ITEM No. 1
TECHNICAL SPECIFICATIONS FOR ENDOSCOPY SET FOR LARGE AND SMALL ANIMAL
1. FLEXIBLE ENDOSCOPE WITH ACCESSORIES-Insertion tube outer diameter > 10.3
mm,instrument channel >2.7 mm, working length >275 cm,deflection of distal tip(up – down) 180/200°
– 100°,deflection of distal tip (left – right)120° - 120°, field of view 140° or more,depth of field 2 – 200
mm, CCD chip size 1/6”, resolution- 480,000 pixels or more, two programmable control keys on
handgrip & four way deflection locks.
Accompanied with accessories -biopsy forceps with double action jaws, grasping forceps with double
action alligator jaws, polypectomy snare , cytology brush, foreign body basket, Injection needle with
retractabletip,carrying case, ETO cap, leakage tester, caps for working channel, Irrigation tube,
instrument oil, bite protector ,cleaning brush, cleaning valve&video connecting cable.
2. URETHROCYSTOSCOPY FOR MALE DOG- Flexible scope with Instrument channel: 1.2 mm.
Direction of view: 0°, Angle of view: 90°, working length: 100 cm, Outer Diameter: 3.0 mm.Following
accessories should be included: grasping forcep, biopsy forcep, stone basket, carrying case , pressure
compensation cap , leakage tester & cleaning brush.
3. ARTHROSCOPE WITH ACCESSORIES- Wide Angle Forward- Oblique Telescope 30°, enlarged
view, diameter 4 mm, length 18 cm, auto-clavable. Fiber optic light transmission incorporated.
The system to include arthroscope sheath, Obturators (Blunt & Sharp),Inflow cannula, suction tube,
probe, hook & retractor, curette of size 1 & 00, elevator sharp, osteotome, Ferris smith rongeur,
rongeur with heavy handle & strong jaws, biopsy forcep with strong jaws, grasping forcepwiith ratchet,
micro fracturing chisel.
4. LIGHT SOURCE- Cold Light Fountain Power LED 175 with integrated high-performance LED and one
light outlet, power supply 110 - 240 VAC, 50/60 Hz including mains Cord &connecting cable. Fibre Optic
Light Cable, extremely heat resistant of length 300 cm
5. CAMERA SYSTEM- Digital, single-chip medical-video-camera compatible with all rigid and flexible
scopes. Image sensor of ½” CCD-chip, picture elements- 752 (H) x 582 (V) pixels per chip (PAL) ,
resolution minimum 450 lines (horizontal) , exposure control of 1/60 sec- 1/10000 sec, signal-to-noise
ratio: > 50 dB & min. sensitivity of 3 Lux (f =1.4), Lens with integrated zoom lens of f=25-50mm .
Instrument coupling device should be compatible with all rigid & flexible endoscopes with standard
eyepiece. Adaptable to an operating microscope by means of a special quick adapter, camera head
fully soak able for sterilization & also gas sterilizable and sterrad compatible. The camera to be
accompanied by 22” LCD monitor for display.
6. Documentation system for digital storage of still images and video sequences, with mains cord,
power supply: 100/240 VAC, 50/60 Hz. Consists of: DVI-D Connecting Cable of Length 2m , USB
Silicone Keyboard with Touchpad with USB Stick, 8GB. The system to be accompanied with
monitor for use with Documentation system (with VGA Connector) along with DVI to HDMI
cable.
Standards & Warranty
1. Manufacturer must be EC certified for Full & Production Quality assurance.
2. Manufacturer must be ISO -13485: 2003 & ISO -9001:2008 certified for quality standards.
3. Manufacturer should quote FDA, CE,UL or BIS approved product, minimum three years from this
ITB.
4. Product should have IEC safety standards where ever listed. All products should be IEC 601-1, CE
certified according to MDD.
5. Manufacture should have minimum 10 installations in Government Institutions with minimum2
installation in the state.
6. Must have a veterinary training centre in India.
7. Manufacturer should have its own full- fledged corporate office along with service facility.
8. All units must be covered under standard warranty for a period of 24 months from the date of
installation.
9. The core operating endoscope like telescopes, single chip camera, light source, and other units
should be from single manufacturer for system compatibility.
F.No.5-13/2016-17/P&S/MKS
ITEM N0.2
Specifications for Thermal Cycler
Fully functional thermal cycler with following features:
1. The system should have 96 well format with at least six separate peltier blocks to provide independent
temperature zones to run six different assays with different annealing temperature at the same time
2. The system should be capable to run upto six separate temperatures in the same plate with user’s
defined time
3. The system should be able to accommodate 0.2ml/0.1 ml PCR tubes as well as microplate
4. The system should support reaction volumes of at least 10 µl
5. The system should have ramp rate ≥3.50C with accuracy of ± 0.50C or less and provision of heating lid to
prevent sample evaporation.
6. The system should have capacity to store at least 500 programmes in the instruments with provision to
store programmes in memory stick also
7. The machine should be duly certified/authorized for PCR process and it should be ISO 9001 & CE
Certified.
8. The system should be supplied essentially with 2kVa online UPS with back-up of at least 1 h on full load
9. System should be supplied equipped with 1.5 ton (five star rating) split air-conditioning unit with
appropriate wall mounted stabilizer(140V to 240V) in order to maintain ambient temperature
10. The system should be operable at 220-240V
11. System should be supplied with at least two years warranty on all part and on-demand repair services
as and when required.
12. Indian user’s list for the quoted model along with complete contact details including telephone no. & e.
mail should be provided.
13. Technical features described above should exclusively be supported by authentic original illustrated
company catalogue of the instrument/equipment quoted. Photocopy or computer print-out of
catalogue will not be accepted.
14. Vendor’s business record is to be authenticated by producing Income Tax return for at least 3
consecutive previous years.
F.No.5-14/2016-17/P&S/AK
ITEM No. 3
Specifications for Portable ultrasound Machine
Equipment should be single man usable, ultra-portable, lightweight & shock resistant dedicated veterinary ultrasound scanner for diagnosis and research use on large and small animals.
• Equipment should have dedicated veterinary applications on reproduction & gynaecology via in-built real
veterinarian software.
• Equipment should have software features including automatic follicle measurement, back fat and muscle
measurements, multispecies gestation tables, recordable user presets & automatic / manual focal zone adjustment.
• Equipment should be housed in sturdy, long life and corrosion free metal / alloy body.
Equipment should be light weight and easy to carry weighing between 1.5kg to 2.0kg including weight of inbuilt, user interchangeable and user extractable rechargeable battery.
• Equipment should have working battery autonomy of minimum 6 hours with nil to full recharge time of not
more than 3.5 hours.
• Equipment should have provision to inter-change the battery by the user in simple way without using any
tools.
