Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2 1/44
Comparison Report on Calibration Factor
Measurement of Power Sensor
Final
GULFMET.EM.RF-S2
Coordinating Laboratory;
Dr. Erkan DANACI, Özlen TUNCEL, Dr. Murat CELEP
TUBITAK UME, National Metrology Institute, Turkey
Contributing Laboratory:
Ahmed Hamad ALJAWAN, Ibrahim ALBORAIH, Khalid Saleh ALDAWOOD
SASO NMCC, National Measurement and Calibration Center, Saudi Standards,
Metrology and Quality Organization, Kingdom of Saudi Arabia
March 12, 2019
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2 2/44
CONTENTS
CONTENTS ....................................................................................................................... 2
1. INTRODUCTION .................................................................................................... 3
2. TRAVELLING STANDARD ..................................................................................... 3
3. PARTICIPANT LABORATORIES ........................................................................... 5
4. TIME SCHEDULE ................................................................................................... 5
5. MEASUREMENT TECHNIQUE .............................................................................. 6
6. DISCUSSION OF THE RESULTS .......................................................................... 6
REFERENCES ................................................................................................................ 15
ANNEX A. COMPARISON MEASUREMENTS REPORT of TÜBİTAK UME .................. 16
ANNEX B. COMPARISON MEASUREMENTS REPORT of SASO NMCC ..................... 37
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2 3/44
1. INTRODUCTION
It was planned to organise a comparison on calibration factor (CF) measurement of Power
Sensor between SASO NMCC and TÜBİTAK UME, on the frame of GULFMET.
Calibration factor is used for characterization of RF power sensors. RF power sensors are used
to measure RF power, and it has a fundamental importance of achieving the traceability of RF
power measurements. The bilateral comparison is performed by measuring a power sensor at
frequencies of 10 MHz, 50 MHz, 1 GHz, 4 GHz, 8 GHz, 12 GHz, 15 GHz, and 18 GHz, and a
power level of 0 dBm.
TÜBİTAK UME is acting as the pilot laboratory. The travelling standard was provided by SASO
NMCC. TÜBİTAK UME is responsible for monitoring standard performance during the circulation,
and the evaluation and reporting of the comparison results.
The comparison was carried out in accordance with the CCEM Guidelines for Planning,
Organizing, Conducting and Reporting Key, Supplementary and Pilot Comparisons [1].
2. TRAVELLING STANDARD
The travelling standard, Agilent N8481A Power Sensor (Figure 1), has identification as follows:
Manufacturer : Agilent
Model : N8481A
Serial No : MY54220011
Top View
Bottom View
Figure 1. Travelling standard is an Agilent N8481A Power Sensor
The N8481A power sensor, henceforth denoted by the Standard, has an N type male connector
and a DC output port for connecting the power sensor to the power meter. The standard has a
power range from -35 dBm to +20 dBm. Within this comparison, only the 0 dBm power level was
measured.
A measured quantity in the comparison is a Calibration Factor (CF). CF is relative to 50 MHz is
defined by;
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2 4/44
CFrel=Pind
Pin
.Pin_ref
Pind_ref
.M (1)
where;
Pind : indicated power on power meter,
Pin : incident RF power to the power sensor (standard) at the measurement frequency,
Pind_ref : indicated power on power meter at 50 MHz,
Pin_ref : incident RF power to the power sensor (standard) at 50 MHz.
M : Vectorial mismatch correction
Participants submitted the measured calibration factor and reflection coefficient results of the
standard at 8 frequencies (10 MHz, 50 MHz, 1 GHz, 4 GHz, 8 GHz, 12 GHz, 15 GHz, and 18
GHz) concerning its calibration factor and reflection coefficient. Both measurands have extended
uncertainty values (coverage factor k=2). 50 MHz is selected to be a reference frequency of
calibration factor measurements.
The travelling standard was chosen for its common usage frequency and power range and it can
be available at most NMIs for RF power measurement. The general specifications of standard are
given in Table 1.
Table 1. The general specifications of N8481A standard
Operating Frequency 10 MHz to 18 GHz
Maximum Power 500 mW (average)
Connector Type Type N male
Maximum SWR 1.37 (10 MHz to 30 MHz)
1.14 (30 MHz to 50 MHz)
1.08 (50 MHz to 2 GHz)
1.16 (2 GHz to 12.4 GHz)
1.23 (12.4 MHz to 18 GHz)
Dimensions Height 30 mm
Width 38 mm
Long 130 mm
Weight 0.181 kg
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2 5/44
3. PARTICIPANT LABORATORIES
The pilot institute for this comparison is TÜBİTAK UME (Turkey). The contact details of the
coordinator are given below:
Pilot Institute : TÜBİTAK Ulusal Metroloji Enstitüsü (UME)
Coordinator : Dr. Erkan DANACI
Tel: +90 262 679 50 00
Fax: +90 262 679 50 01
E-mail: [email protected]
The participating institutes and contact persons with their addresses are given in Table 2.
Table 2. Participants
Country Institute Acronym Shipping Address Contact Person
Turkey
TÜBİTAK
Ulusal Metroloji Enstitüsü
TÜBİTAK UME
TÜBİTAK Ulusal Metroloji Enstitüsü (UME)
TÜBİTAK Gebze Yerleşkesi
Barış Mah. Dr. Zeki Acar Cad. No:1
41470 Gebze-Kocaeli, TURKEY
Dr. Erkan DANACI
Tel: +90 262 679 50 00
Saudi Arabia
SASO
The National Measurement and Calibration Center
SASO NMCC
Saudi Standards, Metrology and Quality Organization of The Kingdom of Saudi Arabia
(SASO)
Riyadh 11471, P.O. Box 3437
KINGDOM of SAUDI ARABIA
Ahmed Hamad AlJawan
Tel: +966 11 252 9741
4. TIME SCHEDULE
The time schedule for the comparison is given in the Table 3. The circulation of travelling
standard was organized to monitor the performance of the travelling standard. Each laboratory
had 2 weeks to carry out the measurements.
Table 3. Circulation Time Schedule
Acronym of Institute
Country Starting Date Time for measurement
and transportation
TÜBİTAK UME Turkey 01.03.2018 – 15.03.2018 2 weeks
SASO NMCC Saudi Arabia 26.06.2018 – 19.07.2018 3 weeks
TÜBİTAK UME Turkey 01.10.2018 – 22.10.2018 3 weeks
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2 6/44
5. MEASUREMENT TECHNIQUE
Although, the main measured parameter was calibration factor, additionally the reflection
coefficient was measured at 10 MHz, 50 MHz, 1 GHz, 4 GHz, 8 GHz, 12 GHz, 15 GHz and 18
GHz frequency points.
Also the quantities given below had been measured;
Ambient temperature
Ambient humidity
Pin depth of the travelling standards
Pin depth of the connected ports of the measurement device
No correction was applied for the ambient temperature and relative humidity.
Pilot and the participant laboratories were used their own VNA for the measurement of reflection
coefficient of travelling standard.
Calibration factor measurement methods of the participants were explained at Appendix A and
Appendix B.
