23
How to interpret and apply MR implant labelling information correctly in practice ePoster, ECR, 2011: How to interpret and apply MR implant labeling information correctly in practice http://dx.doi.org/10.1594/ecr2011/C-1886

How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

  • Upload
    others

  • View
    10

  • Download
    0

Embed Size (px)

Citation preview

Page 1: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

How to interpret and apply MR implant labelling information

correctly in practiceePoster, ECR, 2011:

How to interpret and apply MR implant labeling information correctly in practicehttp://dx.doi.org/10.1594/ecr2011/C-1886

Page 2: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Which procducts are affected?

Innomotion, Innomedic, Germany

• vascular implants

• interventional instruments

• orthopedic implants

• active implants and medical electrical devices

• surgical instruments

robots, monitoring equipment, injectors

stents, filters, clips, valves

catheters, guidewires, endoscopes,

prostheses, fixation devices

clamps, scalpels and other tools

Page 3: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MR device labelling of a stent, 2010

Reference:from Instructions For Use (IFU) of TAXUS® Express2™ and TAXUS®Express2™ Atom™, Boston Scientific, USA

Page 4: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MR environment

Ref.:IEC 60601-2-33, ASTM F2503, E-DIN 6877-1

MR magnet

MR magnet

MR alternating fieldarea

e.g. technical room

0.5 mT lineControlled area

RF shielded window

RF shielded room

(yellow)

MR operator console

Representative example

Page 5: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Current situation of medical devices in conjunction with MRI

MR system manufacturer„The MR system IFU shall

describe the significant RISK associated with the scanning of patients with active or passive implants containing conductive

materials.“

Since 3rd edition of IEC 60601-2-33 2010: When the implant device is labeled as MR safe or MR conditional,

the OPERATOR is then informed via implant IFU about MR

safety and scan conditions. The IFU shall explain that

information can be found in the ACCOMPANYING

DOCUMENTS of the implant MANUFACTURER.“

contraindicated are:• magnetic materials• electrically conductive materials

Page 6: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MR testing issues – summary• Magnetic induced displacement force (static, dynamic)• Magnetic induced torque (static, dynamic)• Radio frequency (RF) and Gradient induced heating• RF and Gradient induced voltages (stimulation, activation)• Gradient induced vibration• Safe operation of the device within the MR environment (dependent

on individual demands)• Safe operation of the MR system => Image quality issues

MR image artifacts and quality issues• Susceptibility artifacts (dependent on material)• RF artifacts of devices in the imaging volume (distortion, signal

shielding and amplification effects)• RF artifacts in the MR image due to RF emission• Signal-to-Noise ratio (SNR)• B0-Field homogeneity (image uniformity/ spatial distortion)• MR signals from remained H-protons (in plastics)

list is not limited to above mentioned issues

• Magnetic induced displacement force (static, dynamic)• Magnetic induced torque (static, dynamic)• Radio frequency (RF) and Gradient induced heating• RF and Gradient induced voltages (stimulation, activation)• Gradient induced vibration• Safe operation of the device within the MR environment (dependent

on individual demands)• Safe operation of the MR system => Image quality issues

MR image artifacts and quality issues• Susceptibility artifacts (dependent on material)• RF artifacts of devices in the imaging volume (distortion, signal

shielding and amplification effects)• RF artifacts in the MR image due to RF emission• Signal-to-Noise ratio (SNR)• B0-Field homogeneity (image uniformity/ spatial distortion)• MR signals from remained H-protons (in plastics)

list is not limited to above mentioned issues

Page 7: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MR Safe—an item that poses no known hazards in allMR environments.

MR Conditional—an item that has been demonstrated to pose no known hazards in a specified MR environment with specified conditions of use.

MR Unsafe—an item that is known to pose hazards in all MR environments.

Definitions and icons extracted, with permission, from “ASTM F2503-08, Standard Practice for Marking Medical Devices and Other Items for Safety in the MR Environment”, copyright ASTM International, 100 Barr Harbor Drive, West Conshohocken, PA 19428. A copy of the complete standard may be purchased from [email protected], website: www.astm.org.

MR Safe—an item that poses no known hazards in allMR environments.

MR Conditional—an item that has been demonstrated to pose no known hazards in a specified MR environment with specified conditions of use.

MR Unsafe—an item that is known to pose hazards in all MR environments.

