New Ideas on Gas Chromatograph SiteMaintenance and Monitoring
By Anwar Sutan
www.i-Vigilant.co.uk
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 2/38
Objectives
• Review current Gas Chromatograph maintenance regimes
– Current practices subect to !i""erent interpretations – #mplementation re$uires skille! technicians
– GC is vulnerable to unsatis"actory per"ormance
– &he use o" automate! system to gather important parameters
– &he use o" log-log chart
– &he use o" live uncertainty value as per"ormance benchmark
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 3/38
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 4/38
Repeatability Test
• &olerance is normally base! on AS&' ()*+, stan!ar!
• ormally !one between / , runs
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 5/38
Repeatability Test
• , runs o" repeatability test on methane
AS&' ()*+, tolerance
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 6/38
Repeatability Test
• +0 hours o" repeatability test 1'ethane (ata2
Calibration gas concentration3 4+.**+
AS&' ()*+, tolerance3 4+.4*+ / 4,.0*+
'inimum measure!3 4+.*45,
'a6imum measure!3 4,.0
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 7/38
Repeatability Test
• +0 hours o" repeatability test 17ropane (ata2
Calibration gas concentration3 ).408
AS&' ()*+, tolerance3 ).588 / ).458
'inimum measure!3 ).40)9
'a6imum measure!3 ).40,)
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 8/38
Repeatability Test
• #s repeatability test really worth !oing:
• GC is capable to per"orm much better than normally use!limits
• 'any people think that i" GC pass repeatability test; it means
– #n "act GC can provi!e a repeate!ly wrong results
• 'ost o" the times; when repeatability test "ails; test was
per"orme! on un-normali<e! basis
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 9/38
Control Charts
• #t is plot o" response "actors an! retention times o" GC
• #t only re"lects the per"ormance o" GC against calibration gas• ormally a contractual re$uirement
• &here are variation on how people !o this
– 8 st! !eviation o" "irst , !ata is set as control limit – 8 st! !eviation o" all !ata is set as control limit
• =hat woul! logical>practical limit be:
– #" GC per"orms well "rom beginning..
– #" GC per"orms poorly "rom beginning..
• =hat woul! pass>"ail criteria be:
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 10/38
Control Charts Simulation
• &o prove a point; consi!er the "ollowing theoretical values
• charts; both o" methane R? have 8 st! !eviation
Current practice say
both GC works
e$ually goo!
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 11/38
Control Charts Simulation
• ?irst control chart has synchronous components change
• Secon! control chart has ran!om component changes
Comparing the two
charts; it is visible
that the top chart is
better than bottom
chart
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 12/38
Control Charts Simulation
• ?irst control chart has synchronous components change
• Secon! control chart has ran!om component changes
%ooking at the secon!
chart in!ivi!ually;
everything looks
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 13/38
Control Charts Simulation
• ?irst control chart has synchronous components change
• Secon! control chart has ran!om component changes
Better 7er"ormance
=orse 7er"ormance
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 14/38
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 15/38
Control Charts
• ealthy GC Control Chart
• e6ane an! eptane per"orm slightly worse than the rest o"
the components
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 16/38
Control Charts
• Actual GC uncertainty
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 17/38
Control Charts
• ?ew points3
– Currently use! control chart looks at components in!ivi!ually an!not as a whole
– &here is no stan!ar!i<e! limit "or the control chart monitoring
• 8 st! !eviation o" the "irst , !ata or all !ata is totally !epen!ent on how
goo! > ba! the GC per"ormance is• &his shoul! not be use! as control limit
– 7er"ormance o" in!ivi!ual component tren! in control chart cannot
be use! to !etermine the actual per"ormance o" the GC
– uncertainty value can be !erive! "rom control chart an! may be abetter representation o" the actual GC per"ormance
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 18/38
Calibration Gas Correctness
• Scenario )3
– Calibration gas has been e6pose! to the winter col! below themi6tureDs !ew point temperature
– Some heavy en!s have migrate! to the bottom o" the bottle
– 'i6ture is non-homo enous an more
– &echnician believes the value written in the certi"icate
– &his can lea! to $uite a signi"icant measurement error
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 19/38
Calibration Gas Correctness
• Scenario 3
– (uring shut !own; a technician put new calibration gas; an! "orgot toput the new component !ata table
– e also "orgot to put a note that new calibration gas has been
installe!
