Upload
api-3714301
View
271
Download
2
Embed Size (px)
Citation preview
Seminar Report
on
MEASUREMENTS IN UNSTEADY FLOWS
Submitted in the partial fulfillment of the requirements for the degree of
Bachelor of Technology in
Aerospace Engineering
By
Jitesh Chauhan 99D01001 Ashish Gupta 99D01003
Moble Benedict 99D01011
Department of Aerospace Engineering
Indian Institute of Technology Bombay
12 November 2002
1
CHAPTER 1
THE DYNAMICS OF CONTINUOUS MEASUREMENT OF
UNSTEADY FLOW
1.1 DYNAMIC PROPERTIES OF SYSTEMS FOR THE CONTINUOUS
MEASUREMENT OF UNSTEADY FLOW
Systems for the continuous measurement of unsteady flow have now been quite
widespread. They are utilized both industrially and in experimental research. There are
two ways of continuously measuring unsteady flow, depending on the task in hand :
(a) By determining the mean value of unsteady flow over some time interval from
t1 to t2.
(b) By continuously measuring rapidly varying flow and determining the
instantaneous value of the flow at any moment from t1 to t2.
The second instance of measuring unsteady flow is the more general.
The devices used measure unsteady flow function according to extremely varied
principles (mechanical, electrical, magnetic, optical and thermal). In all the continuous
measuring systems to be examined the primary transducer emits an electrical signal
which is the subject to appropriate conversion in secondary transducers and conversion
elements.
Flowmeters which do not use pressure difference when metering may be divided
into 2 groups.
(a) Mass flowmeters. Mass flowmeters are instruments applying the Coriolis
principle and gyroscopic effect, and also those using the force of inertia of a moving and
rotating flow of material. These flowmeters directly measure the mass of the flow, their
readings being independent of the physical parameters of the material.
(b) Volumetric flowmeters. These include turbine, electromagnetic, ultrasonic
flowmeters, and also to some extent the thermo-anemometers and ionization flowmeters.
These flowmeters react to the volumetric flow. In order to use their readings to assess the
flow by weight, corrections must be made for the density of the material.
2
The presence of a link of a metering chain is due to the need to perform definite
functions:
(a) The conversion of the unsteady flow rate or flow velocity into a primary
electric signal.
(b) Amplification and conversion of the primary signal.
(c) The recording of the measuring signal.
The structural schemes of most flow meters designed to measure unsteady flow have
three links: a detecting device, a transducer amplifier and a recording instrument. In such
measuring systems the flow acts directly on the detecting element of the emitter.
Continuous metering of unsteady flow may be effected using systems in which
the following measuring principles are applied.
(a) Continuous creation of an electrical intermediate parameter in the flow, whose
variation determines the velocity of the flow. These systems include inductive
flowmeters having a constant magnetic field and ionization flowmeters with
constant ionization of flow.
(b) Continuous creation of a mechanical intermediate parameter in the flow,
whose magnitude is proportional to the mass of the flow. Systems utilizing the
inertial properties of the medium include various flowmeters with spinning rotors
: flowmeters measuring the Coriolis force, gyroscopic and turbo-flowmeters.
(c) Continuous creation of a thermal intermediate parameter in the flow,
variations in which determine the velocity of the flow – calorimetric flowmeters.
(d) Introduction of a heated body into the flow under measurement, whose
thermal balance varies continuously depending on the velocity of the flow –
thermo anemometers.
(e) Introduction into the stream of a body which continuously detects the dynamic
pressure of the medium in motion.
In the discrete measurement of unsteady flow there is a periodic creation of an
intermediate parameter. It could be electrical, mechanical or thermal.
The primary transducers in electrical systems for continuous flow measurements
convert the sensed parameters into electrical signals. The primary elements may be
divided, in terms of power, into generator and modulator elements. No energy from a
3
source of supply is applied to primary generator elements; they form the electrical signal
by deriving the appropriate amount of energy from the flow. Magnetic-turbine type
transmitters may serve as an example of these elements. Modular elements, as distinct
from the generator type, consume power from the sources of supply and accordingly
modulate the size of the voltage (or current) in proportion to the amount of flow. The
electrical turbine transmitter is an example of such elements.
The quality of an element in a continuous measuring system may be characterized
by the magnitude and interrelationship of the input and output signals. On a change in the
input magnitude in the element a transient occurs whose nature depends on the elements
dynamic properties. These are a consequence of the presence in the element of inertial
parts which store energy. After a lapse of a certain time interval an equilibrium relation
is established between the input and output magnitudes which does not vary with time.
This is called the steady state of the element. The relation of the output magnitude to the
input magnitude for the element’s steady working mode is called its static characteristic.
1.2 FREQENCY CHARACTERISTICS OF CONTINUOUS MEASURING
SYSTEMS FOR UNSTEADY FLOW
When determining the percentage errors created by linear measuring systems in
measuring complicated functions, the errors may be considered as the sum of elementary
errors which arise in the individual measurement of the harmonic components of the
frequency spectrum of a complicated signal. This becomes possible through the fact that
in linear elements and linear systems where several disturbing effects are simultaneously
applied, their joint effect equals the sum of effects caused by each of the disturbing
elements in isolation. The properties of such continuous measuring systems are
determined by the amplitude and phase distortions occurring in response to the sinusoidal
varying signals at different frequencies. In this case the amplitude-frequency and the
phase-frequency characteristics may be used to describe the dynamic properties of
continuous measuring systems.