Definitions and icons extracted, with permission, from “ASTM F2503-08, Standard Practice for Marking Medical Devices and Other Items for Safety in the MR Environment”, copyright ASTM International, 100 Barr Harbor Drive, West Conshohocken, PA 19428. A copy of the complete standard may be purchased from [email protected], website: www.astm.org.

The active definitions of MR safety:according to ASTM F2503

(and E-DIN 6877-1)

Page 8: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MR Image Artifacts - MR compatibilityaccording to ASTM F2503

(and E-DIN 6877-1)

An MR image artifact is not considered as a direct safety issue by ASTM F2503-08 Standard Practice for Marking Medical Devices and therefore handled as separate issue, but

– artifact information provide important help for the physicianbefore MR scanning of a patient with implant or other item

– a statement about MR image artifacts produced by an implant/ item should be included in the product labeling/ patient implant card

An MR image artifact is not considered as a direct safety issue by ASTM F2503-08 Standard Practice for Marking Medical Devices and therefore handled as separate issue, but

– artifact information provide important help for the physicianbefore MR scanning of a patient with implant or other item

– a statement about MR image artifacts produced by an implant/ item should be included in the product labeling/ patient implant card

Page 9: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Displacement force and torque

RF-induced heating

MR image artifacts

MR system informationexperimental heating data

Reference:from Instructions For Use (IFU) of TAXUS® Express2™ and TAXUS®Express2™ Atom™, Boston Scientific, USA

MR device labelling of a stent, 2010

Page 10: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Which documents are required?

• exact and reliable data about the implant (name, model, no., etc.)• written original information showing the der MR labeling from Instructions

For Use or other equivalent source e. g. from device manufacturer or

from MR implant online database (e. g. MagResource, mri-safety)

• compatibility data sheet of the MR system (=> system handbook)

• exact and reliable data about the MR coils of the MR system used

• exact and reliable data about the implant (name, model, no., etc.)• written original information showing the der MR labeling from Instructions

For Use or other equivalent source e. g. from device manufacturer or

from MR implant online database (e. g. MagResource, mri-safety)

• compatibility data sheet of the MR system (=> system handbook)

• exact and reliable data about the MR coils of the MR system used

Page 11: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

• www.MagResource.comOnline database with original data of the implant manufacturercurrently over 6500 devices listed

• www.mrisafety.comdata listed in different stage, format, categorized

• own lists (! make sure data is valid)

• www.MagResource.comOnline database with original data of the implant manufacturercurrently over 6500 devices listed

• www.mrisafety.comdata listed in different stage, format, categorized

• own lists (! make sure data is valid)

Database for MR labeling information

Page 12: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MagResource – a comprehensive database for MR safety information

Page 13: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Displacement force and torque

Reference:from Instructions For Use (IFU) of TAXUS® Express2™ and TAXUS®Express2™ Atom™, Boston Scientific, USA

MR device labelling of a stent, 2010

Page 14: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

0

200

400

600

800

1000

1200

1400

1600

0 20 40 60 80 100 120 140 160 180 200

B [m

T]

z-Achse [cm]

Hauptmagnetfeld B bei 1,5 Tesla Philips Intera,

The change = gradient of the static magnetic field is the

responsible parameter for the

static induced attraction force (displacement force)dependent on the MR systemmax.

Gradient Bz

Static magnetically induced forcesStatic magnetically induced forces

Page 15: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Compatibility technical specification sheet acc. IEC 60601-2-33

These examples show the MR system-specific parameter of the spatial gradient magnetic field and its gradient product responsible for the magnetically induced displacement force

Magnetom Verio, 3 TSiemens Medical Solutions, ErlangenIBI = 15 T/m,IBI ·IBI = 27 T2/m

Philips Medical Systems, Best, The Netherlands

Page 16: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MR device labelling of a stent, 2010

Reference:from Instructions For Use (IFU) of TAXUS® Express2™ and TAXUS®Express2™ Atom™, Boston Scientific, USA

Page 17: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Factor of approx. 100 to 1000 between RF frequency (MHz) and gradient frequency (kHz)

dB/dt is dependent on gradient magnetic field vector and object position (x,y,z) within the gradient coils

90 mV

Ref. plot: Xixi Zhang, basic studies project, University of Applied Sciences Fachhochschule Gelsenkirchen and MR:comp GmbH, Gelsenkrichen, Germany