• Equipment should have minimum 6" TFT LCD monitor working on VGA Format with minimum 30cm image
display.
• Equipment should function via full touch screen operation for both left handed and right handed users.
• Equipment should have facility for dual display with ultrasound scanning output both on inbuilt equipment
screen and multi-dimensional view glasses providing either both or one display at a time.
• Equipment should have facility to supply power to multi-dimensional view glasses without need of external
power input source.
• Equipment should have totally waterproof front face and waterproof connector plugs for equipment safety
and durability especially during field use.
• Equipment should come with following pinless, waterproof and all weather probes:
Wideband sector probe with central frequency 5.0MHz, B Mode frequencies of 3.5MHz to 6.0MHz for
applications on abdominal and OB/GYN scanning on small animals especially dogs.
• Equipment should have facility to interchange probe without switching-off the scanner to prevent any
dynamic image loss while scanning with minimum 3 frequency choices for the connected wide band probe.
• Equipment should have boot up time of maximum 20 seconds for immediate use with multiple operating
modes viz. B mode, B/B mode (Dual B) & M mode.
• Equipment should have internal memory capacity of minimum 1000 ultrasound images along with facility to
store and transfer images to external data storage source via USB & Bluetooth and computer connectivity.
• Equipment should have onboard input & output options viz. video out for external screen display, DC out for
multidimensional view glasses, thermal printer, bluetooth connectivity and USB port facility.
• Equipment should have on-screen display for brightness adjustment, date/time/battery gauge, automatic
probe identification among others.
• Equipment should have individual user protocols with facility to export images, clips, reports, patient files to
USB.
• Equipment should have facility for patient file modification, easy patient search, export of patient files,
review of patient files on USB and list of previous reports for each patient.
• Equipment should have minimum 6 depth levels, minimum focus 5 levels, minimum 4 TGC levels, minimum
8 dynamic levels & multiple average of slow, slow medium, medium fast and fast display modes.
• Equipment should have alphanumeric touch keyboard with express keys for modifying user parameters to
adapt and get best image resolution.
• Equipment should have provision to save preset with provision of multiple presets all storable in the same
equipment.
• Equipment should have processing via digital beam-former, real veterinarian software, minimum 16 true
emission channels, multi-zone focusing & 256 shades of gray.
• Equipment should be software upgradeable with evolutive hardware & firmware upgrade (FPGA) free of
charge.
• Equipment should have B Mode imaging with minimum 5 selectable focus zone positions, focus points for
each probe, adjustable region of interest, full screen zoom in minimum 3 selectable steps.
• Equipment should have automatic focus, frequency according to depth and measurement of distance in
both live and frozen image.
• Equipment should have ellipse and trace with touch freeze button, cine-review for frame by frame, loop and
measurements calculations on images from clips and selectable grid of 10mm x 10mm division and body marks.
• Equipment should have multiple image storing formats viz. Jpeg (.jpeg), Bitmap (.bmp), Raw data format
(.imag), Clip storing formats Raw data format(.imag).
• Equipment should be backed by certifications complying CE Marked to Council Directive 93/42/EEC, ISO
13485 / ISO 14001 / ISO 9001 / EMAS / UL. All manufacturer certifications to be produced at the time of delivery/installation.
• Equipment should be supplied in very sturdy easy to carry case with inbuilt battery recharger, minimum
32GB memory stick, ultrasound gel, leather carrier bag including spacer and sun protector.
• Equipment should come with onsite post sale service back–up and free of charge facility to train the user on
equipment's working and use at the time of installation.
F.No.5-15/2016-17/P&S/
ITEM No. 4
Technical Specification of REFRIGERATED WATERBATH INCUBATOR SHAKER
Temperature range: 5°C to 50°C
Shaking Speed: 0-300 RPM
Double walled, stainless steel.
Shaking Tray made of Stainless Steel 1 No.
Speed controlled by speed regulator.
RPM display on indicator and temperature display.
The water bath should have capacity to hold atleast 9 lotus clamps for holding
500 ml flasks alongwith universal platform.
Power supply 220 V, single phase 50 hz AC
Temperature Accuracy: ± 0.5°C
Stabilizer : 3 KVA
Warranty: 2 Years
F.No.5-16/2016-17/P&S/
ITEM No. 5
Technical Specification of DIGITAL ANALYTICAL WEIGHING BALANCE
Weighing capacity of upto 220 gm
Readability : 0.1 mg
80mm diameter pan round weighing platform
3 door draft shield glass chamber
Atleast 8 inches height from pan to top of glass door.
Repeatability: 0.1mg.
Stabilization time 4s
Linearity deviation : 0.3mg
Internal calibration weight
Warranty: atleast 2 Years
F.No.5-17/2016-17/P&S/
ITEM No.6
SPECTROPHOTOMETER (UV-VIS)
1. Optical System: Double Beam
2. Light Source: Deuterium and Tungston or xenon flash lamp
3. Photometric Accuracy: ± 0.005 A at 1.0 A
4. Baseline Stability: ± 0.0003 A
5. Photometric Range: upto 3.5 A
6. Photometric Repeatability: + 0.001 A
7. Spectral Bandwidth: ≤ 1.8 nM
8. Wavelength Accuracy (nM): ± 1.0 nM
9. Wavelength Range (nM): ≤ 190 to ≥ 1100
10. Wavelength Repeatability (nM): ± 0.5 nM
11. Scanning Speed: >2400 nM/Min
12. Cell Holder: ≥ 4 no.
13. Readout Format: Quartz Cuvette
14. Control: Capability to function either as standalone or with PC control
15. Software: Suitable data analysis software
16. Interface: USB and parallel printer interface
17. Cuvette: Quartz (4 no.) to be provided with the system
18. Warranty: 2 years from the date of installation
19. Catalogue: Authentic original company catalogue
20. User List: List of 10 recent installations in India with valid contract no. , mail ID & vender’s business
record (annual return) with 3 years ITR OPTIONAL ACCESSORIES TO BE ESSENTIALLY QUOTED SEPARATELY:
1. Computer: Computer system complete with UPS(Branded computer with i7 processor, 4
GB RAM, 1 TB HDD, 19” LED monitor, DVD R/W, USB ports, Key Board optical mouse and
compatible window-license) with suitable Laser Printer.