The amplitude-frequency characteristic is the relation showing how the amplitude
of the forced oscillations at the link’s output varies with frequency, if constant amplitude
harmonic oscillations are applied to the input.
4
The phase-frequency characteristic is the relation showing the variation with the
frequency of the phase shift between the harmonic oscillations applied to input of the link
and those obtained at the output.
Frequency characteristic are essentially static in nature, since for each value of the
frequency they indicate relations obtaining during steady state operation. The amplitude-
frequency and phase-frequency characteristics show the manner in which a link reacts to
periodic disturbances of different frequencies. The elements of continuous measuring
systems possess certain selective properties, so that out of the entire frequency spectrum
of the variation of flow to be measured, they transmit without distortion only frequencies
located definite bands of frequency range.
1.3 ACCURACY OF MEASUREMENT OF UNSTEADY FLOW
Errors which occur in measuring unsteady flow may be divided into static and
dynamic. Static errors are those which arise during measuring of magnitudes which do
not vary in time. Errors occurring during measurement of magnitudes which do vary in
time are known as dynamic.
Errors may also be classified as those due to the method and those due to the
apparatus, though our opinion is that, when considering continuous measuring systems
for unsteady flow, it is more rational to classify errors into static and dynamic.
One of the most important questions arising in the measurement of unsteady flow
is that of the dynamic accuracy. In most continuous measuring systems the dynamic
accuracy is governed by what is called the “inertia factor” of the system, i.e. the ability of
a measuring system to react rapidly to all variations of flow. The question of dynamic
accuracy in such systems is important, since in many cases dynamic accuracy obtainable
is the principal selection criterion for a system required to measure rapidly varying
magnitudes.
In the measurement of a constant flow (not varying with time) a continuous
measuring system is completely characterized by a static amplification factor (or gain).
The factor is proportional to the tangent of the angle of slope of static characteristic
relative to the abscissa. In the case of measuring unsteady flow, the instantaneous
dynamic factor of the system is not constant in value, but varies with time. Its
5
instantaneous magnitude basically depends on the amplification factor of the individual
harmonics and their relative disposition at a given moment. Since the overall dynamic
amplification factor of the system has definite variation with time, the measuring signal
follows the variation of magnitude to be measured with certain deviations. These
variations in the amplification factor of the system with time also cause the occurrence of
dynamic error when measuring unsteady magnitudes.
A system can measure a varying magnitude without any dynamic error in the case
where its dynamic amplification factor constant. This is possible in the ideal case when
the variation to be measured falls within the distortion-free-pass-band of the system, and
furthermore when, for all frequencies within the pass-band of the system, constant and
equal amplification factors occur in the absence of phase shift. Since in overwhelming
majority of cases when measuring varying magnitudes these conditions cannot be entirely
satisfied, corresponding dynamic errors arise. The magnitude of dynamic error depends
both on the nature of the physical processes underlying the measuring system and on the
parameters of the components (resistance, inductance and capacitance) composing the
measuring system.
The magnitude of errors introduced by the primary element is accordingly
amplified. In the other event of errors being introduced by subsequent elements of the
system, these errors suffer no amplification as in the first case.
The following parameters exercise an indirect influence on the sensing element of
a flowmeter: temperature and pressure of the medium, external magnetic and electrical
fields, and vibrations. The most substantial effect on the reading of a flowmeter is
exercised by the parameters of the substance under measurement via the sensing element.
In the transmission of a measuring signal, noise is superimposed on it, it becomes
necessary to separate signal from noise. This is achieved by the means of filters.
1.4 CALIBRATING CONTINUOUS MEASURING SYSTEM FOR UNSTEADY
FLOW
In order to determine an actual variation of flow with time from the measured
signal obtained, the appropriate transient relations must be available. For this purpose
frequency characteristics may be used for continuous measuring systems which may be
6
considered linear over the measuring range. The use of a statically calibrated relation to
determine the flow quantitatively is possible only when the amplitude-frequency and
phase- frequency characteristics of the system do not vary throughout the measured
frequency range, beginning with static mode. Thus in the general case the dynamic
calibration of the linear continuous measuring systems for unsteady flow boils down to
determining experimentally their frequency characteristics. For this purpose flow pulses
are formed on special installations, the pulses having a definite regular variation in time.
The principle of such a device is shown in Fig.7. A rotary valve is set in motion
by an electric motor having a variable rotation speed. The signal which controls the
electric motor is formed by a special electronic signal generator. The measuring signal
obtained from the flowmeter passes to an oscilloscope, to which is simultaneously
applied an electrical signal corresponding to the true variation of the flow. With such a
device one can determine the frequency characteristics of flowmeters and over the range
of frequencies of pulsating flow which can be generated by means of the rotating valve.
Thus the amplitude and phase distortions can be evaluated from one oscillogram on
which is recorded the reaction of flowmeter to a pulsating flow of increasing frequency
7
CHAPTER 2
CONTINUOUS MEASURING SYSTEMS FOR MASS FLOW The principle of action is based on imparting to the flow material to be measured
an additional movement, in order to excite in it intermediate measuring parameters. The
possibility of creating intermediate parameters in a given case is due to the tendency of a
flow of material (owing to its inherent inertia) to maintain a motion imparted to it.
Depending on the actual additional motion communicated to the flow (by a rotating or
oscillating unit) a Coriolis force, gyroscopic effect or an inertial torque is set up in the
sensitive element, proportional to the mass of flow of the substance. If a vibrating
element is used instead of a rotating unit, the working principle of the instrument is
unchanged.
