge types should show the

same high recoveries. The only proviso is

the polarity of the drug since the more

polar drugs may be poorly retained on

the C18 cartridges.

Salicylic and benzoic acids are polar

compounds with pKa’s close to 3.0; the

structures are shown in Figure 2. They

were extracted by Oasis™ HLB extraction

cartridges with the same SPE method

employed for the amphoteric tetracy-

clines. Table 3 illustrates the importance

of extracting the polar acids in proto-

nated form. When salicylic acid was

extracted from spiked saline at pH 7.0

only 62.5% recovery was obtained. In

contrast, extraction of these acidic com-

pounds in the protonated form at pH <

2 gave 100% recovery. A protonated

acid is less polar than its corresponding

ionic form and is therefore better retained

on a reversed-phase SPE sorbent.

Excellent peak shapes were obtained for

both salicylic and benzoic acid on the

SymmetryShield™ RP8 column which

enabled accurate quantitation of these

compounds in serum or saline after solid-

phase extraction. The chromatography is

shown in Figure 4 for both the extracted

acids and the serum blank. Table 4

shows the recoveries obtained when

salicylic acid and benzoic acid at both

5 µg and 1 µg per mL levels were

extracted from serum using Oasis™ HLB

extraction cartridges. Recoveries greater

than 85% percent and RSDs less than

5% were obtained.

Figure 3: Chromatograms of (A) serum and (B) serum spiked with minocycline andtetracycline using demeclocycline as internal standard. Chromatographic conditions are described in the text.

Table 3: Effect of sample pH during loading on recovery of salicyclic acid.

Salicylic Acid 10 62.5 101pka 2.97, 13.4

Compounds Concentration Load at pH 7 Load at pH < 2 µg/mL Recovery(%) Recovery (%)

1

2

3

0.035

00 5 10 15

Minutes

A

B

AU

Figure 4: Chromatograms of (A) serum and (B) serum spiked with salicylic acid and benzoic acid using nalidixic acid as internal standard. Chromatographic conditions are described in the text.

Column: SymmetryShield™ RP8, 5 µm, 3 mm x 150 mmMobile phase: Acetonitrile: 1% glacial acetic acid (75:25)Flow rate: 0.6 mL/minDetection: UV at 236 nmInjection Volume:10 µL

0.02

00 5

1

23

10 15 20 25 30A

U

B

A

1: Minocycline2: Tetracycline3: Demeclocycline (I.S.)

Peaks:1: Benzoic acid2: Nalidixic acid (I.S.)3: Salicylic acid

Page 6: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

6© 1997 Waters Corporation

Specifications and sorbent batch test

results are reported on a Certificate of

Analysis which is included in each box

of product.

As a result of this stringent quality

control, excellent batch-to-batch repro-

ducibility has been observed for the

Oasis™ HLB sorbent. Figure 5 shows the

high recovery obtained for seven com-

pounds, ranging from the very polar

neutral acetaminophen to the more

lipophilic dipropyl phthalate and

betamethasone valerate and including

basic drugs such as procainamide, pro-

pranolol, doxepin and ranitidine.

0.96% 1.21% 2.02% 3.66% 2.40% 3.06% 3.67%

Batch A Batch B Batch C

Proc

aina

mid

e

% R

ecov

ery

Rani

tidin

e

Ace

tam

inop

hen

Prop

rano

lol

DP-

phth

alat

e

Dox

epin

Beta

met

haso

neVa

lera

te

100

80

60

40

20

0

Figure 5: Batch-to-Batch Reproducibility of Oasis™ HLB Sorbent in 1 cc/30 mg Cartridges

Regardless of the nature of the drug,

greater than 90% recoveries were

observed. More important is the fact that

the recoveries for all drugs were highly

consistent from batch-to-batch. RSDs

≤ 3.67% were obtained across three

different batches of Oasis™ HLB sorbent.

These data along with the

low % RSDs for replicates of six car-

tridges per analysis (Tables 2 and 4)

illustrate the high reproducibility from

batch-to-batch and cartridge-to-car-

tridge, respectively, for the Oasis™ HLB

extraction cartridges.

Table 4: Recoveries of salicylic and benzoic acids extracted from serum using Oasis™ HLBextraction cartridges. The SPE method is described in the text.

Salicylic acid 5.0 91 3.31.0 92 5.0

Benzoic acid 5.0 87 3.01.0 85 3.8

Compound Concentration (%) Recovery (%)RSD(µg/mL)) (n=6)

When the acids were extracted from

acidified serum, using Bond Elut® C18

cartridges and the procedure detailed

in Figure 1, being careful not to let the

cartridges run dry, the results were

similar to those obtained with the Oasis™

HLB extraction cartridges. Greater than

90% recoveries were obtained for sali-

cylic acid at 1 µg/mL on both types

of cartridges.

Because acidic compounds interact

with the SPE sorbents used here princi-

pally by a reversed-phase (hydrophobic

interaction) mechanism, no differences

were expected or observed between

the results obtained using traditional

bonded-silica C18 cartridges and those

using Oasis™ HLB extraction cartridges.

Many methods that have been devel-

oped on silica-based reversed-phase

cartridges can be transferred directly

to Oasis™ HLB extraction cartridges.

The key advantage in doing this is that

flow through Oasis™ HLB extraction

cartridges does not have to be moni-

tored. Cartridge drying as described

above will not lead to low and/or

variable recoveries.

ReproducibilityA high level of reproducibility is built

into the synthetic design and the manu-

facturing process for the Oasis™ HLB

sorbent. Each sorbent batch is quality

controlled by eleven different tests.

Physical property specifications include

pore size, particle size, surface area,

total pore volume and fines content

(% of particle volume with particles

less than 10 micron in size).

Recoveries and relative standard devia-

tion (n=6) are reported for the

solid-phase extraction of three polar

drugs—procainamide, ranitidine and

acetaminophen. A sorbent selectivity test

ensures that each batch has identical

retention characteristics. This selectivity

test involves packing the sorbent in an

HPLC column to measure relative reten-

tions and a capacity factor for a mixture

of acidic (salicylic acid), basic (raniti-

dine) and neutral (caffeine, p-toluamide)

compounds.

Page 7: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

7© 1997 Waters Corporation

ConclusionsOasis™ HLB extraction cartridges are a

universal replacement for, and provide a

solution to the problems associated with,

traditional silica-based reversed-phase

SPE cartridges. The absence of silanol

interaction and the excellent retention of

polar drugs on the Oasis™ HLB extrac-

tion cartridges make it possible to

develop ideal sample preparation

methods without difficulty. Compounds

as different as tetracyclines, basic drugs,

and polar acids were all extracted with

high recoveries using the same simple,

rugged and highly reproducible method.

High and consistent recoveries were

obtained across cartridges and batches

of Oasis™ HLB sorbent regardless of an

analyte’s polarity or chemical function-

ality. These new reversed-phase SPE

cartridges make sample preparation fast

and easy for the analytical chemists who

are quantitating drugs and metabolites in

biological fluids. Oasis™ HLB extraction

cartridges should replace C18 cartridges

in existing methods and should be the

cartridge of choice for developing all

new SPE protocols.

References1. A. A. MacDonald, D. G. Nickell and S. H. DeWitt,

Pharmaceutical News, 3(6) 1996 pp 19–21.

2. GlaxoWellcome sponsors, Nature, 384 (6604

supplement) 1996 pp 1–6.

