Fluoroscan ® InSight FD Mini C-Arm Be smart with dose. Introducing the Low Dose mode Limiting the amount of radiation exposure to patients and healthcare professionals is a key topic in healthcare today. When using fluoroscopy, it’s important to be aware of the different methods used to reduce dose and how different manufacturers chose to use them. The Hologic Fluoroscan InSight FD mini C-arm is designed to use dose efficiently, applying the ALARA “as low as reasonably achievable” – radiation safety principle, to reduce unnecessary radiation exposure to both you and your patients. The low dose mode enables the user to reduce dose by up to 50% compared to auto mode. 4

Fluoroscan InSight FD Mini C-Arm Be smart with dose ......OrthoScan FD Pulse’s Low dose mode at 7.5 pps.1,2 Fluoroscan InSight FD system, when using its high resolution mode at 15

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Fluoroscan InSight FD Mini C-Arm Be smart with dose ......OrthoScan FD Pulse’s Low dose mode at 7.5 pps.1,2 Fluoroscan InSight FD system, when using its high resolution mode at 15

Fluoroscan® InSight™ FD Mini C-Arm

Be smart with dose. Introducing the Low Dose modeLimiting the amount of radiation exposure to patients and healthcare professionals is a key topic in healthcare today. When using fluoroscopy, it’s important to be aware of the different methods used to reduce dose and how different manufacturers chose to use them. The Hologic Fluoroscan InSight FD mini C-arm is designed to use dose efficiently, applying the ALARA – “as low as reasonably achievable” – radiation safety principle, to reduce unnecessary radiation exposure to both you and your patients. The low dose mode enables the user to reduce dose by up to 50% compared to auto mode.4

Page 2: Fluoroscan InSight FD Mini C-Arm Be smart with dose ......OrthoScan FD Pulse’s Low dose mode at 7.5 pps.1,2 Fluoroscan InSight FD system, when using its high resolution mode at 15

Superb image quality at a lower doseThe InSight FD system is designed to be extremely dose efficient, using lower technique factors (kVp and/or mA) than the OrthoScan FD Pulse to achieve the same image quality.3 When imaging smaller anatomy, the system uses up to 21% less kVp to acquire the same image in the auto mode.

A collimator with significant dose savingsIn addition to the dose-saving benefits of the flat detector’s shape, size and rotation capabilities, the system’s collimator can also be used to limit the field of view to 125cm2 and decrease dose by up to 33%.1 The InSight FD system uses up to 30% less dose when collimated compared to OrthoScan FD Pulse with its same sized collimation.1,2

Dual imaging modes to provide optionsThe InSight FD system comes with two pre-set imaging modes that use dose differently. For routine imaging, the standard Auto Mode provides the greatest dose efficiencies. The Auto IQ Mode, which uses slightly more dose, is ideal when there’s a need for lower noise and better definition.

125cm2

Orthoscan

15 cm x 15 cm detector

Hologic

12 cm x 15 cm detector

125cm2

Fluoroscan® InSight™ FD system, when using the Auto mode at 30 fps produces up to 39% less dose compared to OrthoScan FD Pulse’s mode at 30 pps - without the need for pulsed fluoroscopy.1,2

Fluoroscan InSight FD system, when using its low dose mode at 15 fps, produces up to 34% less dose compared to OrthoScan FD Pulse’s Low dose mode at 7.5 pps.1,2

Fluoroscan InSight FD system, when using its high resolution mode at 15 fps, produces up to 5% less dose compared to OrthoScan FD pulse’s high dose mode at 30 pps.1,2

A flat detector tailored for extremity imaging

Mini C-arms are designed for imaging extremities where the anatomy is often long and narrow. We chose a rectangular shape for our detector to best suit the anatomy it is imaging. This is designed to help minimize unnecessary radiation to the patient and staff by not exposing the unused area on the detector. Combine our detector’s ability to rotate with its tailored shape, and you have a system that is uniquely designed for extremity imaging.

0.300

0.200

0.100

0

0.208

R/M

in

0.152

0.133

0.091

OrthoscanMed 15pps

OrthoscanLow 7.5pps

HologicAuto IQ 30fps

HologicAuto 30fps

OrthoscanHigh 30pps

0.220

SKIN ENTRY DOSE1,2

37%LESSDOSE

Hologic Hi-ResAuto 30fps

0.208

0.060

HologicLow 15fps

15

10

5

0

7.33

DA

P µ

Gym

2

5.9

9.02

6.22

OrthoscanMed 15pps

OrthoscanLow 7.5pps

HologicAuto IQ 30fps

HologicAuto 30fps

OrthoscanHigh 30pps

15.27

DOSE AREA PRODUCT1,2

39%LESSDOSE

Hologic Hi-ResAuto 30fps

7.73

2.31

HologicLow 15fps

Pictures show competitor’s images using a square detector. Outline represents the InSight FD system’s rectangular detector. Its rotating capability enables relevant anatomy to be captured.

Page 3: Fluoroscan InSight FD Mini C-Arm Be smart with dose ......OrthoScan FD Pulse’s Low dose mode at 7.5 pps.1,2 Fluoroscan InSight FD system, when using its high resolution mode at 15

Features designed for speed and efficiency When taking a single image, it’s important to optimize radiation used to generate each image.