The most important feature of this group of flowmeters is that they ensure the
direct measuring of the true value of mass flow in a non-steady flow of the substance,
regardless of the properties and the state of the substance (pressure, temperature, density,
viscosity, etc.). A desirable quality of these flowmeters is that they cause only small
pressure losses.
The group under consideration comprises:
(1) continuous measuring systems for mass flow using Coriolis force;
(2) those using gyroscopic couple;
(3) those using inertial moment of a rotating flow.
2.1 SYSTEMS FOR CONTINUOUSLY MEASURING MASS FLOW BY THE
CORIOLIS FORCE
A liquid mass-flowmeter with a rotating T-shaped element acts on the following
principle. The liquid flows through a double T-shaped which revolves at a constant
angular velocity (Fig. 8a). The tube system consists of two parts (1 and 2), joined by
flexible connections. The T-shaped tube is mounted on a torsion element 5, which
permits it to move through a certain angle relative to the rotating case 4, to which the
remainder of the rotating tubing is rigidly fixed. The system is joined to the stationary
tubing by means of sealing muffs. The rotor is at a constant speed by an electric motor
through the gearing 3. Under the action of Coriolis acceleration, as the liquid passes
8
through tube 1 a moment develops whose magnitude is proportional to the flow of liquid.
The moment can be measured by various electrical transmitters, a particular one for these
being the strain gauge, which is glued to the torsion element and measures the
deformation. The measuring signal is tapped off the rotor by slip rings and fed to the
measuring circuit; the magnitude of the measuring signal is in linear proportion to the
mass flow up to the limit of elasticity of the material of which the torsion element is
made.
The Coriolis force exerted on the rotating tubing by a liquid (Fig. 8b) can be
expressed by the following formula.
dP = ρ F dr a,
where ρ is the density of the liquid;
F is the area of cross-section of the tube;
R is the radial distance of the liquid element from the axis of rotation;
a is the Coriolis acceleration, which has the same value for any part of the liquid
present in the tube 1 (a = 2 w v);
v is the speed of liquid in the tube.
The magnitude of the mass flow rate through the tube is G = ρ F v. After the appropriate
substitution we obtain
9
dP = 2 G w dr
The moment about the axis of rotation developed in one radial tube 1 under the action of
Coriolis force is
M = ∫ 2 G w r dr = w (r22 – r1
2) G
It follows from the above formula that the magnitude of the measured moment is
unaffected by the properties of the liquid or the shape of the driving tube. The sole
requirement for this moment to develop is radial motion of the liquid in the rotating
channel, while the rotor can be designed to a great variety of patterns.
The defects inherent in the majority of flowmeters in which the flow is measured
by the magnitude of Coriolis force, are due to the use of sealed rotating joints, and are
entirely absent in an instrument employing an electrical rotor driven through the
hermetically sealed wall of a branch pipe. The design of such an instrument is as follows
(Fig .12). The rotating tube 2, with one central and several radial channels, is mounted in
low friction bearings 3 in the casing 1. The element 2 is made in the form of the rotor of a
synchronous electrical motor 4; it is rotated by means of an electrical field created by the
windings of the stator 5 situated outside the body.
When the rotor revolves at constant speed, the liquid travels along the rotating
radial channels from centre to periphery and then returns to the axis of the rotor along the
stationary channel 6, situated in the casing. During its motion in the rotating radial
channels, a Coriolis acceleration and the corresponding forces develop in the liquid.
Since the liquid, on emerging from the rotor, passes into the stationary channels, the
power spent on imparting to it a Coriolis acceleration is not returned to the rotating
system but is carried off with the stream. This power is proportional to the mass flow of
material, and may be written as
N = G (r22 – r1
2) w2
where G is the mass flow
r2 and r1 are the respective radii of the rotating channels of the rotor
In the flowmeters of the type in which the liquid returns to the rotor axis along the
revolving channels the power required to turn the rotor does not depend the amount of
flow. If conditions are created in such a device that for returning the liquid to the rotor
axis along the stationary channels, thus obviating an electric motor with a sealed shaft, it
10
is impossible to measure the flow by the power with sufficient accuracy because the
revolving rotor is connected through sealed joints in which the power used to overcome
friction has a value which in a number of cases greatly exceeds that needed to create
Coriolis force in the stream.
In the case where the rotor is driven by a rotating magnetic field passing through
the wall of the body, however the flow may be measured directly from the power
consumed to form Coriolis accelerations in the stream. This is due to greatly reduced
mechanical friction of the system.
In a flowmeter working on the principle under discussion, flow can also be
measured directly by the value of the torque developed in the stator of the electric motor.
In this case the stator of the electric motor, situated outside the connecting tube, is
designed to be able to rotate slightly in relation to the axis of the connecting tube.
Rotation of the stator is resisted by a force gauge which detects a torque arising in it.
When Coriolis forces develop on the uniformly rotating vane, a corresponding torque is
set up on it in the opposite direction to that of rotation, and with a value proportional to
the flow rate of material. A torque of equal value and opposite sign arises on the stator of
11
the electric motor and is detected by a force gauge. The measuring signal created at the
latter will therefore be proportional to the flow rate.
2.2 GYROSCOPIC CONTINUOUS MEASURING SYSTEMS FOR MASS FLOW
A gyroscopic mass flowmeter with a revolving rotor has the following structure.
The stream of liquid or gas passes along a special tube of complicated shape, rotating at
constant speed. The gyroscopic effect thus caused, proportional to the mass flow of the
substance, will tend to rotate the whole revolving system relative to the axis B-B (Fig.
19b). The size of this moment will be measured by some sort of transmitter.
