GEO600 Status Update Harald Lück Hannover, April 2005

GEO600 Status Update Harald Lück Hannover, April 2005

  • Upload
    aelwen

  • View
    62

  • Download
    1

Embed Size (px)

DESCRIPTION

GEO600 Status Update Harald Lück Hannover, April 2005. S4. Feb 22nd – March 23rd Two manned shifts/day (5-21 UTC), 1 ‚expert on duty‘ 8-8UTC Overnight SMS alarms to ‚expert on duty‘ Locking status DAQS (DCUs running, frame making, timing, calibration) Temperatures Vacuum - PowerPoint PPT Presentation

Citation preview

  • GEO600 Status UpdateHarald LckHannover, April 2005

  • S4Feb 22nd March 23rdTwo manned shifts/day (5-21 UTC), 1 expert on duty 8-8UTCOvernight SMS alarms to expert on dutyLocking statusDAQS (DCUs running, frame making, timing, calibration)TemperaturesVacuumInstrumental duty cycle 97.5%

  • Typical S4 Sensitivityh [1/sqrt(Hz)]

  • S4Glitchy periods within the first week

  • Glitches occurred mainly during seismically quiet times

  • Curing the glitches(temporarily)

    (What is a jump? :-)

  • Intermediate Mass of Mounting Unit 3 probably touched coil holder frame

  • After unstickng MU3by rotating it

  • BLRMS during S4

  • Strange sine gaussian glitchesh [1/sqrt(Hz)]

  • Strange sine gaussian glitches ctnd.

  • Strange sine gaussian glitches ctnd.

  • Calibration of GEO600Time-domain calibration methodOptimization routine quasi-continuously estimates parameters of optical responseCalibration lines give known differential arm-length disturbanceThese points are fit to a modelUpdate rate is 1Hz

  • Michelson Servo Error PointsPQ

  • On-line optical TF measurementsactuatorCALP and Q

  • Optical&Electronic Gain

  • Calibration

  • Calibrated EP Quadrature Signalsh [1/sqrt(Hz)]

  • Combining hP(t) and hQ(t) IIICreate filters from noise floor estimatesh(t) = Pfilter{hP(t)} + Qfilter{hQ(t)}sQQsPPsPQ

  • Combining hP(t) and hQ(t) resultsGet the best of hP and hQ plus a little extra!h [1/sqrt(Hz)]

  • Steps after S4PR fast autoalignment, Unity gain ~4-5Hz

    Improve seismic long. FF & long.->IM tilt FFIncrease PR factor by changing PR mirror 1.35% -> 0.1%Change laser, increase power to ~12W, i.e. ~5W @ MPRLock interferometer with ~7kW internal power

    TF fluctuates and depends on tuning