The measurements and comparison report were done according to the technical protocol of the
bilateral comparison.
6. DISCUSSION OF THE RESULTS
The results were presented to the pilot laboratory in the format of linear CF value. The chosen
frequencies are 10 MHz, 50 MHz, 1 GHz, 4 GHz, 8 GHz, 12 GHz, 15 GHz and 18 GHz for the
comparison.
The comparison was organised in a single loop of two laboratories.
The uncertainties of measurement were calculated according to the JCGM 100 “Guide to the
Expression of Uncertainty in Measurement” [2] for the coverage probability of approximately 95%.
All contributions to the measurement uncertainty listed in the report submitted by each
participant.
Participant was also asked to provide detailed uncertainty budget and the combined standard
uncertainty for the aforementioned measurand.
The measurement results and associated uncertainties together with the reference values and
associated uncertainties can be found in Tables 4 to 6. UME1, SASO and UME2 representing the
first measurement at TÜBİTAK UME, measurement at SASO NMCC and final measurement at
TÜBİTAK UME respectively. 50 MHz reference value of the two laboratories are given 1.000.
Comparison reference values and associated uncertainties were calculated using the
measurements result of TÜBİTAK UME as given below. TÜBİAK UME was participant of the
CCEM.RF-K8.CL [3].
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2 7/44
Reference Value=Xref=UUME1 + UUME2
2 (2)
Uncertainty of Reference Value=Uref=2.√(UUME1-UUME2
2.√3)
2
+(max(UUMEi)
2)
2
(3)
Where max(UUMEi) is maximum uncertainty value of UME measurements.
Degrees of equivalence, D, is calculated by subtracting the reference value from the each
measurements (Equation (4)) and its uncertainty is calculated according to the Equation (5). The
graphics of D are given in Figure 2-8.
D=Xlab-Xref (4)
UD=√Uref2
+Ulab2
(5)
Degrees of equivalence with respect to the reference value and between each of the
measurements can be found in Table 5.
The En value which is calculated according to Equation (6) stated at ISO / IEC 17043 “Conformity
assessment — General requirements for proficiency testing” [4] Standard for each measurement
and frequencies can be found in Table 6.
En=Xlab-Xref
√Uref2
+Ulab2
(6)
xlab: Participant laboratory measurement result
xref: Pilot laboratory measurement result
Ulab: Participant laboratory measurement uncertainty
Uref: Pilot laboratory measurement uncertainty
The laboratory measurement results can be utilized according to the criteria of En value which is
given below.
If │En│≤ 1 then it is successful
If │En│> 1 then it is unsuccessful
Uncertainty budget provided by the participant and the full set of reported results can be found in
Appendix A and Appendix B.
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
8/44
Table 4. Calibration Factor Measurement results of HP8481A Power Sensor
Laboratory UME1 SASO UME2 Reference
Value
Frequencies CF Unc. CF Unc. CF Unc. xref Uref
10 MHz 0.9829 0.0070 0.9776 0.0083 0.9836 0.0071 0.9833 0.0071
50 MHz
(reference) 1.0000 0.0065 1.0000 0.0084 1.0000 0.0065 1.0000 0.0065
1 GHz 0.9936 0.0063 0.9976 0.0086 0.9936 0.0063 0.9936 0.0063
4 GHz 0.9822 0.0062 0.9853 0.0090 0.9822 0.0063 0.9822 0.0063
8 GHz 0.9796 0.0062 0.9814 0.0127 0.9798 0.0063 0.9797 0.0063
12 GHz 0.9815 0.0064 0.9858 0.0133 0.9818 0.0064 0.9817 0.0064
15 GHz 0.9844 0.0066 0.9884 0.0151 0.9851 0.0067 0.9847 0.0067
18 GHz 0.9940 0.0162 0.9966 0.0267 0.9936 0.0133 0.9938 0.0163
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
9/44
Table 5. Degrees of Equivalence and its uncertainty
Labs UME1 SASO UME2
Frequencies D UD D UD D UD
10 MHz -0.0003 0.0100 -0.0057 0.0109 0.0003 0.0100
1 GHz 0.0000 0.0089 0.0040 0.0106 0.0000 0.0089
4 GHz 0.0000 0.0089 0.0031 0.0110 0.0000 0.0089
8 GHz -0.0001 0.0089 0.0017 0.0142 0.0001 0.0089
12 GHz -0.0001 0.0090 0.0041 0.0148 0.0001 0.0091
15 GHz -0.0004 0.0094 0.0037 0.0165 0.0004 0.0095
18 GHz 0.0002 0.0230 0.0028 0.0313 -0.0002 0.0211
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
10/44
Table 6. En values of the comparison
Labs |En|
Frequencies UME1 SASO UME2
10 MHz 0.03 0.58 0.03
1 GHz 0.00 0.41 0.00
4 GHz 0.00 0.31 0.00
8 GHz 0.01 0.13 0.01
12 GHz 0.01 0.29 0.01
15 GHz 0.04 0.23 0.04
18 GHz 0.01 0.10 0.01
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
11/44
Figure 2. CF at 10 MHz
Figure 3. CF at 1 GHz
UME1
SASO
UME2
-0.030
-0.020
-0.010
0.000
0.010
0.020
0.030
D (
CF
)
Laboratory
CF @10 MHz
UME1SASO
UME2
-0.030
-0.020
-0.010
0.000
0.010
0.020
0.030
D (
CF
)
Laboratory
CF @1 GHz
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
12/44
Figure 4. CF at 4 GHz
Figure 5. CF at 8 GHz
UME1 SASO UME2
-0.030
-0.020
-0.010
0.000
0.010
0.020
0.030
D (
CF
)
Laboratory
CF @4 GHz
UME1 SASO UME2
-0.030
-0.020
-0.010
0.000
0.010
0.020
0.030
D (
CF
)
Laboratory
CF @8 GHz
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
13/44
Figure 6. CF at 12 GHz
Figure 7. CF at 15 GHz
UME1SASO
UME2
-0.030
-0.020
-0.010
0.000
0.010
0.020
0.030
D (
CF
)
Laboratory
CF @12 GHz
UME1SASO UME2
-0.030
-0.020
-0.010
0.000
0.010
0.020
0.030
D (
CF
)
Laboratory
CF @15 GHz
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
14/44
Figure 8. CF at 18 GHz
UME1 SASO UME2
-0.040
-0.030
-0.020
-0.010
0.000
0.010
0.020
0.030
0.040
D (
CF
)
Laboratory
CF @18 GHz
Draft A Report of GULFMET Supplementary Comparison for Calibration Factor Measurement of Power Sensor
GULFMET.EM.RF-S2
15/44
REFERENCES
[1] CCEM Guidelines for Planning. Organizing. Conducting and Reporting Key. Supplementary
and Pilot Comparisons. 2007 (available on the BIPM website:
http://www.bipm.org/utils/common/pdf/CC/CCEM/ccem_guidelines.pdf)
[2] Evaluation of measurement data - Guide to the Expression of Uncertainty in Measurement
(GUM). JCGM 100. First edition. September 2008 (available on the BIPM website:
http://www.bipm.org/utils/common/documents/jcgm/JCGM_100_2008_E.pdf)
[3] CCEM.RF-K8.CL Comparison Calibration Factor of Thermistor Mounts, May 2005
(https://www.bipm.org/utils/common/pdf/final_reports/EM/RF/K8/CCEM.RF-K8.CL.pdf)
[4] ISO / IEC 17043 “Conformity assessment — General requirements for proficiency testing”.