The measuring device works as follows. If a force F (Fig. 19a) is applied to the
axis of the rotating disc, then under the action of the moment of precession the axis will
begin to move in a plane perpendicular to the vector of the force. Conversely, if the disc
rotating about its axis is caused to rotate also in a plane passing through its axis, then a
force F will occur on the axis of the disc.
If the rim of the rotating disc is replaced by an annular tube (Fig. 19b) along
which a liquid is continuously passing, and it is uniformly turned about the axis A-A, a
gyroscopic moment occurs relative to the axis B-B, whose value is proportional to the
mass flow through the annular tube.
12
2.3 TURBO-FLOWMETERS
The working principle of a turbo-flowmeter designed to measure the unsteady
flow is as follows. Located coaxially in the body of the instrument are two impellers,
mounted some distance apart (Fig. 29). Round the periphery of both impellers there are
channels, whose axes are parallel to those of the impellers. There is a certain amount of
radial clearance between the body and the vanes. Liquid enters the body, passes
consecutively through the channels in both the impellers and issue from the instrument
through the outlet tube. The upstream impeller is set in rotation by an electric motor at
constant angular velocity. The second impeller is fixed to an elastic element.
The rotating impeller creates a certain inertial moment in the liquid so that the
flow of liquid under torsion exerts a pressure on the blades of the fixed impeller, causing
it to rotate through a certain angle until equilibrium ensues between the moment acting on
the impeller, and what arises through torsion of the elastic element. The value of the
moment and consequently the deformation of the elastic element, are proportional to the
mass flow of liquid.
All the turbo-flowmeters under discussion, compared with other mass flow
meters, possess the following advantages: (a) small dimensions and weight; (b) low
hydraulic loss. In addition these instruments contain no sliding contacts in the measuring
circuit, which is also an extremely great advantage.
The driving impeller in turbo-flowmeters may be driven by means of an external
rotary magnetic field. In this case the rotating impeller is designed as the rotor of a
synchronous electric motor, the windings of whose stator is arranged externally directly
on the branch pipe of the flowmeter. The rotating magnetic field, creator by the stator
winding, passes through the metal tube of the instrument and sets in rotation the rotor
through which longitudinal channels have been drilled parallel to the flow axis. When the
driving impeller is actuated in this manner, the turbo-flowmeter will have the following
positive qualities.
1. The absence of rotating sealed couplings, which is very important when
measuring the flow of corrosive liquids and liquids under pressure.
13
2. A substantial simplification of the flowmeter, by mounting the stator on the
branch-tube of the flowmeter, because of the absence of any special electric motor
and a corresponding mechanical transmission.
3. A substantial reduction in weight and dimensions of the flowmeter for the same
reasons.
2.4 THE COMPARISON OF MASS FLOWMETERS
In order to create an intermediate measuring parameter in mass-flowmeters, an
additional movement is imparted to the substance to be measured. Through this
movement accelerations and forces arise which are proportional to the mass flow of the
substance under measurement. Mass-flowmeters are made both with rotating and
oscillating sensitive elements. In the instruments having rotating sensitive elements the
formation of the intermediate measuring parameter proceeds continuously, so that they
constitute continuous-measurement instruments. The dynamic ranges of flowmeters
whose sensitive elements are mounted on elastic couplings are governed by the natural
oscillating frequency of the corresponding oscillating systems.
In mass-flowmeters having oscillating sensitive elements, the inertial forces are
produced in the flowing substance in the same manner as with a uniformly revolving
rotor, though their value varies in time according to the law of variation of velocity of the
oscillating element. The force oscillating frequency of the system is chosen in most cases
equal to the natural oscillating frequency of the sensitive element.
14
CHAPTER 3
CONTINOUS MEASURING SYSTEM FOR VOLUMETRIC FLOW Continuous measuring systems for volumetric flow include the following: turbine,
ultrasonic, induction, thermal, those which continuously ionize the flow and those which
mark the flow in various ways.
3.1 TURBINE-TYPE FLOWMETER
Turbine-type flowmeters belong to the velocity flowmeters, in which the kinetic
energy of the stream under investigation is used to produce a torque on the measuring
vane. The measurement of unsteady flow by the speed of rotation of a rotor is affected by
two principal factors: variation in the velocity of the flow, and in its density.
The working principle of turbine-type flowmeters designed to measure unsteady
flow is as follows. A well-balanced light rotor is placed in the flow to be measured and
revolves in low-friction bearings. Under the pressure of the moving flow, the rotor
revolves, doing do at a rate proportional to the speed of flow. The vanes may be made
axial or tangential.
The rotational speed of the rotor may be measured in various ways:
electromagnetically, photo-electrically, by radioactivity etc. To measure the mass flow,
turbine flowmeters must be equipped with density pick-ups and the appropriating
correcting systems.
A. Turbo-magnetic flowmeters:
In this type of flow meter the speed can be measured simply and with great
accuracy by counting the rate at which turbine blades pass a give point, using a magnetic
proximity pickup to produce voltage pluses. By feeding these pulses to an electronic
pulse-rate meter, one can measure flow rate; by accumulating the total number of pulse
during a time interval, the total flow is obtained. These measurements can be made very
accurately because of their digital nature. If an analog voltage signal is desired, the pulses
can be fed to a frequency-to-voltage converter. The two types of turbo-magnetic
flowmeters are shown in the fig below.