– ?ew months later; ne6t technician came in; wante! to start up the GC – e !i! a "orce calibration on the GC
– &his lea!s to a signi"icant measurement error
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 20/38
Composition Comparison
• #n many places un-normalise! composition between two GC is
compare!
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 21/38
Composition Comparison
• =hen normali<e!; in "act all is goo!
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 22/38
Not Enough Confidence in GC
• E6ample3
– GC result shows that the gas is o""-spec – &echnician is usually calle! to check what is wrong with the GC
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 23/38
Current Regime Summary
• Repeatability test most likely pass
• ?ailure normally !ue to wrong proce!ure• Control chart in!ivi!ual limits cannot be use! to !etermine the
per"ormance o" the GC
• urt er ata process ng rom contro c art ea s to uncerta nty
value that can be use! to !etermine GC per"ormance
• Calibration gas correctness check is o"ten misse!
• =rong composition is compare! lea! to "alse !oubts o" GC
• %ittle con"i!ence in GC measurements
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 24/38
So How can We Detect Such Issue
andImprove Our Confidence?
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 25/38
Automate! System
• A!vantages
– &akes away re$uirement to manual entry !ata to a sprea!sheet
• A lot o" hours o" technician time is save!
• Removes operator error
• Better interpretation o" the !ata
– Alarm "lag in case one o" the parameters "aile! – istorical !ata in the "inger tips
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 26/38
Log-log Chart
• othing new about log-log chart; but3
– &he current practice is to look all gases together in the same chart
– Check is per"orme! to see i" the chart looks linear
– &his !oes not necessarily mean that GC is not healthy
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 27/38
Log-log Chart
• %og chart to be mo!i"ie! as per the GC gas path
• =hen put together result "ails 1R F 0.**42• =hen separate! base! on the gas path in the column; all
results pass
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 28/38
Log-log Chart
• =hat can be !etermine! "rom log-log chart:
– #" the GC valve timings are correct
– #" the calibration gas composition is as per the certi"icate
– &his will provi!e with con"i!ence that GC "unctions correctly on that
moment
– nce we have the con"i!ence on the GC; how can we monitor that itmaintains its repro!ucibility:
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 29/38
Live CV uncertainty value
• &his is !erive! "rom repro!ucibility !ata > control chart
• &he !i""erence is3 all !ata are looke! at together as a whole• ncertainty value can be !erive!
• istorical !ata o" uncertainty can be tren!e!
• Check both long term uncertainty value an! short termuncertainty value
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 30/38
Live uncertainty value
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 31/38
Live uncertainty value with Limit
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 32/38
Last Calibration Date
• n automate! system; no up!ate in calibration !ata means3
– Calibration in the GC has "aile!
– &here is a communication "ailure in the system
– Both issue can be easily checke!
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 33/38
Daily Activities
• Check that the status o" the GC is healthy
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 34/38
How About the Control Chart?
• Control chart is integral part o" the uncertainty calculation
• =e can now interpret it better
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 35/38
Analysis Parameters
• #n case o" "ailures; knowing the healthy value o" the analysisparameters will help troubleshooting
– GC ven temperature
– Carrier gas pressure
– Valve actuating pressure
– Carrier gas "low rate 1through each gas path2
– Bri!ge balance
– Sample vent "low
– Sample pressure
• =hen GC per"ormance is poor; one o" these parameters willhave change!
• &his helps pin-pointing where the issue is
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 36/38
8/17/2019 GC Site Maintenance and Monitoring
http://slidepdf.com/reader/full/gc-site-maintenance-and-monitoring 37/38
Conclusions
• (oes the current GC maintenance regime provi!e "ull
con"i!ence o" GC per"ormance:
• =hen implemente! the new i!eas will provi!e the "ollowing
bene"its3
–
– Early "ailure !etection base! on uncertainty values
– igher con"i!ence o" the GC
– %ess maintenance time
– %ess unnecessary errors