2. Cuvettes: Quartz cuvette (Rate/Piece) for 0.5 ml sample volume. 3. Stablizer: 3 KVA branded voltage stabilizer
F.No.5-18/2016-17/P&S/
ITEM No.7
Technical Specification of Thermal cycler (PCR)
1. The system should have 96 well format with at least six separate peltier blocks to
provide independent temperature zones to run six different assays with different
annealing temperature at the same time
2. The system should be capable to run up to six separate temperatures in the same plate
with user’s defined time
3. The system should have ‘Standard’ and ‘Fast’ run modes in a single instrument and
should be able to accommodate 0.2ml/0.1 ml PCR tubes as well as microplate
4. The system should support reaction volumes of at least 10 µl
5. The system should have ramp rate ≥3.50C with accuracy of ± 0. 50C or less and
provision of heating lid to prevent sample evaporation.
6. The system should have capacity to store at least 500 programmes in the instruments
with provision to store programmes in memory stick also
7. The machine should be duly certified/authorized for PCR process
8. The system should be supplied essentially with 2kVa online UPS with back-up of at
least 1 h on full load
9. The system should be operable at 220-240V
10. It should be ISO 9001 & CE Certified.
11. System should be supplied with at least two years warranty on all part and on-
demand repair services as and when required.
12. Indian user’s list for the quoted models along with complete contact details including
telephone no. & e. mail should be provided.
13. Technical features described above should exclusively be supported by authentic
original illustrated company catalog of the instrument/equipment quoted. Photocopy
or computer print-out of catalogue will not be accepted.
14. Vendor’s business record is to be authenticated by producing Income Tax return for
at least 3 consecutive previous years.
Optional: To be essentially quoted separately:
1. System should be equipped with 1.5 ton (five star rating) split air –conditioning unit
with appropriate stabilizer in order to maintain ambient temperature.
F.No.5-19/2016-17/P&S/
ITEM No.8
UV TRANSILLUMINATOR
Specifications:
Wavelength: 254-315 nm
Filter size: 150 X 150mm or more
Higher filter Lifetime
Protection against UV Radiation by cover
Adjustable UV protective shield provision
Adjustable dual intensity selector provision
Lower amount of infrared light
Stainless steel frame
Electrical operation: 230V / 50-60 HZ
Warranty : 02 Year from the date of installation
Catalogue : Original brochure to be provide
User list: 10 recent installations with phone no. & E mail IDs.
F.No.5-20/2016-17/P&S/
ITEM No.9
HIGH SPEED-LOW TEMPERATURE CENTRIFUGE
Specifications:
Must have temperature range of -9°C to 40°C, which can be modifiable during the run operation.
Centrifuge must have a fast cool function where cooling process can be shortened.
Must have a standby cooling mode to enable cooling or heating even when not in operation.
Pre-cool function
Max. speed of Instrument 17,500rpm, max rcf of 30,130g
Standard 1.5/2ml rotor, but has option to accelerate and decelerate gently for sensitive samples using SOFT function.
Built – in condensation drain to eliminate water accumulation.
Time function, Automatic rotor identification, rotor imbalance detection features.
Roters:
Fixed angle aerosol tight rotor for 30 X 1.5 /2ml tubes.
Fixed angle rotor for 6x 15/50ml with adaptors for 15/50ml centrifuge tube & round bottom tubes.
A pack of 1.5ml, 5ml, 15ml, 50ml centrifuge tubes must be quoted with instrument. Warranty: 02 year from the date of installation. Stabilizer: 3 KVA stabilizer servo controlled Optional: 03 KVA online UPS with 01 Hour backup on full load (APC/Numeric/
Emerson with LCD Display)
ITEM No.10 ‘IN VIVO IMAGING SYSTEM FOR SMALL ANIMALS’
1. Should be an automatic system to image small animals (mice & rats).
2. The In-Vivo Optical imaging system should be licensed for biophotonic imaging for small animals such as bioluminescence, fluorescence (filters ranges up to
400nm to 900nm), chem-iluminescence and radioisotopic cerenkov studies for in-vivo and in-vitro use.
3. Complete system inclusive of light tight cabinet, CCD camera, excitation and
emission filters, sample stage, computer workstation, optional gas anesthesia system and other accessories such as animal isolation chamber, ECG monitoring system, startup kit, etc. should be quoted.
4. System should have min of 15 excitation and 7 emission filters. Both Excitation
and emission filter wheels should be motorized and software controlled
5. Fitted with LED 's for photographic imaging
6. Cabinet should be able to accommodate gas anesthesia manifold and its tubing.
Essential accessories/attachment should be quoted.
7. Data generated should be in absolute calibrated data according to the National Institute of Standards and Technology (NIST).
8. Stage movement should be software controlled for different levels of magnifications.
9. Stage temperature should be controllable between 20 deg. cel to 40 deg. cel.
10. Able to accommodate at least 3 nos. of mice at largest FOV
11. Back thinned, back illuminated cooled grade 1 CCD (with min.mega pixel 13µ)
12. Thermoelectrically cooled to atleast -80˚ C (absolute)
13. Quantum efficiency: >85% for 500-700nm range.
14. Data computational device with high end configuration(i7)
15. The system should be supplied with 5 kVA UPS with half an hour backup.
16. Image acquisition and analysis software suitable for fluorescence, bioluminescence and Cerenkov
studies should be licensed for in-vivo and in-vitro studies.
17. Software should be able to spectrally unmix multiple reporters (minimum 3-5) within same animal.
18. Gas anesthesia ports and 5 position manifold with gas flow controller should be supplied.
Complete working configuration inclusive of hardware/software, tubing, valves, should be
supplied (Quote optionally).
19. The company should have complete in house reagent & cell line support manufactured by the same
company required for in vivo experiments; pre optimized on the instrument for demonstration.
20. The company should have minimum of 5-10 installations of similar systems in India. Provide list
of current installations in India.
21. System should have in-built X-Ray module (or model) with the following specs:
• Energy range 10-40 kV
• Typical scan time of 10 sec or better
• Radiation Shielded Cabinet should meet the safety standards set by the U.S. FDA Center for
Devices and Radiological Health (21 CFR 1020.40)
• Radiation Leakage should be <0.1 mR/hr
22. Dynamic contrast enhancement module for optical anatomical images for location of major
internal organs (Quote optionally)
23. The system should come with 1 year standard warranty and additionally 2 years AMC.
24. Company’s authentication with income tax records for atleast three years.
25. Company should provide hands on training to the scientists and students post installation
F.No.5-28/2015-16/P&S/MKS
ITEM No. 11
Technical specifications for A HIGH-END 3D/4D COLOR DOPPLER VETERINARY
ULTRASOUND IMAGING SYSTEM 1. Should be high-end ultrasound system capable of performing Abdominal, Vascular, Cardiac, Small
Parts, and Musculoskeletal scanning in small and large animals.
2. The system should come with species specific (at least for cattle/buffalo, sheep, goat, dog and horse)
extensive calculation packages (atleast Urinary, Gynecology and Car-diac) and integrated 3D/4D
imaging and calculation package for veterinary use.