3. E.S.P. Bouvier, D.M. Martin, P.C. Iraneta,

M. Capparella, D. Phillips, Y-F. Cheng and L. Bean,

Waters Column, VI(3) 1996 pp 1–5.

4. E.S.P. Bouvier, D.M. Martin, P.C. Iraneta,

M. Capparella, Y-F. Cheng and D.J. Phillips, LC:GC,

15(2) 1997 pp 152–158.

5. Y-F. Cheng, D.J. Phillips, and U.D. Neue,

Chromatographia, 44(3/4) 1997 pp 187-190.

6. H. Kalant and W. H. E. Roschlau, Principles of Medical

Pharmacology, B. C. Decker, 1989, p 32.

7. R. C. Heel and G. S. Avery, Drug Data Information

in Drug Treatment, Principles and Practice of Clinical

Pharmacology and Therapeutics, Adis, 1980,

pp 1211–1223.

Ordering Information

Oasis™ HLB Extraction Cartridge 1 cc/30 mg 100/box WAT094225

Oasis™ HLB Extraction Cartridge 3 cc/60 mg 100/box WAT094226

For your free copy of the Oasis™ HLB Extraction CartridgesApplications Notebook, please checkbox 3 on the business reply card.

The Waters Oasis™ HLB Extraction Plate (96-well) is available with 30 mg of sorbent in each well. Each well contains a unique (patent pending) sorbent, a copolymerdesigned to have a Hydrophilic-Lipophilic Balance (HLB), that gives high and reproducible recoveries for acidic, basic, and neutral compounds—even if the sorbent in the wells runs dry.

Oasis™ HLB Extraction Plate (30 mg/96 well) 1/pkg WAT058951

Oasis™ HLB Extraction Plate (30 mg/96 well) 3/pkg WAT058939

Extraction Plate Manifold WAT058941

Description Part No.

To achieve the maximun advantage of the Oasis™ HLB extraction cartridge, we recommend the new Waters Extraction manifold (see page 18).

For more information on Oasis™ HLB Extraction Cartridges andExtraction Plate, please check box 4 on the business reply card.

NEW Oasis™ HLB Extraction Plate

Description Qty Part No.

Page 8: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

8© 1997 Waters Corporation

IntroductionThe process of developing HPLC

methods is a critical step in the work

of an analytical chemist. Frequently, the

investigator encounters a situation where

the selectivity obtainable on a particular

column fails to meet the separation

requirements, even after systematic

optimisation of the mobile phase (1,2).

When this happens the chromatogra-

pher often resorts to use of unusual

mobile phase compositions or additional

mobile phase additives. In this paper,

we demonstrate an alternative

approach: when a separation does

not respond to simple variations in

mobile phase composition, it is often

advantageous to change to a column

that has significantly different selectivity

characteristics.

The SymmetryShield™ RP8Packing MaterialWaters has recently developed a unique

patented reversed-phase packing mate-

rial (SymmetryShield ™ RP8) possessing

selectivity characteristics significantly

different from traditional C8 and C18

materials (3). The modified selectivity

arises from the insertion of a carbamate

group into the alkylsilane bonded phase

(Figure 1). The incorporation of polar

groups into alkylsilane bonded phases

has been shown to significantly modify

their selectivity characteristics, due to

perturbation of both analyte/stationary

phase and mobile phase/stationary

phase interactions (3-7). These polar

functional groups compete with polar

analytes for the interaction with the

surface silanols. Therefore, one

particularly notable feature of all

reversed-phases containing embedded

polar group is the reduction in both

retention and peak tailing of amines.

Improving Method Development Using a New Reversed-Phase High-Performance Liquid Chromatography (HPLC) Column with Unique SelectivityUwe D. Neue, Dorothy J. Phillips, Michael S. Young, Thomas H. Walter, John E. O’Gara, Mark Capparella and Bonnie A. Alden

This reduced retention can be attributed

to a shielding of the silanols by the

embedded polar group.

DiscussionIn the process of methods development,

the investigator often encounters a situa-

tion where the selectivity obtainable on a

particular packing does not entirely

match the requirements of the applica-

tion. In such a situation, the

chromatographer usually struggles for

extended periods of time with unusual

mobile phase compositions or additional

mobile phase additives. An alternative

approach that has proven very valuable

in our laboratory is the use of a column

chemistry that provides similar retention

characteristics but different selectivity

compared to a classical reversed-phase

column. The SymmetryShield™ RP8

column provides this alternative. The

addition of the carbamate group in

the bonded phase significantly changes

the interaction between polar functional

groups in the analyte molecule and the

silanol groups present on all bonded-

phase packings. As a consequence,

switching from a standard C8 or C18

column to a SymmetryShield™ RP8 column

can drastically affect the peak spacing

and the elution order of analytes without

the need to change mobile phase com-

position. This allows for a rapid

comparison between a standard C8

column and the SymmetryShield™ RP8

column.

The orientation of the carbamate group

should be emphasised. It is oriented such

that after a hydrolytic loss of the ligand at

the carbamate function, an interactively

neutral alcohol function is formed.

Although careful hydrolysis studies have

not revealed any stability issues, we pur-

posely designed the phase to avoid any

inadvertent secondary interactions. This is

not the case for other packing materials

available on the market, which also rely

on polar functional groups to prevent the

interaction between analytes and surface

silanols (7).

A selectivity study of SymmetryShield™ RP8

was made using simple substituted ben-

zenes as samples. It was found that,

generally, the methylene group selectivity

was lower on the SymmetryShield™ RP8

column than on a standard Symmetry® C8

column. At the same time, the selectivity

for polar functional groups was signifi-

cantly different for both phases. There was

a reduced retention for very polar func-

tional groups on the SymmetryShield™ RP8

column, especially for amino-, alcohol-

and amide- functions, and an increased

retention for phenols and acids at

acidic pH. This indicates that the

influence of acidic silanols on the

retention of analytes is decreased using

the SymmetryShield™ RP8 column. An

example of the selectivity differences is

shown in Figure 2.

Si

CH3

(CH2) CH3H3C

O O

O

NH

C 7

Figure 1: Structure of the SymmetryShield™ RP8 bonded phase

Page 9: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

9© 1997 Waters Corporation

On the SymmetryShield™ RP8 column, we

find a significant decrease in retention for

benzyl alcohol and benzylamine, and a

slight increase in retention for phenol and

benzoic acid. It is believed that the

reason for these selectivity differences is

due to the fact that water strongly binds to

the polar carbamate group. This tightly

bound layer of water prevents interactions

between the analytes and the underlying

tether function as well as with the surface

silanols of the underlying silica. As a con-

sequence, surface silanols do not play as

much of a role in the retention of analytes

on the SymmetryShield™ RP8 column as

they do on a standard C8 column. Also,

good peak shapes are obtained for polar

analytes, in addition to the differences in

selectivity discussed above. Thus, the

name of the column is quite appropriate:

the surface silanols are “shielded” from

interacting with the polar functional

groups of analytes. Such differences in

interaction can result in significant shifts

in the selectivity of a separation.