The InSight FD system has a true “snapshot” mode that applies the ALARA – “as low as reasonably achievable”– radiation safety principle. The user simply presses the X-ray button or pedal, and the system automatically uses the lowest dose possible to get the image.

Conversely, with Orthoscan’s system, radiation is entirely user dependent, requiring the user to press the X-ray button or pedal until the image stops adjusting and a clear image is displayed, then release. This manual method can result in more dose than necessary to get an optimal image – or if the X-ray switch is released too soon, a poor image that needs to be retaken.

Both can lead to more unnecessary radiation exposure. The automated dose-saving snapshot mode takes the guesswork out of using dose efficiently.

The InSight FD system is also up to 60% faster than the competition when normalizing to reach the appropriate technique factors.4 This saves time and reduces unnecessary radiation.

A system engineered to protect you

When measurements are taken at the head, waist and knee height, the InSight FD system at 30 fps produces, on average, up to 50% less scatter radiation to the operator compared to OrthoScan FD Pulse at 30 fps, in the auto mode.5,6

At the low dose mode at 15 fps, the InSight FD system produces up to 55% lower scatter radiation on average, compared to OrthoScan’s low dose mode at 7.5 pps.

It’s important to protect everyone – the patient, the surgeon and the staff – from unnecessary radiation exposure.

0 1 2 3 4 5 6

Hologic: 2-3 seconds

Orthoscan: 4-5 seconds

Seconds

5

4

3

2

1

0

mR

/hr

Operator Head

2.44

1.5

1.9

Operator Waist

Operator Knees

SCATTER RADIATION5

4.16

OrthoscanHigh 30pps

OrthoscanMed 15pps

HologicAuto IQ 30fps

HologicAuto 30fps

42%LESS

SCATTER

Hologic Hi-ResAuto 15fps

OrthoscanLow 15pps

HologicLow 15 fps

2.1

0.6

1.28

3

2

1

0

mR

/hr Operator Knees

1.060.81.1

2.02

OrthoscanHigh 30pps

OrthoscanMed 15pps

HologicAuto IQ 30fps

HologicAuto 30fps

Hologic Hi-ResAuto 15fps

OrthoscanLow 15pps

HologicLow 15 fps

1.2

0.3

0.96

5

4

3

2

1

0

mR

/hr

Operator Waist

2.35

2.83.2

4.16

OrthoscanHigh 30pps

OrthoscanMed 15pps

HologicAuto IQ 30fps

HologicAuto 30fps

Hologic Hi-ResAuto 15fps

OrthoscanLow 15pps

HologicLow 15 fps

3.5

0.8

1.55

Be smart with dose. Know the facts.

SCATTER RADIATION5,6

Page 4: Fluoroscan InSight FD Mini C-Arm Be smart with dose ......OrthoScan FD Pulse’s Low dose mode at 7.5 pps.1,2 Fluoroscan InSight FD system, when using its high resolution mode at 15

WP-00103 Rev 002 (6/18) © 2018 Hologic. All rights reserved. Hologic, Fluoroscan, InSight, The Science of Sure and associated logos are trademarks and/or registered trademarks of Hologic, Inc. and/or its subsidiaries in the United States and/or other countries. All other trademarks, registered trademarks, and product names are the property of their respective owners. This information is intended for medical professionals in the U.S. and other markets and is not intended as a product solicitation or promotion where such activities are prohibited. Because Hologic materials are distributed through websites, eBroad-casts and tradeshows, it is not always possible to control where such materials appear. For specific information on what products are available for sale in a particular country, please contact your local Hologic representative or write to [email protected].

1. F.X.Masse Associates Inc, Report Apr 6 2018, Air Kerma and Exposure rate measured 5 cm above imaging assembly with a 19mm Al phantom and DAP measurements for Fluoroscan InSight FD, measured for 30 second exposure of 19mm Al phantom. 2. F.X.Masse Associates Inc, Report dated Sep 3 2015 and Jan 14 2016, Exposure rate measured 5 cm above imaging assembly with a 19mm Al phantom and DAP measurements for Fluoroscan InSight FD, measured for 30 second exposure of 19mm Al phantom. 3. F.X. Masse Associates, Inc. Report, September, 2, 2015 and January 14, 2016. High and Low Contrast Resolution. Measurements made with no phantom, 3.3 mm or 19 mm AI phantom. 4. Data on File. 5. “Scatter Measurements for OrthoScan FD pulse ,F.X. Masse Associates, Inc. Report, September, 2, 2015 and January 14, 2016. Scatter Measurements. Measured in a vertical plane one foot from table edge using a 19 mm AI phantom. 6. F.X. Masse Associated, Inc Report dated Apr 6 2018, Scatter Measurements for Fluoroscan Insight FD - Measured in a vertical plane one foot from table edge using a 19mm AI phantom.

Fluoroscan.com | [email protected] | +1.800.321.4659

Fluoroscan® InSight™ FD Mini C-Arm

Be smart with dose. Know the facts.