3. System should have following modes: Real time 2D, M-mode, Pulsed wave Doppler, Color flow
imaging, Power Doppler, Tissue Harmonic Imaging, Anatomical M-mode, Tissue Doppler, 3D and
4D ultrasound imaging with duplex and triplex mode simultaneity.
4. System should have dual live mode or equivalent and dual and/or quad split display
5. System should have minimum 15” High Resolution medical grade flat LCD/LED monitor.
6. System should have speckle reduction technology with noise and artifact reduction for improving
tissue clarity and margin definition.
7. System should have a dynamic range of at least 170dB
8. System should have at least 3 active transducer ports.
9. Multifrequency, multifocal transducers to be supplied with the machine: Convex Ar-ray 2-5MHz (±1
MHz), Linear Array 4-10 MHz (±1 MHz), Phased Array Cardiac 2-5 MHz (±0.5 MHz),
Pediatric Phased Array Cardiac 4-8 MHz (±1 MHz), Vol-ume Convex Transducer 2-6 MHz
(±1 MHz), Linear rectal Transducer 4-6 MHz (±1 MHz), Intracavitary Transducer 4-10 MHz
(±1 MHz).
10. System should have a dedicated port for ECG and should display the ECG simultane-ously with the
ultrasonogram on the screen
11. Biopsy attachment/ needle guide and biopsy needle guide software should be provid-ed with convex
array and linear array transducers.
12. System should have Cine for at least 256 frames and raw data processing with angle correction.
13. The system should have loop storage with active image.
14. System should have facility for pan and zoom on live and frozen image.
15. System should have one touch image optimization & automatic real-time Doppler tracing
16. The system should have Tissue Harmonic Imaging capability on all transducers. Tis-sue Harmonic
Imaging should be available in colour flow imaging, M-Mode, and 3D rendering modes.
17. The system should have tissue Doppler imaging capability with appropriate quantification package.
18. PW Doppler analysis in Auto & Manual mode, Variable sample volume, Selectable wall filter,
Angle correction
19. System should have the facility of Elastography (ARFI and SWEI) with quantification tools
including Strain measurement, shell measurement and Histogram measurement.
20. Inbuilt hard disk with capacity of minimum 500 GB
21. System should have inbuilt DVD/CD RW, minimum four USB 3.0 ports and LAN/ Ethernet port
22. System should be DICOM 3 Compatible with Life time DICOM Licence and up-gradable for
PACS.
23. System should come with Online UPS with at least 15 minutes backup for the whole system.
24. System should be CE or FDA (US) approved
25. The supplier should have installed at least 10 ultrasound machines of the same make in and around
Bareilly region.
26. Accessories to be supplied: B/W Thermal printer of Latest model (with CE or FDA mark), Latest
model colour Laser printer, ECG attachments, Free Software upgrade(s) during the period of
Warranty & AMC, Probe Protection covers / probe connector cover for all probes, External Hard
Disk of at least 1 TB capacity.
27. Warranty of 2 years for the machine should be included in the package.
28. Original Catalogue should be supplied to verify the facts.
F.No.5-31/2015-16/P&S/MKS
ITEM No. 12
HAEMO DIALYZER A.Main Unit
ed at least 10 inch color TFT display
-280-300mm Hg), Venous(-60-
450mmHg) & transmembrane(-60-500mmHg) pressure monitor Facility
Tubing Pump segment (4mm-10 mm)
Should have Volumetric Ultra filtration System, inbuilt on line dialysate fluid filter system for
dialysate delivery
during dialysis
ble detecting & level adjustment facility for venous chambers
sensor with alarm for blood loss < 0.5 ml/min at max dialysate flow
sion (0-10 ml/min),& bolus infusion (5 ml/min) facility
-
700 ml/min) temp (35-39 0 C) and ultra-filtration flow rate(0-4 L/Hr.
er filtration assembly
disinfectant programme
-invasive blood pressure monitoring facility.
management system.
-
up at least for half an hour.
ument should be US FDA approved or CE certified
oom temperature 16-35 0C, Humidity30-75% and training at user site(2units)
B. RO Plant compatible for the instrument
Original Catalogue should be submitted.
F.No.5-62/2015-16/P&S/AK
ITEM No. 13
ULTRASONOGRAPH ALONG WITH TRANSDUCERS
nal Tissue Enhanced Imaging.
articulated arm and best image quality.
iology
features, such as Compass M-Mode and real-time Tissue Velocity Mapping, abdominal, Reproductive
and Contrast Imaging etc.
with various high-density phased ar-ray, convex and linear transducers (up to 9MHz) for cardiac,
abdominal reproductive studies in canine.
capabilities with or without wireless technolo-gy options.
Electronic independent channels.
-mode, Anatomical M-mode, Color M-mode, PW and CW doppler, Steerable CW
doppler.
256 gray shades for sharp contrast resolutions
canine use
external printer unit should be supplied with suitable UPS with a minimum 30 minutes back-up time.
functional safety test along with the technical bid.
-2-37ed1: Medical Electrical Equipment - Part 2-37:
Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment.
F.No.5-49/2015-16/P&S/AM
ITEM No. 14
Technical specifications of cryogenic storage tank (Liquid Nitrogen Silo) :
S.NO. DESCRIPTION PARTICULARS
1. Tank Type Cryogenic Storage Tank
2. Model Vertical
3. Configuration Double walled/Cylindrical
4. Service Intended Liquid Nitrogen
5. Gross Volume Approx. 3300 Ltrs
6. Net Capacity 3000 Ltrs
7. Dimensions
Overall Height Appx. 3100 mm
Outside Diameter Appx. 1700 mm
8. Inner Vessel
Material of Construction SS304
Design & Mnfg. Code ASME Sec. VIII, Div. I
MAWP 28.50 psiG with Operating Pressure of
0.9 kg/cm2
Design Pressure, kg/cm2 3.23 (appx.)
Hydro. Test Pressure, kg/cm2 4.20 (appx.)
Third Party Inspection Agency Any TPI Agency approved by Chief
Controller of Explosive, Nagpur
9. Outer Vessel
Material of Construction Carbon Steel/IS2062 Gr. B
Design & Mnfg. Code CGA-341
MAWP,(External) kg/cm2 1.05
Safety Device Size 40 KF x 3
Safety Device Rating kg/cm2g 0.5
Safety Device Location Outer Shell
Third Party Inspection Agency Any TPI Agency approved by Chief
Controller of Explosive, Nagpur
10. Insulation System
Type of Insulation High Vacuum Multilayer Super
Insulation
11. Performance
Static Evaporation Loss Rate 0.80 % of Full Capacity per day
Stable Warm Vacuum at our Works Better than 5 x 10-3Torr
12. Pressure Building Coil
Tube & Flange Material SS304 / AL
Fin Material Aluminium
13. CCOE site approval & License As per norms of Chief Controller of
Explosive, Nagpur
Description of Technical Specifications
The Liquid Nitrogen Storage Tank shall be double walled cryogenic storage tank having
annular space insulated with high vacuum Multi-Layer Super Insulation (MLSI) with absorber
(vacuum getter) material for vacuum retention.
