Xxxxx

Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions

Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio, David RalphLa Quinta California, San Paulo, Brazil, & London England

Disclosures: Consultant for Abeon, Boston Scientific, Coloplast, NeoTract, Sontec

Page 2: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

Glans Necrosis After Penile Implant Is Rare, but Unfortunately, Incidence Is Increasing Lately!

Only five cases glans ischemia/necrosis found in literatureWe gathered 19 from USA, UK, Saudi ,Germany, Spain and Brazil8 Rods and 11 IPPWe defined risk factors and surgical maneuvers likely to predispose

Page 3: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

5 Risks & 4 Maneuvers Predispose Glans Ischemia

Smoking, Severe ASCVD, Radiation, Previous Implant Surgery, Diabetes

>80% of patients had (3 of 5) 60% of comorbidities

50% of patients had (4 of 5) 80% of comorbidities

Subcoronal Incision, Urethral Injury/Repair, Sliding Procedure, Elastic wrap

90% had subcoronal incision (circumcision)

64% urethral mobilization for sliding technique,

urethral injury or repair

78% had elastic wrap of penis

Page 4: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

Dorsal arteries are compromised by risk factors ie. comorbiditiesPenile rigidity 2º implant must further limit flow of dorsal arteries

Corpus spongiosum compromised by subcoronal incision, urethral mobilization, & elastic wrap with Foley in place

Blood Supply of Glans Penis: Two Sources

Dorsal Arteries & Corpus Spongiosum

Page 5: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

Used for penile lengthening: Average length gain 2.5-3 cm6 patients in this series had sliding procedures for penile lengthAll 6 patients had normal penile Doppler

Egydio PH, Kuehhas FE. BJU 2015; 116: 965

What Is The Sliding Procedure? A Penile Lengthening Surgery Gaining Popularity

Page 6: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

Pre op: POD #1

Glans Necrosis -- Typical PatientDiabetic, Rad RX, IPP Removal, ASCVD, Smoker – 100% Risk Factors

Semi Rigid Rods & Coincident Circumcision

POD 12: POD 14 Removal Rods

The Physician tried to take the patient back

to surgery for removal but the patient talked

him out of it.

Page 7: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

51 y. o. Smoker: ED with “Small Penis”Subcoronal, Sliding Technique, Rods & Elastic Dressing

POD #1 Dusky glans & blisters, note hematoma & elastic dressing

Sliding Procedure Patient with Glans Necrosis

POD # 40 POD #100

3 debriedments w/ new rodsUrethra relocated to shaft

Six Months after 4 operations …Patient using implant but unhappy about position of urethra

This patient was obsessed with penile length. He

desired a longer penis and wanted preservation of

implant at all costs.

Page 8: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

if Many Comorbidities in Patient: BewareDiabetesSmokingPrevious implant removalSevere ASCVDHistory of radiation therapy

Avoid these Maneuvers in these PatientsSubcoronal incision, penile degloving, coincident circumcisionSliding technique for penile lengtheningSimultaneous distal urethral repair with implant preservationOcclusive elastic dressing

>80% of Patients Had >60% of Comorbidities …

≥90% Patients Had Subcoronal Incision …

Future of the sliding technique?

6 of 6 Sliding Procedure Had OK Doppler …

xxxxxxxxxxxxxxxx

Page 9: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

67 y. o, ASCVD, Diabetic, Smoker, Elastic WrapInfrapubic IPP & Coincident Circumcision

Pre Op Doppler OKThe next morning … note hematoma under circumcision

What If We Remove the Implant Immediately?

Page 10: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

Immediate Implant Removal May Reverse Glans Necrosis

Treated with immediate removal and hyperbaric oxygenEtiology? Dorsal arteries compromised and circumcision hematoma

put pressure on urethral blood supply

POD #1 7 Days post removalOne year post emergent removal

Page 11: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

Conclusions19 Cases of Glans Necrosis After Penile Implant

15 cases handled w/ observation & implant preservation … all had glans necrosis11 of 14 (79%) eventually had removal of prosthesis: some up to 6 months later

One year post urgent removal POD #1

4 cases handled with immediate implant removal ... All had no sequelae2 patient subsequently reimplanted penoscrotal without difficulty

Page 12: Glans Necrosis Following Penile Implant: Etiology …Xxxxx Glans Necrosis Following Penile Implant: Etiology and Treatment Suggestions Steven K Wilson, Cesar Mora-Esteves, Paulo Egydio,

What’s The Take Home Message …?When You See an Congested Glans the Next Morning

After Penile Implant.

Preserving Implant Is a Mistake

It Isn’t “Congested” … It’s Ischemic!