International Standardization Organization”. 2010
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
16/44
ANNEX A. COMPARISON MEASUREMENTS REPORT of TÜBİTAK UME
1. PARTICIPANT INFORMATION
Laboratory Name RF and Microwave
Related Person Name Dr. Erkan DANACI
Telephone No +90 262 679 5000
Fax No +90 262 679 5001
E-mail [email protected]
Adress TÜBİTAK UME Gebze Yerleşkesi Barış Mah.Dr. Zeki Acar Cad. No: 1
Gebze TR-41470 Kocaeli TURKEY
2. MEASUREMENT DATE
First measurement : 01.03.2018 – 15.03.2018
Second measurement : 01.10.2018 – 22.10.2018
3. ENVIRONMENTAL CONDITION
Temperature : (23 ± 1) C
Relative Humidity : (45 ± 10) rh%
4. REFERENCES USED IN MEASUREMENT
Instrument Name Manufacturer Type / Model Serial No Traceability
Signal Generator Agilent E8257D MY45140665 UME, G1ZF-0091,
14.09.2018
Power Splitter Weinschel 1870A 9879 UME, G1MD-0002,
10.01.2017
Thermistor Mount (STD) Agilent
Technologies 8478B 3318A25886
UME, G1MD-0009, 12.01.2018
Powermeter (STD) Hewlett Packard
432A 3303U00513 UME, G1MD-0069,
09.05.2017
Powermeter (DUT) Keysight N1914A_A MY56500012 UME, G1MD-0054,
02.05.2017
Powermeter (MON) Keysight N1914A_B MY56500012 UME, G1MD-0054,
02.05.2017
Voltmeter (STD) Hewlett Packard
34401 US36013987 UME, G1LV-0103,
20.06.2018
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
17/44
Instrument Name Manufacturer Type / Model Serial No Traceability
Vector Network Analyzer Rohde & Schwarz
ZVA 50 100147 UME, G1MD-0145,
17.11.2017
Calibration Kit Hewlett Packard HP85054B 3106A01936
UME, G1MD-0145,
17.11.2017
VNA Cables Gore FE0AL0BL025.0
FE0AH0BL025.0
9215621
9215619
UME, G1MD-0145,
17.11.2017
5. MEASUREMENT METHOD
SOL (Short, Open, Load) calibration method is applied to measure reflection coefficient of the
travelling standard (DUT). Male connector of the power sensor was measured as S11. The
measurements were done with VNA Tolls II software and data and uncertainty evaluated by using
excel macro software (VNAUnc_Makro_ReIm3Olcum_r3v1).
In order to measure calibration factor of the DUT, software developed by TÜBİTAK UME was
used. Measurement software flow graph is given in Figure 1. According to the software;
calibration factor of the travelling standard was performed using direct comparison transfer
method at 0 dBm for 3 orientations (0°, 120° and 240°). Standard thermistor mount and the
travelling standard were connected to one of the output port of power splitter while monitor power
sensor was attached to the other output port of the power splitter. Microwave power applied to
the input port of the power splitter and standard power, then monitor power were measured. After
the measurement was finished, standard thermistor mount (STD) was disconnected from power
splitter port and DUT was connected immediately. Then, again, microwave power was applied to
input port and output powers were measured using DUT and monitor power meters. The STD,
DUT parameters and related powers were used to calculate calibration factor of the DUT.
Mismatch correction was applied. “Calibration factor.exe” software was used for the
measurements.
The declared final calibration factors are relative to 50 MHz.
Measurements were performed by Özlen TUNCEL and Erkan DANACI.
The measurements and comparison report were done according to the technical protocol of the
bilateral comparison.
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
18/44
Introducing the New Device
and
Entering the Parameters of
Devices
Start
Selecting the MON and
STD Power Sensors
According to the DUT
Power Sensor
Updating the Parameters of
Older Devices
Selecting the New/Older
Measurement Setup
Selecting the Devices on
Measurement Setup
Entering the Measurement
Parameters
Selecting the Frequency
Measuring MON Line Power at Selected
Frequency
Calculating the to be Applied Power at
Selected Frequency
Last
Frequency?
No
Selecting the Orientation of STD&DUT
Power Sensor
Selecting the Frequency
Measuring MON Line Power at Selected
Frequency
Measuring STD Line Power at Selected
Frequency
Last
Frequency?
No
Selecting the Frequency
Measuring MON Line Power at Selected
Frequency
Measuring DUT Line Power at Selected
Frequency
Last
Frequency?
Last
Orientation?
No
No
Calculating the CF of DUT and İts Uncertainty
Stop
Yes
Yes
Yes
Yes
MON
Power
To Be
Applied
Power
MON
Power
STD
Power
MON
Power
DUT
Power
STD, MON and
DUT Power
Figure 1. Software Flow Diagram
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
19/44
6. MEASUREMENT RESULT
6.1. First Measurement Results at TÜBİTAK UME
Measured pin depth: -0.00180 inches, U = 0.00016 inches (k=2)
Table 1. Reflection coefficient of the travelling standard
Frequency
(GHz)
Reflection Coefficient ( jyxS11 ) of DUT Power Sensor
Real Component
x
Real Component
Uncertainty (k=2)
Imaginary
Component
y
Imaginary
Component
Uncertainty (k=2)
0.01 0.0210 0.0042 -0.1030 0.0042
0.05 0.0095 0.0040 -0.0217 0.0040
1 0.0049 0.0040 -0.0092 0.0040
4 -0.0036 0.0040 0.0074 0.0040
8 -0.0038 0.0040 -0.0127 0.0040
12 -0.0155 0.0080 -0.0177 0.0080
15 0.0441 0.0080 -0.0120 0.0080
18 -0.0283 0.0080 0.0306 0.0080
Table 2. Calibration factor (reference to 50 MHz)
Frequency (GHz) Calibration Factor Calibration Factor
uncertainty (k=2)
0.01 0.9829 0.0070
0.05 1.0000 Reference
1 0.9936 0.0063
4 0.9822 0.0062
8 0.9796 0.0062
12 0.9815 0.0064
15 0.9844 0.0066
18 0.9940 0.0162
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
20/44
6.2. Second Measurement Results at TÜBİTAK UME
Measured pin depth: -0.0018 inches, U (k=2)= 0.00016 inches
Table 3. Reflection coefficient of travelling standard
Frequency
(GHz)
Reflection Coefficient ( jyxS11 ) of DUT Power Sensor
Real Component
x
Real Component
Uncertainty (k=2)
Imaginary
Component
y
Imaginary
Component
Uncertainty (k=2)
0.01 0.0196 0.0042 -0.1046 0.0042
0.05 0.0082 0.004 -0.0219 0.004
1 0.0064 0.004 -0.0054 0.004
4 -0.0036 0.0041 0.0077 0.0041
8 -0.0049 0.0043 -0.0126 0.0043
12 -0.016 0.0081 -0.0187 0.0081
15 0.0463 0.0081 -0.009 0.0081
18 -0.0284 0.0082 0.0303 0.0082
Table 4. Calibration factor (reference to 50 MHz)
Frequency Calibration Factor Calibration Factor uncertainty
(k=2)
0.01 0.9836 0.0071
0.05 1.0000 Reference
1 0.9936 0.0063
4 0.9822 0.0063
8 0.9798 0.0063
12 0.9818 0.0064
15 0.9851 0.0067
18 0.9936 0.0133
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
21/44
7. UNCERTAINTY BUDGET
7.1. Calibration Factor Measurements Uncertainty Budget at First TÜBİTAK UME
Measurement
TÜBİTAK UME measured the travelling standard in 3 orientations. The measured values of
second orientation are given in Table 5-12. Uncertainty of the measurement contains the
reproducibility error of three orientations.