15
Figure 1 Axial Turbine Flowmeter (Magnetic)
Figure 2 Tangential Turbine Flowmeter (Magnetic)
B. Turbo-optical flowmeters:
To measure the speed of a liquid of a sufficient transparency, one may use turbo-
optical flowmeters. In a turbo-optical flowmeters, the rotor is similar to that in the turbo-
magnetic flowmeter, except that the device measuring the speed of rotation of the rotor is
based on an optical principle. As the rotor 1 rotates, its blades intersect a beam of light
16
passing from the incandescent lamp 2 to the photoelement 3, thus alternately opening and
closing the passage of the light to the photoelement (Fig 3).
Figure 3 Turbo-optical flowmeter
At the electrodes of the photoelement pulsating electrical signals are produced as
a result, whose frequency is proportional to the flow of liquid through the flowmeter.
C. Turbo-radioactive flowmeter:
These flowmeters are constructed as follows. A balanced rotor (axial or tangential) is
placed in the flow to be measured and rotates under the pressure of the fluid. In one or
more of the blades radioactive isotopes with gamma-radiation are paced in the form of a
solid insert. The flowmeter is externally screened to prevent radiation. There is an
aperture in the screen for the outlet of the radiation into the external medium.
Opposite the aperture is placed a radiation indicator, coupled to a suitable
recording device. At the moment when the blade with the radioactive isotope is in the line
with the opening in the screen and the indicator, the radiation is detected by a counter.
The frequency of the radiation pulses falling on the counter is governed by the rotational
speed of the rotor, i.e. by the flow of fluid passing through the flowmeter.
17
3.2 ULTRASONIC FLOWMETER
Small-magnitude disturbances are propagated through a fluid at a definite velocity
(the speed of sound) relative to the fluid. If the fluid also has a velocity, the absolute
velocity of pressure-disturbance propagation is the algebraic sum of the two. Since flow
rate is related to fluid velocity, this effect may be used in several ways as the operating
principle of an “ultrasonic” flowmeter. The term ultrasonic refers to the fact that, in
practice, the pressure disturbance usually are short bursts of sine waves whose frequency
is above the range audible to human hearing, about 20,000 Hz. A typical frequency might
be 10 MHz.
The various methods of implementing the above phenomenon all depend on the
existence of transmitters and receivers of acoustic energy. A common approach is to
utilize piezoelectric crystal transducers for both functions. In a transmitter, electrical
energy in the form of a short burst of high-frequency voltage is applied to a crystal,
causing it to vibrate. If the crystal is in contact with the fluid, the vibration will be
communicated to the fluid and propagated through it. The receiver crystal is exposed to
these pressure fluctuations and responds by vibrating. The vibration motion produces an
electrical signal in proportion, according to usual action of piezoelectric displacement
transducers. For crystal to be an efficient transmitter of acoustic energy, its diameter D
must be large compared with the wavelength λ of the oscillation. The conical beam
projected from a circular crystal has a half-angle α given by sin α = 1.2λ/D; thus the
desired small angles also require a small λ/D ratio. Since compactness requires
reasonably small (~ 1 cm) values of D, we need to use wavelength on the order of 1 mm.
Figure 4 shows the most direct application of these principles. With zero flow velocity
the transit time to of pulse from the transmitter to the receiver is given by
cLto =
where L = distance between transmitter and receiver and c = acoustic velocity in fluid. If
the fluid is moving at a velocity V, the transit time t becomes
2cLVt =
and we define ∆t = to – t, then 2cLVt ≈∆
18
Thus, if c and L are known, measurement of ∆t allows calculation of V. While L may be
taken as constant, c varies, for example, with temperature; and since c appears as c2, the
error caused may be significant. Also, ∆t is quite small since V is a fraction of c.
Figure 4 Ultrasonic Flowmeter
Since the measurement of to is not directly provided for in this arrangement, the
modification is required. In the figure 5 two self-excited oscillating systems are created
by using the received pulses to trigger the transmitted pulses in a feedback arrangement.
The pulse repetition frequency in the forward propagation loop is 1/t1 while that in the
backward loop is 1/t2. The frequency difference is ∆f = 1/t1 - 1/t2, and since
t1=L/(c+Vcosθ) and t2=L/(c-Vcosθ), we get
LVf θcos2
=∆
Figure 5 Ultrasonic Flowmeter
This is independent of c and thus not subjected to errors due to changes in c.
19
CHAPTER 4
THERNAL SYSTEM FOR MEASURING UNSTEADY FLOW In thermal automatic monitoring system the rate of flow is measured either by the
cooling of a heated body placed in the flow or by the transfer of heat energy between two
points situated along the flow. There are two main classifications of the system, first
Thermo-anemometers and second, calorimetric flow meter.
4.1.1 THERMO-ANEMOMETER
The principle is based on the relation between the quantity of heat lost by a heated
measuring element and the rate of the ambient flow. This is further classified in two
groups. The first group comprises those in which the thermo element is connected as one
of the arms directly to a measuring bridge. These thermo-elements may be made with
their thermo element at either constant or variable temperature. The rate of flow is
measured by the variation of resistance in the thermo element for a current of constant
magnitude, or by the variation in the strength of the current for a thermo-element with its
resistance and temperature constant.
The second group contains instrument in which the receptor consist of a heated
filament and a thermocouple or thermistor, designed to measure the temperature of the
filament itself. In this instance the rate of flow is measured by the variation in the
filament temperature for constant power or current. The second group as a rule possesses
enormously greater inertia than the first.
The sensitive element of the thermo anemometer is made of platinum, tungsten or
nickel wire or 0.005 to 0.3mm diameter. As the diameter of the filament decreases so
does the inertia.