Design & Construction
i. Inner Vessel
Inner vessel shall be designed, fabricated, tested and inspected as per latest ASME Pressure Vessel
Code, Section VIII, Div. I with service temperature of –1960C suitable for storing cryogenic liquid.
Material of Construction: SS304
ii. Outer Vessel
The outer vessel or vacuum casing shall be designed as per CGA-341 code to withstand 15
psi g external pressure. Material of construction: Carbon steel of suitable grade
iii. Insulation
The annular space between inner vessel and outer vessel shall be insulated with Multilayer Super
Insulation (MLSI) and sealed under high vacuum of better than 5 Micron.
Suitable absorber material - Vacuum Getter shall be incorporated for vacuum retention.
iv. Piping
Suitable product piping shall be provided for product fill, product withdrawal and pressure
building with suitable relief devices and valves. The piping size shall be suitable and material of
construction for pipes shall be SS304. All valves shall be located at convenient height for easy
operation. All the Cryogenic Valves shall be of reputed make such as Herose/ Mack/
Rego/Bestobell.
v. Instrumentation
The tank shall be equipped with a pressure gauge and liquid level differential pressure gauge
of reputed make located at convenient height to read liquid level in the tank.
Safety
The Tank shall be provided with dual safety protection – 2 No. Safety Relief Valves and 1
No. Rupture Discs.
Cleaning
All wetted surfaces shall be thoroughly cleaned for cryogenic service.
Painting
All the external surface of outer vessel shall be thoroughly cleaned and immediately primer
coated. Subsequently the surface shall be painted with two coats of Polyurethane (PU) paint as per
specified shade. The outer surface shall also have painted LOGO and other information as per the
statutory authority’s requirement.
Inspection & Testing
I. Vacuum Leak tightness of inner, outer vessels and all joints to be tested with Helium
Mass Spectrometer Leak Detection system.
II. Hydro-test as per code for Inner Vessel.
III. Static Evaporation Loss Rate test / cold spot test
Statutory Compliance
The tank shall be designed and manufactured in compliance with Static and Mobile Pressure
Vessel Rules (SMPV Rules) of Chief Controller of Explosives (CCOE), Nagpur.
Layout approval and CCOE License as required shall be arranged by supplier.
Installation and Commissioning
Installation, testing and commissioning of the complete system shall be the responsibility of the
supplier. For the evaporation loss rate test to be carried out at site after installation of the storage
tank and distribution header network, the supplier shall arrange required quantity of liquid
nitrogen. After successful completion of the testing, remaining quantity of the liquid nitrogen
inside the tank shall be handed over to the indenter along with the tank. Necessary civil
foundation, fencing with gate etc. shall be provided by supplier.
Training
Suitable training to the operating person/s for safe operation and maintenance of the tank
shall be arranged at plant/site by the supplier.
Documentation
All documentation as per SMPV Rules shall be provided by supplier. Operation and
Maintenance Manual shall be provided.
Warranty: 24 months from the date of commissioning.
F.No.5-36/2015-16/P&S/MKS
ITEM No.15
FLUORESCENCE UPRIGHT RESEARCH MICROSCOPE WITH IMAGE
ANALYSIS SYSTEM OPTICAL SYSTEM- Infinity corrected optical system with 100x-1000x magnification.
EYEPIECE TUBE- Trinocular tube (inclination: 30º) with interpupillary distance:
45-75mm, with eyepiece 10X (F.O.V: 20mm or better).
NOSEPIECE- Reversed type quadruple nosepiece to accommodate 4 objectives at a time. STAGE- Rectangular refocusing type stage, with cross travel 76 (X) x50mm (Y), with specimen holder for 2 slides
at a time.
COARSE/ FINE FOCUSING- Fine: 0.2mm per rotation, coarse: 37mm per rotation (approx)
OBJECTIVES OBJECTIVES N.A W.D Plan Achromat 10X Phase 0.25 10.5mm.
Plan Apochromate 20X Phase 0.40 1.3mm.
Plan Apochromate 40X Phase 0.95 0.66mm, (SPRING-LOADER)
Plan Flour Phase 100X Oil 1.3-1.4 0.66MM, (SPRING-LOADER) CONDENSER- Six position universal condenser for bright field, phase contrast & dark field microscopy
application, NA 0.9
ILLUMINATION SYSTEM- 6v30 watt halogen or 3 watt LED uniform illumination or better X 2pcs
FLUORESCENE ATTACHMENT- With four position fluorescence filter turret with100w
Mercury illumination, with following band pass fluorescence filters .
˃ Filter block blue for FITC/ Acridine Orange/ Auramine (EX465-495,DM505 and EM 515-555nm);
> Filter block green for TRITC/ Rhodamine/ Proium iodide (EX540-565, DM555 and EM 605-655nm);
> Filter block UV for DAPI/Hoechst (EX5330-380, DM400 and EM 435-485nm); CAMERA: Digital color cooled CCD camera with 2/3” CCD chip, with approx. 5.00 to 10.00 million
pixel resolution. live display mode: (5m interlace mode -5.0 frames/sec.; 20 frame per/sec. with ROI & binning); with
peltier cooling 20 ͦc below ambient, binning modes: 2x2,4x4 & digital zoom option ; interval shooting : 5 sec.- 10hr
intervals for different microscopy techniques with fire wire / USB port for attaching camera onto desktop/ laptop
through single wire IMAGE ANALYSIS SOFTWARE: Features for image acquisition and device control up to three dimensional,
live image capture, time lapse imaging, manual measurement & object counting, large image report generator facility,
vector layer & multi dimensional file format etc. DATA PROCESSING UNIT: Branded PC with Intel Core i7 processor, CPU3.40 GHz, 4GB RAM, 500 GB
HDD, optical mouse, DVD writer, 18.5” TFT color monitor, multimedia kit, 1394B fire wire port & cable along with
ups & color inkjet printer.
Suitable UPS with at least 30 mins power back-up Warranty: Three years
F.No.5-26/2015-16/P&S/MKS
ITEM No. 16
VERTICAL ULTRA LOW TEMPERATURE HIGH EFFICIENCY DEEP FREEZER Capacity: 400-450 Liters to store upto approx. 200-250 boxes.
Operating temperature: –86°C at ambient temp of 32°C with 1°C increment.