Model function :
STDGESTDGEGESTD
DUTGEDUTGEGEDUT
D
S
STD
DUTSTDDUT
Cos
Cos
Pm
Pm
P
PCFCF
21
2122
22
+R
Table 5. Calibration factor uncertainty budget form at 10 MHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.99905 0.00043 Normal (k=1) -0.98294 1.81E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
0.992157 0.000048 Normal (k=1) 0.989769 2.22E-09
Power of DUT Power Sensor-PDUT (mW)
1.00146 0.00043 Normal (k=1) 0.98057 1.81E-07
Reproducibility – R 0.00000 0.00285 Normal (k=1) 1.00000 8.12E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.996584 0.000077 Normal (k=1) -0.985372 5.73E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9840 0.0038 Normal (k=2) 0.9980 3.60E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.1051 0.0021 Normal (k=1) -0.0003 4.92E-13
Reflection Coefficient of STD Power sensor - ΓSTD
0.0588 0.0021 Normal (k=1) 0.0008 2.61E-12
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0017 0.0025 Normal (k=1) 0.0067 2.83E-10
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-1.3695 0.0200 Normal (k=1) 0.0003 4.75E-11
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-1.5095 0.0349 Normal (k=1) -0.0002 4.29E-11
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.8387 1.4914 Normal (k=1) 0.0002 5.43E-08
Measured Value 0.9820 Combined Uncertainty 0.0035
Expanded Uncertainty (k=2) 0.0070
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
22/44
Table 6. Calibration factor uncertainty budget form at 50 MHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.99737 0.00043 Normal (k=1) -1.00180 1.87E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.001078 0.000039 Normal (k=1) 0.998087 1.48E-09
Power of DUT Power Sensor-PDUT (mW)
1.00058 0.00043 Normal (k=1) 0.99858 1.87E-07
Reproducibility – R 0.00000 0.00275 Normal (k=1) 1.00000 7.56E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
1.004843 0.000033 Normal (k=1) -0.994348 1.11E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9997 0.0033 Normal (k=2) 0.9995 2.72E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0237 0.0020 Normal (k=1) -0.0024 2.27E-11
Reflection Coefficient of STD Power sensor - ΓSTD
0.0129 0.0020 Normal (k=1) 0.0046 8.45E-11
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0026 0.0025 Normal (k=1) 0.0012 8.73E-12
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-1.15737 0.08463 Normal (k=1) 0.00011 8.27E-11
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-1.78947 0.15495 Normal (k=1) -0.00003 2.03E-11
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.24554 0.97609 Normal (k=1) 0.00008 5.85E-09
Measured Value 0.9992 Combined Uncertainty 0.0033
Expanded Uncertainty (k=2) 0.0065
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
23/44
Table 7. Calibration factor uncertainty budget form at 1 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.99606 0.00043 Normal (k=1) -0.99715 1.85E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.000108 0.000026 Normal (k=1) 0.993118 6.81E-10
Power of DUT Power Sensor-PDUT (mW)
0.99584 0.00043 Normal (k=1) 0.99737 1.85E-07
Reproducibility – R 0.0000 0.0026 Normal (k=1) 1.0000 6.76E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.997577 0.000029 Normal (k=1) -0.995638 8.23E-10
Calibration Factor of the STD Power Sensor - CFSTD
0.9914 0.0033 Normal (k=2) 1.0018 2.73E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0104 0.0020 Normal (k=1) -0.0143 8.16E-10
Reflection Coefficient of STD Power sensor - ΓSTD
0.0231 0.0020 Normal (k=1) -0.0139 7.69E-10
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0077 0.0025 Normal (k=1) -0.0606 2.30E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-1.08177 0.19180 Normal (k=1) -0.00006 1.29E-10
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
2.88295 0.08670 Normal (k=1) 0.00015 1.74E-10
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
0.70302 0.32326 Normal (k=1) 0.00009 9.04E-10
Measured Value 0.9932 Combined Uncertainty 0.0031
Expanded Uncertainty (k=2) 0.0063
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
24/44
Table 8. Calibration factor uncertainty budget form at 4 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.99761 0.00043 Normal (k=1) -0.98382 1.82E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
0.992080 0.000046 Normal (k=1) 0.989304 2.04E-09
Power of DUT Power Sensor-PDUT (mW)
0.99213 0.00043 Normal (k=1) 0.98925 1.81E-07
Reproducibility – R 0.00000 0.00255 Normal (k=1) 1.00000 6.50E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.989665 0.000047 Normal (k=1) -0.991718 2.18E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9838 0.0033 Normal (k=2) 0.9976 2.71E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0083 0.0020 Normal (k=1) 0.0072 2.08E-10
Reflection Coefficient of STD Power sensor - ΓSTD
0.0512 0.0020 Normal (k=1) 0.0122 5.99E-10
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0066 0.0025 Normal (k=1) 0.1041 6.77E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
2.02578 0.24218 Normal (k=1) -0.00009 4.61E-10
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-2.42534 0.03908 Normal (k=1) 0.00021 7.02E-11
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.09559 0.37957 Normal (k=1) 0.00013 2.28E-09
Measured Value 0.9815 Combined Uncertainty 0.0031
Expanded Uncertainty (k=2) 0.0062
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
25/44
Table 9. Calibration factor uncertainty budget form at 8 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.99639 0.00043 Normal (k=1) -0.98227 1.80E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
0.992916 0.000047 Normal (k=1) 0.985711 2.15E-09
Power of DUT Power Sensor-PDUT (mW)
1.00091 0.00043 Normal (k=1) 0.97784 1.80E-07
Reproducibility – R 0.00000 0.00255 Normal (k=1) 1.00000 6.50E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.996522 0.000040 Normal (k=1) -0.982145 1.52E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9762 0.0033 Normal (k=2) 1.0026 2.74E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0133 0.0020 Normal (k=1) -0.0424 7.21E-09
Reflection Coefficient of STD Power sensor - ΓSTD
0.0551 0.0020 Normal (k=1) 0.0402 6.45E-09
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0244 0.0025 Normal (k=1) 0.0676 2.86E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-1.8640 0.1506 Normal (k=1) 0.0003 1.93E-09
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-2.9147 0.0363 Normal (k=1) 0.0014 2.69E-09
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.3417 0.1025 Normal (k=1) 0.0017 3.11E-08