As mentioned above, the

SymmetryShield™ RP8 column often yields

a separation when mobile phase modifi-

cation on a C8 packing has failed to do

so. An example is shown in Figure 3. On

the left, the separation of furazolidone

impurities is shown on a Symmetry® C8

column. It appears that there is only a

single impurity present. However, as

shown in the chromatogram on the right,

the SymmetryShield™ RP8 column cleanly

1

1 2

50 10 15

3

4

2

3

4

Minutes

V0

V0

Symmetry Shield™ RP8

Symmetry® C8

Figure 2: Selectivity differences between SymmetryShield™ RP8 and Symmetry® C8 packings

0.01

0.008

0.006

0.004

0.002

0

0 5 10 15

Minutes

20 25 30 0 5 10 15

Minutes

20 25 30

0.01

0.008

0.006

0.004

0.002

0

Figure 3: Selectivity comparison for furazolidone impurities

Peaks:1. Benzyl alcohol2. Phenol3. Benzoic acid4. Benzylamine

Columns: 3.9 mm x 150 mm SymmetryShield™ RP8, 5 µm and Symmetry® C8 5 µm

Mobile phase: 80% 50 mM KH2PO4, pH 3.0, 20% acetonitrileFlow rate: 1 mL/minDetection: Absorbance at 254 nm

Column: a. Symmetry® C8 5 µm b. SymmetryShield™ RP8, 5 µm 3.9 mm x 150 mm

Mobile Phase: Gradient at 1%/min acetonitrile starting at 10% acetonitrile/ 90% 0.1% acetic acid, pH 3.3

Flow Rate: 0.7 mL/minDetection: Absorbance at 370 nmInjection: 10 µLSample: Furazolidone and its impurities

separates the impurity into two well

resolved peaks. In this case, the

selectivity difference between a

SymmetryShield™ RP8 column and a

Symmetry® C8 column is quite striking.

Another example of selectivity

differences is shown in Figure 4

for the Alberta peptide standards. A

significant reduction in retention is

observed for these peptides on the

SymmetryShield™ RP8 column, as com-

pared to the standard Symmetry® C8

column.

Symmetry® C8SymmetryShield™ RP8

Page 10: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

10© 1997 Waters Corporation

This reduction in retention is due to a

decrease in the interactions of these basic

peptides with the SymmetryShield™ RP8

column compared to the Symmetry® C8

column. This reduced interaction is com-

pletely in line with the expectations

outlined above. It appears that the

SymmetryShield™ RP8 column provides

additional separation capabilities for

peptide samples where the options for

mobile phase manipulation are more

limited.

A further example in Figure 5 shows the

separation of phenoxyacid herbicides

and phenols on a SymmetryShield™ RP8

column. The separation of all 16 EPA

555 analytes is accomplished without

difficulties using this column. The same

separation is not possible using standard

reversed-phase columns. Once again, the

advantage of the different selectivity

achievable with the SymmetryShield™ RP8

column is well demonstrated for this

application.

For separations of compounds with differ-

ences in the pKa, significant variations in

elution order have been observed

between a standard C8 column and the

SymmetryShield™ RP8 column. An

example is the separation of the weak

acid, acetaminophen; the strong base,

codeine and caffeine at pH 3 (Figure 6).

On the Symmetry® C8 column, codeine

elutes after acetaminophen. Using the

identical mobile phase conditions on a

SymmetryShield™ RP8 column, significantly

different retention times were observed.

Most importantly, the elution order of

acetaminophen and codeine changed.

The basic compound codeine exhibits

less retention than acetaminophen on the

SymmetryShield™ RP8 column, while it

elutes after acetaminophen on the

Symmetry® C8 column. This is completely

in line with the expectations outlined

above.

In addition to the selectivity differences

demonstrated, SymmetryShield™ RP8

columns also perform better than standard

reversed-phase columns in highly

aqueous mobile phases.

5

1 2

3

45

.10

.05

0

10

Minutes

AU

FS15

SymmetryShield™ RP8

Figure 4: Difference in selectivity for polar basic peptides

Figure 5: Separation of phenoxyacid herbicides and phenols on a SymmetryShield™ RP8 column

5 10

12

3

4

5

Minutes

15

.10

.05

0

AU

FS

Symmetry® C8

0 20 40

Minutes

Column: SymmetryShield™ RP8, 5µm 3.9 mm x 150 mmMobile phase: A: Phosphate Buffer pH 3.1

B: AcetonitrileGradient: 90% A to 70% A in 8 min, hold for 7 min, then linear to 40% A

at 30 min, then to 10% A at 35 minFlow rate: 1.0 mL/minDetection: Absorbance at 230 nm (0.03 AUFS)Sample: 500 mL of tap water spiked with 5 µg/L each component,

trace enriched by solid-phase extraction (SPE)Injection: 40 µL of SPE eluent diluted 1:3 with buffer

Columns: SymmetryShield™ RP8 5µm and Symmetry® C8, 5µm 3.9 mm x 150 mm

Mobile phase:A. 0.1% TFA in waterB. 0.1% TFA in acetonitrile

Gradient: From 10% to 40% B in 30 min, step to 60% B for 5 min

Flow rate: 0.8 mL/minDetection: Absorbance at 214 nmInjection: 9 µL Alberta Peptide Mixture

Sample: Ac-Arg-Gly-X-X-Gly-Gly-Leu-Gly-LysAmide-X-X-:1: Ala-Gly with free alpha amino group2: Gly-Gly 4: Val-Gly3: Ala-Gly 5: Val-Val

Page 11: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

11© 1997 Waters Corporation

Figure 6: Gradient separation of acetaminophen, codeine and caffeine

0 2 4

2 13

21 3

Minutes6 8

0 2 4Minutes

6 8

Columns: a. SymmetryShield™ RP8 5 µm 3.9 mm x 150 mmb. Symmetry® C18 5 µm, 3.9 mm x 150 mm

Mobile phases: A. 50 mM potassium phosphate, pH 3B. Acetonitrile

Gradient: 5% to 50% acetonitrile in 15 minFlow rate: 1 mL/minDetection: Absorbance at 254 nmInjection volume: 5 µL

ConclusionThe SymmetryShield™ RP8 column repre-

sents a powerful new tool for

reversed-phase separations. The selec-

tivity of this column for compounds with

polar functional groups can be signifi-

cantly different from the selectivity of

standard reversed-phase columns. This

property can be used advantageously

in methods development. This is espe-

cially true in circumstances where a

good separation with standard reversed-

phase columns appears to be difficult to

achieve. The difference in the selectivity

characteristics observed between the

SymmetryShield™ RP8 column and stan-

dard reversed-phase column is primarily

due to the shielding of the surface

silanols by the carbamate group incor-

porated in column. Another advantage

of this column is its good wettability in

highly aqueous mobile phases. The

SymmetryShield™ RP8 column signifi-

cantly enhances the tools available to

the methods development chemist and

opens a new dimension in reversed-

phase selectivity.

This improvement arises because the

embedded polar functional group in the

bonded phase ensures that the packing

remains well wetted even in mobile

phases containing 100% water. In con-

trast, in highly aqueous (> 90% water)

mobile phases, high coverage C8 and

C18 columns often show erratic retention

(8). This has been attributed to poor

wettability of the bonded alkylsilane

groups, preventing the partitioning of

analyte into the stationary phase (9).