Power Consumption: 8.4 KWHr/Day, 350 Watts with Pull down from Ambient to -86°C should be
within 5-6 hrs
Working voltage: 220V/50Hz
It should have fully programmable microprocessor controlled with membrane keypad and eye level
control panel.
System should have stainless steel interior and tough interior body.
Exterior body: Powder Coated exterior constructed on steel
System should achieve energy savings by 130mm-thick poly urethane foam.
Freezer should have five insulated doors giving access to shelves (minimum 4 shelves) which
should be adjustable in steps.
Freezers should have heated air vent and air filter.
Heavy duty lockable castors and lockable outer doors and lids.
Freezer must have battery back-up and PIN security lock for unauthorized tempering.
Audible and visible alarms for temperature, power failure, system failure, battery low etc. and it
should have remote alarm port for connec-tion to and auto dialler.
Freezer must use CFC-FREE, HCFC-FREE non-flammable refrigerants, and refrigeration system
must be energy efficient and hermetical-ly sealed cascade refrigeration system.
Compressor should be capable to run any voltage of normal varying ranges.
Freezer must have ISO 9001- safety requirements and IEC 61010 Electrical safety CE certified.
Freezer must have washable condenser filter indication which should keep fins free of dust.
Instruments with 10-15 Racks, 200-250 Boxes, preferably with carbon dioxide Backup system with
carbon dioxide cylinder
Equipment should be supplied with a compatible 3KVA Voltage stabilizer to be provided with the
equipment with time delay facility to restart after power failure.
Two years warranty on instrument and at least 5 years warranty on compressor after installation
Technical features described above should exclusively be supported by authentic original illustrated
company catalogue of the instru-ment/equipment quoted (Photocopy/Printout/coloured print will
not be accepted).
A list of 5-10 recent installations with valid phone numbers and mail IDs of users.
File No.5-23/2015-16/P&S/MKS
ITEM No. 17
Binocular ophthalmoscope specifications
1. Video Indirect Ophthalmoscope equipped with fully integrated compact light weight camera system with CCD4 mega pixel or more
2. Focus should be adjustable to variable distance atleastupto 800mm
3. It should have an automatic white balance adjustment light boost.
Illumination mirror can be moved independently of viewing assemblies, Spotsize, illumination light axis and observation light axis
4. Processors should be compatible with image capturing device and should be with necessary software for video capturing
5. Table for laptop and printer should be provided
6. A laptop with connecting cable should be provided
7. Colour laser printer with connecting cable to laptop should be provided.
8. Should be supplied a +20D lens, 5 nos of illumination lamp, large and smallscleral depressor and fundus chart.
9. Should be supplied a carry case.
10. Warranty should be atleast 2 years.
F.No.5-131/2014-15/P&S/MS
ITEM No. 18
TECHNICAL SPECIFICATIONS FOR GRADIENT THERMAL CYCLER
Durable, independently controllable dual blocks
Gradient blocks to accommodate 0.2 ml and /0.5 ml and minimum reaction volume should be 10
µL
Universal block to accommodate all kinds of PCR tubes
Should have minimum 12°C gradient range
Gradient temperature range from 30 – 99ºC
Gradient technology should ensure identical ramp rates across the rows in both gradient and
normal operation
Adjustable ramp rate in gradient step to meet critical amplification conditions
Block temperature control range must be ≤ 4°C to ≥ 99°C
Lid Temperature range: 37 - 110 °C and should have over-tightening/pressure protection system
Block Temperature Accuracy: ± 0.2°C or less
Block Homogeneity: ≤ ±0.3° C
Heating rate: ≥ 3 °C/s; Cooling rate: ≥ 2 °C/s
Heating and cooling of block must be through peltier technology having multiple elements
Should be capable of testing temperatures at Denaturation, Annealing & Extension steps
Should have larger display with latest technology screen
Status indicating run, remaining cycle and run time in larger font
Should have latest technology for temperature control and homogeneity
Should have Administrator and user login for enhanced security and file protection
Preprogrammed protocol templates for easy selection
Should have options for customized programming in few steps
USB ports for protocol and data transfer
Auto restart facility in case of power interruption occurs before the completion of reaction
Power save Standby function
Authentic product/company catalogue should be provided for proving technical specifications
5-10 recent installations with name, address and contact numbers of the users should be
provided
Two years complete warranty with 24X7 Service support on call basis
Should be provided with the 3KVA online UPS of standard company with 1hour power backup.
F.No.5-34/2015-16/P&S/MKS
ITEM No. 19
SEVOFLURANE VAPORIZER
1. Sevoflurane vaporizer compatible with Surgivet large animal anaesthesia
machine.
2. The vaporizer should have appropriate back bar for the Surgivet large
animal Anaesthesia Machine
3. It should be temperature, pressure and flow compensated agent specific, of
good quality
F.No.5-58/2015-16/P&S/Ash
ITEM No. 20
ISOFLURANE VAPORISER
1. Isoflurane vaporizer compatible with Surgivet large animal
anaesthesia machine.
2. The vaporizer should have appropriate back bar for the
Surgivet Anaesthesia Machine
3. It should be temperature, pressure and flow compensated
4. It should be agent specific, on spot installation and
demonstration
5. Atleast one year warranty.
F.No.5-57/2015-16/P&S/Ash
ITEM No. 21
TECHNICAL SPECIFICATION OF PHACOEMULSIFICATION APPARATUS/SYSTEM
INCLUDING A SCAN, HAND HELD SLIT LAMP WITH PACHYMETER
Pump-
Peristaltic Pump
Fluidics
i) Closed Fluidic System, No Internal Tubings
ii) Aspiration: 0-50 cc/mm
iii) Maximum vacuum of 500 mmHG
iv) Vacuum and Aspiration Control – Programmable Linear control Or Panel Control
v) Reflux: Gravity Fed; Eliminates Contamination
vi) Venting: Fluid Venting System
Ultrasound:
1. Micro pulse Technology available in both mode- continuous Mode of Ultra sound and
within Pulse mode of
2. Ultrasound power delivery system.
3. Auto Tuning Mode
4. Ultrasound Frequency 40 KHz
5. Handpiece should have ability to drive cold phaco
Different type of pulses are available like short pulses, long pulses, power pulses with
adjustable no of pulse
Micro Pulse (Max 100 Pulses adjustable) in both modes continuous mode and pulse mode of
ultrasound power delivery system.
Programmable Occlusion Mode
Flow, vacuum and ultrasound simultaneously and automatically change as phaco tip sense the
vacuum at phaco tip and machine automatically control Occlusion building and breaking
process.
Tubing Pack Options
- Both reusable & disposable tubing options should be available.
Anterior Vitrectomy
TAC Vitrectomy probe with max cutting 50-700cuts/min, Re-usable, Oscillating,
Autoclavable.