Measured Value 0.9787 Combined Uncertainty 0.0031
Expanded Uncertainty (k=2) 0.0062
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
26/44
Table 10. Calibration factor uncertainty budget form at 12 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.99860 0.00043 Normal (k=1) -0.98171 1.82E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.002851 0.000041 Normal (k=1) 0.977544 1.61E-09
Power of DUT Power Sensor-PDUT (mW)
1.01973 0.00044 Normal (k=1) 0.96136 1.80E-07
Reproducibility – R 0.00000 0.00265 Normal (k=1) 1.00000 7.02E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
1.015498 0.000037 Normal (k=1) -0.965369 1.27E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9708 0.0033 Normal (k=2) 1.0098 2.78E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0236 0.0040 Normal (k=1) -0.0268 1.15E-08
Reflection Coefficient of STD Power sensor - ΓSTD
0.0343 0.0040 Normal (k=1) 0.0572 5.24E-08
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0300 0.0025 Normal (k=1) 0.0444 1.23E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-2.2904 0.1699 Normal (k=1) 0.0012 4.40E-08
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
2.6597 0.1166 Normal (k=1) 0.0005 3.04E-09
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
-2.8958 0.0833 Normal (k=1) 0.0017 2.02E-08
Measured Value 0.9803 Combined Uncertainty 0.0032
Expanded Uncertainty (k=2) 0.0064
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
27/44
Table 11. Calibration factor uncertainty budget form at 15 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.98624 0.00043 Normal (k=1) -0.99693 1.82E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
0.998880 0.000040 Normal (k=1) 0.984316 1.56E-09
Power of DUT Power Sensor-PDUT (mW)
0.99295 0.00043 Normal (k=1) 0.99020 1.83E-07
Reproducibility – R 0.00000 0.00275 Normal (k=1) 1.00000 7.56E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.986531 0.000056 Normal (k=1) -0.996638 3.08E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9659 0.0033 Normal (k=2) 1.0179 2.82E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0457 0.0040 Normal (k=1) -0.0415 2.79E-08
Reflection Coefficient of STD Power sensor - ΓSTD
0.0083 0.0040 Normal (k=1) 0.0573 5.25E-08
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0300 0.0050 Normal (k=1) -0.0473 5.60E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-0.2666 0.0880 Normal (k=1) -0.0019 2.85E-08
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
0.7685 0.4793 Normal (k=1) -0.0001 3.29E-09
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
-0.5233 0.1666 Normal (k=1) -0.0020 1.15E-07
Measured Value 0.9832 Combined Uncertainty 0.0033
Expanded Uncertainty (k=2) 0.0066
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
28/44
Table 12. Calibration factor uncertainty budget form at 18 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.98933 0.00043 Normal (k=1) -0.99580 1.85E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.024125 0.000038 Normal (k=1) 0.961975 1.31E-09
Power of DUT Power Sensor-PDUT (mW)
1.00128 0.00043 Normal (k=1) 0.98392 1.82E-07
Reproducibility – R 0.0000 0.0078 Normal (k=1) 1.0000 6.08E-05
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.992961 0.000051 Normal (k=1) -0.992166 2.56E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9464 0.0035 Normal (k=2) 1.0410 3.32E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0417 0.0040 Normal (k=1) 0.0878 1.24E-07
Reflection Coefficient of STD Power sensor - ΓSTD
0.0706 0.0041 Normal (k=1) -0.0901 1.33E-07
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0571 0.0050 Normal (k=1) -0.0473 5.58E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
2.3165 0.0962 Normal (k=1) -0.0029 7.92E-08
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
0.9993 0.0573 Normal (k=1) -0.0047 7.35E-08
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
1.5007 0.0875 Normal (k=1) -0.0077 4.48E-07
Measured Value 0.9852 Combined Uncertainty 0.0081
Expanded Uncertainty (k=2) 0.0162
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
29/44
7.2. Calibration Factor Measurements Uncertainty Budget at Second TÜBİTAK UME
Measurement
TÜBİTAK UME measured the travelling standard in 3 orientations. The measured values of third
orientation are given in Table 13-20. Uncertainty of the measurement contains the reproducibility
error of three orientations.
Model function :
STDGESTDGEGESTD
DUTGEDUTGEGEDUT
D
S
STD
DUTSTDDUT
Cos
Cos
Pm
Pm
P
PCFCF
21
2122
22
+R
Table 13. Calibration factor uncertainty budget form at 10 MHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
0.99807 0.00043 Normal (k=1) -0.98433 1.81E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.00148 0.00021 Normal (k=1) 0.98098 4.14E-08
Power of DUT Power Sensor-PDUT (mW)
1.00039 0.00043 Normal (k=1) 0.98205 1.81E-07
Reproducibility – R 0.0000 0.0029 Normal (k=1) 1.0000 8.41E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
1.00541 0.00016 Normal (k=1) -0.97715 2.38E-08
Calibration Factor of the STD Power Sensor - CFSTD
0.9840 0.0038 Normal (k=2) 0.9984 3.60E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.1064 0.0021 Normal (k=1) -0.0004 6.57E-13
Reflection Coefficient of STD Power sensor - ΓSTD
0.0588 0.0021 Normal (k=1) 0.0008 2.61E-12
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0017 0.0025 Normal (k=1) 0.0031 6.13E-11
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-1.38556 0.01973 Normal (k=1) 0.00035 4.72E-11
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-1.50954 0.03486 Normal (k=1) -0.00019 4.30E-11
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.83871 1.49137 Normal (k=1) 0.00016 5.70E-08
Measured Value 0.9824 Combined Uncertainty 0.0035
Expanded Uncertainty (k=2) 0.0071
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
30/44
Table 14. Calibration factor uncertainty budget form at 50 MHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
1.00303 0.00043 Normal (k=1) -0.99598 1.87E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.00011 0.00016 Normal (k=1) 0.99889 2.64E-08
Power of DUT Power Sensor-PDUT (mW)
1.00433 0.00044 Normal (k=1) 0.99469 1.87E-07
Reproducibility – R 0.0000 0.0028 Normal (k=1) 1.0000 7.56E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
1.00210 0.000066 Normal (k=1) -0.99690 4.28E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9997 0.0033 Normal (k=2) 0.9993 2.72E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0234 0.0020 Normal (k=1) -0.0026 2.75E-11