Since the SymmetryShield™ RP8 packing

does not exhibit this problem, it can be

used with advantage in the separation

of very polar analytes, which require

mobile phases with a high water

content. These advantageous wetting

properties can be readily demonstrated

using very polar analytes, such as the

antibiotic amoxicillin. The mobile phase

consisted of 99.9% water and 0.1%

glacial acetic acid. Good retention was

achieved when the separation of this

analyte was run on a Symmetry® C8

column. However, when the flow was

stopped for a short time (5 min), then

resumed, the retention of this compound

was lost completely. It was only possible

to regain the original retention of the

compound by reconditioning the sta-

tionary phase first with methanol, then

with the aqueous mobile phase. When

the same experiment is performed using

a SymmetryShield™ RP8 column, no loss

of retention is observed. Fundamentally,

this shows that the surface remains

wetted even when a mobile phase con-

taining 100% water is used. This wetting

phenomenon is due to the polar func-

tionality incorporated into the stationary

phase. As mentioned already in the dis-

cussion of the retention mechanism, the

polar carbamate group of the

SymmetryShield™ RP8 column binds

water very strongly, thereby preventing

the “hydrophobic collapse” observed

with standard reversed-phase columns

in mobile phases with a very high

water content. Consequently, the

SymmetryShield™ RP8 column can be

used advantageously for the analysis

of very polar compounds that require

mobile phase compositions approaching

100% water.

References1. M.Z. El Fallah, Waters Column, Vol. 6 (1) 1995, 1

2. M.Z. El Fallah, in U.D. Neue HPLC Columns –

Technology, Theory and Practical Use, Wiley-Interscience,

New York, 1997

3. J.E. O’Gara, B.A. Alden, T.H. Walter, J. S. Petersen,

C.L. Niederlander and U.D. Neue, Anal. Chem. 67,

1995, 3809

4. B. Buszewski, J. Schmid, K. Albert and E. Bayer,

J. Chromatogr. 552, 1991, 415

5. B. Buszewski, M. Jaroniec and R.K. Gilpin,

J. Chromatogr. A673, 1994, 11

6. B. Buszewski, M. Jaroniec and R.K. Gilpin,

J. Chromatogr. A668, 1994, 293

7. T.L. Ascah, K.M.R. Kallury, C.A. Szafranski,

S.C. Corman and F. Liu, J. Liq. Chrom. & Rel. Technol.

19 (17 & 18), 1996, 3049

8. R.P.W. Scott and C.F. Simpson, J. Chromatogr. 197,

1980, 11

9. Z. Li, S.C. Rutan and S. Dong, Anal. Chem. 68, 1996,

124

SymmetryShield™ RP8

Symmetry® C18

Peaks:2. Codeine1. Acetaminophen3. Caffeine

Peaks:1. Acetaminophen2. Codeine3. Caffeine

Page 12: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

12© 1997 Waters Corporation

Ordering Information

* Requires reusable endfittings** Guard columns require the appropriate Sentry Guard Holder.

(3.9 x 50 mm cartridge columns must use the Universal Guard Holder)

Dimension 2.1 x 150 3.0 x 150 3.9 x 150 4.6 x 150 4.6 x 250

Part No. WAT094245 WAT094243 WAT200655 WAT200662 WAT200670 WAT200675

Sentry™ Guard SymmetryShield™ RP8 5 µm Columns Column**

Dimension 3.9 x 50 3.9 x 150 4.6 x 150 4.6 x 250

Part No. WAT094248 WAT200658 WAT200665 WAT200661

Dimension 2.1 x 50 4.6 x 50 4.6 x 75 4.6 x 100 4.6 x 150

Part No. WAT094257 WAT094260 WAT094263 WAT094266 WAT094269

Dimension 4.6 x 75 4.6 x 100

Part No. WAT094272 WAT094275

Dimension 2.1 x 150 3.0 x 150 3.9 x 150 4.6 x 150 4.6 x 250

Part No. WAT094254 WAT094251 WAT210594 WAT210588 WAT210591

Dimension 3.9 x 150 4.6 x 150 4.6 x 250

Part No. WAT210582 WAT210585 WAT210579

Dimension 4.6 x 150

Part No. WAT094278

Endfittings WAT037525

Integrated Guard Holder

(for Waters steel cartridge columns only) WAT046905

Universal Guard Holder

(for any HPLC column) WAT046910

SymmetryShield™ RP8 5 µm Cartridge Columns*

SymmetryShield™ RP8 3.5 µm Columns

SymmetryShield™ RP8 3.5 µm Cartridge Columns*

SymmetryShield™ RP8 5 µm Method Validation Kits (three columns from 3 different batch materials)

SymmetryShield™ RP8 5 µm Method Validation Kits (three cartridge columns* from 3 different batch materials)

SymmetryShield™ RP8 3.5 µm Method Validation Kits (three columns from 3 different batch materials)

Description Part No.

For complete Symmetry®

ordering information,please check box 5 onthe business reply card.

For your free copy of the Symmetry® Applications Notebook,please check box 6 on the business reply card.

Page 13: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

13© 1997 Waters Corporation

Evaluation of Imitation µBondapak™ C18 Columns

Waters µBondapak™ C18 column

was introduced in 1973 as the first

10 micron particle size bonded

reversed-phase column. Twenty-four

years later, it remains one of the most

widely used HPLC columns, with appli-

cations in many different scientific fields.

The selectivity of the µBondapak™ C18

column is unique, the result of a carefully

controlled combination of hydrophobicity

and silanol activity. µBondapak™ C18

columns are manufactured only at

Waters’ ISO 9002-certified

Material Synthesis Facility in Taunton,

Massachusetts, using proprietary

procedures.

The popularity of Waters

µBondapak™ C18 columns has inspired

many attempts at imitation. To determine

how successful these attempts have

been, we recently characterised three

imitation columns using a sensitive chro-

matographic selectivity test. The results

are described in this article.

ExperimentalThe columns evaluated were all

3.9 x 300 mm steel columns, except for

the 4.6 x 216 mm Imitation C column.

The reference column was an authentic

Waters µBondapak™ C18 3.9 x 300 mm

column. The chromatography was

carried out using a Waters 625 LC

System, a 717plus Autosampler, a

490 Programmable Multiwavelength

Detector, and a Millennium® Workstation

(all from Waters Corporation). The

columns were thermostatted at 23.4

(± 0.5) ˚C using a Mistral® column

thermostat.

Results and DiscussionThe columns were first tested for effi-

ciency using acenaphthene as the

marker, and a mobile phase of

60/40, v/v, acetonitrile/water flowing

at 2.0 mL/min. The results are sum-

marised in Table 1.

The authentic Waters µBondapak™ C18

column was found to be the most effi-

cient of the four columns, and also gave

the best peak symmetry. The imitation

columns were found to have efficiencies

that were 25–30% lower, and all

showed significant peak fronting (USP

Tailing < 1). This is an indication that the

imitation columns are not as well packed

as the authentic Waters µBondapak™

C18 column.

To evaluate the selectivity of the columns

we used a test mixture containing acidic,

basic, and non-ionisable compounds

(see Figure 1). The mobile phase con-

tained methanol (65% v/v) and a pH

720 mM potassium phosphate buffer.

Under these conditions, the strong base,

propranolol (pKa ≈ 9.6) is protonated,

and interacts with ionised silanol groups

through an ion-exchange mechanism.

This causes the propranolol peak to tail

noticeably.

The chromatograms obtained on the

four columns are shown in Figure 2.

The retention factors and several relative

retentions are summarised in Table 2.

The imitation columns clearly do not

match the separation obtained using the

authentic Waters µBondapak™ C18

column. The largest differences are

observed for the strong base, propra-

nolol (marked with arrow). On the

authentic Waters µBondapak™ C18

column, propranolol elutes midway

between toluene and acenaphthene.

On the Imitation A C18 columns,

however, propranolol partially coelutes

with acenaphthene.