Induced voltages in a bipolar pacemaker electrode, Selute Picotip 4034, Guidant, Giessen, Germany

G = 15 mT/m

ts,= 0.1 ms

SL = 150 mT/m/ms

ts = stimulation

Gz G = e.g. 60 mT/m

ts, eff. = e.g. 0.2 ms

Slew rate (SL) = e.g. 300 mT/m/ms

t

Gz+

Gz-

Switched gradient magnetic fields in general contribute negligible to induced heating, but

• if meeting the correct conditions, heating is possible, too• induce voltages (resulting in currents) in conductive structures• can cause unintended stimulation

Switched gradient magnetic fields in general contribute negligible to induced heating, but

• if meeting the correct conditions, heating is possible, too• induce voltages (resulting in currents) in conductive structures• can cause unintended stimulation

Page 18: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

What do I need to consider regarding parameters of the „dB/dt-induced interactions“?

• Most MR labeling are currently not addressing dB/dt-related parameters

Important gradient limit parameters are:• gradient slew rate in T/m/s• gradient amplitude in mT/m• effective stimulation time in ms

derived from fundamental MR testing according to ISO/TS 10974 for active implants. What about passive implants?

=> Standardization committees are currently working on improvements of technical basics and information interfaces

• Most MR labeling are currently not addressing dB/dt-related parameters

Important gradient limit parameters are:• gradient slew rate in T/m/s• gradient amplitude in mT/m• effective stimulation time in ms

derived from fundamental MR testing according to ISO/TS 10974 for active implants. What about passive implants?

=> Standardization committees are currently working on improvements of technical basics and information interfaces

Page 19: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

RF-induced heating

MR system informationexperimental heating data

Reference:from Instructions For Use (IFU) of TAXUS® Express2™ and TAXUS®Express2™ Atom™, Boston Scientific, USA

MR device labelling of a stent, 2010

Page 20: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

SAR distribution (from FDA SAR intercomparison study at 1.5 T, MR:comp)

0dB = 50W/kg

FDA SAR intercomparion protocol; studies project in cooperation with SPEAG, Fachhochschule Gelsenkirchen & MR:comp, T. Malechka et al, 2007

Computer simulation provided the same SAR characteristics for the implant heating at each implant

Computer simulation provided the same SAR characteristics for the implant heating at each implant

Head receive

coil

Page 21: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

What do I need to consider regarding parameters of the „RF-induced interactions“?

• The MR labeling has to contain sufficient information MR system, magnetic field strength, heating of the implant, scan

duration, local and averaged whole body SAR (software displayed, calorimetrical SAR, background SAR), implant location, configuration, orientation, MR coil, pulse sequence, etc. (future: magnetic field measure B1RMS, fixed paramter option))

derived from fundamental MR testing according to ASTM, ISO/TS 10974

However, be aware of the following:The software displayed whole body averaged SAR is in general not appropriate to scale local heating at devices/implants

=> Standardization committees are currently working on improvements of technical basics and information interfaces

• The MR labeling has to contain sufficient information MR system, magnetic field strength, heating of the implant, scan

duration, local and averaged whole body SAR (software displayed, calorimetrical SAR, background SAR), implant location, configuration, orientation, MR coil, pulse sequence, etc. (future: magnetic field measure B1RMS, fixed paramter option))

derived from fundamental MR testing according to ASTM, ISO/TS 10974

However, be aware of the following:The software displayed whole body averaged SAR is in general not appropriate to scale local heating at devices/implants

=> Standardization committees are currently working on improvements of technical basics and information interfaces

Page 22: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

Whole body averaged (WBA) SAR versus local SAR

12

34

5

012345678

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18

1

2

3

4

5

1 2 3 4 5

01

2

3

4

5

6

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18

1

2

3

4

5

SARWBA = 2 W/kg

Two different localSAR distributions

Page 23: How to interpret and apply MR implant labelling ... · IEC 60601-2-33, ASTM F2503, E-DIN 6877-1 MR magnet MR magnet MR alternating field area e.g. technical room Controlled 0.5 mT

MR:comp GmbHMR Safety Testing Laboratory

Buschgrundstrasse 3345894 Gelsenkirchen

Germany

(Local landline) Hotline for MR Safety and MR Application +49 1497730-MR(67)Phone +49 209 149 7730 0_Fax +49 209 149 7730 88Email [email protected]

Testing Services for Magnetic Resonance Safety & Compatibility