Reflection Coefficient of STD Power sensor - ΓSTD
0.0129 0.0020 Normal (k=1) 0.0046 8.44E-11
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0026 0.0025 Normal (k=1) -0.0008 3.79E-12
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-1.212526 0.085526 Normal (k=1) 0.000103 7.73E-11
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-1.789465 0.154950 Normal (k=1) -0.000029 2.03E-11
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.245537 0.976086 Normal (k=1) 0.000074 5.17E-09
Measured Value 0.9990 Combined Uncertainty 0.0033
Expanded Uncertainty (k=2) 0.0065
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
31/44
Table 15. Calibration factor uncertainty budget form at 1 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
1.01231 0.00044 Normal (k=1) -0.98022 1.85E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.00253 0.00016 Normal (k=1) 0.98979 2.58E-08
Power of DUT Power Sensor-PDUT (mW)
1.00383 0.00044 Normal (k=1) 0.98850 1.85E-07
Reproducibility – R 0.0000 0.0026 Normal (k=1) 1.0000 6.76E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.99278 0.000102 Normal (k=1) -0.99950 1.03E-08
Calibration Factor of the STD Power Sensor - CFSTD
0.9914 0.0033 Normal (k=2) 1.0009 2.73E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0084 0.0020 Normal (k=1) -0.0153 9.42E-10
Reflection Coefficient of STD Power sensor - ΓSTD
0.0231 0.0020 Normal (k=1) -0.0139 7.68E-10
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0077 0.0025 Normal (k=1) -0.0579 2.10E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-0.7008544 0.2388411 Normal (k=1) 0.0000003 4.40E-15
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
2.88295 0.08670 Normal (k=1) 0.00015 1.74E-10
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
0.70302 0.32326 Normal (k=1) 0.00015 2.43E-09
Measured Value 0.9923 Combined Uncertainty 0.0031
Expanded Uncertainty (k=2) 0.0063
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
32/44
Table 16. Calibration factor uncertainty budget form at 4 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
1.01859 0.00044 Normal (k=1) -0.963003 1.81E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.006527 0.00012 Normal (k=1) 0.974541 1.42E-08
Power of DUT Power Sensor-PDUT (mW)
1.00171 0.00044 Normal (k=1) 0.979228 1.82E-07
Reproducibility – R 0.0000 0.0026 Normal (k=1) 1.0000 6.76E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.993471 0.000085 Normal (k=1) -0.987349 7.11E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9838 0.0033 Normal (k=2) 0.9971 2.71E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0085 0.0021 Normal (k=1) 0.0074 2.29E-10
Reflection Coefficient of STD Power sensor - ΓSTD
0.0512 0.0020 Normal (k=1) 0.0122 5.98E-10
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0066 0.0025 Normal (k=1) 0.1045 6.83E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
2.008134 0.241176 Normal (k=1) -0.000090 4.72E-10
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-2.425336 0.039081 Normal (k=1) 0.000214 7.01E-11
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.095592 0.379573 Normal (k=1) 0.000124 2.22E-09
Measured Value 0.9809 Combined Uncertainty 0.0031
Expanded Uncertainty (k=2) 0.0063
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
33/44
Table 17. Calibration factor uncertainty budget form at 8 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
1.02738 0.00045 Normal (k=1) -0.95305 1.80E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.01100 0.00018 Normal (k=1) 0.96849 2.93E-08
Power of DUT Power Sensor-PDUT (mW)
1.01470 0.00044 Normal (k=1) 0.96496 1.81E-07
Reproducibility – R 0.0000 0.0026 Normal (k=1) 1.0000 6.76E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.99716 0.000114 Normal (k=1) -0.98194 1.26E-08
Calibration Factor of the STD Power Sensor - CFSTD
0.9762 0.0033 Normal (k=2) 1.0030 2.74E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0135 0.0022 Normal (k=1) -0.0440 8.96E-09
Reflection Coefficient of STD Power sensor - ΓSTD
0.0551 0.0020 Normal (k=1) 0.0402 6.46E-09
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0244 0.0025 Normal (k=1) 0.0664 2.75E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-1.94169 0.15903 Normal (k=1) 0.00025 1.60E-09
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
-2.91466 0.03629 Normal (k=1) 0.00143 2.70E-09
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
2.34170 0.10247 Normal (k=1) 0.00168 2.97E-08
Measured Value 0.9791 Combined Uncertainty 0.0032
Expanded Uncertainty (k=2) 0.0063
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
34/44
Table 18. Calibration factor uncertainty budget form at 12 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
1.03940 0.00045 Normal (k=1) -0.94362 1.81E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.01581 0.00013 Normal (k=1) 0.96553 1.50E-08
Power of DUT Power Sensor-PDUT (mW)
1.03555 0.00045 Normal (k=1) 0.94713 1.81E-07
Reproducibility – R 0.00000 0.00265 Normal (k=1) 1.00000 7.02E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
1.00308 0.000098 Normal (k=1) -0.97779 9.11E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9708 0.0033 Normal (k=2) 1.0103 2.78E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0246 0.0041 Normal (k=1) -0.0262 1.13E-08
Reflection Coefficient of STD Power sensor - ΓSTD
0.0343 0.0040 Normal (k=1) 0.0573 5.25E-08
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0300 0.0025 Normal (k=1) 0.0440 1.21E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-2.27854 0.16456 Normal (k=1) 0.00130 4.56E-08
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
2.65969 0.11659 Normal (k=1) 0.00047 3.04E-09
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
-2.89581 0.08333 Normal (k=1) 0.00177 2.18E-08
Measured Value 0.9808 Combined Uncertainty 0.0032
Expanded Uncertainty (k=2) 0.0064
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
35/44
Table 19. Calibration factor uncertainty budget form at 15 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
1.04669 0.00046 Normal (k=1) -0.94084 1.84E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.01956 0.00018 Normal (k=1) 0.96588 3.19E-08
Power of DUT Power Sensor-PDUT (mW)
1.02080 0.00044 Normal (k=1) 0.96471 1.82E-07
Reproducibility – R 0.00000 0.00275 Normal (k=1) 1.00000 7.56E-06
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.97367 0.000101 Normal (k=1) -1.01140 1.04E-08
Calibration Factor of the STD Power Sensor - CFSTD
0.9659 0.0033 Normal (k=2) 1.0195 2.83E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0472 0.0041 Normal (k=1) -0.0446 3.27E-08