Table 1: Efficiencies and Tailing Factors for Authentic Waters µBondapak™ C18

and Imitation C18 Columns

µBondapak™ C18 30,200 1.01

Imitation A C18 20,800 0.92

Imitation B C18 22,100 0.93

Imitation Ct C18 21,900 0.98

Table 2: Retention Factors and Relative Retentions for Authentic Waters µBondapak™ andImitation C18 Columns

µBondapak™ Imitation A Imitation B Imitation C

Retention Factors

Propyl paraben 1.47 1.29 1.17 0.98

Toluene 2.43 2.40 1.90 1.72

Propranolol 4.52 6.35 1.61 13.1

Acenaphthene 7.60 7.69 6.23 5.09

Dibutyl phthalate 12.3 11.4 9.61 7.85

Relative Retention versus Acenaphthene

Propyl paraben 0.193 0.167 0.188 0.193

Toluene 0.320 0.312 0.305 0.339

Propranolol 0.594 0.825 0.258 2.57

Dibutyl phthalate 1.62 1.48 1.54 1.54

Column Efficiency1 USP Tailing

1 Plates per meter, measured using 5 sigma method

Page 14: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

14© 1997 Waters Corporation

For your copy of the WatersColumn Selection Guide for USP Specifications, please checkbox 7 on the business reply card.

This indicates that the silanol activity of the

alphaBond™ column is higher than that of

an authentic Waters µBondapak™ C18

column. On the Imitation C C18 column,

the silanol activity is so much higher that

propranolol elutes last, as a barely dis-

cernable peak. In contrast, on the

Imitation B C18 column propranolol elutes

before toluene, indicating reduced silanol

activity. Clearly, none of the imitation

columns matches the selectivity of an

authentic Waters µBondapak™ C18

column.

ConclusionsThese results show that there are signifi-

cant differences between authentic

Waters µBondapak™ C18 columns and the

imitation columns. The imitation columns

were found to have lower efficiences, and

produced fronting peaks. In addition, we

found large differences in selectivity, par-

ticularly evident in the retention of basic

compounds relative to neutral compounds.

This is not surprising, because controlling

the retention of basic compounds has

been one of the most difficult problems in

the synthesis of C18–silica packings.

Research has shown that base retention is

affected not only by the bonding

processes, but also by many properties of

the silica support. In the final analysis, no

imitation column matches the performance

of an authentic Waters µBondapak™ C18

column.

Figure 1: Structures of Test Compounds

O

OHO

O OO

O

NH

O

OH

O

HOH2C CH2OH

CH3

CH3C3H7

C4H9

C4H9

Dihydroxyacetone (DHA) Acenaphthene

Proplparaben Propranolol Dibutyl phthalate

CH3

Toluene

10 20 30

Minutes

10 20 30

Minutes

10 20 30

Minutes

10 20 30

Minutes

µBondapak C18

Imitation A C18

Imitation B C18

Imitation C C18

1

2

34

5

6

2

34

56

2

654

3

2

3

4

5

61

1

1

Figure 2: Chromatograms for Selectivity Test Mixture Separated on Authentic WatersµBondapak™ C18 and Imitation Columns

Peaks:

1: DHA2: Propyl paraben3: Toluene4: Acenaphthene5: Dibutyl phthalate6: Propranolol

Peaks:

1: DHA2: Propyl paraben3: Toluene4: Propranolol5: Acenaphthene6: Dibutyl phthalate

Peaks:

1: DHA2: Propyl paraben3: Toluene4: Propranolol5: Acenaphthene6: Dibutyl phthalate

Peaks:

1: DHA2: Propyl paraben3: Propranolol4: Toluene5: Acenaphthene6: Dibutyl phthalate

Page 15: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

15© 1997 Waters Corporation

Ordering information

µBondapak C18 10 µm 125Å 2.1 mm x 300 mm WAT086609

3.9 mm x 150 mm WAT086684

3.9 mm x 300 mm WAT027324

4.6 mm x 150 mm WAT044370

Bondapak C18 15-20 µm 125Å 3.9 mm x 300 mm WAT025875

µBondapak C18 cartridge column 10 µm 125Å 4.6 mm x 250 mm WAT052860

(requires endfittings)

Endfittings (includes 2 sets of reusable endfittings, WAT037525

2 c-clips and o-rings)

Particle PoreColumn Size Size Dimensions Part No.

How to Reach UsSales Offices

Austria and European Export(Central Europe, CIS, Middle East, India and India SubcontinentWaters Ges.m.b.H.

Hietzinger Haupstrasse 145

A-1130 Wien

Tel: 1 877 1807

Fax: 1 877 1808

Australia Waters Australia Pty.

Limited

Unit 3, 38-46 South Street

P.O Box 84

Rydalmere NSW 2116

Australia

Tel: 2 9933 1777

Fax: 2 9898 1455

Belgium and LuxemburgWaters S.A.-N.V.

Raketstraat 60

Rue de la Fusee 60

1130 Brussels

Tel: 02 726 1000

Fax: 02 726 1100

Brazil Waters Comercial LTDA

Avenida Morumbi

8411 - 1 Andar (1st floor)

CEP. 04703-004 Sao Paulo

S.P. Brazil

Tel: 011 55 11 543 7788

Fax: 011 55 11 530 6413

Canada Waters Limited

3687 Nashua Drive, Unit 8

Mississauga, Ontario

L4V 1V5

Tel: 800 252 4752

Fax: 905 678 9237

CIS Representation Office,

Moscow

U1. Miklukho-Miklaia 16/10

117871 Moscow

Tel: 7 095 336 7000

Fax: 7 095 931 9193

Czech RepublicWaters Ges.m.b.H

Ohnivcova 11

14700 Praha 4 Branik

Tel: 42 2 472 8672

Fax: 42 2 472 8023

Denmark Waters A/S

Baldersbuen 46

2640 Hedehusene

Tel: 46 598080

Fax: 46 598585

Finland Waters Oy

Kuparitie 1

00440 Helsinki

Tel: 90 506 4140

Fax: 90 5064 1411

France Waters S.A

B.P. 608

78056 St. Quentin en Yvelines

Cedex

Tel: 1 30487200

Fax: 1 30487211

Germany Waters GmbH

Hauptstraße 87

D-65760 Eschborn

Tel: 06196 40 06 00

Fax: 06196 48 23 88

Hong Kong Waters China Ltd.

Unit 1804, Seaview Commercial

Bldg

21-24 Connaught Road West

Sheung Wan, Hong Kong

Tel: 852 2964 1800

Fax: 852 2549 6802

Hungary Waters kft

Vaci ut 202

H-1138 Budapest

Tel: 36 1 270 5086

Fax: 36 1 270 5087

India Waters Pvt. Ltd.

No. 36A, II Phase

Peenya Industrail Area

Bangalore 560080

Tel: 80 8394799

Fax: 80 3392157

Italy Waters S.p.A.

Via Achille Grandi, 27

20090 Vimodrone MI

Tel: 02 2500565

Fax: 02 2501827

Japan Nihon Waters K.K.

No 5. Koike Bldg.

1-3-12 Kitashinagawa

Shinagawa-ku, Tokyo

Tel: 3 3471 7191

Fax: 3 3471 7116

Malaysia Waters Asia Ltd.

51A Jalan SS25/2

Taman Bukit Emas

47301 Petaling Jaya

Selangor, Malaysia

Tel: 3 704 8600

Fax: 3 704 8599

Mexico Waters S.A. DE C.V.