Phaco Handpiece
Digitized lightweight phaco handpiece – 4 crystal
DIATHERMY SPECIFICATIONS
Power Adjustment: 5 to 100%, in 5% increments
Diathermy Power: 10 Watts into 100-Ohms
Frequency: 386 KHz
Diathermy Type: Bipolar
Power Delivery: Linear, Burst, or Panel
Display
Should have LCDtouch screen panel
Memory
6(add) Surgeons setting can be stored
Foot-Pedal
Programmable Detents and side switches.
Feedback: Vibration Detent
OPHTHALMIC ULTRASOUND - A SCAN
TECHNICAL SPECIFICATIONS
A Scan
Probe: 10 MHz focused transducer with internal fixation light assures proper alignment along
visual axis.
LCD display.
Measurements: Axial length, anterior chamber depth, lens thickness, vitreous
Measurement modes Immersion, contact
Clinical accuracy 0.1 mm
IOL calculation formulas incorporated: SRK, SRK/T, Holladay, Hoffer Q,Haigis
Manual Gain control.
Both Manual and Automatic measuring modes.
HAND HELD SLIT LAMP
TECHNICAL SPECIFICATIONS
1. Should be converging binocular microscope type
2. Should have at least 2 step magnification (10X and 16X).
3. Should have at least two 10X and 16X eyepiece.
4. Should have work distance 60- 80 mm
5. Should have Diopter adjustment range ±5D to ±7D
6. Should have continuously adjustable slit width 0-10mm
7. Should have slit inclination/angle with Horizontal circumference ±30º
8. Should have Green, cobalt blue, color temperature transformation angle.
9. Should be supplied with motorized table.
10. Should have good illumination using White LED
11. Should be battery operated
Pachymeter (To be quoted separately)
1. Should have automatic compensation capability, accurate measurement of central and
peripheral corneal thickness.
2. Should have probe of 15-20 MHz with angle of 40-45 degree.
3. Should have resolution of 3-5µm and measuring range 100 to 1500 µm.
4. Should display ultrasound waveform
F.No.5-21/2015-16/P&S/BS
ITEM No. 22
Operating Microscope Ophthalmic for both anterior and posterior segment
1. Should have apochromatic optics with 45 degree binocular and converging optics
2. Should have working distance of objective lens F = 200mmand 5 step magnification
3. Eye piece should be minimum 10x or 12.5x wide
4. Should have XY coupling
5. Should have red flex switching in/out facility
6. Should have total magnification from at least 4.5x to 22x
7. Should have field of view from at least 15mm to 50mm
8. Should have cold light coaxial illumination by fiber optic light guide
9. Should have tools free design for stand-by bulb change over and for failed bulb replacement.
10. Should have heat absorbing and UV filters.
11. Should be floor standing type with fiber wheels with brake.
12. Should have a minimum vertical stroke of 400mm
13. Should have rust free stand and paddles.
14. Should be operated in 200-240 Vac 50/60 Hz input supply.
15. Should provide sterilization cap and Dust Cover.
16. Warranty –Atleast for two years.
17. Power Backup- to be supplied with online UPS (1 KVA) with atleast ½ hour power backup on
full load.
F.No.5-19/2015-16/P&S/BS
ITEM No. 23
Specifications for CO2 incubator with CO2 cylinder with compatible voltage stabilizer
1. Should be a stackable, direct heat or air-jacketed CO2 incubator
2. Capacity: 150-200 ltrs
3. Temperature range: 3- -
4. CO2 control range from 0-0.2 to 20% with accuracy of ±0.1%.
5. At least 4 adjustable height shelves and a removable humidity pan.
6. HEPA or ULPA filter on CO2 inlet and N2 inlet
7. Alarm systems or sensors to monitor the CO2 level, temperature etc.
8. Should have IR sensor for CO2 regulation
9. Should have inner glass door
10. Should have sterilization facility, air-flow 6-8 cfm.
11. Should have LCD or LED display of the parameters.
12. Should have ISO 9001 and CE certifications
13. User list with contact details for at least last 10 installations in India must be enclosed.
14. Vendor must submit the original brochure for verification of specifications conformity of
quoted
equipment in tender. Photocopy or computer print-out of catalogue will not be accepted.
15. Should be supplied with 2 each of CO2 cylinders (approx. 40 kg or more) and regulators.
16. Free installation, 3 years warranty and on-site repair services as and when required
17. Should be supplied with 5 KVA branded online UPS with 1 hr back up on full load along
with installation.
18. Should be supplied with a 2-ton air conditioner of 4/ 5 star with digital display, along with
a suitable branded voltage stabilizer (wall mounted) and installation, to ensure proper
functioning of equipment in hot summer days
19. AMC/CMC may be covered for 2 years post-warranty period.
F.No.5-96/2015-16/P&S/BS
ITEM No. 24
Cryocans for storage
1. LN2 capacity approx.:33.0 lit.
2. weight of empty container:15kg
3. Outer diameter: 460mm
4. Neck diameter: 51mm
5. Static holding time: 280 days
6. Canisters OD: 38
7. Canisters height: 120mm
8. Number of canisters: 6 with capacity of 700 straws
Cryocans for transport
1. LN2 Capacity approx.:51.5 lit.
2. weight of empty container:15kg
3. Outer diameter: 460mm
4. Neck diameter: 51mm
5. Static holding time: 120 days
Portable Cryocans
1. LN2 Capacity approx.:33.0 lit.
2. weight of empty container:3kg
3. Outer diameter: 300mm
4. Neck diameter: 51mm
5. Static holding time: 30 days
6. Canisters OD: 38
7. Canisters height: 120mm
8. Number of canisters: 6 with capacity of 700 straws
F.No.5-100/2015-16/P&S/BS
ITEM No. 25
Integrated Semen straw filling-sealing and printing machine (0.25 ml) 1. Filling & sealing capacity – 12000-15000 straws per hour, four straws at a
time. 2. Hopper capacity -2000-4000 straws 3. Have liquid level sensor to automatically stop the machine when semen is
finished in the reservoir. 4. Facility to automatically switch-on completely filled straws instead of 3-4 half
filled straws at the end of filling to avoid semen wastage in hoses and half-filled straws.
5. In case of malfunction of machine, it should stop automatically 6. Standardization - ISO certified 7. MS Window compatible dedicated software. 8. Work as a filling & sealing machine only without changing any hardware 9. Ultrasonic sealing facility for the straws. 10. Suitable inkjet printer for permanent & legible printing 11. Automatic bull identification using Bar code. 12. Barcode printing with nozzle size : 40-50 micron, should have facility of 128
condensed bar code. 13. Straw printing should have facility for Barcode, date, bull no., year, batch,
place etc. Also free texts edit line. 14. Work as a printing machine only without changing any hardware. 15. Software for integrated data base for bull, file, administration for every
breed. 16. Printer with efficient permanent viscosity checks facility. 17. Nozzle & other parts of the printer should not be chocked by ink due to
unused period of printing. 18. Printer with efficient automatic nozzle-close system, permanent viscosity
checks, 24 hour ready to use. 19. System software can communicate & regulate ultrasonic generator vacuum
pump, printer & filling- sealing machine with a single command of PC attached with machine.