Reflection Coefficient of STD Power sensor - ΓSTD
0.0083 0.0040 Normal (k=1) 0.0574 5.27E-08
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0300 0.0050 Normal (k=1) -0.0542 7.34E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
-0.19199 0.08587 Normal (k=1) -0.00183 2.48E-08
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
0.76845 0.47933 Normal (k=1) -0.00012 3.30E-09
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
-0.52328 0.16657 Normal (k=1) -0.00195 1.06E-07
Measured Value 0.9848 Combined Uncertainty 0.0033
Expanded Uncertainty (k=2) 0.0067
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex A
36/44
Table 20. Calibration factor uncertainty budget form at 18 GHz
Definition
Expected Value
xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of
the Uncertainty Contribution
Power of STD Power Sensor-PSTD (mW)
1.01951 0.00044 Normal (k=1) -0.97747 1.87E-07
Power of MON Power Sensor @ STD Power Sensor is measuring-PmS (mW)
1.02367 0.00017 Normal (k=1) 0.97350 2.90E-08
Power of DUT Power Sensor-PDUT (mW)
1.00166 0.00043 Normal (k=1) 0.99489 1.87E-07
Reproducibility – R 0.0000 0.0063 Normal (k=1) 1.0000 3.97E-05
Power of MON Power Sensor @ DUT Power Sensor is measuring-PmD (mW)
0.95248 0.000070 Normal (k=1) -1.04626 5.34E-09
Calibration Factor of the STD Power Sensor - CFSTD
0.9464 0.0035 Normal (k=2) 1.0530 3.40E-06
Reflection Coefficient of DUT Power sensor - ΓDUT
0.0415 0.0041 Normal (k=1) 0.0883 1.31E-07
Reflection Coefficient of STD Power sensor - ΓSTD
0.0706 0.0041 Normal (k=1) -0.0911 1.36E-07
Equivalent Reflection Coefficient of Power Splitter - ΓGE
0.0571 0.0050 Normal (k=1) -0.0484 5.87E-08
Phase of Reflection Coefficient of DUT Power Sensor - ΘDUT (rad)
2.3238 0.0987 Normal (k=1) -0.0030 8.62E-08
Phase of Reflection Coefficient of STD Power Sensor - ΘSTD (rad)
0.9993 0.0573 Normal (k=1) -0.0048 7.52E-08
Phase of Equivalent Reflection Coefficient of Power Splitter - ΘGE (rad)
1.5007 0.0875 Normal (k=1) -0.0078 4.61E-07
Measured Value 0.9965 Combined Uncertainty 0.0067
Expanded Uncertainty (k=2) 0.0133
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
37/44
ANNEX B. COMPARISON MEASUREMENTS REPORT of SASO NMCC
1. PARTICIPANT INFORMATION
Laboratory Name RF and Microwave
Contact Person Ahmed Hamad AlJawan
Telephone No +966 11 252 9741
Fax No +966 11 207 6484
E-mail [email protected]
Address
Saudi Standards, Metrology and Quality Organisation
of The Kingdom of Saudi Arabia (SASO)
Riyadh 11471, P.O. Box 3437 Kingdom of Saudi
Arabia
2. MEASUREMENT DATE
26/06/2018- 19/07/2018
3. ENVIRONMENTAL CONDITION
Temperature : (23 ± 2) C
Relative Humidity :(45 ± 15) %rh
4. REFERENCES USED IN MEASUREMENT
Instrument Name
Manufacturer Type / Model Serial No Traceability
Vector Network Analyzer
Keysight N5225A MY51451320
SASO
2016/57/TI-MI 1C
SASO-2016/57/TI-MI 3C
Type N Calibration Kit
Keysight 85054B MY39200420 SASO
2016/57/TI-MI 3C
VNA Cables Keysight 85135-60002 Port1: 50396
Port2: 50397
SAS0
2016/57/TI-MI 3C
Signal Generator
Keysight E8257D MY53400800 SASO
2018-323-TF-TF-3 C
Power Meter Agilent N1914A MY54266033 UME
G1MD-0114 (14-09-2015)
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
38/44
Instrument Name
Manufacturer Type / Model Serial No Traceability
Power Meter Agilent N432A MY54406003 UME
2015-G1MD-0106 (14-09-2015)
Thermistor mount
Agilent 8478B MY55070005 UME
2015G1MD-0100 (04-09-2015)
Power splitter Agilent 11667A MY51357191 UME
2015-G1MD-0115 (16-09-2015)
Power sensor Agilent N8481A MY54220012 Keysight
Vector Network Analyzer
Keysight N5225A MY51451320
SASO
2016/57/TI-MI 1C
SASO-2016/57/TI-MI 3C
5. MEASUREMENT METHOD
Calibration factor measurement of the travelling standard was performed using a direct
comparison method at an applied RF power of 0 dBm. A standard thermistor mount and a
travelling standard were connected to one of the output ports of the power splitter while a monitor
power sensor was attached to the other output port of the power splitter. A microwave power was
applied to the input port of the power splitter while the standard thermistor mount (STD) is
connected to the splitter, then the output power of the splitter was measured using the monitor
power sensor and the STD thermistor mount. After the measurement finished, STD thermistor
mount was disconnected from the power splitter port and the travelling standard (DUT) was
connected immediately. Then, again, a microwave power was applied to the input port of the
splitter and the power at the two output ports was measured using monitor and DUT power
sensors.
The calibration factor of the STD thermistor mount and the measured power values; using
STD, DUT, and monitor power sensors; were used to calculate the calibration factor of the DUT
power sensor, after applying a mismatch correction. Reflection coefficient of the travelling
standard (power sensor) was measured using the Vector Network Analyzer setup after applying a
SOL (Short, Open, Load) calibration method. The declared final calibration factor values of the
travelling standard are relative to the calibration factor at a reference frequency of 50 MHz.
Ahmed ALJAWAN and Ibrahim ALBORAIH have performed the reported measurements.
The measurements and comparison report were done according to the technical protocol of the
bilateral comparison.
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
39/44
6. MEASURMENT RESULT
Measured pin depth: -0.0017 inches, U (k=2) =0.0012 inches
Table 1. Reflection coefficient
Frequency
(GHz) Real Part
Real
Uncertainty
(k=2)
Imaginary
Part
Imaginary
Uncertainty
(k=2)
0.01 0.0205 0.0063 -0.1032 0.0063
0.05 0.0085 0.0060 -0.0218 0.0060
1 0.0051 0.0060 -0.0069 0.0060
4 -0.0045 0.0060 0.0104 0.0060
8 0.0015 0.0060 -0.0154 0.0060
12 -0.0180 0.0140 -0.0246 0.0140
15 0.0509 0.0140 0.0025 0.0140
18 -0.0356 0.0140 0.0317 0.0140
Table 2. Calibration factor (reference to 50 MHz)
Frequency Calibration Factor Calibration Factor
Uncertainty (k=2)
10 MHz 0.9776 0.0083
50 MHz 1.0000 Reference
1 GHz 0.9976 0.0086
4 GHz 0.9853 0.0090
8 GHz 0.9814 0.0127
12 GHz 0.9858 0.0133
15 GHz 0.9884 0.0151
18 GHz 0.9966 0.0267
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
40/44
7. UNCERTAINTY BUDGET
7.1. Calibration Factor measurements uncertainty budget
Model function : CFDUT=CFSTDPDUT
PSTD
PmS
PmDMM MM=
MD
MS
Where MD is mismatch between DUT power sensor and power splitter and MS is mismatch
between STD Power sensor and power splitter.