Adolfo Prieto No. 1634

Col. Del Valle

03100 Mexico, D.F

Tel: 525 524 7636

Fax: 525 524 7643

The NetherlandsWaters Chromatography B.V.

Postbus 379

Florijnstraat 19

4870 AJ ETTEN-LEUR

Tel: 076 508 7200

Fax: 076 508 7280

Norway Waters AS

Hvamstubben 17

2013 Skjetten

Tel: 47 638 46 050

Fax: 47 638 46 051

Peoples Republic of China - BeijingWaters China Ltd.

Suite 709-711, Xinzhong Building

No. 2 Xinzhong Xiejie

Gongti Beilu, Dongcheng District

Beijing, 100027 P.R. China

Tel: 860 10 6592 8917

Fax: 860 10 6506 0344

Poland Waters Sp.zo.o

ul. Lektykarska 25 m 10

01-687 Warszawa

Tel: 48 22 33 44 00

Fax: 48 22 33 09 87

Puerto RicoWaters Technologies Corporation

P.O. Box 6870

Caguas, Puerto Rico 00726-6870

Tel: 787 747 8445, 790 2225

Fax: 787 747 8448

Singapore Waters Asia Ltd.

(Asia Headquarters)

No. 12 Prince Edward Road

#01-03 Podium A

Bestway Building

Singapore 079212

Tel: 225 4855

Fax: 225 4655

Spain Waters Chromatografia S.A.

Entenza 24- planta baja

E-08015 Barcelona

Tel: 3 325 9616

Fax: 3 325 9896

Sweden Waters Sverige AB

Box 485, Turebergsvagen 3

S-191 24 Sollentuna

Tel: 08623 0090

Fax: 08623 0095

Switzerland Waters AG

Dorfstrasse 10

5102 Rupperswil

Tel: 062 889 2030

Fax: 062 889 2059

Taiwan Waters Asia Ltd.

Room #603, 6th Floor, No.

136

Sec. 3, Jen-Ai Road

Taipei, Taiwan, 10628, R.O.C.

Tel: 02 706 8766

Fax: 02 706 9704

UK and Ireland Waters Ltd.

The Boulevard

Blackmoor Lane

Watford, Hertsfordshire

WD1 8YW

Tel: 01 923 816700

Fax: 01 923 219012

All Other CountriesWaters Corporation

34 Maple Street

Milford, MA 01757 USA

Tel: 508 478 2000

Toll-free: 1 800 252 4752

Fax: 508 872 1990

http://www.waters.com/

Page 16: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

16© 1997 Waters Corporation

Papers to be presented at HPLC ‘97

Dorothy J. Phillips, Edouard S. P. Bouvier,Mark Capparella, Yung-Fong Cheng, Pamela Iraneta, Uwe D. Neue andLaura L. Bean

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Yung-Fong Cheng, Dorothy J. Phillips, Uwe D. Neue and Laura Bean

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Method development experts in the pharmaceutical industry are striving to streamline

sample preparation. Considerable time and effort are consumed in choosing the

appropriate solid-phase extraction (SPE) sorbent and extraction protocol. The new Waters

Oasis™ HLB extraction cartridges address the problem of low recovery of polar analytes.

This paper includes the extraction procedures, HPLC methods and the results for several

highly polar drugs.

Dorothy J. Phillips, Mark Capparella, Yung-Fong Cheng, Pamela Iraneta, Uwe D. Neue

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Steroids have many biological functions, making their analysis important in biomedical

applications. The ratio of the amount of a steroid to that of extraneous material is very low in

biological fluids. Therefore, extraction of the steroid from the biological matrix is necessary

prior to HPLC analysis. The solid-phase extraction method development process, giving details

at each step from sample preparation to elution, will be presented.

The development of assays for drugs and their metabolites from serum is time-consuming since

optimisation of the HPLC method and the sample preparation is difficult and tedious. It

would be desirable to have a well thought out strategy for HPLC method development which

can provide good peak shapes and good resolution for a wide range of compounds. In this

paper, a carefully planned HPLC method development protocol using Waters Symmetry®

columns for the separation of chlordiazepoxide and its metabolites (ranging from basic to

uacidic in nature), will be presented.

Simple and Rugged Solid-Phase Extraction (SPE) Method for the Determinationof Tetracyclines in Serum by HPLC Using a Volatile Mobile Phase

Applications of a Polymeric Extraction Sorbent for IsolatingHighly Polar Drugs From Biological Fluids

Development of Methods for Extraction of Neutral Steroids From BiologicalMatrices With a Polymeric Solid-Phase Sorbent

Simple and Fast Method Development For HPLC and Sample Preparation

Recently there has been considerable interest in the stability of silica-based reversed-phase

packings when exposed to high pH mobile phases. The most common failure mode is loss

of efficiency and peak symmetry due to voiding, as a result of dissolution of the silica particles.

In contrast to the results reported by other researchers, we find no correlation between stability

and the method of preparation of the silica particles

Tom Walter, Bonnie Alden and Ray Fisk

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Due to its high sensitivity, high selectivity and high sample throughput, LC/MS/MS has

become the choice for the analysis of drugs and their metabolites in biological fluids such as

serum and urine. This paper reports a simple and reproducible SPE method for the determination

of tetracycline (TC), minocycline (MC) and demeclocycline (DCC) in porcine serum using a

volatile HPLC mobile phase.

Yung-Fong Cheng, Dorothy J. Phillips and Uwe D. Neue

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

High pH Stability of Silica-Based Reversed-Phase HPLC Packings

Page 17: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

17© 1997 Waters Corporation

Yung-Fong Cheng, Dorothy J. Phillips, Uwe D. Neue and Laura Bean

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Sample preparation is a necessary step in the analysis of drugs from biological fluids

including serum and urine. This paper reports a simple, fast and rugged solid-phase extraction

(SPE) method for determination of two basic tricyclic antidepressants and their metabolites

using Waters Oasis™ HLB extraction cartridges. Extraction of other basic compounds using the

same SPE method will also be presented.

Edouard S.P. Bouvier, Randy E. Meirowitzand Uwe D. Neue

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Tom Walter, Bonnie Alden, Mark Capparellaand Randy Meirowitz

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Solid-phase extraction (SPE) is a routinely used chromatographic technique for sample enrich-

ment and sample cleanup. Processing aqueous samples by SPE is predominantly performed

using cartridges containing C18-bonded silica. Waters has introduced a porous polymeric

sorbent for reversed-phase solid-phase extraction. This sorbent possesses greater water-wetting

properties than traditional C18-silica sorbents. We have measured wetting properties for several

hydro-organic mixtures on both C18-silica and the polymeric sorbent and have correlated the

results to performance of the sorbents in solid-phase extraction.

The behavior of reversed-phase HPLC packings is strongly dependent on the interaction of the

mobile phase with the stationary phase. We have developed chromatographic protocols to

observe the wetting of reversed-phase packings, and determined the extent of wetting using

water/methanol and water/acetonitrile mobile phases of varying water content. Results will

be presented for a range of C18- and C8-silica packings showing the effects of different pore

size distributions and bonded-phase coverages.