If, rate of the flow, v varies, both the current strength I and resistant of the thermo
element R vary. Since the output signal in this kind of instrument is usually expressed in
the form of a voltage drop on the measuring bridge, an independent variation of current
and resistance makes such a measurement impossible. To eliminate one of the variables,
the instrument’s power supply circuit must be designed so that the size of the current
passing through heated thermo-element always remains constant, regardless of its
resistance. In this case the rate of flow can be determined by the potential difference at
20
the ends of the thermo- element, the temperature of the flow and the characteristics of the
material of the thermo-element. The main drawback of the thermo-anemometer is that
their readings depend substantially on the temperature of the flow being measured.
The filament is subjected to an aerodynamic load, which depends on the ratio of
the length of the wire to its diameter, and to a shock load occurring when it is bombarded
by solid particles borne along with the flow. If the vibration load does not damage the
filament, it will cause its resistance to pulsate, which may introduce substantial errors
into the measurement. Higher is the temperature of the thermo element, the more
sensitive is the instrument, and less its reading are affected by fluctuation in the
temperature of the flow being measured. The heating range of the thermo-element
normally lies in the range of 400-5000C. If the axis of the heated filament is
perpendicular to the direction of the airstream, the stream exerts the maximum effect on
the filament else it would have some cosθ component. The transfer of the heat depends
on the rate of flow, temperature difference between the body and medium.
4.1.2 HOT WIRE ANEMOMETER
The hot-wire anemometer increases VOUT until the power dissipated in the wire
sensing element, and hence its temperature and resistance, has risen to the point where
the bridge at equilibrium. Air movement past the wire would cool it, but VOUT increases
compensate for the increased dissipation restoring the wire to its equilibrium temperature.
The equilibrium behavior of the system is independent of the heat capacity of the wire.
However, the dynamic response, and noise figure, are both improved by minimizing the
size of the wire. Anemometry will refer to the use of a small, electrically heated element
exposed to a fluid medium for the purpose of measuring a property of that medium.
Normally, the property being measured is the velocity. Since these elements are sensitive
to heat
Transfer between the element and its environment, temperature and composition changes
can also be sensed.
Typical dimensions of the wire sensor are 0.00015 to 0.0002 inches (0.0038 to
0.005 mm) in diameter and 0.040 to 0.080 inches (1.0 to 2.0 mm) long. This is the type of
21
hot wire that has been used for such measurements as turbulence levels in wind tunnels,
flow patterns around models and blade wakes in radial compressors.
4.1.3 HOT FILM ANEMOMETER
The hot film is used in regions where a hot wire probe would quickly break such
as in water flow measurements. The hot-film sensor is essentially a conducting film on a
ceramic substrate. The sensor shown in Figure 2 is a quartz rod with a platinum film on
the surface. Gold plating on the ends of the rod isolates the sensitive area and provides a
heavy metal contact for fastening the sensor to the supports. The metal film thickness on
a typical film sensor is less than 1000 Angstrom units, causing the physical strength and
the effective thermal conductivity to be determined almost entirely by the substrate
material. Most films are made of platinum due to its good oxidation resistance and the
resulting long-term stability. The ruggedness and stability of film sensors have led to
their use for many measurements that have previously been very difficult with the more
fragile and less stable hot wires. This type of anemometer also facilitate advantages over
hot wire anemometer like,
1) Better frequency response (when electronically controlled) than a hot wire of
the same diameter because the sensitive part of the sensor is distributed on the surface
rather than including the entire cross section as with a wire.
22
2) Lower heat conduction to the supports (end loss) for a given length to diameter
ratio due to the low thermal conductivity of the substrate material. A shorter sensing
length can thus be used.
3) More flexibility in sensor configuration. Wedge, conical, parabolic and flat
surface shapes are available.
4) Less susceptible to fouling and easier to clean. A thin quartz coating on the
surface resists accumulation of foreign material. Fouling tends to be a direct function of
size.
figure 3, probe as a bridge branch
23
4.2 PROBE SHAPES
In addition to the cylindrical shape, hot films have been made on cones, wedges,
parabolas, hemispheres, and flat surfaces. Cylindrical film sensors that are cantilever
mounted are also made. This is done by making the cylindrical film sensor from a quartz
tube and running one of the electrical leads through the inside of the tube.
4.2.1 SINGLE ENDED SENSOR
This is an important modification for fluidic applications since they can be made
very small and inserted into very small channels. Also, for omni-directional
measurements (e.g., meteorology applications when the vertical flow can be ignored), it
permits unobstructed flow from all directions.
4.2.2 CONE SHAPE SENSOR
This sensor is used primarily in water applications where its shape is particularly
valuable in preventing lint and other fibrous impurities from getting entangled with
sensor. The cone can be used in relatively contaminated water, while cylindrical sensors
are more applicable when the water has been filtered.
24
figure 5, cone probe
4.2.3 FLUSH MOUNTED PROBE
Flush mounted probe which has been used for sensing the presence of flow with
no obstruction in the fluid passage, detecting whether the boundary layer is laminar or
turbulent, and measurements of shear stress at the wall. It makes a very rugged probe
when compared with other anemometer type sensors.
figure 6, flush probe
4.2.4 WEDGE SHAPED PROBE
The wedge shaped probe shown in Figure 7 has been used for both gaseous and
liquid applications. It is somewhat better than cylindrical sensors when used in
25
contaminated water and is certainly stronger than cylindrical sensors for use in very high
velocity air or water where there is a large load on the sensor due to fluid forces.