Table 3. Uncertainty budget of CF at 10 MHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9877 0.0006 Normal (k=1) -0.9915 3.61E-07
Power for standard monitor sensor (PmS) (mW)
0.9766 0.0004 Normal (k=1) 1.0028 1.52E-07
Power for DUT sensor (PDUT) (mW)
0.9877 0.0029 Normal (k=1) 0.9915 8.13E-06
Power for DUT monitor sensor (PmD) (mW)
0.9830 0.0003 Normal (k=1) -0.9962 8.92E-08
Calibration factor of the standard sensor (CFSTD)
0.9858 0.0058 Normal (k=2) 0.9934 8.30E-06
Mismatch correction (MM)
1.000 0.0006 Normal (k=1) 0.9793 3.17E-07
Measured Value 0.9793 Combined Uncertainty 0.0042
Expanded Uncertainty (k=2) 0.0083
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
41/44
Table 4. Uncertainty budget of CF at 50 MHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9967 0.0007 Normal (k=1) -1.0050 4.58E-07
Power for standard monitor sensor (PmS) (mW)
0.9980 0.0003 Normal (k=1) 1.0037 1.11E-07
Power for DUT sensor (PDUT) (mW)
1.0038 0.0029 Normal (k=1) 0.9978 8.41E-06
Power for DUT monitor sensor (PmD) (mW)
1.0026 0.0003 Normal (k=1) -0.9991 7.64E-08
Calibration factor of the standard sensor (CFSTD)
0.9990 0.0059 Normal (k=2) 1.0027 8.75E-06
Mismatch correction (MM) 1.0000 0.0001 Normal (k=1) 1.0016 1.70E-08
Measured Value 1.0017 Combined Uncertainty 0.0042
Expanded Uncertainty (k=2) 0.0084
Table 5. Uncertainty budget of CF at 1 GHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9917 0.0007 Normal (k=1) -1.0076 5.57E-07
Power for standard monitor sensor (PmS) (mW)
0.9977 0.0004 Normal (k=1) 1.0015 1.81E-07
Power for DUT sensor (PDUT) (mW)
0.9948 0.0029 Normal (k=1) 1.0045 8.34E-06
Power for DUT monitor sensor (PmD) (mW)
0.9936 0.0003 Normal (k=1) -1.0057 6.75E-08
Calibration factor of the standard sensor (CFSTD)
0.9925 0.0061 Normal (k=2) 1.0068 9.43E-06
Mismatch correction (MM) 0.9996 0.0001 Normal (k=1) 0.9997 5.32E-09
Measured Value 0.9992 Combined Uncertainty 0.0043
Expanded Uncertainty (k=2) 0.0086
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
42/44
Table 6. Uncertainty budget of CF at 4 GHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9830 0.0013 Normal (k=1) -1.0041 1.70E-06
Power for standard monitor sensor (PmS) (mW)
0.9836 0.0006 Normal (k=1) 1.0035 4.02E-07
Power for DUT sensor (PDUT) (mW)
0.9778 0.0028 Normal (k=1) 1.0094 8.18E-06
Power for DUT monitor sensor (PmD) (mW)
0.9785 0.0005 Normal (k=1) -1.0087 2.27E-07
Calibration factor of the standard sensor (CFSTD)
0.9864 0.0062 Normal (k=2) 1.0006 9.62E-06
Mismatch correction (MM) 1.0007 0.0001 Normal (k=1) 0.9863 1.03E-08
Measured Value 0.9870 Combined Uncertainty 0.0045
Expanded Uncertainty (k=2) 0.0090
Table 7. Uncertainty budget of CF at 8 GHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9768 0.0012 Normal (k=1) -1.0064 1.55E-06
Power for standard monitor sensor (PmS) (mW)
0.9781 0.0011 Normal (k=1) 1.0051 1.25E-06
Power for DUT sensor (PDUT) (mW)
0.9757 0.0046 Normal (k=1) 1.0075 2.16E-05
Power for DUT monitor sensor (PmD) (mW)
0.9792 0.0010 Normal (k=1) -1.0040 1.09E-06
Calibration factor of the standard sensor (CFSTD)
0.9799 0.0076 Normal (k=2) 1.0032 1.45E-05
Mismatch correction (MM) 1.0055 0.0004 Normal (k=1) 0.9777 1.18E-07
Measured Value 0.9831 Combined Uncertainty 0.0063
Expanded Uncertainty (k=2) 0.0127
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
43/44
Table 8. Uncertainty budget of CF at 12 GHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9740 0.0015 Normal (k=1) -1.0138 2.18E-06
Power for standard monitor sensor (PmS) (mW)
0.9795 0.0010 Normal (k=1) 1.0082 9.93E-07
Power for DUT sensor (PDUT) (mW)
0.9716 0.0029 Normal (k=1) 1.0162 8.49E-06
Power for DUT monitor sensor (PmD) (mW)
0.9684 0.0003 Normal (k=1) -1.0197 1.20E-07
Calibration factor of the standard sensor (CFSTD)
0.9690 0.0110 Normal (k=2) 1.0190 3.14E-05
Mismatch correction (MM) 1.0100 0.0012 Normal (k=1) 0.9777 1.30E-06
Measured Value 0.9874 Combined Uncertainty 0.0067
Expanded Uncertainty (k=2) 0.0133
Table 9. Uncertainty budget of CF at 15 GHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9671 0.0017 Normal (k=1) -1.0237 2.90E-06
Power for standard monitor sensor (PmS) (mW)
0.9774 0.0014 Normal (k=1) 1.0129 1.95E-06
Power for DUT sensor (PDUT) (mW)
1.0061 0.0062 Normal (k=1) 0.9840 3.69E-05
Power for DUT monitor sensor (PmD) (mW)
1.0026 0.0010 Normal (k=1) -0.9875 9.19E-07
Calibration factor of the standard sensor (CFSTD)
0.9698 0.0072 Normal (k=2) 1.0209 1.35E-05
Mismatch correction (MM) 1.0065 0.0010 Normal (k=1) 0.9836 9.23E-07
Measured Value 0.9900 Combined Uncertainty 0.0076
Expanded Uncertainty (k=2) 0.0151
Comparison Report on Calibration Factor Measurement of Power Sensor GULFMET.EM.RF-S2
Annex B
44/44
Table 10. Uncertainty budget of CF at 18 GHz frequency
Definition Expected
Value xi
Standard Uncertainty
u(xi)
Distribution Function
Sensitivity Coefficient
ci
Partial Variance of the
Uncertainty Contribution
Power for standard sensor (PSTD) (mW)
0.9518 0.0012 Normal (k=1) -1.0489 1.59E-06
Power for standard monitor sensor (PmS) (mW)
0.9821 0.0013 Normal (k=1) 1.0164 1.80E-06
Power for DUT sensor (PDUT) (mW)
1.0156 0.0051 Normal (k=1) 0.9829 2.51E-05
Power for DUT monitor sensor (PmD) (mW)
1.0055 0.0006 Normal (k=1) -0.9928 3.57E-07
Calibration factor of the standard sensor (CFSTD)
0.9590 0.0230 Normal (k=2) 1.0409 1.43E-04
Mismatch correction (MM) 0.9987 0.0025 Normal (k=1) 0.9996 6.45E-06
Measured Value 0.9983 Combined Uncertainty 0.0134
Expanded Uncertainty (k=2) 0.0267