Determination of Tricyclic Antidepressants in Serum by a Simple Solid-Phase Extraction (SPE)and a Well Designed HPLC Method

The Effect of Contact Angle and Wetting on Performance in Solid-Phase Extraction

Observations on the Wetting of Reversed-Phase HPLC Packings

Michael S. Young

Waters Corporation, 34 Maple Street,Milford, MA 01757 USA

Recently, a new cartridge has been introduced for reversed-phase solid-phase extraction (SPE).

The Oasis™ HLB cartridge contains a hydrophilic/lipophilic balanced polymeric resin which

has been demonstrated effective for the extraction of polar pharmaceuticals and metabolites.

In this research, the cartridge is shown to be highly effective for the demanding analysis of

water samples for polar pesticides and related compounds. Among the applications discussed

in this presentation are sulfonylurea herbicides, triazine herbicides and metabolites, phenolic

compounds, and water soluble pesticides such as acephate. For most analytes, using a

75 mL water sample and a 1 mL elution volume, detection limits at or below 100 ng/L are

routinely achieved.

Fast HPLC Determination of Pesticides and Related Compounds in Water Using a Novel Polymeric Solid-Phase Extraction (SPE) Cartridge

Page 18: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

18© 1997 Waters Corporation

For our new Gelman Sciences filters brochurewith complete ordering information, pleasecheck box 8 on the business reply card.

What’s New

With the Waters Edge Frequent Buyer Program, you can earn points with every purchase of Gelman Sciences filtration products from Waters. You can then redeemthese points for products, services and merchandise found in the Waters EdgeFrequent Buyer brochure. The points are automatically shipped from Waters with the Gelman Sciences products .

The new extraction manifold has enhanced features designed for use with conventionalsilica-based solid-phase extraction cartridges as well as modifications that allow you totake full advantage of the unique performance characteristics of our new Oasis™ HLBextraction cartridges

Free gifts just for buying Gelman Sciences filters from Waters

Waters Spherisorb® columns are superior in quality and competitively priced. They are available everywhere through the Waters worldwide distribution network. Buyingyour Spherisorb® column from Waters assures you the best packed Spherisorb®

column you can buy.

For the new Waters Spherisorb® columnsbrochure, please check box 10 on the business reply card.

he New Waters Extraction Manifold

Capillary Zone Electrophoresis (CZE) hasbeen demonstrated to be a highly efficientseparation technique for charged ana-lytes. However, in its native form, it isincapable of separating neutral molecules.On the other hand, high-performanceliquid chromatography (HPLC) is anestablished technique with a vast numberof stationary phases available, but it isrestricted in the use of small particles dueto pressure constraints. CapillaryElectroChromatography (CEC) uses anelectric field rather than hydraulic pres-sure to drive the mobile phase through the packed bed. Since the resulting flow is very uniform, very high efficiencies are achieved.

In our next issue, we will explore the fun-damental principles of CEC highlightingseveral exciting applications. The authorof the article is Norman Smith of

lectroChromatography

Waters is now offering a line of Capillary ElectroChromatography products manufactured by Innovatech Ltd. for Waters.

he Waters EDGE Frequent Buyer Program

For information on the Waters ExtractionManifold, please check box 11 on the business reply card.

For information on the Waters Edge FrequentBuyer Program, please check box 9 on the business reply card.

Copies of this article can be obtained now by checking box 12on the business reply card.

aters Spherisorb® Columns

Description Dimension Part No.

Waters Spherisorb® S3 ODS1 50 µm i.d. x 25 cm WAT058919Waters Spherisorb® S3 ODS1 50 µm i.d. x 35 cm WAT058915Waters Spherisorb® S3 ODS1 75 µm i.d. x 25 cm WAT058907Waters Spherisorb® S3 ODS1 75 µm i.d. x 35 cm WAT058911Waters Spherisorb® S3 ODS1 100 µm i.d. x 25 cm WAT058903Waters Spherisorb® S3 ODS1 100 µm i.d. x 35 cm WAT058899Waters Spherisorb® S3 C8 50 µm i.d. x 25 cm WAT058895Waters Spherisorb® S3 C8 75 µm i.d. x 25 cm WAT058891Waters Spherisorb® S3 C8 100 µm i.d. x 25 cm WAT058887

Innovatech, one of the leading innovators in the field of CEC. CEC is a highly promisingtechnique that couples the very high efficiencies associated with CZE with reversed-phasechromatography. The theory predicts a two to three-fold increase in efficiency going froma pressure driven system (HPLC) to an electrically driven system (CEC) at equal columnlength. Due to the differences in the driving mechanism, however, gains of over a factorof 10 can be achieved by going from an HPLC separation to a CEC separation.

E

T

W

T

Page 19: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

Androgens: Testosterone Esters

SPE Spiked Serum Results

Waters Corporation 34 Maple Street, Milford, Massachusetts 01757-3696 U.S.A. 508 478-2000

.008

0

1

2

34

10 20

Minutes

AU

HPLC Analysis Conditions:

Column: Symmetry® C8, 3.5 µm, 4.6 mm x 75 mm with Sentry™ guard column, 3.9 mm x 20 mm

Sample: 20 µL of reconstituted porcine serum extract

Mobile phase: Acetonitrile/water 60:40

Flow rate: 1.5 mL/min

Detection: UV at 254 nm

Drug Concentration µg/mL % Recovery % RSD (n=6)

Testosterone 1.00 98% 1.2%Acetate 0.200 94% 3.7%

Testosterone 1.00 100% 1.9%Propionate 0.200 94% 4.3%

Testosterone 1.00 94% 1.6%Benzoate 0.200 87% 3.2%

Condition1 mL methanol/1 mL water

Elute1 mL methylene chloride:

methanol 50:50 (v/v)

Wash1 mL 5% methanol in water

Load1 mL sample with 800 µLspiked porcine serum and

200 µL of methanol

Evaporate and Reconstitute40 ˚C under nitrogen stream

500 µL methanol with 5 µg/mL of testosterone enanthate

Solid-Phase Extraction Procedure

Oasis™ HLB 1 cc Extraction Cartridge

System:

A Waters 717plus Autosampler and a Waters 625 LC System.

Detection was by UV using a Waters 486 Tunable UV/VIS

Absorbance Detector. Data acquisition was performed using a

Waters Millennium® 2010 Chromatography Manager.

A 20-position vacuum manifold with a vacuum pump was

used to process solid-phase extraction.

Chromatogram of Serum Extracts: A) Blank B) Spiked Sample

© 1997 Waters Corporation

O

O

O

CH3

O

O

O

O

O

O

O

O

O

CH3

CH3

Testosterone acetate

Testosterone propionate

Testosterone benzoate

Testosterone enanthate (I.S.)

Peaks:

1: Testosterone acetate

2: Testosterone propionate

3: Testosterone benzoate

4: Testosterone enanthate (I.S.)

Page 20: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

© 1997 Waters Corporation

Waters 717plus Autosampler

The Waters 717plus Autosampler combines reliability,

versatility and low cost of ownership into a fully automated, all-

electric, easy-to-use instrument that will meet and exceed your

requirements. The 717plus Autosampler incorporates a highly

reliable and extremely precise fluid path design that is versatile

enough to use for all your methods and applications. Advanced

automation capabilities make it easy to develop methods using

automated derivatisation techniques or to automatically add

reference standards or diluents. The optional Heating/Cooling

module allows you to work with a wide range of samples, from

heat labile biological samples to viscous polymer samples,

with exceptional precision and reliability.