4.3 CONSTANT TEMPERATURE ANEMOMETER (CTA)
It works based on the fact that the probe’s resistance will be proportional to the
temperature of the hot wire. The bridge circuit shown in Figure 3 below is set up by
setting the adjustable resistor to the resistance you wish the probe and its leads to have
during operation. (The other two legs of the bridge have identical resistance.) The servo
amplifier tries to keep the error voltage zero (meaning the resistances of the two lower
legs of the bridge match). It will adjust the bridge voltage such that the current through
the probe heats it to the temperature, which gives the selected resistance. When we put
the probe in a flow, the air (or water) flowing over it will try to cool it. In order to
maintain the temperature (resistance) constant, the bridge voltage will have to be
increased, Thus, faster the flow, higher the voltage. A very fine hot wire by itself cannot
respond to changes in fluid velocity at frequencies above about 500 Hz. By compensating
for frequency lag with a non-linear amplifier this response can be increased to values of
300 to 500 kHz.
4.4 CALORIMETRIC FLOW METERS
The principle is based on the heat transfer by the flow of material. This is divided
into three groups,
26
1) Devices operating by the application of constant power to the heater, and with
measurement of the amount by which the flow is heated, depending on the velocity.
2) Those which operate by heating the flow to constant temperature and by
measuring the amount of energy applied to the heater and depending on the velocity of
flow.
3) Those in which the temperature of the heater varies sinusoidal in time. In these
devices the rate of flow is measured by the amount of phase shift of the signals obtained
at the meter and those applied to the heater.
The measuring elements used in calorimetric flow meters are thermocouples or
resistance thermometers, which have relatively low dynamic properties. The measuring
frequency range of these instruments is small. In calorimeters operating with a constant
temperature difference of the flow before and after the heater, the flow is measured by the
amount of power consumed by the heater. When the flow temperature varies, the
measuring bridge goes out of equilibrium and varies the current strength in the heater to a
value at which a pre-set degree of heating of the gas is restored. The inertial errors
created by such a system are governed by the thermal inertia of the temperature
measuring devices employed, and also by the heater operating in a pulsating mode. The
measuring errors substantially decrease as the distance between the heater and the
measuring devices is reduced. The equation for the thermal balance is given as,
q = QCp∆t
where, Q is the flow of gas:
q is the heat consumed in heating the gas
∆t is the temperature difference of the gas before and after the heater.
Cp is the thermal capacity for constant pressure.
Heat is transferred to the flow by an electrical heater, hence,
q = KU2/R2
Q = KU2/RCp∆t2,
Where R is the resistance to be heated, through which passes a current of strength I with a
voltage drop U. K is the thermal equivalent of the electrical energy.
The calorimetric method of measuring the flow of a liquid or a gas can be so
designed that the heating and measuring elements are not introduced into the flow, but
27
are mounted on the outer surface of the flow meter tubing. The flow is measured by the
transfer of heat by the boundary layer of the substance flowing in the tube. It should be
noted that the dynamic sensitivity of this method is poor, since the motion of the
boundary layer follows that of the main flow with a certain time lag, and also because of
the relatively large thermal inertia of the walls of the branch tube. the static response of
such a measuring system is linear.
The mode in which the heater of thermal flow meter operates may be either steady
or pulsating, and in later case an intermediate parameter is periodically formed in the
flow. The readings of the flow meters with continuous heating of the flow, and which
measure by the amplitude method, depends greatly on the pressure of the medium being
measured, since the conditions of heat transfer alter considerably with a change of
pressure. The dynamic properties of calorimetric flow meters are governed by the thermal
inertia of their elements. The dynamic properties of the flow meters improves as the
masses of these elements is reduced, though the strength of the elements also diminished.
The dynamic properties of calorimetric flowmeters are far lower than for thermo
anemometers.
28
CHAPTER 5
CONTINOUS MEASURING SYSTEM WITH CONTINUOSLY
IONIZED FLOW Automatic monitoring systems with continuously ionized flow includes,
1) Flow meters with continuous ionization of the flow by radioactive irradiations.
2) Ionization anemometers with a glow discharge.
The characteristic feature of this group of devices is that their primary transducers
produce an intermediate electrical parameter in the flow, from whose value the velocity
of the flow can be found.
5.1 FLOW METERS WITH CONTINOUS IONIZATION OF THE FLOW BY
RADIOACTIVE IRRADIATION
It consists of two electrodes 1 and 2, attached to the inner surface of the tubing. A
layer of radioactive material is deposited on electrode 2, which emits α or β particles
which ionize the gaseous flow passing between the electrodes. Owing to the potential
difference (100-300V) applied to the electrodes, and ion flux is formed in the inter-
electrode space. The ions so formed travel in the space between the electrodes at a
velocity determined by two components. The component v1 of the velocity of the ions,
which is directed along the lines of force, is determined by the strength of the electrical
field H and the mobility of the ions k (v1 = kH). The second component of the velocity of
the ions is the speed of the flow v.
The gaseous flow carries away from the inter-electrode space a certain quantity of
ions, the number of ions swept away increasing with a rise in the velocity of flow. This
causes diminution in the current in the measuring circuit, and at some value of the rate of
flow, when all the ions are swept away, current will cease altogether in the circuit. These
instruments may be constructed either with a plane or cylindrical ionization chamber. The
ionizer employed may be, e.g., radioactive polonium deposited on one of the electrodes.