20. Easy operation & maintenance 21. Easy to clean 22. All parts that require cleaning should be easily handled without tools. 23. Compact, user friendly & ergonomic. 24. Low noise. 25. Equipment must be quoted with all additional software. 26. Warranty: at least two years. Amount ( in Rs.) require for AMC for further
next five years.
27. Free onsite training & installation 28. Satisfactory working report from users (India & abroad) to be attached 29. Supplier should have their service centre in India to provide after-sale
services. 30. Accessory items: One split AC (2 tons) and Online UPS (5 KVA) with at least
one hour power back up on full load for running this instrument. F.No.5-113/2014-15/P&S/BS
ITEM No. 26 MICRO DIGESTION SYSTEM
Automatic auto sequencing twelve place PC compatible regular series micro
block digestion system with auto lifting facility, suitable for acid digestion of plants, soils, water, food and feed samples with thermo control, regulator. Attached with acid neutralizer which should be fully automatic, four stage with micro intelligent microprocessor.
F.No.5-38/2015-16/P&S/MKS
ITEM No. 27 Multimode Reader
Fully functional programmable absorbance, fluorescence, luminescence and Time Resolved Fluorescence
based microplate reader which supports end point, kinetic and spectral scanning assays. The technology
should be based on monochromator or filter based or hybrid.
1. The system should be capable to provide readings with internal machine software and on-board display on
screen
2. It should also be compatible for PC connectivity with external analysis software control. The analysis
software must be provided essentially.
3. Detection modes - Fluorescence, UV-Visible Absorbance, luminescence, AlphaLISA/Alpha screene
4. Absorbance: Light source - Xenon flash lamp, Wavelength range: 250-999 nm
5. Fluorescence: Wavelength range 270-700 nm, top/bottom reading
6. Filter based systems should provide at least 10 filters to meet entire range of standard fluorescence dyes
like DAPI, Hoechst, FITC, GFP, TRITC, RFP, Rhodamine, PI etc.
7. Temperature controlled microplate chamber for maintaining temperature of 40oC and above
8. Linear shaking/mixing feature for microplate platform
9. Compatible online UPS to provide atleast one hour power backup
10. Branded desktop PC with latest configuration available in the market to install the analysis software and
one black laser printer to print the result reports must be provided essentially with system
11. Warranty: At least two years from the date of installation
12. Original brochure/catalogue with technical details and list of users must be provided with the bid.
F.No.5-88/2015-16/P&S/Ash
ITEM No. 28 ICP Spectrophotometer (ICP-OES)
Specifications
Fully automatic complete functional ICP-OES (Inductively Coupled Plasma-Optical Emission
Spectrometer) system using a polychromator with all essential accessories needed to run the system
(able to analyze all kind of mineral elements including hydride forming elements in biological samples)
with following specifications
Measurement mode Simultaneous
RF generator 27 Mhz or higher
Gas control PC controlled Auto Mass flow controller for all the gas flow
Peristaltic pump with three or more channels
Plasma viewing Dual view
Resolution 0.007-0.009 nm or better
Plasma Torch Demountable/ Semi-demountable
Wavelength range About 160-770 nm or wider
Plasma Tail management By air blower or other suitable device compatible with the system
Optical system Echelle optics
Detector CCD/CMOS/CID
Sample introduction/handling kit Aqueous sample introduction/ handling kit
Safety interlocks: system should be able to constantly monitor water flow, shear gas pressure (if applicable), argon
pressures, sample or compartment door closure and plasma stability, and display the interlock status on the
computer screen. On interruption of interlock, the plasma should immediately shut down.
To be fitted/installed with line protection system and working demonstrated with at least three trainings onsite
Any civil and electrical fittings/changes needed for the installation of the equipment will be responsibility of the
supplier
Essential Accessories
1. Compatible system suitable for simultaneous analysis of hydride forming elements
2. Argon gas cylinders with regulators: 3
3. Self standing fume hood made of non-corrosive metal with provision of duct to outside of lab (As per installation
requirement)
5. Compatible UPS (10KVA online) of ISI marked with at least 2 hrs. backup
6. Pure standards- single/multi-element standard for at least 20-25 elements (including Ca, K, Mg, Na, S, P, Mo, As,
Cd, Cr, Co, Cu, Fe, Mn, Zn, Se, Pb, B, V, Si, Ni, Ru, Ag, Li) with a certificate of analysis for quality, stability and
reliability; Pack size 100 ml, Concentration of 500 ppm or higher
7. Compatible window based software with data recorder and analyzer computer (latest system available at the
time of supply) with laser printer(normal)
Accessories (Rates to be quoted essentially separately for each item)
1. AC Unit (2Tons) split: 1 with compatible voltage wall mounted stabilizer (Range 140-240 V)
2. Argon humidifier, cooling system (if any required) compatible to the instrument
3. Spares: Nebulizers/ spray chambers, injectors, torches designed for quick and easy replacement, tubing for
sample intake and drainage for next two years considering a load about 200 samples per month
4. Auto-sampler (Able to handle 80-100 samples) with 1000 tube
5. Microwave sample digestion system- precise control with high power heating and high pressure capability along
with built in cooling
6. Sample block digester system- acid resistant, Teflon coated blocks with temperature uniformity
Other Essential Requirements:
Original brochure highlighting the desired features/specifications for each instrument component and accessories like ICP-OES, microwave sample digestion system, fume hood, auto sampler, sample block digester system etc.
Address, e-mail id and phone numbers of ICP-OES users (preferably of the quoted mode supplied by your firm) with their performance reports/ feed back
All system operable at 220±5% VAC with online protection system
Warranty: 3yrs for whole machine and its all parts including accessories
Rates of AMC for next three years after warranty period
F.No.5-42/2014-15/P&S/BS
ITEM No. 29
Specifications for ELISA Microplate Reader
1) Should have provision for automatic calibration before each reading
2) Should be able to read flat, U or V bottom microplates
3) Should have motorized door for plate loading & retrieval
4) Should have fast & step modes
5) Should have self-diagnostic capabilities to detect any lamp burnout
6) Should be supplied with desktop computer for recording readings
7) Should have wavelength range of 400-750 nm
8) Should have built in variable speed plate shaking capability
9) Warranty:3 years