Waters 625 LC System

The Waters 625 LC System combines advanced polymeric

technology and low dispersion system volume into a single

liquid chromatograph providing the highest performance available

in a non-metallic system. This instrument is designed to handle a

wide variety of life science applications such as high-resolution

protein purification, microbore peptide mapping, nucleic acid iso-

lations, and purification and analysis of oligosaccharides. These

applications require an array of different column chemistries.

The Waters 625 LC System is optimised for ion-exchange, gel

filtration, affinity, hydrophobic interaction and reversed-phase

column chemistries used in microbore, analytical or microprepara-

tive methods relevant to the biochemist. The Waters 625 LC

System is designed to be easy to operate, and to maintain,

for reliable daily operation.

Millennium® 2010 Chromatography Manager

Advanced computer technology simplifies accessing, interpreting, and reporting your results.

The productivity of today’s chromatography is centered around

the ability to find, generate, summarise, and report results in a

way that meets your needs and complies with any regulatory

requirements. The Millennium® relational database lets you

create your own view filters to customise the way you sort

your information, such as “lotCode”.

• Using the database, you can track a method history giving

you audit trail capability that ensures total security of your

raw data and results. In addition, you can specify “Access

Type” which allows you to “lock” your methods for even

greater security.

• The database keeps track of your instruments, triggering

an alarm when scheduled maintenance is required. The

Millennium® relational database lets you view the injection

information on any calibration point at any particular point

in time on a processing sequence. This gives you a detailed

account of the processing method for that injection, ensuring

an accurate audit trail for regulatory compliance.

• The Millennium® Chromatography Manager can provide

acquisition and control for as many as four chromatographic

systems, including three HP5890 GC systems.

• The Millennium® Report Publisher has the capability to give

you exactly what you want.

The Waters 486 Tunable UV/VIS Absorbance Detector

The performance you expect with the flexibility you need.

The Waters 486 Tunable UV/VIS Absorbance Detector is

the highest performing UV/VIS detector with the flexibility to

accommodate virtually all your HPLC applications. Our

exclusive Taper-Cell® flow cell design eliminates refractive

index effects, providing exceptional baseline stability.

• The 486 detector is tunable from 190 to 600 nm

wavelength range with reproducibility and precision.

• The unique modular cell assemblies for microbore,

analytical, preparative, high-pressure mass spectrometry,

and nonmetallic flow paths give your application flexibility.

The quality management system of Waters manufacturing facility,Milford, Massachusetts, complies with the International StandardISO 9001 Quality Management and Quality Assurance Standards.Waters quality management system is periodically audited by theregistering body to ensure compliance.

Page 21: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

Morphine and Its Metabolites

SPE Spiked Serum Results

Waters Corporation 34 Maple Street, Milford, Massachusetts 01757-3696 U.S.A. 508 478-2000

6

0 5

1

2

3

10Minutes

15

mV

Compound Concentration µg/mL % Recovery % RSD (n=6)

M3G 0.48 90.7 2.00.097 100 3.1

M6G 2.4 92.2 2.80.49 93.1 2.5

Morphine 3.6 102 3.60.73 102 3.2

Chromatogram of Serum Extracts: A) Blank B) Spiked Sample

© 1997 Waters Corporation

O

O O

HO

OHHO

OH

COOH

HO

HNCH3

O

HO

HNCH3

O OOH

HO

OH

COOH

O

HO

HNCH3

Morphine

Morphine - 3 - Glucuronide

Morphine - 6 - Glucuronide

Column: SymmetryShield™ RP8, 3.9 mm x 150 mm

Sample: 100 µL of porcine serum extract

Mobile phase: 20 mM potassium phosphate, pH 6.4

Flow rate: 1.0 mL/min

Detection: Fluorescence: excitation at 280 nm,emission at 355 nm

Condition Cartridge1 mL methanol/1 mL water

Wash Cartridge1 mL water

Load Sample1 mL sample spiked

porcine serum

Equilibrate Cartridge1 mL water

Elute Analytes0.5 mL 3% triethylamine

Solid-Phase Extraction Procedure

Oasis™ HLB 1 cc Extraction Cartridge

Evaporate and ReconstituteAnalytes

No evaporation–sample injected directly

Peaks:

1. Morphine–3 glucuronide

2. Morphine–6 glucuronide

3. Morphine

System

A Waters 717plus Autosampler and a Waters 616 LC System.

Detection was by fluorescence using a Waters Fluorescence

Detector. Data acquisition was performed using a Waters

Millennium® 2010 Chromatography Manager. A 20-position

vacuum manifold with a vacuum pump was used to process

solid-phase extraction.

Page 22: Fast and Easy Solid-Phase Extraction with a New … · Fast and Easy Solid-Phase Extraction with ... A Unique Solution Waters new Oasis“ HLB extraction car- ... Salicylic acid,

© 1997 Waters Corporation

Millennium® 2010 Chromatography Manager

Advanced computer technology simplifies accessing, interpreting, and reporting your results.

The productivity of today’s chromatography is centered around

the ability to find, generate, summarise, and report results in a

way that meets your needs and complies with any regulatory

requirements. The Millennium® relational database lets you

create your own view filters to customise the way you sort your

information, such as “lotCode”.

• Using the database, you can track a method history giving

you audit trail capability that ensures total security of your

raw data and results. In addition, you can specify “Access

Type” which allows you to “lock” your methods for even

greater security.

• The database keeps track of your instruments, triggering an

alarm when scheduled maintenance is required. The

Millennium® relational database lets you view the injection

information on any calibration point at any particular point

in time on a processing sequence. This gives you a detailed

account of the processing method for that injection, ensuring

an accurate audit trail for regulatory compliance.

• The Millennium® Chromatography Manager can provide

acquisition and control for as many as four chromatographic

systems, including three HP5890 GC systems.

• The Millennium® Report Publisher has the capability to give

you exactly what you want.

Waters 717plus Autosampler

The Waters 717plus Autosampler combines reliability,

versatility and low cost of ownership into a fully automated, all-

electric, easy-to-use instrument that will meet and exceed your

requirements. The 717plus Autosampler incorporates a highly

reliable and extremely precise fluid path design that is versatile

enough to use for all your methods and applications. Advanced

automation capabilities make it easy to develop methods using

automated derivatisation techniques or to automatically add

reference standards or diluents. The optional Heating/Cooling

module allows you to work with a wide range of samples, from

heat labile biological samples to viscous polymer samples, with

exceptional precision and reliability.

Waters 616 LC System

Waters has incorporated the latest technological advances into

the new Waters 616 LC system. The 616 is the culmination of

leading innovative technology—fluidic, electronic, and soft-

ware—designed to deliver superior performance in a premier

pumping system. The 616 offers pulse-free flow, low delay

volume, efficienct gradient mixing, precision, accuracy, and

reliability.

Waters Fluorescence Detector

Fluorescence detection offers both greater sensitivity and

selectivity for compounds that fluoresce naturally, or can react

reproducibly with reagents to form stable fluorescent compounds.

Waters Scanning Fluorescence Detector enables you to easily

apply this technology with unbeatable performance and versa-

tility. A dual monochromator design provides independent

programming of both excitation and emission wavelengths to

improve individual analyte detectability. The Waters Fluorescence

Detector offers the sensitivity you need for applications where

very low concentrations of targeted compounds are analysed.

The quality management system of Waters manufacturing facility,Milford, Massachusetts, complies with the International StandardISO 9001 Quality Management and Quality AssuranceStandards. Waters quality management system is periodicallyaudited by the registering body to ensure compliance.