For measuring the flow of substances which corrode the radioactive coating, a device is
used, in which a plate with the isotopes deposited on it is covered with a special
corrosion-proof protective film. This device will be corrosion proof, so that steady
29
radioactive emission will take place over a longer interval than in the case of a
radioactive isotope placed openly in the flow.
The accuracy of this method is comparatively poor, since, the ion flux depends
not only on the rate of flow, but also on the mobility of the ions and the recombination
rate constant, which to a great extent are governed by the state of the gaseous medium.
When measuring the velocity of a gas, the range of flight of the α particles depends on its
temperature, pressure and humidity. The devices may be used to measure pulsating flow
with an average velocity from 0 to 100m/min.
Among the disadvantages of continuous ionization of flow should be included the
direct relation of the readings of the measuring instrument to the magnitude of the supply
voltage and to the decay of the radioactive isotope.
5.2 IONIZATION ANEMOMETERS WITH A GLOW DISCHARGE
Ionization anemometer with a glow discharge, which are also called gas resistance
instruments, are usually made in the form of a probe on which platinum electrodes are
mounted. Under the action of a fairly large potential difference between the two platinum
electrodes, with a clearance of 0.1-0.2m, a glow discharge is obtained. The gaseous flow
whose rate is being measured affects the conditions under which the ionized particles
pass between the electrodes, which are detected as a variation in the resistance of the
discharge gap. The ion flux varies inversely as the rate of flow. If a device is used to
maintain the current between the electrodes constant, the potential difference between the
electrodes varies as the rate of flow. The sensitivity of the glow discharge anemometer is
governed by the blowing out of ions from the spark gap, the effect off the temperature
and pressure of the medium, the effect of gas cooling on the thermo-ionic properties of
the electrodes and mechanical deformation of the electrodes caused by pressure of the
flow or its cooling effect on the electrodes when heated by the discharge. In anemometer
with a fixed current passing through the ionization gap, the source of energy of the glow
discharge may be either d.c. or a.c. sources, which ensure a fixed value of the current
strength of the order of 10mA. A great disadvantage of the flow discharge anemometer
using a d.c. source is the blasting of the material of the cathode and a subsequent voltage
30
increase because of the increased gap. This fault is absent in the a.c. source either at
relatively low or at high frequencies.
The accuracy of the measurement depends to a large extent on the amount and
uniformity of the natural ionization of the flow being measured. The anemometer has low
inertia and high sensitivity. Disadvantages include the unsteadiness of the reading and the
blow out of the charge. The instrument we have been using are used to determine the
local velocity of the flow, so that when measuring the flow through a tube of given
section, the profile of the distribution of velocities must be taken in to account.
31
CHAPTER 6
MISCELLANEOUS
6.1 SYSTEM OF DIGITAL MEASUREMENT OF UNSTEADY FLOW USING
VARIOUS FLOW MARKERS
In several cases, unsteady velocity and flow must be measured in high speed
stream, in to which for a great many reasons measuring elements cannot be introduced,
for these purposes systems are employed which operate by forming measuring markers
directly in the stream being metered. The markers so produced constitute an integral part
of the flow, and experiences all its variations of velocity. The marker can be ionic,
radioactive, optical or thermal. The working principle is based on the flow markers and
on measuring the time of motion of the marker together with the flow between two points
on the tubing. When the flow s measured by means of such devices, the profile of the
distribution of velocities over the channel cross section must be taken care. In order to
measure the velocity, two types of the electronic circuits are employed and this gives
continuous measurement of the time interval during which the flow marker travels with
the stream from its point of formation to that of registration. These are open loop and
closed loop circuits. Different classifications are,
Flow markers and mediums:
Type of markers Method of marker
formation
Medium to which
methods apply
Ionic Spark ionization of flow
Gases at low
temperature and
humidity in the
absence of external
ionization of flow.
Ionic
Ionization of flow by
modulated radioactive
emission
Gases at low
temperature and
humidity in the
absence of external
ionization of flow.
32
Radioactive Introduction of the radio
active isotopes into flow
Gases and liquids
of any physical
properties.
Optical, opaque Introduction of opaque
substance into flow
Transparent liquid
and gases
Optical, radiating
Introduction into flow of
radiating substances of
given spectral composition
High temperature
gases
Thermal Periodic heating of part of
flow
Liquids not at high
temperature
Vortex
Creation of rotating vortex
in flow by periodic
switching of magnetic field
Conducting liquid
Table 1.1
6.2 INSTRUMENTS WHICH MEASURES FLOW BY VELOCITY PRESSURE
For measuring unsteady flow of liquids and gases, devices are used which
determine the flow by the velocity head. They include manometric and immersion
flowmeters.
Immersion flow meters are instruments whose sensitive element is a body which
detects the dynamic pressure of the current flowing about it, and which acts on a
measuring element. These instruments can be divided into systems of constant pressure
drop and systems in which velocity pressure is measured directly. The instruments of this
group used for measuring pulsating flow include low inertia devices in which the velocity
pressure set up by the current on the sensitive element is determined directly. Flow is
measured in these devices by the amount of movement of the body under action of
velocity pressure or by the amount of the corresponding force. Flow meter also devised to
measure flow by the velocity pressure, without introducing any immersed body into the
flow. They measure the amount of velocity pressure on the branch-tube wall as the flow
changes direction at right angle.
Flow meter of the manometric group include instruments in which the pulsating
flow is determined by pressure difference before and after a fixed perforated diaphragm,
33
at two points of Venturi tube inserted in the flow, or obtained by mean of a Pitot tube.
Various differential electrical pressure pick-ups are employed for measuring the pressure
difference: capacitive, inductive, and others.